Comparisons between artificial intelligence computer-aided detection synthesized mammograms and digital mammograms when used alone and in combination with tomosynthesis images in a virtual screening setting

verfasst von: Takayoshi Uematsu, Kazuaki Nakashima, Taiyo Leopoldo Harada, Hatsuko Nasu, Tatsuya Igarashi

Erschienen in: Japanese Journal of Radiology | Ausgabe 1/2023

Abstract

Purpose

To compare the reader performance of artificial intelligence computer-aided detection synthesized mammograms (AI CAD SM) with that of digital mammograms (DM) when used alone or in combination with digital breast tomosynthesis (DBT) images.

Materials and methods

This retrospective multireader (n = 4) study compared the reader performances in 388 cases (84 cancer, 83 benign, and 221 normal or benign cases). The overall accuracy of the breast-based assessment was determined by four radiologists using two sequential reading modes: DM followed by DM + DBT; and AI CAD SM followed by AI CAD SM + DBT. Each breast was rated by each reader using five-category ratings, where 3 or higher was considered positive. The area under the receiver-operating characteristic curve (AUC) and reading time were evaluated.

Results

The mean AUC values for DM, AI CAD SM, DM + DBT, and AI CAD SM + DBT were 0.863, 0.895, 0.886, and 0.902, respectively. The mean AUC of AI CAD SM was significantly higher (P < 0.0001) than that of DM. The mean AUC of AI CAD SM + DBT was higher than that of DM + DBT (P = 0.094). A significant reduction in the reading time was observed after using AI CAD SM + DBT when compared with that after using DM + DBT (P < 0.001).

Conclusion

AI CAD SM + DBT might prove more effective than DM + DBT in a screening setting because of its lower radiation dose, noninferiority, and shorter reading time compared to DM + DBT.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
DM
Digital mammograms
AI
Artificial intelligence
CAD
Computer-aided detection
SM
Synthesized mammograms
DBT
Digital breast tomosynthesis
POM
Probability of malignancy
AUC
Areas under the receiver-operating characteristic curve
DCIS
Ductal carcinoma in situ
CC
Cranio-caudal
MLO
Mediolateral oblique
BI-RADS
Breast Imaging Reporting and Data System
ROC
Receiver-operating characteristic
CI
Confidence intervals
2D
Two dimensional

Introduction

Digital breast tomosynthesis (DBT) images have been successfully used for population-based breast cancer screening. There is evidence of improved overall performance with digital mammograms (DM) when used along with DBT compared to that when used alone [1]. However, the use of DM + DBT implies a twofold increase in the radiation dose, which is not acceptable for the screening of asymptomatic women. Synthesized mammograms (SM) consist of two-dimensional images that are generated from the DBT data to prevent exposure to additional radiation doses; SM + DBT have been used to replace DM + DBT for the screening process [13]. However, DBT is commonly used in combination with DM or SM. This increases in the reading time because the radiologist has to read many slice images generated by the DBT data [4], which could act as a barrier for the implementation of DBT in screening.
In the recent years, deep learning-based artificial intelligence (AI) systems have been quickly evolving in the field of breast imaging and surpassing the performance and clinical value of traditional computer-aided detection (CAD) for mammography [5]. In addition, the use of an AI CAD-enhanced SM enabled radiologists to review the DBT 23.5% at a faster rate, without altering the interpretation performance, when compared to that without CAD [6]. Previous studies have reported that the use of AI CAD SM can automatically enhance suspicious findings in the DBT data, thereby improving the quality of the SM [79]. We hypothesized that newer versions of SM using AI CAD-like technology could improve the quality of images and enhance the performance of the radiologists along with a reduction in the reading time. The aim of this study was to assess the reader performance of new SM and compare it with that of the original full-field DM; the comparisons were made using the mammograms alone or in combination with DBT in a virtual screening setting.

Materials and methods

Study design

This retrospective, fully crossed, completely randomized, multireader, multicase study was approved by the Institutional Review Board of our hospital. Written informed consent to was obtained from all the recruited patients enrolled in this study. FUJIFILM (Tokyo, Japan), which invented the new SM for the reading, provided the equipment and support for this study. The data and materials submitted for publication were always under the control of the researchers who were not FUJIFILM personnel. According to the joint research and development agreement, 500 DM + DBT examinations with a bilateral two view (cranio-caudal/mediolateral oblique CC/MLO) image were performed at ? Hospital between January 2020 and August 2020. All DM + DBT examinations were performed with a narrow-angle DBT system with 15-degree tube motion from 15 projection images (AMULET Innovality; FUJIFILM)). DM and DBT were performed during a single breast compression per view. We decided that the comparison of the reader performance was performed at the per-breast level using only mediolateral oblique (MLO) view images taking the reading time and readers’ attention span during the reading study into consideration.

