01.11.2015 | Original Article

Monocytes in Icelandic horses: comparison between automatic gating by the ADVIA 2120i, a manual gating method and microscopic differentiation

verfasst von: E. Leidinger, G. Prechtl, J. Figl, J. Leidinger

Erschienen in: Comparative Clinical Pathology | Ausgabe 6/2015

Einloggen, um Zugang zu erhalten

Abstract

Inspection of the ADVIA 2120i PEROX cytograms from healthy Icelandic horses revealed a mismatch of the automatic gating (A) for monocytes, resulting in a lower percentage. The raw data of 40 Icelandic horses were transformed to FCS files. Data were imported to WinMDI 2.9 (Purdue Cytometry Laboratories) and gated manually (M). In addition, a manual 200-cell differential (D) was performed. ANOVA was used to test for significant differences. RefVal 2.1 (Toulouse Veterinary School) was used to calculate the reference values by different methods. Significance level was set at P < 0.05. Monocytes were significantly higher by M compared to A and D, neutrophils were significantly lower for A compared to M, but not for A compared to D. Data were used to calculate the RI as follows: neutrophils (A) 37.7–65.6 % (1.7–4.6 × 109/L); neutrophils (M): 33.8–61.8 % (1.6–4.3 × 109/L); lymphocytes (A): 26.0–52.7 % (1.1–4.6 × 109/L), lymphocytes (M): 28.1–55.5 % (1.2–4.0 × 109/L); monocytes (A): 1.5–4.6 % (0.1–0.4 × 109/L); monocytes (M): 3.4–6.9 % (0.2–0.5 × 109/L). The ADVIA 2120i built-in export function and the use of WinMDI allow a very precise recalculation of cell populations. Icelandic horses reportedly have the lowest monocyte number of all horse breeds. This may be possibly caused by higher MPO activity and/or an inaccuracy of the algorithm used.
Literatur
Zurück zum Zitat Bauer N, Nakagawa J, Dunker C, Failing K, Moritz A (2012) Evaluation of the automated hematology analyzer Sysmex XT-2000iVTM compared to the ADVIA® 2120 for its use in dogs, cats, and horses. Part II: accuracy of leukocyte differential and reticulocyte count, impact of anticoagulant and sample aging. J Vet Diag Invest 24:74–89CrossRef Bauer N, Nakagawa J, Dunker C, Failing K, Moritz A (2012) Evaluation of the automated hematology analyzer Sysmex XT-2000iVTM compared to the ADVIA® 2120 for its use in dogs, cats, and horses. Part II: accuracy of leukocyte differential and reticulocyte count, impact of anticoagulant and sample aging. J Vet Diag Invest 24:74–89CrossRef
Zurück zum Zitat Björnsson GB, Sveinsson HJ (2007) The Icelandic horse. 2nd ed.: Edda Publishing, Reykjavik pp 20-33 Björnsson GB, Sveinsson HJ (2007) The Icelandic horse. 2nd ed.: Edda Publishing, Reykjavik pp 20-33
Zurück zum Zitat Clark P, Mogg TD, Tvedten H, Korcal D (2002) Artifactual changes in equine blood following storage, dedected using the Advia 120 hematology analyser. Vet Clin Pathol 31:90–94CrossRefPubMed Clark P, Mogg TD, Tvedten H, Korcal D (2002) Artifactual changes in equine blood following storage, dedected using the Advia 120 hematology analyser. Vet Clin Pathol 31:90–94CrossRefPubMed
Zurück zum Zitat Dura A (2005) Blutzellzählung und -differenzierung bei Pferd und Schwein mit dem Hämatologiesystem ADVIA 120. Gerätevalidierung und Softwareadaptation [Blood cell count and differentation in horses and swines. Instrument validation and software adaption]. Dissertation, Justus-Liebig-Universität, Giessen Dura A (2005) Blutzellzählung und -differenzierung bei Pferd und Schwein mit dem Hämatologiesystem ADVIA 120. Gerätevalidierung und Softwareadaptation [Blood cell count and differentation in horses and swines. Instrument validation and software adaption]. Dissertation, Justus-Liebig-Universität, Giessen
Zurück zum Zitat Flandrin G, Valensi F, Mazon C, Sassier, P. Ann Biol Clin (Paris) (1993) [The progresses in the determination of blood monocytosis and its perspextives of clinical application in pathology];51:787–791 Flandrin G, Valensi F, Mazon C, Sassier, P. Ann Biol Clin (Paris) (1993) [The progresses in the determination of blood monocytosis and its perspextives of clinical application in pathology];51:787–791
Zurück zum Zitat Geffré A, Concordet D, Braun J-P, Trumel C (2011) Reference value advisor: a new freeware set of macroinstructions to calculate reference intervals with Microsoft Excel. Vet Clin Pathol 40:107–112CrossRefPubMed Geffré A, Concordet D, Braun J-P, Trumel C (2011) Reference value advisor: a new freeware set of macroinstructions to calculate reference intervals with Microsoft Excel. Vet Clin Pathol 40:107–112CrossRefPubMed
Zurück zum Zitat Gieseler T (2012) Referenzwerte beim Pferd [Reference intervals in horses]. Dissertation, Vet Med Leipzig Gieseler T (2012) Referenzwerte beim Pferd [Reference intervals in horses]. Dissertation, Vet Med Leipzig
Zurück zum Zitat Giordano A, Rossi G, Pieralisi C, Paltrinieri S (2008) Evaluation of equine hemograms using the Advia 120 as compared with an impedance counter and manual differential count. Vet Clin Pathol 37:21–30CrossRefPubMed Giordano A, Rossi G, Pieralisi C, Paltrinieri S (2008) Evaluation of equine hemograms using the Advia 120 as compared with an impedance counter and manual differential count. Vet Clin Pathol 37:21–30CrossRefPubMed
