Erschienen in: European Journal of Pediatrics 7/2010

01.07.2010 | Original Paper

Use of self-heating gel mattresses eliminates admission hypothermia in infants born below 28 weeks gestation

verfasst von: C. P. Hafis Ibrahim, C. W. Yoxall

Erschienen in: European Journal of Pediatrics | Ausgabe 7/2010

Einloggen, um Zugang zu erhalten

Abstract

Introduction

Hypothermia at birth is strongly associated with mortality and morbidity in pre-term infants.

Background

A local audit showed limited effectiveness of occlusive wrapping in preventing admission hypothermia in very pre-term infants. Self-heating acetate gel mattresses were introduced as a result to prevent hypothermia at birth in infants born at or below 28 weeks gestation.

Methods

A retrospective audit was conducted to evaluate the effectiveness of self-heating acetate gel mattresses at resuscitation of infants born at or below 28 weeks to prevent hypothermia at birth. All infants born at or below 28 weeks gestation during 18 months before and 18 months after self-heating acetate gel mattresses were introduced during resuscitation were included.

Results

One hundred five babies were born when acetate gel mattresses were not used, and 124 were born during the period when they were. Four (3.3%) babies were hypothermic (temperature <36°C) at admission when the mattresses were used compared to 21 (22.6%) babies who were hypothermic during the period it was not (p < 0.001). Hyperthermia (temperature >37°C) rose from 30.1% prior to use of gel mattresses to 49.6% when they were used (p = 0.004).

