Open Access 01.12.2012 | Original Paper

Multicenter assessment of the reproducibility of volumetric radiofrequency-based intravascular ultrasound measurements in coronary lesions that were consecutively stented

verfasst von: Jennifer Huisman, Rasmus Egede, Adam Rdzanek, Dirk Böse, Raimund Erbel, Janusz Kochman, Lisette Okkels Jensen, Job van der Palen, Marc Hartmann, Gary S. Mintz, Clemens von Birgelen

Erschienen in: The International Journal of Cardiovascular Imaging | Ausgabe 8/2012

Abstract

To assess in a multicenter design the between-center reproducibility of volumetric virtual histology intravascular ultrasound (VH-IVUS) measurements with a semi-automated, computer-assisted contour detection system in coronary lesions that were consecutively stented. To evaluate the reproducibility of volumetric VH-IVUS measurements, experienced analysts of 4 European IVUS centers performed independent analyses (in total 8,052 cross-sectional analyses) to obtain volumetric data of 40 coronary segments (length 20.0 ± 0.3 mm) from target lesions prior to percutaneous intervention that were performed in the setting of stable (65%) or unstable angina pectoris (35%). Geometric and compositional VH-IVUS measurements were highly correlated for the different comparisons. Overall intraclass correlation for vessel, lumen, plaque volume and plaque burden was 0.99, 0.92, 0.96, and 0.83, respectively; for fibrous, fibro-lipidic, necrotic core and calcified volumes overall intraclass correlation was 0.96, 0.94, 0.98, and 0.99, respectively. Nevertheless, significant differences for both geometrical and compositional measurements were seen. Of the plaque components, fibrous tissue and necrotic core showed on average the highest measurement reproducibility. A central analysis for VH-IVUS multicenter studies of lesions prior to PCI should be pursued. Moreover, it may be problematical to pool VH-IVUS data of individual trials analyzed by independent centers.

Abkürzungen
IVUS: 

Intravascular ultrasound

VH: 

Virtual histology

PCI: 

Percutaneous coronary intervention

Multicenter assessment of the reproducibility of volumetric radiofrequency-based intravascular ultrasound measurements in coronary lesions that were consecutively stented

Jennifer Huisman1, Rasmus Egede2, Adam Rdzanek3, Dirk Böse4, Raimund Erbel4, Janusz Kochman3, Lisette Okkels Jensen2, Job van der Palen5, 6, Marc Hartmann1, Gary S. Mintz7 and Clemens von Birgelen1, 8  
The International Journal of Cardiovascular ImagingX-Ray Imaging, Echocardiography, Nuclear Cardiology Computed Tomography and Magnetic Resonance Imaging201228:11

DOI: 10.1007/s10554-012-0011-y

Received: 28 September 2011

Accepted: 2 January 2012

Published: 14 January 2012

Abstract

To assess in a multicenter design the between-center reproducibility of volumetric virtual histology intravascular ultrasound (VH-IVUS) measurements with a semi-automated, computer-assisted contour detection system in coronary lesions that were consecutively stented. To evaluate the reproducibility of volumetric VH-IVUS measurements, experienced analysts of 4 European IVUS centers performed independent analyses (in total 8,052 cross-sectional analyses) to obtain volumetric data of 40 coronary segments (length 20.0 ± 0.3 mm) from target lesions prior to percutaneous intervention that were performed in the setting of stable (65%) or unstable angina pectoris (35%). Geometric and compositional VH-IVUS measurements were highly correlated for the different comparisons. Overall intraclass correlation for vessel, lumen, plaque volume and plaque burden was 0.99, 0.92, 0.96, and 0.83, respectively; for fibrous, fibro-lipidic, necrotic core and calcified volumes overall intraclass correlation was 0.96, 0.94, 0.98, and 0.99, respectively. Nevertheless, significant differences for both geometrical and compositional measurements were seen. Of the plaque components, fibrous tissue and necrotic core showed on average the highest measurement reproducibility. A central analysis for VH-IVUS multicenter studies of lesions prior to PCI should be pursued. Moreover, it may be problematical to pool VH-IVUS data of individual trials analyzed by independent centers.

Keywords

Coronary artery disease Intravascular ultrasound Radiofrequency Virtual Histology Measurement reproducibility Percutaneous coronary interventions

Abbreviations

IVUS: 

Intravascular ultrasound

VH: 

Virtual histology

PCI: 

Percutaneous coronary intervention

Introduction

Virtual Histology intravascular ultrasound (VH-IVUS) allows measurements of both atherosclerotic plaque and vessel size and plaque composition in vivo [110]. Plaque composition of target lesions prior to percutaneous coronary intervention (PCI) may impact clinical outcome following PCI [1119]. A relatively large necrotic core volume in target lesions as assessed with VH-IVUS is an independent predictor of creatine kinase and troponin elevation after PCI [11, 15]. Microembolization of such plaque components may lead to myocardial injury during stent implantation [11, 14, 16, 18, 20]. Certain plaque components or phenotypes as assessed with VH-IVUS are related to no-reflow phenomenon post-PCI which may be triggered by such microembolization and adversely affects prognosis [14, 18]. Therefore, volumetric VH-IVUS assessment of target lesions prior to PCI may have the potential to identify lesions at particularly high risk for such complications and may help to tailor such procedures. However, an important pre-requisite for clinical VH-IVUS studies assessing such relations in advanced coronary lesions prior to PCI is an adequate measurement reproducibility [2].

