Factors affecting corneal hysteresis in normal eyes

verfasst von: Kazutaka Kamiya, Mana Hagishima, Fusako Fujimura, Kimiya Shimizu

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 10/2008

Einloggen, um Zugang zu erhalten

Abstract

Background

To evaluate factors affecting corneal hysteresis (CH) in normal eyes.

Methods

We examined 86 normal eyes of 43 healthy volunteers (age, 39.1 ± 14.5 years (mean ± standard deviation); range, 19 to 68 years; gender, 26 men, 60 women; manifest refraction, −2.25 ± 2.89 diopters (D); range, −9.13 to 3.88 D). We quantitatively assessed the value of CH using an Ocular Response Analyzer™ (Reichert Ophthalmic Instruments). We carried out this measurement three times, and the average value was used for statistical analysis. Multiple regression analysis was used to assess the relevant factors of the CH.

Results

The mean CH was 10.2 ± 1.3 mmHg. Explanatory variables relevant to the CH were, in order of magnitude of influence, the central corneal thickness (CCT) (partial regression coefficient B = 0.022, p < 0.0001), and the intraocular pressure (IOP) (B = -0.119, p = 0.04). No significant correlation was seen with other clinical factors such as age, gender, manifest refraction, or mean keratometric readings.

Conclusions

Eyes with thinner CCT and eyes with higher IOP are more predisposed to have lower CH. Refractive surgeons should, from a biomechanical viewpoint, take not only CCT but also IOP into consideration before performing keratorefractive surgery.
Literatur
1.
Zurück zum Zitat Roberts C (2002) Biomechanics of the cornea and wavefront-guided laser refractive surgery. J Refract Surg 18:S589–S592PubMed Roberts C (2002) Biomechanics of the cornea and wavefront-guided laser refractive surgery. J Refract Surg 18:S589–S592PubMed
2.
Zurück zum Zitat Kamiya K, Miyata K, Tokunaga T et al (2004) Structural analysis of the cornea using scanning-slit corneal topography in eyes undergoing excimer laser refractive surgery. Cornea 23:S59–S64PubMedCrossRef Kamiya K, Miyata K, Tokunaga T et al (2004) Structural analysis of the cornea using scanning-slit corneal topography in eyes undergoing excimer laser refractive surgery. Cornea 23:S59–S64PubMedCrossRef
3.
Zurück zum Zitat Jaycock PD, Lobo L, Ibrahim J et al (2005) Interferometric technique to measure biomechanical changes in the cornea induced by refractive surgery. J Cataract Refract Surg 31:175–184PubMedCrossRef Jaycock PD, Lobo L, Ibrahim J et al (2005) Interferometric technique to measure biomechanical changes in the cornea induced by refractive surgery. J Cataract Refract Surg 31:175–184PubMedCrossRef
4.
Zurück zum Zitat Deenadayalu C, Mobasher B, Rajan SD, Hall GW (2006) Refractive change induced by the LASIK flap in a biomechanical finite element model. J Refract Surg 22:286–292PubMed Deenadayalu C, Mobasher B, Rajan SD, Hall GW (2006) Refractive change induced by the LASIK flap in a biomechanical finite element model. J Refract Surg 22:286–292PubMed
5.
Zurück zum Zitat Dupps WJ Jr, Wilson SE (2006) Biomechanics and wound healing in the cornea. Exp Eye Res 83:709–720PubMedCrossRef Dupps WJ Jr, Wilson SE (2006) Biomechanics and wound healing in the cornea. Exp Eye Res 83:709–720PubMedCrossRef
6.
Zurück zum Zitat Orssengo GJ, Pye DC (1999) Determination of the true intraocular pressure and modulus of elasticity of the human cornea in vivo. Bull Math Biol 61:551–572PubMedCrossRef Orssengo GJ, Pye DC (1999) Determination of the true intraocular pressure and modulus of elasticity of the human cornea in vivo. Bull Math Biol 61:551–572PubMedCrossRef
7.
Zurück zum Zitat Liu J, Roberts CJ (2005) Influence of corneal biomechanical properties on intraocular pressure measurement; quantitative analysis. J Cataract Refract Surg 31:146–155PubMedCrossRef Liu J, Roberts CJ (2005) Influence of corneal biomechanical properties on intraocular pressure measurement; quantitative analysis. J Cataract Refract Surg 31:146–155PubMedCrossRef
8.
Zurück zum Zitat Herndon LW (2006) Measuring intraocular pressure-adjustments for corneal thickness and new technologies. Curr Opin Ophthalmol 17:115–119PubMedCrossRef Herndon LW (2006) Measuring intraocular pressure-adjustments for corneal thickness and new technologies. Curr Opin Ophthalmol 17:115–119PubMedCrossRef
9.
Zurück zum Zitat Bryant MR, McDonnell PJ (1996) Constitutive laws for biomechanical modeling of refractive surgery. J Biomech Eng 118:473–481PubMedCrossRef Bryant MR, McDonnell PJ (1996) Constitutive laws for biomechanical modeling of refractive surgery. J Biomech Eng 118:473–481PubMedCrossRef
10.
Zurück zum Zitat Luce DA (2005) Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg 31:156–162PubMedCrossRef Luce DA (2005) Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg 31:156–162PubMedCrossRef
11.
Zurück zum Zitat Lu F, Xu S, Qu J, Shen M, Wang X, Fang H, Wang J (2007) Central corneal thickness and corneal hysteresis during corneal swelling induced by contact lens wear with eye closure. Am J Ophthalmol 143:616–622PubMedCrossRef Lu F, Xu S, Qu J, Shen M, Wang X, Fang H, Wang J (2007) Central corneal thickness and corneal hysteresis during corneal swelling induced by contact lens wear with eye closure. Am J Ophthalmol 143:616–622PubMedCrossRef
