What is the optimal valgus pre-set for intramedullary femoral alignment rods in total knee arthroplasty?

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 11/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

In total knee arthroplasty (TKA), intramedullary guides are often used for aligning the distal femoral cutting block. Because of the highly varying angles between the mechanical axis and the anatomical femoral axis (AMA), different valgus pre-sets have been recommended. The present study investigated the optimal valgus pre-set (measuring the AMA in long-leg radiographs or at 5°, 6°, 7° or 8° valgus) to align the cutting block perpendicularly to the mechanical axis.

Methods

The AMA was preoperatively measured in weight-bearing long-leg radiographs. After alignment of the cutting block by means of an intramedullary rod, deviation of the block from the mechanical femoral axis was measured with a pinless navigation device. The true AMA (tAMA) was calculated by adding the valgus pre-set of the alignment rod to the deviation measured with the navigation device. Mean deviations between the tAMA and (a) the AMA measured by the surgeon, (b) the AMA calculated with the computer software, (c) 5°, (d) 6°, (e) 7° and (f) 8° valgus pre-sets were measured for each patient. The lowest mean differences were determined.

Results

The 40 knees measured showed a mean tAMA of 7.2° valgus (1.7 SD) (range 4°–11.5°). The following mean differences and 95 % limits of agreement were calculated: 2.2 (−1.2, 5.5) to the tAMA for the 5° valgus pre-set, 1.2 (−2.2, 4.5) for 6°, 0.2 (−3.2, 3.5) for 7° and −0.8 (−4.2, 2.5) for 8°. AMA measurements by the surgeon and with the digital medical planning software yielded mean differences of 0.6 (−3.1, 4.3) and 0.4 (−4.1, 4.8), respectively.

Conclusion

In the present setting, the best mean distal femoral cutting block alignment perpendicular to the mechanical femoral axis could be achieved with a valgus pre-set of 7° and not by measuring the AMA. Nevertheless, we recommend conducting weight-bearing radiographs of the entire leg prior to TKA for easy detection of any anatomical varieties, old fractures, long stems of total hip arthroplasties or cement. However, surgeons must be aware that exact coronal component alignment can only be achieved by navigational devices.

