Open Access 14.03.2022 | Original Article

Unenhanced abdominal low-dose CT reconstructed with deep learning-based image reconstruction: image quality and anatomical structure depiction

verfasst von: Tetsuro Kaga, Yoshifumi Noda, Takayuki Mori, Nobuyuki Kawai, Toshiharu Miyoshi, Fuminori Hyodo, Hiroki Kato, Masayuki Matsuo

Erschienen in: Japanese Journal of Radiology | Ausgabe 7/2022

Abstract

Purpose

To evaluate the utility of deep learning-based image reconstruction (DLIR) algorithm in unenhanced abdominal low-dose CT (LDCT).

Materials and methods

Two patient groups were included in this prospective study: 58 consecutive patients who underwent unenhanced abdominal standard-dose CT reconstructed with hybrid iterative reconstruction (SDCT group) and 48 consecutive patients who underwent unenhanced abdominal LDCT reconstructed with high strength level of DLIR (LDCT group). The background noise and signal-to-noise ratio (SNR) of the liver, pancreas, spleen, kidney, abdominal aorta, inferior vena cava, and portal vein were calculated. Two radiologists qualitatively assessed the overall image noise, overall image quality, and abdominal anatomical structures depiction. Quantitative and qualitative parameters and size-specific dose estimates (SSDE) were compared between SDCT and LDCT groups.

Results

The background noise was lower in LDCT group than in SDCT group (P = 0.02). SNRs were higher in LDCT group than in SDCT group (P < 0.001–0.004) except for the liver. Overall image noise was superior in LDCT group than in SDCT group (P < 0.001). Overall image quality was not different between SDCT and LDCT groups (P = 0.25–0.26). Depiction of almost all abdominal anatomical structures was equal to or better in LDCT group than in SDCT group (P < 0.001–0.88). The SSDE was lower in LDCT group (4.0 mGy) than in SDCT group (20.6 mGy) (P < 0.001).

Conclusions

DLIR facilitates substantial radiation dose reduction of > 75% and significantly reduces background noise. DLIR can maintain image quality and anatomical structure depiction in unenhanced abdominal LDCT.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Unenhanced computed tomography (CT) has become indispensable in modern medicine in applications such as screening and diagnosis of acute diseases, assessment of therapeutic response, tumor recurrence, and follow-up for various patients’ conditions [1]. However, repeated CT examinations can lead to excessive medical radiation exposure, which increases the risk of adverse events to clinical staff and patients [2]. Radiation dose reduction strategies have always been a high priority issue in CT examinations. However, they tend to facilitate increased image noise, which can deteriorate image quality and diagnostic performance [3].
Image reconstruction techniques for the acquisition of CT images have been remarkably evolved in recent years [4]. For example, the statistical iterative reconstruction (IR) technique, especially the hybrid-IR technique, has managed to reduce image noise without image quality degradation compared with filtered back projection (FBP) [5]. Consequently, hybrid-IR has now become widely used in clinical setting. Hybrid-IR technique had been applied to low-dose CT (LDCT) protocol because it could significantly reduce image noise. However, LDCT images reconstructed with hybrid-IR were inferior to standard-dose CT (SDCT) images reconstructed with FBP in terms of detectability of low-contrast lesions and spatial resolution [69]. The detectability of low-contrast lesion was crucial for diagnostic imaging, as a result, the clinical use of abdominal LDCT has been limited.
Recently, the deep learning-based image reconstruction (DLIR) technique has been introduced as a next-generation CT image reconstruction method [10, 11]. Noda et al. [12, 13] have reported the impact of DLIR on whole-body and abdominal contrast-enhanced LDCT protocols, and achieved extremely low radiation dose (2.9 mGy and 2.3 mGy in CT dose-index volumes [CTDIvol]), while maintaining effective image quality and lesion detectability. Jensen et al. [14] reported that contrast-enhanced LDCT reconstructed with DLIR achieve 65% radiation dose reduction and while maintaining the detectability of liver lesions compared to SDCT reconstructed with FBP. To the best of our knowledge, however, no study has evaluated the use of a DLIR in an unenhanced abdominal LDCT protocol. Therefore, in this study, we aim to evaluate the usefulness of the DLIR technique in unenhanced abdominal LDCT protocol for the assessment of image quality and to compare with SDCT reconstructed using hybrid-IR technique.

Materials and methods

Phantom study

We performed CT scans of the self-made phantom to assess the objective image quality. This phantom study was conducted under the same scan protocols as shown in “Imaging technique” section without dose modulation, and noise power spectrum (NPS) and modulation transfer function (MTF) were calculated using CT measure software (version 0.98f; Japanese Society of CT Technology, Hiroshima, Japan). NPS was calculated by the radial frequency method with a square circular regions of interest (ROI) [15] on CT image of self-made water phantom. MTF was calculated by circular edge technique [16] using a self-made phantom in which acrylic resin rod was immersed in water.

