Temperature imaging of laser-induced thermotherapy (LITT) by MRI: evaluation of different sequences in phantom

verfasst von: Babak Bazrafshan, Frank Hübner, Parviz Farshid, Renate Hammerstingl, Jijo Paul, Vitali Vogel, Werner Mäntele, Thomas J. Vogl

Erschienen in: Lasers in Medical Science | Ausgabe 1/2014

Einloggen, um Zugang zu erhalten

Abstract

The purpose of this study was to evaluate magnetic resonance (MR) temperature imaging of the laser-induced thermotherapy (LITT) comparing the proton resonance frequency (PRF) and T 1 thermometry methods. LITT was applied to a liver-mimicking acrylamide gel phantom. Temperature rise up to 70 °C was measured using a MR-compatible fiber-optic thermometer. MR imaging was performed by a 1.5-T scanner utilizing fast gradient echo sequences including a segmented echo planar imaging (seg-EPI) sequence for PRF and the following sequences for T 1 method: fast low-angle shot (FLASH), inversion recovery turbo flash (IRTF), saturation recovery turbo flash (SRTF), and true fast imaging (TRUFI). Temperature-induced change of the pixel values in circular regions of interest, selected on images under the temperature probe tip, was recorded. For each sequence, a calibration constant could be determined to be −0.0088 ± 0.0002 ppm °C−1 (EPI), −1.15 ± 0.03 °C−1 (FLASH), −1.49 ± 0.03 °C−1 (IRTF), −1.21 ± 0.03 °C−1 (SRTF), and −2.52 ± 0.12 °C−1 (TRUFI). These constants were evaluated in further LITT experiments in phantom comparing the calculated temperatures with the fiber optic-measured ones; temperature precisions of 0.60 °C (EPI), 0.81 °C (FLASH), 1.85 °C (IRTF), 1.95 °C (SRTF), and 3.36 °C (TRUFI) were obtained. Furthermore, performing the Bland–Altman analysis, temperature accuracy was determined to be 0.23 °C (EPI), 0.31 °C (FLASH), 1.66 °C (IRTF), 1.19 °C (SRTF), and 3.20 °C (TRUFI). In conclusion, the seg-EPI sequence was found to be more convenient for MR temperature imaging of LITT due to its relatively high precision and accuracy. Among the T 1 method sequences, FLASH showed the highest accuracy and robustness.
Fußnoten
1
40 vol% acrylamide/bisacrylamide; mixture ratio 29:1
 