We recently reported significant but small differences for VH-IVUS measurements of mild-to-moderately diseased coronary segments between different European IVUS centers [21]. But so far, no data is available on the between-center reproducibility of VH-IVUS measurements of advanced lesions prior to PCI. Therefore, in the present study we assessed the between-center reproducibility of volumetric VH-IVUS measurements of target lesions, that were consecutively stented, by comparing data from repeated analyses as performed by independent analysts at 4 independent IVUS centers in Europe.

Methods

Study population

In order to assess the between-center reproducibility of segmental volumetric VH-IVUS data analysis of diseased atherosclerotic human coronary arteries, we performed IVUS analyses in 20 mm-long segments of coronary lesions prior to PCI in patients with unstable angina pectoris or stable angina. Coronary segments had to meet the following inclusion criteria: no severe plaque calcification, no major vessel tortuosity, and/or major side-branches. In this multicenter study design, 4 experienced analysts independently performed VH-IVUS analyses of all segments included in the study. Each IVUS center was invited to submit IVUS pullbacks of 10 coronary segments obtained from patients in sinus rhythm in whom a clinically driven cardiac catheterization required IVUS assessment prior to PCI. Participating centers were Odense University Hospital, Odense, Denmark; Essen University Hospital, Essen, Germany; Central University Hospital, Warsaw, Poland; and Thoraxcentrum Twente, Enschede, The Netherlands.

IVUS data acquisition

Acquisition of IVUS data was performed in an ECG-gated way with commercially available phased-array IVUS catheters (Eagle Eye Gold 2.9F 20 MHz, Volcano Corporation, Rancho Cordova, CA, USA) with a dedicated console [22]. The IVUS transducer was advanced ≥10 mm distal to the most distal side-branch. Angiographic cine runs were performed to define the position of the IVUS transducer. After intracoronary injection of 200 μg nitroglycerin, a continuous pullback of the IVUS catheter was performed using a motorized pullback device at 0.5 mm/s (TrackBack-II, Volcano Corporation). IVUS image data were stored on digital video disk (DVD) for offline analysis.

VH-IVUS data analysis

Description of the technique and the validation of VH-IVUS have previously been published [1, 3, 4, 6]. In brief, tissue maps are reconstructed from spectral IVUS radiofrequency data and classify the plaque in four major components: calcium, fibro-lipidic, fibrous, and necrotic core. Tissue components were displayed according to a color code (white, light-green, green, and red, respectively).

Offline VH-IVUS analysis of all cases was performed by one experienced IVUS analyst per center (Fig. 1). The IVUS sequences were provided with a predefined region of interest (20 mm length) that contained a (part of a) target lesion for PCI. Segments were located between two adjacent side-branches and contained no major calcification that could have limited reliable detection of the external vascular boundary.
https://static-content.springer.com/image/art%3A10.1007%2Fs10554-012-0011-y/MediaObjects/10554_2012_11_Fig1_HTML.gif
Fig. 1

Image acquisition and data analysis. Motorized pullbacks at 0.5 mm/s (including image frame acquisition at the time of the R-wave peak) were performed by each IVUS center (AD). These pullback sequences (obtained from 10 patients) were exchanged in order to obtain a ‘pullback pool’ of 40 pullbacks, which each analyst independently analyzed (I). Automated contour detection and, if required, manual correction of the lumen and vessel border were performed on all frames using the VH-images and software (II). Volumetric VH-IVUS data of vessel and plaque geometry and plaque composition were automatically generated for the analyzed segment based on the application of Trapezium method (III)

By use of the VH-images and software, IVUS measurements of geometry and plaque composition were made. The lumen and vessel borders were traced using a semi-automated, computer-assisted contour detection program (Volcano pcVH software program version 2.2; Volcano); to exclude potential between version differences of the analysis software, all four centers used the latest version of the analysis software. The leading edge of the blood-intima acoustic interface and the leading edge of the media-adventitia interface were automatically detected to define the lumen and vessel borders, respectively, and manually corrected as required. Plaque & media was used as a measure of atherosclerotic plaque and was calculated as the difference between vessel and lumen cross-sectional area. For each coronary segment, volumetric data of the vessel, lumen, plaque and plaque burden, as well as fibrous, fibro-lipidic, necrotic core, and calcified tissue were obtained from the application of the trapezium method to the cross-sectional area measurements [21, 23]. The pcVH software program uses this method that considers the exact spacing between image frames to calculate volumes from the cross-sectional area data in the most accurate way. In addition, the relative amount of the four plaque components was calculated. To account for any differences in length, all volumetric data were normalized for 10 mm-long segments [21, 23].

In order to assess the between-center reproducibility of volumetric VH-IVUS analyses, comparisons were performed on a two-observer basis (i.e., center A vs. B, C and D respectively; center B vs. C and D respectively; and center C vs. D).

