Intraoperative detection and localization of cylindrical implants in cone-beam CT image data

verfasst von: Joseph Görres, Michael Brehler, Jochen Franke, Karl Barth, Sven Y. Vetter, Andrés Córdova, Paul A. Grützner, Hans-Peter Meinzer, Ivo Wolf, Diana Nabers

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 6/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Orthopedic fractures are often fixed using metal implants. The correct positioning of cylindrical implants such as surgical screws, rods and guide wires is highly important. Intraoperative 3D imaging is often used to ensure proper implant placement. However, 3D image interaction is time-consuming and requires experience. We developed an automatic method that simplifies and accelerates location assessment of cylindrical implants in 3D images.

Methods

Our approach is composed of three major steps. At first, cylindrical characteristics are detected by analyzing image gradients in small image regions. Next, these characteristics are grouped in a cluster analysis. The clusters represent cylindrical implants and are used to initialize a cylinder-to-image registration. Finally, the two end points are optimized regarding image contrast along the cylinder axis.

Results

A total of 67 images containing 420 cylindrical implants were used for testing. Different anatomical regions (calcaneus, spine) and various image sources (two mobile devices, three reconstruction methods) were investigated. Depending on the evaluation set, the detection performance was between 91.7 and 96.1 % true- positive rate with a false-positive rate between 2.0 and 3.2 %. The end point distance errors ranged from \(1.0 \pm 1.2\) to \(4.3 \pm 2.9\) mm and the orientation errors from \(1.6 \pm 2.2\) to \(2.3 \pm 2.2\) degrees. The average computation time was less than 5 seconds.