2
59 wt%; microsphere diameter, 0.45–3.51 μm
 
3
0.5 mmol gadopentetate dimeglumine (Gd-DTPA) in 1 ml Magnevist solution
 
4
10 wt% solution; diluted in distilled water
 
Literatur
1.
Zurück zum Zitat Vogl TJ, Mack M, Eichler K, Lehnert T, Nabil M (2009) Effect of laser-induced thermotherapy on liver metastases. Expert Rev Anticancer Ther 6(5):769–774CrossRef Vogl TJ, Mack M, Eichler K, Lehnert T, Nabil M (2009) Effect of laser-induced thermotherapy on liver metastases. Expert Rev Anticancer Ther 6(5):769–774CrossRef
2.
Zurück zum Zitat Ferrari FS, Megliola A, Scorzelli A et al (2007) Treatment of small HCC through radiofrequency ablation and laser ablation. Comparison of techniques and long-term results. Radiol Med 112(3):377–393PubMedCrossRef Ferrari FS, Megliola A, Scorzelli A et al (2007) Treatment of small HCC through radiofrequency ablation and laser ablation. Comparison of techniques and long-term results. Radiol Med 112(3):377–393PubMedCrossRef
3.
Zurück zum Zitat Vogl TJ, Lehnert T, Eichler K, Proschek D, Flöter J, Mack MG (2007) Adrenal metastases: CT-guided and MR-thermometry-controlled laser-induced interstitial thermotherapy. Eur Radiol 17(8):2020–2027PubMedCrossRef Vogl TJ, Lehnert T, Eichler K, Proschek D, Flöter J, Mack MG (2007) Adrenal metastases: CT-guided and MR-thermometry-controlled laser-induced interstitial thermotherapy. Eur Radiol 17(8):2020–2027PubMedCrossRef
4.
Zurück zum Zitat Ritz JP, Lehmann KS, Zurbuchen U et al (2007) Improving laser-induced thermotherapy of liver metastases—effects of arterial microembolization and complete blood flow occlusion. Eur J Surg Oncol 33(5):608–615PubMedCrossRef Ritz JP, Lehmann KS, Zurbuchen U et al (2007) Improving laser-induced thermotherapy of liver metastases—effects of arterial microembolization and complete blood flow occlusion. Eur J Surg Oncol 33(5):608–615PubMedCrossRef
5.
Zurück zum Zitat Mensel B, Weigel C, Heidecke CD, Stier A, Hosten N (2005) Laser-induced thermotherapy (LITT) of tumors of the liver in central location: results and complications. Rofo 177(9):1267–1275PubMedCrossRef Mensel B, Weigel C, Heidecke CD, Stier A, Hosten N (2005) Laser-induced thermotherapy (LITT) of tumors of the liver in central location: results and complications. Rofo 177(9):1267–1275PubMedCrossRef
6.
Zurück zum Zitat Vogl TJ, Straub R, Zangos S, Mack MG, Eichler K (2004) MR-guided laser-induced thermotherapy (LITT) of liver tumours: experimental and clinical data. Int J Hyperthermia 20(7):713–724PubMedCrossRef Vogl TJ, Straub R, Zangos S, Mack MG, Eichler K (2004) MR-guided laser-induced thermotherapy (LITT) of liver tumours: experimental and clinical data. Int J Hyperthermia 20(7):713–724PubMedCrossRef
7.
Zurück zum Zitat Vogl TJ, Mack MG, Roggan A et al (1998) Internally cooled power laser for MR-guided interstitial laser-induced thermotherapy of liver lesions: initial clinical results. Radiology 209(2):381–385PubMed Vogl TJ, Mack MG, Roggan A et al (1998) Internally cooled power laser for MR-guided interstitial laser-induced thermotherapy of liver lesions: initial clinical results. Radiology 209(2):381–385PubMed
8.
Zurück zum Zitat Clasen S, Pereira PL (2008) Magnetic resonance guidance for radiofrequency ablation of liver tumors. J Magn Reson Imaging 27(2):421–433PubMedCrossRef Clasen S, Pereira PL (2008) Magnetic resonance guidance for radiofrequency ablation of liver tumors. J Magn Reson Imaging 27(2):421–433PubMedCrossRef
9.
Zurück zum Zitat Veltri A, Fazio GD, Malfitana V et al (2004) Percutaneous US-guided RF thermal ablation for malignant renal tumors: preliminary results in 13 patients. Eur Radiol 14:2303–2310PubMedCrossRef Veltri A, Fazio GD, Malfitana V et al (2004) Percutaneous US-guided RF thermal ablation for malignant renal tumors: preliminary results in 13 patients. Eur Radiol 14:2303–2310PubMedCrossRef
