A new method to measure necrotic core and calcium content in coronary plaques using intravascular ultrasound radiofrequency-based analysis

verfasst von: Eun-Seok Shin, Hector M. Garcia-Garcia, Patrick W. Serruys

Erschienen in: The International Journal of Cardiovascular Imaging | Ausgabe 4/2010

Abstract

Although previous intravascular ultrasound (IVUS) radiofrequency-based analysis data showed acceptable reproducibility for plaque composition, measurements are not easily obtained, particularly that of lumen contour, because of the limited IVUS resolution. The purpose of this study was to compare a new measurement method (Shin’s method) and the conventional measurement method for necrotic core and calcium content in atherosclerotic lesions using Virtual Histology-intravascular ultrasound (VH-IVUS). Fifty-seven patients with unstable angina who underwent elective percutaneous coronary intervention were included. Shin’s method focuses on catheter contour, instead of lumen contour, and vessel contour. Patients ages ranged from 46 to 88 years, and 34 were men. A total of 1,401 frames from 59 culprit lesions were assessed. There were no significant differences in the mean area and volume of necrotic core and dense calcium between the two methods. Correlation coefficients (R) were ≥0.99 for all above mentioned parameters (P < 0.001). Between methods, the absolute differences in mean area and volume of necrotic core were 0.02 ± 0.02 mm² and 0.34 ± 0.29 mm³, respectively, while for mean area and volume of dense calcium, the absolute differences were 0.04 ± 0.07 mm² and 0.36 ± 0.52 mm³, respectively. The reproducibility of Shin’s method was excellent. For area of the necrotic core and dense calcium, the means of the differences between the two measurements were nearly zero, and the reproducibility coefficients were within 1% of the means of the two measurements. Mean analysis time for both measurements was 26.8 ± 6.7 min/segment in the conventional method and 3.3 ± 0.6 min/segment in Shin’s method. Shin’s method for measurement of necrotic core and dense calcium using VH-IVUS demonstrated a good correlation with the conventional method and excellent reproducibility. Also, Shin’s method required a significantly shorter analysis time than the conventional method. Therefore, Shin’s method could replace the conventional method for necrotic core and calcium measurement in atherosclerotic lesions, and it might be useful in the catheterization laboratory for online clinical decision.

Introduction

While there is a well established relationship between degree of stenosis and plaque burden and ischemic symptoms in patients with stable angina, plaque composition, especially necrotic core, is more relevant in terms of occurrence of acute coronary syndromes [14]. Virtual Histology-intravascular ultrasound (VH-IVUS), which permits the analysis of coronary plaque composition in vivo, is increasingly used in clinical research [512]. Despite its demonstrated utility in quantifying coronary plaque components with high predictive accuracy both in vitro and in vivo [13, 14], VH-IVUS is infrequently used routinely in the catheterization laboratory for clinical decision making partly because it is not easy to immediately draw the contours. In particular, lumen contour detection takes a long time and it is especially challenging in patients with high risk lesions (i.e., acute coronary syndrome) because there is no clear separation between luminal thrombi and intima. The aim of this study is to assess a new measurement method (Shin’s method) that readily provides on-line (i.e., immediately after acquisition) necrotic core and calcium measurements to help interventionalists to better assess coronary lesions irrespective of clinical presentation. Therefore, in the present study we compared the new measurement method (Shin’s method) and the conventional measurement method in quantifying necrotic core and calcium in atherosclerotic lesions using VH-IVUS.

Methods

Study patients

In this study, 1,401 frames and 59 segments of culprit lesions were imaged in a total of 57 consecutive patients with unstable angina who underwent coronary artery stenting. Specifically, measurements of necrotic core and dense calcium (i.e., area and volume) by VH-IVUS were considered. Unstable angina was defined according to the Braunwald classification [15], and patients with class IIB and IIIB were included in the study. Patients were excluded if they had an IVUS-detected thrombotic lesion, left main disease and chronic total occlusion. Culprit lesion is defined as the lesion related to the patient’s symptom that was treated with stenting. We determined the analysis times needed for both approaches and report the mean times of analysis. This study was performed with the patient’s written informed consent and approval of the institutional review board.

