Ultra-low-dose CT with model-based iterative reconstruction (MBIR): detection of ground-glass nodules in an anthropomorphic phantom study

verfasst von: Cristiano Rampinelli, Daniela Origgi, Vittoria Vecchi, Luigi Funicelli, Sara Raimondi, Paul Deak, Massimo Bellomi

Erschienen in: La radiologia medica | Ausgabe 7/2015

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The authors sought to evaluate the effect of model-based iterative reconstruction (MBIR) on the sensitivity of ground-glass nodule (GGN) detection at different dose levels.

Materials and methods

Fifty-four artificial GGN were randomly divided into three sets, each positioned in an anthropomorphic phantom. The three sets were evaluated on standard-dose (SD, 350 mA), low-dose (LD, 35 mA) and ultra-low-dose (ULD, 10 mA) CT scans (100 kV, 64 × 0.625 mm, 0.5 s), and each scan was reconstructed twice with filtered back projection (FBP) and MBIR. Three radiologists independently evaluated the scans for GGN presence and size. SD + FBP was considered the reference standard. A region of interest (ROI) was used to calculate signal-to-noise ratio (SNR) and contrast-to-noise ratio normalised to dose (CNRD). McNemar’s test, Bland–Altman analysis and t test were used for statistical assessment (p < 0.05).

Results

The mean diameter of the 54 GGNs was 9.2 mm (range 3.7–17.3 mm). For the three readers, no statistically significant differences were observed in the sensitivity of GGN detection between LD + MBIR, ULD + MBIR and SD + FBP (p > 0.05). Bland–Altman analysis showed a good reader agreement (±1.5 mm) for GGN size between SD + FBP and ULD + MBIR. For low dose and ultra-low dose, the SNR and CNRD were significantly higher with MBIR (p < 0.0001). The effective dose was 97.1 % lower with ultra-low dose (0.15 mSv) than standard dose (5.15 mSv).