Case selection

We excluded patients with past surgical history from a case set and developed the case set of 388 breast images from 198 DM/AI CAD SM + DBT examinations in sequence based on the diagnostic work-up patients, which comprises 84 breast cancers, 83 biopsy-proven benign lesions, and 221 normal or benign cases (BI-RADS score of 1 or 2) [10] with negative results after 1 year follow-up. We included dataset with anonymized images and patient’s data with pathology results if an image-guided biopsy and/or surgery was performed. The distribution and mammographic findings were 83 masses (49 malignant and 34 benign), 16 masses with associated calcifications (15 malignant and 1 benign), 5 focal asymmetry densities (2 malignant and 3 benign), 10 architectural distortions (4 malignant and 6 benign), and 53 calcifications (14 malignant and 39 benign). The lesion size of the malignant cases with mass type, ranged from 6 to 42 mm (median, 19 mm; mean, 20 mm). The verified cancer cases included 71 cases of invasive ductal carcinoma, 8 cases of DCIS, 3 cases of invasive lobular carcinoma, 1 case of tubular carcinoma, and 1 case of invasive micropapillary carcinoma. The breast density distribution ratings of the cases were 206 of 388 (53%) non-dense breasts (including almost entirely fatty or scattered fibroglandular densities) and 182 of 388 (47%) dense breasts (including heterogeneously dense or extremely densities), indicating the tissue density BI-RADS score [10]. The age of the study cohort ranged from 23 to 88 years (median, 52 years; mean, 54 years).

Generation of new SM

We utilized the latest version (under development) of FUJIFILUM image processing approach to generate the AI CAD SM used in this study. The AI algorithm uses a fast and efficient convolutional neural network, ESPNet [11]. They have improved different finding patterns, such as soft tissue densities, linear structures, and calcifications, detected and extracted from DBT data using AI CAD-like technology. They are new pixel-based synthesis algorithm that merges the selected DBT plane based on the AI CAD detections from the DBT data. The lesions detected by AI CAD are not marked or outlined on AI CAD SM.

Image readers

Four Japanese radiologists who were ranked as having an “A” certification from the Japan Central Organization on Quality Assurance of Breast Cancer Screening (i.e., achieved the scores required for a DM reading instructor in Japan) performed the image evaluations. The radiologists were blinded to the case selection. The readers had between 5 and 30 years of experience in mammography (median, 12 years) and their experience with DBT ranged from 2 to 8 years (median, 4 years). None of the radiologists had prior experience with the new SM reading.

Evaluation methods

The evaluations of the images by the readers were conducted in two stages. The examinations were read twice as follows: (a) DM followed by DM + DBT, and (b) AI CAD SM followed by AI CAD SM + DBT. At least 4 week intervals were allowed as wash-out periods. In each reading mode, the radiologists were prompted to rate each breast using the two-dimensional (2D) image alone followed by the rating of the DBT using the five-point forced BI-RADS (1, 2, 3, 4, or 5) scores [10]. The most suspicious finding was marked and rated on the electronic reporting system.
The reading times were automatically recorded per examination using the workstation software from the moment the examination was displayed until the reader document the image findings including the locations and radiological characterizations of the cancers and benign lesions were documented. The readers did not use a timer and were blinded to the reading time.
To acquaint themselves with the workstation and interpretation procedure, all radiologists were trained to review sets of 100 DM + DBT and sets of 200 breast images from 100 DM/AI CAD SM + DBT examinations (not included in the study) before participating in this study.
The radiologists used a reading dual-monitor workstation with an electronic reporting system (Climb-Mammography WS; Climb Medical Systems) and a 5MP DB/DBT-certified diagnostic color display (CCL-S500; JVCKENWOOD) calibrated to the DICOM Grayscale Standard Display Function, for the image interpretation.

