Abstract

We report the case of a 64-year old Caucasian male patient with a tear of the left ventricular driveline just above the driveline-air tube junction. We describe the repair technique and the necessary set of tools.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1749-8090-9-7) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

SS drafted the manuscript. MT was involved in the drafting of the manuscript. DG was involved in the drafting of the manuscript. GD revised the manuscript. RK has given final approval of the version to be published. GT has given final approval of the version to be published. All authors read and approved the final manuscript.
Abkürzungen
TAH
Total artificial heart
HUD
Humanitarian use device
FDA
Food and drug administration.

Background

Tear of a SynCardia-Total Artificial Heart driveline is a rare but potentially life-threatening event. Once it has been identified it has to be sufficiently repaired, otherwise air leak would result into pressure drop, malpositioning of the diaphragm and consequently rapid decrease of artificial ventricle fill volume.

Case presentation

The SynCardia CardioWest TAH is a pneumatically driven, biventricular system used for the orthotopic replacement of the native heart ventricles in cases of end-stage heart failure and refractory cardiogenic schock [1, 2]. Two wire-reinforced, dacron-coated drivelines connect the artificial ventricles to the pneumatic drivers of the external console. Like all extracorporeal components, these drivelines are prone to wear and tear. We describe the repair technique of a left ventricular driveline tear in a SynCardia Total Artificial Heart patient and present the set of tools used.
A 64-years old male, caucasian patient was admitted for therapy of a refractory cardiogenic schock on the grounds of an ischemic cardiomyopathy. Despite innotropic therapy the patient continued to decline. Initially we implemented veno-arterial extracorporeal membrane oxygenation as a bridge to decision and finally implanted the 70-cc SynCardia-TAH as a bridge to transplantion. The postoperative course was uneventful. The patient was switched to the Freedom driver and was discharged at home. The patient was readmitted at postoperative day 172 due to progressive exertion intolerance caused by an air leak at the left ventricular driveline. Radiology showed mid-severe pulmonary congestion. Upon inspection, we noted a small horizontal tear above the connection of the driveline to the air tubes (Figure 1). After cutting through both cable ties that secure the drivelines with the connectors using a Xcelite 170 M general purpose shear cutter (Apex tool group, Sparks, MD, USA) (Figure 2) we cut off the left driveline slightly above the location of the tear with a high leverage cable cutter (Klein Tools, Lincolnshire, IL, USA) (Figure 3) and quickly switched the patient to a Companion Driver operating in a standard Intensive Care Unit Mode. The whole procedure was completed in approximately 5 minutes. Throughout the procedure the patient showed no clinical signs such as syncopal spell. Two days later the patient recompensated fully. He was switched to a Freedom Driver and discharged at home. Mechanical support of the patient is still ongoing.

Conclusions

Driveline tear is a rare event that potentially harbours the risk of insufficient filling of the artificial ventricles. In our case patient became symptomatic in terms of pulmonary congestion and progressive exertion intolerance. Therefore it is important to identify and repair quickly the defect. Other currently available repair techniques include banding the driveline with a silicon tube [3] or sealing it with cyanoacrylate ashesive. However this is only a provisory repair. Since the U.S. Food and Drug Administration (FDA) has approved a Humanitarian Use Device (HUD) designation for the SynCardia temporary Total Artificial Heart to be used for destination therapy in patients not being eligible for transplantation [4] and since the durability of the driveline could limit successful long-term support [5], the aim should be to definitively repair any driveline defect As in this case, the repair can be easily facilitated by cutting the driveline at the location of the tear and reconnecting to the air tubes. During repair special care should be taken to avoid any low-output events. For this reason repair should be performed by a team of two physicians who are responsible for cutting the driveline and switching the patient to the new Driver and one technician who supervises the Driver settings. Furthermore we would recommend in a first step to switch the patient to a Companion Driver in order to control fill volumes, pumping rate and pressure and finally switch to a Freedom Driver.
Written informed consent was obtained from the patient for publication of this case report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.

