Neointimal coverage of zotarolimus-eluting stent at 1, 2, and 3 months’ follow-up: an optical coherence tomography study

verfasst von: Takehiro Hashikata, Taiki Tojo, Sayaka Namba, Lisa Kitasato, Takuya Hashimoto, Ryo Kameda, Takao Shimohama, Minako Yamaoka-Tojo, Junya Ako

Erschienen in: Heart and Vessels | Ausgabe 2/2016

Einloggen, um Zugang zu erhalten

Abstract

Incomplete neointimal coverage and malapposed struts after stenting are associated with increased risk of stent thrombosis. We aimed to evaluate neointimal coverage early after Resolute zotarolimus-eluting stent (R-ZES) implantation using optical coherence tomography (OCT). A total of 20 patients with de novo native coronary lesions with R-ZES were enrolled. Among these patients, 20 stented lesions in 19 patients were evaluated at 1, 2, and 3 months after R-ZES implantation. The strut apposition and neointimal coverage were evaluated by OCT. Neointimal hyperplasia (NIH) thickness and percentage of covered struts and the proportion of incompletely apposed struts were measured at 1-mm intervals. The mean percentages of covered stent struts were over 85 % within 3 months (88.4 ± 6.3 % at 1 month, 95.5 ± 5.5 % at 2 months, 93.6 ± 3.5 % at 3 months). The percentages of incompletely apposed struts were not significantly different among the groups (4.4 ± 4.2 % at 1 month, 1.9 ± 1.9 % at 2 months, 3.1 ± 2.2 % at 3 months, p = 0.51). Mean NIH thickness (38.9 ± 8.1 μm at 1 month, 70.6 ± 18.8 μm at 2 months, 54.1 ± 5.9 at 3 months, p = 0.0016) was thickest in the 2 months group. Most of all OCT findings within 2 months demonstrated neointimal coverage with low signal intensity. The neointimal coverage of ZES-R was over 85 % within 3 months. These data may support shorter requirement of dual antiplatelet therapy duration with R-ZES.
Literatur
1.
Zurück zum Zitat Stefanini GG, Holmes DR Jr (2013) Drug-eluting coronary-artery stents. N Engl J Med 368:254–265CrossRefPubMed Stefanini GG, Holmes DR Jr (2013) Drug-eluting coronary-artery stents. N Engl J Med 368:254–265CrossRefPubMed
2.
Zurück zum Zitat Udipi K, Chen M, Cheng P, Jiang K, Judd D, Caceres A, Melder RJ, Wilcox JN (2008) Development of a novel biocompatible polymer system for extended drug release in a next-generation drug-eluting stent. J Biomed Mater Res A 85:1064–1071CrossRefPubMed Udipi K, Chen M, Cheng P, Jiang K, Judd D, Caceres A, Melder RJ, Wilcox JN (2008) Development of a novel biocompatible polymer system for extended drug release in a next-generation drug-eluting stent. J Biomed Mater Res A 85:1064–1071CrossRefPubMed
3.
Zurück zum Zitat Widimsky P, Motovska Z, Belardi J, Serruys P, Silber S, Windecker S, Neumann FJ (2013) Long-term outcomes of patients receiving zotarolimus-eluting stents in ST elevation myocardial infarction, non-ST elevation acute coronary syndrome, and stable angina: data from the Resolute program. Int J Cardiol 168:3522–3526CrossRefPubMed Widimsky P, Motovska Z, Belardi J, Serruys P, Silber S, Windecker S, Neumann FJ (2013) Long-term outcomes of patients receiving zotarolimus-eluting stents in ST elevation myocardial infarction, non-ST elevation acute coronary syndrome, and stable angina: data from the Resolute program. Int J Cardiol 168:3522–3526CrossRefPubMed
4.
Zurück zum Zitat Jim MH, Yiu KH, Fung RC, Ho HH, Ng AK, Siu CW, Chow WH (2014) Zotarolimus-eluting stent utilization in small-vessel coronary artery disease (ZEUS). Heart Vessels 29:29–34CrossRefPubMed Jim MH, Yiu KH, Fung RC, Ho HH, Ng AK, Siu CW, Chow WH (2014) Zotarolimus-eluting stent utilization in small-vessel coronary artery disease (ZEUS). Heart Vessels 29:29–34CrossRefPubMed
5.
