Gated Multidetector Computed Tomography

A Technique to Reduce Intracranial Aneurysm Clip and Coil Artifacts

verfasst von: A. Kovács, M.D., S. Flacke, H. Tschampa, D. Hadizadeh, S. Greschus, H. Clusmann, R. Kristof, H. Urbach

Erschienen in: Clinical Neuroradiology | Ausgabe 2/2010

Einloggen, um Zugang zu erhalten

Abstract

Background and Purpose:

Streak artifacts caused by aneurysm clips and coils impede image quality in multidetector computed tomography (MDCT). The authors propose a technique to minimize these artifacts by gated data reconstruction and shifting the reconstruction window.

Patients and Methods:

Intracranial CT angiograms were acquired in the follow-up of six patients with clipped and coiled intracranial aneurysms, respectively. Images were reconstructed from four consecutive 45° rotated segments with an acquisition time of 52.5 ms/segment. Data acquisition was gated via an external pacemaker cable-connected to the scanner.

Results:

Artifact orientation could be rotated by shifting the reconstruction window and interesting vessel segments visualized without disturbing streak artifacts. This allowed to assess the posterior communicating artery origin in two cases and a middle cerebral artery aneurysm remnant in another case, respectively. However, due to a higher noise interesting vessel segments were not adjustable in another three patients.

Conclusion:

Gated MDCT is a promising technique to reduce the amount and to change the position of artifacts induced by clips or coils.
Literatur
1.
Zurück zum Zitat Urbach H, Dorenbeck U, von Falkenhausen M, Wilhelm K, Willinek W, Schaller C, Flacke S. Three-dimensional time-of-flight MR angiography at 3 T compared to digital subtraction angiography in the follow-up of ruptured and coiled intracranial aneurysms: a prospective study. Neuroradiology. 2008;50:383–9.CrossRefPubMed Urbach H, Dorenbeck U, von Falkenhausen M, Wilhelm K, Willinek W, Schaller C, Flacke S. Three-dimensional time-of-flight MR angiography at 3 T compared to digital subtraction angiography in the follow-up of ruptured and coiled intracranial aneurysms: a prospective study. Neuroradiology. 2008;50:383–9.CrossRefPubMed
2.
Zurück zum Zitat Willinsky RA, Taylor SM, TerBrugge K, Farb RI, Tomlinson G, Montanera W. Neurologic complications of cerebral angiography: prospective analysis of 2899 procedures and review of the literature. Radiology. 2003;227:522–8.CrossRefPubMed Willinsky RA, Taylor SM, TerBrugge K, Farb RI, Tomlinson G, Montanera W. Neurologic complications of cerebral angiography: prospective analysis of 2899 procedures and review of the literature. Radiology. 2003;227:522–8.CrossRefPubMed
3.
Zurück zum Zitat Kovács A, Probst C, Sommer T, Leiss A, Naehle P, Welz A, Schild H, Flacke S. CT coronary angiography in patients with atrial fibrillation. Fortschr Röntgenstr. 2005;177:1655–62. Kovács A, Probst C, Sommer T, Leiss A, Naehle P, Welz A, Schild H, Flacke S. CT coronary angiography in patients with atrial fibrillation. Fortschr Röntgenstr. 2005;177:1655–62.
4.
Zurück zum Zitat van der Schaaf I, van Leeuwen M, Vlassenbroek A, Velthuis B. Minimizing clip artifacts in multi CT angiography of clipped patients. AJNR. 2006;27:60–6. van der Schaaf I, van Leeuwen M, Vlassenbroek A, Velthuis B. Minimizing clip artifacts in multi CT angiography of clipped patients. AJNR. 2006;27:60–6.
5.
Zurück zum Zitat Mamourian AC, Pluta DJ, Eskey CJ, Merlis AL. Optimizing computed tomography to reduce artifacts from titanium aneurysm clips: an in vitro study. J Neurosurg. 2007;107:1238–43.CrossRefPubMed Mamourian AC, Pluta DJ, Eskey CJ, Merlis AL. Optimizing computed tomography to reduce artifacts from titanium aneurysm clips: an in vitro study. J Neurosurg. 2007;107:1238–43.CrossRefPubMed
6.
Zurück zum Zitat Ertl-Wagner BB, Hoffmann RT, Bruning R, Herrmann K, Snyder B, Blume JD, Reiser MF. Multi-detector row CT angiography of the brain at various kilovoltage settings. Radiology. 2004;231:528–35.CrossRefPubMed Ertl-Wagner BB, Hoffmann RT, Bruning R, Herrmann K, Snyder B, Blume JD, Reiser MF. Multi-detector row CT angiography of the brain at various kilovoltage settings. Radiology. 2004;231:528–35.CrossRefPubMed
7.
Zurück zum Zitat Waaijer A, Prokop M, Velthuis BK, Bakker CJ, de Kort GA, van Leuwen MS. Circle of Willis at CT angiography: dose reduction and image quality – reducing tube voltage and increasing tube current settings. Radiology. 2007;242:832–9. Waaijer A, Prokop M, Velthuis BK, Bakker CJ, de Kort GA, van Leuwen MS. Circle of Willis at CT angiography: dose reduction and image quality – reducing tube voltage and increasing tube current settings. Radiology. 2007;242:832–9.
8.
Zurück zum Zitat Meyer LT, Boll DT. Novel technique for addressing streak artifact in gated dual-source MDCT angiography utilizing ECG-editing. Eur Radiol. 2008;18:2446–8.CrossRefPubMed Meyer LT, Boll DT. Novel technique for addressing streak artifact in gated dual-source MDCT angiography utilizing ECG-editing. Eur Radiol. 2008;18:2446–8.CrossRefPubMed
Metadaten
Titel
Gated Multidetector Computed Tomography
A Technique to Reduce Intracranial Aneurysm Clip and Coil Artifacts
verfasst von
A. Kovács, M.D.
S. Flacke
H. Tschampa
D. Hadizadeh
S. Greschus
H. Clusmann
R. Kristof
H. Urbach
Publikationsdatum
01.06.2010
Verlag
Urban and Vogel
Erschienen in
Clinical Neuroradiology / Ausgabe 2/2010
Print ISSN: 1869-1439
Elektronische ISSN: 1869-1447
DOI
https://doi.org/10.1007/s00062-010-0001-1

Weitere Artikel der Ausgabe 2/2010

Clinical Neuroradiology 2/2010 Zur Ausgabe

New in the Web

iNeuroradiology

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.