Acknowledgements

The authors would like to acknowledge the significant contribution of Markus Leinberger, Managing Director of SynCardia Systems Europe GmbH, for his valuable input in framing the topic.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

SS drafted the manuscript. MT was involved in the drafting of the manuscript. DG was involved in the drafting of the manuscript. GD revised the manuscript. RK has given final approval of the version to be published. GT has given final approval of the version to be published. All authors read and approved the final manuscript.
Anhänge

Authors’ original submitted files for images

Literatur
1.
Zurück zum Zitat Copeland JG, Copeland H, Gustafson M, Mineburg N: Experience with 100 total artificial heart implants. J Thorac Cardiovasc Surg. 2012, 143: 727-734. 10.1016/j.jtcvs.2011.12.002.CrossRefPubMed Copeland JG, Copeland H, Gustafson M, Mineburg N: Experience with 100 total artificial heart implants. J Thorac Cardiovasc Surg. 2012, 143: 727-734. 10.1016/j.jtcvs.2011.12.002.CrossRefPubMed
2.
Zurück zum Zitat Morshuis M, Reiss N, Arusoglu L, Tenderich G, Koerfer R, El-Banayosi A: Implantation of Cardiowest total artificial heart for irreversible acute myocardial infarction. Heart Surg Forum. 2007, 10 (4): 251-256. 10.1532/HSF98.20070706.CrossRef Morshuis M, Reiss N, Arusoglu L, Tenderich G, Koerfer R, El-Banayosi A: Implantation of Cardiowest total artificial heart for irreversible acute myocardial infarction. Heart Surg Forum. 2007, 10 (4): 251-256. 10.1532/HSF98.20070706.CrossRef
3.
Zurück zum Zitat Roussel JC, Sénage T, Baron O, Périgaud C: CardioWest (Jarvic) total artificial heart: a single center experience with 42 patients. Ann Thorac Surg. 2009, 87: 124-130. 10.1016/j.athoracsur.2008.09.048.CrossRefPubMed Roussel JC, Sénage T, Baron O, Périgaud C: CardioWest (Jarvic) total artificial heart: a single center experience with 42 patients. Ann Thorac Surg. 2009, 87: 124-130. 10.1016/j.athoracsur.2008.09.048.CrossRefPubMed
4.
Zurück zum Zitat Slepian MJ, Alemu Y, Soares JS, Smith RG, Einav S, Bluestein D: The SynCardia total artificial heart: in vivo, in vitro and computational modelling studies. J Biomech. 2013, 46: 266-275. 10.1016/j.jbiomech.2012.11.032.CrossRefPubMedPubMedCentral Slepian MJ, Alemu Y, Soares JS, Smith RG, Einav S, Bluestein D: The SynCardia total artificial heart: in vivo, in vitro and computational modelling studies. J Biomech. 2013, 46: 266-275. 10.1016/j.jbiomech.2012.11.032.CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Shah KB, Volman RA, Harton SC, Tang DG, Kasirajan V: Fracture of the total artificial heart pneumatic driveline after transition to the portable driver. J Heart Lung Transplant. 2013, 32 (10): 1041-1043. 10.1016/j.healun.2013.06.014.CrossRefPubMed Shah KB, Volman RA, Harton SC, Tang DG, Kasirajan V: Fracture of the total artificial heart pneumatic driveline after transition to the portable driver. J Heart Lung Transplant. 2013, 32 (10): 1041-1043. 10.1016/j.healun.2013.06.014.CrossRefPubMed
Metadaten
Titel
Repair of left ventricular driveline tear in a SynCardia-total artificial heart patient
verfasst von
Sotirios Spiliopoulos
Magda Tenderich
Dilek Guersoy
Guenes Dogan
Reiner Koerfer
Gero Tenderich
Publikationsdatum
01.12.2014
Verlag
BioMed Central
Erschienen in
Journal of Cardiothoracic Surgery / Ausgabe 1/2014
Elektronische ISSN: 1749-8090
DOI
https://doi.org/10.1186/1749-8090-9-7

Weitere Artikel der Ausgabe 1/2014

Journal of Cardiothoracic Surgery 1/2014 Zur Ausgabe

Wie erfolgreich ist eine Re-Ablation nach Rezidiv?

23.04.2024 Ablationstherapie Nachrichten

Nach der Katheterablation von Vorhofflimmern kommt es bei etwa einem Drittel der Patienten zu Rezidiven, meist binnen eines Jahres. Wie sich spätere Rückfälle auf die Erfolgschancen einer erneuten Ablation auswirken, haben Schweizer Kardiologen erforscht.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Ureterstriktur: Innovative OP-Technik bewährt sich

19.04.2024 EAU 2024 Kongressbericht

Die Ureterstriktur ist eine relativ seltene Komplikation, trotzdem bedarf sie einer differenzierten Versorgung. In komplexen Fällen wird dies durch die roboterassistierte OP-Technik gewährleistet. Erste Resultate ermutigen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.