Reproducibility of quantitative analysis of aortic 4D flow data

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Sonderheft 1/2013

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Background

3D cine phase-contrast CMR ("4D Flow") permits quantitative assessment of anomalous alterations of aortic blood flow. Two hemodynamic parameters that have been used for this purpose is the wall shear stress (WSS), which is known to regulate endothelial cell function, and the normalized flow displacement from the vessel center, which was recently shown to correlate with increased growth rates of ascending aortic dilation [1, 2]. Analysis of these hemodynamic parameters requires that a user 1) positions a 2D plane of interest in the volumetric 4D Flow dataset and 2) delineates the contour of the vascular lumen in this 2D plane. We set out to assess the reproducibility of 4D Flow-based estimation of WSS and normalized flow displacement at these two critical levels of user-interaction. Furthermore, we assessed which of the parameters correlate best with aortic growth.

Methods

25 patients previously studied with 4D Flow imaging were included. Previously reported data on interval aortic growth was available for each subject [2].
CMR velocity data from a plane perpendicular to the ascending aorta just distal to the sinotubular junction was collected independently by two blinded reviewers, and then separately segmented by two blinded observers (Figure 1). Subsequently, the following parameters were calculated: normalized flow displacement, maximum peak-systolic WSS, maximum of systole-averaged WSS, mean peak-systolic WSS, minimal peak-systolic WSS. Normalized flow displacement was calculated as in reference [1]. WSS calculation was performed with propriety software (Flow Tool) [4]. Reproducibility analysis and correlation with interval aortic growth were performed.

Results

Inter-observer correlations with regards to plane selection and contour delineation are reported in Table 1. For the contour delineation, correlation coefficients were 0.97-0.98 for normalized displacement and 0.78-0.96 for the WSS parameters. For the plane positioning, these correlation coefficients were 0.91-0.93 for normalized displacement and 0.51-0.85 for the WSS parameters. Flow displacement best correlated with interval aortic growth (r = 0.65). The range of WSS parameters did not correlate well with growth (r < 0.15).
Table 1
Reproducibility analysis: inter-observer correlations
  
Mean peak-systolic WSS
Max peak-systolic WSS
Min peak-systolic WSS
Max systole-averaged WSS
Min systole-averaged WSS
Normalized Flow Displacement
Observerplanes#1 versus Observerplanes#2
Contours #1
0.71
0.67
0.47
0.74
0.71
0.91
 
Contours #2
0.82
0.85
0.51
0.85
0.70
0.93
Observercontours#1 versus Observercontours#2
Planes #1
0.93
0.87
0.84
0.87
0.88
0.98
 
Planes #2
0.96
0.91
0.81
0.90
0.78
0.97

Conclusions

Normalized flow displacement is a reproducible hemodynamic marker that shows good correlation with interval aortic growth. Reproducibility of contour delineation for WSS analysis was good and in line with previous reports [3]. However, markedly lower reproducibility was found for the plane positioning step of the WSS analysis. Normalized flow displacement should be considered in future work aimed at identifying and risk-stratifying patients who are likely to develop clinically significant aortic dilation based on CMR-estimated hemodynamic parameters.

Funding

Covidien/Radiologic Society of North America Research Scholar Grant 2012-2014.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Metadaten
Titel
Reproducibility of quantitative analysis of aortic 4D flow data
Publikationsdatum
01.01.2013
DOI
https://doi.org/10.1186/1532-429X-15-S1-P126

Weitere Artikel der Sonderheft 1/2013

Journal of Cardiovascular Magnetic Resonance 1/2013 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.