Diagnostic performance of dual-energy CT and subtraction CT for renal lesion detection and characterization

verfasst von: Ali Pourvaziri, Anushri Parakh, Amirkasra Mojtahed, Avinash Kambadakone, Dushyant Vasudeo Sahani

Erschienen in: European Radiology | Ausgabe 12/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To compare the effect of dual-energy CT (DECT) material density datasets on diagnostic performance, readers’ confidence, and interpretation time for renal lesion detection and characterization in comparison to subtraction CT (SCT).

Material and methods

One hundred fourteen patients (69/45 = M/F, mean age = 67 years) who underwent contrast-enhanced DECT between January 2015 and February 2018 for suspected renal mass were included retrospectively. For each patient, three radiologists assessed three image datasets: group A, material density iodine (MDI) + material density water (MDW); group B, SCT only; and group C, SCT + true unenhanced phase + virtual monochromatic images at 65 keV. Readers evaluated image quality (4-point scale), the number of lesions, and likely diagnosis. Reading times were recorded. Quantitatively, iodine concentration (IC from MDI) and delta Hounsfield units (ΔHU) for all lesions were measured. Diagnostic accuracy was compared using the area under the receiver operating characteristic curve (AUC). Image quality and interpretation time were compared with Kruskal-Wallis and t tests.

Results

Study cohort (230 lesions; mean size = 23.63 mm (5–116 mm)) consisted of 60 enhancing, 158 non-enhancing, and 12 lipid-dominant angiomyolipoma lesions. Significantly higher image quality was demonstrated for MDI compared to SCT (mean score = 3.82 vs. 3; p < 0.05). Comparable diagnostic accuracy was observed for group A (AUC = 0.88) and group C (AUC = 0.87) and was higher compared to that for group B (AUC = 0.75). Group A was read faster than group C (41.49 s vs. 71.45 s per exam; p < 0.05). Both IC and ΔHU values had high accuracy (AUC = 0.97) for differentiating enhancing vs. non-enhancing lesions; however, IC enabled differentiation of clear cell renal cell carcinoma from other enhancing lesions with moderate accuracy (AUC = 0.73).

Conclusion

MDI images increase readers’ confidence for renal lesion detection and characterization while providing a more efficient radiologist workflow, irrespective of readers’ experience.

