Surface Damage Versus Tibial Polyethylene Insert Conformity: A Retrieval Study

verfasst von: Markus A. Wimmer, PhD, Michel P. Laurent, PhD, Jeannie D. Haman, PhD, Joshua J. Jacobs, MD, Jorge O. Galante, MD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 7/2012

Einloggen, um Zugang zu erhalten

Abstract

Background

Surface damage of the tibial polyethylene insert in TKA is thought to diminish with increasing conformity, based on computed lower contact stresses. Added constraint from higher conformity may, however, result in greater forces in vivo.

Questions/purposes

We therefore determined whether increased conformity was associated with increased surface pitting, delamination, creep, and polishing in a group of retrieved tibial inserts.

Methods

We compared 38 inserts with a dished articular surface (conforming group) with 31 inserts that were unconstrained and nonconforming in the sagittal plane (less conforming group). The two groups had identical polyethylene composition and processing history. The articulating surfaces were scored for pitting, delamination, deformation/creep, and polishing. Evidence of edge loading and the presence of embedded bone cement were also recorded.

Results

The conforming inserts were associated with higher delamination and pitting scores but lower polishing scores, even after adjusting for the effects of sex, age, insert thickness, and implantation duration. Long implantation duration and male sex were also associated with increased delamination, pitting, and polishing, whereas long shelf life was associated only with increased delamination. The conforming group also had approximately a fourfold greater prevalence of edge loading and approximately a threefold greater prevalence of embedded bone cement. The latter was associated with higher scores and proportions of delamination and pitting.

Conclusions

These findings suggest more conformity may increase surface fatigue damage in TKA. Higher constraint-induced stresses during secondary motions and more possibility for edge loading and bone cement capture on a dished surface may account for these results.

