3T MRI of the knee with optimised isotropic 3D sequences: Accurate delineation of intra-articular pathology without prolonged acquisition times

verfasst von: Osamah M. Abdulaal, Louise Rainford, Peter MacMahon, Eoin Kavanagh, Marie Galligan, James Cashman, Allison McGee

Erschienen in: European Radiology | Ausgabe 11/2017

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To investigate optimised isotropic 3D turbo spin echo (TSE) and gradient echo (GRE)-based pulse sequences for visualisation of articular cartilage lesions within the knee joint.

Methods

Optimisation of experimental imaging sequences was completed using healthy volunteers (n=16) with a 3-Tesla (3T) MRI scanner. Imaging of patients with knee cartilage abnormalities (n=57) was then performed. Acquired sequences included 3D proton density-weighted (PDW) TSE (SPACE) with and without fat-suppression (FS), and T2*W GRE (TrueFISP) sequences, with acquisition times of 6:51, 6:32 and 5:35 min, respectively.

Results

One hundred sixty-one confirmed cartilage lesions were detected and categorised (Grade II n=90, Grade III n=71). The highest sensitivity and specificity for detecting cartilage lesions were obtained with TrueFISP with values of 84.7% and 92%, respectively. Cartilage SNR mean for PDW SPACE-FS was the highest at 72.2. TrueFISP attained the highest CNR means for joint fluid/cartilage (101.5) and joint fluid/ligament (156.5), and the lowest CNR for cartilage/meniscus (48.5). Significant differences were identified across the three sequences for all anatomical structures with respect to SNR and CNR findings (p-value <0.05).

Conclusion

Isotropic TrueFISP at 3T, optimised for acquisition time, accurately detects cartilage defects, although it demonstrated the lowest contrast between cartilage and meniscus.