Key Points

• Material density iodine (MDI) images enable faster interpretation due to high image quality and potentially reduced need for quantitation.
• MDI images increase diagnostic confidence of readers, irrespective of radiologists’ experience.
• High accuracy with dual-energy CT (DECT) can potentially reduce healthcare costs by eliminating the need for additional investigations.
Literatur
1.
Zurück zum Zitat Al Harbi F, Tabatabaeefar L, Jewett MA, Finelli A, O’Malley M, Atri M (2016) Enhancement threshold of small (< 4 cm) solid renal masses on CT. AJR Am J Roentgenol 206:554–558PubMedCrossRef Al Harbi F, Tabatabaeefar L, Jewett MA, Finelli A, O’Malley M, Atri M (2016) Enhancement threshold of small (< 4 cm) solid renal masses on CT. AJR Am J Roentgenol 206:554–558PubMedCrossRef
2.
Zurück zum Zitat Apfaltrer P, Meyer M, Meier C et al (2012) Contrast-enhanced dual-energy CT of gastrointestinal stromal tumors: is iodine-related attenuation a potential indicator of tumor response? Invest Radiol 47:65PubMedCrossRef Apfaltrer P, Meyer M, Meier C et al (2012) Contrast-enhanced dual-energy CT of gastrointestinal stromal tumors: is iodine-related attenuation a potential indicator of tumor response? Invest Radiol 47:65PubMedCrossRef
3.
Zurück zum Zitat Ascenti G, Mileto A, Krauss B et al (2013) Distinguishing enhancing from nonenhancing renal masses with dual-source dual-energy CT: iodine quantification versus standard enhancement measurements. Eur Radiol 23:2288–2295PubMedCrossRef Ascenti G, Mileto A, Krauss B et al (2013) Distinguishing enhancing from nonenhancing renal masses with dual-source dual-energy CT: iodine quantification versus standard enhancement measurements. Eur Radiol 23:2288–2295PubMedCrossRef
4.
Zurück zum Zitat Baerends E, Oostveen LJ, Smit CT et al (2018) Comparing dual energy CT and subtraction CT on a phantom: which one provides the best contrast in iodine maps for sub-centimetre details? Eur Radiol 28:5051–5059PubMedPubMedCentralCrossRef Baerends E, Oostveen LJ, Smit CT et al (2018) Comparing dual energy CT and subtraction CT on a phantom: which one provides the best contrast in iodine maps for sub-centimetre details? Eur Radiol 28:5051–5059PubMedPubMedCentralCrossRef
5.
Zurück zum Zitat Birnbaum BA, Maki DD, Chakraborty DP, Jacobs JE, Babb JS (2002) Renal cyst pseudoenhancement: evaluation with an anthropomorphic body CT phantom. Radiology 225:83–90PubMedCrossRef Birnbaum BA, Maki DD, Chakraborty DP, Jacobs JE, Babb JS (2002) Renal cyst pseudoenhancement: evaluation with an anthropomorphic body CT phantom. Radiology 225:83–90PubMedCrossRef
6.
Zurück zum Zitat Birnbaum BA, Hindman N, Lee J, Babb JS (2007) Multi–detector row CT attenuation measurements: assessment of intra- and interscanner variability with an anthropomorphic body CT phantom. Radiology 242:109–119PubMedCrossRef Birnbaum BA, Hindman N, Lee J, Babb JS (2007) Multi–detector row CT attenuation measurements: assessment of intra- and interscanner variability with an anthropomorphic body CT phantom. Radiology 242:109–119PubMedCrossRef
7.
Zurück zum Zitat Borhani AA, Kulzer M, Iranpour N et al (2017) Comparison of true unenhanced and virtual unenhanced (VUE) attenuation values in abdominopelvic single-source rapid kilovoltage-switching spectral CT. Abdom Radiol (NY) 42:710–717CrossRef Borhani AA, Kulzer M, Iranpour N et al (2017) Comparison of true unenhanced and virtual unenhanced (VUE) attenuation values in abdominopelvic single-source rapid kilovoltage-switching spectral CT. Abdom Radiol (NY) 42:710–717CrossRef
8.
Zurück zum Zitat Cha D, Kim CK, Park JJ, Park BK (2016) Evaluation of hyperdense renal lesions incidentally detected on single-phase post-contrast CT using dual-energy CT. Br J Radiol 89:20150860PubMedPubMedCentralCrossRef Cha D, Kim CK, Park JJ, Park BK (2016) Evaluation of hyperdense renal lesions incidentally detected on single-phase post-contrast CT using dual-energy CT. Br J Radiol 89:20150860PubMedPubMedCentralCrossRef
