Influence of spatial resolution and contrast agent dosage on myocardial T1 relaxation times
verfasst von:
Edyta Blaszczyk, Agnieszka Töpper, Luisa Schmacht, Felix Wanke, Andreas Greiser, Jeanette Schulz-Menger, Florian von Knobelsdorff-Brenkenhoff
Erschienen in:
Magnetic Resonance Materials in Physics, Biology and Medicine
|
Ausgabe 1/2017
Einloggen, um Zugang zu erhalten
Abstract
Objective
Our aim was to study the influence of small variations in spatial resolution and contrast agent dosage on myocardial T1 relaxation time.
Materials and methods
Twenty-nine healthy volunteers underwent cardiovascular magnetic resonance at 3T twice, including a modified look-locker inversion recovery (MOLLI) technique—3(3)3(3)5—for T1 mapping. Native T1 was assessed in three spatial resolutions (voxel size 1.4 × 1.4 × 6, 1.6 × 1.6 × 6, 1.7 × 1.7 × 6 mm3), and postcontrast T1 after 0.1 and 0.2 mmol/kg gadobutrol. Partition coefficient was calculated based on myocardial and blood T1. T1 analysis was done per segment, per slice, and for the whole heart.
Results
Native T1 values did not differ with varying spatial resolution per segment (p = 0.116–0.980), per slice (basal: p = 0.772; middle: p = 0.639; apex: p = 0.276), and globally (p = 0.191). Postcontrast T1 values were significantly lower with higher contrast agent dosage (p < 0.001). The global partition coefficient was 0.43 ± 0.3 for 0.2 and 0.1 mmol gadobutrol (p = 0.079).
Conclusion
Related to the tested MOLLI technique at 3T, very small variations in spatial resolution (voxel sizes between 1.4 × 1.4 × 6 and 1.7 × 1.7 × 6 mm3) remained without effect on the native T1 relaxation times. Postcontrast T1 values were naturally shorter with higher contrast agent dosage while the partition coefficient remained constant. Further studies are necessary to test whether these conclusions hold true for larger matrix sizes and in larger cohorts.