Contrast media volume optimization in high-pitch dual-source CT coronary angiography: feasibility study

verfasst von: Wen Jie Yang, Ke Min Chen, Bo Liu, Li Fang Pang, Huan Zhang, Zi Lai Pan, Fu Hua Yan

Erschienen in: The International Journal of Cardiovascular Imaging | Ausgabe 1/2013

Einloggen, um Zugang zu erhalten

Abstract

To investigate the feasibility of contrast media (CM) volume reduction in dual-source coronary computed tomography angiography high-pitch mode without affecting coronary artery enhancement. Eighty patients were involved in a preliminary experiment with a default injection protocol (60 ml of CM). The age, BMI, test bolus (TB) enhancement peak and the CT values of coronary artery for each patient were recorded and the key factors for determining coronary artery enhancement were investigated. Based on the results of the preliminary experiment, 120 patients were involved in the main experiment with a new injection protocol. For each patient, the CT values and noise of left coronary sinus (LCS), and the distal segment of right coronary artery were measured. In the preliminary experiment, the peak enhancement of TB correlated most strongly with the HU values of coronary artery. Consequently, the new injection protocol was devised to catalog patients into four groups (30, 40, 50 and 60 ml) of CM based on their TB peak enhancement. In the main experiment, the 30 ml CM injection group whose peak attenuation of TB were the highest (30 vs. 40,50,60 ml: 323.0 ± 27.5 vs. 264.2 ± 11.9, 242.1 ± 8.8, 206.2 ± 18.2 HU, p < 0.05), obtained the highest attenuation of LCS (30 vs. 40,50,60 ml: 365.0 ± 41.2 vs. 341.8 ± 40.0, 326.9 ± 34.7, 312.5 ± 38.2 HU p < 0.05). Contrast optimization is feasible in high-pitch DSCT coronary angiography. Certain patients may receive 30 ml of CM without affecting vessel enhancement.
Literatur
1.
Zurück zum Zitat Ehara M, Surmely JF, Kawai M et al (2006) Diagnostic accuracy of 64-slice computed tomography for detecting angiographically significant coronary artery restenosis in an unselected consecutive patient population: comparison with conventional invasive angiography. Circ J 70:564–571PubMedCrossRef Ehara M, Surmely JF, Kawai M et al (2006) Diagnostic accuracy of 64-slice computed tomography for detecting angiographically significant coronary artery restenosis in an unselected consecutive patient population: comparison with conventional invasive angiography. Circ J 70:564–571PubMedCrossRef
2.
Zurück zum Zitat Garcia MJ, Lessick J, Hoffmann MH (2006) CATSCAN study investigators accuracy of 16-row multidetector computed tomography for the assessment of coronary artery stenosis. JAMA 296:403–411PubMedCrossRef Garcia MJ, Lessick J, Hoffmann MH (2006) CATSCAN study investigators accuracy of 16-row multidetector computed tomography for the assessment of coronary artery stenosis. JAMA 296:403–411PubMedCrossRef
3.
Zurück zum Zitat Nikolaou K, Rist C, Wintersperger BJ et al (2006) Clinical value of MDCT in the diagnosis of coronary artery disease in patients with a low pretest likelihood of significant disease. Am J Roentgenol 186:1659–1668CrossRef Nikolaou K, Rist C, Wintersperger BJ et al (2006) Clinical value of MDCT in the diagnosis of coronary artery disease in patients with a low pretest likelihood of significant disease. Am J Roentgenol 186:1659–1668CrossRef
4.
Zurück zum Zitat Muhlenbruch G, Seyfarth T, Soo CS et al (2007) Diagnostic value of 64-slice multidetector row cardiac CTA in symptomatic patients. Eur Radiol 17:603–609PubMedCrossRef Muhlenbruch G, Seyfarth T, Soo CS et al (2007) Diagnostic value of 64-slice multidetector row cardiac CTA in symptomatic patients. Eur Radiol 17:603–609PubMedCrossRef
