Open Access 01.12.2017 | Research article

Validity and reliability of fluoroscopy for digital radiography: a new way to evaluate diaphragmatic mobility

verfasst von: Bruna Estima Leal, Márcia Aparecida Gonçalves, Liseane Gonçalves Lisboa, Larissa Martins Schmitz Linné, Michelle Gonçalves de Souza Tavares, Wellington Pereira Yamaguti, Elaine Paulin

Erschienen in: BMC Pulmonary Medicine | Ausgabe 1/2017

Abstract

Background

Fluoroscopy is considered the most accurate method to evaluate the diaphragm, yet most existing methods for measuring diaphragmatic mobility using fluoroscopy are complex. To assess the validity and reliability of a new evaluation method of diaphragmatic motion using fluoroscopy by digital radiography of healthy adults.

Methods

Twenty-six adults were evaluated, according to the parameters: anthropometry and pulmonary function test. The evaluation of diaphragm mobility by means of fluoroscopy by digital radiography method was randomly conducted by two raters (A and B). The Pearson correlation coefficient and the intraclass correlation coefficient (ICC) were used to assess the concurrent validity. The inter-rater and intra-rater reliability of the measurement of diaphragmatic motion was determined using ICC and a confidence interval of 95%.

Results

There was a relationship in the assessment of the concurrent validity. There was good inter-rater reliability for right hemidiaphragm mobility and moderate reliability for left hemidiaphragm in the first assessment. In the second assessment, there was good reliability for the mobility of both hemidiaphragms. There was good intra-rater reliability in the mobility of both hemidiaphragms for raters A and B.

Conclusion

The evaluation of diaphragmatic motion using fluoroscopy by digital radiography proved to be a valid and reliable method of healthy adults.
Abkürzungen
1st
First evaluation
2nd
Second evaluation
AP
Anteroposterior incidence
BMI
Body mass index
Bpm
Beats per minute
CI
Confidence interval
DMdig
Diaphragmatic mobility digital
DMdist
Diaphragmatic mobility by distance
FEV1 (%predicted)
Estimated percentage of forced expiratory volume in the first second
FEV1
Forced expiratory volume in the first second
FEV1/FVC
Ratio between forced expiratory volume in the first second and forced vital capacity
FVC (%predicted)
Estimated percentage of forced vital capacity
FVC
Forced vital capacity
HR
Heart rate
kg
Kilograms
L
Litre
LH
Left hemidiaphragms
LL
Lower limit
M
Meter
mm
Millimeter
RH
Right hemidiaphragms
RV
Residual volume
SpO2
Oxygen saturation
SVC
Slow vital capacity
TLC
Total lung capacity
UL
Upper limit
ZCC
Intraclass correlation coeficiente

Background

Specifically evaluating the mobility of the diaphragm is important for understanding and diagnosing possible alterations in the muscle, which can be compromised in several ways: due to central or peripheral nervous system dysfunction, muscular disease and thoracic or abdominal disease, resulting in a reduction of mobility or paralysis [16].
In clinical practice it is essential to use valid and reliable methods for assessing diaphragmatic dysfunction, because the use of subjective methods may compromise the results. Therefore, it is extremely important that any assessment method have its validity and reliability tested to ensure that the error in the measurement is reduced [79].
There are several imaging methods that assess diaphragmatic mobility: fluoroscopy [1013], ultrasound [14, 15], computed tomography [16], magnetic resonance [1719], and chest radiography [20, 21]. Each technique has its particularities in the observation of the diaphragm, considering cost, radiation exposure and method availability in the study environment [5, 22].
Considering all the methods for evaluating diaphragmatic mobility, the fluoroscopy is the most accurade method for assessing the diaphragm muscle because it provides dynamic images of the diaphragm and direct visualization of diaphragmatic movements in real time [5]. However, there are no studies confirming the validity and reliability of digital radiography fluoroscopy to assess diaphragmatic excursion.
Diaphragmatic mobility can be measured by the fluoroscopy method in various ways, but those forms reported in the literature are not simple to be obtained [1013]. Some measures require radiographic impression for analysis [13], others, those recorded on video [10, 11], are not always available on fluoroscopy devices, and in other measurements, image calculations involve several complex lines for obtaining the value of diaphragmatic mobility [13].
Since fluoroscopy assesses diaphragm motion in real time [5], and the ways reported in the literature for measuring diaphragm mobility by means of fluoroscopy are complex [1013], and there is no literature studies investigating its validity and reliability, we propose creating a new method and a new measurement procedure that is much more easily obtained in clinical practice, using fluoroscopy by digital radiography.
Thus, the aim of this study was to assess the validity and reliability (inter- and intra-rate) of a new method of evaluation of diaphragmatic mobility using fluoroscopy by digital radiography.

