Skip to main content
Erschienen in: International Journal of Computer Assisted Radiology and Surgery 2/2016

01.02.2016 | Original Article

Patient-specific instrument can achieve same accuracy with less resection time than navigation assistance in periacetabular pelvic tumor surgery: a cadaveric study

verfasst von: Kwok-Chuen Wong, Kwan-Yik Sze, Irene Oi-Ling Wong, Chung-Ming Wong, Shekhar-Madhukar Kumta

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 2/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Inaccurate resection in pelvic tumors can result in compromised margins with increase local recurrence. Navigation-assisted and patient-specific instrument (PSI) techniques have recently been reported in assisting pelvic tumor surgery with the tendency of improving surgical accuracy. We examined and compared the accuracy of transferring a virtual pelvic resection plan to actual surgery using navigation-assisted or PSI technique in a cadaver study.

Methods

We performed CT scan in twelve cadaveric bodies including whole pelvic bones. Either supraacetabular or partial acetabular resection was virtually planned in a hemipelvis using engineering software. The virtual resection plan was transferred to a CT-based navigation system or was used for design and fabrication of PSI. Pelvic resections were performed using navigation assistance in six cadavers and PSI in another six. Post-resection images were co-registered with preoperative planning for comparative analysis of resection accuracy in the two techniques.

Results

The mean average deviation error from the planned resection was no different (\(p=0.19\)) for the navigation and the PSI groups: 1.9 versus 1.4 mm, respectively. The mean time required for the bone resection was greater (\(p=0.0006\)) for the navigation group than for the PSI group: 16.2 versus 1.1 min, respectively.