5.
Zurück zum Zitat Mann JF, Gerstein HC, Yi QL et al (2003) Development of renal disease in people at high cardiovascular risk: results of the HOPE randomized trial. J Am Soc Nephrol 14:641–647PubMedCrossRef Mann JF, Gerstein HC, Yi QL et al (2003) Development of renal disease in people at high cardiovascular risk: results of the HOPE randomized trial. J Am Soc Nephrol 14:641–647PubMedCrossRef
6.
Zurück zum Zitat Finn WF (2006) The clinical and renal consequences of contrast-induced nephropathy. Nephrol Dial Transplant 21:2–10CrossRef Finn WF (2006) The clinical and renal consequences of contrast-induced nephropathy. Nephrol Dial Transplant 21:2–10CrossRef
7.
Zurück zum Zitat From AM, Bartholmai BJ, Williams AW et al (2008) Mortality associated with nephropathy after radiographic contrast exposure. Mayo Clin Proc 83:1095–1100PubMedCrossRef From AM, Bartholmai BJ, Williams AW et al (2008) Mortality associated with nephropathy after radiographic contrast exposure. Mayo Clin Proc 83:1095–1100PubMedCrossRef
8.
Zurück zum Zitat Toprak O (2007) Conflicting and new risk factors for contrast induced nephropathy. J Urol 178:2277–2283PubMedCrossRef Toprak O (2007) Conflicting and new risk factors for contrast induced nephropathy. J Urol 178:2277–2283PubMedCrossRef
9.
Zurück zum Zitat Mann JF, Gerstein HC, Yi QL et al (2003) Development of renal disease in people at high cardiovascular risk: results of the HOPE randomized trial. J Am Soc Nephrol 14:641–647PubMedCrossRef Mann JF, Gerstein HC, Yi QL et al (2003) Development of renal disease in people at high cardiovascular risk: results of the HOPE randomized trial. J Am Soc Nephrol 14:641–647PubMedCrossRef
10.
Zurück zum Zitat Rist C, Nikolaou K, Kirchin MA et al (2006) Contrast bolus optimization for cardiac 16-slice computed tomography: comparison of contrast medium formulations containing 300 and 400 mg of iodine per milliliter. Invest Radiol 41:460–467PubMedCrossRef Rist C, Nikolaou K, Kirchin MA et al (2006) Contrast bolus optimization for cardiac 16-slice computed tomography: comparison of contrast medium formulations containing 300 and 400 mg of iodine per milliliter. Invest Radiol 41:460–467PubMedCrossRef
11.
Zurück zum Zitat Cademartiri F, Mollet NR, van der Lugt A et al (2005) Intravenous contrast material administration at helical 16-detector row CT coronary angiography: effect of iodine concentration on vascular attenuation. Radiology 236:661–665PubMedCrossRef Cademartiri F, Mollet NR, van der Lugt A et al (2005) Intravenous contrast material administration at helical 16-detector row CT coronary angiography: effect of iodine concentration on vascular attenuation. Radiology 236:661–665PubMedCrossRef
12.
Zurück zum Zitat Cademartiri F, de Monye C, Pugliese F et al (2006) High iodine concentration contrast material for noninvasive multislice computed tomography coronary angiography: iopromide 370 versus iomeprol 400. Invest Radiol 41:349–353PubMedCrossRef Cademartiri F, de Monye C, Pugliese F et al (2006) High iodine concentration contrast material for noninvasive multislice computed tomography coronary angiography: iopromide 370 versus iomeprol 400. Invest Radiol 41:349–353PubMedCrossRef
13.
Zurück zum Zitat Cademartiri F, Maffei E, Palumbo AA et al (2008) Influence of intra-coronary enhancement on diagnostic accuracy with 64-slice CT coronary angiography. Eur Radiol 18(3):576–583PubMedCrossRef Cademartiri F, Maffei E, Palumbo AA et al (2008) Influence of intra-coronary enhancement on diagnostic accuracy with 64-slice CT coronary angiography. Eur Radiol 18(3):576–583PubMedCrossRef