Zurück zum Zitat Gorman ME, Villarroel A, Tornquist SJ, Flachsbart J, Warden A, Boeder L (2009) Comparison between manual and automated total nucleated cell counts using the ADVIA 120 for pleural and peritoneal fluid samples from dogs, cats, horses, and alpacas. Vet Clin Pathol 38:388–391CrossRefPubMed Gorman ME, Villarroel A, Tornquist SJ, Flachsbart J, Warden A, Boeder L (2009) Comparison between manual and automated total nucleated cell counts using the ADVIA 120 for pleural and peritoneal fluid samples from dogs, cats, horses, and alpacas. Vet Clin Pathol 38:388–391CrossRefPubMed
Zurück zum Zitat Klenner S, Richartz J, Bauer N, Moritz A (2010) Myeloperoxidase deficiency in dogs observed with the ADVIA120. Tierärztl Praxis Kleint 3:139–46 Klenner S, Richartz J, Bauer N, Moritz A (2010) Myeloperoxidase deficiency in dogs observed with the ADVIA120. Tierärztl Praxis Kleint 3:139–46
Zurück zum Zitat Lumsden JH, Rowe R, Mullen K (1980) Hematology and biochemistry reference values for the light horse. Can J Comp Med 44:32–42PubMedCentralPubMed Lumsden JH, Rowe R, Mullen K (1980) Hematology and biochemistry reference values for the light horse. Can J Comp Med 44:32–42PubMedCentralPubMed
Zurück zum Zitat Moritz A (2000) Der Einsatz lasergestützter Multiparameter-Hämatologiesysteme in der Veterinärmedizin [The use of laser-based multiparameter hematology systems in veterinary medicine]. Dissertation, Justus-Liebig-Universität, Giessen Moritz A (2000) Der Einsatz lasergestützter Multiparameter-Hämatologiesysteme in der Veterinärmedizin [The use of laser-based multiparameter hematology systems in veterinary medicine]. Dissertation, Justus-Liebig-Universität, Giessen
Zurück zum Zitat Rizzi TE, Meinkoth JH, Clinkenbeard KD (2010) Normal hematology of the dog. In: Weiss DJ, Wardrop KJ (eds) Schalm’s veterinary hematology, 6th edn. Blackwell, Ames, pp 799–810 Rizzi TE, Meinkoth JH, Clinkenbeard KD (2010) Normal hematology of the dog. In: Weiss DJ, Wardrop KJ (eds) Schalm’s veterinary hematology, 6th edn. Blackwell, Ames, pp 799–810
Zurück zum Zitat Rümke CL (1959) Variability of results in differential cell counts on blood smears. Triangle 4:154–158 Rümke CL (1959) Variability of results in differential cell counts on blood smears. Triangle 4:154–158
Zurück zum Zitat Ryseff JK (2012) Characterization of errors and predictability of disease using the ADVIA 120 hematology analyser leukocyte differential. Dissertation, Colorado State University Ryseff JK (2012) Characterization of errors and predictability of disease using the ADVIA 120 hematology analyser leukocyte differential. Dissertation, Colorado State University
Zurück zum Zitat Sahraneshin Samani F, Moore JK, Khosravani P, Ebrahimi M (2010) Features of free software packages in flow cytometry: a comparison between four non-commercial software sources. Cytotechnology. doi:10.1007/s10616-013-9609-8 Sahraneshin Samani F, Moore JK, Khosravani P, Ebrahimi M (2010) Features of free software packages in flow cytometry: a comparison between four non-commercial software sources. Cytotechnology. doi:10.​1007/​s10616-013-9609-8
Zurück zum Zitat Seiser M, Strasser A, Hofbauer B (2001) Der Einfluss von Alter und Geschlecht of diagnostisch wesentliche Blutparameter bei Islandpferden. [The influence of age and sex on significant haematological parameters in Icelandic horses]. Tierarztl Prax 29:324–331 Seiser M, Strasser A, Hofbauer B (2001) Der Einfluss von Alter und Geschlecht of diagnostisch wesentliche Blutparameter bei Islandpferden. [The influence of age and sex on significant haematological parameters in Icelandic horses]. Tierarztl Prax 29:324–331
Zurück zum Zitat Takasu M, NAGATANI N, Tozaki T, Kakoi H, Maeda M, Murase T, Mukoyama H (2013) Hematological and biochemical reference values for the endangered Kiso horse. J Equ Sci 24:75–78CrossRef Takasu M, NAGATANI N, Tozaki T, Kakoi H, Maeda M, Murase T, Mukoyama H (2013) Hematological and biochemical reference values for the endangered Kiso horse. J Equ Sci 24:75–78CrossRef
Zurück zum Zitat Tvedten H, Lilliehöok I, Nostell K, Edner A, Bröjer J (2012) Decrease in ADVIA 2120 MPXI and a monocyte identification error occur late after endotoxin treatment of horses. Seattle, WA: Proceedings 47th ASVCP Ann. Meeting. p10 Tvedten H, Lilliehöok I, Nostell K, Edner A, Bröjer J (2012) Decrease in ADVIA 2120 MPXI and a monocyte identification error occur late after endotoxin treatment of horses. Seattle, WA: Proceedings 47th ASVCP Ann. Meeting. p10
Zurück zum Zitat van der Meer W (2006) Blood cell morphology: controversies and alternatives. Dissertation, Radboud University, Nijmegen, The Netherlands van der Meer W (2006) Blood cell morphology: controversies and alternatives. Dissertation, Radboud University, Nijmegen, The Netherlands
Metadaten
Titel
Monocytes in Icelandic horses: comparison between automatic gating by the ADVIA 2120i, a manual gating method and microscopic differentiation
verfasst von
E. Leidinger
G. Prechtl
J. Figl
J. Leidinger
Publikationsdatum
01.11.2015
Verlag
Springer London
Erschienen in
Comparative Clinical Pathology / Ausgabe 6/2015
Print ISSN: 1618-5641
Elektronische ISSN: 1618-565X
DOI
https://doi.org/10.1007/s00580-015-2080-5