Reference standard

An unblinded review of every case per breast was performed by one radiologist (?, a breast imaging radiologist with 30 years of experience in mammogram interpretation and 10 years of experience in tomosynthesis image interpretation), who was not involved in the observer study, to determine the standard pathologic reference and reference mammographic findings. The radiologist also confirmed that there were no discrepancies between the pathologic diagnoses and mammographic findings of 84 breast cancers and 83 biopsy-proven benign lesions. The number of true and false interpretations by the readers was assessed for each image, including the location and radiological characterization of the cancers and benign lesions and the confirmed normal status. This study was modeled in a screening setting: hence, the final assessments of the images were categorized as follows: positive (BI-RADS score, 3–5) and negative (BI-RADS score, 1–2).

Statistical analysis

The mean area under the receiver-operating characteristic (ROC) curve (AUC) was used to assess the accuracy of the reported forced BI-RADS ratings. The reader-averaged AUC was analyzed using the random-reader, random-case model to account for the possible correlations during the evaluation of the breasts in the same patient. Additionally, the correlations between repeated assessments of the same cases by different radiologists using different modalities, and between-reader variabilities were assessed. Bootstrap percentile confidence intervals (CI) and the corresponding estimated P values were computed based on the 100,000 bootstrap samples (1000 resamples of cases by 100 resamples of readers). A linear mixed model with the subject and reader as variable factors was used for inter-method comparisons of the reading time. Statistical inferences for all analyses were performed at a significance level of 0.05. The statistical analyses were performed using the R package for Windows (version 4.0.2, R Foundation for Statistical Computing, Vienna, Austria).

Results

The mean (± standard deviation) compressed breast thickness during DM + DBT was 38.8 mm ± 14.5. The mean glandular dose for a single mammographic view was 1.12 mGy ± 0.32 (standard deviation) for DM and 1.62 mGy ± 0.49 for DBT. These dose levels constitute an average dose reduction of 41% (1.12 mGy/2.74 mGy) for AI CAD SM + DBT as compared with DM + DBT.
The mean AUC values for DM, AI CAD SM, DM + DBT, and AI CAD SM + DBT were 0.863, 0.895, 0.886, and 0.902, respectively. The AUC of AI CAD SM alone was significantly higher than that of DM alone (difference, 0.033; 95% CI 0.013, 0.054; P < 0.0001; Fig. 1). All four readers performed somewhat better with AI CAD SM than with DM (Table 1). The AUC of AI CAD SM + DBT was higher than that of DM + DBT (difference, 0.016; 95% CI − 0.003, 0.035; P = 0.094), and three readers performed somewhat better with AI CAD SM + DBT than with DM + DBT. In addition, the mean AUC of AI CAD SM alone and DM + DBT did not differ significantly (P = 0.312); and three readers performed somewhat better with AI CAD SM alone than with DM + DBT. Except for a single reader, who performed slightly worse (AUC difference, 0.006) with DM + DBT when compared with DM alone, improvements in the performances of the remaining three readers were observed when DBT was made available to them after a review of either the DM or AI CAD SM alone.
Table 1
AUC based on the forced bi-rads ratings from each reader for the breast-level analysis
Reader
AUC-based forced BI-RADS ratings
DM
SM
DM + DBT
SM + DBT
1
0.858
0.906
0.889
0.913
2
0.885
0.893
0.914
0.898
3
0.838
0.908
0.886
0.920
4
0.869
0.878
0.863
0.885
Mean
0.863
0.895
0.886
0.902
DM digital mammograms, SM synthesized mammograms, DBT digital breast tomosynthesis images
The cancer detection sensitivities for DM, AI CAD SM, DM + DBT, and AI CAD SM + DBT were 84.2, 87.5, 87.2, and 88.4%, respectively (Table 2). The detection sensitivity was improved after using AI CAD SM alone compared to DM alone (difference, 0.033; 95% CI − 0.003, 0.068; P = 0.056; Fig. 2). No significant difference in sensitivity was observed between AI CAD SM alone and DM + DBT (P = 0.173), whereas a significant difference was noted between DM alone and DM + DBT (P = 0.034). In the dense breast cases (n = 32), the sensitivities of the 2D images alone were improved in the corresponding sequential reading modes when DBT were made available to them (after DM: difference, 0.055; 95% CI 0.000, 0.109; P = 0.034; after AI CAD SM: difference, 0.031; 95% CI 0.008, 0.063; P < 0.0001).
Table 2
Sensitivities of each reading mode for the malignant lesions and each finding
 
