15.04.2021 | Chest

Radiation dose reduction with the wide-volume scan mode for interstitial lung diseases

verfasst von: Alain Abdo, Elige Karam, Théophraste Henry, Sébastien Leygnac, Karim Haioun, Antoine Khalil, Marie-Pierre Debray

Erschienen in: European Radiology | Ausgabe 10/2021

Einloggen, um Zugang zu erhalten

Abstract

Objectives

The wide-volume mode, available on wide-area detector row CTs, has the advantage of reducing exposure time and radiation dose. It is infrequently used for lung diseases. The purpose of this study is to compare image quality and radiation dose of wide-volume chest CT to those of standard helical CT in the setting of interstitial lung diseases.

Methods

Retrospective monocentric study including 50 consecutive patients referred for follow-up or screening of interstitial lung diseases, requiring prone scan, acquired with the wide-volume mode, in addition to the routine supine scan, acquired with the helical mode. The optimal collimation in wide-volume mode (320 × 0.5mm or 240 × 0.5mm) was chosen according to the length of the thorax. Wide-volume acquisitions were compared to helical acquisitions for radiation dose (CTDIvol, DLP) and image quality, including analysis of normal structures, lesions, overall image quality, and artifacts (Wilcoxon signed-rank test).

Results

Median CTDIvol and DLP with wide volumes (3.1 mGy and 94.6 mGy·cm) were significantly reduced (p < 0.0001) as compared to helical mode (3.7mGy and 122.1 mGy·cm), leading to a median 21% and 32% relative reduction of CTDIvol and DLP, respectively. Image noise and quality were not significantly different between the two modes. Misalignment artifact at the junction of two volumes was occasionally seen in the wide-volume scans and, when present, did not impair the diagnostic quality in the majority of cases.

Conclusions

Wide-volume mode allows 32% radiation dose reduction compared to the standard helical mode and could be used routinely for diagnosis and follow-up of interstitial lung diseases.