Methods

Sample

In this study, 26 apparently healthy adults were included in a convenience sample. They were recruited among students and employees of the Universidade do Estado de Santa Catarina (Brazil), as well as their relatives. The sample size calculation was based on Bonett [23] and Fleiss [24] studies. According to these authors, to assess test reliability, a sample can vary between 15 and 20 participants.
Inclusion criteria for the study were: non-smoking healthy subjects with normal pulmonary function (FVC and FEV1 ≥ 80% predicted and FEV1/FVC ≥ 0.7) without cardiorespiratory or neurological diseases, women who were not pregnant or with suspected pregnancy, participants without a diagnosis of cancer, or disease history or any other alteration that could impair the evaluations. Exclusion criteria were: participants who presented clinical complications of respiratory nature, inability to perform any of the procedures used in the study (lack of understanding or collaboration), clinical complications of the respiratory system and those who requested exclusion from the study. This study was approved (16696413.8.0000.0118) by the Research Ethics Committee of the Universidade do Estado de Santa Catarina (UDESC), Brazil, and all participants provided written informed consent.

Study design

This is a cross-sectional study, with validity and reliability test, which assessed the agreement degree of diaphragmatic mobility by means of X-ray digital fluoroscopy [25].
Initially, anthropometric and pulmonary parameters were evaluated. Following, digital X-rays were scheduled in order to measure diaphragmatic excursion. Before the fluoroscopy examination, training of the diaphragm through diaphragmatic breathing exercise was performed, slow vital capacity (SVC) was also measured before and during the examination.
The examination of diaphragmatic motion by digital radiography fluoroscopy was randomly performed by two radiologists (raters A and B), and subsequently digital diaphragmatic mobility (DMdig) was measured by both raters, aiming to evaluate the intra-rater and inter-rater reliability. Diaphragmatic mobility by distance (DMdist) [20] was measured by rater A, to evaluate the validity of the method.

Collection procedures

Anthropometry

For measurement of body mass and height, a previously calibrated scale and a stadiometer (Welmy® model W200/5) were used respectively. Once the anthropometric values were obtained (weight and height), the body mass index (BMI) was calculated using the equation: body mass/(height) 2 (kg/m2). Subjects were classified according to BMI as underweight (≤18.5 kg/m2), normal (18.5–24.9 kg/m2), overweight (25–29.9 kg/m2) and obese (≥ 30 kg/m2) [26].

Pulmonary function test

The pulmonary function test was performed using a previously calibrated spirometer in accordance with the methods and criteria recommended by the American Thoracic Society [27]. For assessment of forced vital capacity (FVC), forced expiratory volume in the first second (FEV1), and the FEV1/FVC ratio, a portable digital EasyOne® spirometer of the ndd brand was used. The criteria for normal lung test consisted in FVC and FEV1 ≥ 80% predicted and FEV1/FVC ≥ 0.7. Spirometric variables were expressed as absolute values and as percentages of predicted normal values, according to Pereira et al. [28].
Before performing the pulmonary function test, pulse oxygen saturation (SpO2) and heart rate (HR) were measured with the participant in the supine position and at rest with a pulse oximeter (Oximeter, Model MD300C11).

Assessment of diaphragmatic mobility by digital radiography fluoroscopy

Diaphragmatic mobility was assessed through examination of digital X-ray fluoroscopy in anteroposterior incidence (AP) by two raters. To perform the test, a Siemens fluoroscopy device, model Lumino RF Classic was used at a distance of 1.15 m from the image intensifier and X-ray tube.
Subjects were placed on a radioscopic table in a supine position with their feet supported to restrict their movement on the table, and a radiopaque graduation ruler was placed under the subjects’ trunks in a longitudinal direction and in the craniocaudal direction. Before performing the fluoroscopy by digital radiography to evaluate the diaphragmatic motion, training in diaphragmatic breathing was provided. The goal was to develop diaphragmatic proprioception movement and enable the evaluation of diaphragm maximum amplitude during fluoroscopic digital X-ray examination. Afterwards, we measured slow vital capacity (SVC) using a Wright Respirometer Brit.® UK ventilometer, before and during the examination of digital X-ray fluoroscopy with subjects in a supine position. Three manoeuvers were performed before radiographic exposure, and the highest value was recorded for later comparison with what would be assessed during the examination of diaphragmatic mobility. During the examination of digital X-ray fluoroscopy, each rater asked the subjects, while exhaling, to breathe using TLC (total lung capacity) until they approached RV (residual volume), and then, upon exhaling, to breathe from RV until approaching TLC. The values of the SVC manoeuvers obtained were compared to each other (before and during the exam), to determine whether the subjects performed the same respiratory effort before and during the evaluation of diaphragmatic mobility. If there was a difference greater than 10% from the previously obtained value, the examination would be repeated on the same day.
Subjects were evaluated randomly, in a simple raffle, by two expert radiologists (raters A and B) who guided subjects in a standardized way. The raters viewed the diaphragm movement through the display on the fluoroscopy device, which was positioned on the central part of the thorax, viewing both hemidiaphragms at the same time. The images of maximum expiration and inspiration were recorded on the same film, which remained motionless during exams.