Human clinical study

This prospective study was approved by our Institutional Review Board. This study consisted of two patient group sets (Fig. 1). The first group initially included 59 consecutive patients who underwent unenhanced abdominal SDCT from March 2020 to April 2020; however, one patient was excluded because the CT scan was performed in the lateral decubitus position. Thus, the remaining 58 patients were included in this study (SDCT group). Informed consent was waived for this group because all data were retrospectively collected. In the second group, 48 consecutive patients who underwent unenhanced abdominal LDCT from May 2020 to October 2020 were included, and written informed consent was obtained from all patients prior to enrollment in our study because all data were prospectively collected. We did not observe any unexpected events, including technical failure and unstable breath holding among these 48 patients. Therefore, all patients were enrolled in this study (LDCT group). Detailed patient information was obtained from the respective medical records to document the demographics of our sample.

Imaging technique

All unenhanced abdominal CT examinations were performed using a fast kilovoltage-switching dual-energy CT scanner (Revolution CT; GE Healthcare, Milwaukee, WI, USA) that was used in single-energy CT mode. The CT parameters were as follows: X-ray tube voltage, 120 kilovolt peak (kVp); noise index, 7.0 Hounsfield unit (HU) in SDCT group and 14.0 HU in LDCT group [12, 13, 17] based on 5-mm reconstruction thickness; rotation time, 0.5 s; pitch, 0.508:1; table speed, 81.3 mm/s; and detector configuration, 128 × 0.625 mm in both groups.
For the SDCT group, raw data were reconstructed with adaptive statistical iteration reconstruction-Veo (ASiR-V; GE Healthcare) of 40% with 5-mm section thickness and 0% overlap. In contrast, raw data for the LDCT group were reconstructed with DLIR (TrueFidelity™; GE Healthcare) at high strength level with 5-mm section thickness and 0% overlap. In this study, we used a high strength level of DLIR because this level has been proved to strongly reduce background noise compared with low and middle strength levels of DLIR and ASiR-V in several studies [1013, 18]. The CTDIvol, size-specific dose estimates (SSDE), and dose-length product (DLP) were recorded from the dose report.

Quantitative image analysis

A radiologist (T.M. with 2 years of post-training experience in interpreting abdominal CT images) measured the mean CT numbers of the liver, pancreas, spleen, kidney, abdominal aorta, inferior vena cava, and portal vein using a commercially available DICOM viewer by placing ROI on axial images. The CT numbers of the liver were measured using an ROI in the anterior segment, while carefully avoiding large vessels, bile ducts, focal lesions, and artifacts. The CT numbers of the pancreas were measured using an ROI, while carefully avoiding the main pancreatic duct, visible vessels, focal lesions, and artifacts. Furthermore, the CT numbers of the spleen and kidney were measured using an ROI, while carefully avoiding visible vessels, focal lesions, and artifacts. The CT numbers of the abdominal aorta at the level of the first lumbar vertebral body were measured using an ROI that incorporated as much of the vascular lumen as possible, devoid of vascular walls, calcification, thrombi, and artifacts. Finally, the CT numbers of the inferior vena cava at the level of the first lumbar vertebral body and the main portal vein were measured using an ROI that incorporated as much of the vascular lumen as possible, devoid of vascular walls, thrombi, and artifacts.
Background noise was defined as one standard deviation of the mean CT number at the homogeneous anterior abdominal subcutaneous fat tissue. The signal-to-noise ratio (SNR) of each anatomical structure was calculated by dividing the CT number of the corresponding anatomical structure by the background noise.

Qualitative image analysis

Two radiologists (T.M. and T.K., with 2 and 4 years of post-training experience in interpreting abdominal CT images, respectively), who were unaware of the CT protocols, independently and randomly reviewed the images and graded the image quality with respect to the overall image noise and overall image quality using a 5-point rating scale as follows: 5, excellent; 4, good; 3, acceptable; 2, suboptimal; and 1, unacceptable.
The same two radiologists also independently and randomly graded the depiction of the liver, pancreas, gall bladder, spleen, stomach, kidney, adrenal gland, bladder, prostate (if patients were men), uterus (if patients were women), intestinal tract, urinary duct, abdominal aorta, portal vein, and inferior vena cava using a 5-point rating scale as follows: 5, excellent; 4, good; 3, acceptable; 2, suboptimal; and 1, unacceptable.