Level of evidence

Diagnostic study, Level II.
Literatur
1.
Zurück zum Zitat Baier C, Maderbacher G, Springorum HR, Zeman F, Fitz W, Schaumburger J, Grifka J, Beckmann J (2014) No difference in accuracy between pinless and conventional computer-assisted surgery in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 22(8):1819–1826CrossRefPubMed Baier C, Maderbacher G, Springorum HR, Zeman F, Fitz W, Schaumburger J, Grifka J, Beckmann J (2014) No difference in accuracy between pinless and conventional computer-assisted surgery in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 22(8):1819–1826CrossRefPubMed
2.
Zurück zum Zitat Fang DM, Ritter MA, Davis KE (2009) Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty 24:39–43CrossRefPubMed Fang DM, Ritter MA, Davis KE (2009) Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty 24:39–43CrossRefPubMed
3.
Zurück zum Zitat Harding IJ, Crawford RW, Mclardy-Smith P, Murray DW (1999) The importance of femoral intramedullary entry point in knee arthroplasty. Knee 6:207–210CrossRef Harding IJ, Crawford RW, Mclardy-Smith P, Murray DW (1999) The importance of femoral intramedullary entry point in knee arthroplasty. Knee 6:207–210CrossRef
4.
Zurück zum Zitat Huang NF, Dowsey MM, Ee E, Stoney JD, Babazadeh S, Choong PF (2012) Coronal alignment correlates with outcome after total knee arthroplasty: five-year follow-up of a randomized controlled trial. J Arthroplasty 27(9):1737–1741CrossRefPubMed Huang NF, Dowsey MM, Ee E, Stoney JD, Babazadeh S, Choong PF (2012) Coronal alignment correlates with outcome after total knee arthroplasty: five-year follow-up of a randomized controlled trial. J Arthroplasty 27(9):1737–1741CrossRefPubMed
5.
Zurück zum Zitat Huang TW, Hsu WH, Peng KT, Hsu RW (2011) Total knee replacement in patients with significant femoral bowing in the coronal plane: a comparison of conventional and computer-assisted surgery in an Asian population. J Bone Joint Surg Br 93(3):345–350CrossRefPubMed Huang TW, Hsu WH, Peng KT, Hsu RW (2011) Total knee replacement in patients with significant femoral bowing in the coronal plane: a comparison of conventional and computer-assisted surgery in an Asian population. J Bone Joint Surg Br 93(3):345–350CrossRefPubMed
6.
Zurück zum Zitat Jiang CC, Insall JN (1989) Effect of rotation on the axial alignment of the femur. Pitfalls in the use of femoral intramedullary guides in total knee arthroplasty. Clin Orthop Relat Res 248:50–56 Jiang CC, Insall JN (1989) Effect of rotation on the axial alignment of the femur. Pitfalls in the use of femoral intramedullary guides in total knee arthroplasty. Clin Orthop Relat Res 248:50–56
7.
Zurück zum Zitat Kim JM, Hong SH, Kim JM, Lee BS, Kim DE, Kim KA, Bin SI (2015) Femoral shaft bowing in the coronal plane has more significant effect on the coronal alignment of TKA than proximal or distal variations of femoral shape. Knee Surg Sports Traumatol Arthrosc 23(7):1936–1942CrossRefPubMed Kim JM, Hong SH, Kim JM, Lee BS, Kim DE, Kim KA, Bin SI (2015) Femoral shaft bowing in the coronal plane has more significant effect on the coronal alignment of TKA than proximal or distal variations of femoral shape. Knee Surg Sports Traumatol Arthrosc 23(7):1936–1942CrossRefPubMed
8.
Zurück zum Zitat Lotke PA, Ecker ML (1977) Influence of positioning of prosthesis in total knee replacement. J Bone Joint Surg 59A:77–79CrossRef Lotke PA, Ecker ML (1977) Influence of positioning of prosthesis in total knee replacement. J Bone Joint Surg 59A:77–79CrossRef
9.
Zurück zum Zitat Lonner JH, Laird MT, Stuchin SA (1996) Effect of rotation and knee flexion on radiographic alignment in total knee arthroplasties. Clin Orthop Relat Res 331:102–106CrossRef Lonner JH, Laird MT, Stuchin SA (1996) Effect of rotation and knee flexion on radiographic alignment in total knee arthroplasties. Clin Orthop Relat Res 331:102–106CrossRef
10.
