Skip to main content
Erschienen in: Malaria Journal 1/2021

Open Access 01.12.2021 | Research

Performance of a sensitive haemozoin‐based malaria diagnostic test validated for vivax malaria diagnosis in Brazilian Amazon

verfasst von: Gisely Cardoso de Melo, Rebeca Linhares Abreu Netto, Victor Irungu Mwangi, Yanka Evellyn Alves Rodrigues Salazar, Vanderson de Souza Sampaio, Wuelton Marcelo Monteiro, Fernando Fonseca de Almeida e Val, Anne Rocheleau, Priyaleela Thota, Marcus Vinícius Guimarães Lacerda

Erschienen in: Malaria Journal | Ausgabe 1/2021

Abstract

Background

Vivax malaria diagnosis remains a challenge in malaria elimination, with current point of care rapid diagnostic tests (RDT) missing many clinically significant infections because of usually lower peripheral parasitaemia. Haemozoin-detecting assays have been suggested as an alternative to immunoassay platforms but to date have not reached successful field deployment. Haemozoin is a paramagnetic crystal by-product of haemoglobin digestion by malaria parasites and is present in the food vacuole of malaria parasite-infected erythrocytes. This study aimed to compare the diagnostic capability of a new haemozoin-detecting platform, the Gazelle™ device with optical microscopy, RDT and PCR in a vivax malaria-endemic region.

Methods

A comparative, double-blind study evaluating symptomatic malaria patients seeking medical care was conducted at an infectious diseases reference hospital in the western Brazilian Amazon. Optical microscopy, PCR, RDT, and Gazelle™ were used to analyse blood samples. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and Kappa values were calculated.

Results

Out of 300 patients, 24 test results were excluded from the final analysis due to protocol violation (6) and inconclusive and/or irretrievable results (18). Gazelle™ sensitivity was 96.1 % (91.3–98.3) and 72.1 % (65.0–78.3) when compared to optical microscopy and PCR, respectively whereas it was 83.9 % and 62.8 % for RDTs. The platform presented specificity of 100 % (97.4–100), and 99.0 % (94.8–99.9) when compared to optical microscopy, and PCR, respectively, which  was the same for RDTs. Its correct classification rate was 98.2 % when compared to optical microscopy and 82.3 % for PCR; the test’s accuracy when compared to optical microscopy was 98.1 % (96.4–99.7), when compared to RDT was 95.2 % (93.0–97.5), and when compared to PCR was 85.6 % (82.1–89.1). Kappa (95 % CI) values for Gazelle™ were 96.4 (93.2–99.5), 88.2 (82.6–93.8) and 65.3 (57.0–73.6) for optical microscopy, RDT and PCR, respectively.

Conclusions

The Gazelle™ device was shown to have faster, easier, good sensitivity, specificity, and accuracy when compared to microscopy and was superior to RDT, demonstrating to be an alternative for vivax malaria screening particularly in areas where malaria is concomitant with other febrile infections (including dengue fever, zika, chikungunya, Chagas, yellow fever, babesiosis).
Hinweise

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
AUC
Area under curve
DNA
Deoxyribonucleic acid
FAP
Full analysis population
FMT-HVD
Fundação de Medicina Tropical Heitor Vieira Dourado
GZL
Gazelle™
LED
Light emitting diode
LR
Likelihood ratios
LOD
Limit of detection
LAMP
Loop mediated isothermal amplification
MOD
Magnetic-optical detection
NATs
Nucleic acid tests
PPP
Per protocol population
PVP
Plasmodium vivax population
POC
Point of care
PCR
Polymerase chain reaction
PPV
Positive predictive value
RDT
Rapid diagnostic test
ROC
Receiver-operating characteristics
STARD
Standards for reporting of diagnostic accuracy studies
TBS
Thick blood smear