Clinical Relevance

The selection of materials with high fatigue resistance may be particularly important for high-conformity/constraint tibial inserts. In addition, awareness of the benefits and trade-offs with conformity may allow better matching of TKA design to patient.
Literatur
1.
Zurück zum Zitat Andriacchi TP. Dynamics of knee malalignment. Orthop Clin North Am. 1994;25:395–398.PubMed Andriacchi TP. Dynamics of knee malalignment. Orthop Clin North Am. 1994;25:395–398.PubMed
2.
Zurück zum Zitat Archard JF. Contact and rubbing of flat surface. J Appl Phys. 1953;24:981–988.CrossRef Archard JF. Contact and rubbing of flat surface. J Appl Phys. 1953;24:981–988.CrossRef
3.
Zurück zum Zitat Bartel DL, Bicknell VL, Wright TM. The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am. 1986;68:1041–1051.PubMed Bartel DL, Bicknell VL, Wright TM. The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am. 1986;68:1041–1051.PubMed
4.
Zurück zum Zitat Bartel DL, Rawlinson JJ, Burstein AH, Ranawat CS, Flynn WF Jr. Stresses in polyethylene components of contemporary total knee replacements. Clin Orthop Relat Res. 1995;317:76–82.PubMed Bartel DL, Rawlinson JJ, Burstein AH, Ranawat CS, Flynn WF Jr. Stresses in polyethylene components of contemporary total knee replacements. Clin Orthop Relat Res. 1995;317:76–82.PubMed
5.
Zurück zum Zitat Bell CJ, Walker PS, Abeysundera MR, Simmons JM, King PM, Blunn GW. Effect of oxidation on delamination of ultrahigh-molecular-weight polyethylene tibial components. J Arthroplasty. 1998;13:280–290.PubMedCrossRef Bell CJ, Walker PS, Abeysundera MR, Simmons JM, King PM, Blunn GW. Effect of oxidation on delamination of ultrahigh-molecular-weight polyethylene tibial components. J Arthroplasty. 1998;13:280–290.PubMedCrossRef
6.
Zurück zum Zitat Berzins A, Jacobs JJ, Berger R, Ed C, Natarajan R, Andriacchi T, Galante JO. Surface damage in machined ram-extruded and net-shape molded retrieved polyethylene tibial inserts of total knee replacements. J Bone Joint Surg Am. 2002;84:1534–1540.PubMed Berzins A, Jacobs JJ, Berger R, Ed C, Natarajan R, Andriacchi T, Galante JO. Surface damage in machined ram-extruded and net-shape molded retrieved polyethylene tibial inserts of total knee replacements. J Bone Joint Surg Am. 2002;84:1534–1540.PubMed
7.
Zurück zum Zitat Blunn GW, Joshi AB, Lilley PA, Engelbrecht E, Ryd L, Lidgren L, Hardinge K, Nieder E, Walker PS. Polyethylene wear in unicondylar knee prostheses: 106 retrieved Marmor, PCA, and St Georg tibial components compared. Acta Orthop Scand. 1992;63:247–255.PubMedCrossRef Blunn GW, Joshi AB, Lilley PA, Engelbrecht E, Ryd L, Lidgren L, Hardinge K, Nieder E, Walker PS. Polyethylene wear in unicondylar knee prostheses: 106 retrieved Marmor, PCA, and St Georg tibial components compared. Acta Orthop Scand. 1992;63:247–255.PubMedCrossRef
8.
Zurück zum Zitat Blunn GW, Joshi AB, Minns RJ, Lidgren L, Lilley P, Ryd L, Engelbrecht E, Walker PS. Wear in retrieved condylar knee arthroplasties: a comparison of wear in different designs of 280 retrieved condylar knee prostheses. J Arthroplasty. 1997;12:281–290.PubMedCrossRef Blunn GW, Joshi AB, Minns RJ, Lidgren L, Lilley P, Ryd L, Engelbrecht E, Walker PS. Wear in retrieved condylar knee arthroplasties: a comparison of wear in different designs of 280 retrieved condylar knee prostheses. J Arthroplasty. 1997;12:281–290.PubMedCrossRef
9.
Zurück zum Zitat Bragdon CR, O’Connor DO, Lowenstein JD, Jasty M, Syniuta WD. The importance of multidirectional motion on the wear of polyethylene. Proc Inst Mech Eng H. 1996;210:157–165.PubMed Bragdon CR, O’Connor DO, Lowenstein JD, Jasty M, Syniuta WD. The importance of multidirectional motion on the wear of polyethylene. Proc Inst Mech Eng H. 1996;210:157–165.PubMed
10.
Zurück zum Zitat Burstein AH, Wright TM. Biomechanics. In: Insall JN, Windsor RE, Scott WN, Kelly MA, Aglietti P, eds. Surgery of the Knee. 2nd ed. New York, NY: Churchill Livingstone; 1993:43–62. Burstein AH, Wright TM. Biomechanics. In: Insall JN, Windsor RE, Scott WN, Kelly MA, Aglietti P, eds. Surgery of the Knee. 2nd ed. New York, NY: Churchill Livingstone; 1993:43–62.
11.
