01.07.2013 | Glaucoma
Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses
verfasst von:
Alexandra Anton, Matthias Neuburger, Daniel Böhringer, Jens F. Jordan
Erschienen in:
Graefe's Archive for Clinical and Experimental Ophthalmology
|
Ausgabe 7/2013
Einloggen, um Zugang zu erhalten
Abstract
Background
The aim of the present study was to compare the measurement of intraocular pressure (IOP) through a therapeutic soft contact lens with the “native” measurement. We additionally investigate whether a rebound tonometer (RT) or non-contact tonometer (NCT) is more suitable to measure IOP through a bandage contact lens.
Methods
The IOP was determined using each of the two methods, three times successively with (lens measurement) and without (native measurement) a soft contact lens. The Icare tonometer (Icare® TA01i, Icare Finland Oy, 23 subjects) and the Airpuff tonometer (Nidek NT 53OP, Nidek CO., LTD, Hiroishi Gamagori, Aichi, Japan, 16 subjects) were used. We compared the mean values (validity parameter) and standard deviation (precision parameter) of the three individual measurements in each case using the paired t-test. In addition, we conducted a power analysis to estimate the maximum error in the measurement caused by the contact lens (power level set to 0.8).
Results
With the Airpuff tonometer we detected no statistically significant between the lens and the native measurement (15.6 ± 2.6 vs. 15.3 ± 2.6 mmHg; p = 0.42). The power analysis revealed that the maximum error caused by the contact lens was 1.2 mmHg. The Icare tonometry, however, trended toward higher values in the contact lens measurements (17.5 ± 4.3 vs. 16.4 ± 3.5 mmHg in the native measurements; p = 0.05). Interestingly, this difference exhibited a statistically significant correlation with the corneal thickness (0.03 mmHg per μm corneal thickness; p = 0.04).
Conclusion
The use of NCT and RT for IOP measurement over a soft contact lens is feasible. The accuracy appears to be sufficient for the most common clinical applications.