DM (%)
SM (%)
DM + DBT (%)
SM + DBT (%)
Cancer detection (n = 84)
84.2
87.5
87.2
88.4
Non-dense breast (n = 52)
88.0
90.9
89.4
90.4
Dense breast (n = 32)
78.1
82.0
83.6
85.2
Calcification lesion (n = 14)
96.4
92.9
96.4
92.9
DM digital mammograms, SM synthesized mammograms, DBT digital breast tomosynthesis
For the malignant calcification lesions (n = 14), the sensitivities for DM, AI CAD SM, DM + DBT, and AI CAD SM + DBT were 96.4, 92.9, 96.4, and 92.9%, respectively (Table 2). No significant differences were observed between the sensitivities of AI CAD SM and DM alone (P = 0.286) and DM + DBT and AI CAD SM + DBT (P = 0.286).
Similarly, for the verified benign cases (n = 83), the sensitivities for DM, AI CAD SM, DM + DBT, and AI CAD SM + DBT were 53.6, 47.3, 55.7, and 48.8%, respectively (Table 3). Significant differences were observed between DM and AI CAD SM alone (P = 0.012), DM + DBT and AI CAD SM + DBT (P = 0.002), and AI CAD SM alone and DM + DBT (P = 0.001). These findings indicated that the benign recall rates of AI CAD SM were lower than those of DM. The sensitivity of AI CAD SM alone was significantly lower than that of DM + DBT in the dense breast cases.
Table 3
Sensitivities of each reading mode for benign lesions and each finding
 
DM (%)
SM (%)
DM + DBT (%)
SM + DBT (%)
Verified benign lesion detection (n = 83)
53.6
47.3
55.7
48.8
Non-dense breast (n = 42)
54.2
45.2
53.6
44.6
Dense breast (n = 41)
53.0
49.4
57.9
53.0
Calcification lesion (n = 39)
44.9
31.4
44.2
30.8
DM digital mammograms, SM synthesized mammograms, DBT digital breast tomosynthesis images
The sensitivities for DM, AI CAD SM, DM + DBT, and AI CAD SM + DBT in the benign calcification lesions (n = 39) were 44.9, 31.4, 44.2, and 30.8%, respectively (Table 3). The sensitivity of AI CAD SM alone was significantly lower than that of DM alone (P = 0.001). Similarly, the sensitivity of AI CAD SM + DBT was significantly lower than that of DM + DBT (P < 0.0001), and that of AI CAD SM alone was significantly lower than that of DM + DBT (P = 0.002).
The mean reading time per examination for AI CAD SM + DBT (91.87 s; 95% CI 70.46–113.27 s) was significantly shorter (P < 0.001) than that of DM + DBT (99.08 s, 95% CI 77.67–120.48 s; Table 4). Alternatively, the reading time was shorter using DM alone (19.95 s; 95% CI 19.95–62.77 s) when compared to that using AI CAD SM alone (42.37 s; 95% CI 20.96–63.78 s), statistical significance notwithstanding (P = 0.555).
Table 4
Analysis of the reading times
 
Mean (s)
95% CI
P value
vs. DM
vs. SM
vs. DM + DBT
DM
41.36
19.95–62.77
   
SM
42.37
20.96–63.78
0.555
 
< 0.001
DM + DBT
99.08
77.67–120.48
< 0.001
  
SM + DBT
91.87
7.46–113.27
< 0.001
< 0.001
< 0.001
DM digital mammograms, SM synthesized mammograms, DBT digital breast tomosynthesis images, CI   confidence interval

