The effect of posterior corneal flat meridian and astigmatism amount on the total corneal astigmatism estimated from anterior corneal measurements

verfasst von: Youngsub Eom, Su-Yeon Kang, Hyo Myung Kim, Jong Suk Song

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 11/2014

Einloggen, um Zugang zu erhalten

Abstract

Background

To evaluate the effects of posterior corneal astigmatism and the absolute flat meridian difference between anterior and posterior corneal surfaces (AMDAnt-Post) on the estimation of total corneal astigmatism using anterior corneal measurements (simulated keratometry [K]).

Methods

Ninety-nine eyes of 99 healthy participants were enrolled. Anterior, posterior, and total mean corneal power, cylinder power, flat meridian, and vector components J0, and J45 measured by a dual Scheimpflug camera were analyzed. The correlation between the posterior corneal cylinder power, AMDAnt-Post, and the difference in the cylinder power between simulated K and total cornea (cylinder power differenceSimK-Tot) were evaluated.

Results

The cylinder power differenceSimK-Tot was positively correlated with the posterior corneal cylinder power (rho = 0.704 and P < 0.001) and negatively correlated with AMDAnt-Post (rho = −0.717 and P < 0.001). In the multivariate linear regression analysis, anterior corneal J0 was strongly associated with the posterior corneal cylinder power and the AMDAnt-Post. When corneal J0 had a positive value, the cylinder power of simulated K tended to be larger than the total corneal cylinder power. In comparison, the opposite trend was presented in eyes with negative anterior corneal J0. When anterior corneal J0 was larger than 1.0 or smaller than −0.9, the errors from estimating the total corneal cylinder power using anterior corneal measurements tended to be larger than 0.25 D.