Conclusions

In simulated periacetabular pelvic tumor resections, PSI technique enabled surgeons to reproduce the virtual surgical plan with similar accuracy but with less bone resection time when compared with navigation assistance. Further studies are required to investigate the clinical benefits of PSI technique in pelvic tumor surgery.
Literatur
1.
Zurück zum Zitat Jeys L, Matharu GS, Nandra RS, Grimer RJ (2013) Can computer navigation-assisted surgery reduce the risk of an intralesional margin and reduce the rate of local recurrence in patients with a tumour of the pelvis or sacrum? Bone Joint J 95-B(10):1417–1424 Jeys L, Matharu GS, Nandra RS, Grimer RJ (2013) Can computer navigation-assisted surgery reduce the risk of an intralesional margin and reduce the rate of local recurrence in patients with a tumour of the pelvis or sacrum? Bone Joint J 95-B(10):1417–1424
2.
Zurück zum Zitat Fuchs B, Hoekzema N, Larson DR, Inwards CY, Sim FH (2009) Osteosarcoma of the pelvis: outcome analysis of surgical treatment. Clin Orthop Relat Res 467:510–518PubMedCentralCrossRefPubMed Fuchs B, Hoekzema N, Larson DR, Inwards CY, Sim FH (2009) Osteosarcoma of the pelvis: outcome analysis of surgical treatment. Clin Orthop Relat Res 467:510–518PubMedCentralCrossRefPubMed
3.
Zurück zum Zitat Ozaki T, Flege S, Kevric M, Lindner N, Maas R, Delling G, Schwarz R, von Hochstetter AR, Salzer-Kuntschik M, Berdel WE, Jürgens H, Exner GU, Reichardt P, Mayer-Steinacker R, Ewerbeck V, Kotz R, Winkelmann W, Bielack SS (2003) Osteosarcoma of the pelvis: experience of the cooperative osteosarcoma study group. J Clin Oncol 21:334–341CrossRefPubMed Ozaki T, Flege S, Kevric M, Lindner N, Maas R, Delling G, Schwarz R, von Hochstetter AR, Salzer-Kuntschik M, Berdel WE, Jürgens H, Exner GU, Reichardt P, Mayer-Steinacker R, Ewerbeck V, Kotz R, Winkelmann W, Bielack SS (2003) Osteosarcoma of the pelvis: experience of the cooperative osteosarcoma study group. J Clin Oncol 21:334–341CrossRefPubMed
4.
Zurück zum Zitat Cartiaux O, Docquier PL, Paul L, Francq BG, Cornu OH, Delloye C, Raucent B, Dehez B, Banse X (2008) Surgical inaccuracy of tumor resection and reconstruction within the pelvis: an experimental study. Acta Orthop 79:695–702CrossRefPubMed Cartiaux O, Docquier PL, Paul L, Francq BG, Cornu OH, Delloye C, Raucent B, Dehez B, Banse X (2008) Surgical inaccuracy of tumor resection and reconstruction within the pelvis: an experimental study. Acta Orthop 79:695–702CrossRefPubMed
5.
Zurück zum Zitat Aponte-Tinao LA, Ritacco LE, Ayerza MA, Muscolo DL, Farfalli GL (2013) Multiplanar osteotomies guided by navigation in chondrosarcoma of the knee. Orthopedics 36(3):e325–e330CrossRefPubMed Aponte-Tinao LA, Ritacco LE, Ayerza MA, Muscolo DL, Farfalli GL (2013) Multiplanar osteotomies guided by navigation in chondrosarcoma of the knee. Orthopedics 36(3):e325–e330CrossRefPubMed
6.
Zurück zum Zitat Cho HS, Oh JH, Han I, Kim HS (2012) The outcomes of navigation-assisted bone tumour surgery: minimum three-year follow-up. J Bone Joint Surg (Br) 94–B:1414–1420CrossRef Cho HS, Oh JH, Han I, Kim HS (2012) The outcomes of navigation-assisted bone tumour surgery: minimum three-year follow-up. J Bone Joint Surg (Br) 94–B:1414–1420CrossRef
7.
Zurück zum Zitat Ieguchi M, Hoshi M, Takada J, Hidaka N, Nakamura H (2012) Navigation-assisted surgery for bone and soft tissue tumors with bony extension. Clin Orthop Relat Res 470:275–283PubMedCentralCrossRefPubMed Ieguchi M, Hoshi M, Takada J, Hidaka N, Nakamura H (2012) Navigation-assisted surgery for bone and soft tissue tumors with bony extension. Clin Orthop Relat Res 470:275–283PubMedCentralCrossRefPubMed
8.
Zurück zum Zitat Li J, Wang Z, Guo Z, Chen GJ, Yang M, Pei GX (2012) Irregular osteotomy in limb salvage for juxta-articular osteosarcoma under computer-assisted navigation. J Surg Oncol 106(4):411–416CrossRefPubMed Li J, Wang Z, Guo Z, Chen GJ, Yang M, Pei GX (2012) Irregular osteotomy in limb salvage for juxta-articular osteosarcoma under computer-assisted navigation. J Surg Oncol 106(4):411–416CrossRefPubMed
9.
Zurück zum Zitat So TY, Lam YL, Mak KL (2010) Computer-assisted navigation in bone tumor surgery: seamless workflow model and evolution of technique. Clin Orthop Relat Res 468:2985–2991PubMedCentralCrossRefPubMed So TY, Lam YL, Mak KL (2010) Computer-assisted navigation in bone tumor surgery: seamless workflow model and evolution of technique. Clin Orthop Relat Res 468:2985–2991PubMedCentralCrossRefPubMed
10.
Zurück zum Zitat Wong KC, Kumta SM, Chiu KH, Antonio GE, Unwin P, Leung KS (2007) Precision tumour resection and reconstruction using image-guided computer navigation. J Bone Joint Surg (Br) 89–B:943–947CrossRef Wong KC, Kumta SM, Chiu KH, Antonio GE, Unwin P, Leung KS (2007) Precision tumour resection and reconstruction using image-guided computer navigation. J Bone Joint Surg (Br) 89–B:943–947CrossRef
12.
Zurück zum Zitat Lu S, Xu YQ, Lu WW, Ni GX, Li YB, Shi JH, Li DP, Chen GP, Chen YB, Zhang YZ (2009) A novel patient-specific navigational template for cervical pedicle screw placement. Spine (Phila Pa 1976) 34(26):E959–E966CrossRef Lu S, Xu YQ, Lu WW, Ni GX, Li YB, Shi JH, Li DP, Chen GP, Chen YB, Zhang YZ (2009) A novel patient-specific navigational template for cervical pedicle screw placement. Spine (Phila Pa 1976) 34(26):E959–E966CrossRef
13.
