Comparison of the prognostic value of myocardial perfusion imaging using a CZT-SPECT camera with a conventional anger camera
verfasst von:
Ronaldo Lima, MD, PhD, Thais Peclat, MD, Thalita Soares, BMBS, Caio Ferreira, BMBS, Ana Carolina Souza, BMBS, Gabriel Camargo, MD
Erschienen in:
Journal of Nuclear Cardiology
|
Ausgabe 1/2017
Einloggen, um Zugang zu erhalten
Abstract
Background
Recent studies have shown that myocardial perfusion imaging (MPI) in cadmium-zinc-telluride (CZT) cameras allow faster exams with less radiation dose but there are little data comparing its prognosis information with that of dedicated cardiac Na-I SPECT cameras
Objective
The objective of this study is to compare the prognostic value of MPI using an ultrafast protocol with low radiation dose in a CZT-SPECT and a traditional one.
Methods
Group 1 was submitted to a two-day MIBI protocol in a conventional camera, and group 2 was submitted to a 1-day MIBI protocol in CZT camera. MPI were classified as normal or abnormal, and perfusion scores were calculated. Propensity score matching methods were performed
Results
3554 patients were followed during 33±8 months. Groups 1 and 2 had similar distribution of age, gender, body mass index, risk factors, previous revascularization, and use of pharmacological stress. Group 1 had more abnormal scans, higher scores than group 2. Annualized hard events rate was higher in group 1 with normal scans but frequency of revascularization was similar to normal group 2. Patients with abnormal scans had similar event rates in both groups
Conclusion
New protocol of MPI in CZT-SPECT showed similar prognostic results to those obtained in dedicated cardiac Na-I SPECT camera, with lower prevalence of hard events in patients with normal scan.