Digital measurement of diaphragmatic mobility (DMdig)

Diaphragmatic mobility (DMdig) was measured by calculating the distance between the diaphragmatic dome in expiration and inspiration for the right and left hemidiaphragms, based on the method described by Saltiel et al. [20]. Initially, the line of MediWorks 8.4.215 software was calibrated using a radiopaque ruler to correct the magnification caused by the divergence of the rays. This calibration was performed by drawing a line with the computer cursor over a distance of 10 mm on the image of the ruler in the digital radiography, thus determining the actual distance. Then, to measure diaphragmatic mobility, the highest point on the hemidiaphragm in expiration was found and a straight perpendicular line was drawn with the cursor until it met the hemidiaphragm in inspiration finding the distance between the diaphragmatic domes (Fig. 1). This measurement was performed for both right and left hemidiaphragms (RH) and (LH).

Measurement of diaphragmatic mobility by distance (DMdist)

To measure DMdist, rater A identified, on the printed chest radiography, the highest point of the diaphragm during expiration of each hemidiaphragmatic dome, and from this point a line was drawn with a black marker until the hemidiaphragm during inspiration was found. The diaphragm mobility was determined by the distance between the diaphragm dome on expiration and inspiration by the calliper, both on the right side and the left side. To correct the image magnification caused by the divergence of X-rays, a correction formula was used: Corrected mobility (mm) = mobility measure (mm) × 10/graduation on the ruler (mm) [20].

Analysis of the validity

To assess the validity of the method (criterion validity), we analysed the concurrent validity. The competing method was the method of Saltiel et al. [20], which evaluates diaphragm mobility on the printed radiograph by distance (DMdist) by means of a calliper. Validity was assessed by relating the first measurement of DMdig obtained by rater A with the measurement using the DMdist method by the same rater. We evaluated both right and left hemidiaphragms.

Analysis of reliability

Measurements of diaphragmatic mobility were assessed soon after scanning the radiograph. The first measurement was used for the analysis of inter-rater reliability of the method (measurements A1 and B1) in the following manner: a measurement taken by rater A (A1) was correlated with that taken by rater B (B1). Then, a second measurement was performed for the analysis of inter-rater reliability of the measure as follows: rater A measured the digital radiography fluoroscopy exam performed by rater B, obtaining measure AB, and rater B measured the digital radiography fluoroscopy exam performed by rater A obtaining measure BA. The analysis was performed by correlating measures AB with B1 and BA with A1.
The intra-rater reliability of the measurement was assessed by measuring diaphragmatic mobility from the previous examination, after a minimum interval of 7 days and maximum of 20 days from the first measurement25. Both raters A and B measured the digital radiographs one more time (measures A2 and B2) performed at the beginning of the study. Inter-rater reliability was also examined in the second assessment.
Raters A and B did not know who conducted the radiographs and did not have access to the values of the other’s assessments. The results were analysed after completion of all assessments.

Statistical analysis

Data were analysed using SPSS for Windows, version 20.0 (IBM SPSS Statistics, IBM, Armonk, NY, USA) and GraphPad Prism 5.1 program and treated with descriptive analysis (mean and standard deviation) and inferential analysis. Shapiro-Wilk test was used to analyse data normality.
Pearson correlation coefficient and intraclass correlation coefficient (two-way random model, with absolute agreement - ICC[2.1]) were used to assess the correlation between the digital method (DMdig) and the method of distance (DMdist.). The analyses of inter-rater and intra-rater reproducibility were determined using intraclass correlation coefficient (two-way random model, with absolute agreement - ICC[2.1]) and confidence interval (CI) of 95% [8]. Reliability was interpreted as the magnitude of Portney and Watkins coefficient of reliability [8]: ‘poor’ for coefficients under 0.50, ‘moderate’ for coefficients between 0.50 and 0.75, and ‘good’ for coefficients above 0.75. ICC varies from 0.00 to 1.00, and values close to 1.00 show strong reliability. Bland-Altman plot was also used to allow better visualization of agreement between the individual measures [29].
Wilcoxon test was used to compare the values of slow vital capacity before and during radiographic exposures for each rater. Paired T test was used to compare the mobility of the right and left hemidiaphragms. The T test for independent samples was used to compare the values of diaphragmatic mobility between male and female participants. The significance level for statistical treatment was 5% (p < 0.05).

