Here we show that accuracy and precision of kinetic parameters related to atherosclerotic plaque microvascularization can be improved by sampling the AIF at faster time resolution than the plaque enhancement. This feature is implemented in the novel SHILO sequence by interleaving the low spatial/high temporal resolution AIF images with high spatial/low temporal resolution tissue acquisitions. A model AIF [
43] with time resolution 0.1 s over a 5 minutes interval was used to compute tissue uptake curves using a modified Tofts-Kermode [
44] model and a range of kinetic parameters, representative of vessel wall perfusion indices reported in the literature [
45,
46]. Kinetic parameters used in the simulation were
v
p
, the fraction of intra-vascular volume;
K
trans
, expressing the inflow of contrast agent from the plasma to the tissue compartment;
v
e
, the fraction of extra-vascular extra-cellular space. The parameter
K
ep
expressing the backflow of contrast agent from the tissue to the plasma compartment was calculated as
K
trans
/
ve
. 150 different combination of kinetic parameters were used, with
v
p
values 0.001, 0.005, 0.02, 0.05, 0.1,
K
trans
values 0.02, 0.06, 0.1, 0.15, 0.2 min
-1, and
v
e
values 0.1, 0.2, 0.3, 0.4, 0.6, 0.8. Two different sampling schemes were then implemented while fitting for kinetic parameters [
16]: 1) sampling of both AIF and tissue curves at the same increasingly lower rate (referred to as “same time resolution”, STR) to emulate deriving the AIF curve from the same tissue frames 2) sampling of the tissue curves only (referred to as “different time resolution”, DTR) at an increasingly lower rate to emulate the SHILO dual imaging or other dual bolus techniques. These different levels of under-sampling simulate situations in which an increasing number of tissue slices is acquired using both sampling schemes. In the STR scheme both AIF and tissue curves were under-sampled at rates 1.6, 3.2, 6.4, 12.8, 25.6 and 51.2 s (corresponding to multiples from 1 to 32 of the AIF time resolution in SHILO). In the DTR scheme, tissue curves were under-sampled at the same rate used for the STR case. As for AIF, two DTR schemes were investigated with the AIF sampled at 1.6 s and 0.8 s. The first case (1.6 s AIF sampling) corresponds to the temporal resolution of the AIF in the proposed SHILO acquisition. The second case correspond to a hypothetical test bolus acquisition, where the time resolution of the AIF is equal to the acquisition time of one single shot image as mentioned in the above description of the SHILO sequence, without interleaving the acquisition of segments of tissue slices. Comparing both these DTR schemes allows assessing the impact on kinetic parameters estimation of interleaving AIF and tissue acquisitions, such as is proposed in the SHILO sequence, as opposed to the corresponding test-bolus acquisition. For each parameter set and sampling scheme, the simulation was repeated 6 times, each time shifting the sampling grid by one sixth of the sampling interval to account for a variable temporal registration of the image sampling and bolus passage. Finally, kinetic parameters were estimated with the same modified Tofts-Kermode model used to simulate tissue curves [
44], using non linear least squares fitting routines implemented in Matlab (MathWorks, Natick, MA). The average and standard deviation of estimated kinetic parameters across the 6 variable temporal registration cases were recorded. Relative errors, normalized by the true parameter value were recorded and the magnitude taken. The median of relative magnitude errors and standard deviation over the span of kinetic parameters simulated was plotted as a function of tissue slice temporal sampling resolution.