VH-IVUS

Details of the radiofrequency (RF) IVUS technique have been extensively validated [9, 10, 13, 14]. In brief, spectral analysis of VH-IVUS data is used to create tissue maps to classify atherosclerotic plaques into four major components (fibrous: green; fibrofatty: light-green; dense calcium: white; and necrotic core: red). In this study, only dense calcium and necrotic core were considered because they are commonly found in high-risk plaques; both were assessed using the conventional method and Shin’s method.

IVUS data acquisition

IVUS data were acquired with commercially available phased-array IVUS catheters (Eagle Eye Gold 2.9-F 20 MHz, Volcano Corporation, Rancho Cordova, USA) by a dedicated VH-IVUS console (Volcano Corporation, Rancho Cordova, USA). Image acquisition was ECG-gated. After intracoronary injection of 200 μg nitroglycerine, continuous pullback of the IVUS catheter was performed using a motorized pullback device at 0.5 mm/s. Data of both pullbacks were stored on hard disk for off-line analysis.

VH-IVUS data analysis

RF-IVUS analysis was performed off-line by an independent core laboratory (Cardialysis BV, Rotterdam, The Netherlands). The region of interest comprises the most diseased 10-mm segment within the culprit lesion that was contained the largest plaque burden. This approach mimics a real clinical situation in which the interventionalists must make a clinical decision on the most stenotic segment using not only geometrical but also compositional analysis.

Conventional method of analysis

The contours were traced using a semiautomatic contour detection program (PcVH 2.2, Volcano Corporation, Rancho Cordova, USA). The contours of each cross sectional image per frame were manually corrected if required to meet a high standard of accuracy. The lumen contour was detected by tracing the leading edge of the intima, and the vessel contour by tracing the leading edge of the adventitia. For each frame geometrical and compositional analyses were performed; the mean area of the vessel and lumen were obtained and for each segment, the vessel and lumen volumes were generated (Figs. 1, 2).

Shin’s method of analysis

Shin’s method also used the off-line software PcVH 2.2. Instead of measuring the lumen as described above for the conventional method, in Shin’s method the lumen contour is drawn around the IVUS catheter (hereafter termed catheter contour), that is, without following the leading edge of the interface lumen intima. When there is a “flare” (ring-down artifact) around the IVUS catheter, this catheter contour must be drawn away from this flare (Fig. 1). The catheter contour was manually detected. To compare only the influence of the two different approaches (Conventional vs. Shin’s), in terms of dense calcium and necrotic core, the same contour of the vessel was used (Figs. 1, 2).
Using Shin’s method the following information can be obtained:
1.
Necrotic core and dense calcium areas and volumes.
 
2.
Vessel area and volume.
 
However, using Shin’s method the following information can NOT be obtained:
1.
Fibrous and fibrofatty areas and volumes.
 
2.
Lumen and plaque areas and volumes.
 

Data analysis

For all comparisons, two levels were considered, the frame level and the segment level.
First, we compared the two methods (necrotic core and calcium) by one expert analyst. To assess the reproducibility of Shin’s method, eight segments (189 frames) were randomly selected and the same segments were analyzed twice by the same analyst with a 3-week time difference. To this end, the vessel contour was kept from the conventional method, so that only the catheter contour was drawn twice.

Statistical analysis

Analyses were performed with SPSS 16.0 (SPSS Inc., Chicago, IL). Dichotomous data are presented as frequencies. Quantitative data are presented as mean ± SD and compared using Student’s t-test, and linear regression analysis for repeated measures. The Spearman’s correlation coefficient was used to assess associations between measured parameters determined by the two methods. According to Bland and Altman [16], the agreement between two measurements were assessed by determining the mean ± 2 SD of the between measurement differences. A two-sided P value < 0.05 was considered significant.

