Understanding Dry Powder Inhalers: Key Technical and Patient Preference Attributes

verfasst von: Mark L. Levy, Will Carroll, José L. Izquierdo Alonso, Claus Keller, Federico Lavorini, Lauri Lehtimäki

Erschienen in: Advances in Therapy | Ausgabe 10/2019

Abstract

Inhalable medications for patients with asthma and chronic obstructive pulmonary disease (COPD) can be confusing even for health care professionals because of the multitude of available devices each with different operating principles. Dry powder inhalers (DPI) are a valuable option for almost all of the patients with asthma or COPD. Based on recorded patient inspiratory profiles, the peak inspiratory flow requirement of 30 L min−1 of high-resistance devices does not usually pose any practical limitations for the patients. Suboptimal adherence and errors in device handling are common and require continuous checking and patient education in order to avoid these pitfalls of all inhalation therapy. The aim of this opinion paper is to describe the working principles of DPIs and to summarise their key properties in order to help prescribing the correct inhaler for each patient.
Funding: Orion Pharma.
Hinweise

Enhanced Digital Features

To view enhanced digital features for this article go to https://​doi.​org/​10.​6084/​m9.​figshare.​9577574.

Different Types of Inhalers

The history of inhaler technology goes back almost 250 years. The first time the word inhaler was used was related to the ‘Mudge’ Inhaler in 1778 [1]. In modern inhalation therapy, there are four main device types. The first devices resembling nebulisers were introduced in the 1860s, metered dose inhalers (MDI) were developed in the 1950s, dry powder inhalers (DPI) in the 1980s and soft mist inhalers (SMI) after the year 2000.
In 2006, Chrystyn et al. [2] proposed five qualities which are required for an ‘ideal inhaler’. Firstly, the emitted dose must have appropriate aerodynamic characteristics to be as much as possible independent of a patient’s characteristics and clinical condition. Secondly, delivery of the drug must be safe and effective, thus maximising the therapeutic effects locally, while minimising systemic effects. Thirdly, the inhaler must be easy and intuitive to use. Fourthly, the device must be appealing to the patient to promote adherence and proper use. And lastly, ideally, a single device type should include all of the patient’s required inhalable treatments.
Since 2006, there has been an increased recognition of global environmental threats. There has therefore been increasing pressure to include environmental factors in the ideal features. To limit the depletion of the ozone layer, the permitted level of chlorofluorocarbons (CFC) produced was reduced from 150% of calculated levels of production and consumption to zero between 1986 and 2010 by the Montreal Protocol [3]. In response to these changes, the pharmaceutical industry increased its focus on the development of dry powder inhalers [4]. While pressurised MDIs (pMDIs) remain the most commonly prescribed type of inhaler [5], even in the post-CFC era, they still pose an environmental hazard [6], as the modern propellants (hydrofluorocarbons and hydrofluoroalkanes) are roughly 1300 times more potent than CO2 as greenhouse gases. In fact, the carbon footprint of pMDIs has been reported to be over 100-fold greater than that of DPIs [7].
The properties of each device type, including pros and cons, are outlined in Table 1: pMDIs facilitate patient-independent aerosolisation, but require sufficient coordination upon actuation; nebulisers allow delivery during normal breathing patterns, but are cumbersome to carry and need cleaning and servicing; SMIs are portable, but like pMDIs handling and co-ordination for dosing may be difficult for some subjects in comparison with DPIs; finally, DPIs are convenient and lightweight but rely upon patient inhalation technique to aerosolise the drug powder [8]. Taking environmental considerations into account, DPIs, being portable and easy to use, offer an appealing option for pulmonary drug delivery. Choosing the right device for the right patient is not as straightforward as it might first appear. Myths and misconceptions persist—particularly with DPIs—and debunking these will help clinicians pick the best device for their patient. For these reasons, this opinion paper aims to discuss the benefits and drawbacks of DPIs, from both a device and a patient’s perspective.
Table 1
Comparison of characteristics of different types of inhalers
 
DPI
pMDI
BA-MDI
SMI
Nebuliser
No spacer
With spacer
Medicine stored in
Dry powder
Suspension/solution
Suspension/solution
Suspension/solution
Solution
Solution
Propellant required
No
Yes
Yes
Yes
No
No
Built in dose countera
Yes
No/yes
No/yes
No/yes
No/yes
No
Handheld/portable
Yes
Yes
Yes/no
Yes
Yes
No
Single handed use and manual dexterity in loading and using the device
No/Yes
Yes
No
Yes
Yes
No
Breath-actuated
Yes
No
No
Yes
No
No
Co-ordination between actuation and inhalation required
No
Yes
No
No
Yes
No
Inhalation
Fast and deep (forceful inhalation from the very beginning, 4–5 s)
Slow and deep (4–5 s)
Slow and deep (4–5 s)
Slow and deep (4–5 s)
Slow and deep (4–5 s)
Normal breathing
Time needed for one dose delivery
Short
Short
Short
Short
Short
Long
DPI dry powder inhaler, BA-MDI breath-actuated MDI, pMDI pressurised metered dose inhaler, SMI soft mist inhaler
aHighly appreciated by the patients, hence increasingly incorporated and enhanced in devices
This article is based on previously conducted studies and does not contain any studies with human participants or animals performed by any of the authors.