Statistical analysis

Statistical analysis was performed with SPSS 15.0 (SPSS Inc., Chicago, IL, USA). Dichotomous variables are presented as frequencies, while quantitative data are presented as mean ± 1SD (or ±1SEM). Quantitative data were compared by use of a paired t test after Normality of the data was confirmed. A two-sided P-value < 0.05 was considered as significant. According to Bland and Altman, the agreement between two measurements was assessed by determining the mean ± 2SD of the between measurement differences [24]. Intra-class correlations were calculated for comparisons at an individual level. An intra-class correlation >0.90 was considered good.

Results

Study population

A total of 40 coronary segments (6 right; 27 left anterior descending; and 7 left circumflex coronary arteries) were analyzed from 40 patients (age 64 ± 10 years, 75% were male) who were in sinus rhythm during the IVUS pullback. IVUS examination was performed in both patients with stable syndromes (65%) and unstable syndromes (35%). Eighty-eight percent of the patients was on statin therapy prior to the cardiac catheterization, and 35% was diabetics. There was no complication related to IVUS imaging.

VH-IVUS analysis and data

Independent analysis of each of the 40 coronary segments was performed by each of the four analysts and volumetric data was obtained to evaluate the reproducibility of volumetric VH-IVUS measurements (50 ± 14 frames/segment). This approach required the analysis of a total of 8,052 cross-sectional frames (2,013 frames per analyst). Manual correction of the automatic border detection was required in almost all frames for a total analysis time of 60 ± 6 min/segment. The length of the analyzed segments was 20.0 ± 0.3 mm (range: 19.1–20.6 mm). On average, the atherosclerotic segments contained predominantly fibrous (57.4%), fibro-lipidic (13.1%), and necrotic core (19.1%) tissue. Geometric and compositional IVUS data obtained by the analysts of the 4 different centers (A–D) are presented in Table 1.
Table 1

VH-IVUS data from four different centers (A–D)

 

A

B

C

D

ANOVA

Vessel geometrics

     

 Vessel volume (mm3)

162.8 ± 36.3

156.9 ± 37.7

163.1 ± 36.7

159.4 ± 36.4

0.9

 Lumen volume (mm3)

73.8 ± 16.5

74.3 ± 17.3

78.9 ± 18.6

69.8 ± 17.1

0.2

 Plaque volume (mm3)

89.0 ± 25.7

82.6 ± 26.9

84.3 ± 25.0

89.5 ± 26.1

0.7

 Plaque burden (%)

54.2 ± 6.8

52.0 ± 7.7

51.3 ± 7.6

55.8 ± 7.5

0.051*

Plaque composition

     

 Fibrous volume (mm3)

31.0 ± 16.1

27.9 ± 16.1

28.1 ± 14.9

31.7 ± 15.8

0.8

 Fibrous volume (%)

58.0 ± 9.9

56.8 ± 11.0

57.3 ± 10.6

57.6 ± 10.1

1.0

 Fibro-lipidic volume (mm3)

7.0 ± 4.9

6.3 ± 4.9

5.8 ± 4.2

7.3 ± 5.0

0.7

 Fibro-lipidic volume (%)

13.4 ± 9.1

13.1 ± 9.0

12.3 ± 8.2

13.7 ± 8.7

1.0

 Necrotic core volume (mm3)

10.0 ± 6.0

9.2 ± 6.0

9.6 ± 6.2

10.3 ± 6.6

1.0

 Necrotic core volume (%)

18.7 ± 9.0

19.2 ± 9.2

19.5 ± 9.0

18.7 ± 9.0

1.0

 Calcium volume (mm3)

4.9 ± 3.7

4.9 ± 3.7

5.0 ± 3.7

5.1 ± 3.9

1.0

 Calcium volume (%)

9.9 ± 7.7

10.9 ± 8.7

10.8 ± 8.3

9.9 ± 7.9

1.0

Values are normalized to 10 mm segment length and presented as mean ± standard deviation and ANOVA

* Significant difference could be explained by the comparison between center C and D (adjusted P = 0.048 (Holm-Bonferroni correction), the other comparisons were not significant

VH-IVUS measurement variability of plaque and vessel geometry

VH-IVUS measurements of vessel geometry were highly correlated for the different comparisons. Overall intra-class correlation and range of the individual between-center comparisons (within brackets) was for vessel, lumen, plaque volume and plaque burden 0.99 (0.98–1.00), 0.92 (0.84–0.97), 0.96 (0.93–0.98), and 0.83 (0.75–0.90), respectively. Although there were significant differences for vessel, lumen, and plaque volumes, these differences were only moderate in size (Tables 1, 2; Fig. 2). Plaque burden also showed significant differences for the comparisons between two centers. However, the differences were only moderate in size, ranging from 0.8 to 4.6%. Statistical analysis using Anova with Holm-Bonferroni correction did not show significant differences, except for plaque burden measurements of center C versus D (adjusted P-value 0.048), no significant differences were seen (P > 0.1, Table 1). The limits of agreement (i.e., 2 SD of mean difference) were relatively small and similar for the different comparisons; measurement reproducibility of vessel volume was particularly high (Fig. 3a, b). Measurement reproducibility was not influenced by the size of the measured volumes.
Table 2

Measurement differences of VH-IVUS data from four different centers (A–D)