Results

Study patients

Table 1 shows the patients’ characteristics. The mean age was 60.1 ± 8.6 years, and 57.6% of participants were men. The study vessels were the left anterior descending artery in 55.9%, the left circumflex artery in 10.2% and the right coronary artery in 33.9% of cases. The length of the segments was 10.2 ± 0.3 mm with a volumetric plaque burden of 60.7 ± 0.8% (43.1–75.5%). There was no complication related to IVUS imaging.
Table 1
Study population
 
N = 57
Age, years
60.1 ± 8.6
Male
34 (57.6)
Diabetes
16 (27.1)
Hypertension
30 (50.8)
Current smoker
13 (22.0)
Hypercholesterolemia
34 (57.6)
Total cholesterol, mg/dl
190.9 ± 40.2
LDL cholesterol, mg/dl
111.5 ± 38.5
hsCRP (mg/l)
2.2 ± 3.3
Study vessel
 LAD
33 (55.9)
 LCX
6 (10.2)
 RCA
20 (33.9)
Plaque burden, %
60.7 ± 0.8
Values are presented as number (%) or mean ± SD. Hypercholesterolemia defined as ≥180 mg/dl of total cholesterol
LAD left anterior descending artery, LCX left circumflex artery, RCA right coronary artery, hsCRP high-sensitivity C-reactive protein

Comparison between the conventional and Shin’s methods

By design, the mean length of analysis and the mean vessel area are the same (Table 2). There were no significant differences in the mean area and volume of necrotic core and dense calcium between the two methods. The correlation coefficients (R) were 0.99 (P < 0.001) for all compared parameters (Fig. 3). Between methods, the absolute differences in mean area and volume of necrotic core were 0.02 ± 0.02 mm² and 0.34 ± 0.29 mm³, respectively, while for mean area and volume of dense calcium, the absolute differences were 0.04 ± 0.07 mm² and 0.36 ± 0.52 mm³, respectively. Although it did not achieve statistical significance, mean dense calcium was consistently smaller and mean necrotic area was larger in Shin’s method as compared to the conventional method (Fig. 4). A total of three segments in necrotic core and five segments in dense calcium had an absolute difference in volume of more than 1 mm³. The correlation of the mean areas and volumes of necrotic core and dense calcium between Shin’s and the conventional method was expressed by a linear least-squares fit in Fig. 3.
Table 2
Mean area and volume of necrotic core and dense calcium with conventional and Shin’s method
 
Conventional
Shin’s
Absolute △
P
Frame (n = 1401)
 Vessel area, mm²
13.78 ± 3.90
13.78 ± 3.90
0
 
 DC area, mm²
0.41 ± 0.51
0.38 ± 0.48
0.04 ± 0.07
0.07
 NC area, mm²
0.78 ± 0.70
0.80 ± 0.71
0.02 ± 0.02
0.62
Segment (n = 59)
 Lesion length, mm
10.15 ± 0.29
10.15 ± 0.29
0
 
 Vessel volume, mm³
140.65 ± 37.90
140.65 ± 37.90
0
 
 DC volume, mm³
4.35 ± 4.43
3.99 ± 4.16
0.36 ± 0.52
0.66
 NC volume, mm³
8.26 ± 6.62
8.39 ± 6.72
0.34 ± 0.29
0.91
Values are presented as mean ± SD
DC dense calcium, NC necrotic core, P conventional versus Shin’s method, difference
Although fibrous and fibrofatty components are not correct by Shin’s method because this method can not differentiate between lumen contents (usually characterized as fibrous and fibrofatty) and the fibrous and fibrofatty that are part of the plaque contents, here we provide a comparison between conventional vs. Shin’s method: 3.37 ± 2.03 mm2 vs. 4.04 ± 2.16 mm2, P < 0.001 for fibrous area and 0.87 ± 0.93 mm2 vs. 1.68 ± 1.49 mm2, P < 0.001 for fibrofatty area, respectively.