9.
Zurück zum Zitat Chandarana H, Megibow AJ, Cohen BA et al (2011) Iodine quantification with dual-energy CT: phantom study and preliminary experience with renal masses. AJR Am J Roentgenol 196:W693–W700PubMedCrossRef Chandarana H, Megibow AJ, Cohen BA et al (2011) Iodine quantification with dual-energy CT: phantom study and preliminary experience with renal masses. AJR Am J Roentgenol 196:W693–W700PubMedCrossRef
10.
Zurück zum Zitat Chandler A, Wei W, Herron DH, Anderson EF, Johnson VE, Ng CS (2011) Semiautomated motion correction of tumors in lung CT-perfusion studies. Acad Radiol 18:286–293PubMedCrossRef Chandler A, Wei W, Herron DH, Anderson EF, Johnson VE, Ng CS (2011) Semiautomated motion correction of tumors in lung CT-perfusion studies. Acad Radiol 18:286–293PubMedCrossRef
11.
Zurück zum Zitat Chandler A, Wei W, Anderson EF, Herron DH, Ye Z, Ng CS (2012) Validation of motion correction techniques for liver CT perfusion studies. Br J Radiol 85:e514–e522PubMedPubMedCentralCrossRef Chandler A, Wei W, Anderson EF, Herron DH, Ye Z, Ng CS (2012) Validation of motion correction techniques for liver CT perfusion studies. Br J Radiol 85:e514–e522PubMedPubMedCentralCrossRef
12.
Zurück zum Zitat Dillman JR, Caoili EM, Cohan RH, Ellis JH, Francis IR, Schipper MJ (2008) Detection of upper tract urothelial neoplasms: sensitivity of axial, coronal reformatted, and curved-planar reformatted image-types utilizing 16-row multi-detector CT urography. Abdom Imaging 33:707–716PubMedCrossRef Dillman JR, Caoili EM, Cohan RH, Ellis JH, Francis IR, Schipper MJ (2008) Detection of upper tract urothelial neoplasms: sensitivity of axial, coronal reformatted, and curved-planar reformatted image-types utilizing 16-row multi-detector CT urography. Abdom Imaging 33:707–716PubMedCrossRef
13.
14.
Zurück zum Zitat Eisner B, Kambadakone A, Samir A, Sahani D (2010) 619 subtraction ct improves radiologist confidence in evaluation of enhancing renal lesions. J Urol 183:e243–e244 Eisner B, Kambadakone A, Samir A, Sahani D (2010) 619 subtraction ct improves radiologist confidence in evaluation of enhancing renal lesions. J Urol 183:e243–e244
15.
Zurück zum Zitat Feuerlein S, Heye TJ, Bashir MR, Boll DT (2012) Iodine quantification using dual-energy multidetector computed tomography imaging: phantom study assessing the impact of iterative reconstruction schemes and patient habitus on accuracy. Invest Radiol 47:656PubMedCrossRef Feuerlein S, Heye TJ, Bashir MR, Boll DT (2012) Iodine quantification using dual-energy multidetector computed tomography imaging: phantom study assessing the impact of iterative reconstruction schemes and patient habitus on accuracy. Invest Radiol 47:656PubMedCrossRef
16.
Zurück zum Zitat Goodsitt MM, Christodoulou EG, Larson SC (2011) Accuracies of the synthesized monochromatic CT numbers and effective atomic numbers obtained with a rapid kVp switching dual energy CT scanner. Med Phys 38:2222–2232PubMedCrossRef Goodsitt MM, Christodoulou EG, Larson SC (2011) Accuracies of the synthesized monochromatic CT numbers and effective atomic numbers obtained with a rapid kVp switching dual energy CT scanner. Med Phys 38:2222–2232PubMedCrossRef
17.
Zurück zum Zitat Graser A, Johnson TR, Hecht EM et al (2009) Dual-energy CT in patients suspected of having renal masses: can virtual nonenhanced images replace true nonenhanced images? Radiology 252:433–440PubMedCrossRef Graser A, Johnson TR, Hecht EM et al (2009) Dual-energy CT in patients suspected of having renal masses: can virtual nonenhanced images replace true nonenhanced images? Radiology 252:433–440PubMedCrossRef
18.
Zurück zum Zitat Graser A, Becker CR, Staehler M et al (2010) Single-phase dual-energy CT allows for characterization of renal masses as benign or malignant. Invest Radiol 45:399PubMedCrossRef Graser A, Becker CR, Staehler M et al (2010) Single-phase dual-energy CT allows for characterization of renal masses as benign or malignant. Invest Radiol 45:399PubMedCrossRef