Results

Twenty-six healthy adults, 17 women (65.4%) and 9 men (34.6%) were evaluated, with a mean age of 28.19 ± 6.1 years, mean BMI 23.89 ± 4.2, classified as normal, healthy and with normal pulmonary function (Table 1). No subject refused to participate or desisted during assessments.
Table 1
Anthropometric and cardiopulmonary characteristics of the study participants
Variables
Average ± standard deviation variables (n = 26)
Age (years)
28.19 ± 6.1
Body mass (kg)
68.14 ± 16.7
Height (m)
1.68 ± 0.1
BMI (kg.m- 2)
23.89 ± 4.2
HR (bpm)
72.61 ± 9.7
SpO2 (%)
98.35 ± 0.6
FVC (% predicted)
92.85 ± 7.7
FEV1 (% predicted)
94.73 ± 7.1
FEV1/FVC (L)
0.91 ± 0.2
Values were express as mean and standard deviation
n number of subjetcs, kg lbs, m meters, BMI body mass index, HR heart rate, bpm: beats per minute, SpO2 oxygen saturation by pulse, FVC (%predicted): Estimated percentage of FVC, FEV1 (%predicted): Estimated percentage of forced expiratory volume in one second, L liters
A high correlation was found between DMdig and DMdist for the mobility of the right hemidiaphragm (RH) (r = 0.97, p = 0.00) and the left one (LH) (r = 0.88, p = 0. 00). There was good reliability for mobility in both hemidiaphragms (RH: ICC[1, 2] = 0.98, 95% CI = 0.96 to 0.99; LH: ICC[2,1] = 0.93, 95% CI = 0.84 to 0.97) (Fig. 2).
In the inter-raters analysis, in the first assessment there was good reliability for RH and moderate in LH. In the second assessment, there was good reliability for mobility in both hemidiaphragms. There was good intra-rater reliability of mobility in both RH and LH for rater A. Similar results were found in the measurements obtained by rater B in assessing RH and LH. The analyses of inter-rater and intra-rater reliability for digital diaphragmatic mobility of RH and LH are described in Table 2.
Table 2
Inter-rater and intra-rater reliability of the mobility measurement of the right and left hemidiaphragms method
Variables
ICC[2,1]
CI 95%
Right hemidiaphragm
 Inter-rater reliability
1ª assess
0.89
0.76–0.95
2ª assess
0.84
0.68–0.93
 Intra-rater reliability
rater A
0.83
0.66–0.92
rater B
0.89
0.76–0.95
Left hemidiaphragm
 Inter-rater reliability
1ª assess
0.73
0.48–0.87
2ª assess
0.78
0.56–0.89
 Intra-rater reliability
rater A
0.86
0.70–0.93
rater B
0.83
0.65–0.92
ICC [2,1] the intraclass correlation coefficient (two-way random model, with absolute and agreement), CI 95% confidence interval of 95%
To demonstrate higher reliability, we also evaluated inter-rater reproducibility of the measurement, where rater A measured the examination performed by rater B, yielding measure AB and we compared this with measure B1. For this analysis there was good reliability for the ratings of RH and LH (ICC[2,1] = 0.98, 95% CI = 0.96 to 0.99; ICC[2,1] = 0.95, 95% CI = 0.90 to 0.98, respectively). Rater B also measured the examination performed by rater A to obtain measure BA. When comparing it with measure A1, there was good reliability for RH (ICC[2,1] = 0.98, 95% CI = 0.96 to 0.99) and for LH (ICC[2,1] = 0.86, 95% CI = 0.72 to 0.94).
According to Bland-Altman plot, there was good concordance between measures of mobility of both RH and LH, obtained by raters A and B (inter-rater agreement) (Fig. 3), and good concordance between measures of mobility of RH and LH, obtained by each of the raters (Fig. 4), at two different times (intra-rater agreement). Figure 5 shows that there was good concordance between measures of mobility of RH and LH, obtained by raters A and B when rater A measured the test conducted by rater B and when rater B measured the test conducted by rater A. There was good concordance because the difference between the measures is the limits of agreement (upper and lower limits). The mean of the measures between raters in all analyses is close to zero, which indicates reproducibility of the measurements. However, when analysing the inter-rater graphics, there was greater dispersion of the means obtained.
There were no differences between the manoeuvers of slow vital capacity (SVC) performed before and during fluoroscopy examinations by radiography by raters A and B, respectively (before: 3.92 ± 1.5 mm; during: 4.07 ± 1.5 mm, p = 0.57; before: 3.92 ± 1.5 mm; during : 4.12 ± 1.4 mm, p = 0.46).
The mobility measured in the right (RH) and left (LH) hemidiaphragms by the two raters (A and B) showed no statistically significant difference. The mean values of RH and LH mobility when analysed by rater A were 64.8 ± 12.6 mm (30.3 to 96.7 mm) and 64.1 ± 10.8 mm (31.7 to 92.3 mm), respectively and when analysed by rater B, 64.7 ± 12.5 mm (26.1 to 83.3 mm) and 62.9 ± 11.3 mm (27.0 to 81.5 mm), respectively. There was no statistically significant difference in the mobility of RH and LH measured by the two raters: A (p = 0.69) and B (p = 0.41).