Statistical analysis

Statistical analyses were performed using the SPSS statistical software package (version 24.0; IBM Corp., Armonk, NY, USA). The data were tested for normal distribution using the Shapiro–Wilk test. We used unpaired t-test or chi-square test to compare patients’ age, sex, height, body weight, and body mass index between the SDCT and LDCT groups. We used the Mann–Whitney U test to compare the CTDIvol, SSDE, DLP, CT numbers, background noise, SNRs, and confidence ratings for overall image noise, overall image quality, and depiction of abdominal structures between the SDCT and LDCT groups. P values of less than 0.05 were used to denote statistical significance.
Interobserver variability in qualitative analysis was assessed using the ĸ statistics. A ĸ-value of ≤ 0.20 was interpreted as slight agreement, 0.21–0.40 as fair agreement, 0.41–0.60 as moderate agreement, 0.61–0.80 as substantial agreement, and ≥ 0.81 as almost perfect agreement [19].

Results

Phantom study

NPS and MTF are shown in Fig. 2. The NPS curve analysis showed similar spatial frequency profile between LDCT protocol reconstructed with high strength level of DLIR and SDCT protocol reconstructed with ASiR-V of 40%, but the NPS value was higher in LDCT protocol reconstructed with high strength level of DLIR than in SDCT protocol reconstructed with ASiR-V of 40%. MTF10% values were 0.65 in LDCT reconstructed with high strength level of DLIR and 0.65 in SDCT reconstructed with ASiR-V of 40%, that meant both of spatial resolution were almost same.

Patients’ demographics and radiation dose

Patients’ demographics and radiation dose are summarized in Table 1. The SDCT group included 33 men and 25 women (mean age, 62.6 ± 16.0 years; age range, 18–90 years). Among them, three patients were status post hysterectomy, one was status post cholecystectomy, one was status post cystectomy, one was status post prostatectomy, and one was placed bilateral ureteral stent. The LDCT group included 26 men and 22 women (mean age 59.2 ± 16.0 years; age range 19–86 years). Among them, four patients were status post cholecystectomy, four were status post hysterectomy, and one was status post cystectomy. We did not assess the depiction of the corresponding structures in patients who had undergone surgery or treatment. No difference was noted in terms of patients’ age (P = 0.27), sex (P = 0.78), height (P = 0.97), body weight (P = 0.74), and body mass index (P = 0.72) between the SDCT and LDCT groups. The median CTDIvol (14.8 mGy in SDCT group and 3.1 mGy in LDCT group; P < 0.001), median SSDE (20.0 mGy in SDCT group and 4.0 mGy in LDCT group; P < 0.001), and median DLP (810.9 mGy∙cm in SDCT group and 172.8 mGy∙cm in LDCT group: P < 0.001) were lower in the LDCT group than in the SDCT group. The average reduction rates of the CTDIvol, SSDE, and DLP were 78.9%, 80.5%, and 78.9%, respectively.
Table 1
Patients’ demographics and radiation dose
Parameter
SDCT group
LDCT group
P value
Age (y)
62.6 ± 16.0 (18–90)
59.2 ± 16.0 (19–86)
0.27
Men:Women
33:25
26:22
0.78
Height (cm)
162.0 ± 8.8 (145–184)
162.1 ± 11.4 (115–180)
0.97
Body Weight (kg)
57.8 ± 11.5 (36.7–85.9)
58.8 ± 11.7 (27.0–83.0)
0.74
Body Mass Index (kg/cm2)
22.0 ± 0.5 (15.8–29.7)
22.2 ± 0.5 (12.5–28.1)
0.72
CTDIvol (mGy)
14.8 (11.3–17.9)
3.1 (2.8–3.6)
 < 0.001
SSDE (mGy)
20.0 (17.1–23.0)
4.0 (3.7–4.4)
 < 0.001
DLP (mGy∙cm)
810.9 (612.5–990.2)
172.8 (141.5–204.9)
 < 0.001
Data are means ± 1 standard deviation with ranges in parentheses in patients’ demographics. Data are medians with interquartile ranges in parentheses in radiation dose
SDCT standard-dose CT, LDCT low-dose CT, CTDIvol CT dose-index volume, SSDE size-specific dose estimates, DLP dose-length product

Quantitative image analysis

The CT numbers, background noise, and SNRs are summarized in Table 2. Between the SDCT and LDCT groups, the CT numbers of almost all anatomical structures were statistically higher in the LDCT group compared to the SDCT group except for the liver (P < 0.001–0.048). However, the difference in the CT numbers ranged only between 0.9 and 3.4 HU. The background noise was observed to be lower in the LDCT group compared to the SDCT group (P = 0.02). Moreover, the SNRs of almost all anatomical structures were higher in the LDCT group than in the SDCT group except for the liver (P < 0.001–0.004). No difference was found in the CT number (P = 0.45) and SNR (P = 0.07) of the liver between the SDCT and LDCT groups.
Table 2
CT numbers and signal-to-noise ratios of the abdominal anatomical structures and background noise
Anatomical structure
SDCT group
LDCT group
P value
Liver
   