Zurück zum Zitat Collier JP, Mayor MB, McNamara JL, Surprenant VA, Jensen RE. Analysis of the failure of 122 polyethylene inserts from uncemented tibial knee components. Clin Orthop Relat Res. 1991;273:232–242.PubMed Collier JP, Mayor MB, McNamara JL, Surprenant VA, Jensen RE. Analysis of the failure of 122 polyethylene inserts from uncemented tibial knee components. Clin Orthop Relat Res. 1991;273:232–242.PubMed
12.
Zurück zum Zitat Collier JP, Sperling DK, Currier JH, Sutula LC, Saum KA, Mayor MB. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplasty. 1996;11:377–389.PubMedCrossRef Collier JP, Sperling DK, Currier JH, Sutula LC, Saum KA, Mayor MB. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplasty. 1996;11:377–389.PubMedCrossRef
13.
Zurück zum Zitat Costa L, Luda MP, Trossarelli L, Brach del Prever EM, Crova M, Gallinaro P. In vivo UHMWPE biodegradation of retrieved prosthesis. Biomaterials. 1998;19:1371–1385.PubMedCrossRef Costa L, Luda MP, Trossarelli L, Brach del Prever EM, Crova M, Gallinaro P. In vivo UHMWPE biodegradation of retrieved prosthesis. Biomaterials. 1998;19:1371–1385.PubMedCrossRef
14.
Zurück zum Zitat Currier BH, Currier JH, Mayor MB, Lyford KA, Van Citters DW, Collier JP. In vivo oxidation of gamma-barrier-sterilized ultra-high-molecular-weight polyethylene bearings. J Arthroplasty. 2007;22:721–731.PubMedCrossRef Currier BH, Currier JH, Mayor MB, Lyford KA, Van Citters DW, Collier JP. In vivo oxidation of gamma-barrier-sterilized ultra-high-molecular-weight polyethylene bearings. J Arthroplasty. 2007;22:721–731.PubMedCrossRef
15.
Zurück zum Zitat D’Lima DD, Chen PC, Colwell CW Jr. Polyethylene contact stresses, articular congruity, and knee alignment. Clin Orthop Relat Res. 2001;392:232–238.PubMedCrossRef D’Lima DD, Chen PC, Colwell CW Jr. Polyethylene contact stresses, articular congruity, and knee alignment. Clin Orthop Relat Res. 2001;392:232–238.PubMedCrossRef
16.
Zurück zum Zitat Fregly BJ, Marquez-Barrientos C, Banks SA, DesJardins JD. Increased conformity offers diminishing returns for reducing total knee replacement wear. J Biomech Eng. 2010;132:021007.PubMedCrossRef Fregly BJ, Marquez-Barrientos C, Banks SA, DesJardins JD. Increased conformity offers diminishing returns for reducing total knee replacement wear. J Biomech Eng. 2010;132:021007.PubMedCrossRef
17.
Zurück zum Zitat Furman BD, Lipman J, Kligman M, Wright TM, Haas SB. Tibial post wear in posterior-stabilized knee replacements is design-dependent. Clin Orthop Relat Res. 2008;466:2650–2655.PubMedCrossRef Furman BD, Lipman J, Kligman M, Wright TM, Haas SB. Tibial post wear in posterior-stabilized knee replacements is design-dependent. Clin Orthop Relat Res. 2008;466:2650–2655.PubMedCrossRef
18.
Zurück zum Zitat Galvin AL, Kang L, Udofia I, Jennings LM, McEwen HM, Jin Z, Fisher J. Effect of conformity and contact stress on wear in fixed-bearing total knee prostheses. J Biomech. 2009;42:1898–1902.PubMedCrossRef Galvin AL, Kang L, Udofia I, Jennings LM, McEwen HM, Jin Z, Fisher J. Effect of conformity and contact stress on wear in fixed-bearing total knee prostheses. J Biomech. 2009;42:1898–1902.PubMedCrossRef
19.
Zurück zum Zitat Hood RW, Wright TM, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res. 1983;17:829–842.PubMedCrossRef Hood RW, Wright TM, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res. 1983;17:829–842.PubMedCrossRef
20.
Zurück zum Zitat Ingham E, Fisher J. Biological reactions to wear debris in total joint replacement. Proc Inst Mech Eng H. 2000;214:21–37.PubMed Ingham E, Fisher J. Biological reactions to wear debris in total joint replacement. Proc Inst Mech Eng H. 2000;214:21–37.PubMed
21.
Zurück zum Zitat Keeney JA, Eunice S, Pashos G, Wright RW, Clohisy JC. What is the evidence for total knee arthroplasty in young patients? A systematic review of the literature. Clin Orthop Relat Res. 2011;469:574–583.PubMedCrossRef Keeney JA, Eunice S, Pashos G, Wright RW, Clohisy JC. What is the evidence for total knee arthroplasty in young patients? A systematic review of the literature. Clin Orthop Relat Res. 2011;469:574–583.PubMedCrossRef
22.
Zurück zum Zitat Kurtz SM. UHMWPE Biomaterials Handbook. 2nd ed. Boston, MA: Elsevier; 2009. Kurtz SM. UHMWPE Biomaterials Handbook. 2nd ed. Boston, MA: Elsevier; 2009.
