ROC Study of Four LCD Displays Under Typical Medical Center Lighting Conditions

verfasst von: Steve Langer, Ken Fetterly, Jay Mandrekar, Scott Harmsen, Brian Bartholmai, Charles Patton, Alan Bishop, Colin McCannel

Erschienen in: Journal of Imaging Informatics in Medicine | Ausgabe 1/2006

Einloggen, um Zugang zu erhalten

Abstract

Nine observers reviewed a previously assembled library of 320 chest computed radiography (CR) images. Observers participated in four sessions, reading a different 1/4 of the sample on each of four liquid crystal displays: a 2-megapixel (MP) consumer color display, a 2-MP business color display, a 2-MP medical-grade gray display, and a 3-MP gray display. Each display was calibrated according to the DICOM Part 14 standard. The viewing application required observer login, then randomized the order of the subsample seen on the display, and timed the responses of the observer to render a 1–5 judgment on the absence or presence of ILD on chest CRs. Selections of 1–2 were considered negative, 3 was indeterminate, and 4–5 were positive. The order of viewing sessions was also randomized for each observer. The experiment was conducted under controlled lighting, temperature, and sound conditions to mimic conditions typically found in a patient examination room. Lighting was indirect, and illuminance at the display face was 195 ± 8% lux and was monitored over the course of the experiment. The average observer sensitivity for the 2 MP color consumer, 2 MP business color, 2 MP gray, and 3 MP gray displays were 83.7%, 84.1%, 85.5%, and 86.7%, respectively. The only pairwise significant difference was between the 2-MP consumer color and the 2-MP gray (P = 0.05). Effect of order within a session was not signitfficant (P = 0.21): period 1 (84.3%), period 2 (86.2%), period 3 (85.4%), period 4 (84.1%). Observer specificity for the various displays was not statistically significant (P = 0.21). Finally, a timing analysis showed no significant difference between the displays for the user group (P = 0.13), ranging from 5.3 s (2 MP color business) to 5.9 s (3 MP Gray). There was, however, a reduction in time over the study that was significant (P <<< 0.001) for all users; the group average decreased from 6.5 to 4.7 s per image. Physical measurements of the resolution, contrast, and noise properties of the displays were acquired. Most notably, the noise of the displays varied by 3.5× between the lowest and highest noise displays. Differences in display noise were indicative of observer performance. However, the large difference in the magnitude of the noise was not predictive of the small difference (3%) in the observer sensitivity for various displays. This is likely because detection of interstitial lung disease is limited by “““““anatomical noise””” rather than display or x-ray image noise.
Literatur
2.
Zurück zum Zitat Blume H, Steven PM, Bobb M, Ho AM, Stevens F, Muller S, Roehrig H, Fan J: Characterization of high-resolution liquid–crystal displays for medical images, Medical Imaging 2002: Visualization, Image-Guided Procedures, and Display, Proc of SPIE vol. 4681 (2002). Blume H, Steven PM, Bobb M, Ho AM, Stevens F, Muller S, Roehrig H, Fan J: Characterization of high-resolution liquid–crystal displays for medical images, Medical Imaging 2002: Visualization, Image-Guided Procedures, and Display, Proc of SPIE vol. 4681 (2002).
3.
Zurück zum Zitat Digital Imaging and Communications in Medicine, Part 14: Grayscale Standard Display Function, published by National Electrical Manufacturers Association, 2001. Digital Imaging and Communications in Medicine, Part 14: Grayscale Standard Display Function, published by National Electrical Manufacturers Association, 2001.
4.
Zurück zum Zitat Fetterly, KA, Schueler, BA 2003Performance evaluation of a ‘dual-side read’ computed radiography systemMed Phys3018431854CrossRefPubMed Fetterly, KA, Schueler, BA 2003Performance evaluation of a ‘dual-side read’ computed radiography systemMed Phys3018431854CrossRefPubMed
5.
Zurück zum Zitat Fleiss, JL, Levin, B, Paik, MC 2003Statistical Methods for Rates and Proportions, 3rd edn. WileyNew York Fleiss, JL, Levin, B, Paik, MC 2003Statistical Methods for Rates and Proportions, 3rd edn. WileyNew York
6.
Zurück zum Zitat Hsu, JC 1996Multiple Comparisons: Theory and MethodsChapman and HallLondon Hsu, JC 1996Multiple Comparisons: Theory and MethodsChapman and HallLondon
7.
Zurück zum Zitat Kramer, CY 1956Extension of multiple range tests to group means with unequal numbers of replicationsBiometrics12307310 Kramer, CY 1956Extension of multiple range tests to group means with unequal numbers of replicationsBiometrics12307310
8.
Zurück zum Zitat Langer, S, Bartholmai, B, Fetterly, K, Erickson, B, Harmson, S, Ryan, W, Andriole, K 2004SCAR 3'rd R&D Symposium: Efficacy of 5 megapixel CRT vs. 3 megapixel LCD displayJournal of Digital Imaging17149157CrossRefPubMed Langer, S, Bartholmai, B, Fetterly, K, Erickson, B, Harmson, S, Ryan, W, Andriole, K 2004SCAR 3'rd R&D Symposium: Efficacy of 5 megapixel CRT vs. 3 megapixel LCD displayJournal of Digital Imaging17149157CrossRefPubMed
9.
Zurück zum Zitat McNemar, Q 1947Note on the sampling error of the difference between correlated proportions or percentagesPsychometrika12153157CrossRef McNemar, Q 1947Note on the sampling error of the difference between correlated proportions or percentagesPsychometrika12153157CrossRef
10.
Zurück zum Zitat Tukey, JW 1994The problem of multiple comparisonsBraun, HI eds. The Collected Works of John W. Tukey, Volume VIII. Chapter 1Chapman and HallNew York1300 Tukey, JW 1994The problem of multiple comparisonsBraun, HI eds. The Collected Works of John W. Tukey, Volume VIII. Chapter 1Chapman and HallNew York1300
Metadaten
Titel
ROC Study of Four LCD Displays Under Typical Medical Center Lighting Conditions
verfasst von
Steve Langer
Ken Fetterly
Jay Mandrekar
Scott Harmsen
Brian Bartholmai
Charles Patton
Alan Bishop
Colin McCannel
Publikationsdatum
01.01.2006
Erschienen in
Journal of Imaging Informatics in Medicine / Ausgabe 1/2006
Print ISSN: 2948-2925
Elektronische ISSN: 2948-2933
DOI
https://doi.org/10.1007/s10278-005-8149-y

Weitere Artikel der Ausgabe 1/2006

Journal of Digital Imaging 1/2006 Zur Ausgabe

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Medizinstudium Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.