Discussion

The latest version of image processing technology was used to generate a novel SM in this study. The new things about it are as follows. First, the use of this new SM (AI CAD SM) alone increased the reader performance for breast cancer detection in terms of the overall accuracy when compared to DM alone. Second, the reading time was significantly decreased, and the accuracy was improved after using AI CAD SM + DBT when compared to that observed after using DM + DBT. Furthermore, the cancer detection sensitivity was markedly improved after using AI CAD SM alone when compared to that observed after using DM alone (nearly significant at P = 0.056). To the best of our knowledge, this is the first study to show that the detection accuracy of the AI CAD SM alone was equivalent to that of DM + DBT. Moreover, the benign recall rates of the AI CAD SM including calcification lesions were lower than those obtained with DM or DM + DBT. One explanation is that mass margins and calcification morphologies are accurately highlighted on AI CAD SM, and this may improve the perception of benign lesions. These results indicated that AI CAD SM maintained the performance benefits of the DBT examinations and could be used as standalone 2D mammograms with higher performance compared to DM alone. Thus, the latest version of the image processing algorithm for generating the AI CAD SM might meet the demand of radiologists.
The AI CAD SM, which was based on the deep convolutional neural network algorithm that processed individual DBT data, presented the suspicious findings that were enhanced by CAD and naturally blended into the corresponding standard AI CAD SM to improve the image quality. That result in the AI CAD SM as concurrent CAD for DBT, which accounts for the significant reduction in the reading time for AI CAD SM + DBT compared to DM + DBT. The reading time for DM alone was 1.01 s shorter than that with AI CAD SM alone (statistical significance notwithstanding). This might be because the readers might not have been able to reach the plateau in their learning curves with regard to the AI CAD SM, despite the training provided before the study. Nonetheless, AI CAD SM alone or in combination with DBT demonstrated a higher performance than DM (alone or DM in combination with DBT).
Evaluation of the performance of the readers based on the breast density is important for screening mammography. In the case of dense breasts, the cancer detection sensitivity of AI CAD SM or AI CAD SM + DBT was better than that of DM or DM + DBT, statistical significance notwithstanding. This finding is in line with those reported previously [1217]. The sensitivity of AI CAD SM alone in the verified benign cases was significantly lower than that of DM + DBT, which might account for the significantly higher cancer detection accuracy with AI CAD SM when compared to DM. The AI CAD SM generated from DBT data eliminating overlapping breast tissue have power of resistance to dense breasts to easily distinguish malignant lesions and benign lesions. Thus, AI CAD SM should be used proactively in women with dense breasts.
Evaluation of the performance of the readers based on the calcification lesions is important, particularly when using the AI CAD SM.
The sensitivities of DM, AI CAD SM, DM + DBT, and AI CAD SM + DBT were similar for malignant calcification lesions. In the benign calcification lesions, the sensitivities of AI CAD SM alone or AI CAD SM + DBT were significantly lower than those of DM alone or DM + DBT. In addition, the sensitivity of AI CAD SM alone was significantly lower than that of DM + DBT. These findings indicated that AI CAD SM was significantly superior to DM in terms of the specificity and sensitivity for the detection of calcification lesions in a screening setting. Previous studies have reported that the detection and interpretation of calcifications, especially amorphous calcifications, using SM was challenging [13, 1822]. However, this study showed that the new algorism enhances the contrast of the detail of interest without creating an artifact and suppressing the structure noise thereby improving the visualization of the calcifications on the AI CAD SM. Furthermore, in a diagnostic setting, indeterminate calcifications such as amorphous calcifications are always evaluated using spot compression magnification views to further examine their features during routine clinical practice and to plan their management. Thus, it is important for SM promotion activities not to worry too much for calcification detection and diagnosis. If not using SM in breast imaging, we cannot benefit from the SM faculty forever.
This study has several limitations. First, this study used so an enriched data with malignant cases; hence, the results may not be generalized to a real-world screening program. Second, this study used only the MLO view; access to the CC view could slightly increase the sensitivity and specificity of the results [23]. Thus, further validation using two views might be necessary. Finally, this is a retrospective, single-institution study with a single vendor prototype system comprising cases and radiologists from one country. These study characteristics potentially limit the generalizability of our results for the designing of prospective assessments across a diverse group of radiologists, patients, and institutions. Therefore, the findings of this study should be considered as preliminary results; however, the current SM generated from DBT is already commonly enhanced with suspicious lesions detected by AI CAD [24]. Nevertheless, to the best of our knowledge, this is the first study to assess the reader performance of a new SM and compare it with the original full-field DM when used alone or in combination with DBT in a virtual screening setting.
In conclusion, the use of the AI CAD SM + DBT could allow for a more effective screening program with higher performance, especially in terms of an increase in the detection accuracy and a reduction of in the reading time as well as radiation dose, when compared to the DM + DBT. The AI CAD SM might be considered as standalone 2D mammograms with higher performance compared to the DM.

Acknowledgements

We thank the Department of Medical Statistics, Satista Co., Ltd. for the statistical analysis review. We also thank Enago for the English language review.