10.
Zurück zum Zitat Xu HX, Xie XY, Lu MD et al (2004) Ultrasound-guided percutaneous thermal ablation of hepatocellular carcinoma using microwave and radiofrequency ablation. Clin Radiol 59(1):53–61PubMedCrossRef Xu HX, Xie XY, Lu MD et al (2004) Ultrasound-guided percutaneous thermal ablation of hepatocellular carcinoma using microwave and radiofrequency ablation. Clin Radiol 59(1):53–61PubMedCrossRef
11.
Zurück zum Zitat Martin RC, Scoggins CR, McMasters KM (2010) Safety and efficacy of microwave ablation of hepatic tumors: a prospective review of a 5-year experience. Ann Surg Oncol 17(1):171–178PubMedCrossRef Martin RC, Scoggins CR, McMasters KM (2010) Safety and efficacy of microwave ablation of hepatic tumors: a prospective review of a 5-year experience. Ann Surg Oncol 17(1):171–178PubMedCrossRef
12.
Zurück zum Zitat Wolf FJ, Grand DJ, Machan JT, Dipetrillo TA, Mayo-Smith WW, Dupuy DE (2008) Microwave ablation of lung malignancies: effectiveness, CT findings, and safety in 50 patients. Radiology 247(3):871–879PubMedCrossRef Wolf FJ, Grand DJ, Machan JT, Dipetrillo TA, Mayo-Smith WW, Dupuy DE (2008) Microwave ablation of lung malignancies: effectiveness, CT findings, and safety in 50 patients. Radiology 247(3):871–879PubMedCrossRef
13.
Zurück zum Zitat Pauly KB, Diederich CJ, Rieke V et al (2006) Magnetic resonance-guided high-intensity ultrasound ablation of the prostate. Top Magn Reson Imaging 17(3):195–207PubMedCrossRef Pauly KB, Diederich CJ, Rieke V et al (2006) Magnetic resonance-guided high-intensity ultrasound ablation of the prostate. Top Magn Reson Imaging 17(3):195–207PubMedCrossRef
14.
Zurück zum Zitat Köhrmann KU, Michel MS, Gaa J, Marlinghaus E, Alken P (2002) High intensity focused ultrasound as noninvasive therapy for multilocal renal cell carcinoma: case study and review of the literature. J Urol 167(6):2397–2403PubMedCrossRef Köhrmann KU, Michel MS, Gaa J, Marlinghaus E, Alken P (2002) High intensity focused ultrasound as noninvasive therapy for multilocal renal cell carcinoma: case study and review of the literature. J Urol 167(6):2397–2403PubMedCrossRef
16.
Zurück zum Zitat Klingenberg M, Bohris C, Niemz MH, Bille JF, Kurek R, Wallwiener D (2000) Multifibre application in laser-induced interstitial thermometry under on-line MR control. Lasers Med Sci 15:6–14CrossRef Klingenberg M, Bohris C, Niemz MH, Bille JF, Kurek R, Wallwiener D (2000) Multifibre application in laser-induced interstitial thermometry under on-line MR control. Lasers Med Sci 15:6–14CrossRef
17.
Zurück zum Zitat Peters RD, Hinks RS, Henkelman RM (1998) Ex vivo tissue-type independence in proton-resonance frequency shift MR thermometry. Magn Reson Med 40:454–459PubMedCrossRef Peters RD, Hinks RS, Henkelman RM (1998) Ex vivo tissue-type independence in proton-resonance frequency shift MR thermometry. Magn Reson Med 40:454–459PubMedCrossRef
18.
Zurück zum Zitat Peller M, Muacevic A, Reinl H et al (2004) MRT-gestützte Thermometrie in der regionalen Tiefenhyperthermie und interstitiellen Laserthermotherapie. Radiologe 44:310–319PubMedCrossRef Peller M, Muacevic A, Reinl H et al (2004) MRT-gestützte Thermometrie in der regionalen Tiefenhyperthermie und interstitiellen Laserthermotherapie. Radiologe 44:310–319PubMedCrossRef
19.
Zurück zum Zitat Thomson S (1991) Pathologic analysis of photothermal and photomechanical effects of laser–tissue interactions. Photochem Photobiol 53(6):825–835 Thomson S (1991) Pathologic analysis of photothermal and photomechanical effects of laser–tissue interactions. Photochem Photobiol 53(6):825–835
20.