Discussion

This study demonstrated that the digital X-ray fluoroscopy method is valid and reliable for measuring mobility of the left and right hemidiaphragms in healthy adults. In clinical practice, the use of reliable instruments is essential for ensuring reliable results [9].
Fluoroscopy is the most accurate method for evaluating diaphragm dysfunction as it assesses the diaphragmatic motion in real time [5], but some measures require a video be made of the image [10, 11], which is not always available in fluoroscopy equipment, or they require calculations involving complex procedures to determine diaphragmatic motion [13].
Thus, the proposal of this study is innovative because it verifies the validity and the reliability of a new method of fluoroscopy and a new way of measuring diaphragmatic movement. For this, we applied a common resource used in clinical practice, which is the X-ray, and we innovated by using fluoroscopy associated with digital radiography. This method is easy to apply and measure, has a low cost because it is not necessary to print the radiography, and it can be another tool for professionals in the health assessment of patient’s diaphragmatic mobility.
Our proposal was to perform a digital measurement using software that is routinely used in medical radiological practice, but this is a unique way for evaluating diaphragmatic mobility in the scientific community. The measurement of diaphragmatic motion using the scanned image (DMdig) is simple and practical to perform, and proved to be reliable. We emphasize that the digitization of the exam is a technology that generates practicality, quality in analysis and makes storage easier in examinations, requiring no physical space. Moreover, it enables sharing among several professionals for the consultation and discussion of clinical cases.
This study found good reliability in the method analyses of diaphragmatic mobility for both raters A and B. There was good inter-rater reliability for the analysis of measure for RH and LH. Thus, we find that digital X-ray fluoroscopy is a reliable method for measuring diaphragmatic mobility, because a measure is considered reliable if the ICC is greater than 0.75 [8].
In the inter-rater analysis of the Bland-Altman plot in Fig. 3, a greater dispersion of the measures was observed. However, in Fig. 5 when rater A measured the test conducted by rater B and when rater B measured the test conducted by rater A, there was lower dispersion of data. Possibly, this difference in dispersion of data between the two analyses was due to the fact that, in the first measurement (Fig. 3), the rater performed the test and also assessed the measure.
In Fig. 5 the raters only had to perform the digital measurement of diaphragmatic mobility (DMdig) in the previously undertaken test. Therefore, we suggest that the same rater perform the measurement of diaphragmatic motion when comparing pre- and post-treatment, ensuring a lower bias between measurements, as our results show lower dispersion of data in intra-rater reliability.
When we observe the correlation between measures in the Bland-Altman plots (Figs. 3, 4 and 5), there was better agreement for the RH. This difference may have happened due to the proximity of the LH to the heart. Considering that the heart is a dynamic organ with involuntary contraction, at the time of the expiratory pause, a cardiac muscle contraction may have occurred during the evaluation by one of the raters, thereby altering the position of the left hemidiaphragm and consequently the location of the highest expiration point. However, it is important to note that the evaluation method of diaphragmatic mobility is reliable to assess both right and left hemidiaphragms, but the rater should choose the RH, if possible, because the agreement was better.
This study found a high variability in the amount of diaphragmatic motion (from 30.3 to 96.7 mm), similar to other studies [14, 15, 21]. Kantarci et al. [30] found variability of 25 to 84 mm for RH, and of 24 to 81 mm for LH, while in Gerscovich et al. [4], variability from 16.7 to 92 mm was found for RH, and 38 to 96 mm for LH. This variability may be related to BMI and age, because according to Kantarci et al. [30], subjects with low weight and younger than 30 have lower diaphragmatic mobility.
In this study there was no difference between diaphragmatic mobility for RH and LH, a result similar to data obtained in other studies by our group [20]. We also found no difference in diaphragmatic excursion between men and women. The studies by Bousseges et al. [15] and Kantarci et al. [30] found differences between diaphragmatic mobility between men and women. However, Grams et al. [14] and Saltiel et al. [20] found no such differences. This is possibly due to the sample size because the studies by Bousseges et al. [15] and Kantarci et al. [30] had a large number of participants (210 and 164 subjects, respectively), whereas Grams et al. [14] and Saltiel et al. [20] had a smaller number of about 40 subjects.
The mean values of RH mobility were 64.90 ± 12.9 mm, and 63.95 ± 11.7 mm for LH. According to Simon et al. [21], most healthy adults have diaphragm mobility ≥ 30 mm. Other studies have found amounts similar to those in our research of diaphragmatic mobility in healthy adults [12, 14]. There are, however, no reference values for diaphragmatic mobility in the literature.
When performing an examination, it is relevant to consider the acceptability and repeatability of the measurement. In the evaluation of diaphragmatic mobility by fluoroscopy, digital radiography is possible by analysing the acceptability of the examination to verify SVL; however, due to radiation it is not feasible to evaluate the repeatability of the measure. Despite this methodological limitation, it is possible to ensure the maximum movement of the diaphragm through the standardization of the evaluator’s guidelines, previous diaphragmatic training, and by obtaining the same SVL during the test.
In clinical practice, a change in diaphragm movement is often found, with a reduction in mobility or muscle paralysis, which may be due to several factors such as: changes in the muscle itself or neuromuscular conduction, dysfunction in the central nervous system, muscular dystrophy, muscle injury by trauma, phrenic nerve injury, chronic obstructive pulmonary disease, thoracic or abdominal diseases that interfere with their mobility, such as atelectasis or pleural effusions. The study of evaluation methods for diaphragmatic motion is important to better detect and diagnose diaphragm dysfunction [25]. A limitation of our study was to evaluate diaphragmatic mobility only in healthy subjects. Therefore, new studies for the validation of the method in patient populations would be relevant, aiming at creating a classification scale of diaphragm dysfunction severity.
Since the diaphragm is the most important respiratory muscle for pulmonary ventilation, it is essential to understand the functional condition of this muscle and its possible changes to establish therapeutic strategies to restore or improve its mobility and thus provide improved functionality and quality of life for the patient [13, 6].