19.
Zurück zum Zitat Herts BR, Silverman SG, Hindman NM et al (2018) Management of the incidental renal mass on CT: a white paper of the ACR Incidental Findings Committee. J Am Coll Radiol 15:264–273PubMedCrossRef Herts BR, Silverman SG, Hindman NM et al (2018) Management of the incidental renal mass on CT: a white paper of the ACR Incidental Findings Committee. J Am Coll Radiol 15:264–273PubMedCrossRef
20.
Zurück zum Zitat Hock LM, Lynch J, Balaji KC (2002) Increasing incidence of all stages of kidney cancer in the last 2 decades in the United States: an analysis of surveillance, epidemiology and end results program data. J Urol 167:57–60PubMedCrossRef Hock LM, Lynch J, Balaji KC (2002) Increasing incidence of all stages of kidney cancer in the last 2 decades in the United States: an analysis of surveillance, epidemiology and end results program data. J Urol 167:57–60PubMedCrossRef
21.
Zurück zum Zitat Hollingsworth JM, Miller DC, Daignault S, Hollenbeck BK (2006) Rising incidence of small renal masses: a need to reassess treatment effect. J Natl Cancer Inst 98:1331–1334PubMedCrossRef Hollingsworth JM, Miller DC, Daignault S, Hollenbeck BK (2006) Rising incidence of small renal masses: a need to reassess treatment effect. J Natl Cancer Inst 98:1331–1334PubMedCrossRef
22.
Zurück zum Zitat Jamis-Dow CA, Choyke PL, Jennings SB, Linehan WM, Thakore KN, Walther MM (1996) Small (< or = 3-cm) renal masses: detection with CT versus US and pathologic correlation. Radiology 198:785–788PubMedCrossRef Jamis-Dow CA, Choyke PL, Jennings SB, Linehan WM, Thakore KN, Walther MM (1996) Small (< or = 3-cm) renal masses: detection with CT versus US and pathologic correlation. Radiology 198:785–788PubMedCrossRef
23.
Zurück zum Zitat Jayson M, Sanders H (1998) Increased incidence of serendipitously discovered renal cell carcinoma. Urology 51:203–205PubMedCrossRef Jayson M, Sanders H (1998) Increased incidence of serendipitously discovered renal cell carcinoma. Urology 51:203–205PubMedCrossRef
24.
Zurück zum Zitat Jinzaki M, Tanimoto A, Mukai M et al (2000) Double-phase helical CT of small renal parenchymal neoplasms: correlation with pathologic findings and tumor angiogenesis. J Comput Assist Tomogr 24:835PubMedCrossRef Jinzaki M, Tanimoto A, Mukai M et al (2000) Double-phase helical CT of small renal parenchymal neoplasms: correlation with pathologic findings and tumor angiogenesis. J Comput Assist Tomogr 24:835PubMedCrossRef
25.
Zurück zum Zitat Kambadakone A, Arasu VA, Samir AE et al (2012) Qualitative assessment of enhancement in a renal mass: contribution of subtraction CT. J Comput Assist Tomogr 36:381PubMedCrossRef Kambadakone A, Arasu VA, Samir AE et al (2012) Qualitative assessment of enhancement in a renal mass: contribution of subtraction CT. J Comput Assist Tomogr 36:381PubMedCrossRef
26.
Zurück zum Zitat Kaza RK, Caoili EM, Cohan RH, Platt JF (2011) Distinguishing enhancing from nonenhancing renal lesions with fast kilovoltage-switching dual-energy CT. AJR Am J Roentgenol 197:1375–1381PubMedCrossRef Kaza RK, Caoili EM, Cohan RH, Platt JF (2011) Distinguishing enhancing from nonenhancing renal lesions with fast kilovoltage-switching dual-energy CT. AJR Am J Roentgenol 197:1375–1381PubMedCrossRef
27.
Zurück zum Zitat Koonce JD, Vliegenthart R, Schoepf UJ et al (2014) Accuracy of dual-energy computed tomography for the measurement of iodine concentration using cardiac CT protocols: validation in a phantom model. Eur Radiol 24:512–518PubMedCrossRef Koonce JD, Vliegenthart R, Schoepf UJ et al (2014) Accuracy of dual-energy computed tomography for the measurement of iodine concentration using cardiac CT protocols: validation in a phantom model. Eur Radiol 24:512–518PubMedCrossRef