Conclusions

An automatic method was developed and tested that obviates the need for 3D image interaction during intraoperative assessment of cylindrical orthopedic implants. The required time for working with the viewing software of cone-beam CT device is drastically reduced and leads to a shorter time under anesthesia for the patient.
Literatur
1.
Zurück zum Zitat Barth K, Dennerlein F, Brunner T, Fieselmann A, Graumann R (2013) Infinite impulse response filtering for cone beam tomography. In: Proceedings of SPIE 8668, medical imaging 2013, physics of medical, imaging, 86682U Barth K, Dennerlein F, Brunner T, Fieselmann A, Graumann R (2013) Infinite impulse response filtering for cone beam tomography. In: Proceedings of SPIE 8668, medical imaging 2013, physics of medical, imaging, 86682U
2.
Zurück zum Zitat Chaperon T, Goulette F (2001) Extracting cylinders in full 3D data using a random sampling method and the Gaussian image. In: Vision modeling and visualization, pp 35–42 Chaperon T, Goulette F (2001) Extracting cylinders in full 3D data using a random sampling method and the Gaussian image. In: Vision modeling and visualization, pp 35–42
3.
Zurück zum Zitat Duda RO, Hart PE (1972) Use of the hough transformation to detect lines and curves in pictures. Commun ACM 15(1):11–15CrossRef Duda RO, Hart PE (1972) Use of the hough transformation to detect lines and curves in pictures. Commun ACM 15(1):11–15CrossRef
4.
Zurück zum Zitat Ester M, Kriegel HP, Sander J, Xu X (1996) A density-based algorithm for discovering clusters in large spatial databases with noise. In: Knowledge discovery and data mining, pp 226–231 Ester M, Kriegel HP, Sander J, Xu X (1996) A density-based algorithm for discovering clusters in large spatial databases with noise. In: Knowledge discovery and data mining, pp 226–231
5.
Zurück zum Zitat Fischler MA, Bolles RC (1981) Random sample consensus: a paradigm for model fitting with applications to image analysis and automated cartography. Commun ACM 24(6):381–395CrossRef Fischler MA, Bolles RC (1981) Random sample consensus: a paradigm for model fitting with applications to image analysis and automated cartography. Commun ACM 24(6):381–395CrossRef
6.
Zurück zum Zitat Franke J, von Recum J, Suda AJ, Grützner PA, Wendl K (2012) Intraoperative three-dimensional imaging in the treatment of acute unstable syndesmotic injuries. J Bone Joint Surg 94(15):1386–1390PubMedCrossRef Franke J, von Recum J, Suda AJ, Grützner PA, Wendl K (2012) Intraoperative three-dimensional imaging in the treatment of acute unstable syndesmotic injuries. J Bone Joint Surg 94(15):1386–1390PubMedCrossRef
7.
Zurück zum Zitat Franke J, Recum J, Wendl K, Grützner P (2013) Intraoperative three-dimensional imaging—beneficial or necessary? Der Unfallchirurg 116(2):185–190PubMedCrossRef Franke J, Recum J, Wendl K, Grützner P (2013) Intraoperative three-dimensional imaging—beneficial or necessary? Der Unfallchirurg 116(2):185–190PubMedCrossRef
8.
Zurück zum Zitat Geerling J, Kendoff D, Citak M, Zech S, Gardner MJ, Hüfner T, Krettek C, Richter M (2009) Intraoperative 3d imaging in calcaneal fracture care-clinical implications and decision making. J Trauma Acute Care Surg 66(3):768–773CrossRef Geerling J, Kendoff D, Citak M, Zech S, Gardner MJ, Hüfner T, Krettek C, Richter M (2009) Intraoperative 3d imaging in calcaneal fracture care-clinical implications and decision making. J Trauma Acute Care Surg 66(3):768–773CrossRef
9.
Zurück zum Zitat Kendoff D, Citak M, Gardner MJ, Stübig T, Krettek C, Hüfner T (2009) Intraoperative 3d imaging: value and consequences in 248 cases. J Trauma Acute Care Surg 66(1):232–238CrossRef Kendoff D, Citak M, Gardner MJ, Stübig T, Krettek C, Hüfner T (2009) Intraoperative 3d imaging: value and consequences in 248 cases. J Trauma Acute Care Surg 66(1):232–238CrossRef
10.
Zurück zum Zitat Meyer E, Raupach R, Lell M, Schmidt B, Kachelrieß M (2012) Frequency split metal artifact reduction (fsmar) in computed tomography. Med Phys 39(4):1904–1916 Meyer E, Raupach R, Lell M, Schmidt B, Kachelrieß M (2012) Frequency split metal artifact reduction (fsmar) in computed tomography. Med Phys 39(4):1904–1916
11.
Zurück zum Zitat Nolden M, Zelzer S, Seitel A, Wald D, Müller M, Franz AM, Maleike D, Fangerau M, Baumhauer M, Maier-Hein L, Maier-Hein KH, Meinzer HP, Wolf I (2013) The medical imaging interaction toolkit: challenges and advances. Int J Comput Assist Radiol Surg 8(4):607–620PubMedCrossRef Nolden M, Zelzer S, Seitel A, Wald D, Müller M, Franz AM, Maleike D, Fangerau M, Baumhauer M, Maier-Hein L, Maier-Hein KH, Meinzer HP, Wolf I (2013) The medical imaging interaction toolkit: challenges and advances. Int J Comput Assist Radiol Surg 8(4):607–620PubMedCrossRef
12.
