Skip to main content
Erschienen in: Hernia 4/2019

Open Access 12.04.2019 | How-I-Do-It

Tentacle-shaped mesh for fixation-free repair of umbilical hernias

verfasst von: G. Amato, G. Romano, A. Agrusa, G. L. Canu, E. Gulotta, E. Erdas, P. G. Calò

Erschienen in: Hernia | Ausgabe 4/2019

Abstract

Purpose

Mesh fixation and broad overlap represent an open issue in umbilical hernia repair. A proprietary-designed implant with tentacle straps at its boundary has been developed to ensure a suture-free repair and a broader coverage of the abdominal wall. The study describes the results of umbilical hernia procedures carried out with the tentacle-shaped implant and the related surgical technique.

Methods

A proprietary tentacle-shaped flat mesh having a central body with integrated radiating arms at its edge was used to repair large umbilical hernias in 62 patients. The implant was placed in preperitoneal sublay. The friction of the straps, crossing the abdominal wall thanks to a special needle passer, was intended to assure adequate grip to hold the implant in place assuring a fixation-free procedure and broad overlap of the hernia defect.

Results

In a mean follow-up of 48 months (range 10–62 months), 4 seromas and 2 ischemia of the navel skin occurred. No infections, hematomas, chronic pain, mesh dislocation, or recurrence has been reported.

Conclusions

The tentacle strap system of the prosthesis effectively ensured an easier implant placement avoiding the need for suturing the mesh. The arms of the implant ensured a proper orientation and stabilization of the mesh in association with a broad defect overlap. The specifically developed surgical procedure showed a quick postoperative recovery, a very low complication rate, and no recurrences even in the long term.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Umbilical hernia is a rather frequent pathology, having an incidence of 10–14% compared to all hernias. For decades, it was performed with direct suture technique with recurrences up to 54% [1, 2, 3]. In recent years, the use of synthetic implants has contributed to a drastic reduction in recurrences [4, 5, 6]. Nevertheless, prosthetic umbilical hernia repair is affected by frequent, specific postoperative complications consequent to mesh fixation and insufficient overlap of the mesh over the hernia opening, which can lead to tissue tearing, mesh migration, and recurrences [7]. To eliminate these problems, a prosthetic device has been developed, composed of an oval flat mesh with incorporated tentacle straps at its boundary, that allows a simple, fast "fixation free" positioning of the implant, assuring broad coverage of the umbilical region. This implant is delivered by mean of a proprietary passer that allows simple, quick deployment of the device distant from the hernia edge. The mesh is held in place without fixation by the tentacle straps that are pulled out from the preperitoneal space to the subcutaneous space passing through the rectus muscles and fascia. The present report aims to demonstrate the effectiveness of the use of an adequately sized tentacle-shaped mesh in patients suffering from umbilical hernia defects larger than 2 cm in diameter.

