01.03.2015 | Scientific Article

Quality-controlled dose reduction of full-leg radiography in patients with knee malalignment

verfasst von: Jost Karsten Kloth, Regina Neumann, Eva von Stillfried, Wolfram Stiller, Hans-Ulrich Kauczor, Volker Ewerbeck, Marc-André Weber

Erschienen in: Skeletal Radiology | Ausgabe 3/2015

Einloggen, um Zugang zu erhalten

Abstract

Objective

Digital plain radiographs of the full leg are frequently performed examinations of children and young adults. Thus, the objective of this work was to reduce the radiation exposure dependent on specific indications, and to determine objective quality-control criteria to ensure accurate assessment.

Materials and methods

Institutional review board approval and informed consent of all participants were obtained. In this prospective, randomized controlled, blinded, two-armed single-center study, 288 evaluable patients underwent plain radiography of the full leg with standard and reduced doses. The evaluation of the plain radiographs was conducted using the following criteria: mechanical axis, leg length, and maturation of the epiphyseal plate. Two blinded radiologists evaluated these criteria using a score ranging from 1 (definitely assessable) to 4 (not assessable). If a single criterion had been evaluated with a score of 3 or more points or all criteria with 2 points, the radiograph was scored as “not assessable”. The study was designed as a non-inferiority trial.

Results

Eleven (3.8 %) examined X-rays were scored as not assessable. The rate of non-assessable radiographs with 33 % reduced dose was significantly not inferior to the rate of non-assessable radiographs with standard dose. The evaluation of the quality criteria was dose independent.