Reproducibility of Shin’s method

Table 3 shows the reproducibility of Shin’s method. The linear least-squares fit of the relationship between the two measurements for necrotic core area and dense calcium area were R = 0.999, P < 0.001 and R = 1.0, P < 0.001, respectively (Fig. 5). For areas of the necrotic core and dense calcium, the means of the differences between the two measurements were 0.004 ± 0.019 and 0.002 ± 0.013 mm². The limits of agreement for area of the necrotic core and dense calcium were −0.034 to 0.042 mm² and −0.024 to 0.051 mm². The reproducibility coefficients were within 1% of the means of the two measurements for area and volume of the necrotic core and dense calcium.
Table 3
Reproducibility of necrotic core and dense calcium by Shin’s method
 
Shin’s 1
Shin’s 2
Absolute △
P
Frame (n = 189)
 Vessel area, mm²
14.66 ± 3.73
14.66 ± 3.73
0
 
 DC area, mm²
0.39 ± 0.48
0.38 ± 0.48
0.01 ± 0.01
0.96
 NC area, mm²
0.82 ± 0.57
0.81 ± 0.56
0.01 ± 0.02
0.95
Segment (n = 8)
 Lesion length, mm
10.24 ± 0.28
10.24 ± 0.28
0
 
 Vessel volume, mm³
150.42 ± 35.2
150.42 ± 35.2
0
 
 DC volume, mm³
4.03 ± 3.78
4.00 ± 3.78
0.03 ± 0.02
0.99
 NC volume, mm³
8.39 ± 4.58
8.35 ± 4.57
0.06 ± 0.06
0.99
Values are presented as mean ± SD
DC dense calcium, NC necrotic core, P conventional versus Shin’s method, difference

Analysis time

To assess the measurement analysis time difference between the two methods, 12 segments were randomly selected. Manual correction of the automatic contour detection was required in almost all frames for a total analysis time of 26.8 ± 6.7 min/segment in the conventional method. However, analysis time was 3.3 ± 0.6 min/segment for Shin’s method (Table 4).
Table 4
Comparison of analysis time between the conventional method and Shin’s method (12 segments)
 
Conventional
Shin’s
Absolute △
P
Analyzed length, mm
10.1 ± 0.2
10.1 ± 0.2
0
1.0
Analysis time, min
26.8 ± 6.7
3.3 ± 0.6
23.5 ± 6.2
<0.001
Values are presented as mean ± SD. P Conventional vs. Shin’s method, difference