Materials and methods

The institutional Ethics Committee approved this study. A cohort of 62 patients with hernia defect larger than 2 cm (medium and large defects according to the European Hernia Society classification) forms the body of this study. All patients, consecutively operated, underwent umbilical hernia repair with a specially designed implant composed of a central oval core and eight straps arising from the boundary edge at regular distances (Freedom Octomesh VHR XS produced by Insightra Medical Inc., USA). The body of the mesh used for the repair of navel hernias had an oval shape, sized 15 × 12 cm. The straps of the implant, positioned at the edge of the mesh, measure 2 cm in width and 15 cm in length. The straps are intended to be delivered laterally from the defect borders, from the preperitoneal space across the abdominal-wall structures holding the implant in place by friction. The implant is manufactured with lightweight, large-porous polypropylene mesh type 1, with a density of 70 g/m2 and pores > 10 µm. The procedure was performed in 62 patients (24 male and 38 female), with a mean age of 41 years (range 25–63), a median BMI of 29.40 (range 24–36). General anesthesia was administered in all individuals. A protocol-based antibiotic prophylaxis, as well as analgesic medication, were followed for 5 days postop. Forty-eight repaired umbilical hernias were primary protrusions, while 14 were recurrent (among these 3 multi-recurrent). Mean hernia defect size: 3.5 cm (range 2–5 cm). To carry out the procedure, a semilunar skin incision was made below the edge of the umbilical hollow (Fig. 1a, b). Once the hernia sac was dissected from the adhesion with the surrounding tissue, it was returned with its contents to the abdominal cavity (Figs. 2, 3a, b). Next, the peritoneal sheath was separated from the posterior rectus sheath with blunt dissection as far as possible from hernia opening creating a broad space for the placement of the implant (Fig. 4a, b). The tentacle mesh was then prepared to be delivered (Fig. 5). At this stage, thanks to a special needle passer specifically designed for this purpose, each strap could easily be deployed across the abdominal-wall tissue, entering the preperitoneal space distal from the hernia opening (Fig. 6a). In all 62 patients, the tentacle-shaped implant was placed in preperitoneal sublay below the posterior rectus sheath (Fig. 6b). Each implant was inserted to ensure adequate coverage and tension-free repair. After fascia closure, mesh tentacles were pulled high all together thus allowing spontaneous deployment of the mesh body above the peritoneum (Fig. 7a). Due to the variability of hernias repaired, duration of the mesh placement was determined starting from the return of the protrusion into the abdominal cavity and ending with the placement of the last tentacle. After closure of the linea alba, all 8 straps were trimmed leaving a ca 2 cm-long stump in the subcutaneous layer (Fig. 7b). In 42 of the operated patients, the redundant skin of the navel was excised to carry out an omphaloplasty through skin closure with total intradermal introverting stitches (Fig. 8a). This way, on completed wound healing, the navel could achieve an esthetically reasonable aspect (Fig. 8b). Postoperative follow-ups were at 1, 3, 6, and 12 months, and every subsequent year. The follow-up also included physical examination and real-time ultrasound (US) to look for implant and tentacle position. Further details concerning patient demographics, surgical treatment and related results are shown in Table 1.
Table 1
Patients demographics, treatment details, and results
Patient characteristics
Hernia types
Number of patients: 62
Primary umbilical hernia: 48 (16 incarcerated)
Gender: 24 male–38 female
Recurrent umbilical hernia: 14 (3 multi-recurrent– 5 incarcerated)
Mean age 41 years (25–63)
Mean hernia defect size: 3.5 cm ( 2–5 cm)
Mean BMI: 29.40 (24–36)
Total: 62
Procedure outcomes
Postoperative complications:
Mean mesh overlap: 6 cm (5–7 cm)
Wound infection: 0 (0%)
Time needed for mesh placement and strap positioning: mean 6 min (4–8 min)
Seroma: 4 (6%)a
Intraoperative complications: 0
Navel ischemia after omphaloplasty: 2 (3%)a
Hospital stay: 1 day (day surgery)
Recurrence: 0 (0%)
Mean follow-up length: 48 months (10–62 months)
Total complication rate: 15.4%
aResolved within 15 days (conservative therapy)

Results

In the cohort of 62 patients who underwent umbilical hernia repair with the tentacle-shaped implant, broad coverage of the abdominal wall far beyond the hernia opening was achieved. We measured a mean overlap of ca. 6 cm (range 5–7 cm) assessed from the hernia border to the lateral edge of the implant. The mean time needed to properly deliver all tentacles and deploy the mesh was 6 min (range 4–8 min). No suture fixation of the implant was necessary. The procedures were carried out in day surgery and the patients could be discharged on the same day of the operation. In the postoperative period, we observed 4 (6%) seromas and 2 ischemia of the navel skin after reconstruction. These minor complications were resolved within 15 days with conservative therapy. There were no wound infections, hematoma, chronic pain, or nerve-entrapment symptoms (Table 1). US scans during follow-up showed that all tentacles laid in the correct position. The placement of the tentacles would have unveiled an eventual implant dislocation, but no mesh dislocation was documented. During follow-up, no patient was lost. Patients were questioned regarding pain, any limitations to abdominal-wall mobility and overall satisfaction. In particular, no discomfort due to strap placement in the subcutaneous fat was reported. For all 62 patients, the mean follow-up was 48 months (range 10–62 months). No mesh dislocation or recurrence was reported. No significant pain or discomfort was described by the patients, even in the long term.