Conclusions

The detection of GGN with MBIR at low-dose and ultra-low-dose CT does not differ significantly from standard-dose CT with FBP in an anthropomorphic phantom.
Literatur
1.
Zurück zum Zitat Brenner DJ, Hall EJ (2007) Computed tomography–an increasing source of radiation exposure. N Engl J Med 357:2277–2384PubMedCrossRef Brenner DJ, Hall EJ (2007) Computed tomography–an increasing source of radiation exposure. N Engl J Med 357:2277–2384PubMedCrossRef
2.
Zurück zum Zitat Sarma A, Heilbrun ME, Conner KE et al (2012) Radiation and chest CT scan examinations: what do we know? Chest 142:750–760PubMedCrossRef Sarma A, Heilbrun ME, Conner KE et al (2012) Radiation and chest CT scan examinations: what do we know? Chest 142:750–760PubMedCrossRef
3.
Zurück zum Zitat Sodickson A, Baeyens PF, Andriole KP et al (2009) Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology 251(1):175–184PubMedCrossRef Sodickson A, Baeyens PF, Andriole KP et al (2009) Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology 251(1):175–184PubMedCrossRef
4.
Zurück zum Zitat Team National Lung Screening Trial Research, Church TR, Black WC, Aberle DR, Berg CD, Clingan KL et al (2013) Results of initial low-dose computed tomographic screening for lung cancer. N Engl J Med 368:1980–1991CrossRef Team National Lung Screening Trial Research, Church TR, Black WC, Aberle DR, Berg CD, Clingan KL et al (2013) Results of initial low-dose computed tomographic screening for lung cancer. N Engl J Med 368:1980–1991CrossRef
5.
Zurück zum Zitat Dohan A, Soyer P (2012) Low-dose abdominal CT for diagnosing appendicitis. N Engl J Med 367(5):477–479PubMedCrossRef Dohan A, Soyer P (2012) Low-dose abdominal CT for diagnosing appendicitis. N Engl J Med 367(5):477–479PubMedCrossRef
6.
Zurück zum Zitat Macari M, Bini EJ, Xue X et al (2002) Colorectal neoplasms: prospective comparison of thin-section low-dose multi-detector row CT colonography and conventional colonoscopy for detection. Radiology 224(2):383–392PubMedCrossRef Macari M, Bini EJ, Xue X et al (2002) Colorectal neoplasms: prospective comparison of thin-section low-dose multi-detector row CT colonography and conventional colonoscopy for detection. Radiology 224(2):383–392PubMedCrossRef
7.
Zurück zum Zitat Albert JM (2013) Radiation risk from CT: implications for cancer screening. Am J Roentgenol 201:W81–W87CrossRef Albert JM (2013) Radiation risk from CT: implications for cancer screening. Am J Roentgenol 201:W81–W87CrossRef
8.
Zurück zum Zitat Team National Lung Screening Trial Research, Aberle DR, Adams AM, Berg CD, Black WC, Clapp JD et al (2011) Reduced lung-cancer mortality with low-dose computed tomographic screening. N Engl J Med 365:395–409CrossRef Team National Lung Screening Trial Research, Aberle DR, Adams AM, Berg CD, Black WC, Clapp JD et al (2011) Reduced lung-cancer mortality with low-dose computed tomographic screening. N Engl J Med 365:395–409CrossRef
9.
Zurück zum Zitat Naidich DP, Bankier AA, MacMahon H et al (2013) Recommendations for the management of subsolid pulmonary nodules detected at CT: a statement from the Fleischner Society. Radiology 266(1):304–317PubMedCrossRef Naidich DP, Bankier AA, MacMahon H et al (2013) Recommendations for the management of subsolid pulmonary nodules detected at CT: a statement from the Fleischner Society. Radiology 266(1):304–317PubMedCrossRef
10.
Zurück zum Zitat Godoy MC, Naidich DP (2012) Overview and strategic management of subsolid pulmonary nodules. J Thorac Imaging 27(4):240–248PubMedCrossRef Godoy MC, Naidich DP (2012) Overview and strategic management of subsolid pulmonary nodules. J Thorac Imaging 27(4):240–248PubMedCrossRef
11.
Zurück zum Zitat Henschke CI, Yankelevitz DF, Mirtcheva R et al (2002) CT screening for lung cancer: frequency and significance of part-solid and nonsolid nodules. Am J Roentgenol 178(5):1053–1057CrossRef Henschke CI, Yankelevitz DF, Mirtcheva R et al (2002) CT screening for lung cancer: frequency and significance of part-solid and nonsolid nodules. Am J Roentgenol 178(5):1053–1057CrossRef
12.
Zurück zum Zitat Christe A, Charimo-Torrente J, Roychoudhury K et al (2013) Accuracy of low-dose computed tomography (CT) for detecting and characterizing the most common CT-patterns of pulmonary disease. Eur J Radiol 82(3):e142–e150PubMedCrossRef Christe A, Charimo-Torrente J, Roychoudhury K et al (2013) Accuracy of low-dose computed tomography (CT) for detecting and characterizing the most common CT-patterns of pulmonary disease. Eur J Radiol 82(3):e142–e150PubMedCrossRef
13.
Zurück zum Zitat Silva AC, Lawder HJ, Hara A et al (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. Am J Roentgenol 194(1):191–199CrossRef Silva AC, Lawder HJ, Hara A et al (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. Am J Roentgenol 194(1):191–199CrossRef
14.
Zurück zum Zitat Lee Y, Jin KN, Lee NK (2012) Low-dose computed tomography of the chest using iterative reconstruction versus filtered back projection: comparison of image quality. J Comput Assist Tomogr 36(5):512–517PubMedCrossRef Lee Y, Jin KN, Lee NK (2012) Low-dose computed tomography of the chest using iterative reconstruction versus filtered back projection: comparison of image quality. J Comput Assist Tomogr 36(5):512–517PubMedCrossRef
15.
Zurück zum Zitat Miéville FA, Gudinchet F, Brunelle F et al (2013) Iterative reconstruction methods in two different MDCT scanners: physical metrics and 4-alternative forced-choice detectability experiments–a phantom approach. Phys Med 29(1):99–110PubMedCrossRef Miéville FA, Gudinchet F, Brunelle F et al (2013) Iterative reconstruction methods in two different MDCT scanners: physical metrics and 4-alternative forced-choice detectability experiments–a phantom approach. Phys Med 29(1):99–110PubMedCrossRef
16.
Zurück zum Zitat Xu Y, He W, Chen H et al (2013) Impact of the adaptive statistical iterative reconstruction technique on image quality in ultra-low-dose CT. Clin Radiol 68(9):902–908PubMedCrossRef Xu Y, He W, Chen H et al (2013) Impact of the adaptive statistical iterative reconstruction technique on image quality in ultra-low-dose CT. Clin Radiol 68(9):902–908PubMedCrossRef