 

A versus*

B versus*

C versus*

B

C

D

C

D

D

Vessel geometry

      

 Δ Vessel volume (mm3)

5.9 ± 0.8

P < 0.01

−0.3 ± 0.5

P = 0.5

3.4 ± 0.5

P < 0.01

−6.2 ± 0.8

P < 0.01

−2.5 ± 0.8

P < 0.01

3.8 ± 0.6

P < 0.01

 Δ Lumen volume (mm3)

−0.6 ± 0.7

P = 0.4

−5.1 ± 0.9

P < 0.01

3.9 ± 0.7

P < 0.01

−4.5 ± 0.9

P < 0.01

4.5 ± 0.8

P < 0.01

9.0 ± 0.9

P < 0.01

 Δ Plaquee volume (mm3)

6.5 ± 0.9

P < 0.01

4.8 ± 0.9

P < 0.01

−0.5 ± 0.8

P = 0.5

−1.7 ± 1.2

P = 0.2

−7.0 ± 1.2

P < 0.01

−5.3 ± 0.9

P < 0.01

 Δ Plaque burden (%)

2.2 ± 0.5

P < 0.01

3.0 ± 0.5

P < 0.01

−1.6 ± 0.5

P < 0.01

0.8 ± 0.6

P = 0.2

−3.8 ± 0.6

P < 0.01

−4.6 ± 0.5

P < 0.01

Plaque composition

      

 Δ Fibrous volume (mm3)

3.0 ± 0.6

P < 0.01

2.9 ± 0.5

P < 0.01

−0.7 ± 0.4

P = 0.1

−0.1 ± 0.7

P = 0.9

−3.7 ± 0.7

P < 0.01

−3.6 ± 0.5

P < 0.01

 Δ Fibrous volume (%)

1.2 ± 0.3

P < 0.01

0.7 ± 0.3

P < 0.01

0.4 ± 0.2

P = 0.02

−0.5 ± 0.2

P = 0.04

−0.8 ± 0.2

P < 0.01

−0.3 ± 0.2

P = 0.2

 Δ Fibro-lipidic volume (mm3)

0.8 ± 0.2

P < 0.01

1.2 ± 0.3

P < 0.01

−0.3 ± 0.2

P = 0.1

0.4 ± 0.2

P = 0.1

−1.0 ± 0.3

P < 0.01

−1.5 ± 0.2

P < 0.01

 Δ Fibro-lipidic volume (%)

0.3 ± 0.3

P = 0.3

1.0 ± 0.2

P < 0.01

−0.3 ± 0.2

P = 0.1

0.7 ± 0.3

P = 0.02

−0.7 ± 0.3

P = 0.05

−1.4 ± 0.2

P < 0.01

 Δ Necrotic core volume (mm3)

0.7 ± 0.2

P < 0.01

0.4 ± 0.2

P = 0.02

−0.4 ± 0.2

P = 0.08

−0.3 ± 0.2

P = 0.06

−1.1 ± 0.2

P < 0.01

−0.8 ± 0.2

P < 0.01

 Δ Necrotic core volume (%)

−0.5 ± 0.2

P = 0.05

−0.8 ± 0.2

P < 0.01

−0.00 ± 0.2

P = 1.0

−0.3 ± 0.2

P = 0.1

0.5 ± 0.2

P = 0.05

0.8 ± 0.2

P < 0.01

 Δ Calcium volume (mm3)

0.02 ± 0.1

P = 0.7

−0.1 ± 0.1

P = 0.3

−0.2 ± 0.1

P = 0.03

−0.1 ± 0.1

P = 0.2

−0.2 ± 0.1

P = 0.03

−0.1 ± 0.1

P = 0.2

 Δ Calcium volume (%)

−1.0 ± 0.2

P < 0.01

−1.0 ± 0.2

P < 0.01

−0.1 ± 0.1

P = 0.6

0.1 ± 0.2

P = 0.7

1.0 ± 0.2

P < 0.01

0.9 ± 0.2

P < 0.01

Values are normalized to 10 mm segment length and presented as mean ± standard error of the mean. * Two-sided paired t test

https://static-content.springer.com/image/art%3A10.1007%2Fs10554-012-0011-y/MediaObjects/10554_2012_11_Fig2_HTML.jpg
Fig. 2

Example of differences in contour detection in a single frame. The same coronary segment was analyzed by 2 different centers; please note that differences in contour detection resulted in differences in plaque composition. Center A (Greyscale (a) and VH-IVUS (b) analysis) versus Center B (c and d, resp)

https://static-content.springer.com/image/art%3A10.1007%2Fs10554-012-0011-y/MediaObjects/10554_2012_11_Fig3a_HTML.gif
https://static-content.springer.com/image/art%3A10.1007%2Fs10554-012-0011-y/MediaObjects/10554_2012_11_Fig3b_HTML.gif
Fig. 3

a Agreement of repeated VH-IVUS measurements of geometric volumes. Agreement of repeated VH-IVUS measurements of geometry between center A versus B (left), center A versus C (mid) and center A versus D (right). Δ = difference. b Agreement of repeated VH-IVUS measurements of geometric volumes. Agreement of repeated VH-IVUS measurements of geometry between center B versus C (left), center B versus D (mid) and center C versus D (right). Δ = difference