Key Points

• Retrospective monocentric study showed that wide-volume scan mode reduces radiation dose by 32% in comparison to helical mode for chest CT in the setting of interstitial lung diseases.
• Mild misalignment may be observed at the junction between volumes with the wide-volume mode, without decrease of image quality in the majority of cases and without impairing diagnostic quality.
• Wide-volume mode could be used routinely for the diagnosis and follow-up of interstitial lung diseases.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Hodnett PA, Naidich DP (2013) Fibrosing interstitial lung disease: a practical high-resolution computed tomography-based approach to diagnosis and management and a review of the literature. Am J Respir Crit Care Med 188(2):141–149CrossRef Hodnett PA, Naidich DP (2013) Fibrosing interstitial lung disease: a practical high-resolution computed tomography-based approach to diagnosis and management and a review of the literature. Am J Respir Crit Care Med 188(2):141–149CrossRef
2.
Zurück zum Zitat Elicker BM, Kallianos KG, Henry TS (2017) The role of high-resolution computed tomography in the follow-up of diffuse lung disease. Eur Respir Rev 26(144) Elicker BM, Kallianos KG, Henry TS (2017) The role of high-resolution computed tomography in the follow-up of diffuse lung disease. Eur Respir Rev 26(144)
3.
Zurück zum Zitat Ravenel JG, Scalzetti EM, Huda W, Garrisi W (2001) Radiation exposure and image quality in chest CT examinations. AJR Am J Roentgenol 177(2):279–284CrossRef Ravenel JG, Scalzetti EM, Huda W, Garrisi W (2001) Radiation exposure and image quality in chest CT examinations. AJR Am J Roentgenol 177(2):279–284CrossRef
4.
Zurück zum Zitat McCollough CH, Primak AN, Braun N, Kofler J, Yu L, Christner J (2009) Strategies for reducing radiation dose in CT. Radiol Clin North Am 47(1):27–40CrossRef McCollough CH, Primak AN, Braun N, Kofler J, Yu L, Christner J (2009) Strategies for reducing radiation dose in CT. Radiol Clin North Am 47(1):27–40CrossRef
5.
Zurück zum Zitat Kubo T (2019) Vendor free basics of radiation dose reduction techniques for CT. Eur J Radiol 110:14–21CrossRef Kubo T (2019) Vendor free basics of radiation dose reduction techniques for CT. Eur J Radiol 110:14–21CrossRef
6.
Zurück zum Zitat Padole A, Khawaja RDA, Kalra MK, Singh S (2015) CT radiation dose and iterative reconstruction techniques. AJR Am J Roentgenol 204(4):W384–W392CrossRef Padole A, Khawaja RDA, Kalra MK, Singh S (2015) CT radiation dose and iterative reconstruction techniques. AJR Am J Roentgenol 204(4):W384–W392CrossRef
7.
Zurück zum Zitat Singh S, Gilman MD, Shepard JO (2011) Adaptive statistical iterative reconstruction technique for radiation dose reduction in chest CT: a pilot study. Radiology. 259(2):565–573CrossRef Singh S, Gilman MD, Shepard JO (2011) Adaptive statistical iterative reconstruction technique for radiation dose reduction in chest CT: a pilot study. Radiology. 259(2):565–573CrossRef
8.
Zurück zum Zitat Ichikawa Y, Kitagawa K, Nagasawa N, Murashima S, Sakuma H (2013) CT of the chest with model-based, fully iterative reconstruction: comparison with adaptive statistical iterative reconstruction. BMC Med Imaging 13(1):1CrossRef Ichikawa Y, Kitagawa K, Nagasawa N, Murashima S, Sakuma H (2013) CT of the chest with model-based, fully iterative reconstruction: comparison with adaptive statistical iterative reconstruction. BMC Med Imaging 13(1):1CrossRef
9.
Zurück zum Zitat Larke FJ, Kruger RL (2011) Cagnon CH et al Estimated radiation dose associated with low-dose chest CT of average-size participants in the national lung screening trial. AJR Am J Roentgenol 197(5):1165–1169CrossRef Larke FJ, Kruger RL (2011) Cagnon CH et al Estimated radiation dose associated with low-dose chest CT of average-size participants in the national lung screening trial. AJR Am J Roentgenol 197(5):1165–1169CrossRef
10.
Zurück zum Zitat Prendki V, Scheffler M, Huttner B et al (2018) Low-dose computed tomography for the diagnosis of pneumonia in elderly patients: a prospective, interventional cohort study. Eur Respir J 51(5) Prendki V, Scheffler M, Huttner B et al (2018) Low-dose computed tomography for the diagnosis of pneumonia in elderly patients: a prospective, interventional cohort study. Eur Respir J 51(5)
11.
Zurück zum Zitat Rampinelli C, Origgi D, Bellomi M (2012) Low-dose CT: technique, reading methods and image interpretation. Cancer Imaging 12(3):548–556CrossRef Rampinelli C, Origgi D, Bellomi M (2012) Low-dose CT: technique, reading methods and image interpretation. Cancer Imaging 12(3):548–556CrossRef
12.
Zurück zum Zitat Raghu G, Remy-Jardin M, Myers JL et al (2018) Diagnosis of idiopathic pulmonary fibrosis An Official ATS/ERS/JRS/ALAT Clinical practice guideline. Am J Respir Crit Care Med 198(5):e44–e68CrossRef Raghu G, Remy-Jardin M, Myers JL et al (2018) Diagnosis of idiopathic pulmonary fibrosis An Official ATS/ERS/JRS/ALAT Clinical practice guideline. Am J Respir Crit Care Med 198(5):e44–e68CrossRef
13.
