Influence of bone density and implant drill diameter on the resulting axial force and temperature development in implant burs and artificial bone: an in vitro study

verfasst von: Stephan Christian Möhlhenrich, Mustapha Abouridouane, Nicole Heussen, Ali Modabber, Fritz Klocke, Frank Hölzle

Erschienen in: Oral and Maxillofacial Surgery | Ausgabe 2/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The aim of this study was to determine how the bone density affects the temperature development in artificial bone and drill.

Methods

Ten single drills with diameters of 2.2, 2.8, 3.5, and 4.2 mm were used on four artificial bone blocks (density I–IV), with constant speed and external irrigation. Temperature measurement in blocks and drills was done by infrared camera. The resultant axial force was measured, and light microscopic examinations of the drills were performed before and after preparation.

Results

The block density has a greater influence on resulting axial force than the drill diameter (D1 = 2.2 mm, 4.11 ± 0.64 N; 4.2 mm, 9.69 ± 0.78 N vs. D4 = 2.2 mm, 0.5 ± 0.18 N; 4.2 mm, 1.23 ± 0.08 N). For the narrowest drill, a decrease in bone density caused a significant temperature increase in the bone and drill. However, for the thickest drill, no thermal differences were found in the bone but were seen in the drill itself (D1 = 2.8 mm vs. D4 = 2.8 mm; bone p < 0.0001, drill p < 0.0001; D1 = 4.2 mm vs. D4 = 4.2 mm; bone p = 0.5366, drill p = 0.0411). An increase in the drill diameter in the highest bone density led to a significant thermal increase in the bone and drill. However, for the lowest bone density, thermal changes were observed only in the bone (D1 = 2.8 mm vs. D1 = 4.2 mm; bone p < 0.0001, drill p < 0.0001; D4 = 2.8 mm vs. D4 = 4.2 mm; bone p < 0.0102, drill p = 0.1784).

