Skip to main content
Erschienen in: European Journal of Trauma and Emergency Surgery 3/2012

01.06.2012 | Original Article

Radiation exposure during 3D fluoroscopy of the knee: an experimental study

verfasst von: T. Tharmviboonsri, K. Riansuwan, A. Nitising, B. Mahaisavariya

Erschienen in: European Journal of Trauma and Emergency Surgery | Ausgabe 3/2012

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To measure the radiation exposure at the knee and surrounding area during the use of 3D fluoroscopy.

Methods

An experimental study was conducted by using a human cadaveric knee as a focus point for the 3D fluoroscope. An isocentric C-arm fluoroscope machine was applied on the lateral side of the knee. The radiation dosage at the focus point and surrounding area was measured. The mean radiation exposure in each location was compared between low- and high-resolution scanning.

Results

The mean radiation sustained at the focus point was 44.0 ± 5.6 μSv and 20.0 ± 1.0 μSv in high- and low-resolution scanning, respectively. Radiation exposure on the opposite side of the C-arm machine was found to be higher than that on the other locations with the same distance from the focus point. In low-resolution scanning, radiation could not be detected beyond 75 cm from the focus point at the proximal, distal and same side of the machine. Radiation could be measured at a distance of up to 1.25 m on the opposite side of the machine. In high-resolution scanning, radiation could be measured at a distance of up to 1 m at the proximal, distal and same side of the C-arm, but up to 1.5 m on the opposite side.