Declarations

Ethical approval

The study was approved by the Institutional Review Board of our hospital.
Written informed consent to was obtained from all the prospectively recruited patients enrolled in this study.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Alabousi M, Wadera A, Kashif Al-Ghita M, Kashef Al-Ghetaa R, Salameh JP, Pozdnyakov A, et al. Performance of digital breast tomosynthesis, synthetic mammography, and digital mammography in breast cancer screening: a systematic review and meta-analysis. J Natl Cancer Inst. 2021;113:680–90.CrossRef Alabousi M, Wadera A, Kashif Al-Ghita M, Kashef Al-Ghetaa R, Salameh JP, Pozdnyakov A, et al. Performance of digital breast tomosynthesis, synthetic mammography, and digital mammography in breast cancer screening: a systematic review and meta-analysis. J Natl Cancer Inst. 2021;113:680–90.CrossRef
2.
Zurück zum Zitat Heywang-Köbrunner SH, Jänsch A, Hacker A, Weinand S, Vogelmann T. Digital breast tomosynthesis (DBT) plus synthesised two-dimensional mammography (s2D) in breast cancer screening is associated with higher cancer detection and lower recalls compared to digital mammography (DM) alone: results of a systematic review and meta-analysis. Eur Radiol. 2022;32:2301–12.CrossRef Heywang-Köbrunner SH, Jänsch A, Hacker A, Weinand S, Vogelmann T. Digital breast tomosynthesis (DBT) plus synthesised two-dimensional mammography (s2D) in breast cancer screening is associated with higher cancer detection and lower recalls compared to digital mammography (DM) alone: results of a systematic review and meta-analysis. Eur Radiol. 2022;32:2301–12.CrossRef
3.
Zurück zum Zitat Zeng B, Yu K, Gao L, Zeng X, Zhou Q. Breast cancer screening using synthesized two-dimensional mammography: a systematic review and meta-analysis. Breast. 2021;59:270–8.CrossRef Zeng B, Yu K, Gao L, Zeng X, Zhou Q. Breast cancer screening using synthesized two-dimensional mammography: a systematic review and meta-analysis. Breast. 2021;59:270–8.CrossRef
4.
Zurück zum Zitat Skaane P, Bandos AI, Gullien R, Eben EB, Ekseth U, Haakenaasen U, et al. Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology. 2013;267:47–56.CrossRef Skaane P, Bandos AI, Gullien R, Eben EB, Ekseth U, Haakenaasen U, et al. Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology. 2013;267:47–56.CrossRef
5.
Zurück zum Zitat Lehman CD, Wellman RD, Buist DS, Kerlikowske K, Tosteson AN, Miglioretti DL. Diagnostic accuracy of digital screening mammography with and without computer-aided detection. JAMA Intern Med. 2015;175:1828–37.CrossRef Lehman CD, Wellman RD, Buist DS, Kerlikowske K, Tosteson AN, Miglioretti DL. Diagnostic accuracy of digital screening mammography with and without computer-aided detection. JAMA Intern Med. 2015;175:1828–37.CrossRef
6.
Zurück zum Zitat Baldelli P, Cardarelli P, Flanagan F, Maguire S, Phelan N, Tomasi S, et al. Evaluation of microcalcification contrast in clinical images for digital mammography and synthetic mammography. Eur J Radiol. 2021;140: 109751.CrossRef Baldelli P, Cardarelli P, Flanagan F, Maguire S, Phelan N, Tomasi S, et al. Evaluation of microcalcification contrast in clinical images for digital mammography and synthetic mammography. Eur J Radiol. 2021;140: 109751.CrossRef
7.
Zurück zum Zitat Benedikt RA, Boatsman JE, Swann CA, Kirkpatrick AD, Toledano AY. Concurrent computer-aided detection improves reading time of digital breast tomosynthesis and maintains interpretation performance in a multireader multicase study. AJR Am J Roentgenol. 2018;210:685–94.CrossRef Benedikt RA, Boatsman JE, Swann CA, Kirkpatrick AD, Toledano AY. Concurrent computer-aided detection improves reading time of digital breast tomosynthesis and maintains interpretation performance in a multireader multicase study. AJR Am J Roentgenol. 2018;210:685–94.CrossRef
8.