VH-IVUS measurement variability of plaque composition

VH-IVUS measurements of plaque composition were highly correlated for the different comparisons. Overall intra-class correlation and range of the individual between-center comparisons (within brackets) for (absolute) fibrous, fibro-lipidic, necrotic core and calcified volume was 0.96 (0.94–0.98), 0.94 (0.91–0.97), 0.98 (0.96–0.99), and 0.99 (0.98–0.99), respectively. There were significant differences for all plaque components (Tables 1, 2). The limits of agreement (i.e., 2 SD of mean difference) for compositional data were higher compared to geometrical data, but similar for the different comparisons (Fig. 4a, b). Fibrous tissue and necrotic core showed the highest measurement reproducibility. Smaller fibrous and fibro-lipidic volumes tended to show somewhat higher between-center variability.
https://static-content.springer.com/image/art%3A10.1007%2Fs10554-012-0011-y/MediaObjects/10554_2012_11_Fig4a_HTML.gif
https://static-content.springer.com/image/art%3A10.1007%2Fs10554-012-0011-y/MediaObjects/10554_2012_11_Fig4b_HTML.gif
Fig. 4

a Agreement of repeated VH-IVUS measurements of compositional volumes. Agreement of repeated VH-IVUS measurements of plaque composition between center A versus B (left), center A versus C (mid) and center A versus D (right). Δ = difference. b Agreement of repeated VH-IVUS measurements of compositional volumes. Agreement of repeated VH-IVUS measurements of plaque composition between center B versus C (left), center B versus D (mid) and center C versus D (right). Δ = difference

Discussion

In the present study, we evaluated the between-center reproducibility of VH-IVUS measurements of target lesions prior to PCI. VH-IVUS measurements of plaque and vessel dimensions and plaque composition were highly correlated for the different comparisons (overall ICC 0.83–0.99). However, we found significant between-center differences for vessel dimensions (relative differences of ≤4%, ≤12%, ≤9%, and ≤9%; vessel, lumen, and plaque volume and plaque burden, respectively) and plaque composition (relative differences of ≤15%, ≤25%, ≤11%, and ≤6%; fibrous, fibro-lipidic, necrotic core, and calcified plaque components, respectively). The highest measurement reproducibility for plaque components was found for fibrous tissue and necrotic core.

Rationale of VH-IVUS assessment of coronary lesions prior to PCI

Percutaneous coronary interventions may cause microembolization of unstable plaque material, which can lead to no-reflow phenomenon, myocardial micro-infarcts with cardiac marker release and impaired left ventricular function, and significantly impaired clinical outcome [11, 14, 16, 18, 20]. Several VH-IVUS studies demonstrated that differences in plaque composition of target lesions are related to such post-PCI events [1119]. Hong et al. showed that relative necrotic core volumes were significantly larger in patients with elevated troponin levels following PCI (19.8 ± 10.4% vs. 12.8 ± 8.4%, P = 0.015) [15]. Böse et al. reported that cardiac marker release following PCI was much greater in plaques with a large necrotic core volume (>10.81 mm³; P < 0.001) with a trend for a lower increase of cardiac markers in patients treated with statins before PCI [11]. Certain plaque components or phenotypes as assessed with VH-IVUS are related to no-reflow phenomenon post-PCI which may be triggered by such microembolization and adversely effects prognosis [14, 18]. Hong et al. reported that relative necrotic core volume serves as an independent predictor of no-reflow post-stenting in patients with acute coronary syndrome (odds ratio = 1.126; 95%CI 1.045–1.214, P = 0.002) [14].

Therefore, volumetric VH-IVUS assessment of target lesions prior to PCI may have the potential to identify lesions at particularly high risk for such complications and may help to tailor interventions (i.e., embolic protection device, direct stenting vs. pre-dilatation etc.) and may help to optimize systemic medical therapy (i.e., therapy with very high dose statins, or novel anti-atherosclerotic or platelet inhibiting drugs) [25, 26]. However, only large multicenter trials will be able to determine the clinical value of the assessment of plaque composition with VH-IVUS prior to PCI. An important pre-requisite for clinical VH-IVUS studies assessing such relations in advanced coronary lesions, is an adequate measurement reproducibility of VH-IVUS especially when pooled IVUS data is used in multicenter trials or registries [2].

Measurement reproducibility and studies with multicenter design

Several prior single-center studies demonstrated sufficient reproducibility of VH-IVUS measurements between observers, catheters, and repeated pullbacks [2, 8, 9, 21, 23].

Rodriguez-Granillo et al. assessed the measurement variability of VH-IVUS cross-sectional data obtained form 16 mild coronary plaques in a single-center design. The relative intra-observer difference was <11% for plaque cross-sectional area and compositional measurements showed a variability up to 24% [9]. Prasad et al. assessed volumetric VH-IVUS measurement variability of 16 significant coronary plaques (mean plaque burden of 55%) prior to PCI and reported high agreement for geometrical and compositional measurements (Spearman’s correlation >0.8) in a single-center design [8]. A relatively high intra- and interobserver reproducibility of volumetric geometric and compositional VH-IVUS data was previously described by Hartmann et al. in a series of 33 mild-to-moderate diseased coronary segments [2].

