Skip to main content
Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology 4/2018

01.02.2018 | Glaucoma

The STARflo glaucoma implant: preliminary 12 months results

verfasst von: Sofia Fili, Peter Wölfelschneider, Markus Kohlhaas

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 4/2018

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The STARflo glaucoma implant is a drainage-free system designed for the operative treatment of refractory open-angle glaucoma. The purpose of this study is to investigate the safety and efficacy of the STARflo implant. The results of 12 months follow-up are presented in patients with open-angle glaucoma, PEX–glaucoma, and congenital glaucoma.

Methods

A prospective study with a total of over 29 patients (36 eyes), in which the suprachoroidal implantation of the STARflo was indicated for long-term intraocular pressure reduction, is still in process and evaluation.

Results

The reduction of intraocular pressure was satisfying after 12 months with an average intraocular pressure of 15.0 (± 2.5) mmHg in comparison to 21.08 (± 7.29) mmHg preoperatively. There was a significant 66.7% reduction in local glaucoma drops 12 months postoperatively (p < 0.001). Twenty-five percent of the eyes needed further surgery to regulate the intraocular pressure. No intraoperative complications have been reported after the implantation of STARflo. Postoperative complications include hyphema and fibrin formation in the anterior chamber, which were completely resorbed after the first postoperative month. A postoperative intraocular pressure (IOP) elevation was the most common complication after the first 3 months.

Conclusions

The STARflo implant appears to be a safe and effective alternative treatment method for patients with refractory open-angle glaucoma compared to conventional glaucoma surgery.
Literatur
1.
Zurück zum Zitat Pourjavan S, Collignon NJM (2014) STARflo. In: Samples, John R, Ahmed, Iqbal Ike K (eds) Surgical Innovations in Glaucoma. Springer Science+Business Media, New York, pp 235–251 Pourjavan S, Collignon NJM (2014) STARflo. In: Samples, John R, Ahmed, Iqbal Ike K (eds) Surgical Innovations in Glaucoma. Springer Science+Business Media, New York, pp 235–251
4.
Zurück zum Zitat Cseke I, Vámosi P, Bausz M (2016) STARflo glaucoma implant: early experience in Hungary. Rom J Ophthalmol 60:14–17PubMed Cseke I, Vámosi P, Bausz M (2016) STARflo glaucoma implant: early experience in Hungary. Rom J Ophthalmol 60:14–17PubMed
6.
Zurück zum Zitat Pourjavan S, Collignon N, De Groot V (2013) STARflo glaucoma implant 12 month clinical results. Acta Ophthalmol 91:252 Pourjavan S, Collignon N, De Groot V (2013) STARflo glaucoma implant 12 month clinical results. Acta Ophthalmol 91:252
7.
Zurück zum Zitat Garcia-Feijoo J, Rau M, Ahmed I, Antonio A, Grabner G, Ianchulev T (2012) Safety and efficacy of CyPass micro-stent as a stand-alone treatment for open angle glaucoma: worldwide clinical experience. European Society of Cataract and Refractive Surgery Annual Meeting Sep 8-12, 2012; Milan, Italy Garcia-Feijoo J, Rau M, Ahmed I, Antonio A, Grabner G, Ianchulev T (2012) Safety and efficacy of CyPass micro-stent as a stand-alone treatment for open angle glaucoma: worldwide clinical experience. European Society of Cataract and Refractive Surgery Annual Meeting Sep 8-12, 2012; Milan, Italy
8.
Zurück zum Zitat Ichhpujani P, Moster MR (2014) SOLX suprachoroidal shunt. In: Samples JR, Ahmed II (eds) Surgical innovations in glaucoma. Springer Science+Business Media, New York, pp 253–256CrossRef Ichhpujani P, Moster MR (2014) SOLX suprachoroidal shunt. In: Samples JR, Ahmed II (eds) Surgical innovations in glaucoma. Springer Science+Business Media, New York, pp 253–256CrossRef
9.
Zurück zum Zitat Melamed S, Ben Simon GJ, Goldenfeld M, Simon G (2009) Efficacy and safety of gold micro shunt implantation to the supraciliary space in patients with glaucoma: a pilot study. Arch Ophthalmol 127:264–269CrossRefPubMed Melamed S, Ben Simon GJ, Goldenfeld M, Simon G (2009) Efficacy and safety of gold micro shunt implantation to the supraciliary space in patients with glaucoma: a pilot study. Arch Ophthalmol 127:264–269CrossRefPubMed
10.
Zurück zum Zitat Gedde SJ, Schiffman JC, Feuer WJ, Herndon LW, Brandt JD, Budenz DL, Tube Versus Trabeculectomy Study Group (2012) Treatment outcomes in the tube versus trabeculectomy (TVT) study after five years of follow-up. Am J Ophthalmol 153:789–803CrossRefPubMedPubMedCentral Gedde SJ, Schiffman JC, Feuer WJ, Herndon LW, Brandt JD, Budenz DL, Tube Versus Trabeculectomy Study Group (2012) Treatment outcomes in the tube versus trabeculectomy (TVT) study after five years of follow-up. Am J Ophthalmol 153:789–803CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Christakis PG, Tsai JC, Kalenak JW, Zurakowski D, Cantor LB, Kammer JA, Ahmed II (2013) The Ahmed versus Baerveldt study: three-year treatment outcomes. Ophthalmology 120:2232–2240CrossRefPubMed Christakis PG, Tsai JC, Kalenak JW, Zurakowski D, Cantor LB, Kammer JA, Ahmed II (2013) The Ahmed versus Baerveldt study: three-year treatment outcomes. Ophthalmology 120:2232–2240CrossRefPubMed
Metadaten
Titel
The STARflo glaucoma implant: preliminary 12 months results
verfasst von
Sofia Fili
Peter Wölfelschneider
Markus Kohlhaas
Publikationsdatum
01.02.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 4/2018
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-018-3916-x

Weitere Artikel der Ausgabe 4/2018