Background

Malaria is the leading cause of death from parasitic infection worldwide [1]. Achieving control and elimination of malaria requires opportune and accurate determination of Plasmodium infection and, to an extent, differentiation of species, with the capacity for rapid screening becoming increasingly important for elimination. While malaria caused by Plasmodium falciparum, the principal cause of malaria mortality worldwide, is relatively well diagnosed in most areas using antigen-detecting rapid diagnostic tests (RDTs), other species are less well detected, particularly in remote settings [2]. Plasmodium vivax, the second most prevalent species of the six human malaria-causing Plasmodium species, is concentrated in Asia, Central and South America, the Western Pacific and the Horn of Africa [3] where it also imparts a significant health burden [4]. In 2019, 89.3 % of malaria cases reported in Brazil were of vivax malaria [5]. In many areas where both P. falciparum and Plasmodium vivax cause disease, malaria control efforts have effectively reduced the number of P. falciparum infections, but there has been an increase in the proportion of infections attributed to P. vivax [6].
Diagnosis, control and elimination of P. vivax is complicated by hard-to-detect low densities of circulating blood-stage parasites in many symptomatic patients and by its undetectable latent liver stage (hypnozoites) that results in relapses if not specifically targeted with 8-aminoquinolone drugs. Parasite densities of P. vivax are commonly much lower than P. falciparum in symptomatic infections [7], limiting the sensitivity of RDTs [8].
Because RDTs have poor sensitivity and molecular-based diagnosis are expensive for detecting some Plasmodium species [9], light microscopy remains the gold standard for malaria diagnosis in most parts of the malaria-endemic world [10]. However, its use depends on the availability of a well-functioning light microscope, electrical power, clean glass slides, immersion oil with appropriate optical properties, freshly filtered reagents for Giemsa staining, and importantly, a skilled microscopist. It is time-consuming and its accuracy diminishes in low parasitaemia [10].
Similarly, the performance of an RDT can be affected by multiple host and parasite factors [11]. Most RDTs rely on detection of parasite-derived protein histidine-rich protein 2 (PfHRP2), expressed only in P. falciparum, and the parasite’s metabolic enzymes, lactate dehydrogenase (pLDH) and aldolase, which are common to all species. HRP2-detecting tests are of limited utility in some regions because of their inability to detect P. falciparum parasites with hrp2 and hrp3 gene deletions [12]. The sensitivity of pLDH or aldolase-based RDTs for the detection of P. vivax infection is generally lower than that of PfHRP2-based RDTs for P. falciparum [8, 13]. Although polymerase chain reaction (PCR) is highly sensitive and specific for the detection and identification of malarial parasites [14], it requires sophisticated laboratory equipment and highly trained personnel [15]. For these reasons, the use of PCR is effectively limited to research settings and is not a viable option in the field. The loop-mediated isothermal amplification (LAMP) technique is the most commonly used isothermal strategy for malaria diagnosis. Commercial LAMP systems mainly rely on turbidity measurements, as these nucleic acid testing (NAT) reactions produce a large amount of DNA that reacts to produce magnesium pyrophosphate precipitate as a by-product [16, 17]. However, the existent commercially available LAMP-based malaria diagnostics are not yet fully portable, requiring laboratory instruments, high power use, specific reaction temperature settings, previous sample processing, and specialized training. At the moment, LAMP instruments are only bench-top apparatus and require miniaturization to fully meet the desirable qualities of an ideal commercial malaria diagnostic device [18]. Faster, easier and more sensitive diagnostics are indispensable for screening and diagnosis of P. vivax, for successful case management and in achieving malaria elimination [19].
Haemozoin is a paramagnetic crystal by-product of haemoglobin digestion by malaria parasites [20]. Because haemozoin is a highly specific biomarker for malaria infection, present in all Plasmodium species but in no other circulating pathogens, its detection is a promising approach to accurate diagnosis, especially in low parasitaemia. Attempts to develop portable, magneto-optical, point-of-care diagnostic technologies that detect haemozoin have been made but no commercial product has yet emerged. Earlier efforts to use haemozoin as a biomarker for malaria were largely unsuccessful [21, 22]. In addition, the parasite stage targeted influenced the results, such as immature forms that were established to have little or no detectable haemozoin [23]. Other methods lacked adequate specificity, minimum detection levels, had increased errors when detecting by resonance due to increased noise in the system, or were expensive [18, 2022, 24]. The current device is vastly improved. It uses a very small volume of peripheral blood, has a lysis step that releases all the haemozoin into solution, and has powerful magnets that produce a magnetic field to align the haemozoin to inhibit the transmission of light through the solution. However, the size and morphology of haemozoin crystals may hold the key to identify accurately and differentiate species of malaria. Kumar et al.. observed that the sensitivity and specificity of Gazelle™ was 98 and 97 % compared to light microscopy, 82 and 99 % to PCR, and 78 and 99 % to RDT, respectively, in a predominantly P. falciparum endemic area [25].
Here, are reported the results from a study that assessed the accuracy of Gazelle™ (Hemex Health, USA), in detecting vivax malaria cases. The device is a late-stage portable prototype using magneto-optic technology for malaria diagnosis developed to detect haemozoin. The performance of Gazelle™ was compared with three current diagnostic methods: microscopy, RDTs and PCR.

Methods

A comparative, double-blind study evaluating symptomatic individuals with suspected malaria, comparing optical microscopy, PCR, RDT, and Gazelle™ was carried out between June and October 2019 at Fundação de Medicina Tropical Doutor Heitor Vieira Dourado (FMT-HVD). The hospital is a reference centre for infectious disease care in Manaus, western Brazilian Amazon. Approximately 30 % of all cases of malaria in Manaus municipality are diagnosed at the FMT-HVD.

Study subjects

Both male and female individuals aged ≥ 18 years seeking care at the FMT-HVD outpatient clinic were considered for enrolment. Per protocol, subjects were recruited, screened and enrolled by the investigator’s team at the study site, after the hospital’s routine malaria screening. All patients were informed of the study objectives and risks of participation. Participants were given adequate time to read the informed consent form (ICF) after which they were requested to sign it. Exclusion criteria were severe malaria, pregnancy and use of anti-malarials in the 30 days prior to study enrolment.

Sample size

The sample size was calculated based on the assumption of 95 % confidence interval, an estimated 10 % accuracy (95 % CI width ≤ 10 %) and allowed for up to 10 % of subjects to be excluded because of incomplete or missing data. This translated to a minimum of 100 positive and 200 negative cases of malaria.

Sample analysis

On study admission, immediately before drug administration, trained laboratory technicians collected whole blood samples in EDTA. Microscopy was the method used to confirm vivax malaria status of the participants before recruiting them into the study. Blinded study microscopists analysed all samples after slide preparation (thick and thin smears). Collected blood samples were stored at 4 °C and tested by RDT and Gazelle™ within utmost 4 h of collection. Aliquots of the whole blood samples were stored at − 20 °C until needed for DNA extraction for the subsequent molecular diagnosis of malaria by PCR analysis. All testers were blinded to the malaria status of the samples.

Gazelle™

The Gazelle™ technology is housed in a portable, point-of-care device (Fig. 1) and detects malaria infections based on the magneto-optical concept for the detection of paramagnetic haemozoin, a biomarker that appears very early during the life cycle of the malaria parasite. This being a prototype device, at the moment it cannot distinguish between P. falciparum and P. vivax haemozoin. When haemozoin, which is partly composed of iron, is subjected to an intense magnetic field the haemozoin crystals align in a manner that increases the opacity of the blood sample (Fig. 2). The opacity of the sample is measured in both a high magnetic field and a very low or no magnetic field using a LED and photodetector to determine the presence of haemozoin. The decrease of light reaching the photodetector in the presence of the high magnetic field is directly proportional to the amount of haemozoin present. If there is no change with variation in the magnetic field, then there is no haemozoin detected. An algorithm uses the optical information to determine the presence or absence of malaria.
For Gazelle™ testing, blood samples were prepared by pipetting 80 µL of malaria diluent (2 % Triton in water) and 15 µL of blood into a cuvette (Fig. 3). The cuvette containing the diluent/blood mixture is then loaded into the reader that automatically analyses the samples. Results appear on the Gazelle™ screen in about one minute and can either be stored in the reader (has internal storage), locally printed, or transmitted through Bluetooth or Wi-Fi connection. Another feature of this diagnostic device is that it can operate on either electric or lithium battery power, making it versatile for point-of-care use in remote locations with no electricity. This device can be recharged using a standard micro-USB charger similar to that used for android phones. Voltage drift threshold was set at 0.9 V for this study. The algorithm has limits to prevent the use of samples with too high voltage drift. The reader and disposables do not require special storage and can operate in temperatures of up to 45 °C [25].