From Powder to Aerosol

Despite their apparent simplicity, dry powder inhalers are sophisticated devices. In order to deliver active drug to the respiratory tract, the user must inhale through the device. This inhalation provides energy which breaks up the compacted drug powder, a process called de-agglomeration, and transports this de-agglomerated drug into the lung [9]. Most DPIs contain a micronised drug blended with carrier particles such as lactose that prevent aggregation and provide sufficient flowability. To enhance the aerosolisation of the drug particles, fine lactose particles are often added to the carrier lactose to saturate high-energy binding sites on the carrier lactose and form easily detachable aggregates with the drug particles. Both mechanisms contribute to higher fine particle dose, which is defined as the dose able to penetrate deep into the lungs [1013].
When the patient activates the DPI and inhales, airflow through the device creates shear forces and airflow turbulence. As the powder is emitted from the inhaler, the drug particles separate from the carrier particles and are carried deep into the lungs, while the larger carrier particles end up in the oropharynx and are swallowed/cleared [8]. Thus, the level of deposition into the lungs is determined by a combination of two factors: the patient’s inspiratory effort and the device’s properties (e.g. shape of the mouth piece, flow resistance and flow) [14] [15]. To achieve efficient aerosol dispersion, drug formulations have been developed in conjunction with varying DPI design elements, such as swirl chambers, grids and lacunas, to subsequently present varying degrees of airflow resistance. Hence, each DPI results in different pressure-drop and flow-rate characteristics. In general, the flow rate though a DPI is proportional to the square root of the pressure drop that the patient develops across it. The constant of proportionality is termed the device resistance [16].
Counterintuitively, higher resistance is often an advantage with DPIs. With the exception of the Turbuhaler (AstraZeneca), medium-to-high-resistance inhalers require lower patient inspiratory flow rates for optimal performance (Fig. 1) [1722]. Generally, higher-resistance devices generate more turbulent flow, more effectively converting the work done by the patient to energy available for de-agglomeration of the drug particles [2325]. In contrast, the Turbuhaler uses a built-in cyclone to de-agglomerate the formulation, and therefore has a minimum threshold for inspiratory flow rate to effectively aerosolise the drug particles [26]. In breath-actuated DPI devices, a certain pre-determined threshold in inspiratory flow rate has to be achieved before the drug is released from the inhaler. This prevents drug release at low flow rates that would not be high enough to allow effective de-agglomeration and would therefore result in low pulmonary deposition. Devices at the lower end of the resistance spectrum use less manipulation of flow patterns through the inhaler, and therefore rely more on the patient generating a flow rate that achieves sufficient de-agglomeration. With these principles in mind, patients with impaired pulmonary function could benefit from devices with higher resistance [21].
European and US pharmacopoeia recommend the use of a standardised pressure drop of 4 kPa for in vitro characterisation of DPI products. As this was originally intended as a method in quality control in the context of batch release, it is ill suited to produce data relevant to performance with patients. Recent publications are instead reporting the use of fixed ranges of inspiratory flow rates or patient population-specific percentiles when trying to simulate real-life conditions [18, 20, 27]. Haidl et al. [28] have suggested criteria for a successful inhalation manoeuvre, namely sufficient device-specific inhalation flow rate, flow acceleration, inhalation volume and inhaled volume after actuation. Based on this review, the peak inspiratory flow rate (PIF) of 30 L min−1 is sufficient for various DPIs of different operating principles, e.g. Easyhaler (reservoir; Orion), Handihaler (capsule; Boehringer–Ingelheim) and Ellipta (blister; GlaxoSmithKline). Patients’ inspiratory efforts have been shown to vary between age ranges (from paediatric to geriatric) and indications (from asthma to COPD) [22, 29, 30]. However, the expected PIF of 30 L min−1 seems to be well within reasonable limits. For example, when PIF was studied on the Easyhaler (a relatively high-resistance device compared with the Symbicort Turbuhaler) in adult and paediatric patients with asthma, as well as adults with COPD, the 10th, 50th and 90th percentiles for PIF were 44.7 L min−1, 61.1 L min−1 and 74.8 L min−1 [18], respectively. Further, the dosing has been shown to be consistent for PIF values of 28.3–74.8 L min−1 [31, 32]. (Fig. 2).

Inhaler Use in Practice: Pitfalls and Successes

Drugs are only effective if they can reach their site of action. In pulmonary delivery, the inhaler adds an extra step and a further potential barrier. Clinicians working with patients will be only too aware that there are many possible ways to use an inhaler device incorrectly. Education is vital and continuous re-education is also required. Recently, this has been addressed by third parties (e.g. https://​www.​rightbreathe.​com), who provide commercially independent training materials and educational videos on the use of inhalers, but more work is still needed. Melani et al. studied the user-associated mistakes made by 1664 adults with asthma and COPD who used MDIs and DPIs at home. The most common mistakes made with DPIs were exhaling into the inhaler before or after actuation and a lack of holding breath after the inhalation [33]. Sandler et al. showed that relatively few inhaler-naïve adults were able to use an inhaler [either the Spiromax (Teva), Easyhaler or Turbuhaler] correctly without training (just over a third), but following careful guidance by a healthcare provider (HCP), > 95% can master the technical aspects required for successful inhalation [34]. Patient education is required at diagnosis, at regular intervals, and certainly if the inhaler device is changed. Continuous checking and coaching also promotes the HCP–patient partnership and the active involvement of the patient and reduces unscheduled health care utilisation in asthma and COPD [35, 36]. Mistakes in the use of the inhaler have been shown to affect outcomes, but no conclusive evidence has been demonstrated using direct comparisons between inhaler types or devices [3740]. Taking into account that inhaler use has not improved during the last 40 years, substantial effort is needed from both HCPs and inhaler designers to rectify this issue [41].
The range of different medications available for administration via inhalers is vast and may be a source of confusion, for HCPs and especially for lay people. Patients with asthma are often satisfied with their reliever, short-acting beta agonist (SABA)-only treatment, because of their quick onset of action and immediate relief of symptoms. This may lead patients to believe that their asthma is controlled by their reliever therapy; however regular SABA-use is associated with poor outcomes [42, 43]. Dispensing three or more canisters of SABA a year is associated with increased hospital attendance [44], and more than 12 a year with risk of asthma deaths, [45] even in patients with minimal symptoms [46, 47]. This highlights the need to educate patients regarding the functions of the different drugs that they are prescribed. In the most recent 2019 Global Initiative for Asthma (GINA) report, a trend to simplify the treatment regimen can be seen [48]. The formerly used step 1 (SABA-only) treatment regimen was disregarded. Instead, with safety as the main priority, GINA now recommends that all patients with asthma should be prescribed inhaled steroids intermittently (in mild asthma) or continuously. GINA no longer recommends SABA as first-line treatment and recommends a combination of low-dose inhaled corticosteroids (ICS) and formoterol as needed in mild cases (less than 2 episodes of symptoms a month) and as an alternative to regular ICS in Step 2 of the GINA’s stepwise asthma management approach. This treatment strategy has been shown to reduce serious exacerbations when compared with SABA-only and to be non-inferior when compared with daily ICS and SABA as needed [49, 50]. Furthermore, a real-world comparison of as-needed ICS-formoterol in GINA’s stepwise asthma management approach, Step 1 and 2 patients with as-needed albuterol (salbutamol) demonstrated significantly lower asthma exacerbations in the as-needed ICS-formoterol group [51]. As steps 1 and 2, according to the new GINA strategy, can now be managed with single combined treatment as needed it will simplify the treatment. It also makes use of the incorrect practices by patients with infrequent symptoms not using their regular medication [52] and their urge for reliever therapy to control their disease. Therefore, the new GINA report takes a step towards a Maintenance and Reliever Therapy (MART) regimen, in which control and reliever therapy is combined in a single product. The concept of simplifying the treatment regimen has previously been employed in other fields of medicine. For example, when treating hypertension and dyslipidemia, medication may be combined to a single tablet to increase patient adherence [53]. It is worth mentioning that the studies cited in the new GINA report were conducted only with budesonide-formoterol combination therapy, and the clinical data have been extrapolated to also include other ICS. The recommendations are currently off-label and specifically include low-dose ICS-formoterol preparations to take advantage of the rapid action of formoterol.
In a cost–benefit analysis by Haahtela et al., Finnish asthma-related healthcare costs decreased by 14% between 1987 and 2013 despite a 3-fold increase in the number of diagnoses and an increase in medication costs [54]. As with the latest GINA guidelines, one of the cornerstones of the Finnish asthma program was to start inhaled corticosteroids as first-line treatment for all patients. The results from the program suggest that increased disease control not only improves patients’ quality of life but also provides wider societal benefits (e.g. economic benefits and improvements in integrated care) [55].
When compared with other types of inhalers, DPIs have a markedly different risk profile concerning the ambient environment. While they resist microbiological contamination extremely well, they are often susceptible to changes in their physical composition. For example, changes in crystallinity or agglomeration due to moisture may result in decreased performance of the product [56]. When applying for marketing approval, all inhalers must have evidence that they work well if stored according to their summary of product characteristics. Therefore, patient training should not be constrained to the use of an inhaler, but should also include handling and storing of the device.