Discussion

Umbilical hernia is commonly perceived as a disease requiring low attention from the surgical community. Despite this, many patients who undergo surgical repair of umbilical protrusion suffer from adverse events that could be avoided by improving the operative strategy using appropriate prosthetic devices. Among the complications that affect umbilical hernia repair, recurrence represents a controversial subject in literature. Usually, small defects (< 2 cm) are repaired open, with direct suture of the gap, while larger defects can be treated open or laparoscopic with the deployment of an implant in onlay or sublay fashion. Some studies involving large cohorts of patients indicate a recurrence rate of 1–10% after prosthetic umbilical hernia repair [8, 9]. Nevertheless, other reports specify an incidence of recurrences between 11 and 18% [4, 10]. Independent of these incongruences, in prosthetic umbilical repair, several factors are involved in the pathogenesis of recurrence: defect size, BMI, smoking, pregnancy, and reduced mesh overlap. A broad mesh coverage in a small surrounding like the umbilical area represents one of the most important problems that seems to be rarely debated in literature. Actually, the prosthetic coverage of a repaired umbilical defect is seldom larger than a few cm. This seems to be a major issue that, especially in obese patients with large defects, can facilitate the reappraisal of the protrusion. Another significant challenge is represented by mesh fixation at the boundary of the umbilical area, which is often extremely difficult and can lead to postoperative complications like tissue tear, hematoma and/or mesh dislodgment [11]. Furthermore, the position of the mesh used for reinforcing the abdominal wall also plays an important role. In the literature, it is well acknowledged that a deeper mesh placement helps in avoiding recurrences [12, 13]. For this reason, a preperitoneal sublay deployment of an implant seems to be more indicated to prevent recurrences. To overcome all these issues, a proprietary tentacle-shaped implant in different sizes has been developed for the surgical treatment of ventral and incisional hernia. The inherent friction force exerted by the tentacle straps crossing the abdominal-wall tissue has been already tested in an experimental trial on a porcine model to demonstrate the effectivity of the principle of mesh fixation through friction [14]. The tentacle mesh used to repair umbilical hernias in the above-described patient cohort seems to possess suitable characteristics to adequately resolve the problems connected to the surgical treatment of this frequent protrusion. Thanks to a well-established surgical strategy, achieving a broad dissection between posterior rectus sheath and peritoneal sheath does not appear challenging if carried out with careful separation of these elements. Once the preperitoneal planes are dissected, the delivery of the tentacle mesh by means of the proprietary passer is quick and easy; it allows the tentacles to pass from the posterior rectus sheath towards the subcutaneous layer crossing the abdominal wall. Piercing the abdominal wall with the passer, from the subcutaneous layer until the posterior rectus sheath, is quite easy. Actually, despite being carried out blind, the forefinger dome interposed between the posterior rectus sheath and the detached peritoneum consents a safe introduction and guidance of the passer needle through the preperitoneal interstitium as far as the wound opening. This maneuver also allows deployment of the mesh edges distal from the hernia border. Despite the distance from the hernia defect, there is no need for fixating the borders of the implant, the mesh is simply held in place by the friction exerted by the straps passing the abdominal wall structure laterally from the defect. The 3 mm-large passer needle tunneled across the tissue forces the 2 cm large strap to roll along its axis. Therefore, the friction exerted by its rolled contour into the small tunnel allows stable positioning of the arm within the tissues. This principle of physics has already been proven in the literature and is a suitable way to assure a fixation-free but firm positioning of polypropylene strips used for the treatment of female genital prolapse [1519]. This kind of fixation-free implant stabilization is carried out under complete operator visualization, unlike point-to-point sutures carried out in a deep, narrow, and dark space. Delivery of the tentacle straps with the needle passer eliminates point fixation, thus simplifying the hernia repair procedure that can be performed in a shorter time. The limited mean time of 6 min needed for mesh placement and strap positioning seems to further confirm that the surgical repair can be carried out safely and quickly. The stabilizing feature of the tentacle strap system leads us to define its fixation-free effect as “freexation”. The surgical approach with the tentacle-shaped implant also revealed sharply reduced postoperative complications: the use of the tentacle mesh showed a reasonable postoperative complication rate of 9.7%. In the described patient sample, adverse events were represented only by 4 seromas, and 2 navel ischemia. The latter is a typical occurrence when, in large defects, the umbilical skin is tightly adherent to the hernia sack and after dissection remains with limited vascular support even if adequately resected before omphaloplasty. Thanks to the use of the tentacle-shaped mesh, among the patient sample, a mean mesh overlap from the hernia border ranging from 5 to 7 cm was assessed, which in this kind of procedure can be well defined as a broad overlap. It is the wide overlap assured by the tentacle mesh that is presumed to be the reason for the lack of recurrences in the examined patient cohort. In addition, postoperatively, all patients reported low pain that within 1 week sensibly diminished. Postoperative low pain seems to be linked to the non-fixation of the mesh to the myotendineal (and highly sensitive) surrounding of the umbilical region. In postoperative check-ups, easy localization of the straps with ultrasound scan was helpful, in particular in terms of recurrence prediction during patient follow-up which, to date, has not happened.