Zurück zum Zitat Taniwaki M, Stefanini GG, Silber S, Richardt G, Vranckx P, Serruys PW, Buszman PE, Kelbaek H, Windecker S (2014) Four year clinical outcomes and predictors of repeat revascularization in patients treated with new generation drug-eluting stents in the RESOLUTE all comers randomized trial. J Am Coll Cardiol 63:1617–1625CrossRefPubMed Taniwaki M, Stefanini GG, Silber S, Richardt G, Vranckx P, Serruys PW, Buszman PE, Kelbaek H, Windecker S (2014) Four year clinical outcomes and predictors of repeat revascularization in patients treated with new generation drug-eluting stents in the RESOLUTE all comers randomized trial. J Am Coll Cardiol 63:1617–1625CrossRefPubMed
6.
Zurück zum Zitat Silber S, Kirtane AJ, Belardi JA, Liu M, Brar S, Rothman M, Windecker S (2014) Lack of association between dual antiplatelet therapy use and stent thrombosis between 1 and 12 months following resolute zotarolimus-eluting stent implantation. Eur Heart J 35:1949–1956CrossRefPubMed Silber S, Kirtane AJ, Belardi JA, Liu M, Brar S, Rothman M, Windecker S (2014) Lack of association between dual antiplatelet therapy use and stent thrombosis between 1 and 12 months following resolute zotarolimus-eluting stent implantation. Eur Heart J 35:1949–1956CrossRefPubMed
7.
Zurück zum Zitat Guagliumi G, Sirbu V, Musumeci G, Gerber R, Biondi-Zoccai G, Ikejima H, Ladich E, Lortkipanidze N, Matiashvili A, Valsecchi O, Virmani R, Stone GW (2012) Examination of the in vivo mechanisms of late drug-eluting stent thrombosis: findings from optical coherence tomography and intravascular ultrasound imaging. JACC Cardiovasc Interv 5:12–20CrossRefPubMed Guagliumi G, Sirbu V, Musumeci G, Gerber R, Biondi-Zoccai G, Ikejima H, Ladich E, Lortkipanidze N, Matiashvili A, Valsecchi O, Virmani R, Stone GW (2012) Examination of the in vivo mechanisms of late drug-eluting stent thrombosis: findings from optical coherence tomography and intravascular ultrasound imaging. JACC Cardiovasc Interv 5:12–20CrossRefPubMed
8.
Zurück zum Zitat Finn AV, Joner M, Nakazawa G, Kolodgie F, Newell J, John MC, Gold HK, Virmani R (2007) Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization. Circulation 115:2435–2441CrossRefPubMed Finn AV, Joner M, Nakazawa G, Kolodgie F, Newell J, John MC, Gold HK, Virmani R (2007) Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization. Circulation 115:2435–2441CrossRefPubMed
9.
Zurück zum Zitat Gutierrez-Chico JL, van Geuns RJ, Regar E, van der Giessen WJ, Kelbaek H, Saunamaki K, Escaned J, Gonzalo N, di Mario C, Borgia F, Nuesch E, Garcia-Garcia HM, Silber S, Windecker S, Serruys PW (2011) Tissue coverage of a hydrophilic polymer-coated zotarolimus-eluting stent vs. a fluoropolymer-coated everolimus-eluting stent at 13-month follow-up: an optical coherence tomography substudy from the RESOLUTE all comers trial. Eur Heart J 32:2454–2463PubMedCentralCrossRefPubMed Gutierrez-Chico JL, van Geuns RJ, Regar E, van der Giessen WJ, Kelbaek H, Saunamaki K, Escaned J, Gonzalo N, di Mario C, Borgia F, Nuesch E, Garcia-Garcia HM, Silber S, Windecker S, Serruys PW (2011) Tissue coverage of a hydrophilic polymer-coated zotarolimus-eluting stent vs. a fluoropolymer-coated everolimus-eluting stent at 13-month follow-up: an optical coherence tomography substudy from the RESOLUTE all comers trial. Eur Heart J 32:2454–2463PubMedCentralCrossRefPubMed
10.
Zurück zum Zitat Ikuta S, Kobuke K, Iwanaga Y, Nakauchi Y, Yamaji K, Miyazaki S (2014) Difference in neointimal coverage at chronic stage between bare metal stent and sirolimus-eluting stent evaluated at stent-strut level by optical coherence tomography. Heart Vessels 29:320–327CrossRefPubMed Ikuta S, Kobuke K, Iwanaga Y, Nakauchi Y, Yamaji K, Miyazaki S (2014) Difference in neointimal coverage at chronic stage between bare metal stent and sirolimus-eluting stent evaluated at stent-strut level by optical coherence tomography. Heart Vessels 29:320–327CrossRefPubMed