Inhalers are for Patients

From the patient’s perspective, the requirements for an ideal inhaler can be summarised with six E’s (Fig. 3). An inhaler must be Effective; i.e. able to deliver the aerosols with particle sizes of respirable range and independent, as much as possible, of the inhalation volume and flow rate. This minimises the variation between patients, but also guarantees effective treatment when a single patient’s clinical parameters vary during treatment. Successful delivery straight to the site of action also allows the drug to act with minimal adverse effects. An inhaler must also be Efficient, i.e. easy to handle and able to be used with relatively few handling steps; these properties contribute to compliance and error-free handling. An Engaging design promotes patient satisfaction and adherence, leading to better treatment outcomes. Despite all the training and efforts of the HCPs, patients will continue to make errors with their inhalers; therefore, a well-designed inhaler is Error-tolerant and will minimise the effects of at least the most common user errors. An inhaler should also be Easy-to-teach; when teaching the use of an inhaler is easy and fast, the HCP can check the patient’s inhaler technique more easily and retrain the patient in using their device. Improvements in patients’ technical proficiency are likely to result in better treatment outcomes as the doses are more consistently delivered to the site of action. Chronic diseases as asthma and COPD are not only problematic for patients themselves, but provide a considerable societal burden; therefore, an ideal inhaler is also Easy-to-switch to when needed (e.g. at the population level, for cost-effectiveness reasons).
Patients of all ages have been reported to be particularly poorly adherent to inhaled medication [57, 58]. To address this problem, many companies have developed web-based information platforms (e.g. https://​www.​wehale.​life/​ [Orion pharma] and https://​www.​asthma.​com/​;Glaxo-Smith Kline) and inhaler-mounted smart devices, to give advice and feedback to the patient. Current guidelines on asthma and COPD promote the active involvement of the patient when choosing the inhaler. Satisfaction with an inhaler has been shown to relate to adherence and asthma control [59]. Differences in satisfaction between inhalers have been reported to arise from convenience-related factors such as portability, weight and size, and from usability factors such as preparation and use [60]. Several studies have reported therapeutic equivalence and successful switches between the inhalers (e.g. from Turbuhaler to Aerolizer, Spiromax, Easyhaler and Novolizer) [6164]. An example of a large scale switch in real-life practice comes from Norway, where the health authority (Statens legemiddelverk) included the Bufomix Easyhaler and Duoresp Spiromax in its switch regimen of DPIs containing budesonide/formoterol in Norwegian pharmacies, starting on July 1, 2018 [65].

Conclusions

In the market, there are many devices which are capable of delivering the drugs to the lungs of the patient. The myriad of devices and complex physics involved in the delivery makes it difficult for clinicians to choose the right device for their patients. For the majority of patients, there is a sufficiently effective combination of drug molecules available for use. However, poor adherence and incompetence with the device remain as frequent problems in inhalation therapy. Most patients are able to use nebulisers and pMDIs as well as high- and low-resistance DPIs. There is rarely a single best device for any given patient, and the choice should be made with the patient to take into account the parameters which are important to the actual user, such as size, portability or environmental factors. These properties valued by the patients may be medically irrelevant but through adherence could be the defining factor in the success of the treatment.

Acknowledgements

Funding

Funding for this study, the Rapid Service Fee, and the open access charge was provided by Orion Pharma, Orion Corporation.

Medical Writing and Editorial Assistance

Writing and editing assistance, including preparation of a draft manuscript under the direction and guidance of the authors, incorporation of author feedback, and manuscript submission, was provided by Dr. David Griffiths (Bioscript, UK), and Ville Vartiainen, MD, PhD, MSc (Orion Pharma, Orion Corporation, Finland).

Authorship

All named authors meet the International Committee of Medical Journal Editors (ICMJE) criteria for authorship for this article, had full access to the articles reviewed in this manuscript, take complete responsibility for the integrity and accuracy of the work as a whole, and have given their approval for this version to be published.

Disclosures

Mark L. Levy reports personal fees from Clement Clarke International, Teva, AstraZeneca, Chiesi, Conzorcio Futuro In Ricerca, Soar Beyond, Orion Pharma, Napp Pharmaceuticals, National Services for Health Improvement, Novartis Pharmaceuticals, Trudel Pharmaceuticals and GlaxoSmithKlein. He has had non-financial support from the Global Initiative on Asthma (GINA), and from Whole Systems Integrated Care (WSIC) North West London outside the submitted work. He was the clinical Lead for the UK National Review of Asthma Deaths (2011-2014). Will Carroll has received speaker fees and honoraria for advisory board work from GlaxoSmithKline, Novartis, Orion Pharma and Trudell Medical International. José L. Izquierdo Alonso has received honoraria for speaking engagements and consultancy fees from AstraZeneca, Bayer, Boehringer-Ingelheim, Chiesi, GlaxoSmithKline, Menarini, Novartis, Orion Pharma, Pfizer, Sandoz and Teva. Claus Keller has received a speaker honorarium from AstraZeneca, Berlin Chemie, GlaxoSmithKline and Boehringer Ingelheim. Federico Lavorini has received grants for reasearch or fees for speaking and advisory boards from AstraZeneca, Boehirnger Ingelheim, Chiesi, GlaxoSmithKline, Menarini International, Novartis, Orion Pharma, and Trudell Medical International. Lauri Lehtimäki has received honoraria from AstraZeneca, Boehringer-Ingelheim, Chiesi, GSK, Novartis, Mundipharma, Orion Pharma, Sanofi and Teva.