Conclusion

Posterior corneal astigmatism should be considered for more accurate corneal astigmatism predictions, especially in eyes with anterior corneal astigmatism greater than 2.0 D of with-the-rule astigmatism or greater than 1.8 D of against-the-rule astigmatism.
Literatur
2.
Zurück zum Zitat Koch DD, Ali SF, Weikert MP, Shirayama M, Jenkins R, Wang L (2012) Contribution of posterior corneal astigmatism to total corneal astigmatism. J Cataract Refract Surg 38:2080–2087PubMedCrossRef Koch DD, Ali SF, Weikert MP, Shirayama M, Jenkins R, Wang L (2012) Contribution of posterior corneal astigmatism to total corneal astigmatism. J Cataract Refract Surg 38:2080–2087PubMedCrossRef
3.
Zurück zum Zitat Chen D, Lam AK (2007) Intrasession and intersession repeatability of the Pentacam system on posterior corneal assessment in the normal human eye. J Cataract Refract Surg 33:448–454PubMedCrossRef Chen D, Lam AK (2007) Intrasession and intersession repeatability of the Pentacam system on posterior corneal assessment in the normal human eye. J Cataract Refract Surg 33:448–454PubMedCrossRef
4.
Zurück zum Zitat Ho JD, Tsai CY, Tsai RJ, Kuo LL, Tsai IL, Liou SW (2008) Validity of the keratometric index: evaluation by the Pentacam rotating Scheimpflug camera. J Cataract Refract Surg 34:137–145PubMedCrossRef Ho JD, Tsai CY, Tsai RJ, Kuo LL, Tsai IL, Liou SW (2008) Validity of the keratometric index: evaluation by the Pentacam rotating Scheimpflug camera. J Cataract Refract Surg 34:137–145PubMedCrossRef
5.
Zurück zum Zitat Aramberri J, Araiz L, Garcia A, Illarramendi I, Olmos J, Oyanarte I, Romay A, Vigara I (2012) Dual versus single Scheimpflug camera for anterior segment analysis: precision and agreement. J Cataract Refract Surg 38:1934–1949PubMedCrossRef Aramberri J, Araiz L, Garcia A, Illarramendi I, Olmos J, Oyanarte I, Romay A, Vigara I (2012) Dual versus single Scheimpflug camera for anterior segment analysis: precision and agreement. J Cataract Refract Surg 38:1934–1949PubMedCrossRef
6.
Zurück zum Zitat Wang L, Shirayama M, Koch DD (2010) Repeatability of corneal power and wavefront aberration measurements with a dual-Scheimpflug Placido corneal topographer. J Cataract Refract Surg 36:425–430 Wang L, Shirayama M, Koch DD (2010) Repeatability of corneal power and wavefront aberration measurements with a dual-Scheimpflug Placido corneal topographer. J Cataract Refract Surg 36:425–430
7.
Zurück zum Zitat Dubbelman M, Sicam VA, van der Heijde RG (2007) The contribution of the posterior surface to the coma aberration of the human cornea. J Vis 7(10):11–18CrossRef Dubbelman M, Sicam VA, van der Heijde RG (2007) The contribution of the posterior surface to the coma aberration of the human cornea. J Vis 7(10):11–18CrossRef
8.
Zurück zum Zitat Ho JD, Liou SW, Tsai RJ, Tsai CY (2010) Effects of aging on anterior and posterior corneal astigmatism. Cornea 29:632–637PubMed Ho JD, Liou SW, Tsai RJ, Tsai CY (2010) Effects of aging on anterior and posterior corneal astigmatism. Cornea 29:632–637PubMed
9.
Zurück zum Zitat Pinero DP, Camps VJ, Caravaca-ArensE, Perez-CambrodiRJ, Artola A (2014) Estimation of the central corneal power in keratoconus: theoretical and clinical assessment of the error of the keratometric approach. Cornea 33:274–279 Pinero DP, Camps VJ, Caravaca-ArensE, Perez-CambrodiRJ, Artola A (2014) Estimation of the central corneal power in keratoconus: theoretical and clinical assessment of the error of the keratometric approach. Cornea 33:274–279
10.
Zurück zum Zitat Wang L, Mahmoud AM, Anderson BL, Koch DD, Roberts CJ (2011) Total corneal power estimation: ray tracing method versus gaussian optics formula. Invest Ophthalmol Vis Sci 52:1716–1722PubMedCrossRef Wang L, Mahmoud AM, Anderson BL, Koch DD, Roberts CJ (2011) Total corneal power estimation: ray tracing method versus gaussian optics formula. Invest Ophthalmol Vis Sci 52:1716–1722PubMedCrossRef
11.
Zurück zum Zitat Thibos LN, Wheeler W, Horner D (1997) Power vectors: an application of Fourier analysis to the description and statistical analysis of refractive error. Optom Vis Sci 74:367–375PubMedCrossRef Thibos LN, Wheeler W, Horner D (1997) Power vectors: an application of Fourier analysis to the description and statistical analysis of refractive error. Optom Vis Sci 74:367–375PubMedCrossRef