Zurück zum Zitat Chen MY, Arai AE (2013) submillisievert median radiation dose for coronary angiography with a second-generation. Radiology. 267(1):76–85CrossRef Chen MY, Arai AE (2013) submillisievert median radiation dose for coronary angiography with a second-generation. Radiology. 267(1):76–85CrossRef
14.
Zurück zum Zitat Ryu YJ, Kim WS, Choi YH et al (2015) Pediatric chest CT: wide-volume and helical scan modes in 320-MDCT. AJR Am J Roentgenol 205(6):1315–1321CrossRef Ryu YJ, Kim WS, Choi YH et al (2015) Pediatric chest CT: wide-volume and helical scan modes in 320-MDCT. AJR Am J Roentgenol 205(6):1315–1321CrossRef
15.
Zurück zum Zitat Gervaise A, Louis M, Batch T et al (2010) Dose reduction at CT of the lumbar spine using a 320- detector row scanner: initial results. J Radiol 91(7–8):779–785CrossRef Gervaise A, Louis M, Batch T et al (2010) Dose reduction at CT of the lumbar spine using a 320- detector row scanner: initial results. J Radiol 91(7–8):779–785CrossRef
16.
Zurück zum Zitat Roy C, Quin R, Labani A, Leyendecker P, Mertz L, Ohana M (2017) Quantitative and qualitative evaluation of the wide volume (WV) versus helical acquisition on a 320-detector row computed tomography for the exploration of the abdominopelvic region in adults. Radiol Diagn Imaging 1(3):1–7CrossRef Roy C, Quin R, Labani A, Leyendecker P, Mertz L, Ohana M (2017) Quantitative and qualitative evaluation of the wide volume (WV) versus helical acquisition on a 320-detector row computed tomography for the exploration of the abdominopelvic region in adults. Radiol Diagn Imaging 1(3):1–7CrossRef
17.
Zurück zum Zitat Jeon SK, Choi YH, Cheon JE et al (2018) Unenhanced 320-row multidetector computed tomography of the brain in children: comparison of image quality and radiation dose among wide-volume, one-shot volume, and helical scan modes. Pediatr Radiol 48(4):594–601CrossRef Jeon SK, Choi YH, Cheon JE et al (2018) Unenhanced 320-row multidetector computed tomography of the brain in children: comparison of image quality and radiation dose among wide-volume, one-shot volume, and helical scan modes. Pediatr Radiol 48(4):594–601CrossRef
18.
Zurück zum Zitat Honda O, Takenaka D, Matsuki M et al (2012) Image quality of 320-detector row wide-volume computed tomography with diffuse lung diseases: comparison with 64-detector row helical CT. J Comput Assist Tomogr 36(5):505–511CrossRef Honda O, Takenaka D, Matsuki M et al (2012) Image quality of 320-detector row wide-volume computed tomography with diffuse lung diseases: comparison with 64-detector row helical CT. J Comput Assist Tomogr 36(5):505–511CrossRef
19.
Zurück zum Zitat Meyer E, Labani A, Schaeffer M et al (2019) Wide-volume versus helical acquisition in unenhanced chest CT: prospective intra-patient comparison of diagnostic accuracy and radiation dose in an ultra-low-dose setting. Eur Radiol 29(12):6858–6866CrossRef Meyer E, Labani A, Schaeffer M et al (2019) Wide-volume versus helical acquisition in unenhanced chest CT: prospective intra-patient comparison of diagnostic accuracy and radiation dose in an ultra-low-dose setting. Eur Radiol 29(12):6858–6866CrossRef
20.
Zurück zum Zitat Koo TK, Li MY (2016) A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 15(2):155–163CrossRef Koo TK, Li MY (2016) A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 15(2):155–163CrossRef
21.
Zurück zum Zitat Perisinakis K, Papadakis AE, Damilakis J (2009) The effect of x-ray beam quality and geometry on radiation utilization efficiency in multidetector CT imaging. Med Phys 36(4):1258–1266CrossRef Perisinakis K, Papadakis AE, Damilakis J (2009) The effect of x-ray beam quality and geometry on radiation utilization efficiency in multidetector CT imaging. Med Phys 36(4):1258–1266CrossRef
22.
Zurück zum Zitat Lambert JW, Phillips ED, Villanueva-Meyer JE, Nardo L, Facchetti L, Gould RG (2017) Axial or helical? Considerations for wide collimation CT scanners capable of volumetric imaging in both modes. Med Phys 44(11):5718–5725CrossRef Lambert JW, Phillips ED, Villanueva-Meyer JE, Nardo L, Facchetti L, Gould RG (2017) Axial or helical? Considerations for wide collimation CT scanners capable of volumetric imaging in both modes. Med Phys 44(11):5718–5725CrossRef
23.
Zurück zum Zitat Liao YL, Lai NK, Tyan YS, Chuang KS, Tsai HY (2013) Automatic tube current modulation for volume scan in a 320-detector row CT scanner. Radiat Meas 56:328–332CrossRef Liao YL, Lai NK, Tyan YS, Chuang KS, Tsai HY (2013) Automatic tube current modulation for volume scan in a 320-detector row CT scanner. Radiat Meas 56:328–332CrossRef
Metadaten
Titel
Radiation dose reduction with the wide-volume scan mode for interstitial lung diseases
verfasst von
Alain Abdo
Elige Karam
Théophraste Henry
Sébastien Leygnac
Karim Haioun
Antoine Khalil
Marie-Pierre Debray
Publikationsdatum
15.04.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 10/2021
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-021-07862-5

Weitere Artikel der Ausgabe 10/2021

European Radiology 10/2021 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.