Compliance with Ethics Guidelines

This article does not contain any studies with human participants or animals performed by any of the authors.

Data Availability

Data sharing is not applicable to this article as no data sets were generated or analyzed during the current study.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://​creativecommons.​org/​licenses/​by-nc/​4.​0/​), which permits any noncommercial use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

e.Med Allgemeinmedizin

Kombi-Abonnement

Mit e.Med Allgemeinmedizin erhalten Sie Zugang zu allen CME-Fortbildungen und Premium-Inhalten der allgemeinmedizinischen Zeitschriften, inklusive einer gedruckten Allgemeinmedizin-Zeitschrift Ihrer Wahl.

Literatur
2.
Zurück zum Zitat Chrystyn H. Closer to an “ideal inhaler” with the Easyhaler: an innovative dry powder inhaler. Clin Drug Investig. 2006;26:175–83.CrossRefPubMed Chrystyn H. Closer to an “ideal inhaler” with the Easyhaler: an innovative dry powder inhaler. Clin Drug Investig. 2006;26:175–83.CrossRefPubMed
4.
5.
Zurück zum Zitat Lavorini F, Corrigan CJ, Barnes PJ, Dekhuijzen PRN, Levy ML, Pedersen S, et al. Retail sales of inhalation devices in European countries: so much for a global policy. Respir Med. 2011;105:1099–103.CrossRefPubMed Lavorini F, Corrigan CJ, Barnes PJ, Dekhuijzen PRN, Levy ML, Pedersen S, et al. Retail sales of inhalation devices in European countries: so much for a global policy. Respir Med. 2011;105:1099–103.CrossRefPubMed
6.
Zurück zum Zitat Sellers WFS. Asthma pressurised metered dose inhaler performance: propellant effect studies in delivery systems. Allergy, Asthma Clin Immunol. 2017;13:30.CrossRef Sellers WFS. Asthma pressurised metered dose inhaler performance: propellant effect studies in delivery systems. Allergy, Asthma Clin Immunol. 2017;13:30.CrossRef
8.
Zurück zum Zitat Laube BL, Janssens HM, de Jongh FHC, Devadason SG, Dhand R, Diot P, et al. What the pulmonary specialist should know about the new inhalation therapies. Eur Respir J. 2011;37:1308–417.CrossRefPubMed Laube BL, Janssens HM, de Jongh FHC, Devadason SG, Dhand R, Diot P, et al. What the pulmonary specialist should know about the new inhalation therapies. Eur Respir J. 2011;37:1308–417.CrossRefPubMed
9.
Zurück zum Zitat Frijlink HW, De Boer AH. Dry powder inhalers for pulmonary drug delivery. Expert Opin Drug Deliv. 2004;1:67–86.CrossRefPubMed Frijlink HW, De Boer AH. Dry powder inhalers for pulmonary drug delivery. Expert Opin Drug Deliv. 2004;1:67–86.CrossRefPubMed
10.
Zurück zum Zitat Kinnunen H, Hebbink G, Peters H, Shur J, Price R. An investigation into the effect of fine lactose particles on the fluidization behaviour and aerosolization performance of carrier-based dry powder inhaler formulations. AAPS PharmSciTech. 2014;15:898–909.CrossRefPubMedPubMedCentral Kinnunen H, Hebbink G, Peters H, Shur J, Price R. An investigation into the effect of fine lactose particles on the fluidization behaviour and aerosolization performance of carrier-based dry powder inhaler formulations. AAPS PharmSciTech. 2014;15:898–909.CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Ho R, Muresan AS, Hebbink GA, Heng JYY. Influence of fines on the surface energy heterogeneity of lactose for pulmonary drug delivery. Int J Pharm. 2010;388:88–94.CrossRefPubMed Ho R, Muresan AS, Hebbink GA, Heng JYY. Influence of fines on the surface energy heterogeneity of lactose for pulmonary drug delivery. Int J Pharm. 2010;388:88–94.CrossRefPubMed
12.
Zurück zum Zitat Zeng XM, Martin GP, Tee SK, Ghoush AA, Marriott C. Effects of particle size and adding sequence of fine lactose on the deposition of salbutamol sulphate from a dry powder formulation. Int J Pharm. 1999;182:133–44.CrossRefPubMed Zeng XM, Martin GP, Tee SK, Ghoush AA, Marriott C. Effects of particle size and adding sequence of fine lactose on the deposition of salbutamol sulphate from a dry powder formulation. Int J Pharm. 1999;182:133–44.CrossRefPubMed
13.
Zurück zum Zitat Adi H, Larson I, Chiou H, Young P, Traini D, Stewart P. Role of Agglomeration in the Dispersion of Salmeterol Xinafoate from Mixtures for Inhalation with Differing Drug to Fine Lactose Ratios. J Pharm Sci. 2008;97:3140–52.CrossRefPubMed Adi H, Larson I, Chiou H, Young P, Traini D, Stewart P. Role of Agglomeration in the Dispersion of Salmeterol Xinafoate from Mixtures for Inhalation with Differing Drug to Fine Lactose Ratios. J Pharm Sci. 2008;97:3140–52.CrossRefPubMed
14.
Zurück zum Zitat Telko MJ, Hickey AJ. Dry powder inhaler formulation. Respir Care. 2005;50:1209–27.PubMed Telko MJ, Hickey AJ. Dry powder inhaler formulation. Respir Care. 2005;50:1209–27.PubMed
15.
Zurück zum Zitat Weers J, Clark A. The Impact of Inspiratory Flow Rate on Drug Delivery to the Lungs with Dry Powder Inhalers. Pharm Res. United States. 2017;34:507–28. Weers J, Clark A. The Impact of Inspiratory Flow Rate on Drug Delivery to the Lungs with Dry Powder Inhalers. Pharm Res. United States. 2017;34:507–28.
16.
Zurück zum Zitat Azouz W, Chrystyn H. Clarifying the dilemmas about inhalation techniques for dry powder inhalers: integrating science with clinical practice. Prim Care Respir J. 2012;21:208–13.CrossRefPubMed Azouz W, Chrystyn H. Clarifying the dilemmas about inhalation techniques for dry powder inhalers: integrating science with clinical practice. Prim Care Respir J. 2012;21:208–13.CrossRefPubMed