23.
Zurück zum Zitat Kurtz SM, Bartel DL, Rimnac CM. Postirradiation aging affects stress and strain in polyethylene components. Clin Orthop Relat Res. 1998;350:209–220.PubMedCrossRef Kurtz SM, Bartel DL, Rimnac CM. Postirradiation aging affects stress and strain in polyethylene components. Clin Orthop Relat Res. 1998;350:209–220.PubMedCrossRef
24.
Zurück zum Zitat Landy MM, Walker PS. Wear of ultra-high-molecular-weight polyethylene components of 90 retrieved knee prostheses. J Arthroplasty. 1988;3:S73–S85.PubMedCrossRef Landy MM, Walker PS. Wear of ultra-high-molecular-weight polyethylene components of 90 retrieved knee prostheses. J Arthroplasty. 1988;3:S73–S85.PubMedCrossRef
25.
Zurück zum Zitat Laurent MP, Johnson TS, Yao JQ, Blanchard CR, Crowninshield RD. In vitro lateral versus medial wear of a knee prosthesis. Wear. 2003;255:1101–1106.CrossRef Laurent MP, Johnson TS, Yao JQ, Blanchard CR, Crowninshield RD. In vitro lateral versus medial wear of a knee prosthesis. Wear. 2003;255:1101–1106.CrossRef
26.
Zurück zum Zitat Lavernia CJ, Sierra RJ, Hungerford DS, Krackow K. Activity level and wear in total knee arthroplasty: a study of autopsy retrieved specimens. J Arthroplasty. 2001;16:446–453.PubMedCrossRef Lavernia CJ, Sierra RJ, Hungerford DS, Krackow K. Activity level and wear in total knee arthroplasty: a study of autopsy retrieved specimens. J Arthroplasty. 2001;16:446–453.PubMedCrossRef
27.
Zurück zum Zitat Liau JJ, Cheng CK, Huang CH, Lo WH. The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol, Avon). 2002;17:140–146.CrossRef Liau JJ, Cheng CK, Huang CH, Lo WH. The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol, Avon). 2002;17:140–146.CrossRef
28.
Zurück zum Zitat Lidgren L, Robertsson O, eds. The Swedish Arthroplasty Registry–Annual Report 2008. Lund, Sweden: Wallin & Dalholm AB; 2008. Lidgren L, Robertsson O, eds. The Swedish Arthroplasty Registry–Annual Report 2008. Lund, Sweden: Wallin & Dalholm AB; 2008.
29.
Zurück zum Zitat McClung CD, Zahiri CA, Higa JK, Amstutz HC, Schmalzried TP. Relationship between body mass index and activity in hip or knee arthroplasty patients. J Orthop Res. 2000;18:35–39.PubMedCrossRef McClung CD, Zahiri CA, Higa JK, Amstutz HC, Schmalzried TP. Relationship between body mass index and activity in hip or knee arthroplasty patients. J Orthop Res. 2000;18:35–39.PubMedCrossRef
30.
Zurück zum Zitat McDowell MA, Fryar CD, Ogden CL. Anthropometric reference data for children and adults: United States, 1988–1994. Vital Health Stat 11. 2009;249:1–68.PubMed McDowell MA, Fryar CD, Ogden CL. Anthropometric reference data for children and adults: United States, 1988–1994. Vital Health Stat 11. 2009;249:1–68.PubMed
31.
Zurück zum Zitat McGloughlin TM, Kavanagh AG. Wear of ultra-high molecular weight polyethylene (UHMWPE) in total knee prostheses: a review of key influences. Proc Inst Mech Eng H. 2000;214:349–359.PubMedCrossRef McGloughlin TM, Kavanagh AG. Wear of ultra-high molecular weight polyethylene (UHMWPE) in total knee prostheses: a review of key influences. Proc Inst Mech Eng H. 2000;214:349–359.PubMedCrossRef
32.
Zurück zum Zitat Medel FJ, Kurtz SM, Parvizi J, Klein GR, Kraay MJ, Rimnac CM. In vivo oxidation contributes to delamination but not pitting in polyethylene components for total knee arthroplasty. J Arthroplasty. 2011;26:802–810.PubMedCrossRef Medel FJ, Kurtz SM, Parvizi J, Klein GR, Kraay MJ, Rimnac CM. In vivo oxidation contributes to delamination but not pitting in polyethylene components for total knee arthroplasty. J Arthroplasty. 2011;26:802–810.PubMedCrossRef
33.
Zurück zum Zitat Naudie DD, Ammeen DJ, Engh GA, Rorabeck CH. Wear and osteolysis around total knee arthroplasty. J Am Acad Orthop Surg. 2007;15:53–64.PubMed Naudie DD, Ammeen DJ, Engh GA, Rorabeck CH. Wear and osteolysis around total knee arthroplasty. J Am Acad Orthop Surg. 2007;15:53–64.PubMed
34.
Zurück zum Zitat Pascau A, Guardia B, Puertolas JA, Gómez-Barrena E. Knee model of hydrodynamic lubrication during the gait cycle and the influence of prosthetic joint conformity. J Orthop Sci. 2009;14:68–75.PubMedCrossRef Pascau A, Guardia B, Puertolas JA, Gómez-Barrena E. Knee model of hydrodynamic lubrication during the gait cycle and the influence of prosthetic joint conformity. J Orthop Sci. 2009;14:68–75.PubMedCrossRef