 CT number
58.6 (55.1–63.3)
60.3 (53.2–63.3)
0.45
 SNR
9.9 (8.2–11.6)
11.3 (9.1–12.8)
0.07
Pancreas
   
 CT number
45.7 (42.6–48.8)
48.7 (43.5–51.7)
0.02
 SNR
7.5 (6.8–8.9)
8.9 (7.8–10.3)
0.002
Spleen
   
 CT number
50.1 (48.3–52.3)
51.0 (49.4–54.4)
0.048
 SNR
8.7 (7.8–9.8)
10.0 (8.3–11.3)
0.004
Kidney
   
 CT number
37.7 (35.8–39.5)
38.8 (37.4–40.8)
0.001
 SNR
6.3 (5.8–7.3)
7.7 (6.3–8.7)
0.003
Abdominal aorta
   
 CT number
44.3 (40.4–46.5)
47.7 (44.9–50.4)
 < 0.001
 SNR
7.6 (6.2–8.6)
9.2 (7.7–10.7)
 < 0.001
Inferior vena cava
   
 CT number
43.8 (40.1–47.3)
46.6 (43.5–49.7)
0.002
 SNR
7.6 (6.1–8.7)
9.2 (7.5–10.3)
0.002
Portal vein
   
 CT number
41.3 (39.4–44.0)
43.6 (41.2–46.1)
0.002
 SNR
7.2 (6.1–8.0)
8.6 (7.0–9.3)
0.003
Background noise
6.0 (5.1–6.4)
5.1 (4.6–6.1)
0.02
Data are medians with interquartile ranges in parentheses
SDCT standard-dose CT, LDCT low-dose CT, SNR signal-to-noise ratio

Qualitative image analysis

The rating scores for overall image noise, overall image quality, and depiction of abdominal structure are shown in Table 3. The rating scores for the overall image noise were higher in the LDCT than in the SDCT group (P < 0.001 for both radiologists). No difference was observed in the rating scores for overall image quality between the SDCT and LDCT groups (P = 0.26 and 0.25 for radiologists 1 and 2, respectively). The ĸ-values ranged from 0.36 to 0.67, indicating fair to moderate agreement between the two radiologists.
Table 3
The rating scores for image noise, overall image quality, and depiction of abdominal anatomical structures in each reconstruction algorithm
Anatomical structure
SDCT group
LDCT group
P value
Radiologist 1
Radiologist 2
κ value
Radiologist 1
Radiologist 2
κ value
Radiologist 1
Radiologist 2
Overall image noise
4 (4–4)
4 (4–4)
0.67
4 (4–5)
5 (4–5)
0.39
 < 0.001
 < 0.001
Overall image quality
4 (4–4)
4 (4–4)
0.6
4 (4–4)
4 (4–4)
0.36
0.26
0.25
Depiction of anatomical structure
 