Conclusions

Thermal development depends on bone density with increasing density causing a temperature rise. However, this effect is reduced with increasing drill diameter. This may be important with regard to bone reactions and also in terms of tool wear.
Literatur
1.
Zurück zum Zitat Eriksson A, Albrektsson T, Grane B, McQueen D (1982) Thermal injury to bone. A vital-microscopic description of heat effects. Int J Oral Surg 11:115–121CrossRefPubMed Eriksson A, Albrektsson T, Grane B, McQueen D (1982) Thermal injury to bone. A vital-microscopic description of heat effects. Int J Oral Surg 11:115–121CrossRefPubMed
2.
Zurück zum Zitat Eriksson AR, Albrektsson T (1983) Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent 50:101–107CrossRefPubMed Eriksson AR, Albrektsson T (1983) Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent 50:101–107CrossRefPubMed
3.
Zurück zum Zitat Eriksson RA, Adell R (1986) Temperatures during drilling for the placement of implants using the osseointegration technique. J Oral Maxillofac Surg 44:4–7CrossRefPubMed Eriksson RA, Adell R (1986) Temperatures during drilling for the placement of implants using the osseointegration technique. J Oral Maxillofac Surg 44:4–7CrossRefPubMed
4.
Zurück zum Zitat Eriksson RA, Albrektsson T (1984) The effect of heat on bone regeneration: an experimental study in the rabbit using the bone growth chamber. J Oral Maxillofac Surg 42:705–711CrossRefPubMed Eriksson RA, Albrektsson T (1984) The effect of heat on bone regeneration: an experimental study in the rabbit using the bone growth chamber. J Oral Maxillofac Surg 42:705–711CrossRefPubMed
5.
Zurück zum Zitat Möhlhenrich SC, Modabber A, Steiner T, Mitchell DA, Hölzle F (2015) Heat generation and drill wear during dental implant site preparation: systematic review. Br J Oral Maxillofac Surg 53:679–689CrossRefPubMed Möhlhenrich SC, Modabber A, Steiner T, Mitchell DA, Hölzle F (2015) Heat generation and drill wear during dental implant site preparation: systematic review. Br J Oral Maxillofac Surg 53:679–689CrossRefPubMed
6.
Zurück zum Zitat Horch HH, Keiditsch E (1980) Morphological findings on the tissue lesion and bone regeneration after laser osteotomy. Dtsch Zahnarztl Z 35:22–24PubMed Horch HH, Keiditsch E (1980) Morphological findings on the tissue lesion and bone regeneration after laser osteotomy. Dtsch Zahnarztl Z 35:22–24PubMed
7.
Zurück zum Zitat D’Hodet B, Ney T, Mohlmann H, Luckenbach A (1987) Temperaturmessungen mit Hilfe der Infrarottechnik bei enossalen Fräsungen für dentale Implantatate. Z Zahnärztl Implantol 3:123–130 D’Hodet B, Ney T, Mohlmann H, Luckenbach A (1987) Temperaturmessungen mit Hilfe der Infrarottechnik bei enossalen Fräsungen für dentale Implantatate. Z Zahnärztl Implantol 3:123–130
8.
Zurück zum Zitat Misic T, Markovic A, Todorovic A, Colic S, Miodrag S, Milicic B (2011) An in vitro study of temperature changes in type 4 bone during implant placement: bone condensing versus bone drilling. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 112:28–33. doi:10.1016/j.tripleo.2010.08.010 CrossRefPubMed Misic T, Markovic A, Todorovic A, Colic S, Miodrag S, Milicic B (2011) An in vitro study of temperature changes in type 4 bone during implant placement: bone condensing versus bone drilling. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 112:28–33. doi:10.​1016/​j.​tripleo.​2010.​08.​010 CrossRefPubMed
9.
Zurück zum Zitat Oliveira N, Alaejos-Algarra F, Mareque-Bueno J, Ferres-Padro E, Hernandez-Alfaro F (2012) Thermal changes and drill wear in bovine bone during implant site preparation. A comparative in vitro study: twisted stainless steel and ceramic drills. Clin Oral Implants Res 23:963–969. doi:10.1111/j.1600-0501.2011.02248.x CrossRefPubMed Oliveira N, Alaejos-Algarra F, Mareque-Bueno J, Ferres-Padro E, Hernandez-Alfaro F (2012) Thermal changes and drill wear in bovine bone during implant site preparation. A comparative in vitro study: twisted stainless steel and ceramic drills. Clin Oral Implants Res 23:963–969. doi:10.​1111/​j.​1600-0501.​2011.​02248.​x CrossRefPubMed
11.
Zurück zum Zitat Allsobrook OF, Leichter J, Holborrow D, Swain M (2011) Descriptive study of the longevity of dental implant surgery drills. Clin Implant Dent Relat Res 13:244–254CrossRefPubMed Allsobrook OF, Leichter J, Holborrow D, Swain M (2011) Descriptive study of the longevity of dental implant surgery drills. Clin Implant Dent Relat Res 13:244–254CrossRefPubMed
12.
Zurück zum Zitat Bulloch SE, Olsen RG, Bulloch B (2012) Comparison of heat generation between internally guided (cannulated) single drill and traditional sequential drilling with and without a drill guide for dental implants. Int J Oral Maxillofac Implants 27:1456–1460PubMed Bulloch SE, Olsen RG, Bulloch B (2012) Comparison of heat generation between internally guided (cannulated) single drill and traditional sequential drilling with and without a drill guide for dental implants. Int J Oral Maxillofac Implants 27:1456–1460PubMed
13.
Zurück zum Zitat Misch CE (1989) Bone classification, training keys to implant success. Dent Today 8:39–44PubMed Misch CE (1989) Bone classification, training keys to implant success. Dent Today 8:39–44PubMed
