Mesh fixation with fibrin glue (Tissucol/Tisseel®) in hernia repair dependent on the mesh structure—is there an optimum fibrin–mesh combination?—Investigations on a biomechanical model

verfasst von: Christine Schug-Pass, Hans Lippert, Ferdinand Köckerling

Erschienen in: Langenbeck's Archives of Surgery | Ausgabe 5/2010

Einloggen, um Zugang zu erhalten

Abstract

Background

Because of its hemostatic and adhesive properties, fibrin glue has been used in many areas of surgical treatment in recent years. One example is hernia repair, where fibrin gluing has become increasingly established as an alternative method for mesh fixation. Clinically, fixation with fibrin glue shows a reduced postoperative complication rate compared to other fixation methods (staples, sutures), particularly with regard to pain.

Materials and methods

Six different lightweight meshes were tested: TiMesh light, TiMesh extralight, Parietene light, Ultrapro, Optilene LP, and BARD Soft Mesh. Two millimeters Tissucol was used for fixation. Five meshes from each group were tested on muscular tissue with and without fibrin glue. The defined defect was 4.5 cm in diameter. The biomechanical measurements were taken in a standardized way using a materials testing machine. The minimum fixation strength required was 32 N, calculated from a corresponding model.

Results

The fixation strength measurements without fibrin glue gave a mean value for all 30 meshes of 2.98 N with a SD of 0.92 N. This was far below the 32 N required. With fibrin glue, the mean of all the measurements (30 meshes) was 61.86 ± 23.0 N (min 34.9 N, max 97.3 N). The lowest value was recorded for Ultrapro (34.9 ± 12.5 N). All the other meshes had a significantly higher fixation strength when fixed with fibrin glue than Ultrapro (p = 0.001). The best results were found for Optilene LP, and this was significantly better than all the other meshes (97.3 ± 8.9 N; p < 0.001).