35.
Zurück zum Zitat Popoola OO, Yao JQ, Johnson TS, Blanchard CR. Wear, delamination, and fatigue resistance of melt-annealed highly crosslinked UHMWPE cruciate-retaining knee inserts under activities of daily living. J Orthop Res. 2010;28:1120–1126.PubMedCrossRef Popoola OO, Yao JQ, Johnson TS, Blanchard CR. Wear, delamination, and fatigue resistance of melt-annealed highly crosslinked UHMWPE cruciate-retaining knee inserts under activities of daily living. J Orthop Res. 2010;28:1120–1126.PubMedCrossRef
36.
Zurück zum Zitat Popov VL. Contact Mechanics and Friction: Physical Principles and Applications. Berlin, Germany: Springer Verlag; 2010.CrossRef Popov VL. Contact Mechanics and Friction: Physical Principles and Applications. Berlin, Germany: Springer Verlag; 2010.CrossRef
37.
Zurück zum Zitat Puloski SK, McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB. Tibial post wear in posterior stabilized total knee arthroplasty—an unrecognized source of polyethylene debris. J Bone Joint Surg Am. 2001;83:390–397.PubMed Puloski SK, McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB. Tibial post wear in posterior stabilized total knee arthroplasty—an unrecognized source of polyethylene debris. J Bone Joint Surg Am. 2001;83:390–397.PubMed
38.
Zurück zum Zitat Revell PA. The combined role of wear particles, macrophages and lymphocytes in the loosening of total joint prostheses. J R Soc Interface. 2008;5:1263–1278.PubMedCrossRef Revell PA. The combined role of wear particles, macrophages and lymphocytes in the loosening of total joint prostheses. J R Soc Interface. 2008;5:1263–1278.PubMedCrossRef
39.
Zurück zum Zitat Robertsson O, Dunbar M, Pehrsson T, Knutson K, Lidgren L. Patient satisfaction after knee arthroplasty: a report on 27,372 knees operated on between 1981 and 1995 in Sweden. Acta Orthop Scand. 2000;71:262–267.PubMedCrossRef Robertsson O, Dunbar M, Pehrsson T, Knutson K, Lidgren L. Patient satisfaction after knee arthroplasty: a report on 27,372 knees operated on between 1981 and 1995 in Sweden. Acta Orthop Scand. 2000;71:262–267.PubMedCrossRef
40.
Zurück zum Zitat Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM. Insall Award Paper. Why are total knee arthroplasties failing today? Clin Orthop Relat Res. 2002;404:7–13.PubMedCrossRef Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM. Insall Award Paper. Why are total knee arthroplasties failing today? Clin Orthop Relat Res. 2002;404:7–13.PubMedCrossRef
41.
Zurück zum Zitat Surace MF, Berzins A, Urban RM, Jacobs JJ, Berger RA, Natarajan RN, Andriacchi TP, Galante JO. Coventry Award Paper. Backsurface wear and deformation in polyethylene tibial inserts retrieved postmortem. Clin Orthop Relat Res. 2002;404:14–23.PubMedCrossRef Surace MF, Berzins A, Urban RM, Jacobs JJ, Berger RA, Natarajan RN, Andriacchi TP, Galante JO. Coventry Award Paper. Backsurface wear and deformation in polyethylene tibial inserts retrieved postmortem. Clin Orthop Relat Res. 2002;404:14–23.PubMedCrossRef
42.
Zurück zum Zitat Toohey KS, Blanchet TA, Heckelman DD. Effect of accelerated aging conditions on resultant depth-dependent oxidation and wear resistance of UHMWPE joint replacement bearing materials. Wear. 2003;255:1076–1084.CrossRef Toohey KS, Blanchet TA, Heckelman DD. Effect of accelerated aging conditions on resultant depth-dependent oxidation and wear resistance of UHMWPE joint replacement bearing materials. Wear. 2003;255:1076–1084.CrossRef
43.
Zurück zum Zitat Wang A, Stark C, Dumbleton JH. Mechanistic and morphological origins of ultra-high molecular weight polyethylene wear debris in total joint replacement prostheses. Proc Inst Mech Eng H. 1996;210:141–155.PubMedCrossRef Wang A, Stark C, Dumbleton JH. Mechanistic and morphological origins of ultra-high molecular weight polyethylene wear debris in total joint replacement prostheses. Proc Inst Mech Eng H. 1996;210:141–155.PubMedCrossRef
44.
Zurück zum Zitat Wasielewski RC. The causes of insert backside wear in total knee arthroplasty. Clin Orthop Relat Res. 2002;404:232–246.PubMedCrossRef Wasielewski RC. The causes of insert backside wear in total knee arthroplasty. Clin Orthop Relat Res. 2002;404:232–246.PubMedCrossRef