Conclusions

Full-leg plain radiography in patients with knee malalignment can be performed at 33 % reduced dose without loss of relevant diagnostic information.
Literatur
1.
Zurück zum Zitat Sabharwal S, Zhao C, McKeon JJ, McClemens E, Edgar M, Behrens F. Computed radiographic measurement of limb-length discrepancy. Full-length standing anteroposterior radiograph compared with scanogram. J Bone Joint Surg Am Vol. 2006;88(10):2243–51.CrossRef Sabharwal S, Zhao C, McKeon JJ, McClemens E, Edgar M, Behrens F. Computed radiographic measurement of limb-length discrepancy. Full-length standing anteroposterior radiograph compared with scanogram. J Bone Joint Surg Am Vol. 2006;88(10):2243–51.CrossRef
2.
Zurück zum Zitat Odenbring S, Berggren AM, Peil L. Roentgenographic assessment of the hip-knee-ankle axis in medial gonarthrosis. A study of reproducibility. Clin Orthop Relat Res. 1993;289:195–6.PubMed Odenbring S, Berggren AM, Peil L. Roentgenographic assessment of the hip-knee-ankle axis in medial gonarthrosis. A study of reproducibility. Clin Orthop Relat Res. 1993;289:195–6.PubMed
3.
Zurück zum Zitat Hinman RS, May RL, Crossley KM. Is there an alternative to the full-leg radiograph for determining knee joint alignment in osteoarthritis? Arthritis Rheum. 2006;55(2):306–13.PubMedCrossRef Hinman RS, May RL, Crossley KM. Is there an alternative to the full-leg radiograph for determining knee joint alignment in osteoarthritis? Arthritis Rheum. 2006;55(2):306–13.PubMedCrossRef
4.
Zurück zum Zitat Hankemeier S, Gosling T, Richter M, Hufner T, Hochhausen C, Krettek C. Computer-assisted analysis of lower limb geometry: higher intraobserver reliability compared to conventional method. Comput Aided Surg: Off J Int Soc Comput Aided Surg. 2006;11(2):81–6.CrossRef Hankemeier S, Gosling T, Richter M, Hufner T, Hochhausen C, Krettek C. Computer-assisted analysis of lower limb geometry: higher intraobserver reliability compared to conventional method. Comput Aided Surg: Off J Int Soc Comput Aided Surg. 2006;11(2):81–6.CrossRef
5.
Zurück zum Zitat Chao EY, Sim FH. Computer-aided preoperative planning in knee osteotomy. Lowa Orthop J. 1995;15:4–18. Chao EY, Sim FH. Computer-aided preoperative planning in knee osteotomy. Lowa Orthop J. 1995;15:4–18.
6.
Zurück zum Zitat Gorman TM, Vanderwerff R, Pond M, MacWilliams B, Santora SD. Mechanical axis following staple epiphysiodesis for limb-length inequality. J Bone Joint Surg Am Vol. 2009;91(10):2430–9.CrossRef Gorman TM, Vanderwerff R, Pond M, MacWilliams B, Santora SD. Mechanical axis following staple epiphysiodesis for limb-length inequality. J Bone Joint Surg Am Vol. 2009;91(10):2430–9.CrossRef
7.
Zurück zum Zitat Keppler P, Suger G, Kinzl L, Strecker W. [Osteotomies in malalignments of the lower extremities]. Der Chirurg; Zeitschrift fur alle Gebiete der operativen Medizen. 2002; 73(10):982–989 Keppler P, Suger G, Kinzl L, Strecker W. [Osteotomies in malalignments of the lower extremities]. Der Chirurg; Zeitschrift fur alle Gebiete der operativen Medizen. 2002; 73(10):982–989
8.
Zurück zum Zitat Anderson M, Messner MB, Green WT. Distribution of lengths of the normal femur and tibia in children from one to eighteen years of age. J Bone Joint Surg Am Vol. 1964;46:1197–202. Anderson M, Messner MB, Green WT. Distribution of lengths of the normal femur and tibia in children from one to eighteen years of age. J Bone Joint Surg Am Vol. 1964;46:1197–202.
9.
Zurück zum Zitat Hamers S, Freyschmidt J, Neitzel U. Digital radiography with a large-scale electronic flat-panel detector vs. screen-film radiography: observer preference in clinical skeletal diagnostics. Eur Radiol. 2001;11(9):1753–9.PubMedCrossRef Hamers S, Freyschmidt J, Neitzel U. Digital radiography with a large-scale electronic flat-panel detector vs. screen-film radiography: observer preference in clinical skeletal diagnostics. Eur Radiol. 2001;11(9):1753–9.PubMedCrossRef
10.
Zurück zum Zitat Bertolini M, Nitrosi A, Rivetti S, Lanconelli N, Pattacini P, Ginocchi V, et al. A comparison of digital radiography systems in terms of effective detective quantum efficiency. Med Phys. 2012;39(5):2617–27.PubMedCrossRef Bertolini M, Nitrosi A, Rivetti S, Lanconelli N, Pattacini P, Ginocchi V, et al. A comparison of digital radiography systems in terms of effective detective quantum efficiency. Med Phys. 2012;39(5):2617–27.PubMedCrossRef