Conclusion

Given the adequate stability and superior biocompatibility of lightweight large pore monofilament polypropylene meshes, heavyweight polypropylene meshes should no longer be used. It is possible to achieve adequate fixation of the meshes using fibrin glue. However, careful consideration should be given to the particular structure of the mesh in each case. Not every mesh is equally suitable for this purpose.
Literatur
2.
3.
Zurück zum Zitat Katkhouda N, Mavor E, Friedlander MH, Mason RJ, Kiyabu M, Grant SW, Achanta K, Kirkman EL, Narayanan K, Essani R (2001) Use of fibrin sealant for prosthetic mesh fixation in laparoscopic extraperitoneal inguinal hernia repair. Ann Surg 233:18–25CrossRefPubMed Katkhouda N, Mavor E, Friedlander MH, Mason RJ, Kiyabu M, Grant SW, Achanta K, Kirkman EL, Narayanan K, Essani R (2001) Use of fibrin sealant for prosthetic mesh fixation in laparoscopic extraperitoneal inguinal hernia repair. Ann Surg 233:18–25CrossRefPubMed
4.
Zurück zum Zitat Katkhouda N (2004) A new technique for laparoscopic hernia repair using fibrin sealant. Surg Technol Int 12:120–126PubMed Katkhouda N (2004) A new technique for laparoscopic hernia repair using fibrin sealant. Surg Technol Int 12:120–126PubMed
5.
Zurück zum Zitat Olmi S, Addis A, Domeneghini C, Scaini A, Croce E (2007) Experimental comparison of type of tissucol dilution and composite mesh (Parietex) for laparoscopic repair of groin and abdominal hernia: observational study conducted in a university laboratory. Hernia 11:211–215CrossRefPubMed Olmi S, Addis A, Domeneghini C, Scaini A, Croce E (2007) Experimental comparison of type of tissucol dilution and composite mesh (Parietex) for laparoscopic repair of groin and abdominal hernia: observational study conducted in a university laboratory. Hernia 11:211–215CrossRefPubMed
6.
Zurück zum Zitat ten Hallers EJ, Jansen JA, Marres HA, Rakhorst G, Verkerke GJ (2007) Histological assessment of titanium and polypropylene fiber mesh implantation with and without fibrin tissue glue. J Biomed Mater Res A 80:372–380PubMed ten Hallers EJ, Jansen JA, Marres HA, Rakhorst G, Verkerke GJ (2007) Histological assessment of titanium and polypropylene fiber mesh implantation with and without fibrin tissue glue. J Biomed Mater Res A 80:372–380PubMed
7.
Zurück zum Zitat Topart P, Vandenbroucke F, Lozach P (2005) Tisseel versus tack staples as mesh fixation in totally extraperitoneal laparoscopic repair of groin hernias: a retrospective analysis. Surg Endosc 19:724–727CrossRefPubMed Topart P, Vandenbroucke F, Lozach P (2005) Tisseel versus tack staples as mesh fixation in totally extraperitoneal laparoscopic repair of groin hernias: a retrospective analysis. Surg Endosc 19:724–727CrossRefPubMed
8.
Zurück zum Zitat Schwab R, Willms A, Kröger A, Becker HP (2006) Less chronic pain following mesh fixation using a fibrin sealant in TEP inguinal hernia repair. Hernia 10:272–277CrossRefPubMed Schwab R, Willms A, Kröger A, Becker HP (2006) Less chronic pain following mesh fixation using a fibrin sealant in TEP inguinal hernia repair. Hernia 10:272–277CrossRefPubMed
9.
Zurück zum Zitat Lovisetto F, Zonta S, Rota E, Mazzilli M, Bardone M, Bottero L, Faillance G, Longoni M (2007) Use of human fibrin glue (Tissucol) versus staples for mesh fixation in laparoscopic transabdominal preperitoneal hernioplasty. Ann Surg 245:222–231CrossRefPubMed Lovisetto F, Zonta S, Rota E, Mazzilli M, Bardone M, Bottero L, Faillance G, Longoni M (2007) Use of human fibrin glue (Tissucol) versus staples for mesh fixation in laparoscopic transabdominal preperitoneal hernioplasty. Ann Surg 245:222–231CrossRefPubMed
10.