Discussions

This report introduces a new measurement method (Shin’s method) for necrotic core and dense calcium using VH-IVUS. The main findings of this study are: (1) necrotic core and dense calcium areas and volumes determined by Shin’s method show a good correlation to those derived from the conventional method; (2) Shin’s method has a high reproducibility, and (3) analysis time is shorter for Shin’s method.
The introduction of RF-based IVUS analysis permits quantitative assessment of atherosclerotic plaque composition [17]. This is of crucial relevance because necrotic core is a key determinant of plaque vulnerability. This imaging modality has been extensively scrutinized and studied in different clinical contexts, and most studies using this technique have shown consistency as compared with previous pathological studies [13, 14]. However, this imaging technique is very much influenced by the precision of contour detection, which in turn is highly time-consuming thus rendering its clinical use for online clinical decisions today not possible and not widely accepted. There was a need therefore to explore a different manner of necrotic core quantification that would allow information to be quickly obtained as accurate as possible and with excellent reproducibility.
Use of Shin’s method does not require training, and more rapid contour detection around the catheter can be done. In the future, with the implementation of an automatic catheter contour detection program, no user intervention will be needed. This catheter contour must be drawn beyond the ring-down artifact (i.e., ‘flare’) so that this is not misclassified as necrotic core or dense calcium.
Although in the conventional method we would have included thrombi that are undetected by IVUS, recent papers showed that thrombus was always classified as fibrous and fibrofatty tissues by VH-IVUS [18, 19], and therefore even when the luminal contour would have included some thrombi, the necrotic core and calcium measurement using Shin’s method is not affected by a variable contour detection.
Although this study showed that the sum of necrotic core and calcium was very similar using both methods, the mean dense calcium area was consistently smaller and the mean necrotic area was larger in Shin’s method as compared to the conventional method. In particular, a total of five segments in dense calcium and three segments in necrotic core had an absolute difference in volume of more than 1 mm³. Possible explanations for this difference between two methods are as follows: (1) Thin plaque is treated as a grey medial stripe by VH-IVUS in the conventional method. However, Shin’s method is not influenced by the artificial thickness of the medial stripe. Therefore, Shin’s method shows more necrotic core even though there is a thin plaque (Fig. 6). (2) In the case of mixed plaques at the lumen, in the conventional method, the proximity of the contour with the plaque, will characterize the adjacent tissue to the contour using one of the two classification trees of the virtual histology system, resulting in more calcium or more necrotic core at that area (Fig. 6). In Shin’s method, a longer distance will be present from the catheter contour to the plaque, and therefore the second classification tree would be applied resulting in a slightly different tissue characterization. In a previous study, in vivo VH-IVUS images suggested an increase in dense calcium in the acute coronary syndrome group when compared with the stable angina group. Although in general the in vivo VH-IVUS data correlated well with in vitro histology (predictive accuracy of 96.5%), the RF data analysis overestimated the frequency of calcifications. The authors mentioned that one possible explanation was that the artifact is colored with white because the software for VH analysis is obliged to assign one of the four colors for each pixel [13].
Previous data with the conventional method showed acceptable reproducibility for plaque composition; likewise, repeated measurements for necrotic core and calcium with Shin’s method were highly correlated. Awareness of inter-observer differences may be particularly important in serial pharmacological intervention trials where small changes in plaque composition (i.e., necrotic core) may be expected [20, 21]. Therefore, Shin’s method may be a more adequate system to assess the change of necrotic core in longitudinal studies.
Shin’s method for measurement of necrotic core and dense calcium using VH-IVUS demonstrated a good correlation with the conventional method and excellent reproducibility. In the future, since Shin’s method could be automatically applied using an automated catheter contour detection with a computer program, and analysis time will be further decreased. This would also mean that the technique can be used in the catheterization laboratory to help operators in decision making based on content of necrotic core or calcium.

Limitations

This study has several limitations. (1) Because patients in the present study all suffered unstable angina, we cannot know how well the two methods would correlate in other atherosclerotic lesions. (2) The study did not include a validation component as therefore the accuracy of Shin’s method remains to be determined (3) Shin’s method does not show the lumen area and plaque area. Therefore, greyscale should remain the gold standard to assess plaque size changes (geometrical analysis), while virtual histology should be used to obtain compositional analysis, so as complementary information to greyscale. (4) Shin’s method does not provide correct quantification of fibrous or fibrofatty tissues.

Conclusions

Shin’s method for measurement of necrotic core and dense calcium using VH-IVUS demonstrated a good correlation with the conventional method and a high reproducibility. Also, Shin’s method required a significantly shorter analysis time than the conventional method. Therefore, Shin’s method could replace the conventional method for necrotic core and calcium measurement in atherosclerotic lesions, and it might be useful in the catheterization laboratory for online clinical decision.