RDT

Venous blood samples were tested using the Malaria Ag Pf/Pan RDT (SD Bioline, Standard Diagnostics Inc., Gyeonggi-do, Republic of Korea) according to the manufacturer’s instructions. The RDT kits were stored at the recommended temperatures and were used before the expiration date.

Microscopy

Thick blood smears were prepared, stained by Giemsa [26] and examined by two experienced microscopists, who counted the number of parasites per 200 or 500 leukocytes, depending on parasite density. Parasite density (parasites/µL) was determined by counting the number of parasites per leukocyte in high magnification fields with the assumption of 6,000 leukocytes/µL of blood [27]. In the event of a discrepancy > 10 % between the two, a third reading by an independently trained microscopist was performed using the Obare Method Calculator (version 1.0) to resolve the discrepancy [28].

PCR

Extraction of total DNA from 200 µL of whole blood was performed using the QIAamp DNA Blood Mini Kit (Qiagen®, USA), according to the manufacturer’s protocol. QMAL Taqman qPCR was used to detect Plasmodium species by targeting a conserved region of the 18 S rRNA gene in the extracted DNA samples. The following probes and primers were used to analyse the extracted DNA samples: Fw-TTA GAT TGC TTC CTT CAG TRC CTT ATG; Rev-GT TGA GTC AAA TTA AGC CGC AA; FAM - TCA ATT CTT TTA ACT TTC TCG CTT GCG CGA - BHQ1 (wobble R = A/G) [29]. To quantify the 18 S rRNA gene copy numbers in each experiment, three dilutions of plasmids containing the respective targeted region were included in triplicates (102, 104, 106 copies/µL). Briefly, 4 µL of DNA samples and plasmids were used in a total reaction volume of 12 µL to screen for malaria cases.
All qPCR assays were run in the 7500 Fast Real-Time PCR System (Applied Biosystems, CA, USA). The primer and probe sequences, the composition of reaction mixes, PCR profiles and the detection limit for each assay used were as described by Almeida et al. [7].

STARD adherence

STARD (Standards for Reporting of Diagnostic Accuracy Studies) shown in Fig. 4 details study test results, including an analysis of accuracy compared with reference standards for microscopy and PCR (true and false positives and negatives), and comparison of results with RDT as a non-reference comparator (concordance and discordance between Gazelle™ and RDT results).

Ethical considerations

The Ethics Review Board of the Fundação de Medicina Tropical Dr. Heitor Vieira Dourado (FMT-HVD) approved the study (approval number 3.374.472/2019). Subjects were informed of the study objectives and gave their consent for participation through written and signed informed consent forms before study procedures. Subjects diagnosed with malaria were treated according to the Brazilian Ministry of Health guidelines [30].

Analysis populations

Table 1 describes subject enrolment and defines three analysis populations. The full analysis population (FAP) includes all enrolled subjects tested. The per protocol population (PPP) includes all subjects with conclusive Gazelle™ test results. The P. vivax population (PVP) excludes all subjects with P. falciparum or mixed infections. The PVP is the primary population analysed in this study.
Table 1
Enrolment and analysis populations
Candidate subjects presenting for enrolment in the study
300
Total number of subjects enrolled and tested (full analysis population)
300
Subjects excluded from full analysis population
18
Reason: Data file not retrievable
1
Reason: Inconclusive Gazelle™ test result
17
Subject tested negative for malaria by other methods
12
Subject tested positive for P. vivax by other methods
3
Subject tested positive for P. falciparum by other methods
2
Number of subjects included in full data analysis (per protocol population)
282
Additional subjects excluded from the P. vivax data analysis subset
6
Reason: Subject tested positive for P. falciparuma
3
Reason: Subject tested positive for P. falciparum and P. vivax
3
Number of subjects included inP. vivaxdata analysis sub-set (P. vivaxpopulation)
276
aThese 3 subjects infected with P. falciparum are in addition to the 2 subjects with P. falciparum excluded from the full analysis due to inconclusive Gazelle™ test results

Statistical analysis

Gazelle™ diagnostic performance was measured by determining its sensitivity, specificity, likelihood ratios (LR), positive predictive value (PPV), negative predictive value (NPV), and detection accuracy. Gazelle™ was compared to RDT, to optical microscopy and PCR. Likelihood ratios for positive (LR+) and negative (LR−) test results were considered good when LR + was > 10, and LR− <0.1 [31]. Diagnostic accuracy was assessed through receiver-operating characteristics (ROC) curves and areas under the curve (AUC) interpreted as follows: 0.9-1.0, excellent; 0.8–0.9, very good; 0.7–0.8, good; 0.6–0.7, sufficient; 0.5–0.6, bad; <0.5, test not useful [32].
The two-graph ROC analysis enables the visualizing of sensitivity and specificity curves on a single graph, according to the range of values of the new test given the reference test. This provides a visualization for the specificity-sensitivity ‘optimal’’ trade-off in determining the best and acceptable cut-off for the new test to be used in practice against the already established gold standards available. The parasite detection parameters for the Gazelle™ test against microscopy and PCR were determined by a two-graph ROC analysis [32]. Results were considered statistically significant for p < 0.05. Analyses were performed on STATA (v13.1, StataCorp, College Station, TX, USA).