Zurück zum Zitat Otsu N (1975) A threshold selection method from gray-level histograms. Automatica 11:285–296CrossRef Otsu N (1975) A threshold selection method from gray-level histograms. Automatica 11:285–296CrossRef
13.
Zurück zum Zitat Pauwels R, Jacobs R, Bosmans H, Pittayapat P, Kosalagood P, Silkosessak O, Panmekiate S (2013) Automated implant segmentation in cone-beam ct using edge detection and particle counting. Int J Comput Assist Radiol Surg (Epub ahead of print) Pauwels R, Jacobs R, Bosmans H, Pittayapat P, Kosalagood P, Silkosessak O, Panmekiate S (2013) Automated implant segmentation in cone-beam ct using edge detection and particle counting. Int J Comput Assist Radiol Surg (Epub ahead of print)
14.
Zurück zum Zitat Qiu W, Yuchi M, Ding M, Tessier D, Fenster A (2013) Needle segmentation using 3d hough transform in 3d trus guided prostate transperineal therapy. Med Phys 40(4):042902PubMedCrossRef Qiu W, Yuchi M, Ding M, Tessier D, Fenster A (2013) Needle segmentation using 3d hough transform in 3d trus guided prostate transperineal therapy. Med Phys 40(4):042902PubMedCrossRef
15.
Zurück zum Zitat Rabbani T, van den Heuvel F (2005) Efficient hough transform for automatic detection of cylinders in point clouds. In: ISPRS WG III/3, III/4, V/3 Workshop laser, scanning, pp 60–65 Rabbani T, van den Heuvel F (2005) Efficient hough transform for automatic detection of cylinders in point clouds. In: ISPRS WG III/3, III/4, V/3 Workshop laser, scanning, pp 60–65
16.
Zurück zum Zitat Richter M, Zech S (2009) Intraoperative 3-dimensional imaging in foot and ankle trauma-experience with a second-generation device (arcadis-3d). J Orthop Trauma 23(3):213–220PubMedCrossRef Richter M, Zech S (2009) Intraoperative 3-dimensional imaging in foot and ankle trauma-experience with a second-generation device (arcadis-3d). J Orthop Trauma 23(3):213–220PubMedCrossRef
17.
Zurück zum Zitat Roberts KS (1988) A new representation for a line. In: Computer vision and pattern recognition, pp 635–640 Roberts KS (1988) A new representation for a line. In: Computer vision and pattern recognition, pp 635–640
18.
Zurück zum Zitat Rübberdt A, Feil R, Stengel D, Spranger N, Mutze S, Wich M, Ekkernkamp A (2006) The clinical use of the iso-c (3d) imaging system in calcaneus fracture surgery]. Der Unfallchirurg 109(2):112–118PubMedCrossRef Rübberdt A, Feil R, Stengel D, Spranger N, Mutze S, Wich M, Ekkernkamp A (2006) The clinical use of the iso-c (3d) imaging system in calcaneus fracture surgery]. Der Unfallchirurg 109(2):112–118PubMedCrossRef
19.
Zurück zum Zitat Sanders R (2000) Current concepts review—displaced intra-articular fractures of the calcaneus. J Bone Joint Surg 82(2):225–250PubMed Sanders R (2000) Current concepts review—displaced intra-articular fractures of the calcaneus. J Bone Joint Surg 82(2):225–250PubMed
20.
Zurück zum Zitat Schnabel R, Wahl R, Klein R (2007) Efficient ransac for point-cloud shape detection. Comput Graph Forum 26:214–226CrossRef Schnabel R, Wahl R, Klein R (2007) Efficient ransac for point-cloud shape detection. Comput Graph Forum 26:214–226CrossRef
21.
Zurück zum Zitat Styner M, Brechbühler C, Székely G, Gerig G (2000) Parametric estimate of intensity inhomogeneities applied to MRI. IEEE Trans Med Imaging 19:153–165PubMedCrossRef Styner M, Brechbühler C, Székely G, Gerig G (2000) Parametric estimate of intensity inhomogeneities applied to MRI. IEEE Trans Med Imaging 19:153–165PubMedCrossRef
22.
Zurück zum Zitat Vosselman G, Gorte BG, Sithole G, Rabbani T (2004) Recognising structure in laser scanner point clouds. Int Arch Photogramm Remote Sens Sp Inf Sci 46:33–38 Vosselman G, Gorte BG, Sithole G, Rabbani T (2004) Recognising structure in laser scanner point clouds. Int Arch Photogramm Remote Sens Sp Inf Sci 46:33–38
23.
Zurück zum Zitat Zhou H, Qiu W, Ding M, Zhang S (2008) Automatic needle segmentation in 3d ultrasound images using 3d improved hough transform. In: Proceedings of SPIE 6918, medical imaging 2008, image-guided procedures and modeling, 691821 Zhou H, Qiu W, Ding M, Zhang S (2008) Automatic needle segmentation in 3d ultrasound images using 3d improved hough transform. In: Proceedings of SPIE 6918, medical imaging 2008, image-guided procedures and modeling, 691821
Metadaten
Titel
Intraoperative detection and localization of cylindrical implants in cone-beam CT image data
verfasst von
Joseph Görres
Michael Brehler
Jochen Franke
Karl Barth
Sven Y. Vetter
Andrés Córdova
Paul A. Grützner
Hans-Peter Meinzer
Ivo Wolf
Diana Nabers
Publikationsdatum
01.11.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 6/2014
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-014-0998-8

Weitere Artikel der Ausgabe 6/2014

International Journal of Computer Assisted Radiology and Surgery 6/2014 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.