Zurück zum Zitat Vogl TJ, Eichler K, Straub R et al (2001) Laser-induced thermotherapy of malignant liver tumors: general principals, equipment(s), procedure(s)—side effects, complications and results. Eur J Ultrasound 13(2):117–127PubMedCrossRef Vogl TJ, Eichler K, Straub R et al (2001) Laser-induced thermotherapy of malignant liver tumors: general principals, equipment(s), procedure(s)—side effects, complications and results. Eur J Ultrasound 13(2):117–127PubMedCrossRef
21.
Zurück zum Zitat Vogl TJ, Straub R, Eichler K, Woitaschek D, Mack MG (2002) Malignant liver tumors treated with MR imaging-guided laser-induced thermotherapy: experience with complications in 899 patients (2520 lesions). Radiology 225:367–377PubMedCrossRef Vogl TJ, Straub R, Eichler K, Woitaschek D, Mack MG (2002) Malignant liver tumors treated with MR imaging-guided laser-induced thermotherapy: experience with complications in 899 patients (2520 lesions). Radiology 225:367–377PubMedCrossRef
22.
Zurück zum Zitat Arienti V, Pretolani S, Pacella CM et al (2008) Complications of laser ablation for hepatocellular carcinoma: a multicenter study. Radiology 246(3):947–955PubMedCrossRef Arienti V, Pretolani S, Pacella CM et al (2008) Complications of laser ablation for hepatocellular carcinoma: a multicenter study. Radiology 246(3):947–955PubMedCrossRef
23.
Zurück zum Zitat Puccini S, Bär NK, Bublat M, Kahn T, Busse H (2003) Simulations of thermal tissue coagulation and their value for the planning and monitoring of laser-induced interstitial thermotherapy (LITT). Magn Reson Med 49:351–362PubMedCrossRef Puccini S, Bär NK, Bublat M, Kahn T, Busse H (2003) Simulations of thermal tissue coagulation and their value for the planning and monitoring of laser-induced interstitial thermotherapy (LITT). Magn Reson Med 49:351–362PubMedCrossRef
24.
Zurück zum Zitat de Senneville BD, Mougenot C, Quesson B, Dragonu I, Grenier N, Moonen CT (2007) MR thermometry for monitoring tumor ablation. Eur Radiol 17(9):2401–2410PubMedCrossRef de Senneville BD, Mougenot C, Quesson B, Dragonu I, Grenier N, Moonen CT (2007) MR thermometry for monitoring tumor ablation. Eur Radiol 17(9):2401–2410PubMedCrossRef
25.
Zurück zum Zitat Weidensteiner C, Quesson B, Caire-Gana B et al (2003) Real-time MR temperature mapping of rabbit liver in vivo during thermal ablation. Magn Reson Med 50(2):322–330PubMedCrossRef Weidensteiner C, Quesson B, Caire-Gana B et al (2003) Real-time MR temperature mapping of rabbit liver in vivo during thermal ablation. Magn Reson Med 50(2):322–330PubMedCrossRef
26.
Zurück zum Zitat Roujol S, Ries M, Quesson B, Moonen C, de Senneville BD (2010) Real-time MR-thermometry and dosimetry for interventional guidance on abdominal organs. Magn Reson Med 63(4):1080–1087PubMedCrossRef Roujol S, Ries M, Quesson B, Moonen C, de Senneville BD (2010) Real-time MR-thermometry and dosimetry for interventional guidance on abdominal organs. Magn Reson Med 63(4):1080–1087PubMedCrossRef
27.
Zurück zum Zitat Detti V, Grenier D, Perrin E, Beuf O (2011) Assessment of radiofrequency self-heating around a metallic wire with MR T1-based thermometry. Magn Reson Med 66(2):448–455PubMedCrossRef Detti V, Grenier D, Perrin E, Beuf O (2011) Assessment of radiofrequency self-heating around a metallic wire with MR T1-based thermometry. Magn Reson Med 66(2):448–455PubMedCrossRef
28.
Zurück zum Zitat Hey S, de Smet M, Stehning C et al (2012) Simultaneous T1 measurements and proton resonance frequency shift based thermometry using variable flip angles. Magn Reson Med 67(2):457–463PubMedCrossRef Hey S, de Smet M, Stehning C et al (2012) Simultaneous T1 measurements and proton resonance frequency shift based thermometry using variable flip angles. Magn Reson Med 67(2):457–463PubMedCrossRef
29.