28.
Zurück zum Zitat Mahmood U, Horvat N, Horvat JV et al (2018) Rapid switching kVp dual energy CT: value of reconstructed dual energy CT images and organ dose assessment in multiphasic liver CT exams. Eur J Radiol 102:102–108PubMedPubMedCentralCrossRef Mahmood U, Horvat N, Horvat JV et al (2018) Rapid switching kVp dual energy CT: value of reconstructed dual energy CT images and organ dose assessment in multiphasic liver CT exams. Eur J Radiol 102:102–108PubMedPubMedCentralCrossRef
29.
Zurück zum Zitat Manoharan D, Sharma S, Das CJ, Kumar R, Singh G, Kumar P (2018) Single-acquisition triple-bolus dual-energy CT protocol for comprehensive evaluation of renal masses: a single-center randomized noninferiority trial. AJR Am J Roentgenol 211:W1–W11CrossRef Manoharan D, Sharma S, Das CJ, Kumar R, Singh G, Kumar P (2018) Single-acquisition triple-bolus dual-energy CT protocol for comprehensive evaluation of renal masses: a single-center randomized noninferiority trial. AJR Am J Roentgenol 211:W1–W11CrossRef
30.
Zurück zum Zitat Marin D, Davis D, Roy Choudhury K et al (2017) Characterization of small focal renal lesions: diagnostic accuracy with single-phase contrast-enhanced dual-energy CT with material attenuation analysis compared with conventional attenuation measurements. Radiology 284:737–747PubMedCrossRef Marin D, Davis D, Roy Choudhury K et al (2017) Characterization of small focal renal lesions: diagnostic accuracy with single-phase contrast-enhanced dual-energy CT with material attenuation analysis compared with conventional attenuation measurements. Radiology 284:737–747PubMedCrossRef
31.
Zurück zum Zitat Matsumoto K, Jinzaki M, Tanami Y, Ueno A, Yamada M, Kuribayashi S (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 259:257–262PubMedCrossRef Matsumoto K, Jinzaki M, Tanami Y, Ueno A, Yamada M, Kuribayashi S (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 259:257–262PubMedCrossRef
32.
Zurück zum Zitat McCarthy CJ, Baliyan V, Kordbacheh H, Sajjad Z, Sahani D, Kambadakone A (2016) Radiology of renal stone disease. Int J Surg 36:638–646PubMedCrossRef McCarthy CJ, Baliyan V, Kordbacheh H, Sajjad Z, Sahani D, Kambadakone A (2016) Radiology of renal stone disease. Int J Surg 36:638–646PubMedCrossRef
33.
Zurück zum Zitat Mileto A, Nelson RC, Samei E et al (2014) Impact of dual-energy multi–detector row CT with virtual monochromatic imaging on renal cyst pseudoenhancement: in vitro and in vivo study. Radiology 272:767–776PubMedCrossRef Mileto A, Nelson RC, Samei E et al (2014) Impact of dual-energy multi–detector row CT with virtual monochromatic imaging on renal cyst pseudoenhancement: in vitro and in vivo study. Radiology 272:767–776PubMedCrossRef
34.
Zurück zum Zitat Mileto A, Marin D, Ramirez-Giraldo et al (2014b) Accuracy of contrast-enhanced dual-energy MDCT for the assessment of iodine uptake in renal lesions. AJR Am J Roentgenol 202:W466–W474PubMedCrossRef Mileto A, Marin D, Ramirez-Giraldo et al (2014b) Accuracy of contrast-enhanced dual-energy MDCT for the assessment of iodine uptake in renal lesions. AJR Am J Roentgenol 202:W466–W474PubMedCrossRef
35.
Zurück zum Zitat Mohr B, Brink M, Oostveen LJ, Schuijf JD, Prokop M (2016) Lung iodine mapping by subtraction with image registration allowing for tissue sliding. Proc. SPIE 9784, Medical Imaging 2016: Image Processing, 978442 Mohr B, Brink M, Oostveen LJ, Schuijf JD, Prokop M (2016) Lung iodine mapping by subtraction with image registration allowing for tissue sliding. Proc. SPIE 9784, Medical Imaging 2016: Image Processing, 978442
36.
Zurück zum Zitat Patel BN, Bibbey A, Choudhury KR, Leder RA, Nelson RC, Marin D (2017) Characterization of small (< 4 cm) focal renal lesions: diagnostic accuracy of spectral analysis using single-phase contrast-enhanced dual-energy CT. AJR Am J Roentgenol 209:815–825PubMedCrossRef Patel BN, Bibbey A, Choudhury KR, Leder RA, Nelson RC, Marin D (2017) Characterization of small (< 4 cm) focal renal lesions: diagnostic accuracy of spectral analysis using single-phase contrast-enhanced dual-energy CT. AJR Am J Roentgenol 209:815–825PubMedCrossRef