However, the aforementioned studies assessed VH-IVUS measurement variability of analysts from a single center or core lab with the same training background. In a previous multicenter study by this group of investigators, we demonstrated in coronary segments with mild-to-moderate coronary atherosclerosis a higher measurement variability and small but significant measurement differences between analysts from 4 different IVUS centers with different training programs [21]. In the current study, in much more advanced atherosclerotic lesions, VH-IVUS measurements of vessel geometry and plaque composition showed, despite good correlation between the four IVUS centers, significant differences for both vessel geometry and plaque composition. These differences could affect the results of multicenter trials with VH-IVUS.

Implications for multicenter VH-IVUS studies of coronary lesions prior to PCI

VH-IVUS measurement differences may have resulted from systematic over or underestimation of the lumen and/or vessel borders, differences in the interpolation of the vessel contour at the site of side-branches or behind calcified areas, and misinterpretation of thrombus and soft plaque adjacent to the lumen, which could theoretically have introduced larger measurement variability. The nature of such advanced lesions prior to PCI with more calcification, complex lesions and the IVUS probe against vessel wall/plaque, may result in increased difficulties of border detection in contrast with mild-to-moderate diseased coronary segments. It is likely that small differences or failures in contour detection will translate into (larger) errors of measurements of plaque composition. Therefore, our present study underlines the necessity to analyze the VH-IVUS data of advanced coronary lesions prior to PCI at a single site or by an experienced core-lab to avoid measurement errors and misinterpretation of acquired data.

Recently, Shin et al. described a simplified contour detection method for the assessment of plaque composition with on VH-IVUS [27]. In Shin’s method the lumen contour is drawn around the IVUS catheter, that is, without following the leading edge of the interface lumen intima. With this method only calcium and necrotic core can be assessed and related to total vessel area [27]. Shin et al. were able to demonstrate that coronary lesions with a greater amount of dense calcium volume and percent necrotic core to external elastic membrane show more post-PCI myocardial injury [28]. Nevertheless, exclusive application of this method is limited by incompleteness of the geometrical and compositional data (e.g., no data on lumen and plaque size, and no information on fibrous or fibro-lipidic tissue).

Limitations

The present study was performed in an offline setting in coronary segments with advanced coronary atherosclerosis prior to PCI; therefore, our findings cannot be extrapolated to VH-IVUS reproducibility studies in an online setting in the catheterization laboratory or following stent implantation. Similar to other studies with IVUS and VH-IVUS, we excluded very tortuous and severely calcified vessels that could have led to non-uniform pullbacks and/or inability to detect the (external) vessel contour [9, 21, 23]. We used the pcVH software (available in all four centers) to perform IVUS image analyses in entire coronary segments. Meanwhile a newer computer software is available, however, this does not affect the importance of the present study. The purpose of out present study was to examine systematic between-center differences when analyzing lumen and vessel contours with a semi-automated IVUS analysis program in this specific set of coronary lesions; it was not our aim to test a specific software release. In addition, the between-center differences of the present study were mainly explained by differences in manual contour editing, which is still generally not abandoned by the most recent IVUS analysis software. The variability between different IVUS catheters or pullback devices, repeated pullbacks, and independent selections of the segment of interest were not addressed in our present multicenter study, as this study focused on the assessment of measurement reproducibility in predefined coronary segments that contained target lesions prior to PCI.

Conclusions

A central analysis for VH-IVUS multicenter studies of lesions prior to PCI should be pursued. Moreover, it may be problematical to pool VH-IVUS data of individual trials that are analyzed by independent IVUS centers.

Conflict of interest

Dr. G.S. Mintz is a consultant to Volcano Corp. No other relevant conflict of interest was disclosed. This investigator-initiated study was not funded.