Results

Demographics

From the 300 patients that were tested, 24 samples were excluded from the final analysis: 1 had no results because of a corrupted data file, which was irretrievable; 17 due to inconclusive results (voltage drift, cartridge and/or blood contamination); 3 were infected with P. falciparum; and, 3 with mixed infections of P. vivax and P. falciparum (Table 1). Since this was a double-blind study, no repeat tests were performed on the 18 inconclusive and/or irretrievable test results that were excluded. The on-site study team were blind to the Gazelle™ test results when doing the tests.
Of the 276 subjects whose test results were analysed per protocol, 163 (59.1 %) were male. The study participants were aged between 18 and 79 years with a mean age of 40.4 (± 18.8) years. Out of the total number of subjects included in the P. vivax data analysis, 168 (60.9 %) reported one or more previous episodes of malaria prior to the current infection. The mean parasite density was 6,220 parasites/µL (± 8,254). Patient demographic and malaria infection characteristics are presented in Table 2.
Table 2
Demographic and malaria parasite characteristics – Plasmodium vivax population
Characteristic
Value, n (%)
Gender
 Male
163 (59.1)
 Female
113 (40.9)
Age (mean, SD)
40.4 (± 18.8)
Previous malaria episode
 Yes
168 (60.9)
 No
108 (39.1)
 Parasite density per µL (mean, SD)
6,220 (± 8,254)
 Parasite range (min, max)
18–73, 815
Parasitaemia density distribution (parasite/µL)
 < 1,000
34 (26.1)
 1,000–10,000
90 (69.2)
 > 10,000
6 (4.6)
n number of subjects, min  minimum value, max  maximum value, SD  standard deviation

Malaria detection

Malaria detection characteristics such as mean parasite density, range and density distribution are shown in Table 2. Of the 276 subjects in the PVP, malaria was detected by RDT in 109 (39.5 %) subjects, by microscopy in 130 (47.1 %) subjects, by PCR in 172 (62.3 %) subjects, and by Gazelle™ in 125 (45.3 %) subjects.

Microscopy as gold standard versus Gazelle™ and RDT

When microscopy was positive for malaria, Gazelle™ produced a positive result for 96.2 % of the samples (125 of 130 tests). On the other hand, when microscopy was negative for malaria, the Gazelle™ result was 100 % consistent (146 of 146 tests). A ROC analysis of Gazelle™ against microscopy showed an accuracy of 98.3 % (Fig. 5a). In the five instances when the Gazelle™ result was negative and microscopy was positive, the microscopically observed parasite counts were low, ranging from 18 to 174 parasites/µL. For the tests where microscopy was positive and RDT was negative, the range of parasitaemia varied between 18 and 2,805 parasites/µL.
Sensitivity and specificity of Gazelle™ test results compared with optical microscopy were 96.2 and 100 %, whereas for RDTs they were 83.9 and 100 %, respectively (Table 3). PPV for Gazelle™ and RDTs, when compared to microscopy, were both 100 % whereas the NPV was 96.7 % for Gazelle™ and 87.4 % for RDTs, respectively. The overall accuracy of Gazelle™ was 98.2 % whereas it was 92.4 % for RDTs (Table 3).
Table 3
Performance of Gazelle™ compared to RDT with microscopy as gold standard
n = 276
Gazelle™
RDT
Sensitivity
96.2 %
125/130
83.9 %
109/130
Specificity
100 %
146/146
100 %
146/146
Positive predictive value
100 %
100 %
Negative predictive value
96.7 %
87.4 %
Likelihood ratio
of a negative test
0.04
0.16
Accuracy
(ROC analysis)
98.2 %
271/276
92.4 %
255/276

PCR as gold standard versus Gazelle™, RDT and microscopy

When PCR was positive for malaria, Gazelle™ was positive 72.1 % of the time (124 of 172 tests). When PCR was negative for malaria, the Gazelle™ was negative 99.0 % of the time (103 of 104 tests). Further ROC analysis of Gazelle™ against PCR showed an AUC of 0.837, meaning that the Gazelle™ test was slightly less sensitive than PCR but very specific and with very good accuracy (Fig. 5b). Of the 48 tests positive by PCR and negative by Gazelle™, all were negative by RDT, and 5 were positive by microscopy.
Sensitivity and specificity of Gazelle™ test results compared with PCR were 72.1 and 99 %, whereas, for RDTs they were 62.8 and 99 %, and for microscopy, they were 75 and 99 %, respectively (Table 4). The overall accuracy of Gazelle™ was 82.3 % whereas it was 84.1 % for microscopy and 76.5 % for RDTs.
Table 4
Performance of Gazelle™ compared to microscopy and RDT with PCR as gold standard
n = 276
Gazelle™
Microscopy
RDT
Sensitivity
72.1 %
124/172
75.0 %
129/172
62.8 %
108/172
Specificity
99.0 %
103/104
99.0 %
103/104
99.0 %
103/104
Positive predictive value
99.2 %
99.2 %
99.1 %
Negative predictive value
68.2 %
70.6 %
61.7 %
Likelihood ratio
of a positive test
75.0
78.0
65.3
Likelihood ratio
of a negative test
0.28
0.25
0.38
Accuracy
(ROC analysis)
82.3 %
84.1 %
76.5 %
For the sub-set of 48 tests where the PCR result was positive and Gazelle™ was negative, the average PCR count was 326.65 DNA copies/µL (or only 0.014 % of the average PCR count) with a median of 19.73 DNA copies/µL.