18.
Zurück zum Zitat Malmberg LP, Everard ML, Haikarainen J, Lähelmä S. Evaluation of in vitro and in vivo flow rate dependency of budesonide/formoterol Easyhaler®. J Aerosol Med Pulm Drug Deliv. 2014;27:329–40.CrossRefPubMedPubMedCentral Malmberg LP, Everard ML, Haikarainen J, Lähelmä S. Evaluation of in vitro and in vivo flow rate dependency of budesonide/formoterol Easyhaler®. J Aerosol Med Pulm Drug Deliv. 2014;27:329–40.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Delvadia R, Hindle M, Longest PW, Byron PR. In vitro tests for aerosol deposition II: IVIVCs for different dry powder inhalers in normal adults. J Aerosol Med Pulm Drug Deliv. 2013;26:138–44.CrossRefPubMed Delvadia R, Hindle M, Longest PW, Byron PR. In vitro tests for aerosol deposition II: IVIVCs for different dry powder inhalers in normal adults. J Aerosol Med Pulm Drug Deliv. 2013;26:138–44.CrossRefPubMed
20.
Zurück zum Zitat Buttini F, Brambilla G, Copelli D, Sisti V, Balducci AG, Bettini R, et al. Effect of flow rate on in vitro aerodynamic performance of NEXThaler® in comparison with diskus® and Turbohaler® dry powder inhalers. J Aerosol Med Pulm Drug Deliv. 2016;29:167–78.CrossRefPubMedPubMedCentral Buttini F, Brambilla G, Copelli D, Sisti V, Balducci AG, Bettini R, et al. Effect of flow rate on in vitro aerodynamic performance of NEXThaler® in comparison with diskus® and Turbohaler® dry powder inhalers. J Aerosol Med Pulm Drug Deliv. 2016;29:167–78.CrossRefPubMedPubMedCentral
22.
Zurück zum Zitat Ghosh S, Ohar JA, Drummond MB. Peak Inspiratory Flow Rate in Chronic Obstructive Pulmonary Disease: implications for Dry Powder Inhalers. J Aerosol Med Pulm Drug Deliv. 2017;30:381–7.CrossRefPubMedPubMedCentral Ghosh S, Ohar JA, Drummond MB. Peak Inspiratory Flow Rate in Chronic Obstructive Pulmonary Disease: implications for Dry Powder Inhalers. J Aerosol Med Pulm Drug Deliv. 2017;30:381–7.CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Voss A, Finlay WH. Deagglomeration of dry powder pharmaceutical aerosols. Int J Pharm. 2002;248:39–50CrossRefPubMed Voss A, Finlay WH. Deagglomeration of dry powder pharmaceutical aerosols. Int J Pharm. 2002;248:39–50CrossRefPubMed
24.
Zurück zum Zitat Mäkelä M, Chrystyn H, Lavorini F, Canonica GW. Back to Basics in Asthma and COPD: optimising the Patient Journey. Eur Med J Respir. 2018;6:47–54. Mäkelä M, Chrystyn H, Lavorini F, Canonica GW. Back to Basics in Asthma and COPD: optimising the Patient Journey. Eur Med J Respir. 2018;6:47–54.
25.
Zurück zum Zitat Clark AR, Hollingworth AM. The relationship between powder inhaler resistance and peak inspiratory conditions in healthy volunteers–implications for in vitro testing. J Aerosol Med. 1993;6:99–110.CrossRefPubMed Clark AR, Hollingworth AM. The relationship between powder inhaler resistance and peak inspiratory conditions in healthy volunteers–implications for in vitro testing. J Aerosol Med. 1993;6:99–110.CrossRefPubMed
26.
Zurück zum Zitat Palander A, Mattila T, Karhu M, Muttonen E. In vitro Comparison of Three Salbutamol-Containing Multidose Dry Powder Inhalers. Clin Drug Investig. 2000;20:25–33.CrossRef Palander A, Mattila T, Karhu M, Muttonen E. In vitro Comparison of Three Salbutamol-Containing Multidose Dry Powder Inhalers. Clin Drug Investig. 2000;20:25–33.CrossRef
27.
Zurück zum Zitat Jõgi R, Lähelmä S, Vahteristo M, Happonen A, Haikarainen J. In vitro flow rate dependency of delivered dose and fine particle dose of salmeterol/fluticasone propionate easyhaler and seretide diskus with patient flow rates collected in a randomized controlled trial. J Aerosol Med Pulm Drug Deliv. 2018;31:jamp.2018.1463. Jõgi R, Lähelmä S, Vahteristo M, Happonen A, Haikarainen J. In vitro flow rate dependency of delivered dose and fine particle dose of salmeterol/fluticasone propionate easyhaler and seretide diskus with patient flow rates collected in a randomized controlled trial. J Aerosol Med Pulm Drug Deliv. 2018;31:jamp.2018.1463.
28.
Zurück zum Zitat Haidl P, Heindl S, Siemon K, Bernacka M, Cloes RM. Inhalation device requirements for patients’ inhalation maneuvers. Respir Med. 2016;118:65–75.CrossRefPubMed Haidl P, Heindl S, Siemon K, Bernacka M, Cloes RM. Inhalation device requirements for patients’ inhalation maneuvers. Respir Med. 2016;118:65–75.CrossRefPubMed
29.
Zurück zum Zitat Azouz W, Chetcuti P, Hosker HSR, Saralaya D, Stephenson J, Chrystyn H. The inhalation characteristics of patients when they use different dry powder inhalers. J Aerosol Med Pulm Drug Deliv. 2015;28:35–42.CrossRefPubMed Azouz W, Chetcuti P, Hosker HSR, Saralaya D, Stephenson J, Chrystyn H. The inhalation characteristics of patients when they use different dry powder inhalers. J Aerosol Med Pulm Drug Deliv. 2015;28:35–42.CrossRefPubMed
30.
Zurück zum Zitat Malmberg LP, Rytilä P, Happonen P, Haahtela T. Inspiratory flows through dry powder inhaler in chronic obstructive pulmonary disease: age and gender rather than severity matters. Int J Chron Obstruct Pulmon Dis. 2010;5:257–62.CrossRefPubMedPubMedCentral Malmberg LP, Rytilä P, Happonen P, Haahtela T. Inspiratory flows through dry powder inhaler in chronic obstructive pulmonary disease: age and gender rather than severity matters. Int J Chron Obstruct Pulmon Dis. 2010;5:257–62.CrossRefPubMedPubMedCentral