14.
Zurück zum Zitat Fleischmann D (2003) High-concentration contrast media in MDCT angiography:principles and rationale. Eur Radiol 13(Suppl. 3):N39–N43 Fleischmann D (2003) High-concentration contrast media in MDCT angiography:principles and rationale. Eur Radiol 13(Suppl. 3):N39–N43
15.
Zurück zum Zitat Sakai S, Yabuuchi H, Chishaki A et al (2010) Effect of cardiac function on aortic peak time and peak enhancement during coronary CT angiography. Eur J Radiol 75(2):173–177. (Epub 2009 May 13) Sakai S, Yabuuchi H, Chishaki A et al (2010) Effect of cardiac function on aortic peak time and peak enhancement during coronary CT angiography. Eur J Radiol 75(2):173–177. (Epub 2009 May 13)
16.
Zurück zum Zitat Dewey M (2009) Coronary CT Angiography. Springer-Verlag, Berlin Heidelberg, ISBN 9783540798439 Dewey M (2009) Coronary CT Angiography. Springer-Verlag, Berlin Heidelberg, ISBN 9783540798439
17.
Zurück zum Zitat Rybicki FJ, Otero HJ, Steigner ML et al (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging 24:535–546PubMedCrossRef Rybicki FJ, Otero HJ, Steigner ML et al (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging 24:535–546PubMedCrossRef
18.
Zurück zum Zitat Steigner ML, Otero HJ, Cai T et al (2009) Narrowing the phase window width in prospectively ECG-gated single heart beat 320-detector row coronary CT angiography. Int J Cardiovasc Imaging 25(1):85–90PubMedCrossRef Steigner ML, Otero HJ, Cai T et al (2009) Narrowing the phase window width in prospectively ECG-gated single heart beat 320-detector row coronary CT angiography. Int J Cardiovasc Imaging 25(1):85–90PubMedCrossRef
19.
Zurück zum Zitat Kumamaru KK, Steigner ML, Soga S et al (2011) Coronary enhancement for prospective ECG-gated single R–R axial 320-detector row coronary computed tomography angiography: comparison of 60 and 80 ml iopamidol-370 contrast medium injection. Am J Roentgenol 197(4):844–850CrossRef Kumamaru KK, Steigner ML, Soga S et al (2011) Coronary enhancement for prospective ECG-gated single R–R axial 320-detector row coronary computed tomography angiography: comparison of 60 and 80 ml iopamidol-370 contrast medium injection. Am J Roentgenol 197(4):844–850CrossRef
20.
Zurück zum Zitat Van Hoe L, Marchal G, Baert AL et al (1995) Determination of scan delay-time in spiral CT-angiography: utility of a test bolus injection. J Comput Assist Tomogr 19(2):216–220PubMedCrossRef Van Hoe L, Marchal G, Baert AL et al (1995) Determination of scan delay-time in spiral CT-angiography: utility of a test bolus injection. J Comput Assist Tomogr 19(2):216–220PubMedCrossRef
21.
Zurück zum Zitat Nakaura T, Awai K, Yauaga Y et al (2008) Contrast injection protocols for coronary computed tomography angiography using a 64-detector scanner: comparison between patient weight adjusted-and fixed iodine-dose protocols. Invest Radiol 43:512–519PubMedCrossRef Nakaura T, Awai K, Yauaga Y et al (2008) Contrast injection protocols for coronary computed tomography angiography using a 64-detector scanner: comparison between patient weight adjusted-and fixed iodine-dose protocols. Invest Radiol 43:512–519PubMedCrossRef
22.
Zurück zum Zitat Yamamuro M, Tadamura E, Kanao S et al (2007) Coronary angiography by 64-detector row computed tomography using low dose of contrast material with saline chaser: influence of total injection volume on vessel attenuation. J Comput Assist Tomogr 31:272–280PubMedCrossRef Yamamuro M, Tadamura E, Kanao S et al (2007) Coronary angiography by 64-detector row computed tomography using low dose of contrast material with saline chaser: influence of total injection volume on vessel attenuation. J Comput Assist Tomogr 31:272–280PubMedCrossRef