Weitere Artikel der Ausgabe 6/2015

Comparative Clinical Pathology 6/2015 Zur Ausgabe

Neu im Fachgebiet Pathologie

Molekularpathologische Untersuchungen im Wandel der Zeit

Open Access Biomarker Leitthema

Um auch an kleinen Gewebeproben zuverlässige und reproduzierbare Ergebnisse zu gewährleisten ist eine strenge Qualitätskontrolle in jedem Schritt des Arbeitsablaufs erforderlich. Eine nicht ordnungsgemäße Prüfung oder Behandlung des …

Vergleichende Pathologie in der onkologischen Forschung

Pathologie Leitthema

Die vergleichende experimentelle Pathologie („comparative experimental pathology“) ist ein Fachbereich an der Schnittstelle von Human- und Veterinärmedizin. Sie widmet sich der vergleichenden Erforschung von Gemeinsamkeiten und Unterschieden von …

Gastrointestinale Stromatumoren

Open Access GIST CME-Artikel

Gastrointestinale Stromatumoren (GIST) stellen seit über 20 Jahren ein Paradigma für die zielgerichtete Therapie mit Tyrosinkinaseinhibitoren dar. Eine elementare Voraussetzung für eine mögliche neoadjuvante oder adjuvante Behandlung bei …

Personalisierte Medizin in der Onkologie

Aufgrund des erheblichen technologischen Fortschritts in der molekularen und genetischen Diagnostik sowie zunehmender Erkenntnisse über die molekulare Pathogenese von Krankheiten hat in den letzten zwei Jahrzehnten ein grundlegender …