Skip to main content
Erschienen in: Surgical Endoscopy 11/2012

01.11.2012

Radiofrequency energy antenna coupling to common laparoscopic instruments: practical implications

verfasst von: Edward L. Jones, Thomas N. Robinson, Jennifer R. McHenry, Christina L. Dunn, Paul N. Montero, Henry R. Govekar, Greg V. Stiegmann

Erschienen in: Surgical Endoscopy | Ausgabe 11/2012

Einloggen, um Zugang zu erhalten

Abstract

Background

Electromagnetic coupling can occur between the monopolar “Bovie” instrument and other laparoscopic instruments without direct contact by a phenomenon termed antenna coupling. The purpose of this study was to determine if, and to what extent, radiofrequency energy couples to other common laparoscopic instruments and to describe practical steps that can minimize the magnitude of antenna coupling.

Methods

In a laparoscopic simulator, monopolar radiofrequency energy was delivered to an L-hook. The tips of standard, nonelectrical laparoscopic instruments (either an unlit 10 mm telescope or a 5 mm grasper) were placed adjacent to bovine liver tissue and were never in contact with the active electrode. Thermal imaging quantified the change in tissue temperature nearest the tip of the telescope or grasper at the end of a 5 s activation of the active electrode.

Results

A 5 s activation (30 watts, coagulation mode, 4 cm separation between instruments) increased tissue temperature compared with baseline adjacent to the grasper tip (2.2 ± 2.2 °C; p = 0.013) and telescope tip (38.2 ± 8.0 °C; p < 0.001). The laparoscopic telescope tip increased tissue temperature more than the laparoscopic grasper tip (p < 0.001). Lowering the generator power from 30 to 15 Watts decreased the heat generated at the telescope tip (38.2 ± 8.0 vs. 13.5 ± 7.5 °C; p < 0.001). Complete separation of the camera/light cords and the active electrode cord decreased the heat generated near the telescope tip compared with parallel bundling of the cords (38.2 ± 8.0 vs. 15.7 ± 11.6 °C; p < 0.001).