Open Access

This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://​creativecommons.​org/​licenses/​by-nc/​2.​0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Virmani R, Burke AP, Farb A et al (2006) Pathology of the vulnerable plaque. J Am Coll Cardiol 47:C13–C18CrossRefPubMed Virmani R, Burke AP, Farb A et al (2006) Pathology of the vulnerable plaque. J Am Coll Cardiol 47:C13–C18CrossRefPubMed
2.
Zurück zum Zitat Schoenhagen P, Nissen SE (2005) Assessing coronary plaque burden and plaque vulnerability: atherosclerosis imaging with IVUS and emerging noninvasive modalities. Am Heart Hosp J 1:164–169CrossRef Schoenhagen P, Nissen SE (2005) Assessing coronary plaque burden and plaque vulnerability: atherosclerosis imaging with IVUS and emerging noninvasive modalities. Am Heart Hosp J 1:164–169CrossRef
3.
Zurück zum Zitat Schoenhagen P, Nissen SE (2003) Coronary atherosclerotic disease burden: an emerging endpoint in progression/regression studies using intravascular ultrasound. Curr Drug Targets Cardiovasc Haematol Disord 3:218–226CrossRefPubMed Schoenhagen P, Nissen SE (2003) Coronary atherosclerotic disease burden: an emerging endpoint in progression/regression studies using intravascular ultrasound. Curr Drug Targets Cardiovasc Haematol Disord 3:218–226CrossRefPubMed
4.
Zurück zum Zitat Schaar JA, de Korte CL, Mastik F et al (2003) Intravascular palpography for high-risk vulnerable plaque assessment. Herz 28:488–495CrossRefPubMed Schaar JA, de Korte CL, Mastik F et al (2003) Intravascular palpography for high-risk vulnerable plaque assessment. Herz 28:488–495CrossRefPubMed
5.
Zurück zum Zitat Van Mieghem CA, McFadden EP, de Feyter PJ et al (2006) Noninvasive detection of subclinical coronary atherosclerosis coupled with assessment of changes in plaque characteristics using novel invasive imaging modalities: the Integrated Biomarker and Imaging Study (IBIS). J Am Coll Cardiol 47:1134–1142CrossRefPubMed Van Mieghem CA, McFadden EP, de Feyter PJ et al (2006) Noninvasive detection of subclinical coronary atherosclerosis coupled with assessment of changes in plaque characteristics using novel invasive imaging modalities: the Integrated Biomarker and Imaging Study (IBIS). J Am Coll Cardiol 47:1134–1142CrossRefPubMed
6.
Zurück zum Zitat Garcia-Garcia HM, Gonzalo N, Regar E et al (2009) Virtual histology and optical coherence tomography: from research to a broad clinical application. Heart 95:1362–1374CrossRefPubMed Garcia-Garcia HM, Gonzalo N, Regar E et al (2009) Virtual histology and optical coherence tomography: from research to a broad clinical application. Heart 95:1362–1374CrossRefPubMed
7.
Zurück zum Zitat Valgimigli M, Rodriguez-Granillo GA, Garcia-Garcia HM et al (2006) Distance from the ostium as an independent determinant of coronary plaque composition in vivo: an intravascular ultrasound study based radiofrequency data analysis in humans. Eur Heart J 27:655–663CrossRefPubMed Valgimigli M, Rodriguez-Granillo GA, Garcia-Garcia HM et al (2006) Distance from the ostium as an independent determinant of coronary plaque composition in vivo: an intravascular ultrasound study based radiofrequency data analysis in humans. Eur Heart J 27:655–663CrossRefPubMed
8.
Zurück zum Zitat Rodriguez-Granillo GA, Garcia-Garcia HM, Valgimigli M et al (2006) Global characterization of coronary plaque rupture phenotype using three-vessel intravascular ultrasound radiofrequency data analysis. Eur Heart J 27:1921–1927CrossRefPubMed Rodriguez-Granillo GA, Garcia-Garcia HM, Valgimigli M et al (2006) Global characterization of coronary plaque rupture phenotype using three-vessel intravascular ultrasound radiofrequency data analysis. Eur Heart J 27:1921–1927CrossRefPubMed
9.
Zurück zum Zitat Mehta SK, McCrary JR, Frutkin AD et al (2007) Intravascular ultrasound radiofrequency analysis of coronary atherosclerosis: an emerging technology for the assessment of vulnerable plaque. Eur Heart J 28:1283–1288CrossRefPubMed Mehta SK, McCrary JR, Frutkin AD et al (2007) Intravascular ultrasound radiofrequency analysis of coronary atherosclerosis: an emerging technology for the assessment of vulnerable plaque. Eur Heart J 28:1283–1288CrossRefPubMed
11.
Zurück zum Zitat Hong M-K, Mintz GS, Lee CW et al (2007) Comparison of virtual histology to intravascular ultrasound of culprit coronary lesions in acute coronary syndrome and target coronary lesions in stable angina pectoris. The American Journal of Cardiology 100:953–959CrossRefPubMed Hong M-K, Mintz GS, Lee CW et al (2007) Comparison of virtual histology to intravascular ultrasound of culprit coronary lesions in acute coronary syndrome and target coronary lesions in stable angina pectoris. The American Journal of Cardiology 100:953–959CrossRefPubMed