12.
Zurück zum Zitat Lam A, Chen D, Chiu R, Chui WS (2007) Comparison of IOP measurements between ORA and GAT in normal Chinese. Optom Vis Sci 84:909–914PubMed Lam A, Chen D, Chiu R, Chui WS (2007) Comparison of IOP measurements between ORA and GAT in normal Chinese. Optom Vis Sci 84:909–914PubMed
13.
Zurück zum Zitat Broman AT, Congdon NG, Bandeen-Roche K, Quigley HA (2007) Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of intraocular pressure. J Glaucoma 16:581–588PubMedCrossRef Broman AT, Congdon NG, Bandeen-Roche K, Quigley HA (2007) Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of intraocular pressure. J Glaucoma 16:581–588PubMedCrossRef
14.
Zurück zum Zitat Shah S, Laiquzzaman M, Cunliffe I, Mantry S (2006) The use of the Reichert ocular response analyser to establish the relationship between ocular hysteresis, corneal resistance factor and central corneal thickness in normal eyes. Cont Lens Anterior Eye 29:257–262PubMedCrossRef Shah S, Laiquzzaman M, Cunliffe I, Mantry S (2006) The use of the Reichert ocular response analyser to establish the relationship between ocular hysteresis, corneal resistance factor and central corneal thickness in normal eyes. Cont Lens Anterior Eye 29:257–262PubMedCrossRef
15.
Zurück zum Zitat Ortiz D, Piñero D, Shabayek MH, Arnalich-Montiel F, Alió JL (2007) Corneal biomechanical properties in normal, post-laser in situ keratomileusis, and keratoconic eyes. J Cataract Refract Surg 33:1371–1375PubMedCrossRef Ortiz D, Piñero D, Shabayek MH, Arnalich-Montiel F, Alió JL (2007) Corneal biomechanical properties in normal, post-laser in situ keratomileusis, and keratoconic eyes. J Cataract Refract Surg 33:1371–1375PubMedCrossRef
16.
Zurück zum Zitat Pepose JS, Feigenbaum SK, Qazi MA, Sanderson JP, Roberts CJ (2007) Changes in corneal biomechanics and intraocular pressure following LASIK using static, dynamic, and noncontact tonometry. Am J Ophthalmol 143:39–47PubMedCrossRef Pepose JS, Feigenbaum SK, Qazi MA, Sanderson JP, Roberts CJ (2007) Changes in corneal biomechanics and intraocular pressure following LASIK using static, dynamic, and noncontact tonometry. Am J Ophthalmol 143:39–47PubMedCrossRef
17.
Zurück zum Zitat Shah S, Laiquzzaman M, Bhojwani R, Mantry S, Cunliffe I (2007) Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes. Invest Ophthalmol Vis Sci 48:3026–3031PubMedCrossRef Shah S, Laiquzzaman M, Bhojwani R, Mantry S, Cunliffe I (2007) Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes. Invest Ophthalmol Vis Sci 48:3026–3031PubMedCrossRef
18.
Zurück zum Zitat Kotecha A, Elsheikh A, Roberts CR, Zhu H, Garway-Heath DF (2006) Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci 47:5337–5347PubMedCrossRef Kotecha A, Elsheikh A, Roberts CR, Zhu H, Garway-Heath DF (2006) Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci 47:5337–5347PubMedCrossRef
19.
Zurück zum Zitat Toshino A, Uno T, Ohashi Y et al (2005) Transient keratectasia caused by intraocular pressure elevation after laser in situ keratomileusis. J Cataract Refract Surg 31:202–204PubMedCrossRef Toshino A, Uno T, Ohashi Y et al (2005) Transient keratectasia caused by intraocular pressure elevation after laser in situ keratomileusis. J Cataract Refract Surg 31:202–204PubMedCrossRef
20.
Zurück zum Zitat Hiatt JA, Wachler BS, Grant C (2005) Reversal of laser in situ keratomileusis-induced ectasia with intraocular pressure reduction. J Cataract Refract Surg 31:1652–1655PubMedCrossRef Hiatt JA, Wachler BS, Grant C (2005) Reversal of laser in situ keratomileusis-induced ectasia with intraocular pressure reduction. J Cataract Refract Surg 31:1652–1655PubMedCrossRef
21.
Zurück zum Zitat Kirwan C, O’Keefe M, Lanigan B (2006) Corneal hysteresis and intraocular pressure measurement in children using the reichert ocular response analyser. Am J Ophthalmol 142:990–992 Kirwan C, O’Keefe M, Lanigan B (2006) Corneal hysteresis and intraocular pressure measurement in children using the reichert ocular response analyser. Am J Ophthalmol 142:990–992
22.
Zurück zum Zitat Luce D (2006) Methodology for cornea compensated IOP and corneal resistance factor for Reichert ocular response analyzer. ARVO abstract 2266. Invest Ophthalmol Vis Sci Luce D (2006) Methodology for cornea compensated IOP and corneal resistance factor for Reichert ocular response analyzer. ARVO abstract 2266. Invest Ophthalmol Vis Sci
Metadaten
Titel
Factors affecting corneal hysteresis in normal eyes
verfasst von
Kazutaka Kamiya
Mana Hagishima
Fusako Fujimura
Kimiya Shimizu
Publikationsdatum
01.10.2008
Verlag
Springer-Verlag
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 10/2008
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-008-0864-x

Weitere Artikel der Ausgabe 10/2008

Graefe's Archive for Clinical and Experimental Ophthalmology 10/2008 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.