23.
Zurück zum Zitat Tatsugami F, Matsuki M, Inada Y et al (2010) Feasibility of low-volume injections of contrast material with a body weight-adapted iodine-dose protocol in 320-detector row coronary CT angiography. Acad Radiol 17(2):207–211. (Epub 2009 Nov 17) Tatsugami F, Matsuki M, Inada Y et al (2010) Feasibility of low-volume injections of contrast material with a body weight-adapted iodine-dose protocol in 320-detector row coronary CT angiography. Acad Radiol 17(2):207–211. (Epub 2009 Nov 17)
24.
Zurück zum Zitat Ho LM, Nelson RC, DeLong DM (2007) Determining contrast medium dose and rate on basis of lean body weight: does this strategy improve patient-to patient uniformity of hepatic enhancement during multi-detector row CT? Radiology 243:431–437PubMedCrossRef Ho LM, Nelson RC, DeLong DM (2007) Determining contrast medium dose and rate on basis of lean body weight: does this strategy improve patient-to patient uniformity of hepatic enhancement during multi-detector row CT? Radiology 243:431–437PubMedCrossRef
25.
Zurück zum Zitat Hittmair K, Fleischmann D (2001) Accuracy of predicting and controlling time-dependent aortic enhancement from a test bolus injection. J Comput Assist Tomogr 25(2):287–294PubMedCrossRef Hittmair K, Fleischmann D (2001) Accuracy of predicting and controlling time-dependent aortic enhancement from a test bolus injection. J Comput Assist Tomogr 25(2):287–294PubMedCrossRef
26.
Zurück zum Zitat Steigner MT, Mitsouras D, Whitmore AG et al (2010) Iodinated contrast opacification gradients in normal coronary arteries imaged with prospectively ECG-gated single heart beat 320-detector row computed tomography. Circ Cardiovasc Imaging 3(2):179–186PubMedCrossRef Steigner MT, Mitsouras D, Whitmore AG et al (2010) Iodinated contrast opacification gradients in normal coronary arteries imaged with prospectively ECG-gated single heart beat 320-detector row computed tomography. Circ Cardiovasc Imaging 3(2):179–186PubMedCrossRef
27.
Zurück zum Zitat Rybicki FJ, Melchionna S, Mitsouras D et al (2009) Prediction of coronary artery plaque progression and potential rupture from 320-detector row prospectively ECG-gated single heart beat CT angiography: Lattice Boltzmann evaluation of endothelial shear stress. Int J Cardiovasc Imaging 25:289–299CrossRef Rybicki FJ, Melchionna S, Mitsouras D et al (2009) Prediction of coronary artery plaque progression and potential rupture from 320-detector row prospectively ECG-gated single heart beat CT angiography: Lattice Boltzmann evaluation of endothelial shear stress. Int J Cardiovasc Imaging 25:289–299CrossRef
28.
Zurück zum Zitat Melchionna S, Bernaschi M, Succi S et al (2010) Hydrokinetic approach to large-scale cardiovascular blood flow. Comput Phys Commun 181:462–472CrossRef Melchionna S, Bernaschi M, Succi S et al (2010) Hydrokinetic approach to large-scale cardiovascular blood flow. Comput Phys Commun 181:462–472CrossRef
Metadaten
Titel
Contrast media volume optimization in high-pitch dual-source CT coronary angiography: feasibility study
verfasst von
Wen Jie Yang
Ke Min Chen
Bo Liu
Li Fang Pang
Huan Zhang
Zi Lai Pan
Fu Hua Yan
Publikationsdatum
01.01.2013
Verlag
Springer Netherlands
Erschienen in
The International Journal of Cardiovascular Imaging / Ausgabe 1/2013
Print ISSN: 1569-5794
Elektronische ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-012-0068-7

Weitere Artikel der Ausgabe 1/2013

The International Journal of Cardiovascular Imaging 1/2013 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.