Zurück zum Zitat Germain D, Vahala E, Ehnholm GJ et al (2002) MR temperature measurement in liver tissue at 0.23 T with a steady-state free precession sequence. Magn Reson Med 47(5):940–947PubMedCrossRef Germain D, Vahala E, Ehnholm GJ et al (2002) MR temperature measurement in liver tissue at 0.23 T with a steady-state free precession sequence. Magn Reson Med 47(5):940–947PubMedCrossRef
30.
Zurück zum Zitat Lenz C, Klarhöfer M, Scheffler K, Winter L, Sommer G (2011) Assessing extracranial tumors using diffusion-weighted whole-body MRI. Z Med Phys 21(2):79–90PubMedCrossRef Lenz C, Klarhöfer M, Scheffler K, Winter L, Sommer G (2011) Assessing extracranial tumors using diffusion-weighted whole-body MRI. Z Med Phys 21(2):79–90PubMedCrossRef
31.
Zurück zum Zitat Pandeya GD, Klaessens JHGM, Greuter MJW et al (2011) Feasibility of computed tomography based thermometry during interstitial laser heating in bovine liver. Eur Radiol 21:1733–1738PubMedCentralPubMedCrossRef Pandeya GD, Klaessens JHGM, Greuter MJW et al (2011) Feasibility of computed tomography based thermometry during interstitial laser heating in bovine liver. Eur Radiol 21:1733–1738PubMedCentralPubMedCrossRef
32.
Zurück zum Zitat de Senneville BD, Roujol S, Moonen C, Ries M (2010) Motion correction in MR thermometry of abdominal organs: a comparison of the referenceless vs. the multibaseline approach. Magn Reson Med 64(5):1373–1381PubMedCrossRef de Senneville BD, Roujol S, Moonen C, Ries M (2010) Motion correction in MR thermometry of abdominal organs: a comparison of the referenceless vs. the multibaseline approach. Magn Reson Med 64(5):1373–1381PubMedCrossRef
33.
Zurück zum Zitat Delabrousse E, Salomir R, Birer A et al (2010) Automatic temperature control for MR-guided interstitial ultrasound ablation in liver using a percutaneous applicator: ex vivo and in vivo initial studies. Magn Reson Med 63(3):667–679PubMedCrossRef Delabrousse E, Salomir R, Birer A et al (2010) Automatic temperature control for MR-guided interstitial ultrasound ablation in liver using a percutaneous applicator: ex vivo and in vivo initial studies. Magn Reson Med 63(3):667–679PubMedCrossRef
34.
Zurück zum Zitat Seror O, Lepetit-Coiffé M, Le Bail B et al (2008) Real time monitoring of radiofrequency ablation based on MR thermometry and thermal dose in the pig liver in vivo. Eur Radiol 18(2):408–416PubMedCrossRef Seror O, Lepetit-Coiffé M, Le Bail B et al (2008) Real time monitoring of radiofrequency ablation based on MR thermometry and thermal dose in the pig liver in vivo. Eur Radiol 18(2):408–416PubMedCrossRef
35.
Zurück zum Zitat Rieke V, Vigen KK, Sommer G, Daniel BL, Pauly JM, Butts K (2004) Referenceless PRF shift thermometry. Magn Reson Med 51(6):1223–1231PubMedCrossRef Rieke V, Vigen KK, Sommer G, Daniel BL, Pauly JM, Butts K (2004) Referenceless PRF shift thermometry. Magn Reson Med 51(6):1223–1231PubMedCrossRef
36.
Zurück zum Zitat Peters RD, Chan E, Trachtenberg J et al (2000) Magnetic resonance thermometry for predicting thermal damage: an application of interstitial laser coagulation in an in vivo canine prostate model. Magn Reson Med 44(6):873–883PubMedCrossRef Peters RD, Chan E, Trachtenberg J et al (2000) Magnetic resonance thermometry for predicting thermal damage: an application of interstitial laser coagulation in an in vivo canine prostate model. Magn Reson Med 44(6):873–883PubMedCrossRef
37.
Zurück zum Zitat Vogl TJ, Huebner F, Naguib NN et al (2012) MR-based thermometry of laser induced thermotherapy: temperature accuracy and temporal resolution in vitro at 0.2 and 1.5 T magnetic field strengths. Lasers Surg Med 44(3):257–265PubMedCrossRef Vogl TJ, Huebner F, Naguib NN et al (2012) MR-based thermometry of laser induced thermotherapy: temperature accuracy and temporal resolution in vitro at 0.2 and 1.5 T magnetic field strengths. Lasers Surg Med 44(3):257–265PubMedCrossRef