Open Access

This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://​creativecommons.​org/​licenses/​by-nc/​2.​0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Literatur
  1. Garcia-Garcia HM, Mintz GS, Lerman A, Vince DG, Margolis MP, van Es GA, Morel MA, Nair A, Virmani R, Burke AP, Stone GW, Serruys PW (2009) Tissue characterisation using intravascular radiofrequency data analysis: recommendations for acquisition, analysis, interpretation and reporting. Eurointervention 5(2):177–189PubMedView Article
  2. Hartmann M, Mattern ES, Huisman J, van Houwelingen GK, de Man FH, Stoel MG, Danse PW, Louwerenburg HW, von Birgelen C (2009) Reproducibility of volumetric intravascular ultrasound radiofrequency-based analysis of coronary plaque composition in vivo. Int J Cardiovasc Imaging 25(1):13–23PubMedView Article
  3. König A, Margolis MP, Virmani R, Holmes D, Klauss V (2008) Technology insight: in vivo coronary plaque classification by intravascular ultrasonography radiofrequency analysis. Nat Clin Pract Cardiovasc Med 5(4):219–229PubMedView Article
  4. König A, Klauss V (2007) Virtual histology. Heart 93(8):977–982PubMedView Article
  5. Mintz GS, Nissen SE, Anderson WD, Bailey SR, Erbel R, Fitzgerald PJ, Pinto FJ, Rosenfield K, Siegel RJ, Tuzcu EM, Yock PG (2001) American college of cardiology clinical expert consensus document on standards for acquisition, measurement and reporting of intravascular ultrasound studies (IVUS). A report of the american college of cardiology task force on clinical expert consensus documents. J Am Coll Cardiol 37(5):1478–1492PubMedView Article
  6. Nair A, Kuban BD, Tuzcu EM, Schoenhagen P, Nissen SE, Vince DG (2002) Coronary plaque classification with intravascular ultrasound radiofrequency data analysis. Circulation 106(17):2200–2206PubMedView Article
  7. Nasu K, Tsuchikane E, Katoh O, Vince DG, Virmani R, Surmely JF, Murata A, Takeda Y, Ito T, Ehara M, Matsubara T, Terashima M, Suzuki T (2006) Accuracy of in vivo coronary plaque morphology assessment: a validation study of in vivo virtual histology compared with in vitro histopathology. J Am Coll Cardiol 47(12):2405–2412PubMedView Article
  8. Prasad A, Cipher DJ, Prasad A, Mohandas A, Roesle M, Brilakis ES, Banerjee S (2008) Reproducibility of intravascular ultrasound virtual histology analysis. Cardiovasc Revasc Med 9(2):71–77PubMedView Article
  9. Rodriguez-Granillo GA, Vaina S, Garcia–Garcia HM, Valgimigli M, Duckers E, van Geuns RJ, Regar E, van der Giessen WJ, Bressers M, Goedhart D, Morel MA, de Feyter PJ, Serruys PW (2006) Reproducibility of intravascular ultrasound radiofrequency data analysis: implications for the design of longitudinal studies. Int J Cardiovasc Imaging 22(5):621–631PubMedView Article
  10. Van Herck J, De Meyer G, Ennekens G, Van Herck P, Herman A, Vrints C (2009) Validation of in vivo plaque characterisation by virtual histology in a rabbit model of atherosclerosis. EuroIntervention 5(1):149–156PubMedView Article
  11. Böse D, von Birgelen C, Zhou XY, Schmermund A, Philipp S, Sack S, Konorza T, Mohlenkamp S, Leineweber K, Kleinbongard P, Wijns W, Heusch G, Erbel R (2008) Impact of atherosclerotic plaque composition on coronary microembolization during percutaneous coronary interventions. Basic Res Cardiol 103(6):587–597PubMedView Article
  12. Garcia–Garcia HM, Shen Z, Piazza N (2009) Study of restenosis in drug eluting stents: new insights from greyscale intravascular ultrasound and virtual histology. EuroIntervention 5(Suppl D):D84–D92
  13. Garcia–Garcia HM, Gonzalo N, Kukreja N, Alfonso F (2008) Greyscale intravascular ultrasound and IVUS-radiofrequency tissue characterisation to improve understanding of the mechanisms of coronary stent thrombosis in drug-eluting stents. EuroIntervention 4(Suppl C):C33–C38
  14. Hong YJ, Jeong MH, Choi YH, Ko JS, Lee MG, Kang WY, Lee SE, Kim SH, Park KH, Sim DS, Yoon NS, Youn HJ, Kim KH, Park HW, Kim JH, Ahn Y, Cho JG, Park JC, Kang JC (2011) Impact of plaque components on no-reflow phenomenon after stent deployment in patients with acute coronary syndrome: a virtual histology-intravascular ultrasound analysis. Eur Heart J 32(16):2059–2066PubMedView Article
  15. Hong YJ, Mintz GS, Kim SW, Lee SY, Okabe T, Pichard AD, Satler LF, Waksman R, Kent KM, Suddath WO, Weissman NJ (2009) Impact of plaque composition on cardiac troponin elevation after percutaneous coronary intervention: an ultrasound analysis. JACC Cardiovasc Imaging 2(4):458–468PubMedView Article
  16. Kawaguchi R, Oshima S, Jingu M, Tsurugaya H, Toyama T, Hoshizaki H, Taniguchi K (2007) Usefulness of virtual histology intravascular ultrasound to predict distal embolization for ST-segment elevation myocardial infarction. J Am Coll Cardiol 50(17):1641–1646PubMedView Article
  17. Kim SH, Hong MK, Park DW, Lee SW, Kim YH, Lee CW, Kim JJ, Park SW, Park SJ (2009) Impact of plaque characteristics analyzed by intravascular ultrasound on long-term clinical outcomes. Am J Cardiol 103(9):1221–1226PubMedView Article
  18. Nakamura T, Kubo N, Ako J, Momomura S (2007) Angiographic no-reflow phenomenon and plaque characteristics by virtual histology intravascular ultrasound in patients with acute myocardial infarction. J Interv Cardiol 20(5):335–339PubMedView Article
  19. Rodriguez-Granillo GA, Garcia–Garcia HM, Mc Fadden EP, Valgimigli M, Aoki J, de Feyter PJ, Serruys PW (2005) In vivo intravascular ultrasound-derived thin-cap fibroatheroma detection using ultrasound radiofrequency data analysis. J Am Coll Cardiol 46(11):2038–2042PubMedView Article
  20. Kawamoto T, Okura H, Koyama Y, Toda I, Taguchi H, Tamita K, Yamamuro A, Yoshimura Y, Neishi Y, Toyota E, Yoshida K (2007) The relationship between coronary plaque characteristics and small embolic particles during coronary stent implantation. J Am Coll Cardiol 50(17):1635–1640PubMedView Article
  21. Huisman J, Egede R, Rdzanek A, Böse D, Erbel R, Kochman J, Jensen LO, van der Palen J, Hartmann M, Mintz GS, von Birgelen C (2010) Between-centre reproducibility of volumetric intravascular ultrasound radiofrequency-based analyses in mild-to-moderate coronary atherosclerosis: an international multicentre study. Eurointervention 5:925–931PubMedView Article
  22. Bruining N, von Birgelen C, de Feyter PJ, Ligthart J, Li W, Serruys PW, Roelandt JR (1998) ECG-gated versus nongated three-dimensional intracoronary ultrasound analysis: implications for volumetric measurements. Cathet Cardiovasc Diagn 43(3):254–260PubMedView Article
  23. Huisman J, Hartmann M, Mattern ES, Mintz GS, Basalus MW, van Houwelingen GK, Verhorst PM, von Birgelen C (2010) Impact of analyzing less image frames per segment for radiofrequency-based volumetric intravascular ultrasound measurements in mild-to-moderate coronary atherosclerosis. Int J Cardiovasc Imaging 26(5):487–497PubMedView Article
  24. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMedView Article
  25. Huisman J, Hartmann M, von Birgelen C (2011) Ultrasound and light: friend or foe? On the role of intravascular ultrasound in the era of optical coherence tomography. Int J Cardiovasc Imaging 27(2):209–214PubMedView Article
  26. von Birgelen C, Huisman J, Hartmann M (2010) Is size really all that matters? Remarks on size and necrotic core content of atherosclerotic plaques. Int J Cardiovasc Imaging 26(2):173–176View Article
  27. Shin ES, Garcia–Garcia HM, Serruys PW (2010) A new method to measure necrotic core and calcium content in coronary plaques using intravascular ultrasound radiofrequency-based analysis. Int J Cardiovasc Imaging 26(4):387–396PubMedView Article
  28. Shin ES, Garcia–Garcia HM, Garg S, Park J, Kim SJ, Serruys PW (2011) The assessment of Shin’s method for the prediction of creatinine kinase-MB elevation after percutaneous coronary intervention: an intravascular ultrasound study. Int J Cardiovasc Imaging 27(6):883–892PubMedView Article
Metadaten
Titel
Multicenter assessment of the reproducibility of volumetric radiofrequency-based intravascular ultrasound measurements in coronary lesions that were consecutively stented
verfasst von
Jennifer Huisman
Rasmus Egede
Adam Rdzanek
Dirk Böse
Raimund Erbel
Janusz Kochman
Lisette Okkels Jensen
Job van der Palen
Marc Hartmann
Gary S. Mintz
Clemens von Birgelen
Publikationsdatum
01.12.2012
Verlag
Springer Netherlands
Erschienen in
The International Journal of Cardiovascular Imaging / Ausgabe 8/2012
Print ISSN: 1569-5794
Elektronische ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-012-0011-y