Conclusion

Radiation exposure during 3D fluoroscopy of the knee decreases with increasing distance from the focus point. A higher number of scans in the high-resolution mode causes greater radiation exposure. In isocentric 3D fluoroscopy of the knee, a safe zone is located at least 1.5 m away from the focus point.
Literatur
1.
Zurück zum Zitat Atesok K, Finkelstein J, Khoury A, Peyser A, Weil Y, Liebergall M, et al. The use of intraoperative three-dimensional imaging (ISO-C-3D) in fixation of intraarticular fractures. Injury. 2007;38:1163–9.PubMedCrossRef Atesok K, Finkelstein J, Khoury A, Peyser A, Weil Y, Liebergall M, et al. The use of intraoperative three-dimensional imaging (ISO-C-3D) in fixation of intraarticular fractures. Injury. 2007;38:1163–9.PubMedCrossRef
2.
Zurück zum Zitat Kendoff D, Gardner MJ, Citak M, Kfuri M Jr, Thumes B, Krettek C, et al. Value of 3D fluoroscopic imaging of acetabular fractures comparison to 2D fluoroscopy and CT imaging. Arch Orthop Trauma Surg. 2008;128:599–605.PubMedCrossRef Kendoff D, Gardner MJ, Citak M, Kfuri M Jr, Thumes B, Krettek C, et al. Value of 3D fluoroscopic imaging of acetabular fractures comparison to 2D fluoroscopy and CT imaging. Arch Orthop Trauma Surg. 2008;128:599–605.PubMedCrossRef
3.
Zurück zum Zitat Rübberdt A, Feil R, Stengel D, Spranger N, Mutze S, Wich M, et al. The clinical use of the ISO-C(3D) imaging system in calcaneus fracture surgery. Unfallchirurg. 2006;109:112–8.PubMedCrossRef Rübberdt A, Feil R, Stengel D, Spranger N, Mutze S, Wich M, et al. The clinical use of the ISO-C(3D) imaging system in calcaneus fracture surgery. Unfallchirurg. 2006;109:112–8.PubMedCrossRef
4.
Zurück zum Zitat Citak M, Kendoff D, Kfuri M Jr, Pearle A, Krettek C, Hüfner T. Accuracy analysis of Iso-C3D versus fluoroscopy-based navigated retrograde drilling of osteochondral lesions: a pilot study. J Bone Joint Surg Br. 2007;89(3):323–6.PubMedCrossRef Citak M, Kendoff D, Kfuri M Jr, Pearle A, Krettek C, Hüfner T. Accuracy analysis of Iso-C3D versus fluoroscopy-based navigated retrograde drilling of osteochondral lesions: a pilot study. J Bone Joint Surg Br. 2007;89(3):323–6.PubMedCrossRef
5.
Zurück zum Zitat Richter M, Geerling J, Zech S, Goesling T, Krettek C. Intraoperative three-dimensional imaging with a motorized mobile C-arm (SIREMOBIL ISO-C-3D) in foot and ankle trauma care: a preliminary report. J Orthop Trauma. 2005;19:259–66.PubMedCrossRef Richter M, Geerling J, Zech S, Goesling T, Krettek C. Intraoperative three-dimensional imaging with a motorized mobile C-arm (SIREMOBIL ISO-C-3D) in foot and ankle trauma care: a preliminary report. J Orthop Trauma. 2005;19:259–66.PubMedCrossRef
6.
Zurück zum Zitat Holly LT, Foley KT. Percutaneous placement of posterior cervical screws using three-dimensional fluoroscopy. Spine. 2006;31:536–40.PubMedCrossRef Holly LT, Foley KT. Percutaneous placement of posterior cervical screws using three-dimensional fluoroscopy. Spine. 2006;31:536–40.PubMedCrossRef
7.
Zurück zum Zitat Summers LE, Kouri JG, Yang M, Patrick Jacob R. Odontoid screw placement using isocentric 3-dimensional C-arm fluoroscopy. J Spinal Disord Tech. 2008;21:45–8.PubMedCrossRef Summers LE, Kouri JG, Yang M, Patrick Jacob R. Odontoid screw placement using isocentric 3-dimensional C-arm fluoroscopy. J Spinal Disord Tech. 2008;21:45–8.PubMedCrossRef
8.
Zurück zum Zitat Patel VV, Dwyer A, Estes S, Burger E. Intraoperative 3-dimensional reconstructed multiplanar fluoroscopic imaging for immediate evaluation of spinal decompression. J Spinal Disord Tech. 2008;21:209–12.PubMedCrossRef Patel VV, Dwyer A, Estes S, Burger E. Intraoperative 3-dimensional reconstructed multiplanar fluoroscopic imaging for immediate evaluation of spinal decompression. J Spinal Disord Tech. 2008;21:209–12.PubMedCrossRef
9.
Zurück zum Zitat Giachino AA, Cheng M. Irradiation of the surgeon during pinning of femoral fractures. J Bone Joint Surg Br. 1980;62-B(2):227–9.PubMed Giachino AA, Cheng M. Irradiation of the surgeon during pinning of femoral fractures. J Bone Joint Surg Br. 1980;62-B(2):227–9.PubMed
10.
Zurück zum Zitat Riley SA. Radiation exposure from fluoroscopy during orthopedic surgical procedures. Clin Orthop Relat Res. 1989;248:257–60.PubMed Riley SA. Radiation exposure from fluoroscopy during orthopedic surgical procedures. Clin Orthop Relat Res. 1989;248:257–60.PubMed
11.
Zurück zum Zitat Mehlman CT, DiPasquale TG. Radiation exposure to the orthopaedic surgical team during fluoroscopy: “how far away is far enough?”. J Orthop Trauma. 1997;11:392–8.PubMedCrossRef Mehlman CT, DiPasquale TG. Radiation exposure to the orthopaedic surgical team during fluoroscopy: “how far away is far enough?”. J Orthop Trauma. 1997;11:392–8.PubMedCrossRef
12.
Zurück zum Zitat Miller ME, Davis ML, MacClean CR, Davis JG, Smith BL, Humphries JR. Radiation exposure and associated risks to operating-room personnel during use of fluoroscopic guidance for selected orthopaedic surgical procedures. J Bone Joint Surg Am. 1983;65(1):1–4.PubMed Miller ME, Davis ML, MacClean CR, Davis JG, Smith BL, Humphries JR. Radiation exposure and associated risks to operating-room personnel during use of fluoroscopic guidance for selected orthopaedic surgical procedures. J Bone Joint Surg Am. 1983;65(1):1–4.PubMed
13.
Zurück zum Zitat Theocharopoulos N, Perisinakis K, Damilakis J, Papadokostakis G, Hadjipavlou A, Gourtsoyiannis N. Occupational exposure from common fluoroscopic projections used in orthopaedic surgery. J Bone Joint Surg Am. 2003;89-A(9):1698–703. Theocharopoulos N, Perisinakis K, Damilakis J, Papadokostakis G, Hadjipavlou A, Gourtsoyiannis N. Occupational exposure from common fluoroscopic projections used in orthopaedic surgery. J Bone Joint Surg Am. 2003;89-A(9):1698–703.
14.
Zurück zum Zitat International Commission on Radiological Protection (ICRP). The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007;37(2-4):81–100. International Commission on Radiological Protection (ICRP). The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007;37(2-4):81–100.
Metadaten
Titel
Radiation exposure during 3D fluoroscopy of the knee: an experimental study
verfasst von
T. Tharmviboonsri
K. Riansuwan
A. Nitising
B. Mahaisavariya
Publikationsdatum
01.06.2012
Verlag
Springer-Verlag
Erschienen in
European Journal of Trauma and Emergency Surgery / Ausgabe 3/2012
Print ISSN: 1863-9933
Elektronische ISSN: 1863-9941
DOI
https://doi.org/10.1007/s00068-011-0165-7

Weitere Artikel der Ausgabe 3/2012

European Journal of Trauma and Emergency Surgery 3/2012 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders scheint das auf weibliche Kranke zuzutreffen, wie eine Studie zeigt.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.