Conclusions

Self-heating acetate gel mattresses are highly effective in reducing admission hypothermia in infants born at or below 28 weeks gestation. The use of these mattresses is associated with a significant increase in hyperthermia.
Literatur
1.
Zurück zum Zitat Bergman NJ, Linley LL, Fawcus SR (2004) Randomized controlled trial of skin-to-skin contact from birth versus conventional incubator for physiological stabilization in 1200- to 2199-gram newborns. Acta Paediatr 93:779–785CrossRefPubMed Bergman NJ, Linley LL, Fawcus SR (2004) Randomized controlled trial of skin-to-skin contact from birth versus conventional incubator for physiological stabilization in 1200- to 2199-gram newborns. Acta Paediatr 93:779–785CrossRefPubMed
2.
Zurück zum Zitat Carmichael A, McCullough S, Kempley ST (2007) Critical dependence of acetate thermal mattress on gel activation temperature. Arch Dis Child Fetal Neonatal Ed 92:F44–F45CrossRefPubMed Carmichael A, McCullough S, Kempley ST (2007) Critical dependence of acetate thermal mattress on gel activation temperature. Arch Dis Child Fetal Neonatal Ed 92:F44–F45CrossRefPubMed
3.
Zurück zum Zitat Chaput de Saintonge DM, Cross KW, Shathorn MK, Lewis SR et al (1979) Hats for the newborn infant. Br Med J 2:570–571CrossRefPubMed Chaput de Saintonge DM, Cross KW, Shathorn MK, Lewis SR et al (1979) Hats for the newborn infant. Br Med J 2:570–571CrossRefPubMed
4.
Zurück zum Zitat Confidential Enquiry into Stillbirths and Deaths in Infancy (2003) PROJECT 27/28, an enquiry into quality of care and its effect on the survival of babies born at 27–28 weeks. Maternal and Child Health Research Consortium, London Confidential Enquiry into Stillbirths and Deaths in Infancy (2003) PROJECT 27/28, an enquiry into quality of care and its effect on the survival of babies born at 27–28 weeks. Maternal and Child Health Research Consortium, London
5.
Zurück zum Zitat Costeloe K, Hennessy E, Gibson AT, Marlow N et al (2000) The EPICure study: outcomes to discharge from hospital for infants born at the threshold of viability. Pediatrics 106:659–671CrossRefPubMed Costeloe K, Hennessy E, Gibson AT, Marlow N et al (2000) The EPICure study: outcomes to discharge from hospital for infants born at the threshold of viability. Pediatrics 106:659–671CrossRefPubMed
6.
Zurück zum Zitat Craig JV, Lancaster GA, Williamson PR, Smyth RL (2000) Temperature measured at the axilla compared with rectum in children and young people: systematic review. Br Med J 320:1174–1178CrossRef Craig JV, Lancaster GA, Williamson PR, Smyth RL (2000) Temperature measured at the axilla compared with rectum in children and young people: systematic review. Br Med J 320:1174–1178CrossRef
7.
Zurück zum Zitat Falzon A, Grech V, Caruana B, Magro A et al (2003) How reliable is axillary temperature measurement? Acta Paediatr 92:309–313CrossRefPubMed Falzon A, Grech V, Caruana B, Magro A et al (2003) How reliable is axillary temperature measurement? Acta Paediatr 92:309–313CrossRefPubMed
8.
Zurück zum Zitat Ibrahim CPH, Yoxall CW (2009) Use of plastic bags to prevent hypothermia at birth in preterm infants—do they work at lower gestations? Acta Paedtr 98:256–260CrossRef Ibrahim CPH, Yoxall CW (2009) Use of plastic bags to prevent hypothermia at birth in preterm infants—do they work at lower gestations? Acta Paedtr 98:256–260CrossRef
9.
Zurück zum Zitat Knobel RB, Wimmer JE Jr, Holbert D (2005) Heat loss prevention for preterm infants in the delivery room. J Perinatol 25:304–308CrossRefPubMed Knobel RB, Wimmer JE Jr, Holbert D (2005) Heat loss prevention for preterm infants in the delivery room. J Perinatol 25:304–308CrossRefPubMed
10.
Zurück zum Zitat L'Herault J, Petroff L, Jeffrey J (2001) The effectiveness of a thermal mattress in stabilizing and maintaining body temperature during the transport of very low-birth weight newborns. Appl Nurs Res 14:210–219CrossRefPubMed L'Herault J, Petroff L, Jeffrey J (2001) The effectiveness of a thermal mattress in stabilizing and maintaining body temperature during the transport of very low-birth weight newborns. Appl Nurs Res 14:210–219CrossRefPubMed
11.
Zurück zum Zitat Lofgren O, Jacobson L (1979) The influence of different electrode temperatures on recorded transcutaneous pO2 level. Pediatrics 64:892–897PubMed Lofgren O, Jacobson L (1979) The influence of different electrode temperatures on recorded transcutaneous pO2 level. Pediatrics 64:892–897PubMed
12.
Zurück zum Zitat Mathew B, Lakshminrusimha S, Cominsky K, Schroder E et al (2007) Vinyl bags prevent hypothermia at birth in preterm infants. Indian J Pediatr 74:249–253CrossRefPubMed Mathew B, Lakshminrusimha S, Cominsky K, Schroder E et al (2007) Vinyl bags prevent hypothermia at birth in preterm infants. Indian J Pediatr 74:249–253CrossRefPubMed
13.
Zurück zum Zitat McCall EM, Alderdice FA, Halliday HL, Jenkins JG et al (2008) Interventions to prevent hypothermia at birth in preterm and/or low birthweight infants. Cochrane Database of Systematic Reviews 2008, Issue 1. Art. No.: CD004210. doi:10.1002/14651858.CD004210.pub3 McCall EM, Alderdice FA, Halliday HL, Jenkins JG et al (2008) Interventions to prevent hypothermia at birth in preterm and/or low birthweight infants. Cochrane Database of Systematic Reviews 2008, Issue 1. Art. No.: CD004210. doi:10.​1002/​14651858.​CD004210.​pub3
14.
Zurück zum Zitat Vohra S, Frent G, Campbell V, Abbott M et al (1999) Effect of polyethylene occlusive skin wrapping on heat loss in very low birth weight infants at delivery: a randomized trial. J Pediatr 134:547–551CrossRefPubMed Vohra S, Frent G, Campbell V, Abbott M et al (1999) Effect of polyethylene occlusive skin wrapping on heat loss in very low birth weight infants at delivery: a randomized trial. J Pediatr 134:547–551CrossRefPubMed
15.
Zurück zum Zitat Vohra S, Roberts RS, Zhang B, Janes M et al (2004) Heat Loss Prevention (HeLP) in the delivery room: a randomized controlled trial of polyethylene occlusive skin wrapping in very preterm infants. J Pediatr 145:750–753CrossRefPubMed Vohra S, Roberts RS, Zhang B, Janes M et al (2004) Heat Loss Prevention (HeLP) in the delivery room: a randomized controlled trial of polyethylene occlusive skin wrapping in very preterm infants. J Pediatr 145:750–753CrossRefPubMed
16.
Zurück zum Zitat Wang W, Dow KE, Flavin MP (2008) Hyperthermia amplifies brain cytokine and reactive oxygen species response in a model of perinatal inflammation. Neurosci Lett 445:233–235CrossRefPubMed Wang W, Dow KE, Flavin MP (2008) Hyperthermia amplifies brain cytokine and reactive oxygen species response in a model of perinatal inflammation. Neurosci Lett 445:233–235CrossRefPubMed
17.
Zurück zum Zitat Watkinson M (2006) Temperature control of premature infants in the delivery room. Clin Perinatol 33:43–53CrossRefPubMed Watkinson M (2006) Temperature control of premature infants in the delivery room. Clin Perinatol 33:43–53CrossRefPubMed
Metadaten
Titel
Use of self-heating gel mattresses eliminates admission hypothermia in infants born below 28 weeks gestation
verfasst von
C. P. Hafis Ibrahim
C. W. Yoxall
Publikationsdatum
01.07.2010
Verlag
Springer-Verlag
Erschienen in
European Journal of Pediatrics / Ausgabe 7/2010
Print ISSN: 0340-6199
Elektronische ISSN: 1432-1076
DOI
https://doi.org/10.1007/s00431-009-1113-y

Weitere Artikel der Ausgabe 7/2010

European Journal of Pediatrics 7/2010 Zur Ausgabe

Neuer Typ-1-Diabetes bei Kindern am Wochenende eher übersehen

23.04.2024 Typ-1-Diabetes Nachrichten

Wenn Kinder an Werktagen zum Arzt gehen, werden neu auftretender Typ-1-Diabetes und diabetische Ketoazidosen häufiger erkannt als bei Arztbesuchen an Wochenenden oder Feiertagen.

Neue Studienergebnisse zur Myopiekontrolle mit Atropin

22.04.2024 Fehlsichtigkeit Nachrichten

Augentropfen mit niedrig dosiertem Atropin können helfen, das Fortschreiten einer Kurzsichtigkeit bei Kindern zumindest zu verlangsamen, wie die Ergebnisse einer aktuellen Studie mit verschiedenen Dosierungen zeigen.

Spinale Muskelatrophie: Neugeborenen-Screening lohnt sich

18.04.2024 Spinale Muskelatrophien Nachrichten

Seit 2021 ist die Untersuchung auf spinale Muskelatrophie Teil des Neugeborenen-Screenings in Deutschland. Eine Studie liefert weitere Evidenz für den Nutzen der Maßnahme.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Pädiatrie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.