21.10.2022 | Original Article

Robotic system for accurate percutaneous puncture guided by 3D–2D ultrasound

verfasst von: Yanping Lin, Shihang Chen, Wangjie Xu, Xiaoxiao Zhu, Qixin Cao

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 2/2023

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Ultrasound (US)-guided robotic systems can reduce the reliance on the experience and skills of surgeons and enable automatic and accurate percutaneous puncture. Two-dimensional (2D) and three-dimensional (3D) US guidance have various advantages and disadvantages. The planned puncture path in the US data directly affects the puncture quality and tissue injury risk. It is difficult to define the optimal path in 2D US images and achieve accurate and safe puncture under the guidance of 3D US volume. This study aims to propose a robotic system guided by 3D–2D US to realize accurate and safe percutaneous puncture.

Methods

We proposed a 3D–2D US-guided percutaneous puncture robotic system by integrating a 3D US scanning robotic system and a 2D US-guided puncture robotic system. The optimal spatial puncture path that targets the lesion and avoids other important tissues was determined in the 3D US volume reconstructed through robotic US scanning. Thereafter, the puncture robot was placed at the puncture site determined according to the planned path. The optimal path was mapped to the 2D US image taken at the puncture site. Finally, the 2D US image and puncture path were used to guide the robot in performing an accurate and safe percutaneous puncture.

Results

The proposed robotic system based on the guidance of 3D–2D US exhibits the advantages of both 3D US and 2D US to improve the accuracy of percutaneous puncture and reduce the risk of tissue injury. The experimental results of phantom puncture demonstrate that the mean puncture accuracy of the system is 1.09 ± 0.35 mm, and the puncture success rate with single needle insertion is 100%.