Discussion

Accuracy is essential for rapid and timely diagnosis and treatment, identifying and interrupting malaria transmission in the community. Low parasite density infection continues to challenge diagnostic accuracy and obstruct malaria elimination efforts. Highly sensitive, rapid and affordable diagnostic tools are urgently needed to detect and improve the clinical management of P. vivax infections in endemic regions.
Gazelle™, a novel point-of-care, haemozoin-based malaria diagnostic device was evaluated, and compared to microscopy, RDT, and PCR. Sensitivity and specificity of Gazelle™ was nearly as sensitive and was as specific as expert microscopy and considerably more sensitive than RDTs, the most used diagnostic tools at community health centres. In addition, Gazelle™ had advantages in terms of speed, with a turn-around time of 90 s. Although relatively low cost, microscopy is rarely used in remote areas because infrastructure and skilled personnel are lacking, and electrical power supplies are often erratic or unreliable [33]. Furthermore, microscopy is labour-intensive, time-consuming and operator-dependent compared to RDTs [34, 35].
From this field trial study, the Gazelle™ outperformed RDTs, detecting P. vivax parasite densities down to 72 parasites/µL. These results affirmed the superiority of the test against widely used conventional diagnostic methods. Studies by Kumar et al. [25] demonstrated that the Gazelle™ test device detected as low as 50 cultured P. falciparum parasites/µL with 95 % accuracy; as low as 35 P. vivax parasites/µL from Indian patients could be detected with 100 % accuracy by the device. The difference in the limit of detection (LOD) between this study and the Indian study may be due to differences in parasite burden and parasite characteristics  of the P. vivax strains prevalent in the two regions. Dilution experiments elsewhere, applying the magneto-optical concept used in this device, suggest a LOD for Gazelle™ could go to as low as potentially < 20 parasites/µL with higher blood volume and trophozoite/schizont parasite stages [36]. These LODs, which correspond to approximately 20 pg/µL of haemozoin, are better than that of RDTs and compare favourably with the LOD achievable by expert microscopy [36, 37]. The cost of the test on the device is about $1 per test, and very much comparable to the cost of a single RDT and microscopy [25].
An attempt to determine the device cut-off limit relative to microscopy and PCR diagnostic techniques was also done. The parasitaemia and PCR copies raw data obtained from laboratory analysis ranged into thousands (parasitaemia) and millions (PCR copies). The identification of a precise cut-off using a cut-off graph plot was however based on too many points with little gain in accuracy. Subsequently, intervals of 500 were used, making 500,000 PCR copies/µL as the maximum value. With this modification, the “optimal cut-off” for GazelleTM test use was determined to be slightly lower than 100 parasites/µL and 1,000 copies/µL relative to microscopy and PCR, respectively. Further studies in different malaria transmission settings are necessary to validate the proposed “optimal cut-off” and LOD.
The current Gazelle™ prototype provides only qualitative malaria diagnosis without species differentiation, with detection of P. falciparum demonstrated elsewhere [25]. In its current form, it can be used in POC (point of care) to quickly screen for malaria infections from other tropical diseases that present with similar symptoms, including dengue, zika, chikungunya, Chagas, yellow fever, babesiosis. The successful incorporation of haemozoin’s magnetic properties and magneto-optics into a portable diagnostic device should help in the fast and accurate diagnosis of malaria cases in the community, while the capacity for species differentiation currently under development is necessary for management in areas with multiple endemic species. Future improved versions of similar Magnetic-optical detection (MOD) -based diagnostics systems are likely to be more specific, possibly using differing haemozoin crystal morphology to distinguish species [38]. Perhaps the size and morphology of haemozoin crystals may hold the key to accurately identifying and differentiating parasite developmental stages and species. The short time to result and higher sensitivity for P. vivax have potential to address the inadequacies of current RDTs in community-based vivax management and may prove useful in scenarios such as airports or borders in elimination settings where large numbers of people must be rapidly screened each day.
For operational purposes, the samples used for all diagnostic tests were venous blood. For malaria diagnosis in the field peripheral capillary blood is used, which could be a limitation of the study’s current sample collection and analysis design in that malaria parasite concentrations in capillary and venous blood differ. However, the diagnostic device can use small volumes of blood such as those obtained from finger prick (15 µL) for analysis. Despite all tests having been performed  with venous samples, future studies will need to address this issue by comparing the diagnostic potential of Gazelle™’s device  using blood from both collection sites. Expanding the capacity to diagnose and differentiate between species should increase  the Gazelle™ use as a POC testing device in field settings where mixed infections occur, or in areas where several species are in circulation. The results obtained in this study could be restricted to the malaria strains present in the Brazilian Amazon, and the application of this device in other countries should be addressed. Despite these limitations, results from this study support the use of Gazelle™ in areas with limited qualified human resources and electrical power supply, and thereby increase capacity of malaria diagnosis in areas where transmission mostly occurs.

Conclusions

This study demonstrated that Gazelle™, the first commercially developed haemozoin-detecting instrument for the diagnosis of malaria, outperformed RDTs and was nearly as sensitive and as specific as expert microscopy for P. vivax. Gazelle™ demonstrated a limit of detection below that of the comparator rapid test. Its ease of use, short time to result, and accuracy hold potential for rapid screening for P. vivax in remote resource-poor locations where PCR and microscopy are not feasible and cannot immediately access to modern medical attention, but further development is required to differentiate species where necessary to guide treatment. In addition, its speed, storage capacity, rechargeable battery, cost-efficiency, and alternative to lack of microscopists makes it very practical for use in field settings. Even so, more studies are required to determine the device’s performance against different malaria species (and co-infections), using peripheral blood samples and in the field where energy and cold chains are not available.

Acknowledgements

We are grateful to all subjects who participated in the study. We also acknowledge the outpatient clinic staff at FMT-HVD for assistance in patient management.

Declarations

The FMT-HVD Ethics Committee (EC), Manaus, Brazil, approved this study (3.374.472/2019).