31.
Zurück zum Zitat Haikarainen J, Selroos O, Löytänä T, Metsärinne S, Happonen A, Rytilä P. Budesonide/Formoterol Easyhaler®: performance under simulated real-life conditions. Pulm Ther. 2017;3:125–38.CrossRef Haikarainen J, Selroos O, Löytänä T, Metsärinne S, Happonen A, Rytilä P. Budesonide/Formoterol Easyhaler®: performance under simulated real-life conditions. Pulm Ther. 2017;3:125–38.CrossRef
32.
Zurück zum Zitat Abadelah M, Hazim F, Chrystyn H, Bagherisadeghi G, Rahmoune H, Larhrib H. Effect of maximum inhalation flow and inhaled volume on formoterol drug deposition in vitro from an Easyhaler® dry powder inhaler. Eur J Pharm Sci. 2017;104:180–7.CrossRefPubMed Abadelah M, Hazim F, Chrystyn H, Bagherisadeghi G, Rahmoune H, Larhrib H. Effect of maximum inhalation flow and inhaled volume on formoterol drug deposition in vitro from an Easyhaler® dry powder inhaler. Eur J Pharm Sci. 2017;104:180–7.CrossRefPubMed
33.
Zurück zum Zitat Melani AS, Bonavia M, Cilenti V, Cinti C, Lodi M, Martucci P, et al. Inhaler mishandling remains common in real life and is associated with reduced disease control. Respir Med. 2011;105:930–8.CrossRefPubMed Melani AS, Bonavia M, Cilenti V, Cinti C, Lodi M, Martucci P, et al. Inhaler mishandling remains common in real life and is associated with reduced disease control. Respir Med. 2011;105:930–8.CrossRefPubMed
34.
Zurück zum Zitat Sandler N, Holländer J, Långström D, Santtila P, Saukkonen A, Torvinen S. Evaluation of inhaler handling-errors, inhaler perception and preference with Spiromax, Easyhaler and Turbuhaler devices among healthy Finnish volunteers: a single site, single visit crossover study (Finhaler). BMJ Open Respir Res. 2016;3(1):e000119.CrossRefPubMedPubMedCentral Sandler N, Holländer J, Långström D, Santtila P, Saukkonen A, Torvinen S. Evaluation of inhaler handling-errors, inhaler perception and preference with Spiromax, Easyhaler and Turbuhaler devices among healthy Finnish volunteers: a single site, single visit crossover study (Finhaler). BMJ Open Respir Res. 2016;3(1):e000119.CrossRefPubMedPubMedCentral
35.
Zurück zum Zitat Takemura M, Kobayashi M, Kimura K, Mitsui K, Masui H, Koyama M, et al. Repeated instruction on inhalation technique improves adherence to the therapeutic regimen in asthma. J Asthma. 2010;47:202–8.CrossRefPubMed Takemura M, Kobayashi M, Kimura K, Mitsui K, Masui H, Koyama M, et al. Repeated instruction on inhalation technique improves adherence to the therapeutic regimen in asthma. J Asthma. 2010;47:202–8.CrossRefPubMed
36.
Zurück zum Zitat Takemura M, Mitsui K, Itotani R, Ishitoko M, Suzuki S, Matsumoto M, et al. Relationships between repeated instruction on inhalation therapy, medication adherence, and health status in chronic obstructive pulmonary disease. Int J Chron Obstruct Pulmon Dis. 2011;6:97–104.CrossRefPubMedPubMedCentral Takemura M, Mitsui K, Itotani R, Ishitoko M, Suzuki S, Matsumoto M, et al. Relationships between repeated instruction on inhalation therapy, medication adherence, and health status in chronic obstructive pulmonary disease. Int J Chron Obstruct Pulmon Dis. 2011;6:97–104.CrossRefPubMedPubMedCentral
37.
Zurück zum Zitat Price DB, Román-Rodríguez M, McQueen RB, Bosnic-Anticevich S, Carter V, Gruffydd-Jones K, et al. Inhaler Errors in the CRITIKAL Study: type, frequency, and association with asthma outcomes. J Allergy Clin Immunol Pract. 2017;5:1071–81.e9.CrossRefPubMed Price DB, Román-Rodríguez M, McQueen RB, Bosnic-Anticevich S, Carter V, Gruffydd-Jones K, et al. Inhaler Errors in the CRITIKAL Study: type, frequency, and association with asthma outcomes. J Allergy Clin Immunol Pract. 2017;5:1071–81.e9.CrossRefPubMed
38.
Zurück zum Zitat Price D, Bosnic-Anticevich S, Briggs A, Chrystyn H, Rand C, Inhaler Error Steering Committee, et al. Inhaler competence in asthma: common errors, barriers to use and recommended solutions. Respir Med. 2013;107:37–46.CrossRefPubMed Price D, Bosnic-Anticevich S, Briggs A, Chrystyn H, Rand C, Inhaler Error Steering Committee, et al. Inhaler competence in asthma: common errors, barriers to use and recommended solutions. Respir Med. 2013;107:37–46.CrossRefPubMed
39.
Zurück zum Zitat Chrystyn H, van der Palen J, Sharma R, Barnes N, Delafont B, Mahajan A, et al. Device errors in asthma and COPD: systematic literature review and meta-analysis. NPJ Prim care Respir Med. 2017;27:22.CrossRefPubMedPubMedCentral Chrystyn H, van der Palen J, Sharma R, Barnes N, Delafont B, Mahajan A, et al. Device errors in asthma and COPD: systematic literature review and meta-analysis. NPJ Prim care Respir Med. 2017;27:22.CrossRefPubMedPubMedCentral
40.
Zurück zum Zitat Lavorini F, Magnan A, Dubus JC, Voshaar T, Corbetta L, Broeders M, et al. Effect of incorrect use of dry powder inhalers on management of patients with asthma and COPD. Respir Med. 2008;102:593–604.CrossRefPubMed Lavorini F, Magnan A, Dubus JC, Voshaar T, Corbetta L, Broeders M, et al. Effect of incorrect use of dry powder inhalers on management of patients with asthma and COPD. Respir Med. 2008;102:593–604.CrossRefPubMed
41.
