Pencil beam excitation using a 2D spatially selective adiabatic T2-prep for imaging the left coronary arterial system

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Sonderheft 1/2014

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Background

2D spatially selective radiofrequency (RF) pulses can be used to constrain the location from which an MR signal is obtained. This may lead to more time-efficient data collection by reducing the field of view (FoV) or may improve image quality by suppressing artefacts from outside the area of interest. Meanwhile, T2-Preparation, or T2-Prep, is a magnetization preparation scheme used to improve blood/myocardium contrast. We propose incorporating a "pencil-beam" 2D RF pulse into a T2-Prep module, so as to produce a "2D T2-Prep" that combines T2-weighting with an intrinsic spatial selectivity.

Methods

The first RF pulse of a +90°,180°,180°,-90° adiabatic T2-Prep was replaced with a jinc-shaped RF pulse and spiral gradients (Figure 1), which excites a cylindrical volume. Meanwhile, the final RF pulse (-90°) remains non-selective. It thus restores the excited cylinder, while simultaneously tipping down magnetization outside of the cylinder, which is then spoiled. After numerical simulations, phantoms were scanned (not shown), and volume targeted 3D images of the left coronary arterial system were acquired. In experimentally optimizing the protocol, 5 volunteers were scanned. All images were acquired on a 1.5T Siemens Aera using an ECG triggered and navigator gated 3D segmented k-space Cartesian gradient echo sequence, with FoV 432 × 288 mm, matrix size 288 × 192, 1.5 mm reconstructed slice thickness, 24 mm volume thickness, 8 k-space lines/heartbeat, 40 ms T2-Prep, 30° water selective RF excitation pulses, TE/TR/Tacq: 5.95/10.86/86.0 ms. Full FoV images were acquired with both the conventional T2-Prep and the 2D T2-Prep. The 2D T2-Prep scan was then repeated with signal foldover and a reduced FoV (rFoV: 120 × 120 mm, matrix size 80 × 80). All images were reformatted and analyzed using Soap-Bubble. The 3 T2-Preps (conventional, 2D, and 2D rFoV) were compared using vessel sharpness measurements, contrast-to-noise (CNR), and signal-to-noise (SNR) quantification in the blood and myocardium.

Results

The simulated excitation profile of the 2D pulse can be seen in Figure 1, with sample in vivo images shown in Figure 2. In comparing the T2-Preps, the 2D T2-Prep's mean blood and myocardium SNRs (580.2 and 191.2, respectively) stayed within the same range as the conventional T2-Prep's (606.1 and 191.4). The 2D T2-Prep also preserved blood-myocardium CNR (388.1) as compared to the conventional T2-Prep (415.8), demonstrating that the 2D T2-Prep adequately maintains T2-weighting. Vessel sharpness was also comparable (58.4% using the conventional T2-Prep vs. 58.5% using the 2D T2-Prep). In going to a reduced FoV, the mean blood and myocardium SNR decreased slightly (453.9 and 141.8), as did CNR (312.0) and vessel sharpness (56.4%), but scan time was reduced by 60%.

Conclusions

A 2D T2-Prep shows promise to both reduce respiratory motion artefacts and decrease scan time, while preserving the T2-weighting of a conventional T2-Prep. It should thus be considered as a potential tool in accelerated cardiac imaging.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Metadaten
Titel
Pencil beam excitation using a 2D spatially selective adiabatic T2-prep for imaging the left coronary arterial system
Publikationsdatum
01.01.2014
DOI
https://doi.org/10.1186/1532-429X-16-S1-W34

Weitere Artikel der Sonderheft 1/2014

Journal of Cardiovascular Magnetic Resonance 1/2014 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.