Zurück zum Zitat Mac-Thiong JM, Labelle H, Rooze M, Feipel V, Aubin CE (2003) Evaluation of a transpedicular drill guide for pedicle screw placement in the thoracic spine. Eur Spine J 12:542–547PubMedCentralCrossRefPubMed Mac-Thiong JM, Labelle H, Rooze M, Feipel V, Aubin CE (2003) Evaluation of a transpedicular drill guide for pedicle screw placement in the thoracic spine. Eur Spine J 12:542–547PubMedCentralCrossRefPubMed
14.
Zurück zum Zitat Miyake J, Murase T, Moritomo H, Sugamoto K, Yoshikawa H (2011) Distal radius osteotomy with volar locking plates based on computer simulation. Clin Orthop Relat Res 469(6):1766–1773PubMedCentralCrossRefPubMed Miyake J, Murase T, Moritomo H, Sugamoto K, Yoshikawa H (2011) Distal radius osteotomy with volar locking plates based on computer simulation. Clin Orthop Relat Res 469(6):1766–1773PubMedCentralCrossRefPubMed
15.
Zurück zum Zitat Murase T, Oka K, Moritomo H, Goto A, Yoshikawa H, Sugamoto K (2008) Three-dimensional corrective osteotomy of malunited fractures of the upper extremity with use of a computer simulation system. J Bone Joint Surg Am 90(11):2375–2389CrossRefPubMed Murase T, Oka K, Moritomo H, Goto A, Yoshikawa H, Sugamoto K (2008) Three-dimensional corrective osteotomy of malunited fractures of the upper extremity with use of a computer simulation system. J Bone Joint Surg Am 90(11):2375–2389CrossRefPubMed
16.
Zurück zum Zitat Lionberger DR, Crocker CL, Chen V (2014) Patient Specific Instrumentation. J Arthroplasty 29(9):1699–1704CrossRefPubMed Lionberger DR, Crocker CL, Chen V (2014) Patient Specific Instrumentation. J Arthroplasty 29(9):1699–1704CrossRefPubMed
17.
Zurück zum Zitat Ng VY, DeClaire JH, Berend KR, Gulick BC, Lombardi AV Jr (2012) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470(1):99–107PubMedCentralCrossRefPubMed Ng VY, DeClaire JH, Berend KR, Gulick BC, Lombardi AV Jr (2012) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470(1):99–107PubMedCentralCrossRefPubMed
18.
Zurück zum Zitat Bellanova L, Paul L, Docquier PL (2013) Surgical guides (patient specific instruments) for pediatric tibial bone sarcoma resection and allograft reconstruction. Sarcoma 2013:787653PubMedCentralCrossRefPubMed Bellanova L, Paul L, Docquier PL (2013) Surgical guides (patient specific instruments) for pediatric tibial bone sarcoma resection and allograft reconstruction. Sarcoma 2013:787653PubMedCentralCrossRefPubMed
19.
Zurück zum Zitat Cartiaux O, Paul L, Francq BG, Banse X, Docquier PL (2014) Improved accuracy with 3D planning and patient-specific instruments during simulated pelvic bone tumor surgery. Ann Biomed Eng 42(1):13–205CrossRef Cartiaux O, Paul L, Francq BG, Banse X, Docquier PL (2014) Improved accuracy with 3D planning and patient-specific instruments during simulated pelvic bone tumor surgery. Ann Biomed Eng 42(1):13–205CrossRef
20.
Zurück zum Zitat Wong KC, Kumta SM, Sze KY, Wong CM (2012) Use of a patient specific CAD/CAM surgical jig in extremity bone tumor resection and custom prosthetic reconstruction. Comput Aided Surg 17(6):284–293CrossRefPubMed Wong KC, Kumta SM, Sze KY, Wong CM (2012) Use of a patient specific CAD/CAM surgical jig in extremity bone tumor resection and custom prosthetic reconstruction. Comput Aided Surg 17(6):284–293CrossRefPubMed
21.
Zurück zum Zitat Wong KC, Kumta SM, Leung KS, Ng KW, Ng EW, Lee KS (2010) Integration of CAD/CAM planning into computer assisted orthopaedic surgery. Comput Aided Surg 15:65–74CrossRefPubMed Wong KC, Kumta SM, Leung KS, Ng KW, Ng EW, Lee KS (2010) Integration of CAD/CAM planning into computer assisted orthopaedic surgery. Comput Aided Surg 15:65–74CrossRefPubMed
22.
Zurück zum Zitat Khan FA, Lipman JD, Pearle AD, Boland PJ, Healey JH (2013) Surgical technique: computer-generated custom jigs improve accuracy of wide resection of bone tumors. Clin Orthop Relat Res 471(6):2007–2016PubMedCentralCrossRefPubMed Khan FA, Lipman JD, Pearle AD, Boland PJ, Healey JH (2013) Surgical technique: computer-generated custom jigs improve accuracy of wide resection of bone tumors. Clin Orthop Relat Res 471(6):2007–2016PubMedCentralCrossRefPubMed
23.
Zurück zum Zitat Cartiaux O, Banse X, Paul L, Francq BG, Aubin CÉ, Docquier PL (2013) Computer-assisted planning and navigation improves cutting accuracy during simulated bone tumor surgery of the pelvis. Comput Aided Surg 18(1–2):19–26CrossRefPubMed Cartiaux O, Banse X, Paul L, Francq BG, Aubin CÉ, Docquier PL (2013) Computer-assisted planning and navigation improves cutting accuracy during simulated bone tumor surgery of the pelvis. Comput Aided Surg 18(1–2):19–26CrossRefPubMed
24.
Zurück zum Zitat Ritacco LE, Milano FE, Farfalli GL, Ayerza MA, Muscolo DL, Aponte-Tinao LA (2013) Accuracy of 3-D planning and navigation in bone tumor resection. Orthopedics 36(7):e942–e950CrossRefPubMed Ritacco LE, Milano FE, Farfalli GL, Ayerza MA, Muscolo DL, Aponte-Tinao LA (2013) Accuracy of 3-D planning and navigation in bone tumor resection. Orthopedics 36(7):e942–e950CrossRefPubMed
Metadaten
Titel
Patient-specific instrument can achieve same accuracy with less resection time than navigation assistance in periacetabular pelvic tumor surgery: a cadaveric study
verfasst von
Kwok-Chuen Wong
Kwan-Yik Sze
Irene Oi-Ling Wong
Chung-Ming Wong
Shekhar-Madhukar Kumta
Publikationsdatum
01.02.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 2/2016
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-015-1250-x

Weitere Artikel der Ausgabe 2/2016

International Journal of Computer Assisted Radiology and Surgery 2/2016 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.