Key points

• Cartilage is better visualised with 3D TrueFISP than 3D SPACE sequences.
• 3D TrueFISP is a reliable sequence for detecting low- and high-grade cartilage defects.
• 3D TrueFISP at 3T provides excellent contrast between cartilage and joint fluid.
Literatur
1.
Zurück zum Zitat Crema MD, Nogueira-Barbosa MH, Roemer FW et al (2013) Three-dimensional turbo spin-echo magnetic resonance imaging (MRI) and semiquantitative assessment of knee osteoarthritis: comparison with two-dimensional routine MRI. Osteoarthritis Cartilage 21:428–433CrossRef Crema MD, Nogueira-Barbosa MH, Roemer FW et al (2013) Three-dimensional turbo spin-echo magnetic resonance imaging (MRI) and semiquantitative assessment of knee osteoarthritis: comparison with two-dimensional routine MRI. Osteoarthritis Cartilage 21:428–433CrossRef
2.
Zurück zum Zitat Milewski MD, Smitaman E, Moukaddam H et al (2012) Comparison of 3D vs. 2D fast spin echo imaging for evaluation of articular cartilage in the knee on a 3T system scientific research. Eur J Radiol 81:1637–1643CrossRef Milewski MD, Smitaman E, Moukaddam H et al (2012) Comparison of 3D vs. 2D fast spin echo imaging for evaluation of articular cartilage in the knee on a 3T system scientific research. Eur J Radiol 81:1637–1643CrossRef
3.
Zurück zum Zitat Crema MD, Roemer FW, Marra MD et al (2011) Articular cartilage in the knee: current MR imaging techniques and applications in clinical practice and research. Radiographics 31:37–62CrossRef Crema MD, Roemer FW, Marra MD et al (2011) Articular cartilage in the knee: current MR imaging techniques and applications in clinical practice and research. Radiographics 31:37–62CrossRef
4.
Zurück zum Zitat Hafezi-Nejad N, Guermazi A, Roemer FW et al (2016) prediction of medial tibiofemoral compartment joint space loss progression using volumetric cartilage measurments: data from the FNIH OA biomakers consortium. Eur Radiol 27:464–473CrossRef Hafezi-Nejad N, Guermazi A, Roemer FW et al (2016) prediction of medial tibiofemoral compartment joint space loss progression using volumetric cartilage measurments: data from the FNIH OA biomakers consortium. Eur Radiol 27:464–473CrossRef
5.
Zurück zum Zitat Alizai H, Roemer FW, Hayashi D, Crema MD, Felson DT, Guermazi A (2015) An update on risk factors for cartilage loss in knee osteoarthritis assessed using MRI-based semiquantitative grading methods. Eur Radiol 25:883–893CrossRef Alizai H, Roemer FW, Hayashi D, Crema MD, Felson DT, Guermazi A (2015) An update on risk factors for cartilage loss in knee osteoarthritis assessed using MRI-based semiquantitative grading methods. Eur Radiol 25:883–893CrossRef
6.
Zurück zum Zitat Shapiro L, Staroswiecki E, Gold G (2010) Magnetic resonance imaging of the knee: optimizing 3 Tesla imaging. Semin Roentgenol 45:238–249CrossRef Shapiro L, Staroswiecki E, Gold G (2010) Magnetic resonance imaging of the knee: optimizing 3 Tesla imaging. Semin Roentgenol 45:238–249CrossRef
7.
Zurück zum Zitat Magee T, Williams D (2006) 3.0-T MRI of meniscal tears. AJR Am J Roentgenol 187:371–375CrossRef Magee T, Williams D (2006) 3.0-T MRI of meniscal tears. AJR Am J Roentgenol 187:371–375CrossRef
8.
Zurück zum Zitat Notohamiprodjo M, Horng A, Kuschel B et al (2012) 3D-imaging of the knee with an optimized 3D-FSE-sequence and a 15-channel knee-coil. Eur J Radiol 81:3441–3449CrossRef Notohamiprodjo M, Horng A, Kuschel B et al (2012) 3D-imaging of the knee with an optimized 3D-FSE-sequence and a 15-channel knee-coil. Eur J Radiol 81:3441–3449CrossRef
9.
Zurück zum Zitat Kijowski R, Davis KW, Blankenbaker DG, Woods MA, Del Rio AM, De Smet AA (2012) Evaluation of the menisci of the knee joint using three-dimensional isotropic resolution fast spin-echo imaging: diagnostic performance in 250 patients with surgical correlation. Skeletal Radiol 41:169–178CrossRef Kijowski R, Davis KW, Blankenbaker DG, Woods MA, Del Rio AM, De Smet AA (2012) Evaluation of the menisci of the knee joint using three-dimensional isotropic resolution fast spin-echo imaging: diagnostic performance in 250 patients with surgical correlation. Skeletal Radiol 41:169–178CrossRef
10.
Zurück zum Zitat Jung JY, Yoon YC, Kim HR, Choe BK, Wang JH, Jung JY (2013) Knee derangements: comparison of isotropic 3D fast spin-echo, isotropic 3D balanced fast field-echo, and conventional 2D fast spin-echo MR imaging. Radiology 268:802–813CrossRef Jung JY, Yoon YC, Kim HR, Choe BK, Wang JH, Jung JY (2013) Knee derangements: comparison of isotropic 3D fast spin-echo, isotropic 3D balanced fast field-echo, and conventional 2D fast spin-echo MR imaging. Radiology 268:802–813CrossRef
11.
Zurück zum Zitat Kanal E, Barkovich AJ, Bell C (2007) ACR gudiance document for safe MR practice: 2007. AJR Am J RoentgeonI 188:1447–1474CrossRef Kanal E, Barkovich AJ, Bell C (2007) ACR gudiance document for safe MR practice: 2007. AJR Am J RoentgeonI 188:1447–1474CrossRef
12.