11.
Zurück zum Zitat Tada T, Byrne RA, Schuster T, Cuni R, Kitabata H, Tiroch K, Dirninger A, Gratze F, Kaspar K, Zenker G, Joner M, Schomig A, Kastrati A (2013) Early vascular healing with rapid breakdown biodegradable polymer sirolimus-eluting versus durable polymer everolimus-eluting stents assessed by optical coherence tomography. Cardiovasc Revasc Med 14:84–89CrossRefPubMed Tada T, Byrne RA, Schuster T, Cuni R, Kitabata H, Tiroch K, Dirninger A, Gratze F, Kaspar K, Zenker G, Joner M, Schomig A, Kastrati A (2013) Early vascular healing with rapid breakdown biodegradable polymer sirolimus-eluting versus durable polymer everolimus-eluting stents assessed by optical coherence tomography. Cardiovasc Revasc Med 14:84–89CrossRefPubMed
12.
Zurück zum Zitat Takano M, Inami S, Jang IK, Yamamoto M, Murakami D, Seimiya K, Ohba T, Mizuno K (2007) Evaluation by optical coherence tomography of neointimal coverage of sirolimus-eluting stent 3 months after implantation. Am J Cardiol 99:1033–1038CrossRefPubMed Takano M, Inami S, Jang IK, Yamamoto M, Murakami D, Seimiya K, Ohba T, Mizuno K (2007) Evaluation by optical coherence tomography of neointimal coverage of sirolimus-eluting stent 3 months after implantation. Am J Cardiol 99:1033–1038CrossRefPubMed
13.
Zurück zum Zitat Guagliumi G, Musumeci G, Sirbu V, Bezerra HG, Suzuki N, Fiocca L, Matiashvili A, Lortkipanidze N, Trivisonno A, Valsecchi O, Biondi-Zoccai G, Costa MA (2010) Optical coherence tomography assessment of in vivo vascular response after implantation of overlapping bare-metal and drug-eluting stents. JACC Cardiovasc Interv 3:531–539CrossRefPubMed Guagliumi G, Musumeci G, Sirbu V, Bezerra HG, Suzuki N, Fiocca L, Matiashvili A, Lortkipanidze N, Trivisonno A, Valsecchi O, Biondi-Zoccai G, Costa MA (2010) Optical coherence tomography assessment of in vivo vascular response after implantation of overlapping bare-metal and drug-eluting stents. JACC Cardiovasc Interv 3:531–539CrossRefPubMed
14.
Zurück zum Zitat Kim JS, Jang IK, Kim JS, Kim TH, Takano M, Kume T, Hur NW, Ko YG, Choi D, Hong MK, Jang Y (2009) Optical coherence tomography evaluation of zotarolimus-eluting stents at 9-month follow-up: comparison with sirolimus-eluting stents. Heart 95:1907–1912PubMedCentralCrossRefPubMed Kim JS, Jang IK, Kim JS, Kim TH, Takano M, Kume T, Hur NW, Ko YG, Choi D, Hong MK, Jang Y (2009) Optical coherence tomography evaluation of zotarolimus-eluting stents at 9-month follow-up: comparison with sirolimus-eluting stents. Heart 95:1907–1912PubMedCentralCrossRefPubMed
15.
Zurück zum Zitat Kume T, Akasaka T, Kawamoto T, Ogasawara Y, Watanabe N, Toyota E, Neishi Y, Sukmawan R, Sadahira Y, Yoshida K (2006) Assessment of coronary arterial thrombus by optical coherence tomography. Am J Cardiol 97:1713–1717CrossRefPubMed Kume T, Akasaka T, Kawamoto T, Ogasawara Y, Watanabe N, Toyota E, Neishi Y, Sukmawan R, Sadahira Y, Yoshida K (2006) Assessment of coronary arterial thrombus by optical coherence tomography. Am J Cardiol 97:1713–1717CrossRefPubMed
16.
Zurück zum Zitat Templin C, Meyer M, Muller MF, Djonov V, Hlushchuk R, Dimova I, Flueckiger S, Kronen P, Sidler M, Klein K, Nicholls F, Ghadri JR, Weber K, Paunovic D, Corti R, Hoerstrup SP, Luscher TF, Landmesser U (2010) Coronary optical frequency domain imaging (OFDI) for in vivo evaluation of stent healing: comparison with light and electron microscopy. Eur Heart J 31:1792–1801PubMedCentralCrossRefPubMed Templin C, Meyer M, Muller MF, Djonov V, Hlushchuk R, Dimova I, Flueckiger S, Kronen P, Sidler M, Klein K, Nicholls F, Ghadri JR, Weber K, Paunovic D, Corti R, Hoerstrup SP, Luscher TF, Landmesser U (2010) Coronary optical frequency domain imaging (OFDI) for in vivo evaluation of stent healing: comparison with light and electron microscopy. Eur Heart J 31:1792–1801PubMedCentralCrossRefPubMed
