Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses

verfasst von: Alexandra Anton, Matthias Neuburger, Daniel Böhringer, Jens F. Jordan

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 7/2013

Einloggen, um Zugang zu erhalten

Abstract

Background

The aim of the present study was to compare the measurement of intraocular pressure (IOP) through a therapeutic soft contact lens with the “native” measurement. We additionally investigate whether a rebound tonometer (RT) or non-contact tonometer (NCT) is more suitable to measure IOP through a bandage contact lens.

Methods

The IOP was determined using each of the two methods, three times successively with (lens measurement) and without (native measurement) a soft contact lens. The Icare tonometer (Icare® TA01i, Icare Finland Oy, 23 subjects) and the Airpuff tonometer (Nidek NT 53OP, Nidek CO., LTD, Hiroishi Gamagori, Aichi, Japan, 16 subjects) were used. We compared the mean values (validity parameter) and standard deviation (precision parameter) of the three individual measurements in each case using the paired t-test. In addition, we conducted a power analysis to estimate the maximum error in the measurement caused by the contact lens (power level set to 0.8).

Results

With the Airpuff tonometer we detected no statistically significant between the lens and the native measurement (15.6 ± 2.6 vs. 15.3 ± 2.6 mmHg; p = 0.42). The power analysis revealed that the maximum error caused by the contact lens was 1.2 mmHg. The Icare tonometry, however, trended toward higher values in the contact lens measurements (17.5 ± 4.3 vs. 16.4 ± 3.5 mmHg in the native measurements; p = 0.05). Interestingly, this difference exhibited a statistically significant correlation with the corneal thickness (0.03 mmHg per μm corneal thickness; p = 0.04).