Zurück zum Zitat Sanchis J, Gich I, Pedersen S, Aerosol Drug Management Improvement Team (ADMIT). Systematic review of errors in inhaler use: has patient technique improved over time? Chest. 2016;150:394–406.CrossRefPubMed Sanchis J, Gich I, Pedersen S, Aerosol Drug Management Improvement Team (ADMIT). Systematic review of errors in inhaler use: has patient technique improved over time? Chest. 2016;150:394–406.CrossRefPubMed
42.
Zurück zum Zitat Hancox RJ, Cowan JO, Flannery EM, Herbison GP, McLachlan CR, Taylor DR. Bronchodilator tolerance and rebound bronchoconstriction during regular inhaled beta-agonist treatment. Respir Med. 2000;94:767–71.CrossRefPubMed Hancox RJ, Cowan JO, Flannery EM, Herbison GP, McLachlan CR, Taylor DR. Bronchodilator tolerance and rebound bronchoconstriction during regular inhaled beta-agonist treatment. Respir Med. 2000;94:767–71.CrossRefPubMed
43.
Zurück zum Zitat Aldridge RE, Hancox RJ, Robin Taylor D, Cowan JO, Winn MC, Frampton CM, et al. Effects of terbutaline and budesonide on sputum cells and bronchial hyperresponsiveness in asthma. Am J Respir Crit Care Med. 2000;161:1459–64.CrossRefPubMed Aldridge RE, Hancox RJ, Robin Taylor D, Cowan JO, Winn MC, Frampton CM, et al. Effects of terbutaline and budesonide on sputum cells and bronchial hyperresponsiveness in asthma. Am J Respir Crit Care Med. 2000;161:1459–64.CrossRefPubMed
44.
Zurück zum Zitat Stanford RH, Shah MB, D’Souza AO, Dhamane AD, Schatz M. Short-acting beta-agonist use and its ability to predict future asthma-related outcomes. Ann Allergy Asthma Immunol. 2012;109:403–7.CrossRefPubMed Stanford RH, Shah MB, D’Souza AO, Dhamane AD, Schatz M. Short-acting beta-agonist use and its ability to predict future asthma-related outcomes. Ann Allergy Asthma Immunol. 2012;109:403–7.CrossRefPubMed
45.
Zurück zum Zitat Suissa S, Ernst P, Boivin JF, Horwitz RI, Habbick B, Cockroft D, et al. A cohort analysis of excess mortality in asthma and the use of inhaled beta-agonists. Am J Respir Crit Care Med. 1994;149:604–10.CrossRefPubMed Suissa S, Ernst P, Boivin JF, Horwitz RI, Habbick B, Cockroft D, et al. A cohort analysis of excess mortality in asthma and the use of inhaled beta-agonists. Am J Respir Crit Care Med. 1994;149:604–10.CrossRefPubMed
47.
Zurück zum Zitat Dusser D, Montani D, Chanez P, de Blic J, Delacourt C, Deschildre A, et al. Mild asthma: an expert review on epidemiology, clinical characteristics and treatment recommendations. Allergy. 2007;62:591–604.CrossRefPubMed Dusser D, Montani D, Chanez P, de Blic J, Delacourt C, Deschildre A, et al. Mild asthma: an expert review on epidemiology, clinical characteristics and treatment recommendations. Allergy. 2007;62:591–604.CrossRefPubMed
49.
Zurück zum Zitat O’Byrne PM, FitzGerald JM, Bateman ED, Barnes PJ, Zhong N, Keen C, Jorup C, Lamarca R, Ivanov S, Reddel HK. Inhaled combined budesonide–formoterol as needed in mild asthma. New Engl J Med. 2018;378(20):1865–76.CrossRefPubMed O’Byrne PM, FitzGerald JM, Bateman ED, Barnes PJ, Zhong N, Keen C, Jorup C, Lamarca R, Ivanov S, Reddel HK. Inhaled combined budesonide–formoterol as needed in mild asthma. New Engl J Med. 2018;378(20):1865–76.CrossRefPubMed
50.
Zurück zum Zitat Bateman E, Reddel H, O’Byrne P, Barnes P, Zhong N, Keen C, et al. As-needed budesonide-formoterol versus maintenance budesonide in mild asthma. N Engl J Med. 2018;378:1877–87.CrossRefPubMed Bateman E, Reddel H, O’Byrne P, Barnes P, Zhong N, Keen C, et al. As-needed budesonide-formoterol versus maintenance budesonide in mild asthma. N Engl J Med. 2018;378:1877–87.CrossRefPubMed
51.
Zurück zum Zitat Beasley R, Holliday M, Reddel HK, Braithwaite I, Ebmeier S, Hancox RJ, et al. Controlled trial of budesonide-formoterol as needed for mild asthma. N Engl J Med. 2019;380:2020–30.CrossRefPubMed Beasley R, Holliday M, Reddel HK, Braithwaite I, Ebmeier S, Hancox RJ, et al. Controlled trial of budesonide-formoterol as needed for mild asthma. N Engl J Med. 2019;380:2020–30.CrossRefPubMed
52.
Zurück zum Zitat Price D, Fletcher M, van der Molen T. Asthma control and management in 8,000 European patients: the REcognise Asthma and LInk to Symptoms and Experience (REALISE) survey. NPJ Prim care Respir Med. 2014;24:14009.CrossRefPubMedPubMedCentral Price D, Fletcher M, van der Molen T. Asthma control and management in 8,000 European patients: the REcognise Asthma and LInk to Symptoms and Experience (REALISE) survey. NPJ Prim care Respir Med. 2014;24:14009.CrossRefPubMedPubMedCentral
53.
Zurück zum Zitat Schulz M, Krueger K, Schuessel K, Friedland K, Laufs U, Mueller WE, et al. Medication adherence and persistence according to different antihypertensive drug classes: a retrospective cohort study of 255,500 patients. Int J Cardiol. 2016;220:668–76.CrossRefPubMed Schulz M, Krueger K, Schuessel K, Friedland K, Laufs U, Mueller WE, et al. Medication adherence and persistence according to different antihypertensive drug classes: a retrospective cohort study of 255,500 patients. Int J Cardiol. 2016;220:668–76.CrossRefPubMed
54.
Zurück zum Zitat Haahtela T, Herse F, Karjalainen J, Klaukka T, Linna M, Leskelä R-L, et al. The Finnish experience to save asthma costs by improving care in 1987-2013. J Allergy Clin Immunol. 2017;139(408–414):e2. Haahtela T, Herse F, Karjalainen J, Klaukka T, Linna M, Leskelä R-L, et al. The Finnish experience to save asthma costs by improving care in 1987-2013. J Allergy Clin Immunol. 2017;139(408–414):e2.
55.
