Open Access 01.01.2014 | Workshop presentation

Clinical evaluation of accelerated cardiac cine imaging using iterative k-t-sparse SENSE

verfasst von: Bradley D Allen, Maria Carr, Michael O Zenge, Michaela Schmidt, Mariappan S Nadar, Bruce S Spottiswoode, Jeremy D Collins, James C Carr

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Sonderheft 1/2014

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Background

We sought to quantitatively and qualitatively evaluate a clinical implementation of accelerated MR acquisitions using iterative segmented k-t-sparse Cartesian SENSE balanced steady-state free-precession (SSFP) cinegraphic imaging.

Methods

IRB approval was obtained. Twenty patients (age: 54.8 ± 14 years, M:F = 15:5) undergoing non-emergent CMR assessment for myocardial pathology were consecutively recruited. Nine healthy volunteers (age: 44.3 ± 14 years, M:F = 6:3) were also imaged. CMR was performed at 1.5T (MAGNETOM Aera, Siemens Healthcare, Erlangen, Germany). The examination included acquisition of standard segmented SSFP (iPAT2) (GRAPPA accel factor 2, TR 40 msec, 2.1 × 2.1 × 10 mm3) cine and two accelerated segmented SSFP acquisitions (TPAT accel factor 4, TR 37.7 msec, 2.1 × 2.1 × 6 mm3), one with an investigational prototype inline iterative k-t-sparse SENSE reconstruction with L1 regularization along one spatial and temporal dimension (TPAT4i) (1)and the other with conventional SENSE reconstruction (TPAT4). Each technique was used to acquire a three- (3Ch), four-chamber (4Ch), and short axis (SA) series in identical slice positions (Figure 1), with SA coverage of the entire left ventricle (LV) with 10 mm interslice gaps. Individual slice scan times were recorded. Quantitative LV functional analysis was performed. A reviewer blinded to acquisition type scored images for overall image quality, noise, and artifacts using a 5-point Likert scale. Continuous variables were compared between groups using a paired t-test, and ordinal variables were compared using a Wilcoxon signed-rank test.

Results

In a combined analysis of patients and volunteers, there was no significant difference between LV ejection fraction between iPAT2 and TPAT4i (p = 0.48) or TPAT4 (p = 0.34). The iPAT2 technique found marginally lower end-diastolic volume (159 ± 54 ml vs. 169 ± 54 ml, p = 0.01), end-systolic volume (75 ± 33 ml vs. 80 ± 31 ml, p = 0.04), stroke volume (80 ± 35 ml vs. 89 ± 31 ml, p = 0.03) compared to TPAT4i. Single-slice scan times in both TPAT4i(3.29 ± 0.6 sec) and TPAT4 (3.0 ± 0.6 sec) acquisitions were significantly shorter relative to iPAT2 (8.4 ± 1.7 sec, p < 0.001 for both). Qualitative review showed significantly higher image quality and lower noise using both iPAT2 and TPAT4i acquisitions relative to TPAT4 (quality: p < 0.001, noise: p < 0.001). The iPAT2 acquisition was scored higher than TPAT4i in quality (4.8 ± 0.4 vs. 4.6 ± 0.6, p = 0.005), noise (4.9 ± 0.3 vs. 4.6 ± 0.6, p < 0.001), and artifact (4.8 ± 0.4 vs. 4.6 ± 0.6, p = 0.04), although the magnitude of difference is relatively small. (Figure 2)

Conclusions

Acceleration using iterative k-t-sparse SENSE techniques can be successfully applied in CMR to reduce scan times by > 50% while maintaining diagnostic image quality and quantitative accuracy in LV systolic function assessment.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Liu J, et al: ISMRM 20th Annual Meeting. 2012, Melbourne, Australia, 4249- Liu J, et al: ISMRM 20th Annual Meeting. 2012, Melbourne, Australia, 4249-
Metadaten
Titel
Clinical evaluation of accelerated cardiac cine imaging using iterative k-t-sparse SENSE
verfasst von
Bradley D Allen
Maria Carr
Michael O Zenge
Michaela Schmidt
Mariappan S Nadar
Bruce S Spottiswoode
Jeremy D Collins
James C Carr
Publikationsdatum
01.01.2014
Verlag
BioMed Central
DOI
https://doi.org/10.1186/1532-429X-16-S1-W13

Weitere Artikel der Sonderheft 1/2014

Journal of Cardiovascular Magnetic Resonance 1/2014 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.