Zurück zum Zitat Lustig S, Fleury C, Servien E, Demey G, Neyret P, Donell ST (2011) The effect of pelvic movement on the accuracy of hip centre location acquired using an imageless navigation system. Int Orthop 35(11):1605–1610CrossRefPubMedPubMedCentral Lustig S, Fleury C, Servien E, Demey G, Neyret P, Donell ST (2011) The effect of pelvic movement on the accuracy of hip centre location acquired using an imageless navigation system. Int Orthop 35(11):1605–1610CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Maderbacher G, Keshmiri A, Schaumburger J, Springorum HR, Zeman F, Grifka J, Baier C (2014) Accuracy of bony landmarks for restoring the natural joint line in revision knee surgery: an MRI study. Int Orthop 38:1173–1181CrossRefPubMedPubMedCentral Maderbacher G, Keshmiri A, Schaumburger J, Springorum HR, Zeman F, Grifka J, Baier C (2014) Accuracy of bony landmarks for restoring the natural joint line in revision knee surgery: an MRI study. Int Orthop 38:1173–1181CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Maderbacher G, Schaumburger J, Baier C, Zeman F, Springorum HR, Birkenbach AM, Grifka J, Keshmiri A (2015) Appropriate sagittal femoral component alignment cannot be ensured by intramedullary alignment rods. Knee Surg Sports Traumatol Arthrosc. doi:10.1007/s00167-015-3541-8 Maderbacher G, Schaumburger J, Baier C, Zeman F, Springorum HR, Birkenbach AM, Grifka J, Keshmiri A (2015) Appropriate sagittal femoral component alignment cannot be ensured by intramedullary alignment rods. Knee Surg Sports Traumatol Arthrosc. doi:10.​1007/​s00167-015-3541-8
13.
Zurück zum Zitat Maderbacher G, Schaumburger J, Baier C, Zeman F, Springorum HR, Dornia C, Grifka J, Keshmiri A (2014) Predicting knee rotation by the projection overlap of the proximal fibula and tibia in long-leg radiographs. Knee Surg Sports Traumatol Arthrosc 22(12):2982–2988CrossRefPubMed Maderbacher G, Schaumburger J, Baier C, Zeman F, Springorum HR, Dornia C, Grifka J, Keshmiri A (2014) Predicting knee rotation by the projection overlap of the proximal fibula and tibia in long-leg radiographs. Knee Surg Sports Traumatol Arthrosc 22(12):2982–2988CrossRefPubMed
14.
Zurück zum Zitat Maderbacher G, Schaumburger J, Keshmiri A, Barthel M, Springorum HR, Craiovan B, Grifka J, Baier C (2014) Pinless navigation in total knee arthroplasty—navigation reduced by the maximum? Int Orthop 39:455–460CrossRefPubMed Maderbacher G, Schaumburger J, Keshmiri A, Barthel M, Springorum HR, Craiovan B, Grifka J, Baier C (2014) Pinless navigation in total knee arthroplasty—navigation reduced by the maximum? Int Orthop 39:455–460CrossRefPubMed
15.
Zurück zum Zitat Maestro A, Harwin SF, Sandoval MG, Vaquero DH, Murcia A (1998) Influence of intramedullary versus extramedullary alignment guides on final total knee arthroplasty component position: a radiographic analysis. J Arthroplasty 13(5):552–558CrossRefPubMed Maestro A, Harwin SF, Sandoval MG, Vaquero DH, Murcia A (1998) Influence of intramedullary versus extramedullary alignment guides on final total knee arthroplasty component position: a radiographic analysis. J Arthroplasty 13(5):552–558CrossRefPubMed
16.
Zurück zum Zitat Mihalko WM, Boyle J, Clark LD, Krackow KA (2005) The variability of intramedullary alignment of the femoral component during total knee arthroplasty. J Arthroplasty 20(1):25–28CrossRefPubMed Mihalko WM, Boyle J, Clark LD, Krackow KA (2005) The variability of intramedullary alignment of the femoral component during total knee arthroplasty. J Arthroplasty 20(1):25–28CrossRefPubMed
17.
Zurück zum Zitat Moreland JR (1988) Mechanisms of failure in total knee arthroplasty. Clin Orthop Relat Res 226:49–64 Moreland JR (1988) Mechanisms of failure in total knee arthroplasty. Clin Orthop Relat Res 226:49–64
18.
Zurück zum Zitat Mont MA, Urquhart MA, Hungerford DS, Krackow KA (1997) Intramedullary goniometer can improve alignment in knee arthroplasty surgery. J Arthroplasty 12:332–336CrossRefPubMed Mont MA, Urquhart MA, Hungerford DS, Krackow KA (1997) Intramedullary goniometer can improve alignment in knee arthroplasty surgery. J Arthroplasty 12:332–336CrossRefPubMed
19.
Zurück zum Zitat Novotny J, Gonzalez MH, Amirouche FM, Li YC (2001) Geometric analysis of potential error in using femoral intramedullary guides in total knee arthroplasty. J Arthroplasty 16(5):641–647CrossRefPubMed Novotny J, Gonzalez MH, Amirouche FM, Li YC (2001) Geometric analysis of potential error in using femoral intramedullary guides in total knee arthroplasty. J Arthroplasty 16(5):641–647CrossRefPubMed
20.
Zurück zum Zitat Nuño-Siebrecht N, Tanzer M, Bobyn JD (2000) Potential errors in axial alignment using intramedullary instrumentation for total knee arthroplasty. J Arthroplasty 15(2):228–230CrossRefPubMed Nuño-Siebrecht N, Tanzer M, Bobyn JD (2000) Potential errors in axial alignment using intramedullary instrumentation for total knee arthroplasty. J Arthroplasty 15(2):228–230CrossRefPubMed
21.