Zurück zum Zitat Burki TK. Asthma control: learning from Finland's success. Lancet Respir Med. 2019;7:207–8.CrossRefPubMed Burki TK. Asthma control: learning from Finland's success. Lancet Respir Med. 2019;7:207–8.CrossRefPubMed
56.
Zurück zum Zitat Janson C, Lööf T, Telg G, Stratelis G, Nilsson F. Difference in resistance to humidity between commonly used dry powder inhalers: an in vitro study. NPJ Prim Care Respir Med. 2016;26:16053.CrossRefPubMedPubMedCentral Janson C, Lööf T, Telg G, Stratelis G, Nilsson F. Difference in resistance to humidity between commonly used dry powder inhalers: an in vitro study. NPJ Prim Care Respir Med. 2016;26:16053.CrossRefPubMedPubMedCentral
57.
Zurück zum Zitat Sulaiman I, Cushen B, Greene G, Seheult J, Seow D, Rawat F, et al. Objective assessment of adherence to inhalers by patients with chronic obstructive pulmonary disease. Am J Respir Crit Care Med. 2017;195:1333–43.CrossRefPubMed Sulaiman I, Cushen B, Greene G, Seheult J, Seow D, Rawat F, et al. Objective assessment of adherence to inhalers by patients with chronic obstructive pulmonary disease. Am J Respir Crit Care Med. 2017;195:1333–43.CrossRefPubMed
58.
Zurück zum Zitat Papi A, Ryan D, Soriano JB, Chrystyn H, Bjermer L, Rodriguez-Roisin R, et al. Relationship of inhaled corticosteroid adherence to asthma exacerbations in patients with moderate-to-severe asthma. J Allergy Clin Immunol Pract. 2018;6:1989–1998.e3.CrossRefPubMed Papi A, Ryan D, Soriano JB, Chrystyn H, Bjermer L, Rodriguez-Roisin R, et al. Relationship of inhaled corticosteroid adherence to asthma exacerbations in patients with moderate-to-severe asthma. J Allergy Clin Immunol Pract. 2018;6:1989–1998.e3.CrossRefPubMed
59.
Zurück zum Zitat Plaza V, Giner J, Calle M, Rytilä P, Campo C, Ribó P, et al. Impact of patient satisfaction with his or her inhaler on adherence and asthma control. Allergy Asthma Proc. 2018;39:437–45.CrossRefPubMedPubMedCentral Plaza V, Giner J, Calle M, Rytilä P, Campo C, Ribó P, et al. Impact of patient satisfaction with his or her inhaler on adherence and asthma control. Allergy Asthma Proc. 2018;39:437–45.CrossRefPubMedPubMedCentral
60.
Zurück zum Zitat Valero A, Ribó P, Maíz L, Barbero E, Calle M, Campo C, et al. Asthma patient satisfaction with different dry powder inhalers. Expert Rev Respir Med. 2019;13:133–8.CrossRefPubMed Valero A, Ribó P, Maíz L, Barbero E, Calle M, Campo C, et al. Asthma patient satisfaction with different dry powder inhalers. Expert Rev Respir Med. 2019;13:133–8.CrossRefPubMed
61.
Zurück zum Zitat Lötvall J, Mellén A, Arvidsson P, Palmqvist M, Radielovic P, Kottakis J, et al. Similar bronchodilation with formoterol delivered by aerolizer or turbuhaler. Can Respir J. 1999;6:412–6.CrossRefPubMed Lötvall J, Mellén A, Arvidsson P, Palmqvist M, Radielovic P, Kottakis J, et al. Similar bronchodilation with formoterol delivered by aerolizer or turbuhaler. Can Respir J. 1999;6:412–6.CrossRefPubMed
62.
Zurück zum Zitat Virchow JC, Rodriguez-Roisin R, Papi A, Shah TP, Gopalan G. A randomized, double-blinded, double-dummy efficacy and safety study of budesonide–formoterol Spiromax® compared to budesonide–formoterol Turbuhaler® in adults and adolescents with persistent asthma. BMC Pulm Med. 2016;16:42.CrossRefPubMedPubMedCentral Virchow JC, Rodriguez-Roisin R, Papi A, Shah TP, Gopalan G. A randomized, double-blinded, double-dummy efficacy and safety study of budesonide–formoterol Spiromax® compared to budesonide–formoterol Turbuhaler® in adults and adolescents with persistent asthma. BMC Pulm Med. 2016;16:42.CrossRefPubMedPubMedCentral
63.
Zurück zum Zitat Zetterström O, Lähelmä S, Keski-Karhu J, Silvasti M, Östling-Kulling E, Ahonen A. Salbutamol via Easyhaler®is at least as effective as salbutamol via Turbuhaler®in the treatment of histamine-induced bronchoconstriction. Respir Med. 2000;94:1097–102.CrossRefPubMed Zetterström O, Lähelmä S, Keski-Karhu J, Silvasti M, Östling-Kulling E, Ahonen A. Salbutamol via Easyhaler®is at least as effective as salbutamol via Turbuhaler®in the treatment of histamine-induced bronchoconstriction. Respir Med. 2000;94:1097–102.CrossRefPubMed
64.
Zurück zum Zitat Chuchalin AG, Kremer H-J, Metzenauer P, O’Keefe E, Hermann R. Clinical equivalence trial on budesonide delivered either by the novolizer® multidose dry powder inhaler or the turbuhaler® in asthmatic patients. Respiration. 2002;69:502–8.CrossRefPubMed Chuchalin AG, Kremer H-J, Metzenauer P, O’Keefe E, Hermann R. Clinical equivalence trial on budesonide delivered either by the novolizer® multidose dry powder inhaler or the turbuhaler® in asthmatic patients. Respiration. 2002;69:502–8.CrossRefPubMed
65.
Zurück zum Zitat Statens legemiddelverk. Inhalasjonpulver på byttelisten – vedtak gyldig. 2018. Statens legemiddelverk. Inhalasjonpulver på byttelisten – vedtak gyldig. 2018.
Metadaten
Titel
Understanding Dry Powder Inhalers: Key Technical and Patient Preference Attributes
verfasst von
Mark L. Levy
Will Carroll
José L. Izquierdo Alonso
Claus Keller
Federico Lavorini
Lauri Lehtimäki
Publikationsdatum
02.09.2019
Verlag
Springer Healthcare
Erschienen in
Advances in Therapy / Ausgabe 10/2019
Print ISSN: 0741-238X
Elektronische ISSN: 1865-8652
DOI
https://doi.org/10.1007/s12325-019-01066-6

Weitere Artikel der Ausgabe 10/2019

Advances in Therapy 10/2019 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.