Competing interests

PT and AR are employees of Hemex Health. The other authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat WHO. World Malaria Report. Geneva, World Health Organization, 2019. WHO. World Malaria Report. Geneva, World Health Organization, 2019.
2.
Zurück zum Zitat Tanizaki R, Kato Y, Iwagami M, Kutsuna S, Ujiie M, Takeshita N, et al. Performance of rapid diagnostic tests for Plasmodium ovale malaria in Japanese travellers. Trop Med Health. 2014;42:149–53.CrossRef Tanizaki R, Kato Y, Iwagami M, Kutsuna S, Ujiie M, Takeshita N, et al. Performance of rapid diagnostic tests for Plasmodium ovale malaria in Japanese travellers. Trop Med Health. 2014;42:149–53.CrossRef
3.
Zurück zum Zitat Twohig KA, Pfeffer DA, Baird JK, Price RN, Zimmerman PA, Hay SI, et al. Growing evidence of Plasmodium vivax across malaria-endemic Africa. PLoS Negl Trop Dis. 2019;13:e0007140.CrossRef Twohig KA, Pfeffer DA, Baird JK, Price RN, Zimmerman PA, Hay SI, et al. Growing evidence of Plasmodium vivax across malaria-endemic Africa. PLoS Negl Trop Dis. 2019;13:e0007140.CrossRef
4.
Zurück zum Zitat Price RN, Commons RJ, Battle KE, Thriemer K, Mendis K. Plasmodium vivax in the era of the shrinking P. falciparum map. Trends Parasitol. 2020;36:560–70.CrossRef Price RN, Commons RJ, Battle KE, Thriemer K, Mendis K. Plasmodium vivax in the era of the shrinking P. falciparum map. Trends Parasitol. 2020;36:560–70.CrossRef
5.
Zurück zum Zitat WHO. World Malaria Report: 20 years of global progress and challenges. Geneva, World Health Organization. 2020. 238 p. WHO. World Malaria Report: 20 years of global progress and challenges. Geneva, World Health Organization. 2020. 238 p.
6.
Zurück zum Zitat Rizvi I, Tripathi D, Zaman S, Zaidi N, Chughtai A, Beg M. Complications associated with Plasmodium vivax malaria: a retrospective study from a tertiary care hospital based in western Uttar Pradesh, India. Ann Afr Med. 2013;12:155–9.CrossRef Rizvi I, Tripathi D, Zaman S, Zaidi N, Chughtai A, Beg M. Complications associated with Plasmodium vivax malaria: a retrospective study from a tertiary care hospital based in western Uttar Pradesh, India. Ann Afr Med. 2013;12:155–9.CrossRef
7.
Zurück zum Zitat Almeida ACG, Kuehn A, Castro AJM, Vitor-Silva S, Figueiredo EFG, Brasil LW, et al. High proportions of asymptomatic and submicroscopic Plasmodium vivax infections in a peri-urban area of low transmission in the Brazilian Amazon. Parasit Vectors. 2018;11:194.CrossRef Almeida ACG, Kuehn A, Castro AJM, Vitor-Silva S, Figueiredo EFG, Brasil LW, et al. High proportions of asymptomatic and submicroscopic Plasmodium vivax infections in a peri-urban area of low transmission in the Brazilian Amazon. Parasit Vectors. 2018;11:194.CrossRef
8.
Zurück zum Zitat Abba K, Kirkham AJ, Olliaro PL, Deeks JJ, Donegan S, Garner P, et al. Rapid diagnostic tests for diagnosing uncomplicated nonfalciparum or Plasmodium vivax malaria in endemic countries. Cochrane Database Syst Rev. 2014;2:CD011431. Abba K, Kirkham AJ, Olliaro PL, Deeks JJ, Donegan S, Garner P, et al. Rapid diagnostic tests for diagnosing uncomplicated nonfalciparum or Plasmodium vivax malaria in endemic countries. Cochrane Database Syst Rev. 2014;2:CD011431.
9.
Zurück zum Zitat WHO. Guidelines for the treatment of malaria. 3rd edition. Annex 4, Grading of Recommendations Assessment, Development and Evaluation (GRADE) for assessing the quality of evidence. Geneva: World Health Organization; 2015. WHO. Guidelines for the treatment of malaria. 3rd edition. Annex 4, Grading of Recommendations Assessment, Development and Evaluation (GRADE) for assessing the quality of evidence. Geneva: World Health Organization; 2015.
10.
Zurück zum Zitat Mukry SN, Saud M, Sufaida G, Shaikh K, Naz A, Shamsi TS. Laboratory diagnosis of malaria: comparison of manual and automated diagnostic tests. Can J Infect Dis Med Microbiol. 2017;2017:9286392.CrossRef Mukry SN, Saud M, Sufaida G, Shaikh K, Naz A, Shamsi TS. Laboratory diagnosis of malaria: comparison of manual and automated diagnostic tests. Can J Infect Dis Med Microbiol. 2017;2017:9286392.CrossRef
11.
Zurück zum Zitat Fransisca L, Kusnanto JH, Satoto TBT, Sebayang B, Supriyanto, Andriyan E, et al. Comparison of rapid diagnostic test Plasmotec Malaria-3, microscopy, and quantitative real-time PCR for diagnoses of Plasmodium falciparum and Plasmodium vivax infections in Mimika Regency, Papua, Indonesia. Malar J. 2015;14:103.CrossRef Fransisca L, Kusnanto JH, Satoto TBT, Sebayang B, Supriyanto, Andriyan E, et al. Comparison of rapid diagnostic test Plasmotec Malaria-3, microscopy, and quantitative real-time PCR for diagnoses of Plasmodium falciparum and Plasmodium vivax infections in Mimika Regency, Papua, Indonesia. Malar J. 2015;14:103.CrossRef
12.
Zurück zum Zitat Thomson R, Beshir KB, Cunningham J, Baiden F, Bharmal J, Bruxvoort KJ, et al. pfhrp2 and pfhrp3 gene deletions that affect malaria rapid diagnostic tests for Plasmodium falciparum: analysis of archived blood samples from 3 African countries. J Infect Dis. 2019;220:1444–52.CrossRef Thomson R, Beshir KB, Cunningham J, Baiden F, Bharmal J, Bruxvoort KJ, et al. pfhrp2 and pfhrp3 gene deletions that affect malaria rapid diagnostic tests for Plasmodium falciparum: analysis of archived blood samples from 3 African countries. J Infect Dis. 2019;220:1444–52.CrossRef