 Liver
5 (4–5)
5 (4–5)
0.68
5 (4–5)
5 (4–5)
0.45
0.13
0.11
 Pancreas
4 (3–4)
4 (4–4)
0.69
4 (4–5)
4 (4–4)
0.58
0.28
0.06
 Gall bladder
4 (3–5)
3 (4–4)
0.61
4 (3–5)
3 (4–4)
0.69
0.62
0.42
 Spleen
5 (4–5)
5 (4–5)
0.71
5 (4–5)
5 (4–5)
0.43
0.43
0.30
 Stomach
4 (4–5)
4 (4–4)
0.6
5 (4–5)
4 (4–5)
0.40
0.14
0.04
 Kidney
4 (4–5)
4 (4–5)
0.65
5 (4–5)
4.5 (4–5)
0.38
0.04
0.046
 Adrenal gland
4 (3–4)
4 (3–4)
0.66
4 (4–4)
4 (4–4)
0.56
0.09
 < 0.001
 Bladder
4 (4–5)
4 (3–4)
0.45
4 (4–5)
4 (4–4)
0.27
0.35
0.002
 Prostate
4 (4–5)
4 (4–4)
0.39
4 (4–4)
4 (3–4)
0.78
0.03
0.47
 Uterus
4 (4–5)
4 (3–4)
0.54
4 (3–5)
4 (4–5)
0.51
0.64
0.09
 Intestinal tract
4 (3–5)
4 (3–4)
0.73
4 (4–4)
4 (4–4)
0.77
0.03
0.005
 Urinary duct
4 (3–4)
4 (3–4)
0.84
4 (3–4)
4 (4–4)
0.69
0.03
0.004
 Abdominal aorta
4 (4–5)
4 (4–5)
0.56
4.5 (4–5)
5 (4–5)
0.51
0.42
0.88
 Portal vein
3 (2–4)
3 (3–4)
0.68
3 (3–4)
4 (3–4)
0.54
0.36
0.01
 Inferior vena cava
5 (4–5)
4 (4–5)
0.63
4 (4–5)
4 (4–5)
0.50
0.69
0.58
Data are medians with interquartile ranges in parentheses
SDCT standard dose CT, LDCT low-dose CT
The depiction scores of the liver, pancreas, gall bladder, spleen, uterus, abdominal aorta, and inferior vena cava were not different between the SDCT and LDCT groups for both radiologists (P = 0.06–0.88). In contrast, the depiction scores of the kidney, intestinal tract, and urinary duct were higher in the LDCT group than in the SDCT group for both radiologists (P = 0.004–0.046) (Fig. 3), whereas the scores of the stomach, adrenal gland, bladder, and portal vein were higher in the LDCT group than in the SDCT group only for radiologist 2 (P < 0.001–0.04) (Fig. 4). The depiction score of the prostate was found to be higher in the SDCT group than in the LDCT group only for radiologist 1 (P = 0.03). The ĸ-values ranged from 0.27 to 0.84, indicating fair to almost perfect agreement between the two radiologists.

Discussion

Our study has demonstrated that unenhanced abdominal LDCT protocol with DLIR algorithm allows > 75% radiation dose reduction in comparison to the SDCT protocol with hybrid-IR. In phantom study, the NPS value was higher in LDCT protocol reconstructed with DLIR algorithm than SDCT protocol reconstructed with ASiR-V of 40%. Additionally in human clinical study, LDCT protocol with DLIR algorithm was maintaining good image quality and depiction of abdominal anatomical structures.
To verify the utility of DLIR in LDCT protocol, several phantom studies had performed by comparing hybrid- or model based IR algorithms and showed that DLIR demonstrate less noise and higher detectability in same dose reduction level compared with the other reconstruction algorithms [20, 21]. In our phantom study, the NPS value was higher in LDCT reconstructed with DLIR than in SDCT reconstructed with ASiR-V of 40% though spatial resolutions were almost same. Nevertheless, overall image quality and noise level of LDCT protocol with DLIR algorithm was equivalent or better than those of SDCT protocol with ASiR-V of 40% in human clinical study. The following points could be given as reasons, the trivial discrepancy of NPS between the two protocols, preserved spatial resolution, and maintained image texture. As a result, radiologists’ subjective acceptance was kept even in LDCT with DLIR protocol.
DLIR managed to reduce image noise in the LDCT protocol in this study and this finding is consistent with the results of a previous study which compared image noise in the LDCT protocol with hybrid-IR and DLIR in patient [13, 14]. The vendor-specific DLIR was developed with high-quality FBP datasets to learn how to differentiate noise from signals and to suppress noise without changing the image texture [22]. In previous studies, SDCT images reconstructed with DLIR were superior to SDCT images reconstructed with hybrid-IR technique in terms of image noise, SNR, and contrast-to-noise ratio [10, 11, 18]. Both quantitative and qualitative image noise were significantly reduced in the LDCT group compared with the SDCT group in this study. In general, one of the most crucial issues in the LDCT protocol is noise reduction; however, it can be suggested that the development of DLIR has allowed us to overcome this issue. In addition, the overall image quality of the LDCT group was comparable with the SDCT group. We believe that DLIR can achieve effective noise reduction but also maintain high levels of overall image quality even in the LDCT protocol.
The depiction of abdominal structures in the LDCT group was equal to or greater than in the SDCT group. Maintaining a normal organ depiction is a minimum requirement for clinical use; thus, we believe that LDCT images reconstructed with DLIR can satisfy this requirement. The notable point is that even small or narrow organs, such as the adrenal gland and the urinary duct, were clearly depicted in the LDCT group (Figs. 3 and 4). Indeed, Noda et al. [13] have already reported that there were no differences in lesion detectability between the SDCT images reconstructed with hybrid-IR and LDCT images reconstructed with high strength level of DLIR in contrast-enhanced whole-body CT. Similarly, Singh et al. [17] reported that all clinically important lesions could be detected on contrast-enhanced LDCT with DLIR. Moreover, Jensen et al. [14] reported the utility of contrast-enhanced LDCT with DLIR for detecting liver lesions. Although we did not evaluate the diagnostic ability, we believe that our results provide solid evidence for the feasibility of unenhanced abdominal LDCT using DLIR. Additionally, our study has the superiority compared with the previous studies in that we can show the feasibility of unenhanced CT, because unenhanced CT is more broadly used in clinical setting than contrast-enhanced CT.
DLIR has only recently become available. However, the hybrid-IR technique is still the mainstream for CT image reconstruction. The application of the hybrid-IR technique to the LDCT protocol had been considered; therefore, multiple studies have already investigated the efficiency of this method [68, 2325]. Most of the studies concluded that LDCT reconstructed with the hybrid-IR technique could maintain an efficient image quality but undermine spatial resolution and the detectability of low-contrast lesions compared to SDCT reconstructed with FBP. Thus, the clinical use of abdominal LDCT reconstructed with hybrid-IR had been limited. Our study suggests that the advent of DLIR can improve LDCT image quality and make it comparable, or even superior, to the image quality of SDCT reconstructed with hybrid-IR technique. The clinical application of LDCT was expected in the scene like screening.
This study has several limitations. First, the study population was relatively small, and our investigation was carried out at a single institution. Second, the patients’ body size was small. Third, we were not able to compare lesion detectability between the SDCT and LDCT groups because we could not define reference standard. Finally, we only used a CT scanner from a single vendor. Therefore, further clinical studies on larger populations are required to validate our results and evaluate lesion detectability of LDCT with DLIR protocol for other types of CT scanners.