12.
Zurück zum Zitat Fujii K, Carlier SHG, Mintz GS et al (2005) Association of plaque characterization by intravascular ultrasound virtual histology and arterial remodeling. The American Journal of Cardiology 96:1476–1483CrossRefPubMed Fujii K, Carlier SHG, Mintz GS et al (2005) Association of plaque characterization by intravascular ultrasound virtual histology and arterial remodeling. The American Journal of Cardiology 96:1476–1483CrossRefPubMed
13.
Zurück zum Zitat Nasu K, Tsuchikane E, Katoh O et al (2006) Accuracy of in vivo coronary plaque morphology assessment: a validation study of in vivo virtual histology compared with in vitro histopathology. J Am Coll Cardiol 47:2405–2412CrossRefPubMed Nasu K, Tsuchikane E, Katoh O et al (2006) Accuracy of in vivo coronary plaque morphology assessment: a validation study of in vivo virtual histology compared with in vitro histopathology. J Am Coll Cardiol 47:2405–2412CrossRefPubMed
14.
Zurück zum Zitat Nair A, Kuban BD, Tuzcu EM et al (2002) Coronary plaque classification with intravascular ultrasound radiofrequency data analysis. Circulation 106:2200–2206CrossRefPubMed Nair A, Kuban BD, Tuzcu EM et al (2002) Coronary plaque classification with intravascular ultrasound radiofrequency data analysis. Circulation 106:2200–2206CrossRefPubMed
15.
Zurück zum Zitat Braunwald E, Antman EM, Beasley JW et al (2002) ACC/AHA guideline update for the management of patients with unstable angina and non-ST-segment elevation myocardial infarction–2002: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (committee on the management of patients with unstable Angina). Circulation 106:1893–1900CrossRefPubMed Braunwald E, Antman EM, Beasley JW et al (2002) ACC/AHA guideline update for the management of patients with unstable angina and non-ST-segment elevation myocardial infarction–2002: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (committee on the management of patients with unstable Angina). Circulation 106:1893–1900CrossRefPubMed
16.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
17.
Zurück zum Zitat Mintz GS, Maehara A (2009) Serial intravascular ultrasound assessment of atherosclerosis progression and regression. Circ J 73:1557–1560CrossRefPubMed Mintz GS, Maehara A (2009) Serial intravascular ultrasound assessment of atherosclerosis progression and regression. Circ J 73:1557–1560CrossRefPubMed
18.
Zurück zum Zitat Nasu K, Tsuchikane E, Katoh O et al (2008) Impact of intramural thrombus in coronary arteries on the accuracy of tissue characterization by in vivo intravascular ultrasound radiofrequency data analysis. Am J Cardiol 101:1079–1083PubMed Nasu K, Tsuchikane E, Katoh O et al (2008) Impact of intramural thrombus in coronary arteries on the accuracy of tissue characterization by in vivo intravascular ultrasound radiofrequency data analysis. Am J Cardiol 101:1079–1083PubMed
19.
Zurück zum Zitat Frutkin AD, Mehta SK, McCrary JR et al (2007) Limitations to the use of virtual histology-intravascular ultrasound to detect vulnerable plaque. Eur Heart J 28:1783–1784CrossRefPubMed Frutkin AD, Mehta SK, McCrary JR et al (2007) Limitations to the use of virtual histology-intravascular ultrasound to detect vulnerable plaque. Eur Heart J 28:1783–1784CrossRefPubMed
20.
Zurück zum Zitat Serruys PW, Garcia-Garcia HM, Buszman P et al (2008) Effects of the direct lipoprotein-associated phospholipase A2 inhibitor darapladib on human coronary atherosclerotic plaque. Circulation 118:1172–1182CrossRefPubMed Serruys PW, Garcia-Garcia HM, Buszman P et al (2008) Effects of the direct lipoprotein-associated phospholipase A2 inhibitor darapladib on human coronary atherosclerotic plaque. Circulation 118:1172–1182CrossRefPubMed
21.
Zurück zum Zitat Kawasaki M, Sano K, Okubo M et al (2005) Volumetric quantitative analysis of tissue characteristics of coronary plaques after statin therapy using three-dimensional integrated backscatter intravascular ultrasound. J Am Coll Cardiol 45:1946–1953CrossRefPubMed Kawasaki M, Sano K, Okubo M et al (2005) Volumetric quantitative analysis of tissue characteristics of coronary plaques after statin therapy using three-dimensional integrated backscatter intravascular ultrasound. J Am Coll Cardiol 45:1946–1953CrossRefPubMed
Metadaten
Titel
A new method to measure necrotic core and calcium content in coronary plaques using intravascular ultrasound radiofrequency-based analysis
verfasst von
Eun-Seok Shin
Hector M. Garcia-Garcia
Patrick W. Serruys
Publikationsdatum
01.04.2010
Verlag
Springer Netherlands
Erschienen in
The International Journal of Cardiovascular Imaging / Ausgabe 4/2010
Print ISSN: 1569-5794
Elektronische ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-009-9567-6