13.
Zurück zum Zitat Jimenez A, Rees-Channer RR, Perera R, Gamboa D, Chiodini PL, González IJ, et al. Analytical sensitivity of current best-in-class malaria rapid diagnostic tests. Malar J. 2017;16:128.CrossRef Jimenez A, Rees-Channer RR, Perera R, Gamboa D, Chiodini PL, González IJ, et al. Analytical sensitivity of current best-in-class malaria rapid diagnostic tests. Malar J. 2017;16:128.CrossRef
14.
Zurück zum Zitat Hänscheid T, Grobusch MP. How useful is PCR in the diagnosis of malaria? Trends Parasitol. 2002;18:395–8.CrossRef Hänscheid T, Grobusch MP. How useful is PCR in the diagnosis of malaria? Trends Parasitol. 2002;18:395–8.CrossRef
15.
Zurück zum Zitat Poon LLM, Wong BWY, Ma EHT, Chan KH, Chow LMC, Abeyewickreme W, et al. Sensitive and inexpensive molecular test for falciparum malaria: detecting Plasmodium falciparum DNA directly from heat-treated blood by loop-mediated isothermal amplification. Clin Chem. 2006;52:303–6.CrossRef Poon LLM, Wong BWY, Ma EHT, Chan KH, Chow LMC, Abeyewickreme W, et al. Sensitive and inexpensive molecular test for falciparum malaria: detecting Plasmodium falciparum DNA directly from heat-treated blood by loop-mediated isothermal amplification. Clin Chem. 2006;52:303–6.CrossRef
16.
Zurück zum Zitat Mori Y, Nagamine K, Tomita N, Notomi T. Detection of loop-mediated isothermal amplification reaction by turbidity derived from magnesium pyrophosphate formation. Biochem Biophys Res Commun. 2001;289:150–4.CrossRef Mori Y, Nagamine K, Tomita N, Notomi T. Detection of loop-mediated isothermal amplification reaction by turbidity derived from magnesium pyrophosphate formation. Biochem Biophys Res Commun. 2001;289:150–4.CrossRef
17.
Zurück zum Zitat Tomita N, Mori Y, Kanda H, Notomi T. Loop-mediated isothermal amplification (LAMP) of gene sequences and simple visual detection of products. Nat Protoc. 2008;3:877–82.CrossRef Tomita N, Mori Y, Kanda H, Notomi T. Loop-mediated isothermal amplification (LAMP) of gene sequences and simple visual detection of products. Nat Protoc. 2008;3:877–82.CrossRef
18.
Zurück zum Zitat Kolluri N, Klapperich CM, Cabodi M. Towards lab-on-a-chip diagnostics for malaria elimination. Lab Chip. 2017;18:75–94.CrossRef Kolluri N, Klapperich CM, Cabodi M. Towards lab-on-a-chip diagnostics for malaria elimination. Lab Chip. 2017;18:75–94.CrossRef
19.
Zurück zum Zitat malERA Refresh Consultative Panel on Tools for Malaria Elimination. malERA: An updated research agenda for diagnostics, drugs, vaccines, and vector control in malaria elimination and eradication. PLoS Med. 2017;14:e1002455.CrossRef malERA Refresh Consultative Panel on Tools for Malaria Elimination. malERA: An updated research agenda for diagnostics, drugs, vaccines, and vector control in malaria elimination and eradication. PLoS Med. 2017;14:e1002455.CrossRef
20.
Zurück zum Zitat Rifaie-Graham O, Pollard J, Raccio S, Balog S, Rusch S, Hernández-Castañeda MA, et al. Hemozoin-catalyzed precipitation polymerization as an assay for malaria diagnosis. Nat Commun. 2019;10:1369.CrossRef Rifaie-Graham O, Pollard J, Raccio S, Balog S, Rusch S, Hernández-Castañeda MA, et al. Hemozoin-catalyzed precipitation polymerization as an assay for malaria diagnosis. Nat Commun. 2019;10:1369.CrossRef
21.
Zurück zum Zitat Delahunt C, Horning MP, Wilson BK, Proctor JL, Hegg MC. Limitations of haemozoin-based diagnosis of Plasmodium falciparum using dark-field microscopy. Malar J. 2014;13:147.CrossRef Delahunt C, Horning MP, Wilson BK, Proctor JL, Hegg MC. Limitations of haemozoin-based diagnosis of Plasmodium falciparum using dark-field microscopy. Malar J. 2014;13:147.CrossRef
22.
Zurück zum Zitat Rebelo M, Grenho R, Orban A, Hänscheid T. Transdermal diagnosis of malaria using vapor nanobubbles. Emerg Infect Dis. 2016;22:343–4.CrossRef Rebelo M, Grenho R, Orban A, Hänscheid T. Transdermal diagnosis of malaria using vapor nanobubbles. Emerg Infect Dis. 2016;22:343–4.CrossRef
23.
Zurück zum Zitat Rebelo M, Shapiro HM, Amaral T, Melo-Cristino J, Hänscheid T. Haemozoin detection in infected erythrocytes for Plasmodium falciparum malaria diagnosis-prospects and limitations. Acta Trop. 2012;123:58–61.CrossRef Rebelo M, Shapiro HM, Amaral T, Melo-Cristino J, Hänscheid T. Haemozoin detection in infected erythrocytes for Plasmodium falciparum malaria diagnosis-prospects and limitations. Acta Trop. 2012;123:58–61.CrossRef
24.
Zurück zum Zitat Kim W, Ozdemir SK, Zhu J, Faraz M, Coban C, Yang L. Detection and size measurement of individual hemozoin nanocrystals in aquatic environment using a whispering gallery mode resonator. Opt Express. 2012;20:29426–46.CrossRef Kim W, Ozdemir SK, Zhu J, Faraz M, Coban C, Yang L. Detection and size measurement of individual hemozoin nanocrystals in aquatic environment using a whispering gallery mode resonator. Opt Express. 2012;20:29426–46.CrossRef
25.
Zurück zum Zitat Kumar R, Verma AK, Shrivas S, Thota P, Singh MP, Rajasubramaniam S, et al. First successful field evaluation of new, one-minute haemozoin-based malaria diagnostic device. EClinicalMedicine. 2020;22:100347.CrossRef Kumar R, Verma AK, Shrivas S, Thota P, Singh MP, Rajasubramaniam S, et al. First successful field evaluation of new, one-minute haemozoin-based malaria diagnostic device. EClinicalMedicine. 2020;22:100347.CrossRef
26.