Zurück zum Zitat Olmi S, Scaini A, Erba L, Guaglio M, Croce E (2007) Quantification of pain in laparoscopic transabdominal preperitoneal (TAPP) inguinal hernioplasty identifies marked differences between prosthesis fixation systems. Surgery 142:40–46CrossRefPubMed Olmi S, Scaini A, Erba L, Guaglio M, Croce E (2007) Quantification of pain in laparoscopic transabdominal preperitoneal (TAPP) inguinal hernioplasty identifies marked differences between prosthesis fixation systems. Surgery 142:40–46CrossRefPubMed
11.
Zurück zum Zitat Lau H (2005) Fibrin sealant versus mechanical stapling for mesh fixation during endoscopic extraperitoneal inguinal hernioplasty. Ann Surg 242:670–675CrossRefPubMed Lau H (2005) Fibrin sealant versus mechanical stapling for mesh fixation during endoscopic extraperitoneal inguinal hernioplasty. Ann Surg 242:670–675CrossRefPubMed
12.
Zurück zum Zitat Klinge U, Klosterhalfen B, Birkenhauer V, Junge K, Conze J, Schumpelick V (2002) Impact of polymer pore size on the interface scar formation in a rat model. J Surg Res 103:208–214CrossRefPubMed Klinge U, Klosterhalfen B, Birkenhauer V, Junge K, Conze J, Schumpelick V (2002) Impact of polymer pore size on the interface scar formation in a rat model. J Surg Res 103:208–214CrossRefPubMed
13.
Zurück zum Zitat Scheidbach H, Tamme C, Tannapfel A, Lippert H, Köckerling F (2004) In vivo studies comparing the biocompatibility of various polypropylene meshes and their handling properties during endoscopic total extraperitoneal (TEP) patchplasty: an experimental study in pigs. Surg Endosc 18:211–220CrossRefPubMed Scheidbach H, Tamme C, Tannapfel A, Lippert H, Köckerling F (2004) In vivo studies comparing the biocompatibility of various polypropylene meshes and their handling properties during endoscopic total extraperitoneal (TEP) patchplasty: an experimental study in pigs. Surg Endosc 18:211–220CrossRefPubMed
14.
Zurück zum Zitat Weyhe D, Belyaev O, Müller C, Meurer K, Bauer KH, Papapostolou G, Uhl W (2007) Improving outcomes in hernia repair by the use of light meshes—a comparison of different implant constructions based on a critical appraisal of the literature. World J Surg 31:234–244CrossRefPubMed Weyhe D, Belyaev O, Müller C, Meurer K, Bauer KH, Papapostolou G, Uhl W (2007) Improving outcomes in hernia repair by the use of light meshes—a comparison of different implant constructions based on a critical appraisal of the literature. World J Surg 31:234–244CrossRefPubMed
15.
Zurück zum Zitat Schug-Paß C, Tamme C, Sommerer F, Tannapfel A, Lippert H, Köckerling F (2008) A lightweight, partially absorbable mesh (Ultrapro) for endoscopic hernia repair: experimental biocompatibility results obtained in a porcine model. Surg Endosc 22:1100–1106CrossRefPubMed Schug-Paß C, Tamme C, Sommerer F, Tannapfel A, Lippert H, Köckerling F (2008) A lightweight, partially absorbable mesh (Ultrapro) for endoscopic hernia repair: experimental biocompatibility results obtained in a porcine model. Surg Endosc 22:1100–1106CrossRefPubMed
16.
Zurück zum Zitat Nienhujs S, Staal E, Strobbe L, Rosman C, Groenewoud H, Bleichrodt R (2007) Chronic pain after mesh repair of inguinal hernia: a systematic review. Am J Surg 194:394–400CrossRef Nienhujs S, Staal E, Strobbe L, Rosman C, Groenewoud H, Bleichrodt R (2007) Chronic pain after mesh repair of inguinal hernia: a systematic review. Am J Surg 194:394–400CrossRef
17.
Zurück zum Zitat Tamme C, Garde N, Klingler A, Hampe C, Wunder R, Köckerling F (2005) Totally extraperitoneal inguinal hernioplasty with titanium—coated lightweight polypropylene mesh: early results. Surg Endosc 19:1125–1129CrossRefPubMed Tamme C, Garde N, Klingler A, Hampe C, Wunder R, Köckerling F (2005) Totally extraperitoneal inguinal hernioplasty with titanium—coated lightweight polypropylene mesh: early results. Surg Endosc 19:1125–1129CrossRefPubMed
18.
Zurück zum Zitat Klinge U, Conze J, Limberg W, Brücker C, Ottinger AP, Schumpelick V (1996) Pathophysiology of the abdominal wall. Chirurg 67:229–233PubMed Klinge U, Conze J, Limberg W, Brücker C, Ottinger AP, Schumpelick V (1996) Pathophysiology of the abdominal wall. Chirurg 67:229–233PubMed