Weitere Artikel der Ausgabe 4/2010

The International Journal of Cardiovascular Imaging 4/2010 Zur Ausgabe

Case in Point

Don’t stop at two

„Jeder Fall von plötzlichem Tod muss obduziert werden!“

17.05.2024 Plötzlicher Herztod Nachrichten

Ein signifikanter Anteil der Fälle von plötzlichem Herztod ist genetisch bedingt. Um ihre Verwandten vor diesem Schicksal zu bewahren, sollten jüngere Personen, die plötzlich unerwartet versterben, ausnahmslos einer Autopsie unterzogen werden.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Schlechtere Vorhofflimmern-Prognose bei kleinem linken Ventrikel

17.05.2024 Vorhofflimmern Nachrichten

Nicht nur ein vergrößerter, sondern auch ein kleiner linker Ventrikel ist bei Vorhofflimmern mit einer erhöhten Komplikationsrate assoziiert. Der Zusammenhang besteht nach Daten aus China unabhängig von anderen Risikofaktoren.

Semaglutid bei Herzinsuffizienz: Wie erklärt sich die Wirksamkeit?

17.05.2024 Herzinsuffizienz Nachrichten

Bei adipösen Patienten mit Herzinsuffizienz des HFpEF-Phänotyps ist Semaglutid von symptomatischem Nutzen. Resultiert dieser Benefit allein aus der Gewichtsreduktion oder auch aus spezifischen Effekten auf die Herzinsuffizienz-Pathogenese? Eine neue Analyse gibt Aufschluss.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.