Conclusion

The results showed that fluoroscopy by digital radiography method is a valid and reliable instrument to assess the mobility of right and left hemidiaphragms of healthy adults.

Acknowledgments

We appreciate the partnership with the clinic Lâmina Medicina Diagnóstica.
Study carried out in the Laboratório de Fisioterapia Respiratória (LAFIR), Universidade do Estado de Santa Catarina (UDESC), Florianópolis - SC, Brazil.

Funding

Not applicable.

Availability of data and materials

Due to the privacy policy of patients, we can’t openly release the dataset of questionnaire survey underlying the conclusions of the paper available in public. A truncated dataset after eliminating all potentially identifiable features may be provided on an individual request basis.

Authors’ contributions

This study presents a new tool to evaluate the diaphragmatic mobility that has proved to be valid and reproducible. We believe it will be widely used in clinical practice, for the ease and speed of acquisition of the measure. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.
No patient identifiable material is presented.
The study was approved by the Research Ethics Committee of the Universidade do Estado de Santa Catarina (UDESC), Brazil, and all subjects provided written informed consent.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
4.
Zurück zum Zitat Gerscovich EO, Cronan M, McGahan JP, Jain K, Jones CD, McDonald C. Ultrasonographic evaluation of diaphragmatic motion. J Ultrasound Med. 2001;20(6):597–604.CrossRefPubMed Gerscovich EO, Cronan M, McGahan JP, Jain K, Jones CD, McDonald C. Ultrasonographic evaluation of diaphragmatic motion. J Ultrasound Med. 2001;20(6):597–604.CrossRefPubMed
5.
Zurück zum Zitat Gierada DS, Slone RM, Fleishman MJ. Imaging evaluation of the diaphragm. Chest SurgClin N Am. 1998;8(2):237–80. Gierada DS, Slone RM, Fleishman MJ. Imaging evaluation of the diaphragm. Chest SurgClin N Am. 1998;8(2):237–80.
6.
Zurück zum Zitat Reid WD, Dechman G. Considerations when testing and training the respiratory muscles. PhysTher. 1995;75(11):971–82. Reid WD, Dechman G. Considerations when testing and training the respiratory muscles. PhysTher. 1995;75(11):971–82.
7.
Zurück zum Zitat Gadotti IC, Vieira ER, Magee DJ. Importance and clarification of measurement properties in rehabilitation. Rev bras Fisioter. 2006;10(2):137–46.CrossRef Gadotti IC, Vieira ER, Magee DJ. Importance and clarification of measurement properties in rehabilitation. Rev bras Fisioter. 2006;10(2):137–46.CrossRef
8.
Zurück zum Zitat Portney GL, Watkins PM. Reliability. In: Portney GL, Watkins PM, editors. Foundations of clinical research application to pratice. New Jersey: Pearson Prentice Hall; 2008. p. 61–75. Portney GL, Watkins PM. Reliability. In: Portney GL, Watkins PM, editors. Foundations of clinical research application to pratice. New Jersey: Pearson Prentice Hall; 2008. p. 61–75.
10.
Zurück zum Zitat Yi LC, Nascimento OA, Jardim JR. Reliability of an analysis method for measuring diaphragm excursion by means of direct visualization with videofluoroscopy. ArchBronconeumol. 2011;47(6):310–4. Yi LC, Nascimento OA, Jardim JR. Reliability of an analysis method for measuring diaphragm excursion by means of direct visualization with videofluoroscopy. ArchBronconeumol. 2011;47(6):310–4.