12.
Zurück zum Zitat Holladay JT, Moran JR, Kezirian GM (2001) Analysis of aggregate surgically induced refractive change, prediction error, and intraocular astigmatism. J Cataract Refract Surg 27:61–79PubMedCrossRef Holladay JT, Moran JR, Kezirian GM (2001) Analysis of aggregate surgically induced refractive change, prediction error, and intraocular astigmatism. J Cataract Refract Surg 27:61–79PubMedCrossRef
13.
Zurück zum Zitat Hoffmann PC, Auel S, Hutz WW (2011) Results of higher power toric intraocular lens implantation. J Cataract Refract Surg 37:1411–1418PubMedCrossRef Hoffmann PC, Auel S, Hutz WW (2011) Results of higher power toric intraocular lens implantation. J Cataract Refract Surg 37:1411–1418PubMedCrossRef
14.
Zurück zum Zitat Bauer NJ, de Vries NE, Webers CA, Hendrikse F, Nuijts RM (2008) Astigmatism management in cataract surgery with the AcrySof toric intraocular lens. J Cataract Refract Surg 34:1483–1488PubMedCrossRef Bauer NJ, de Vries NE, Webers CA, Hendrikse F, Nuijts RM (2008) Astigmatism management in cataract surgery with the AcrySof toric intraocular lens. J Cataract Refract Surg 34:1483–1488PubMedCrossRef
15.
Zurück zum Zitat Tejedor J, Guirao A (2013) Agreement between refractive and corneal astigmatism in pseudophakic eyes. Cornea 32:783–790PubMedCrossRef Tejedor J, Guirao A (2013) Agreement between refractive and corneal astigmatism in pseudophakic eyes. Cornea 32:783–790PubMedCrossRef
16.
Zurück zum Zitat Ho JD, Tsai CY, Liou SW (2009) Accuracy of corneal astigmatism estimation by neglecting the posterior corneal surface measurement. Am J Ophthalmol 147:788–795, 795 e781-782PubMedCrossRef Ho JD, Tsai CY, Liou SW (2009) Accuracy of corneal astigmatism estimation by neglecting the posterior corneal surface measurement. Am J Ophthalmol 147:788–795, 795 e781-782PubMedCrossRef
17.
Zurück zum Zitat Lowe RF, Clark BA (1973) Posterior corneal curvature. Correlations in normal eyes and in eyes involved with primary angle-closure glaucoma. Br J Ophthalmol 57:464–470 Lowe RF, Clark BA (1973) Posterior corneal curvature. Correlations in normal eyes and in eyes involved with primary angle-closure glaucoma. Br J Ophthalmol 57:464–470
18.
Zurück zum Zitat Dubbelman M, Weeber HA, van der Heijde RG, Volker-DiebenHJ (2002) Radius and asphericity of the posterior corneal surface determined by corrected Scheimpflug photography. Acta Ophthalmol Scand 80:379–383 Dubbelman M, Weeber HA, van der Heijde RG, Volker-DiebenHJ (2002) Radius and asphericity of the posterior corneal surface determined by corrected Scheimpflug photography. Acta Ophthalmol Scand 80:379–383
19.
Zurück zum Zitat Liu YC, Chou P, Wojciechowski R, Lin PY, Liu CJ, Chen SJ, Liu JH, Hsu WM, Cheng CY (2011) Power vector analysis of refractive, corneal, and internal astigmatism in an elderly Chinese population: the Shihpai Eye Study. Invest Ophthalmol Vis Sci 52:9651–9657PubMedCrossRef Liu YC, Chou P, Wojciechowski R, Lin PY, Liu CJ, Chen SJ, Liu JH, Hsu WM, Cheng CY (2011) Power vector analysis of refractive, corneal, and internal astigmatism in an elderly Chinese population: the Shihpai Eye Study. Invest Ophthalmol Vis Sci 52:9651–9657PubMedCrossRef
20.
Zurück zum Zitat Eom Y, Nam KT, Kang SY, Kim HM, Song JS (2013) Axis difference between corneal and internal astigmatism to consider for toric intraocular lenses. Am J Ophthalmol 156(1112):1119, e1112 Eom Y, Nam KT, Kang SY, Kim HM, Song JS (2013) Axis difference between corneal and internal astigmatism to consider for toric intraocular lenses. Am J Ophthalmol 156(1112):1119, e1112
Metadaten
Titel
The effect of posterior corneal flat meridian and astigmatism amount on the total corneal astigmatism estimated from anterior corneal measurements
verfasst von
Youngsub Eom
Su-Yeon Kang
Hyo Myung Kim
Jong Suk Song
Publikationsdatum
01.11.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 11/2014
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-014-2737-9

Weitere Artikel der Ausgabe 11/2014

Graefe's Archive for Clinical and Experimental Ophthalmology 11/2014 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.