38.
Zurück zum Zitat Bazrafshan B, Hübner F, Farshid P et al (2011) A liver-mimicking MRI phantom for thermal ablation experiments. Med Phys 38(5):2674–2684PubMedCrossRef Bazrafshan B, Hübner F, Farshid P et al (2011) A liver-mimicking MRI phantom for thermal ablation experiments. Med Phys 38(5):2674–2684PubMedCrossRef
39.
Zurück zum Zitat de Bazelaire CM, Duhamel GD, Rofsky NM, Alsop DC (2004) MR imaging relaxation times of abdominal and pelvic tissues measured in vivo at 3.0 T: preliminary results. Radiology 230(3):652–659PubMedCrossRef de Bazelaire CM, Duhamel GD, Rofsky NM, Alsop DC (2004) MR imaging relaxation times of abdominal and pelvic tissues measured in vivo at 3.0 T: preliminary results. Radiology 230(3):652–659PubMedCrossRef
40.
Zurück zum Zitat Ritz JP, Roggan A, Isbert C, Müller G, Buhr HJ, Germer CT (2001) Optical properties of native and coagulated porcine liver tissue between 400 and 2400 nm. Lasers Surg Med 29(3):205–212PubMedCrossRef Ritz JP, Roggan A, Isbert C, Müller G, Buhr HJ, Germer CT (2001) Optical properties of native and coagulated porcine liver tissue between 400 and 2400 nm. Lasers Surg Med 29(3):205–212PubMedCrossRef
41.
Zurück zum Zitat Kickhefel A, Roland J, Weiss C, Schick F (2010) Accuracy of real-time MR temperature mapping in the brain: a comparison of fast sequences. Phys Med 26(4):192–201PubMedCrossRef Kickhefel A, Roland J, Weiss C, Schick F (2010) Accuracy of real-time MR temperature mapping in the brain: a comparison of fast sequences. Phys Med 26(4):192–201PubMedCrossRef
42.
Zurück zum Zitat Olsrud J, Wirestam R, Brockstedt S et al (1998) MRI thermometry in phantoms by use of the proton resonance frequency shift method: application to interstitial laser thermotherapy. Phys Med Biol 43(9):2597–2613PubMedCrossRef Olsrud J, Wirestam R, Brockstedt S et al (1998) MRI thermometry in phantoms by use of the proton resonance frequency shift method: application to interstitial laser thermotherapy. Phys Med Biol 43(9):2597–2613PubMedCrossRef
43.
Zurück zum Zitat McDonald M, Lochhead S, Chopra R, Bronskill MJ (2004) Multi-modality tissue-mimicking phantom for thermal therapy. Phys Med Biol 49(13):2767–2778PubMedCrossRef McDonald M, Lochhead S, Chopra R, Bronskill MJ (2004) Multi-modality tissue-mimicking phantom for thermal therapy. Phys Med Biol 49(13):2767–2778PubMedCrossRef
44.
Zurück zum Zitat d’Arcy JA, Collins DJ, Murphy PS, Rowland IJ, Leach MO (1999) MR thermography using radial sampling. Magn Reson Mater Phy 8(Supplement 1):154 d’Arcy JA, Collins DJ, Murphy PS, Rowland IJ, Leach MO (1999) MR thermography using radial sampling. Magn Reson Mater Phy 8(Supplement 1):154
45.
Zurück zum Zitat Chung YC, Duerk JL, Shankaranarayanan A, Hampke M, Merkle EM, Lewin JS (1999) Temperature measurement using echo-shifted FLASH at low field for interventional MRI. J Magn Reson Imaging 9(1):138–145PubMedCrossRef Chung YC, Duerk JL, Shankaranarayanan A, Hampke M, Merkle EM, Lewin JS (1999) Temperature measurement using echo-shifted FLASH at low field for interventional MRI. J Magn Reson Imaging 9(1):138–145PubMedCrossRef
Metadaten
Titel
Temperature imaging of laser-induced thermotherapy (LITT) by MRI: evaluation of different sequences in phantom
verfasst von
Babak Bazrafshan
Frank Hübner
Parviz Farshid
Renate Hammerstingl
Jijo Paul
Vitali Vogel
Werner Mäntele
Thomas J. Vogl
Publikationsdatum
01.01.2014
Verlag
Springer London
Erschienen in
Lasers in Medical Science / Ausgabe 1/2014
Print ISSN: 0268-8921
Elektronische ISSN: 1435-604X
DOI
https://doi.org/10.1007/s10103-013-1306-5

Weitere Artikel der Ausgabe 1/2014

Lasers in Medical Science 1/2014 Zur Ausgabe