Conclusions

Umbilical hernia repair with the described tentacle-shaped implant appears to yield improved results compared with conventional pure tissue or conventional prosthetic repair procedures. Aside from a quick, safe placement technique and the reduced rate of adverse events, usually reported in meshes that need fixation, a more significant benefit is determined by the very broad coverage assured by the implant shape and the related surgical technique. We deem it was the main reason why no recurrences were reported in the patients of the cohort. Overall, the advantages of using this proprietary designed implant are clearly demonstrated and result in fixation-free mesh placement, simplified procedure, broader coverage of the abdominal wall, shortening of operative time, and reduced intra- and early postoperative complications.

Acknowledgement

This research did not receive any specific grant.

Compliance with ethical standards

Conflict of interest

The corresponding author is the developer of the implant and related surgical technique. All other authors have no conflict of interest.

Ethical approval

Approval from institutional ethics committee was obtained for this study.

Human and animal rights

The study including human participants has been performed in accordance with the ethical standards of the Declaration of Helsinki and its later amendments.
Informed consent was obtained from all patients prior to all surgical procedures.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
1.
Zurück zum Zitat Musca AA (2014) Umbilical and ventral herniorrhaphy: a review of 1000 cases. Part 1: characteristics which predict surgical results and predisposing factors. Int Surg 48:169–179 Musca AA (2014) Umbilical and ventral herniorrhaphy: a review of 1000 cases. Part 1: characteristics which predict surgical results and predisposing factors. Int Surg 48:169–179
2.
Zurück zum Zitat Rosenthal RA (1994) Small-bowel disorders and abdominal wall hernia in the elderly patients. Surg Clin North Am 74:261–291CrossRefPubMed Rosenthal RA (1994) Small-bowel disorders and abdominal wall hernia in the elderly patients. Surg Clin North Am 74:261–291CrossRefPubMed
3.
Zurück zum Zitat Schumacher OP, Peiper C, Lorken M, Schumpelick V (2003) Langzeitergebnisse der Nabelhernienreparation nach Spitzy (Nahtverschluss). Chirurg 74:50–54 Schumacher OP, Peiper C, Lorken M, Schumpelick V (2003) Langzeitergebnisse der Nabelhernienreparation nach Spitzy (Nahtverschluss). Chirurg 74:50–54
4.
Zurück zum Zitat Arroyo A, Garcia P, Perez F, Andreu J, Candela F, Calpena R (2001) Randomised clinical trial comparing suture and mesh repair of umbilical hernia in adult. Br J Surg 88:1321–1323CrossRefPubMed Arroyo A, Garcia P, Perez F, Andreu J, Candela F, Calpena R (2001) Randomised clinical trial comparing suture and mesh repair of umbilical hernia in adult. Br J Surg 88:1321–1323CrossRefPubMed
5.
6.
Zurück zum Zitat Kurzer M, Belsham PA, Kark A (2004) Tension-free mesh repair of umbilical hernia as a day case using local anaesthesia. Hernia 8:104–107CrossRefPubMed Kurzer M, Belsham PA, Kark A (2004) Tension-free mesh repair of umbilical hernia as a day case using local anaesthesia. Hernia 8:104–107CrossRefPubMed