45.
Zurück zum Zitat Wasielewski RC, Galante JO, Leighty RM, Natarajan RN, Rosenberg AG. Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty. Clin Orthop Relat Res. 1994;299:31–43.PubMed Wasielewski RC, Galante JO, Leighty RM, Natarajan RN, Rosenberg AG. Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty. Clin Orthop Relat Res. 1994;299:31–43.PubMed
46.
Zurück zum Zitat Wasielewski RC, Parks N, Williams I, Surprenant H, Collier JP, Engh G. Tibial insert undersurface as a contributing source of polyethylene wear debris. Clin Orthop Relat Res. 1997;345:53–59.PubMedCrossRef Wasielewski RC, Parks N, Williams I, Surprenant H, Collier JP, Engh G. Tibial insert undersurface as a contributing source of polyethylene wear debris. Clin Orthop Relat Res. 1997;345:53–59.PubMedCrossRef
47.
Zurück zum Zitat Willie BM, Foot LJ, Prall MW, Bloebaum RD. Surface damage analysis of retrieved highly crosslinked polyethylene tibial components after short-term implantation. J Biomed Mater Res B Appl Biomater. 2008;85:114–124.PubMed Willie BM, Foot LJ, Prall MW, Bloebaum RD. Surface damage analysis of retrieved highly crosslinked polyethylene tibial components after short-term implantation. J Biomed Mater Res B Appl Biomater. 2008;85:114–124.PubMed
48.
Zurück zum Zitat Wimmer MA, Andriacchi TP. Tractive forces during rolling motion of the knee: implications for wear in total knee replacement. J Biomech. 1997;30:131–137.PubMedCrossRef Wimmer MA, Andriacchi TP. Tractive forces during rolling motion of the knee: implications for wear in total knee replacement. J Biomech. 1997;30:131–137.PubMedCrossRef
49.
Zurück zum Zitat Wimmer MA, Andriacchi TP, Natarajan RN, Loos J, Karlhuber M, Petermann J, Schneider E, Rosenberg AG. A striated pattern of wear in ultrahigh-molecular-weight polyethylene components of Miller-Galante total knee arthroplasty. J Arthroplasty. 1998;13:8–16.PubMedCrossRef Wimmer MA, Andriacchi TP, Natarajan RN, Loos J, Karlhuber M, Petermann J, Schneider E, Rosenberg AG. A striated pattern of wear in ultrahigh-molecular-weight polyethylene components of Miller-Galante total knee arthroplasty. J Arthroplasty. 1998;13:8–16.PubMedCrossRef
50.
Zurück zum Zitat Wrona M, Mayor MB, Collier JP, Jensen RE. The correlation between fusion defects and damage in tibial polyethylene bearings. Clin Orthop Relat Res. 1994;299:92–103.PubMed Wrona M, Mayor MB, Collier JP, Jensen RE. The correlation between fusion defects and damage in tibial polyethylene bearings. Clin Orthop Relat Res. 1994;299:92–103.PubMed
51.
Zurück zum Zitat Yao JQ, Blanchard CR, Lu X, Laurent MP, Johnson TS, Gilbertson LN, Swarts DF, Crowninshield RD. Improved resistance to wear, delamination, and posterior loading fatigue damage of electron beam irradiated, melt-annealed, highly crosslinked UHMWPE knee inserts. In: Kurtz SM, Gsell RA, Martell J, eds. Standard Technical Publication 1445: Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene for Joint Replacements. West Conshohocken, PA: ASTM International; 2004:59–72.CrossRef Yao JQ, Blanchard CR, Lu X, Laurent MP, Johnson TS, Gilbertson LN, Swarts DF, Crowninshield RD. Improved resistance to wear, delamination, and posterior loading fatigue damage of electron beam irradiated, melt-annealed, highly crosslinked UHMWPE knee inserts. In: Kurtz SM, Gsell RA, Martell J, eds. Standard Technical Publication 1445: Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene for Joint Replacements. West Conshohocken, PA: ASTM International; 2004:59–72.CrossRef
Metadaten
Titel
Surface Damage Versus Tibial Polyethylene Insert Conformity: A Retrieval Study
verfasst von
Markus A. Wimmer, PhD
Michel P. Laurent, PhD
Jeannie D. Haman, PhD
Joshua J. Jacobs, MD
Jorge O. Galante, MD
Publikationsdatum
01.07.2012
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 7/2012
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-012-2274-y

Weitere Artikel der Ausgabe 7/2012

Clinical Orthopaedics and Related Research® 7/2012 Zur Ausgabe

Orthopaedic Case of the Month

Femur Mass in a 43-Year-Old Woman

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.