Conclusions

DLIR allows substantial radiation dose reduction of > 75%, and it significantly reduces the background noise. Additionally, this study reveals that DLIR can maintain adequate image quality and anatomical structure depiction in unenhanced abdominal LDCT.

Declarations

Conflict of interest

The authors declare that they have no conflict of interest.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Pandharipande PV, Reisner AT, Binder WD, Zaheer A, Gunn ML, Linnau KF, et al. CT in the emergency department: a real-time study of changes in physician decision making. Radiology. 2016;278(3):812–21.CrossRef Pandharipande PV, Reisner AT, Binder WD, Zaheer A, Gunn ML, Linnau KF, et al. CT in the emergency department: a real-time study of changes in physician decision making. Radiology. 2016;278(3):812–21.CrossRef
2.
Zurück zum Zitat Nickoloff EL, Alderson PO. Radiation exposures to patients from CT: reality, public perception, and policy. AJR Am J Roentgenol. 2001;177(2):285–7.CrossRef Nickoloff EL, Alderson PO. Radiation exposures to patients from CT: reality, public perception, and policy. AJR Am J Roentgenol. 2001;177(2):285–7.CrossRef
3.
Zurück zum Zitat Baskan O, Erol C, Ozbek H, Paksoy Y. Effect of radiation dose reduction on image quality in adult head CT with noise-suppressing reconstruction system with a 256 slice MDCT. J Appl Clin Med Phys. 2015;16(3):5360.CrossRef Baskan O, Erol C, Ozbek H, Paksoy Y. Effect of radiation dose reduction on image quality in adult head CT with noise-suppressing reconstruction system with a 256 slice MDCT. J Appl Clin Med Phys. 2015;16(3):5360.CrossRef
4.
Zurück zum Zitat Willemink MJ, Noel PB. The evolution of image reconstruction for CT-from filtered back projection to artificial intelligence. Eur Radiol. 2019;29(5):2185–95.CrossRef Willemink MJ, Noel PB. The evolution of image reconstruction for CT-from filtered back projection to artificial intelligence. Eur Radiol. 2019;29(5):2185–95.CrossRef
5.
Zurück zum Zitat Chen LH, Jin C, Li JY, Wang GL, Jia YJ, Duan HF, et al. Image quality comparison of two adaptive statistical iterative reconstruction (ASiR, ASiR-V) algorithms and filtered back projection in routine liver CT. Br J Radiol. 2018;91(1088):20170655.CrossRef Chen LH, Jin C, Li JY, Wang GL, Jia YJ, Duan HF, et al. Image quality comparison of two adaptive statistical iterative reconstruction (ASiR, ASiR-V) algorithms and filtered back projection in routine liver CT. Br J Radiol. 2018;91(1088):20170655.CrossRef
6.
Zurück zum Zitat Baker ME, Dong F, Primak A, Obuchowski NA, Einstein D, Gandhi N, et al. Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver. AJR Am J Roentgenol. 2012;199(1):8–18.CrossRef Baker ME, Dong F, Primak A, Obuchowski NA, Einstein D, Gandhi N, et al. Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver. AJR Am J Roentgenol. 2012;199(1):8–18.CrossRef
7.
Zurück zum Zitat McCollough CH, Yu L, Kofler JM, Leng S, Zhang Y, Li Z, et al. Degradation of CT low-contrast spatial resolution due to the use of iterative reconstruction and reduced dose levels. Radiology. 2015;276(2):499–506.CrossRef McCollough CH, Yu L, Kofler JM, Leng S, Zhang Y, Li Z, et al. Degradation of CT low-contrast spatial resolution due to the use of iterative reconstruction and reduced dose levels. Radiology. 2015;276(2):499–506.CrossRef
8.
Zurück zum Zitat Schindera ST, Odedra D, Raza SA, Kim TK, Jang HJ, Szucs-Farkas Z, et al. Iterative reconstruction algorithm for CT: can radiation dose be decreased while low-contrast detectability is preserved? Radiology. 2013;269(2):511–8.CrossRef Schindera ST, Odedra D, Raza SA, Kim TK, Jang HJ, Szucs-Farkas Z, et al. Iterative reconstruction algorithm for CT: can radiation dose be decreased while low-contrast detectability is preserved? Radiology. 2013;269(2):511–8.CrossRef