Zurück zum Zitat Oswald MH, Jakob RP, Schneider E, Hoogewoud HM (1993) Radiological analysis of normal axial alignment of femur and tibia in view of total knee arthroplasty. J Arthroplasty 8(4):419–426CrossRefPubMed Oswald MH, Jakob RP, Schneider E, Hoogewoud HM (1993) Radiological analysis of normal axial alignment of femur and tibia in view of total knee arthroplasty. J Arthroplasty 8(4):419–426CrossRefPubMed
22.
Zurück zum Zitat Pitto RP, Graydon AJ, Bradley L, Malak SF, Walker CG, Anderson IA (2006) Accuracy of a computer-assisted navigation system for total knee replacement. J Bone Joint Surg Br 88:601–605CrossRefPubMed Pitto RP, Graydon AJ, Bradley L, Malak SF, Walker CG, Anderson IA (2006) Accuracy of a computer-assisted navigation system for total knee replacement. J Bone Joint Surg Br 88:601–605CrossRefPubMed
23.
Zurück zum Zitat Radtke S, Becher C, Noll Y, Ostermeier S (2010) Effect of limb rotation on radiographic alignment in total knee arthroplasties. Arch Orthop Trauma Surg 130:451–457CrossRefPubMed Radtke S, Becher C, Noll Y, Ostermeier S (2010) Effect of limb rotation on radiographic alignment in total knee arthroplasties. Arch Orthop Trauma Surg 130:451–457CrossRefPubMed
24.
Zurück zum Zitat Reed SC, Gollish J (1997) The accuracy of femoral intramedullary guides in total knee arthroplasty. J Arthroplasty 12(6):677–682CrossRefPubMed Reed SC, Gollish J (1997) The accuracy of femoral intramedullary guides in total knee arthroplasty. J Arthroplasty 12(6):677–682CrossRefPubMed
25.
Zurück zum Zitat Ritter MA, Faris PM, Keating EM, Meding JB (1994) Postoperative alignment of total knee replacement—its effect on survival. Clin Orthop Relat Res 299:153–156 Ritter MA, Faris PM, Keating EM, Meding JB (1994) Postoperative alignment of total knee replacement—its effect on survival. Clin Orthop Relat Res 299:153–156
26.
Zurück zum Zitat Tingart M, Lüring C, Bäthis H, Beckmann J, Grifka J, Perlick L (2008) Computer-assisted total knee arthroplasty versus the conventional technique: how precise is navigation in clinical routine? Knee Surg Sports Traumatol Arthrosc 16:44–50CrossRefPubMed Tingart M, Lüring C, Bäthis H, Beckmann J, Grifka J, Perlick L (2008) Computer-assisted total knee arthroplasty versus the conventional technique: how precise is navigation in clinical routine? Knee Surg Sports Traumatol Arthrosc 16:44–50CrossRefPubMed
27.
Zurück zum Zitat Wasielewski RC, Galante JO, Leighty RM, Natarajan RN, Rosenberg AG (1994) Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty. Clin Orthop Relat Res 299:31–43 Wasielewski RC, Galante JO, Leighty RM, Natarajan RN, Rosenberg AG (1994) Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty. Clin Orthop Relat Res 299:31–43
28.
Zurück zum Zitat Windsor RE, Scuderi GR, Moran MC, Insall JN (1989) Mechanisms of failure of the femoral and tibial components in total knee arthroplasty. Clin Orthop Relat Res 248:15–20 Windsor RE, Scuderi GR, Moran MC, Insall JN (1989) Mechanisms of failure of the femoral and tibial components in total knee arthroplasty. Clin Orthop Relat Res 248:15–20
29.
Zurück zum Zitat Woolson ST, Harris AH, Wagner DW, Giori NJ (2014) Component alignment during total knee arthroplasty with use of standard or custom instrumentation: a randomized clinical trial using computed tomography for postoperative alignment measurement. J Bone Joint Surg Am 96(5):366–372CrossRefPubMed Woolson ST, Harris AH, Wagner DW, Giori NJ (2014) Component alignment during total knee arthroplasty with use of standard or custom instrumentation: a randomized clinical trial using computed tomography for postoperative alignment measurement. J Bone Joint Surg Am 96(5):366–372CrossRefPubMed
30.
Zurück zum Zitat Yau WP, Leung A, Liu KG, Yan CH, Wong LLS, Chiu KY (2007) Interobserver and intra-observer errors in obtaining visually selected anatomical landmarks during registration process in non-image-based navigation-assisted total knee arthroplasty. J Arthroplasty 22:1150–1161CrossRefPubMed Yau WP, Leung A, Liu KG, Yan CH, Wong LLS, Chiu KY (2007) Interobserver and intra-observer errors in obtaining visually selected anatomical landmarks during registration process in non-image-based navigation-assisted total knee arthroplasty. J Arthroplasty 22:1150–1161CrossRefPubMed
Metadaten
Titel
What is the optimal valgus pre-set for intramedullary femoral alignment rods in total knee arthroplasty?
Publikationsdatum
06.05.2016
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 11/2017
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-016-4141-y

Weitere Artikel der Ausgabe 11/2017

Knee Surgery, Sports Traumatology, Arthroscopy 11/2017 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.