15.
Zurück zum Zitat Gehrke SA, Bettach R, Taschieri S, Boukhris G, Corbella S, Del Fabbro M (2013) Temperature changes in cortical bone after implant site preparation using a single bur versus multiple drilling steps: an in vitro investigation. Clin Implant Dent Relat Res. doi:10.1111/cid.12172 Gehrke SA, Bettach R, Taschieri S, Boukhris G, Corbella S, Del Fabbro M (2013) Temperature changes in cortical bone after implant site preparation using a single bur versus multiple drilling steps: an in vitro investigation. Clin Implant Dent Relat Res. doi:10.​1111/​cid.​12172
16.
Zurück zum Zitat Strbac GD, Giannis K, Unger E, Mittlbock M, Watzek G, Zechner W (2014) A novel standardized bone model for thermal evaluation of bone osteotomies with various irrigation methods. Clin Oral Implants Res 25:622–631. doi:10.1111/clr.12090 CrossRefPubMed Strbac GD, Giannis K, Unger E, Mittlbock M, Watzek G, Zechner W (2014) A novel standardized bone model for thermal evaluation of bone osteotomies with various irrigation methods. Clin Oral Implants Res 25:622–631. doi:10.​1111/​clr.​12090 CrossRefPubMed
19.
Zurück zum Zitat Ercoli C, Funkenbusch PD, Lee HJ, Moss ME, Graser GN (2004) The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants: a study of drill durability. Int J Oral Maxillofac Implants 19:335–349PubMed Ercoli C, Funkenbusch PD, Lee HJ, Moss ME, Graser GN (2004) The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants: a study of drill durability. Int J Oral Maxillofac Implants 19:335–349PubMed
20.
Zurück zum Zitat Scarano A, Carinci F, Quaranta A, Di Iorio D, Assenza B, Piattelli A (2007) Effects of bur wear during implant site preparation: an in vitro study. Int J Immunopathol Pharmacol 20:23–26PubMed Scarano A, Carinci F, Quaranta A, Di Iorio D, Assenza B, Piattelli A (2007) Effects of bur wear during implant site preparation: an in vitro study. Int J Immunopathol Pharmacol 20:23–26PubMed
21.
Zurück zum Zitat Li B (2012) A review of tool wear estimation using theoretical analysis and numerical simulation technologies. Int J Refract Met Hard Mater 35:143–151CrossRef Li B (2012) A review of tool wear estimation using theoretical analysis and numerical simulation technologies. Int J Refract Met Hard Mater 35:143–151CrossRef
22.
Zurück zum Zitat Tabassum A, Meijer GJ, Wolke JG, Jansen JA (2009) Influence of the surgical technique and surface roughness on the primary stability of an implant in artificial bone with a density equivalent to maxillary bone: a laboratory study. Clin Oral Implants Res 20:327–332. doi:10.1111/j.1600-0501.2008.01692.x CrossRefPubMed Tabassum A, Meijer GJ, Wolke JG, Jansen JA (2009) Influence of the surgical technique and surface roughness on the primary stability of an implant in artificial bone with a density equivalent to maxillary bone: a laboratory study. Clin Oral Implants Res 20:327–332. doi:10.​1111/​j.​1600-0501.​2008.​01692.​x CrossRefPubMed
23.
Zurück zum Zitat Tabassum A, Meijer GJ, Wolke JG, Jansen JA (2010) Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study. Clin Oral Implants Res 21:213–220. doi:10.1111/j.1600-0501.2009.01823.x CrossRefPubMed Tabassum A, Meijer GJ, Wolke JG, Jansen JA (2010) Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study. Clin Oral Implants Res 21:213–220. doi:10.​1111/​j.​1600-0501.​2009.​01823.​x CrossRefPubMed
25.
Zurück zum Zitat Hsu JT, Huang HL, Chang CH, Tsai MT, Hung WC, Fuh LJ (2013) Relationship of three-dimensional bone-to-implant contact to primary implant stability and peri-implant bone strain in immediate loading: microcomputed tomographic and in vitro analyses. Int J Oral Maxillofac Implants 28:367–374, 10.11607/jomi.2407CrossRefPubMed Hsu JT, Huang HL, Chang CH, Tsai MT, Hung WC, Fuh LJ (2013) Relationship of three-dimensional bone-to-implant contact to primary implant stability and peri-implant bone strain in immediate loading: microcomputed tomographic and in vitro analyses. Int J Oral Maxillofac Implants 28:367–374, 10.11607/jomi.2407CrossRefPubMed
26.
Zurück zum Zitat Möhlhenrich SC, Heussen N, Loberg C, Goloborodko E, Holzle F, Modabber A (2015) Three-dimensional evaluation of implant bed preparation and the influence on primary implant stability after using 2 different surgical techniques. J Oral Maxillofac Surg. doi:10.1016/j.joms.2015.03.071 PubMed Möhlhenrich SC, Heussen N, Loberg C, Goloborodko E, Holzle F, Modabber A (2015) Three-dimensional evaluation of implant bed preparation and the influence on primary implant stability after using 2 different surgical techniques. J Oral Maxillofac Surg. doi:10.​1016/​j.​joms.​2015.​03.​071 PubMed
28.
Zurück zum Zitat Stelzle F, Frenkel C, Riemann M, Knipfer C, Stockmann P, Nkenke E (2014) The effect of load on heat production, thermal effects and expenditure of time during implant site preparation—an experimental ex vivo comparison between piezosurgery and conventional drilling. Clin Oral Implants Res 25:e140–e148. doi:10.1111/clr.12077 CrossRefPubMed Stelzle F, Frenkel C, Riemann M, Knipfer C, Stockmann P, Nkenke E (2014) The effect of load on heat production, thermal effects and expenditure of time during implant site preparation—an experimental ex vivo comparison between piezosurgery and conventional drilling. Clin Oral Implants Res 25:e140–e148. doi:10.​1111/​clr.​12077 CrossRefPubMed