11.
Zurück zum Zitat Kleinman B, Frey K, VanDrunen M, Sheikh T, DiPinto D, Mason R, Smith T. Motion of the diaphragm in patients with chronic obstructive pulmonary disease while spontaneously breathing versus during positive pressure breathing after anesthesia and neuromuscular blockade. Anestesiology. 2002;97:298–305.CrossRef Kleinman B, Frey K, VanDrunen M, Sheikh T, DiPinto D, Mason R, Smith T. Motion of the diaphragm in patients with chronic obstructive pulmonary disease while spontaneously breathing versus during positive pressure breathing after anesthesia and neuromuscular blockade. Anestesiology. 2002;97:298–305.CrossRef
12.
Zurück zum Zitat Unal O, Arslan H, Uzum K, Osbay B, Sakaria ME. Evaluation of diaphragmatic movement with MR fluoroscopy in chronic obstructive pulmonary disease. J Clin Imaging. 2000;24:347–50.CrossRef Unal O, Arslan H, Uzum K, Osbay B, Sakaria ME. Evaluation of diaphragmatic movement with MR fluoroscopy in chronic obstructive pulmonary disease. J Clin Imaging. 2000;24:347–50.CrossRef
13.
Zurück zum Zitat Verschakelen JA, Deschepper K, Jiang TX, Demedts M. Diaphragmatic displacement measured by fluoroscopy and derived by respitrace. J Appl Physiol. 1989;67(2):694–8.PubMed Verschakelen JA, Deschepper K, Jiang TX, Demedts M. Diaphragmatic displacement measured by fluoroscopy and derived by respitrace. J Appl Physiol. 1989;67(2):694–8.PubMed
14.
Zurück zum Zitat Grams ST, Saltiel RV, Pedrini A, Mayer AF, Schivinski CIS, Paulin E. Assesment of the reproducibility of the indirect ultrasound method of measuring diaphragm mobility. Clin Physiol Funct Imaging. 2014;34:18–25.CrossRefPubMed Grams ST, Saltiel RV, Pedrini A, Mayer AF, Schivinski CIS, Paulin E. Assesment of the reproducibility of the indirect ultrasound method of measuring diaphragm mobility. Clin Physiol Funct Imaging. 2014;34:18–25.CrossRefPubMed
16.
Zurück zum Zitat Leung JC, Nance ML, Schwab CW, Miller Jr WT. Thickening of the diaphragm: a new computed tomography sign of diaphragm injury. J Thorac Imaging. 1999;14(2):126–9.CrossRefPubMed Leung JC, Nance ML, Schwab CW, Miller Jr WT. Thickening of the diaphragm: a new computed tomography sign of diaphragm injury. J Thorac Imaging. 1999;14(2):126–9.CrossRefPubMed
17.
Zurück zum Zitat Iwasawa T, Yoshiike Y, Saito K, Kagei S, Gotoh T, Matsubara S. Paradoxical motion of the hemidiaphragm in patients with emphysema. J Thorac Imaging. 2000;15(3):191–5.CrossRefPubMed Iwasawa T, Yoshiike Y, Saito K, Kagei S, Gotoh T, Matsubara S. Paradoxical motion of the hemidiaphragm in patients with emphysema. J Thorac Imaging. 2000;15(3):191–5.CrossRefPubMed
18.
Zurück zum Zitat Plathow C, Ley S, Fink C, Puderbach M, Heilmann M, Zuna I, Kauczor HU. Evaluation of chest motion and volumetry during the breathing cycle by dynamic MRI in healthy subjects: comparison with pulmonary function tests. Investig Radiol. 2004;39(4):202–9.CrossRef Plathow C, Ley S, Fink C, Puderbach M, Heilmann M, Zuna I, Kauczor HU. Evaluation of chest motion and volumetry during the breathing cycle by dynamic MRI in healthy subjects: comparison with pulmonary function tests. Investig Radiol. 2004;39(4):202–9.CrossRef
19.