Zurück zum Zitat Balleyguier C, Arfi-Rouche J, Levy L, Toubiana PR, Cohen-Scali F, Toledano AY, et al. Improving digital breast tomosynthesis reading time: a pilot multi-reader, multi-case study using concurrent computer-aided detection (CAD). Eur J Radiol. 2017;103:83–9.CrossRef Balleyguier C, Arfi-Rouche J, Levy L, Toubiana PR, Cohen-Scali F, Toledano AY, et al. Improving digital breast tomosynthesis reading time: a pilot multi-reader, multi-case study using concurrent computer-aided detection (CAD). Eur J Radiol. 2017;103:83–9.CrossRef
9.
Zurück zum Zitat James JJ, Giannotti E, Chen Y. Evaluation of a computer-aided detection (CAD)-enhanced 2D synthetic mammogram: comparison with standard synthetic 2D mammograms and conventional 2D digital mammography. Clin Radiol. 2018;73:886–92.CrossRef James JJ, Giannotti E, Chen Y. Evaluation of a computer-aided detection (CAD)-enhanced 2D synthetic mammogram: comparison with standard synthetic 2D mammograms and conventional 2D digital mammography. Clin Radiol. 2018;73:886–92.CrossRef
10.
Zurück zum Zitat American College of Radiology. Breast imaging reporting and data system (BI-RADS). 5th ed. Reston: ACR; 2013. American College of Radiology. Breast imaging reporting and data system (BI-RADS). 5th ed. Reston: ACR; 2013.
12.
Zurück zum Zitat Choi JS, Han BK, Ko EY, Ko ES, Hahn SY, Shin JH, et al. Comparison between two-dimensional synthetic mammography reconstructed from digital breast tomosynthesis and full-field digital mammography for the detection of T1 breast cancer. Eur Radiol. 2016;26:2538–46.CrossRef Choi JS, Han BK, Ko EY, Ko ES, Hahn SY, Shin JH, et al. Comparison between two-dimensional synthetic mammography reconstructed from digital breast tomosynthesis and full-field digital mammography for the detection of T1 breast cancer. Eur Radiol. 2016;26:2538–46.CrossRef
13.
Zurück zum Zitat Mariscotti G, Durando M, Houssami N, Fasciano M, Tagliafico A, Bosco D, et al. Comparison of synthetic mammography, reconstructed from digital breast tomosynthesis, and digital mammography: evaluation of lesion conspicuity and BI-RADS assessment categories. Breast Cancer Res Treat. 2017;166:765–73.CrossRef Mariscotti G, Durando M, Houssami N, Fasciano M, Tagliafico A, Bosco D, et al. Comparison of synthetic mammography, reconstructed from digital breast tomosynthesis, and digital mammography: evaluation of lesion conspicuity and BI-RADS assessment categories. Breast Cancer Res Treat. 2017;166:765–73.CrossRef
14.
Zurück zum Zitat Mesurolle B, El Khoury M, Travade A, Bagard C, Pétrou A, Monghal C. Is there any added value to substitute the 2D digital MLO projection for a MLO tomosynthesis projection and its synthetic view when a 2D standard digital mammography is used in a one-stop-shop immediate reading mammography screening? Eur Radiol. 2021;31:9529–39.CrossRef Mesurolle B, El Khoury M, Travade A, Bagard C, Pétrou A, Monghal C. Is there any added value to substitute the 2D digital MLO projection for a MLO tomosynthesis projection and its synthetic view when a 2D standard digital mammography is used in a one-stop-shop immediate reading mammography screening? Eur Radiol. 2021;31:9529–39.CrossRef
15.
Zurück zum Zitat Østerås BH, Martinsen ACT, Gullien R, Skaane P. Digital mammography versus breast tomosynthesis: impact of breast density on diagnostic performance in population-based screening. Radiology. 2019;293:60–8.CrossRef Østerås BH, Martinsen ACT, Gullien R, Skaane P. Digital mammography versus breast tomosynthesis: impact of breast density on diagnostic performance in population-based screening. Radiology. 2019;293:60–8.CrossRef
16.
Zurück zum Zitat Lowry KP, Coley RY, Miglioretti DL, Kerlikowske K, Henderson LM, Onega T, et al. Screening performance of digital breast tomosynthesis vs digital mammography in community practice by patient age, screening round, and breast density. JAMA Netw Open. 2020;3: e2011792.CrossRef Lowry KP, Coley RY, Miglioretti DL, Kerlikowske K, Henderson LM, Onega T, et al. Screening performance of digital breast tomosynthesis vs digital mammography in community practice by patient age, screening round, and breast density. JAMA Netw Open. 2020;3: e2011792.CrossRef
17.
