Purpose
To report on our initial experiences with a new method of real-time virtual sonography (RVS)-guided 11-gauge vacuum-assisted breast biopsy for lesions that were initially detected with breast MRI.
Materials and methods
RVS-guided 11-gauge vacuum-assisted biopsy is performed when a lesion with suspicious characteristics is initially detected with breast MRI and is occult on mammography, sonography, and physical examination. Live sonographic images were co-registered to the previously loaded second-look spine contrast-enhanced breast MRI volume data to correlate the sonography and MR images.
Results
Six lesions were examined in six consecutive patients scheduled to undergo RVS-guided 11-gauge vacuum-assisted biopsy. One patient was removed from the study because of non-visualization of the lesion in the second-look spine contrast-enhanced breast MRI. Five patients with non-mass enhancement lesions were biopsied. The lesions ranged in size from 9 to 13 mm (mean 11 mm). The average procedural time, including the sonography and MR image co-registration time, was 25 min. All biopsies resulted in tissue retrieval. One was fibroadenomatous nodules, and those of four were fibrocystic changes. There were no complications during or after the procedures.
Conclusion
RVS-guided 11-gauge vacuum-assisted breast biopsies provide a safe and effective method for the examination of suspicious lesions initially detected with MRI.