Conclusions

Commonly used laparoscopic instruments couple monopolar radiofrequency energy without direct contact with the active electrode, a phenomenon that results in heat transfer from a nonelectrically active instrument tip to adjacent tissue. Practical steps to minimize heat transfer resulting from antenna coupling include reducing the monopolar generator power setting and avoiding of parallel bundling of the telescope and active electrode cords.
Literatur
1.
Zurück zum Zitat Nduka CC, Super PA, Monson JR, Darzi AW (1994) Cause and prevention of electrosurgical injuries in laparoscopy. J Am Coll Surg 179:161–170PubMed Nduka CC, Super PA, Monson JR, Darzi AW (1994) Cause and prevention of electrosurgical injuries in laparoscopy. J Am Coll Surg 179:161–170PubMed
2.
Zurück zum Zitat Hulka JF, Levy BS, Parker WH, Phillips JM (1997) Laparoscopic-assisted vaginal hysterectomy: American Association of Gynecologic Laparoscopists’ 1995 membership survey. J Am Assoc Gynecol Laparosc 4:167–171PubMedCrossRef Hulka JF, Levy BS, Parker WH, Phillips JM (1997) Laparoscopic-assisted vaginal hysterectomy: American Association of Gynecologic Laparoscopists’ 1995 membership survey. J Am Assoc Gynecol Laparosc 4:167–171PubMedCrossRef
3.
Zurück zum Zitat Feder BJ (2006) Surgical device poses a rare but serious peril. New York Times Feder BJ (2006) Surgical device poses a rare but serious peril. New York Times
4.
Zurück zum Zitat IEEE ASC (2009) American National Standard Dictionary of Electromagnetic Compatibility (EMC) including electromagnetic environmental effects (E3). Institute of Electrical and Electronics Engineers (IEEE) Inc, Washington, DC IEEE ASC (2009) American National Standard Dictionary of Electromagnetic Compatibility (EMC) including electromagnetic environmental effects (E3). Institute of Electrical and Electronics Engineers (IEEE) Inc, Washington, DC
5.
Zurück zum Zitat IEEE S (2005) IEEE Guide for instrumentation and control equipment grounding in generating stations. Institute of Electrical and Electronics Engineers (IEEE) Inc, Washington, DC IEEE S (2005) IEEE Guide for instrumentation and control equipment grounding in generating stations. Institute of Electrical and Electronics Engineers (IEEE) Inc, Washington, DC
6.
Zurück zum Zitat Montero PN, Robinson TN, Weaver JS, Stiegmann GV (2009) Insulation failure in laparoscopic instruments. Surg Endosc 24:462–465PubMedCrossRef Montero PN, Robinson TN, Weaver JS, Stiegmann GV (2009) Insulation failure in laparoscopic instruments. Surg Endosc 24:462–465PubMedCrossRef
7.
Zurück zum Zitat Robinson TN, Pavlovsky KR, Looney H, Stiegmann GV, McGreevy FT (2010) Surgeon-controlled factors that reduce monopolar electrosurgery capacitive coupling during laparoscopy. Surg Laparosc Endosc Percutan Tech 20:317–320PubMedCrossRef Robinson TN, Pavlovsky KR, Looney H, Stiegmann GV, McGreevy FT (2010) Surgeon-controlled factors that reduce monopolar electrosurgery capacitive coupling during laparoscopy. Surg Laparosc Endosc Percutan Tech 20:317–320PubMedCrossRef
8.
Zurück zum Zitat Voyles CR, Tucker RD (1992) Education and engineering solutions for potential problems with laparoscopic monopolar electrosurgery. Am J Surg 164:57–62PubMedCrossRef Voyles CR, Tucker RD (1992) Education and engineering solutions for potential problems with laparoscopic monopolar electrosurgery. Am J Surg 164:57–62PubMedCrossRef
9.
Zurück zum Zitat Govekar HR, Robinson TN, Stiegmann GV, McGreevy FT (2011) Residual heat of laparoscopic energy devices: how long must the surgeon wait to touch additional tissue? Surg Endosc 25(11):3499–3502PubMedCrossRef Govekar HR, Robinson TN, Stiegmann GV, McGreevy FT (2011) Residual heat of laparoscopic energy devices: how long must the surgeon wait to touch additional tissue? Surg Endosc 25(11):3499–3502PubMedCrossRef
10.
Zurück zum Zitat Wu MP, Ou CS, Chen SL, Yen EY, Rowbotham R (2000) Complications and recommended practices for electrosurgery in laparoscopy. Am J Surg 179:67–73PubMedCrossRef Wu MP, Ou CS, Chen SL, Yen EY, Rowbotham R (2000) Complications and recommended practices for electrosurgery in laparoscopy. Am J Surg 179:67–73PubMedCrossRef
11.
Zurück zum Zitat Robinson TN, Barnes K, Govekar HR, Stiegmann GV, Dunn C, McGreevy FT (2012) Antenna coupling: a novel mechanism of radiofrequency electrosurgery complication: practical implications. Ann Surg (In press) Robinson TN, Barnes K, Govekar HR, Stiegmann GV, Dunn C, McGreevy FT (2012) Antenna coupling: a novel mechanism of radiofrequency electrosurgery complication: practical implications. Ann Surg (In press)
12.
Zurück zum Zitat Abu-Rafea B, Vilos GA, Al-Obeed O, Al-Sheikh A, Vilos AG, Al-Mandeel H (2011) Monopolar electrosurgery through single-port laparoscopy: a potential hidden hazard for bowel burns. J Minim Invasive Gynecol 18:734–740PubMedCrossRef Abu-Rafea B, Vilos GA, Al-Obeed O, Al-Sheikh A, Vilos AG, Al-Mandeel H (2011) Monopolar electrosurgery through single-port laparoscopy: a potential hidden hazard for bowel burns. J Minim Invasive Gynecol 18:734–740PubMedCrossRef
Metadaten
Titel
Radiofrequency energy antenna coupling to common laparoscopic instruments: practical implications
verfasst von
Edward L. Jones
Thomas N. Robinson
Jennifer R. McHenry
Christina L. Dunn
Paul N. Montero
Henry R. Govekar
Greg V. Stiegmann
Publikationsdatum
01.11.2012
Verlag
Springer-Verlag
Erschienen in
Surgical Endoscopy / Ausgabe 11/2012
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-012-2312-6

Weitere Artikel der Ausgabe 11/2012

Surgical Endoscopy 11/2012 Zur Ausgabe

Deutlich weniger Infektionen: Wundprotektoren schützen!

08.05.2024 Postoperative Wundinfektion Nachrichten

Der Einsatz von Wundprotektoren bei offenen Eingriffen am unteren Gastrointestinaltrakt schützt vor Infektionen im Op.-Gebiet – und dient darüber hinaus der besseren Sicht. Das bestätigt mit großer Robustheit eine randomisierte Studie im Fachblatt JAMA Surgery.

Chirurginnen und Chirurgen sind stark suizidgefährdet

07.05.2024 Suizid Nachrichten

Der belastende Arbeitsalltag wirkt sich negativ auf die psychische Gesundheit der Angehörigen ärztlicher Berufsgruppen aus. Chirurginnen und Chirurgen bilden da keine Ausnahme, im Gegenteil.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Medizinstudium Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Echinokokkose medikamentös behandeln oder operieren?

06.05.2024 DCK 2024 Kongressbericht

Die Therapie von Echinokokkosen sollte immer in spezialisierten Zentren erfolgen. Eine symptomlose Echinokokkose kann – egal ob von Hunde- oder Fuchsbandwurm ausgelöst – konservativ erfolgen. Wenn eine Op. nötig ist, kann es sinnvoll sein, vorher Zysten zu leeren und zu desinfizieren. 

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.