Conclusion

A percutaneous puncture robotic system based on 3D–2D US guidance was proposed and tested successfully. The experimental results demonstrate the feasibility of the proposed system for achieving accurate and safe robotic percutaneous puncture.
Literatur
1.
Zurück zum Zitat Soni NJ, Franco-Sadud R, Schnobrich D, Dancel R, Tierney DM, Salame G, Restrepo MI, McHardy P (2016) Ultrasound guidance for lumbar puncture. Neurol Clin Pract 6(4):358–368CrossRef Soni NJ, Franco-Sadud R, Schnobrich D, Dancel R, Tierney DM, Salame G, Restrepo MI, McHardy P (2016) Ultrasound guidance for lumbar puncture. Neurol Clin Pract 6(4):358–368CrossRef
2.
Zurück zum Zitat Mahmoud MZ, Aslam M, Alsaadi M, Fagiri MA, Alonazi B (2018) Evolution of Robot-assisted ultrasound-guided breast biopsy systems. J Radiat Res Appl Si 11(1):89–97 Mahmoud MZ, Aslam M, Alsaadi M, Fagiri MA, Alonazi B (2018) Evolution of Robot-assisted ultrasound-guided breast biopsy systems. J Radiat Res Appl Si 11(1):89–97
3.
Zurück zum Zitat Priester AM, Natarajan S, Culjat MO (2013) Robotic ultrasound systems in medicine. IEEE Trans Ultrason Ferroelectr Freq Control 60(3):507–523CrossRef Priester AM, Natarajan S, Culjat MO (2013) Robotic ultrasound systems in medicine. IEEE Trans Ultrason Ferroelectr Freq Control 60(3):507–523CrossRef
4.
Zurück zum Zitat Hemmerling TM, Taddei R, Wehbe M, Cyr S, Zaouter C, Morse J (2013) First robotic ultrasound-guided nerve blocks in humans using the magellan system. Anesth Analg 116(2):491–494CrossRef Hemmerling TM, Taddei R, Wehbe M, Cyr S, Zaouter C, Morse J (2013) First robotic ultrasound-guided nerve blocks in humans using the magellan system. Anesth Analg 116(2):491–494CrossRef
5.
Zurück zum Zitat Hong J, Dohi T, Hashizume M, Konishi K, Hata N (2004) An ultrasound-driven needle-insertion robot for percutaneous cholecystostomy. Phys Med Biol 49(3):441–455CrossRef Hong J, Dohi T, Hashizume M, Konishi K, Hata N (2004) An ultrasound-driven needle-insertion robot for percutaneous cholecystostomy. Phys Med Biol 49(3):441–455CrossRef
6.
Zurück zum Zitat Kobayashi Y, Hong J, Hamano R, Okada K, Fujie MG, Hashizume M (2012) Development of a needle insertion manipulator for central venous catheterization. Int J Med Robot Comp 8(1):34–44CrossRef Kobayashi Y, Hong J, Hamano R, Okada K, Fujie MG, Hashizume M (2012) Development of a needle insertion manipulator for central venous catheterization. Int J Med Robot Comp 8(1):34–44CrossRef
7.
Zurück zum Zitat Morse J, Terrasini N, Wehbe M, Philippona C, Zaouter C, Cyr S, Hemmerling TM (2014) Comparison of success rates, learning curves, and inter-subject performance variability of robot-assisted and manual ultrasound-guided nerve block needle guidance in simulation. Br J Anaesth 112(6):1092–1097CrossRef Morse J, Terrasini N, Wehbe M, Philippona C, Zaouter C, Cyr S, Hemmerling TM (2014) Comparison of success rates, learning curves, and inter-subject performance variability of robot-assisted and manual ultrasound-guided nerve block needle guidance in simulation. Br J Anaesth 112(6):1092–1097CrossRef
8.
Zurück zum Zitat Takacs B, Haidegger T (2020) Autonomous applied robotics: ultrasound-based robot-assisted needle insertion system concept and development. In: IEEE international conference of system of systems engineering (SoSE), pp 487–492 Takacs B, Haidegger T (2020) Autonomous applied robotics: ultrasound-based robot-assisted needle insertion system concept and development. In: IEEE international conference of system of systems engineering (SoSE), pp 487–492
9.
Zurück zum Zitat Neubach Z, Shoham M (2010) Ultrasound-guided robot for flexible needle steering. IEEE T Bio-Med Eng 57(4):799–805CrossRef Neubach Z, Shoham M (2010) Ultrasound-guided robot for flexible needle steering. IEEE T Bio-Med Eng 57(4):799–805CrossRef
10.
Zurück zum Zitat Orhan SO, Yildirim MC, Bebek O (2016) Design and modeling of a parallel robot for ultrasound guided percutaneous needle interventions. In: 41st annual conference of the IEEE industrial electronics society (IECON), pp 5002–5007 Orhan SO, Yildirim MC, Bebek O (2016) Design and modeling of a parallel robot for ultrasound guided percutaneous needle interventions. In: 41st annual conference of the IEEE industrial electronics society (IECON), pp 5002–5007
11.
Zurück zum Zitat Neshat H, Cool DW, Barker K, Gardi L, Kakani N, Fenster A (2013) A 3D ultrasound scanning system for image guided liver interventions. Med Phys 10(1118/1):4824326 Neshat H, Cool DW, Barker K, Gardi L, Kakani N, Fenster A (2013) A 3D ultrasound scanning system for image guided liver interventions. Med Phys 10(1118/1):4824326
13.
Zurück zum Zitat Agel FM, Lorenzo MJV (2005) Tridimensional (3D) ultrasonography. Rev Esp Enferm Dig 97(2):125–130 Agel FM, Lorenzo MJV (2005) Tridimensional (3D) ultrasonography. Rev Esp Enferm Dig 97(2):125–130
14.