17.
Zurück zum Zitat Kim S, Kim JS, Shin DH, Kim BK, Ko YG, Choi D, Cho YK, Nam CW, Hur SH, Jang Y, Hong MK (2013) Comparison of early strut coverage between zotarolimus- and everolimus-eluting stents using optical coherence tomography. Am J Cardiol 111:1–5CrossRefPubMed Kim S, Kim JS, Shin DH, Kim BK, Ko YG, Choi D, Cho YK, Nam CW, Hur SH, Jang Y, Hong MK (2013) Comparison of early strut coverage between zotarolimus- and everolimus-eluting stents using optical coherence tomography. Am J Cardiol 111:1–5CrossRefPubMed
18.
Zurück zum Zitat Kim JS, Kim BK, Jang IK, Shin DH, Ko YG, Choi D, Hong MK, Cho YK, Nam CW, Hur SH, Choi JH, Song YB, Hahn JY, Choi SH, Gwon HC, Jang Y (2012) ComparisOn of neointimal coVerage betwEen zotaRolimus-eluting stent and everolimus-eluting stent using Optical Coherence Tomography (COVER OCT). Am Heart J 163:601–607CrossRefPubMed Kim JS, Kim BK, Jang IK, Shin DH, Ko YG, Choi D, Hong MK, Cho YK, Nam CW, Hur SH, Choi JH, Song YB, Hahn JY, Choi SH, Gwon HC, Jang Y (2012) ComparisOn of neointimal coVerage betwEen zotaRolimus-eluting stent and everolimus-eluting stent using Optical Coherence Tomography (COVER OCT). Am Heart J 163:601–607CrossRefPubMed
19.
Zurück zum Zitat Nishinari M, Shimohama T, Tojo T, Shiono T, Shinagawa H, Kameda R, Aoyama N, Izumi T (2013) Two-week interval optical coherence tomography: imaging evidence on neointimal coverage completion after implantation of the endeavor zotarolimus-eluting stent. Catheter Cardiovasc Interv 82:E871–E878CrossRefPubMed Nishinari M, Shimohama T, Tojo T, Shiono T, Shinagawa H, Kameda R, Aoyama N, Izumi T (2013) Two-week interval optical coherence tomography: imaging evidence on neointimal coverage completion after implantation of the endeavor zotarolimus-eluting stent. Catheter Cardiovasc Interv 82:E871–E878CrossRefPubMed
20.
Zurück zum Zitat Teramoto T, Ikeno F, Otake H, Lyons JK, van Beusekom HM, Fearon WF, Yeung AC (2010) Intriguing peri-strut low-intensity area detected by optical coherence tomography after coronary stent deployment. Circ J 74:1257–1259CrossRefPubMed Teramoto T, Ikeno F, Otake H, Lyons JK, van Beusekom HM, Fearon WF, Yeung AC (2010) Intriguing peri-strut low-intensity area detected by optical coherence tomography after coronary stent deployment. Circ J 74:1257–1259CrossRefPubMed
21.
Zurück zum Zitat Gonzalo N, Serruys PW, Okamura T, van Beusekom HM, Garcia-Garcia HM, van Soest G, van der Giessen W, Regar E (2009) Optical coherence tomography patterns of stent restenosis. Am Heart J 158(284–293):1 Gonzalo N, Serruys PW, Okamura T, van Beusekom HM, Garcia-Garcia HM, van Soest G, van der Giessen W, Regar E (2009) Optical coherence tomography patterns of stent restenosis. Am Heart J 158(284–293):1
Metadaten
Titel
Neointimal coverage of zotarolimus-eluting stent at 1, 2, and 3 months’ follow-up: an optical coherence tomography study
verfasst von
Takehiro Hashikata
Taiki Tojo
Sayaka Namba
Lisa Kitasato
Takuya Hashimoto
Ryo Kameda
Takao Shimohama
Minako Yamaoka-Tojo
Junya Ako
Publikationsdatum
01.02.2016
Verlag
Springer Japan
Erschienen in
Heart and Vessels / Ausgabe 2/2016
Print ISSN: 0910-8327
Elektronische ISSN: 1615-2573
DOI
https://doi.org/10.1007/s00380-014-0598-0

Weitere Artikel der Ausgabe 2/2016

Heart and Vessels 2/2016 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.