Purpose

The objective of this study was to develop, optimize, and validate a method for the simultaneous quantification of U-47700, AH-7921, U-49900, U-50488, MT-45, W-18, and W-15 in whole blood using liquid chromatography–tandem mass spectrometry (LC–MS/MS).

Methods

Blood samples (500 µL) were fortified with mixed calibrator or quality control (25 µL) and internal standard (10 µL) solutions. Analytes were isolated via a solid-phase extraction procedure. Analysis was performed using an Agilent 1290 Infinity II liquid chromatograph coupled to an Agilent 6470 triple quadrupole mass spectrometer. The method was validated in accordance with Scientific Working Group for Forensic Toxicology (SWGTOX) standard practices for method validation in forensic toxicology and applied to the analysis of postmortem blood specimens.

Results

Lower limits of quantification were 0.25–1 ng/mL and the upper limits of quantification were 100 ng/mL. The coefficients of determination (R2) for the calibration curves were > 0.99. Analytical bias, within-run imprecision, and between-run imprecision were within ± 15%, ≤ 16%, and ≤ 17%, respectively. All analytes were found to be stable at room temperature for 24 h, refrigerated (4 °C) for 72 h, and in the autosampler (4 °C) for 72 h. Authentic blood samples (n = 30) were analyzed using the validated method. Mean (range) U-47700 concentrations were 214 (3.2–1448) ng/mL in 15 cases.

Conclusions

A quantification method for seven synthetic opioids (U-47700, AH-7921, U-49900, U-50488, MT-45, W-18, and W-15) in whole blood was developed, optimized, and validated in accordance with SWGTOX standard practices for method validation in forensic toxicology. This sensitive method was successfully applied to postmortem casework.
Literatur
5.
Zurück zum Zitat Dosaka-Akita K, Tortella FC, Holaday JW, Long JB (1993) The kappa opioid agonist U-50,488 H antagonizes respiratory effects of mu opioid receptor agonists in conscious rats. J Pharmacol Exp Ther 264:631–637PubMed Dosaka-Akita K, Tortella FC, Holaday JW, Long JB (1993) The kappa opioid agonist U-50,488 H antagonizes respiratory effects of mu opioid receptor agonists in conscious rats. J Pharmacol Exp Ther 264:631–637PubMed
11.
22.
Zurück zum Zitat Morrow JB, Ropero-Miller JD, Catlin ML, Winokur AD, Cadwallader AB, Staymates JL, Williams SR, McGrath JG, Logan BK, McCormick MM, Nolte KB, Gilson TP, Menendez MJ, Goldberger BA (2018) The opioid epidemic: moving toward an integrated, holistic analytical response. J Anal Toxicol. https://doi.org/10.1093/jat/bky049 Morrow JB, Ropero-Miller JD, Catlin ML, Winokur AD, Cadwallader AB, Staymates JL, Williams SR, McGrath JG, Logan BK, McCormick MM, Nolte KB, Gilson TP, Menendez MJ, Goldberger BA (2018) The opioid epidemic: moving toward an integrated, holistic analytical response. J Anal Toxicol. https://​doi.​org/​10.​1093/​jat/​bky049
28.
Zurück zum Zitat Zaney ME, Shoff EN, Hime GW, Boland DM (2017) Detection of U-47700 in blood and drug paraphernalia from postmortem cases. Society of Forensic Toxicologists Annual Meeting, Boca Raton Zaney ME, Shoff EN, Hime GW, Boland DM (2017) Detection of U-47700 in blood and drug paraphernalia from postmortem cases. Society of Forensic Toxicologists Annual Meeting, Boca Raton
29.
Zurück zum Zitat Hartman RL, Salamone ME, Nemeth DJ, Gallagher TM, Fellows JS, Kadlec LM, Beno JM (2017) Furanyl fentanyl, butyryl fentanyl, and U-47700 in blood and urine of drivers arrested for driving under the influence of drugs (DUID). Society of Forensic Toxicologists Annual Meeting, Boca Raton Hartman RL, Salamone ME, Nemeth DJ, Gallagher TM, Fellows JS, Kadlec LM, Beno JM (2017) Furanyl fentanyl, butyryl fentanyl, and U-47700 in blood and urine of drivers arrested for driving under the influence of drugs (DUID). Society of Forensic Toxicologists Annual Meeting, Boca Raton
Metadaten
Titel
Quantification of seven novel synthetic opioids in blood using LC–MS/MS
verfasst von
Janna Lowry
Michael T. Truver
Madeleine J. Swortwood
Publikationsdatum
22.11.2018
Verlag
Springer Japan
Erschienen in
Forensic Toxicology / Ausgabe 1/2019
Print ISSN: 1860-8965
Elektronische ISSN: 1860-8973
DOI
https://doi.org/10.1007/s11419-018-0458-1

Weitere Artikel der Ausgabe 1/2019

Forensic Toxicology 1/2019 Zur Ausgabe

Neu im Fachgebiet Rechtsmedizin

Molekularpathologische Untersuchungen im Wandel der Zeit

Open Access Biomarker Leitthema

Um auch an kleinen Gewebeproben zuverlässige und reproduzierbare Ergebnisse zu gewährleisten ist eine strenge Qualitätskontrolle in jedem Schritt des Arbeitsablaufs erforderlich. Eine nicht ordnungsgemäße Prüfung oder Behandlung des …

Vergleichende Pathologie in der onkologischen Forschung

Pathologie Leitthema

Die vergleichende experimentelle Pathologie („comparative experimental pathology“) ist ein Fachbereich an der Schnittstelle von Human- und Veterinärmedizin. Sie widmet sich der vergleichenden Erforschung von Gemeinsamkeiten und Unterschieden von …

Gastrointestinale Stromatumoren

Open Access GIST CME-Artikel

Gastrointestinale Stromatumoren (GIST) stellen seit über 20 Jahren ein Paradigma für die zielgerichtete Therapie mit Tyrosinkinaseinhibitoren dar. Eine elementare Voraussetzung für eine mögliche neoadjuvante oder adjuvante Behandlung bei …

Personalisierte Medizin in der Onkologie

Aufgrund des erheblichen technologischen Fortschritts in der molekularen und genetischen Diagnostik sowie zunehmender Erkenntnisse über die molekulare Pathogenese von Krankheiten hat in den letzten zwei Jahrzehnten ein grundlegender …