37.
Zurück zum Zitat Pelgrim GJ, van Hamersvelt RW, Willemink MJ et al (2017) Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT. Eur Radiol 27:3904–3912PubMedPubMedCentralCrossRef Pelgrim GJ, van Hamersvelt RW, Willemink MJ et al (2017) Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT. Eur Radiol 27:3904–3912PubMedPubMedCentralCrossRef
39.
Zurück zum Zitat Savci G, Yazici Z, Sahin N, Akgöz S, Tuncel E (2006) Value of chemical shift subtraction MRI in characterization of adrenal masses. AJR Am J Roentgenol 186:130–135PubMedCrossRef Savci G, Yazici Z, Sahin N, Akgöz S, Tuncel E (2006) Value of chemical shift subtraction MRI in characterization of adrenal masses. AJR Am J Roentgenol 186:130–135PubMedCrossRef
40.
Zurück zum Zitat Siegel CL, Fisher AJ, Bennett HF (1999) Interobserver variability in determining enhancement of renal masses on helical CT. AJR Am J Roentgenol 172:1207–1212PubMedCrossRef Siegel CL, Fisher AJ, Bennett HF (1999) Interobserver variability in determining enhancement of renal masses on helical CT. AJR Am J Roentgenol 172:1207–1212PubMedCrossRef
41.
Zurück zum Zitat Tada S, Yamagishi J, Kobayashi H, Hata Y, Kobari T (1983) The incidence of simple renal cyst by computed tomography. Clin Radiol 34:437–439PubMedCrossRef Tada S, Yamagishi J, Kobayashi H, Hata Y, Kobari T (1983) The incidence of simple renal cyst by computed tomography. Clin Radiol 34:437–439PubMedCrossRef
42.
Zurück zum Zitat Vasudevan A, Davies RJ, Shannon BA, Cohen RJ (2006) Incidental renal tumours: the frequency of benign lesions and the role of preoperative core biopsy. BJU Int 97:946–949PubMedCrossRef Vasudevan A, Davies RJ, Shannon BA, Cohen RJ (2006) Incidental renal tumours: the frequency of benign lesions and the role of preoperative core biopsy. BJU Int 97:946–949PubMedCrossRef
43.
Zurück zum Zitat Yoshioka K, Tanaka R, Takagi H et al (2016) Diagnostic accuracy of a modified subtraction coronary CT angiography method with short breath-holding time: a feasibility study. Br J Radiol 89:20160489PubMedPubMedCentralCrossRef Yoshioka K, Tanaka R, Takagi H et al (2016) Diagnostic accuracy of a modified subtraction coronary CT angiography method with short breath-holding time: a feasibility study. Br J Radiol 89:20160489PubMedPubMedCentralCrossRef
44.
Zurück zum Zitat Yu L, Christner JA, Leng S, Wang J, Fletcher JG, McCollough CH (2011) Virtual monochromatic imaging in dual-source dual-energy CT: radiation dose and image quality. Med Phys 38:6371–6379PubMedPubMedCentralCrossRef Yu L, Christner JA, Leng S, Wang J, Fletcher JG, McCollough CH (2011) Virtual monochromatic imaging in dual-source dual-energy CT: radiation dose and image quality. Med Phys 38:6371–6379PubMedPubMedCentralCrossRef
45.
Zurück zum Zitat Zarzour JG, Milner D, Valentin R et al (2017) Quantitative iodine content threshold for discrimination of renal cell carcinomas using rapid kV-switching dual-energy CT. Abdom Radiol (NY) 42:727–734CrossRef Zarzour JG, Milner D, Valentin R et al (2017) Quantitative iodine content threshold for discrimination of renal cell carcinomas using rapid kV-switching dual-energy CT. Abdom Radiol (NY) 42:727–734CrossRef
Metadaten
Titel
Diagnostic performance of dual-energy CT and subtraction CT for renal lesion detection and characterization
verfasst von
Ali Pourvaziri
Anushri Parakh
Amirkasra Mojtahed
Avinash Kambadakone
Dushyant Vasudeo Sahani
Publikationsdatum
27.05.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 12/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-019-06224-6

Weitere Artikel der Ausgabe 12/2019

European Radiology 12/2019 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.