19.
Zurück zum Zitat Koch CA, Greenlee SM, Larson DR, Harrington JR, Farley DR (2006) Randomized prospective study of totally extraperitoneal inguinal hernia repair: fixation versus no fixation of mesh. ISLS 10:457–460 Koch CA, Greenlee SM, Larson DR, Harrington JR, Farley DR (2006) Randomized prospective study of totally extraperitoneal inguinal hernia repair: fixation versus no fixation of mesh. ISLS 10:457–460
20.
Zurück zum Zitat Scheidbach H, Tannapfel A, Schmidt U, Lippert H, Köckerling F (2004) Influence of titanium coating on the biocompatibility of heavyweight polypropylene mesh. An animal experimental model. Eur Surg Res 36:313–317 Scheidbach H, Tannapfel A, Schmidt U, Lippert H, Köckerling F (2004) Influence of titanium coating on the biocompatibility of heavyweight polypropylene mesh. An animal experimental model. Eur Surg Res 36:313–317
21.
Zurück zum Zitat Schwab R, Schumacher O, Junge K, Binnebösel M, Klinge U, Becker HP, Schumpelick V (2008) Biomechanical analysis of mesh fixation in TAPP and TEP hernia repair. Surg Endosc 22:731–738CrossRefPubMed Schwab R, Schumacher O, Junge K, Binnebösel M, Klinge U, Becker HP, Schumpelick V (2008) Biomechanical analysis of mesh fixation in TAPP and TEP hernia repair. Surg Endosc 22:731–738CrossRefPubMed
22.
Zurück zum Zitat Binnebösel M, Rosch R, Junge K, Flanagan TC, Schwab R, Schumpelick V, Klinge U (2007) Biomechanical analyses of overlap and mesh dislocation in an incisional hernia model in vitro. Surgery 142:365–371CrossRefPubMed Binnebösel M, Rosch R, Junge K, Flanagan TC, Schwab R, Schumpelick V, Klinge U (2007) Biomechanical analyses of overlap and mesh dislocation in an incisional hernia model in vitro. Surgery 142:365–371CrossRefPubMed
23.
Zurück zum Zitat Knook MTT, van Rosmalen AC, Yoder BE, Kleinrensink GJ, Snijders CJ, Looman CWN, van Steesel CJ (2001) Optimal mesh size for endoscopic inguinal hernia repair. Surg Endosc 15:1471–1477CrossRefPubMed Knook MTT, van Rosmalen AC, Yoder BE, Kleinrensink GJ, Snijders CJ, Looman CWN, van Steesel CJ (2001) Optimal mesh size for endoscopic inguinal hernia repair. Surg Endosc 15:1471–1477CrossRefPubMed
24.
Zurück zum Zitat Petter-Puchner AH, Fortelny R, Mittermayr R, Öhlinger W, Redl H (2005) Fibrin sealing versus stapling of hernia meshes in an onlay model in the rat. Hernia 9:322–329CrossRefPubMed Petter-Puchner AH, Fortelny R, Mittermayr R, Öhlinger W, Redl H (2005) Fibrin sealing versus stapling of hernia meshes in an onlay model in the rat. Hernia 9:322–329CrossRefPubMed
Metadaten
Titel
Mesh fixation with fibrin glue (Tissucol/Tisseel®) in hernia repair dependent on the mesh structure—is there an optimum fibrin–mesh combination?—Investigations on a biomechanical model
verfasst von
Christine Schug-Pass
Hans Lippert
Ferdinand Köckerling
Publikationsdatum
01.06.2010
Verlag
Springer-Verlag
Erschienen in
Langenbeck's Archives of Surgery / Ausgabe 5/2010
Print ISSN: 1435-2443
Elektronische ISSN: 1435-2451
DOI
https://doi.org/10.1007/s00423-009-0466-z

Weitere Artikel der Ausgabe 5/2010

Langenbeck's Archives of Surgery 5/2010 Zur Ausgabe

Wie erfolgreich ist eine Re-Ablation nach Rezidiv?

23.04.2024 Ablationstherapie Nachrichten

Nach der Katheterablation von Vorhofflimmern kommt es bei etwa einem Drittel der Patienten zu Rezidiven, meist binnen eines Jahres. Wie sich spätere Rückfälle auf die Erfolgschancen einer erneuten Ablation auswirken, haben Schweizer Kardiologen erforscht.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Ureterstriktur: Innovative OP-Technik bewährt sich

19.04.2024 EAU 2024 Kongressbericht

Die Ureterstriktur ist eine relativ seltene Komplikation, trotzdem bedarf sie einer differenzierten Versorgung. In komplexen Fällen wird dies durch die roboterassistierte OP-Technik gewährleistet. Erste Resultate ermutigen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.