9.
Zurück zum Zitat Yu L, Vrieze TJ, Leng S, Fletcher JG, McCollough CH. Technical Note: Measuring contrast- and noise-dependent spatial resolution of an iterative reconstruction method in CT using ensemble averaging. Med Phys. 2015;42(5):2261–7.CrossRef Yu L, Vrieze TJ, Leng S, Fletcher JG, McCollough CH. Technical Note: Measuring contrast- and noise-dependent spatial resolution of an iterative reconstruction method in CT using ensemble averaging. Med Phys. 2015;42(5):2261–7.CrossRef
10.
Zurück zum Zitat Jensen CT, Liu X, Tamm EP, Chandler AG, Sun J, Morani AC, et al. Image quality assessment of abdominal CT by use of new deep learning image reconstruction: initial experience. AJR Am J Roentgenol. 2020;215(1):50–7.CrossRef Jensen CT, Liu X, Tamm EP, Chandler AG, Sun J, Morani AC, et al. Image quality assessment of abdominal CT by use of new deep learning image reconstruction: initial experience. AJR Am J Roentgenol. 2020;215(1):50–7.CrossRef
11.
Zurück zum Zitat Kaga T, Noda Y, Fujimoto K, Suto T, Kawai N, Miyoshi T, et al. Deep-learning-based image reconstruction in dynamic contrast-enhanced abdominal CT: image quality and lesion detection among reconstruction strength levels. Clin Radiol. 2021;76(9):710 e15-710 e24.CrossRef Kaga T, Noda Y, Fujimoto K, Suto T, Kawai N, Miyoshi T, et al. Deep-learning-based image reconstruction in dynamic contrast-enhanced abdominal CT: image quality and lesion detection among reconstruction strength levels. Clin Radiol. 2021;76(9):710 e15-710 e24.CrossRef
12.
Zurück zum Zitat Noda Y, Iritani Y, Kawai N, Miyoshi T, Ishihara T, Hyodo F, et al. Deep learning image reconstruction for pancreatic low-dose computed tomography: comparison with hybrid iterative reconstruction. Abdom Radiol (NY). 2021;46(9):4238–44.CrossRef Noda Y, Iritani Y, Kawai N, Miyoshi T, Ishihara T, Hyodo F, et al. Deep learning image reconstruction for pancreatic low-dose computed tomography: comparison with hybrid iterative reconstruction. Abdom Radiol (NY). 2021;46(9):4238–44.CrossRef
13.
Zurück zum Zitat Noda Y, Kaga T, Kawai N, Miyoshi T, Kawada H, Hyodo F, et al. Low-dose whole-body CT using deep learning image reconstruction: image quality and lesion detection. Br J Radiol. 2021;94(1121):20201329.CrossRef Noda Y, Kaga T, Kawai N, Miyoshi T, Kawada H, Hyodo F, et al. Low-dose whole-body CT using deep learning image reconstruction: image quality and lesion detection. Br J Radiol. 2021;94(1121):20201329.CrossRef
15.
Zurück zum Zitat Kijewski MF, Judy PF. The noise power spectrum of CT images. Phys Med Biol. 1987;32(5):565–75.CrossRef Kijewski MF, Judy PF. The noise power spectrum of CT images. Phys Med Biol. 1987;32(5):565–75.CrossRef
16.
Zurück zum Zitat Richard S, Husarik DB, Yadava G, Murphy SN, Samei E. Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys. 2012;39(7):4115–22.CrossRef Richard S, Husarik DB, Yadava G, Murphy SN, Samei E. Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys. 2012;39(7):4115–22.CrossRef
17.
Zurück zum Zitat Singh R, Digumarthy SR, Muse VV, Kambadakone AR, Blake MA, Tabari A, et al. Image quality and lesion detection on deep learning reconstruction and iterative reconstruction of submillisievert chest and abdominal CT. AJR Am J Roentgenol. 2020;214(3):566–73.CrossRef Singh R, Digumarthy SR, Muse VV, Kambadakone AR, Blake MA, Tabari A, et al. Image quality and lesion detection on deep learning reconstruction and iterative reconstruction of submillisievert chest and abdominal CT. AJR Am J Roentgenol. 2020;214(3):566–73.CrossRef