Zurück zum Zitat WHO. Malaria Microscopy Quality Assurance Manual – Version 2. Geneva, World Health Organization, 2016. WHO. Malaria Microscopy Quality Assurance Manual – Version 2. Geneva, World Health Organization, 2016.
27.
Zurück zum Zitat Ministério da Saúde. Manual de Diagnóstico Laboratorial da Malária [Internet]. 2009 [cited 2020 Feb 27]. Ministério da Saúde. Manual de Diagnóstico Laboratorial da Malária [Internet]. 2009 [cited 2020 Feb 27].
28.
Zurück zum Zitat Obare Method Calculator. Worldwide Antimalarial Resistance Network [Internet]. [cited 2020 Feb 14]. Obare Method Calculator. Worldwide Antimalarial Resistance Network [Internet]. [cited 2020 Feb 14].
29.
Zurück zum Zitat Wampfler R, Mwingira F, Javati S, Robinson L, Betuela I, Siba P, et al. Strategies for detection of Plasmodium species gametocytes. PLoS ONE. 2013;8:e76316.CrossRef Wampfler R, Mwingira F, Javati S, Robinson L, Betuela I, Siba P, et al. Strategies for detection of Plasmodium species gametocytes. PLoS ONE. 2013;8:e76316.CrossRef
30.
Zurück zum Zitat Ministério da Saúde. Guia prático de tratamento da malária no Brasil [Internet]. 2010 [cited 2020 Feb 14]. Ministério da Saúde. Guia prático de tratamento da malária no Brasil [Internet]. 2010 [cited 2020 Feb 14].
31.
Zurück zum Zitat Florkowski CM. Sensitivity, specificity, receiver-operating characteristic (ROC) curves and likelihood ratios: communicating the performance of diagnostic tests. Clin Biochem Rev. 2008;29(Suppl 1):83-7. Florkowski CM. Sensitivity, specificity, receiver-operating characteristic (ROC) curves and likelihood ratios: communicating the performance of diagnostic tests. Clin Biochem Rev. 2008;29(Suppl 1):83-7.
32.
Zurück zum Zitat Domenica Cappellini M, Motta I. Anemia in clinical practice-definition and classification: does hemoglobin change with aging? Semin Hematol. 2015;52:261–9.CrossRef Domenica Cappellini M, Motta I. Anemia in clinical practice-definition and classification: does hemoglobin change with aging? Semin Hematol. 2015;52:261–9.CrossRef
33.
Zurück zum Zitat Bell D, Fleurent AE, Hegg MC, Boomgard JD, McConnico CC. Development of new malaria diagnostics: matching performance and need. Malar J. 2016;15:406.CrossRef Bell D, Fleurent AE, Hegg MC, Boomgard JD, McConnico CC. Development of new malaria diagnostics: matching performance and need. Malar J. 2016;15:406.CrossRef
34.
Zurück zum Zitat Jain P, Chakma B, Patra S, Goswami P. Potential biomarkers and their applications for rapid and reliable detection of malaria. BioMed Res Int. 2014;2014:852645.CrossRef Jain P, Chakma B, Patra S, Goswami P. Potential biomarkers and their applications for rapid and reliable detection of malaria. BioMed Res Int. 2014;2014:852645.CrossRef
35.
Zurück zum Zitat Bousema T, Okell L, Felger I, Drakeley C. Asymptomatic malaria infections: detectability, transmissibility and public health relevance. Nat Rev Microbiol. 2014;Vol. 12:12:833–40.CrossRef Bousema T, Okell L, Felger I, Drakeley C. Asymptomatic malaria infections: detectability, transmissibility and public health relevance. Nat Rev Microbiol. 2014;Vol. 12:12:833–40.CrossRef
36.
Zurück zum Zitat Orbán Á, Butykai Á, Molnár A, Pröhle Z, Fülöp G, Zelles T, et al. Evaluation of a novel magneto-optical method for the detection of malaria parasites. PLoS ONE 2014;9:e96981.CrossRef Orbán Á, Butykai Á, Molnár A, Pröhle Z, Fülöp G, Zelles T, et al. Evaluation of a novel magneto-optical method for the detection of malaria parasites. PLoS ONE 2014;9:e96981.CrossRef
37.
Zurück zum Zitat Butykai A, Orbán A, Kocsis V, Szaller D, Bordács S, Tátrai-Szekeres E, et al. Malaria pigment crystals as magnetic micro-rotors: key for high-sensitivity diagnosis. Sci Rep. 2013;3:1431.CrossRef Butykai A, Orbán A, Kocsis V, Szaller D, Bordács S, Tátrai-Szekeres E, et al. Malaria pigment crystals as magnetic micro-rotors: key for high-sensitivity diagnosis. Sci Rep. 2013;3:1431.CrossRef
38.
Zurück zum Zitat Noland GS, Briones N, Sullivan DJ. The shape and size of hemozoin crystals distinguishes diverse Plasmodium species. Mol Biochem Parasitol. 2003;130:91–9.CrossRef Noland GS, Briones N, Sullivan DJ. The shape and size of hemozoin crystals distinguishes diverse Plasmodium species. Mol Biochem Parasitol. 2003;130:91–9.CrossRef
Metadaten
Titel
Performance of a sensitive haemozoin‐based malaria diagnostic test validated for vivax malaria diagnosis in Brazilian Amazon
verfasst von
Gisely Cardoso de Melo
Rebeca Linhares Abreu Netto
Victor Irungu Mwangi
Yanka Evellyn Alves Rodrigues Salazar
Vanderson de Souza Sampaio
Wuelton Marcelo Monteiro
Fernando Fonseca de Almeida e Val
Anne Rocheleau
Priyaleela Thota
Marcus Vinícius Guimarães Lacerda
Publikationsdatum
01.12.2021
Verlag
BioMed Central
Erschienen in
Malaria Journal / Ausgabe 1/2021
Elektronische ISSN: 1475-2875
DOI
https://doi.org/10.1186/s12936-021-03688-0

Weitere Artikel der Ausgabe 1/2021

Malaria Journal 1/2021 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Reizdarmsyndrom: Diäten wirksamer als Medikamente

29.04.2024 Reizdarmsyndrom Nachrichten

Bei Reizdarmsyndrom scheinen Diäten, wie etwa die FODMAP-arme oder die kohlenhydratreduzierte Ernährung, effektiver als eine medikamentöse Therapie zu sein. Das hat eine Studie aus Schweden ergeben, die die drei Therapieoptionen im direkten Vergleich analysierte.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.