17.
Zurück zum Zitat Katsura M, Matsuda I, Akahane M et al (2013) Model-based iterative reconstruction technique for ultralow-dose chest CT: comparison of pulmonary nodule detectability with the adaptive statistical iterative reconstruction technique. Invest Radiol 48(4):206–212PubMed Katsura M, Matsuda I, Akahane M et al (2013) Model-based iterative reconstruction technique for ultralow-dose chest CT: comparison of pulmonary nodule detectability with the adaptive statistical iterative reconstruction technique. Invest Radiol 48(4):206–212PubMed
18.
Zurück zum Zitat Ichikawa Y, Kitagawa K, Nagasawa N et al (2013) CT of the chest with model-based, fully iterative reconstruction: comparison with adaptive statistical iterative reconstruction. BMC Med Imaging 9:13–27 Ichikawa Y, Kitagawa K, Nagasawa N et al (2013) CT of the chest with model-based, fully iterative reconstruction: comparison with adaptive statistical iterative reconstruction. BMC Med Imaging 9:13–27
19.
Zurück zum Zitat Neroladaki A, Botsikas D, Boudabbous S et al (2013) Computed tomography of the chest with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination: preliminary observations. Eur Radiol 23(2):360–366PubMedCrossRef Neroladaki A, Botsikas D, Boudabbous S et al (2013) Computed tomography of the chest with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination: preliminary observations. Eur Radiol 23(2):360–366PubMedCrossRef
20.
Zurück zum Zitat Huda W, Mettler FA (2011) Volume CT dose index and dose-length product displayed during CT: what good are they? Radiology 258:236–242PubMedCrossRef Huda W, Mettler FA (2011) Volume CT dose index and dose-length product displayed during CT: what good are they? Radiology 258:236–242PubMedCrossRef
21.
Zurück zum Zitat The 2007 Recommendations of the International Commission on Radiological Protection (2007) ICRP publication 103. Ann ICRP 37(2–4):1–332 The 2007 Recommendations of the International Commission on Radiological Protection (2007) ICRP publication 103. Ann ICRP 37(2–4):1–332
22.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMedCrossRef Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMedCrossRef
23.
Zurück zum Zitat Kim H, Park CM, Song YS et al (2014) Influence of radiation dose and iterative reconstruction algorithms for measurement accuracy and reproducibility of pulmonary nodule volume: a phantom study. Eur J Radiol 83(5):848–857PubMedCrossRef Kim H, Park CM, Song YS et al (2014) Influence of radiation dose and iterative reconstruction algorithms for measurement accuracy and reproducibility of pulmonary nodule volume: a phantom study. Eur J Radiol 83(5):848–857PubMedCrossRef
24.
Zurück zum Zitat Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M (2008) Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 248(1):254–263PubMedCrossRef Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M (2008) Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 248(1):254–263PubMedCrossRef
25.
Zurück zum Zitat Yamada Y, Jinzaki M, Tanami Y et al (2012) Model-based iterative reconstruction technique for ultralow-dose computed tomography of the lung: a pilot study. Invest Radiol 47(8):482–489PubMedCrossRef Yamada Y, Jinzaki M, Tanami Y et al (2012) Model-based iterative reconstruction technique for ultralow-dose computed tomography of the lung: a pilot study. Invest Radiol 47(8):482–489PubMedCrossRef
26.
Zurück zum Zitat Revel MP, Bissery A, Bienvenu M et al (2004) Are two-dimensional CT measurements of small noncalcified pulmonary nodules reliable? Radiology 231(2):453–458PubMedCrossRef Revel MP, Bissery A, Bienvenu M et al (2004) Are two-dimensional CT measurements of small noncalcified pulmonary nodules reliable? Radiology 231(2):453–458PubMedCrossRef
27.
Zurück zum Zitat Kim H, Park CM, Kim SH et al (2014) Persistent pulmonary subsolid nodules: model-based iterative reconstruction for nodule classification and measurement variability on low-dose CT. Eur Radiol 24(11):2700–2708PubMedCrossRef Kim H, Park CM, Kim SH et al (2014) Persistent pulmonary subsolid nodules: model-based iterative reconstruction for nodule classification and measurement variability on low-dose CT. Eur Radiol 24(11):2700–2708PubMedCrossRef
28.
Zurück zum Zitat Silva AC, Lawder HJ, Hara A et al (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. Am J Roentgenol 194(1):191–199CrossRef Silva AC, Lawder HJ, Hara A et al (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. Am J Roentgenol 194(1):191–199CrossRef
29.
Zurück zum Zitat Katsura M, Matsuda I, Akahane M et al (2012) Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique. Eur Radiol 22(8):1613–1623PubMedCrossRef Katsura M, Matsuda I, Akahane M et al (2012) Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique. Eur Radiol 22(8):1613–1623PubMedCrossRef
30.
Zurück zum Zitat Doo KW, Kang EY, Yong HS et al (2014) Accuracy of lung nodule volumetry in low-dose CT with iterative reconstruction:an anthropomorphic thoracic phantom study. Br J Radiol 87(1041):20130644PubMedCrossRef Doo KW, Kang EY, Yong HS et al (2014) Accuracy of lung nodule volumetry in low-dose CT with iterative reconstruction:an anthropomorphic thoracic phantom study. Br J Radiol 87(1041):20130644PubMedCrossRef
Metadaten
Titel
Ultra-low-dose CT with model-based iterative reconstruction (MBIR): detection of ground-glass nodules in an anthropomorphic phantom study
verfasst von
Cristiano Rampinelli
Daniela Origgi
Vittoria Vecchi
Luigi Funicelli
Sara Raimondi
Paul Deak
Massimo Bellomi
Publikationsdatum
01.07.2015
Verlag
Springer Milan
Erschienen in
La radiologia medica / Ausgabe 7/2015
Print ISSN: 0033-8362
Elektronische ISSN: 1826-6983
DOI
https://doi.org/10.1007/s11547-015-0505-5

Weitere Artikel der Ausgabe 7/2015

La radiologia medica 7/2015 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.