7.
Zurück zum Zitat Bisgaard T, Kehlet H, Bay-Nielsen M, Iversen M, Rosenberg GJ, Jørgensen LN (2011) A nationwide study on readmission, morbidity, and mortality after umbilical and epigastric hernia repair. Hernia 15(5):541–546 Bisgaard T, Kehlet H, Bay-Nielsen M, Iversen M, Rosenberg GJ, Jørgensen LN (2011) A nationwide study on readmission, morbidity, and mortality after umbilical and epigastric hernia repair. Hernia 15(5):541–546
8.
Zurück zum Zitat Berrevoet F, D’Hont F, Rogiers X, Troisi R, de Hemptinne B (2011) Open intraperitoneal versus retromuscular mesh repair for umbilical hernias less than 3 cm diameter. Am J Surg 201:85–90CrossRefPubMed Berrevoet F, D’Hont F, Rogiers X, Troisi R, de Hemptinne B (2011) Open intraperitoneal versus retromuscular mesh repair for umbilical hernias less than 3 cm diameter. Am J Surg 201:85–90CrossRefPubMed
9.
Zurück zum Zitat Christoffersen MW, Helgstrand F, Rosenberg J, Kehlet H, Strandfelt P, Bisgaard T (2015) Long-term recurrence and chronic pain after repair for small umbilical or epigastric hernias: a regional cohort study. Am J Surg 209:725–732CrossRefPubMed Christoffersen MW, Helgstrand F, Rosenberg J, Kehlet H, Strandfelt P, Bisgaard T (2015) Long-term recurrence and chronic pain after repair for small umbilical or epigastric hernias: a regional cohort study. Am J Surg 209:725–732CrossRefPubMed
10.
Zurück zum Zitat Venclauskas L. Jokubauskas M, Zilinskas J, Zviniene K, Kiudelis M (2017) Long-term follow-up results of umbilical hernia repair. Wideochir Inne Tech Maloinwazyjne 12(4):350–356. Venclauskas L. Jokubauskas M, Zilinskas J, Zviniene K, Kiudelis M (2017) Long-term follow-up results of umbilical hernia repair. Wideochir Inne Tech Maloinwazyjne 12(4):350–356.
11.
Zurück zum Zitat Amato G, Romano G, Goetze T, Salamone G, Agrusa A, Gulotta G, Paolucci V (2011) New mesh shape and improved implantation procedure to simplify and standardize open ventral hernia repair: a preliminary report. Hernia 15(6):659–665 Amato G, Romano G, Goetze T, Salamone G, Agrusa A, Gulotta G, Paolucci V (2011) New mesh shape and improved implantation procedure to simplify and standardize open ventral hernia repair: a preliminary report. Hernia 15(6):659–665
12.
Zurück zum Zitat Romano G, Calò PG, Erdas E, Medas F, Gordini L, Podda F, Amato G (2017) Fixation-free incisional hernia repair in the elderly: our experience with a tentacle-shaped implant. Aging Clin Exp Res 29(Suppl 1):173–177 Romano G, Calò PG, Erdas E, Medas F, Gordini L, Podda F, Amato G (2017) Fixation-free incisional hernia repair in the elderly: our experience with a tentacle-shaped implant. Aging Clin Exp Res 29(Suppl 1):173–177
13.
Zurück zum Zitat Amato G, Romano G, Goetze T, Gordini L, Erdas E, Medas F, Calò P (2017) Long-Term Results of Fixation-Free Incisional Hernia Repair Using a Tentacle-Shaped Implant. Surg Technol Int 30:175–181 Amato G, Romano G, Goetze T, Gordini L, Erdas E, Medas F, Calò P (2017) Long-Term Results of Fixation-Free Incisional Hernia Repair Using a Tentacle-Shaped Implant. Surg Technol Int 30:175–181