Zurück zum Zitat Kotani T, Minami S, Takahashi K, Isobe K, Nakata Y, Takaso M, Inoue M, Maruta T. An analysis of chest wall and diaphragm motions in patients with idiopathic scoliosis using dynamic breathing MRI. Revista Spine. 2004;29(3):298–302.CrossRef Kotani T, Minami S, Takahashi K, Isobe K, Nakata Y, Takaso M, Inoue M, Maruta T. An analysis of chest wall and diaphragm motions in patients with idiopathic scoliosis using dynamic breathing MRI. Revista Spine. 2004;29(3):298–302.CrossRef
20.
Zurück zum Zitat Saltiel RV, Grams ST, Pedrini A, Paulin E. High reliability of measure of diaphragmatic mobility by radiographic method in healthy individuals. Braz J PhysTher. 2013;17(2):128–36. doi: 10.1590/S1413-35552012005000076. Saltiel RV, Grams ST, Pedrini A, Paulin E. High reliability of measure of diaphragmatic mobility by radiographic method in healthy individuals. Braz J PhysTher. 2013;17(2):128–36. doi: 10.​1590/​S1413-3555201200500007​6.
23.
Zurück zum Zitat Bonett DG. Sample size requirements for testing and estimating coefficient alpha. J Educ Behav Stat Winter. 2002;27(4):335–40.CrossRef Bonett DG. Sample size requirements for testing and estimating coefficient alpha. J Educ Behav Stat Winter. 2002;27(4):335–40.CrossRef
24.
Zurück zum Zitat Fleiss JL. Design and analysis of clinical experiments. New York: Wiley; 1999. p. 2–27.CrossRef Fleiss JL. Design and analysis of clinical experiments. New York: Wiley; 1999. p. 2–27.CrossRef
25.
Zurück zum Zitat Hulley SB. Delineando a pesquisa clínica: uma abordagem epidemiológica. Porto Alegre: Artmed; 2008. p. 23–4. Hulley SB. Delineando a pesquisa clínica: uma abordagem epidemiológica. Porto Alegre: Artmed; 2008. p. 23–4.
26.
Zurück zum Zitat World Health Organization. WHO obesity technical report series. obesity: preventing and managing the global epidemic. report of a world health organization consultation. Geneva: World Health Organization; 2000. p. 284–56. World Health Organization. WHO obesity technical report series. obesity: preventing and managing the global epidemic. report of a world health organization consultation. Geneva: World Health Organization; 2000. p. 284–56.
30.
Zurück zum Zitat Kantarci F, Mihmanli I, DemireL MK, Harmanci K, Akman C, Aydogan F, et al. Normal diaphragmatic motion and the effects of body composition: determination with m-mode sonography. J Ultrasound Med. 2004;23(2):255–60.CrossRefPubMed Kantarci F, Mihmanli I, DemireL MK, Harmanci K, Akman C, Aydogan F, et al. Normal diaphragmatic motion and the effects of body composition: determination with m-mode sonography. J Ultrasound Med. 2004;23(2):255–60.CrossRefPubMed
Metadaten
Titel
Validity and reliability of fluoroscopy for digital radiography: a new way to evaluate diaphragmatic mobility
verfasst von
Bruna Estima Leal
Márcia Aparecida Gonçalves
Liseane Gonçalves Lisboa
Larissa Martins Schmitz Linné
Michelle Gonçalves de Souza Tavares
Wellington Pereira Yamaguti
Elaine Paulin
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
BMC Pulmonary Medicine / Ausgabe 1/2017
Elektronische ISSN: 1471-2466
DOI
https://doi.org/10.1186/s12890-017-0402-x

Weitere Artikel der Ausgabe 1/2017

BMC Pulmonary Medicine 1/2017 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.