11.
Zurück zum Zitat Samei E, Ranger NT, MacKenzie A, Honey ID, Dobbins 3rd JT, Ravin CE. Detector or system? Extending the concept of detective quantum efficiency to characterize the performance of digital radiographic imaging systems. Radiology. 2008;249(3):926–37.PubMedCentralPubMedCrossRef Samei E, Ranger NT, MacKenzie A, Honey ID, Dobbins 3rd JT, Ravin CE. Detector or system? Extending the concept of detective quantum efficiency to characterize the performance of digital radiographic imaging systems. Radiology. 2008;249(3):926–37.PubMedCentralPubMedCrossRef
12.
Zurück zum Zitat Andriole KP, Ruckdeschel TG, Flynn MJ, Hangiandreou NJ, Jones AK, Krupinski E, et al. ACR-AAPM-SIIM practice guideline for digital radiography. J Digit Imaging. 2013;26(1):26–37.PubMedCentralPubMedCrossRef Andriole KP, Ruckdeschel TG, Flynn MJ, Hangiandreou NJ, Jones AK, Krupinski E, et al. ACR-AAPM-SIIM practice guideline for digital radiography. J Digit Imaging. 2013;26(1):26–37.PubMedCentralPubMedCrossRef
13.
Zurück zum Zitat Veit R, Guggenberger R, Nosske D, Brix G. Diagnostic reference levels for X-ray examinations: update 2010. Radiologe. 2010;50(10):907–12.PubMedCrossRef Veit R, Guggenberger R, Nosske D, Brix G. Diagnostic reference levels for X-ray examinations: update 2010. Radiologe. 2010;50(10):907–12.PubMedCrossRef
14.
Zurück zum Zitat Association gm. guidelines for quality management in radiology. 2007. Association gm. guidelines for quality management in radiology. 2007.
15.
Zurück zum Zitat Sharma L, Song J, Felson DT, Cahue S, Shamiyeh E, Dunlop DD. The role of knee alignment in disease progression and functional decline in knee osteoarthritis. JAMA: J Am Med Assoc. 2001;286(2):188–95.CrossRef Sharma L, Song J, Felson DT, Cahue S, Shamiyeh E, Dunlop DD. The role of knee alignment in disease progression and functional decline in knee osteoarthritis. JAMA: J Am Med Assoc. 2001;286(2):188–95.CrossRef
16.
Zurück zum Zitat Crema MD, Roemer FW, Felson DT, Englund M, Wang K, Jarraya M, et al. Factors associated with meniscal extrusion in knees with or at risk for osteoarthritis: the Multicenter Osteoarthritis study. Radiology. 2012;264(2):494–503.PubMedCentralPubMedCrossRef Crema MD, Roemer FW, Felson DT, Englund M, Wang K, Jarraya M, et al. Factors associated with meniscal extrusion in knees with or at risk for osteoarthritis: the Multicenter Osteoarthritis study. Radiology. 2012;264(2):494–503.PubMedCentralPubMedCrossRef
18.
Zurück zum Zitat Dunnett CW, Gent M. Significance testing to establish equivalence between treatments, with special reference to data in the form of 2×2 tables. Biometrics. 1977;33(4):593–602.PubMedCrossRef Dunnett CW, Gent M. Significance testing to establish equivalence between treatments, with special reference to data in the form of 2×2 tables. Biometrics. 1977;33(4):593–602.PubMedCrossRef
19.
Zurück zum Zitat Farrington CP, Manning G. Test statistics and sample size formulae for comparative binomial trials with null hypothesis of non-zero risk difference or non-unity relative risk. Stat Med. 1990;9(12):1447–54.PubMedCrossRef Farrington CP, Manning G. Test statistics and sample size formulae for comparative binomial trials with null hypothesis of non-zero risk difference or non-unity relative risk. Stat Med. 1990;9(12):1447–54.PubMedCrossRef
20.
Zurück zum Zitat Kloth JK, Wiedenhoefer B, Stiller W, Burkholder I, Kauczor HU, Ewerbeck V, et al. [Modern digital plain radiography of the whole spine in scoliosis patients--dose reduction and quality criteria]. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2013; 185(1):48–54 Kloth JK, Wiedenhoefer B, Stiller W, Burkholder I, Kauczor HU, Ewerbeck V, et al. [Modern digital plain radiography of the whole spine in scoliosis patients--dose reduction and quality criteria]. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2013; 185(1):48–54
21.