18.
Zurück zum Zitat Park C, Choo KS, Jung Y, Jeong HS, Hwang JY, Yun MS. CT iterative vs deep learning reconstruction: comparison of noise and sharpness. Eur Radiol. 2021;31(5):3156–64.CrossRef Park C, Choo KS, Jung Y, Jeong HS, Hwang JY, Yun MS. CT iterative vs deep learning reconstruction: comparison of noise and sharpness. Eur Radiol. 2021;31(5):3156–64.CrossRef
19.
Zurück zum Zitat Sim J, Wright CC. The kappa statistic in reliability studies: use, interpretation, and sample size requirements. Phys Ther. 2005;85(3):257–68.CrossRef Sim J, Wright CC. The kappa statistic in reliability studies: use, interpretation, and sample size requirements. Phys Ther. 2005;85(3):257–68.CrossRef
21.
Zurück zum Zitat Greffier J, Dabli D, Hamard A, Belaouni A, Akessoul P, Frandon J, et al. Effect of a new deep learning image reconstruction algorithm for abdominal computed tomography imaging on image quality and dose reduction compared with two iterative reconstruction algorithms: a phantom study. Quant Imaging Med Surg. 2022;12(1):229–43.CrossRef Greffier J, Dabli D, Hamard A, Belaouni A, Akessoul P, Frandon J, et al. Effect of a new deep learning image reconstruction algorithm for abdominal computed tomography imaging on image quality and dose reduction compared with two iterative reconstruction algorithms: a phantom study. Quant Imaging Med Surg. 2022;12(1):229–43.CrossRef
23.
Zurück zum Zitat Gervaise A, Osemont B, Louis M, Lecocq S, Teixeira P, Blum A. Standard dose versus low-dose abdominal and pelvic CT: comparison between filtered back projection versus adaptive iterative dose reduction 3D. Diagn Interv Imaging. 2014;95(1):47–53.CrossRef Gervaise A, Osemont B, Louis M, Lecocq S, Teixeira P, Blum A. Standard dose versus low-dose abdominal and pelvic CT: comparison between filtered back projection versus adaptive iterative dose reduction 3D. Diagn Interv Imaging. 2014;95(1):47–53.CrossRef
24.
Zurück zum Zitat Pooler BD, Lubner MG, Kim DH, Chen OT, Li K, Chen GH, et al. Prospective evaluation of reduced dose computed tomography for the detection of low-contrast liver lesions: direct comparison with concurrent standard dose imaging. Eur Radiol. 2017;27(5):2055–66.CrossRef Pooler BD, Lubner MG, Kim DH, Chen OT, Li K, Chen GH, et al. Prospective evaluation of reduced dose computed tomography for the detection of low-contrast liver lesions: direct comparison with concurrent standard dose imaging. Eur Radiol. 2017;27(5):2055–66.CrossRef
25.
Zurück zum Zitat Sagara Y, Hara AK, Pavlicek W, Silva AC, Paden RG, Wu Q. Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol. 2010;195(3):713–9.CrossRef Sagara Y, Hara AK, Pavlicek W, Silva AC, Paden RG, Wu Q. Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol. 2010;195(3):713–9.CrossRef
Metadaten
Titel
Unenhanced abdominal low-dose CT reconstructed with deep learning-based image reconstruction: image quality and anatomical structure depiction
verfasst von
Tetsuro Kaga
Yoshifumi Noda
Takayuki Mori
Nobuyuki Kawai
Toshiharu Miyoshi
Fuminori Hyodo
Hiroki Kato
Masayuki Matsuo
Publikationsdatum
14.03.2022
Verlag
Springer Nature Singapore
Erschienen in
Japanese Journal of Radiology / Ausgabe 7/2022
Print ISSN: 1867-1071
Elektronische ISSN: 1867-108X
DOI
https://doi.org/10.1007/s11604-022-01259-0

Weitere Artikel der Ausgabe 7/2022

Japanese Journal of Radiology 7/2022 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.