29.
Zurück zum Zitat Scarano A, Piattelli A, Assenza B, Carinci F, Di Donato L, Romani GL, Merla A (2011) Infrared thermographic evaluation of temperature modifications induced during implant site preparation with cylindrical versus conical drills. Clin Implant Dent Relat Res 13:319–323. doi:10.1111/j.1708-8208.2009.00209.x CrossRefPubMed Scarano A, Piattelli A, Assenza B, Carinci F, Di Donato L, Romani GL, Merla A (2011) Infrared thermographic evaluation of temperature modifications induced during implant site preparation with cylindrical versus conical drills. Clin Implant Dent Relat Res 13:319–323. doi:10.​1111/​j.​1708-8208.​2009.​00209.​x CrossRefPubMed
30.
Zurück zum Zitat Strbac GD, Giannis K, Unger E, Mittlbock M, Vasak C, Watzek G, Zechner W (2013) Drilling- and withdrawing-related thermal changes during implant site osteotomies. Clin Implant Dent Relat Res. doi:10.1111/cid.12091 Strbac GD, Giannis K, Unger E, Mittlbock M, Vasak C, Watzek G, Zechner W (2013) Drilling- and withdrawing-related thermal changes during implant site osteotomies. Clin Implant Dent Relat Res. doi:10.​1111/​cid.​12091
31.
Zurück zum Zitat Lekholm U, Zarb G (1985) Patient selection and preparation. In: Branemark P-I, Zarb GA, Albrektsson T (eds) Tissue integrated prostheses: osseointegration in clinical dentistry, Chicago, Quintessence., pp 199–209 Lekholm U, Zarb G (1985) Patient selection and preparation. In: Branemark P-I, Zarb GA, Albrektsson T (eds) Tissue integrated prostheses: osseointegration in clinical dentistry, Chicago, Quintessence., pp 199–209
32.
Zurück zum Zitat Davidson SR, James DF (2000) Measurement of thermal conductivity of bovine cortical bone. Med Eng Phys 22:741–747CrossRefPubMed Davidson SR, James DF (2000) Measurement of thermal conductivity of bovine cortical bone. Med Eng Phys 22:741–747CrossRefPubMed
33.
Zurück zum Zitat Brisman DL (1996) The effect of speed, pressure, and time on bone temperature during the drilling of implant sites. Int J Oral Maxillofac Implants 11:35–37PubMed Brisman DL (1996) The effect of speed, pressure, and time on bone temperature during the drilling of implant sites. Int J Oral Maxillofac Implants 11:35–37PubMed
34.
Zurück zum Zitat Bachus KN, Rondina MT, Hutchinson DT (2000) The effects of drilling force on cortical temperatures and their duration: an in vitro study. Med Eng Phys 22:685–691CrossRefPubMed Bachus KN, Rondina MT, Hutchinson DT (2000) The effects of drilling force on cortical temperatures and their duration: an in vitro study. Med Eng Phys 22:685–691CrossRefPubMed
35.
Zurück zum Zitat Albrektsson T (1980) The healing of autologous bone grafts after varying degrees of surgical trauma. A microscopic and histochemical study in the rabbit. J Bone Joint Surg (Br) 62:403–410 Albrektsson T (1980) The healing of autologous bone grafts after varying degrees of surgical trauma. A microscopic and histochemical study in the rabbit. J Bone Joint Surg (Br) 62:403–410
36.
Zurück zum Zitat Singal RK, Singal M, Singal R (2008) Wear in life of cutting tools. In: Fundamentals of machining and machine tools, vol 1. I K International Publishing House Pvt Ltd, India, pp 230–254 Singal RK, Singal M, Singal R (2008) Wear in life of cutting tools. In: Fundamentals of machining and machine tools, vol 1. I K International Publishing House Pvt Ltd, India, pp 230–254
37.
Zurück zum Zitat Molinari A, Nouari M (2002) Modeling of tool wear by diffusion in metal cutting. Wear 252:135–149CrossRef Molinari A, Nouari M (2002) Modeling of tool wear by diffusion in metal cutting. Wear 252:135–149CrossRef
Metadaten
Titel
Influence of bone density and implant drill diameter on the resulting axial force and temperature development in implant burs and artificial bone: an in vitro study
verfasst von
Stephan Christian Möhlhenrich
Mustapha Abouridouane
Nicole Heussen
Ali Modabber
Fritz Klocke
Frank Hölzle
Publikationsdatum
20.11.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Oral and Maxillofacial Surgery / Ausgabe 2/2016
Print ISSN: 1865-1550
Elektronische ISSN: 1865-1569
DOI
https://doi.org/10.1007/s10006-015-0536-z

Weitere Artikel der Ausgabe 2/2016

Oral and Maxillofacial Surgery 2/2016 Zur Ausgabe

Wie erfolgreich ist eine Re-Ablation nach Rezidiv?

23.04.2024 Ablationstherapie Nachrichten

Nach der Katheterablation von Vorhofflimmern kommt es bei etwa einem Drittel der Patienten zu Rezidiven, meist binnen eines Jahres. Wie sich spätere Rückfälle auf die Erfolgschancen einer erneuten Ablation auswirken, haben Schweizer Kardiologen erforscht.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Ureterstriktur: Innovative OP-Technik bewährt sich

19.04.2024 EAU 2024 Kongressbericht

Die Ureterstriktur ist eine relativ seltene Komplikation, trotzdem bedarf sie einer differenzierten Versorgung. In komplexen Fällen wird dies durch die roboterassistierte OP-Technik gewährleistet. Erste Resultate ermutigen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.