Zurück zum Zitat Rafferty EA, Park JM, Philpotts LE, Poplack SP, Sumkin JH, Halpern EF, et al. Diagnostic accuracy and recall rates for digital mammography and digital mammography combined with one-view and two-view tomosynthesis: results of an enriched reader study. AJR Am J Roentgenol. 2014;202:273–81.CrossRef Rafferty EA, Park JM, Philpotts LE, Poplack SP, Sumkin JH, Halpern EF, et al. Diagnostic accuracy and recall rates for digital mammography and digital mammography combined with one-view and two-view tomosynthesis: results of an enriched reader study. AJR Am J Roentgenol. 2014;202:273–81.CrossRef
18.
Zurück zum Zitat Durand MA. Synthesized mammography: clinical evidence, appearance, and implementation. Diagnostics (Basel). 2018;8:22.CrossRef Durand MA. Synthesized mammography: clinical evidence, appearance, and implementation. Diagnostics (Basel). 2018;8:22.CrossRef
19.
Zurück zum Zitat Zuckerman SP, Sprague BL, Weaver DL, Herschorn SD, Conant EF. Survey results regarding uptake and impact of synthetic digital mammography with tomosynthesis in the screening setting. J Am Coll Radiol. 2020;17:31–7.CrossRef Zuckerman SP, Sprague BL, Weaver DL, Herschorn SD, Conant EF. Survey results regarding uptake and impact of synthetic digital mammography with tomosynthesis in the screening setting. J Am Coll Radiol. 2020;17:31–7.CrossRef
20.
Zurück zum Zitat Gilbert FJ, Tucker L, Gillan MG, et al. Accuracy of digital breast tomosynthesis for depicting breast cancer subgroups in a UK retrospective reading study (TOMMY Trial). Radiology. 2015;277:697–706.CrossRef Gilbert FJ, Tucker L, Gillan MG, et al. Accuracy of digital breast tomosynthesis for depicting breast cancer subgroups in a UK retrospective reading study (TOMMY Trial). Radiology. 2015;277:697–706.CrossRef
21.
Zurück zum Zitat Zuckerman SP, Maidment ADA, Weinstein SP, McDonald ES, Conant EF. Imaging with synthesized 2d mammography: differences, advantages, and pitfalls compared with digital mammography. AJR Am J Roentgenol. 2017;209:222–9.CrossRef Zuckerman SP, Maidment ADA, Weinstein SP, McDonald ES, Conant EF. Imaging with synthesized 2d mammography: differences, advantages, and pitfalls compared with digital mammography. AJR Am J Roentgenol. 2017;209:222–9.CrossRef
22.
Zurück zum Zitat Nelson JS, Wells JR, Baker JA, Samei E. How does c-view image quality compare with conventional 2D FFDM? Med Phys. 2016;43:2538.CrossRef Nelson JS, Wells JR, Baker JA, Samei E. How does c-view image quality compare with conventional 2D FFDM? Med Phys. 2016;43:2538.CrossRef
23.
Zurück zum Zitat Zackrisson S, Lång K, Rosso A, Johnson K, Dustler M, Förnvik D, et al. One-view breast tomosynthesis versus two-view mammography in the Malmö breast tomosynthesis screening trial (MBTST): a prospective, population-based, diagnostic accuracy study. Lancet Oncol. 2018;19:1493–503.CrossRef Zackrisson S, Lång K, Rosso A, Johnson K, Dustler M, Förnvik D, et al. One-view breast tomosynthesis versus two-view mammography in the Malmö breast tomosynthesis screening trial (MBTST): a prospective, population-based, diagnostic accuracy study. Lancet Oncol. 2018;19:1493–503.CrossRef
24.
Zurück zum Zitat Chan HP, Helvie MA. Using single-view wide-angle dbt with AI for breast cancer screening. Radiology. 2021;300:537–8.CrossRef Chan HP, Helvie MA. Using single-view wide-angle dbt with AI for breast cancer screening. Radiology. 2021;300:537–8.CrossRef
Metadaten
Titel
Comparisons between artificial intelligence computer-aided detection synthesized mammograms and digital mammograms when used alone and in combination with tomosynthesis images in a virtual screening setting
verfasst von
Takayoshi Uematsu
Kazuaki Nakashima
Taiyo Leopoldo Harada
Hatsuko Nasu
Tatsuya Igarashi
Publikationsdatum
07.09.2022
Verlag
Springer Nature Singapore
Erschienen in
Japanese Journal of Radiology / Ausgabe 1/2023
Print ISSN: 1867-1071
Elektronische ISSN: 1867-108X
DOI
https://doi.org/10.1007/s11604-022-01327-5

Weitere Artikel der Ausgabe 1/2023

Japanese Journal of Radiology 1/2023 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.