Weitere Artikel der Ausgabe 8/2012

The International Journal of Cardiovascular Imaging 8/2012 Zur Ausgabe

Das Risiko für Vorhofflimmern in der Bevölkerung steigt

02.05.2024 Vorhofflimmern Nachrichten

Das Risiko, im Lauf des Lebens an Vorhofflimmern zu erkranken, ist in den vergangenen 20 Jahren gestiegen: Laut dänischen Zahlen wird es drei von zehn Personen treffen. Das hat Folgen weit über die Schlaganfallgefährdung hinaus.

VHF-Ablation nützt wohl nur bei reduzierter Auswurfleistung

02.05.2024 Ablationstherapie Nachrichten

Ob die Katheterablation von Vorhofflimmern bei Patienten mit Herzinsuffizienz die Komplikationsraten senkt, scheint davon abzuhängen, ob die Auswurfleistung erhalten ist oder nicht. Das legen die Ergebnisse einer Metaanalyse nahe.

Weniger Extremitätenischämien mit dualer Plättchenhemmung

02.05.2024 Thrombozytenaggregationshemmer Nachrichten

Eine Behandlung mit Ticagrelor zusätzlich zu ASS kann das Risiko für Revaskularisierungen und Amputationen von Extremitäten bei Diabetikern mit stabiler KHK deutlich reduzieren, vor allem für solche mit PAVK. Dafür spricht eine Auswertung der Interventionsstudie THEMIS.

Beutel versus Maschine: Beste Beatmungstechnik bei Herzstillstand gesucht

02.05.2024 Kardiopulmonale Reanimation Nachrichten

Stehen die Chancen auf eine Rückkehr der Spontanzirkulation nach Herz-Kreislauf-Stillstand bei manueller oder maschineller Beatmung besser? Und unterscheidet sich das neurologische Outcome nach der Reanimation? Das belgische Herzstillstand-Register liefert die Daten für einen direkten Vergleich zwischen Beutel und Beatmungsgerät.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.