Improved precision in SASHA T1 mapping with a variable flip angle readout

verfasst von: Kelvin Chow, Bruce S Spottiswoode, Joseph J Pagano, Richard B Thompson

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Sonderheft 1/2014

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Background

The SAturation-recovery single-SHot Acquisition (SASHA) T1 mapping sequence has excellent accuracy independent of T1, T2, heart rate, and flip angle [1], which are known dependencies of the more commonly used MOdified Look-Locker Inversion-recovery (MOLLI) sequence. However, SASHA has a greater T1 variability (poorer precision) compared to MOLLI. A two-parameter fit, with assumed ideal saturation, has been shown to improve precision compared to the standard three-parameter fit used for SASHA, but at the expense of introducing systematic errors [2]. We propose that a variable flip angle (VFA) readout will reduce these systematic errors and thereby allow the improved precision of a two-parameter fit while maintaining the accuracy of the three-parameter fit.

Methods

A VFA scheme was empirically designed with Bloch equation simulations to minimize two-parameter fit errors with SASHA data, consisting of scaling the prescribed flip angle for the first 45 pulses by sin(x) for π/90 < × < π/2. The first 5 data acquisitions in the pulse train were discarded, matching the number of dummy pulses with linear catalyzation in the standard SASHA sequence. SASHA, SASHA-VFA, and MOLLI T1 imaging was performed on 4 healthy volunteers (Siemens Aera 1.5T) on a mid-ventricular short-axis slice with typical bSSFP imaging readout parameters: 1.01/2.44 ms TE/TR, 8 mm slice thickness, 112 × 192 matrix size, 270 × 360 mm2 field of view, rate 2 GRAPPA with 24 in-place ACS reference lines, 78% phase resolution, and 7/8 partial Fourier for a total imaging duration of ~175 ms. SASHA datasets were acquired with 9 images having equally spaced TIs from 165-780 ms following BIR-4 saturation, plus a non-saturated image. Standard SASHA was acquired with 5 (dummy) linear catalyzation pulses and SASHA-VFA was acquired with sinusoidal scaling described above, both with a target flip angle of 70°. MOLLI data was acquired with a 5-(3)-3 configuration, 120 ms TI start, 80 ms TI increment, 35° flip angle, and a tan/tanh adiabatic inversion pulse [3]. T1 pixel map were generated and the mean and standard deviation calculated for an ROI enclosing the entire LV myocardium.

Results

Two-parameter SASHA overestimated myocardial T1 as compared to the three-parameter fit but with reduced variability (Table 1). Two-parameter SASHA-VFA showed similar mean T1 values to three-parameter SASHA and with substantially reduced T1 variability. Image artifacts from the bSSFP readout were consistently reduced with the SASHA-VFA sequence compared to the standard SASHA sequence, which may also contribute to the improved variability performance (Figure 1).
Table 1
Mean, standard deviation, and coefficient of variation of myocardial T1 values in 4 healthy subjects
 
Mean Myocardial T1 (ms)
Standard Deviation of Myocardial T1 (ms)
Coefficient of Variation of Myocardial T1 (%)
SASHA (3-parameter fit)
1165 ± 15
78 ± 12
6.9 ± 1.0
SASHA (2-parameter fit)
1177 ± 29
58 ± 5
4.9 ± 0.3
SASHA-VFA (2-parameter fit)
1163 ± 19
47 ± 5
4.1 ± 0.5
MOLLI
996 ± 12
43 ± 4
4.3 ± 0.3
Values are reported as mean ± standard deviation across subjects.

Conclusions

The SASHA sequence with VFA readout significantly reduces T1 variability and reduces image artifacts. The current study suggests that two-parameter SASHA-VFA maintains the accuracy of standard three-parameter SASHA with significantly reduced T1 variability, similar to the MOLLI sequence.

Funding

Canadian Institutes of Health Research, Women and Children's Health Research Institute, Alberta Innovates - Health Solutions.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Chow K, et al: MRM. 2013, doi:10.1002/mrm.24878 Chow K, et al: MRM. 2013, doi:10.1002/mrm.24878
2.
Zurück zum Zitat Kellman P, et al: ISMRM. 2013, 21: 1394- Kellman P, et al: ISMRM. 2013, 21: 1394-
3.
Zurück zum Zitat Kellman P, et al: MRM. 2013, doi:10.1002/mrm.24793 Kellman P, et al: MRM. 2013, doi:10.1002/mrm.24793
Metadaten
Titel
Improved precision in SASHA T1 mapping with a variable flip angle readout
verfasst von
Kelvin Chow
Bruce S Spottiswoode
Joseph J Pagano
Richard B Thompson
Publikationsdatum
01.01.2014
Verlag
BioMed Central
DOI
https://doi.org/10.1186/1532-429X-16-S1-M9

Weitere Artikel der Sonderheft 1/2014

Journal of Cardiovascular Magnetic Resonance 1/2014 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.