Zurück zum Zitat Boctor EM, Choti MA, Burdette EC, Webster RJ (2008) Three-dimensional ultrasound-guided robotic needle placement: an experimental evacuation. Int J Med Robot Comp 4(2):180–191CrossRef Boctor EM, Choti MA, Burdette EC, Webster RJ (2008) Three-dimensional ultrasound-guided robotic needle placement: an experimental evacuation. Int J Med Robot Comp 4(2):180–191CrossRef
15.
Zurück zum Zitat An CY, Hsu YL, Tseng CS (2019) An ultrasound-guided robotic HIFU ablation system with respiration induced displacement and time delay compensation. J Med Biol Eng 39(5):796–805CrossRef An CY, Hsu YL, Tseng CS (2019) An ultrasound-guided robotic HIFU ablation system with respiration induced displacement and time delay compensation. J Med Biol Eng 39(5):796–805CrossRef
16.
Zurück zum Zitat Boctor EM, Webster RJ, Choti MA, Taylor RH , Fichtinger G (2014) Robotically assisted ablative treatment guided by freehand 3D ultrasound. In: International congress and exhibition on computer assisted radiology and surgery, pp 503–508 Boctor EM, Webster RJ, Choti MA, Taylor RH , Fichtinger G (2014) Robotically assisted ablative treatment guided by freehand 3D ultrasound. In: International congress and exhibition on computer assisted radiology and surgery, pp 503–508
17.
18.
Zurück zum Zitat Esteban J, Simson W, Witzig SR, Rienmuller A, Virga S, Frisch B, Zettinig O, Sakara D, Ryang YM, Navab N, Hennersperger C (2018) Robotic ultrasound-guided facet joint insertion. Int J Comput Assist Radiol Surg 13(6):895–904CrossRef Esteban J, Simson W, Witzig SR, Rienmuller A, Virga S, Frisch B, Zettinig O, Sakara D, Ryang YM, Navab N, Hennersperger C (2018) Robotic ultrasound-guided facet joint insertion. Int J Comput Assist Radiol Surg 13(6):895–904CrossRef
19.
Zurück zum Zitat Hungr N, Baumann M, Long JA, Troccaz J (2012) A 3-D ultrasound robotic prostate brachytherapy system with prostate motion tracking. Ieee T Robot 28(6):1382–1397CrossRef Hungr N, Baumann M, Long JA, Troccaz J (2012) A 3-D ultrasound robotic prostate brachytherapy system with prostate motion tracking. Ieee T Robot 28(6):1382–1397CrossRef
20.
Zurück zum Zitat Kim C, Chang D, Petrisor D, Chirikjian G, Han M, Stoianovici D (2013) Ultrasound probe and needle-guide calibration for robotic ultrasound scanning and needle targeting. IEEE Trans Bio-Med Eng 60(6):1728–1734CrossRef Kim C, Chang D, Petrisor D, Chirikjian G, Han M, Stoianovici D (2013) Ultrasound probe and needle-guide calibration for robotic ultrasound scanning and needle targeting. IEEE Trans Bio-Med Eng 60(6):1728–1734CrossRef
21.
Zurück zum Zitat Chen ZP, Huang QH (2018) Real-time freehand 3D ultrasound imaging. Comput Methods Biomech Biomed Eng Imaging Visual 6(1):74–83CrossRef Chen ZP, Huang QH (2018) Real-time freehand 3D ultrasound imaging. Comput Methods Biomech Biomed Eng Imaging Visual 6(1):74–83CrossRef
22.
Zurück zum Zitat Huang QH, Huang YP, Hu W, Li XL (2015) Bezier interpolation for 3-D freehand ultrasound. IEEE Trans Hum Mach Syst 45(3):385–392CrossRef Huang QH, Huang YP, Hu W, Li XL (2015) Bezier interpolation for 3-D freehand ultrasound. IEEE Trans Hum Mach Syst 45(3):385–392CrossRef
23.
Zurück zum Zitat Saftoiu A, Gheonea DI (2009) Tridimensional (3D) endoscopic ultrasound - a pictorial review. J Gastrointest Liver 18(4):501–505 Saftoiu A, Gheonea DI (2009) Tridimensional (3D) endoscopic ultrasound - a pictorial review. J Gastrointest Liver 18(4):501–505
25.
Zurück zum Zitat Chen SH, Li ZJ, Lin YP, Wang F, Cao QX (2021) Automatic ultrasound scanning robotic system with optical waveguide-based force measurement. Int J Comput Assist Radiol Surg 16(6):1015–1025CrossRef Chen SH, Li ZJ, Lin YP, Wang F, Cao QX (2021) Automatic ultrasound scanning robotic system with optical waveguide-based force measurement. Int J Comput Assist Radiol Surg 16(6):1015–1025CrossRef
26.
Zurück zum Zitat Chen SH, Wang F, Lin YP, Shi QS, Wang YL (2021) Ultrasound-guided needle insertion robotic system for percutaneous puncture. Int J Comput Assist Radiol Surg 16(3):475–484CrossRef Chen SH, Wang F, Lin YP, Shi QS, Wang YL (2021) Ultrasound-guided needle insertion robotic system for percutaneous puncture. Int J Comput Assist Radiol Surg 16(3):475–484CrossRef
27.
Zurück zum Zitat Chen SH, Lin YP, Li ZJ, Wang F, Cao QX (2022) Automatic and accurate needle detection in 2D ultrasound during robotic assisted needle insertion process. Int J Comput Assist Radiol Surg 17(2):295–303CrossRef Chen SH, Lin YP, Li ZJ, Wang F, Cao QX (2022) Automatic and accurate needle detection in 2D ultrasound during robotic assisted needle insertion process. Int J Comput Assist Radiol Surg 17(2):295–303CrossRef
Metadaten
Titel
Robotic system for accurate percutaneous puncture guided by 3D–2D ultrasound
verfasst von
Yanping Lin
Shihang Chen
Wangjie Xu
Xiaoxiao Zhu
Qixin Cao
Publikationsdatum
21.10.2022
Verlag
Springer International Publishing
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 2/2023
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-022-02766-1

Weitere Artikel der Ausgabe 2/2023

International Journal of Computer Assisted Radiology and Surgery 2/2023 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.