Zurück zum Zitat Kropil P, Fenk R, Fritz LB, Blondin D, Kobbe G, Modder U, et al. Comparison of whole-body 64-slice multidetector computed tomography and conventional radiography in staging of multiple myeloma. Eur Radiol. 2008;18(1):51–8.PubMedCrossRef Kropil P, Fenk R, Fritz LB, Blondin D, Kobbe G, Modder U, et al. Comparison of whole-body 64-slice multidetector computed tomography and conventional radiography in staging of multiple myeloma. Eur Radiol. 2008;18(1):51–8.PubMedCrossRef
22.
Zurück zum Zitat Wang H, Chen B, Chow SC. Sample size determination based on rank tests in clinical trials. J Biopharm Stat. 2003;13(4):735–51.PubMedCrossRef Wang H, Chen B, Chow SC. Sample size determination based on rank tests in clinical trials. J Biopharm Stat. 2003;13(4):735–51.PubMedCrossRef
23.
Zurück zum Zitat Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159–74.PubMedCrossRef Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159–74.PubMedCrossRef
24.
Zurück zum Zitat Gosch D, Gosch K, Kahn T. [Conversion coefficients for estimation of effective dose to patients from dose area product during fluoroscopy X-ray examinations]. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2007; 179(10):1035–1042 Gosch D, Gosch K, Kahn T. [Conversion coefficients for estimation of effective dose to patients from dose area product during fluoroscopy X-ray examinations]. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2007; 179(10):1035–1042
25.
Zurück zum Zitat Brink JA, Amis Jr ES. Image wisely: a campaign to increase awareness about adult radiation protection. Radiology. 2010;257(3):601–2.PubMedCrossRef Brink JA, Amis Jr ES. Image wisely: a campaign to increase awareness about adult radiation protection. Radiology. 2010;257(3):601–2.PubMedCrossRef
26.
Zurück zum Zitat Blount WP, Clarke GR. Control of bone growth by epiphyseal stapling; a preliminary report. J Bone Joint Surg Am Vol. 1949;31A(3):464–78. Blount WP, Clarke GR. Control of bone growth by epiphyseal stapling; a preliminary report. J Bone Joint Surg Am Vol. 1949;31A(3):464–78.
27.
Zurück zum Zitat Grieser T, Baldauf AQ, Ludwig K. Radiation dose reduction in scoliosis patients: low-dose full-spine radiography with digital flat panel detector and image stitching system. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2011; 183(7):645–649 Grieser T, Baldauf AQ, Ludwig K. Radiation dose reduction in scoliosis patients: low-dose full-spine radiography with digital flat panel detector and image stitching system. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2011; 183(7):645–649
28.
Zurück zum Zitat Dietrich TJ, Pfirrmann CW, Schwab A, Pankalla K, Buck FM. Comparison of radiation dose, workflow, patient comfort and financial break-even of standard digital radiography and a novel biplanar low-dose X-ray system for upright full-length lower limb and whole spine radiography. Skeletal Radiol. 2013;42(7):959–67.PubMedCrossRef Dietrich TJ, Pfirrmann CW, Schwab A, Pankalla K, Buck FM. Comparison of radiation dose, workflow, patient comfort and financial break-even of standard digital radiography and a novel biplanar low-dose X-ray system for upright full-length lower limb and whole spine radiography. Skeletal Radiol. 2013;42(7):959–67.PubMedCrossRef
29.
Zurück zum Zitat Escott BG, Ravi B, Weathermon AC, Acharya J, Gordon CL, Babyn PS, et al. EOS Low-dose radiography: a reliable and accurate upright assessment of lower-limb lengths. J Bone Joint Surg Am Vol. 2013;95(23):e1831–7.CrossRef Escott BG, Ravi B, Weathermon AC, Acharya J, Gordon CL, Babyn PS, et al. EOS Low-dose radiography: a reliable and accurate upright assessment of lower-limb lengths. J Bone Joint Surg Am Vol. 2013;95(23):e1831–7.CrossRef
Metadaten
Titel
Quality-controlled dose reduction of full-leg radiography in patients with knee malalignment
verfasst von
Jost Karsten Kloth
Regina Neumann
Eva von Stillfried
Wolfram Stiller
Hans-Ulrich Kauczor
Volker Ewerbeck
Marc-André Weber
Publikationsdatum
01.03.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Skeletal Radiology / Ausgabe 3/2015
Print ISSN: 0364-2348
Elektronische ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-014-2004-5

Weitere Artikel der Ausgabe 3/2015

Skeletal Radiology 3/2015 Zur Ausgabe

Test Yourself: Answer

Lateral leg pain

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.