Conclusion

The use of NCT and RT for IOP measurement over a soft contact lens is feasible. The accuracy appears to be sufficient for the most common clinical applications.
Literatur
1.
Zurück zum Zitat Martinez-de-la-Casa JM, Jimenez-Santos M, Saenz-Frances F, Matilla-Rodero M, Mendez-Hernandez C, Herrero-Vanrell R, Garcia-Feijoo J (2011) Performance of the rebound, noncontact and goldmann applanation tonometers in routine clinical practice. Acta Ophthalmol 89:676–680PubMedCrossRef Martinez-de-la-Casa JM, Jimenez-Santos M, Saenz-Frances F, Matilla-Rodero M, Mendez-Hernandez C, Herrero-Vanrell R, Garcia-Feijoo J (2011) Performance of the rebound, noncontact and goldmann applanation tonometers in routine clinical practice. Acta Ophthalmol 89:676–680PubMedCrossRef
2.
Zurück zum Zitat Jablonski KS, Rosentreter A, Gaki S, Lappas A, Dietlein TS (2012) Clinical Use of a New Position-independent Rebound Tonometer. J. Glaucoma. PMID: 23172572 Jablonski KS, Rosentreter A, Gaki S, Lappas A, Dietlein TS (2012) Clinical Use of a New Position-independent Rebound Tonometer. J. Glaucoma. PMID: 23172572
3.
Zurück zum Zitat Abraham LM, Epasinghe NCR, Selva D, Casson R (2008) Comparison of the ICare rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists. Eye (Lond) 22:503–506CrossRef Abraham LM, Epasinghe NCR, Selva D, Casson R (2008) Comparison of the ICare rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists. Eye (Lond) 22:503–506CrossRef
4.
Zurück zum Zitat Marini M, Da Pozzo S, Accardo A, Canziani T (2011) Comparing applanation tonometry and rebound tonometry in glaucomatous and ocular hypertensive eyes. Eur J Ophthalmol 21:258–263PubMedCrossRef Marini M, Da Pozzo S, Accardo A, Canziani T (2011) Comparing applanation tonometry and rebound tonometry in glaucomatous and ocular hypertensive eyes. Eur J Ophthalmol 21:258–263PubMedCrossRef
5.
Zurück zum Zitat Gandhi NG, Prakalapakorn SG, El-Dairi MA, Jones SK, Freedman SF (2012) Icare ONE rebound versus Goldmann applanation tonometry in children with known or suspected glaucoma. Am J Ophthalmol 154:843–849.e1PubMedCrossRef Gandhi NG, Prakalapakorn SG, El-Dairi MA, Jones SK, Freedman SF (2012) Icare ONE rebound versus Goldmann applanation tonometry in children with known or suspected glaucoma. Am J Ophthalmol 154:843–849.e1PubMedCrossRef
6.
Zurück zum Zitat Halkiadakis I, Stratos A, Stergiopoulos G, Patsea E, Skouriotis S, Mitropoulos P, Papaconstantinou D, Georgopoulos G (2012) Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects. Graefes Arch Clin Exp Ophthalmol 250:1207–1211PubMedCrossRef Halkiadakis I, Stratos A, Stergiopoulos G, Patsea E, Skouriotis S, Mitropoulos P, Papaconstantinou D, Georgopoulos G (2012) Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects. Graefes Arch Clin Exp Ophthalmol 250:1207–1211PubMedCrossRef
7.
Zurück zum Zitat Flemmons MS, Hsiao Y-C, Dzau J, Asrani S, Jones S, Freedman SF (2011) Home tonometry for management of pediatric glaucoma. Am J Ophthalmol 152:470–478.e2PubMedCrossRef Flemmons MS, Hsiao Y-C, Dzau J, Asrani S, Jones S, Freedman SF (2011) Home tonometry for management of pediatric glaucoma. Am J Ophthalmol 152:470–478.e2PubMedCrossRef
8.
Zurück zum Zitat Blackmore SJ (2010) The use of contact lenses in the treatment of persistent epithelial defects. Cont Lens Anterior Eye 33:239–244PubMedCrossRef Blackmore SJ (2010) The use of contact lenses in the treatment of persistent epithelial defects. Cont Lens Anterior Eye 33:239–244PubMedCrossRef
9.
Zurück zum Zitat Sugimoto-Takeuchi R, Yamamoto R, Kuwayama Y, Kinoshita S (1991) Effect of intraocular pressure measurement through therapeutic soft contact lenses by noncontact tonometer. Nippon Ganka Gakkai Zasshi 95:869–872PubMed Sugimoto-Takeuchi R, Yamamoto R, Kuwayama Y, Kinoshita S (1991) Effect of intraocular pressure measurement through therapeutic soft contact lenses by noncontact tonometer. Nippon Ganka Gakkai Zasshi 95:869–872PubMed
10.
Zurück zum Zitat Firat PG, Cankaya C, Doganay S, Cavdar M, Duman S, Ozsoy E, Koc B (2012) The influence of soft contact lenses on the intraocular pressure measurement. Eye (Lond) 26:278–282CrossRef Firat PG, Cankaya C, Doganay S, Cavdar M, Duman S, Ozsoy E, Koc B (2012) The influence of soft contact lenses on the intraocular pressure measurement. Eye (Lond) 26:278–282CrossRef
11.
Zurück zum Zitat Liu Y-C, Huang J-Y, Wang I-J, Hu F-R, Hou Y-C (2011) Intraocular pressure measurement with the noncontact tonometer through soft contact lenses. J Glaucoma 20:179–182PubMedCrossRef Liu Y-C, Huang J-Y, Wang I-J, Hu F-R, Hou Y-C (2011) Intraocular pressure measurement with the noncontact tonometer through soft contact lenses. J Glaucoma 20:179–182PubMedCrossRef
12.
Zurück zum Zitat Patel S, Illahi W (2004) Non-contact tonometry over soft contact lenses: effect of contact lens power on the measurement of intra-ocular pressure. Cont Lens Anterior Eye 27:33–37PubMedCrossRef Patel S, Illahi W (2004) Non-contact tonometry over soft contact lenses: effect of contact lens power on the measurement of intra-ocular pressure. Cont Lens Anterior Eye 27:33–37PubMedCrossRef
13.
Zurück zum Zitat Patel S, Stevenson G (2009) Influence of lens material and intra-ocular pressure on the outcome of non-contact tonometry over soft contact lenses. Cont Lens Anterior Eye 32:68–72PubMedCrossRef Patel S, Stevenson G (2009) Influence of lens material and intra-ocular pressure on the outcome of non-contact tonometry over soft contact lenses. Cont Lens Anterior Eye 32:68–72PubMedCrossRef
14.
Zurück zum Zitat Zeri F, Calcatelli P, Donini B, Lupelli L, Zarrilli L, Swann PG (2011) The effect of hydrogel and silicone hydrogel contact lenses on the measurement of intraocular pressure with rebound tonometry. Cont Lens Anterior Eye 34:260–265PubMedCrossRef Zeri F, Calcatelli P, Donini B, Lupelli L, Zarrilli L, Swann PG (2011) The effect of hydrogel and silicone hydrogel contact lenses on the measurement of intraocular pressure with rebound tonometry. Cont Lens Anterior Eye 34:260–265PubMedCrossRef
15.
Zurück zum Zitat Khan JA, Graham CE (1991) Effect of contact lens removal or displacement on intraocular pressure. Arch Ophthalmol 109:825–828PubMedCrossRef Khan JA, Graham CE (1991) Effect of contact lens removal or displacement on intraocular pressure. Arch Ophthalmol 109:825–828PubMedCrossRef
16.
Zurück zum Zitat Rüfer F (2011) Sources of error in Goldmann applanation tonometry. Ophthalmologe 108:546–552PubMedCrossRef Rüfer F (2011) Sources of error in Goldmann applanation tonometry. Ophthalmologe 108:546–552PubMedCrossRef
17.
Zurück zum Zitat Neuburger M, Maier P, Böhringer D, Reinhard T, FJordan J (2012) The Impact of Corneal Edema on Intraocular Pressure Measurements Using Goldmann Applanation Tonometry, Tono-Pen XL, iCare, and ORA: An In Vitro Model. J glaucoma. PMID 22366704 Neuburger M, Maier P, Böhringer D, Reinhard T, FJordan J (2012) The Impact of Corneal Edema on Intraocular Pressure Measurements Using Goldmann Applanation Tonometry, Tono-Pen XL, iCare, and ORA: An In Vitro Model. J glaucoma. PMID 22366704
Metadaten
Titel
Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses
verfasst von
Alexandra Anton
Matthias Neuburger
Daniel Böhringer
Jens F. Jordan
Publikationsdatum
01.07.2013
Verlag
Springer-Verlag
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 7/2013
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-013-2329-0

Weitere Artikel der Ausgabe 7/2013

Graefe's Archive for Clinical and Experimental Ophthalmology 7/2013 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.