14.
Zurück zum Zitat Amato G, Romano G, Agrusa A, Cassata G, Salamone G, Gulotta G (2010) Prosthetic strap system for simplified ventral hernia repair: results of a porcine experimental model. Hernia 14:389–395CrossRefPubMed Amato G, Romano G, Agrusa A, Cassata G, Salamone G, Gulotta G (2010) Prosthetic strap system for simplified ventral hernia repair: results of a porcine experimental model. Hernia 14:389–395CrossRefPubMed
15.
Zurück zum Zitat Fatton B, Amblard J, Debodinance P, Cosson M, Jacquetin B (2007) Transvaginal repair of genital prolapse:preliminary results of a new tension-free vaginal mesh (Prolift technique)-a case series multicentric study. Int Urogynecol J Pelvic Floor Dysfunct 18:743–752CrossRefPubMed Fatton B, Amblard J, Debodinance P, Cosson M, Jacquetin B (2007) Transvaginal repair of genital prolapse:preliminary results of a new tension-free vaginal mesh (Prolift technique)-a case series multicentric study. Int Urogynecol J Pelvic Floor Dysfunct 18:743–752CrossRefPubMed
16.
Zurück zum Zitat Altman D, Väyrynen T, Engh ME, Axelsen S, Falconer C (2008) For the Nordic Transvaginal Mesh Group. Short-term outcome after transvaginal mesh repair of pelvic organ prolapse. Int Urogynecol J Pelvic Floor Dysfunct 19(6):787–793 Altman D, Väyrynen T, Engh ME, Axelsen S, Falconer C (2008) For the Nordic Transvaginal Mesh Group. Short-term outcome after transvaginal mesh repair of pelvic organ prolapse. Int Urogynecol J Pelvic Floor Dysfunct 19(6):787–793
17.
Zurück zum Zitat Abdel-Fattah M, Ramsay I (2008) West of Scotland Study Group. Retrospective multicentre study of the new minimally invasive mesh repair devices for pelvic organ prolapse. BJOG 115(1):22–30 Abdel-Fattah M, Ramsay I (2008) West of Scotland Study Group. Retrospective multicentre study of the new minimally invasive mesh repair devices for pelvic organ prolapse. BJOG 115(1):22–30
18.
Zurück zum Zitat Malik B, Lambaudie E, Collinet P, Dubois P, Cosson M (2007) Mechanical resistance of synthetic meshes for incontinence or prolapse surgery. Int Urogynecol J 18:183–187 Malik B, Lambaudie E, Collinet P, Dubois P, Cosson M (2007) Mechanical resistance of synthetic meshes for incontinence or prolapse surgery. Int Urogynecol J 18:183–187
19.
Zurück zum Zitat Alcalay M, Livne M, Shidlovsky D, Hod E. (2009) Pullout force of polypropylene mesh deployed by endofast reliant fastener—a comparative study in a sheep model. ICS Congress Abstract 543 Alcalay M, Livne M, Shidlovsky D, Hod E. (2009) Pullout force of polypropylene mesh deployed by endofast reliant fastener—a comparative study in a sheep model. ICS Congress Abstract 543
Metadaten
Titel
Tentacle-shaped mesh for fixation-free repair of umbilical hernias
verfasst von
G. Amato
G. Romano
A. Agrusa
G. L. Canu
E. Gulotta
E. Erdas
P. G. Calò
Publikationsdatum
12.04.2019
Verlag
Springer Paris
Erschienen in
Hernia / Ausgabe 4/2019
Print ISSN: 1265-4906
Elektronische ISSN: 1248-9204
DOI
https://doi.org/10.1007/s10029-019-01950-8

Weitere Artikel der Ausgabe 4/2019

Hernia 4/2019 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.