Zurück zum Zitat Brennan PC, McEntee M, Evanoff M, Phillips P, O'Connor WT, Manning DJ (2007) Ambient lighting: effect of illumination on soft-copy viewing of radiographs of the wrist. AJR Am J Roentgenol 188:W177–W180CrossRef Brennan PC, McEntee M, Evanoff M, Phillips P, O'Connor WT, Manning DJ (2007) Ambient lighting: effect of illumination on soft-copy viewing of radiographs of the wrist. AJR Am J Roentgenol 188:W177–W180CrossRef
13.
Zurück zum Zitat Tokuda O, Harada Y, Shiraishi G et al (2012) MRI of the anatomical structures of the knee: the proton density-weighted fast spin-echo sequence vs the proton density-weighted fast-recovery fast spin-echo sequence. Br J Radiol 85:e686–e693CrossRef Tokuda O, Harada Y, Shiraishi G et al (2012) MRI of the anatomical structures of the knee: the proton density-weighted fast spin-echo sequence vs the proton density-weighted fast-recovery fast spin-echo sequence. Br J Radiol 85:e686–e693CrossRef
14.
Zurück zum Zitat Sutter R, Hodek R, Fucentesen SF, Nittka M, Pfirrmann CW (2013) Total knee arthroplasty MRI featurig slice-encoding for metal artifact correction: reduction of artifacts for STIR and proton density-weighted sequences. AJR Am J Roentgeonl 201:1315–1324CrossRef Sutter R, Hodek R, Fucentesen SF, Nittka M, Pfirrmann CW (2013) Total knee arthroplasty MRI featurig slice-encoding for metal artifact correction: reduction of artifacts for STIR and proton density-weighted sequences. AJR Am J Roentgeonl 201:1315–1324CrossRef
15.
Zurück zum Zitat Ristow O, Steinbach L, Sabo G et al (2009) Isotropic 3D fast spin-echo imaging versus standard 2D imaging at 3.0 T of the knee--image quality and diagnostic performance. Eur Radiol 19:1263–1272CrossRef Ristow O, Steinbach L, Sabo G et al (2009) Isotropic 3D fast spin-echo imaging versus standard 2D imaging at 3.0 T of the knee--image quality and diagnostic performance. Eur Radiol 19:1263–1272CrossRef
16.
Zurück zum Zitat Van Dyck P, Vanhevel F, Vanhoenacker FM et al (2015) Morphological MR imaging of the articular cartilage of the knee at 3 T-comparison of standard and novel 3D sequences. Insights Imag 6:285–293CrossRef Van Dyck P, Vanhevel F, Vanhoenacker FM et al (2015) Morphological MR imaging of the articular cartilage of the knee at 3 T-comparison of standard and novel 3D sequences. Insights Imag 6:285–293CrossRef
17.
Zurück zum Zitat Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRef Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRef
18.
Zurück zum Zitat Vasey MW, Thayer JF (1987) The continuing problem of false positives in repeated measures ANOVA in psychophysiology: a multivariate solution. Psychophysiology 24:479–486CrossRef Vasey MW, Thayer JF (1987) The continuing problem of false positives in repeated measures ANOVA in psychophysiology: a multivariate solution. Psychophysiology 24:479–486CrossRef
19.
Zurück zum Zitat Moriya S, Miki Y, Kanagaki M, Matsuno Y, Miyati T (2014) 90 degrees -flip-angle three-dimensional double-echo steady-state (3D-DESS) magnetic resonance imaging of the knee: isovoxel cartilage imaging at 3T. Eur J Radiol 83:1429–1432CrossRef Moriya S, Miki Y, Kanagaki M, Matsuno Y, Miyati T (2014) 90 degrees -flip-angle three-dimensional double-echo steady-state (3D-DESS) magnetic resonance imaging of the knee: isovoxel cartilage imaging at 3T. Eur J Radiol 83:1429–1432CrossRef
20.
Zurück zum Zitat Rehnitz C, Kupfer J, Streich NA et al (2014) Comparison of biochemical cartilage imaging techniques at 3 T MRI. Osteoarthr Cartilage 22:1732–1742CrossRef Rehnitz C, Kupfer J, Streich NA et al (2014) Comparison of biochemical cartilage imaging techniques at 3 T MRI. Osteoarthr Cartilage 22:1732–1742CrossRef
21.
Zurück zum Zitat Duc SR, Pfirrmann CW, Koch PP, Zanetti M, Hodler J (2008) Internal knee derangement assessed with 3-minute three-dimensional isovoxel true FISP MR sequence: preliminary study. Radiology 246:526–535CrossRef Duc SR, Pfirrmann CW, Koch PP, Zanetti M, Hodler J (2008) Internal knee derangement assessed with 3-minute three-dimensional isovoxel true FISP MR sequence: preliminary study. Radiology 246:526–535CrossRef
22.
Zurück zum Zitat Duc SR, Pfirrmann CW, Schmid MR et al (2007) Articular cartilage defects detected with 3D water-excitation true FISP: prospective comparison with sequences commonly used for knee imaging. Radiology 245:216–223CrossRef Duc SR, Pfirrmann CW, Schmid MR et al (2007) Articular cartilage defects detected with 3D water-excitation true FISP: prospective comparison with sequences commonly used for knee imaging. Radiology 245:216–223CrossRef
23.
Zurück zum Zitat Friedrich KM, Reiter G, Kaiser B et al (2011) High-resolution cartilage imaging of the knee at 3T: basic evaluation of modern isotropic 3D MR-sequences. Eur J Radiol 78:398–405CrossRef Friedrich KM, Reiter G, Kaiser B et al (2011) High-resolution cartilage imaging of the knee at 3T: basic evaluation of modern isotropic 3D MR-sequences. Eur J Radiol 78:398–405CrossRef
24.
Zurück zum Zitat Thorlund JB, Juhl CB, Roos EM, Lohmander LS (2015) Arthroscopic surgery for degenerative knee: systematic review and meta-analysis of benefits and harms. Br J Sports Med 49:1229–1235CrossRef Thorlund JB, Juhl CB, Roos EM, Lohmander LS (2015) Arthroscopic surgery for degenerative knee: systematic review and meta-analysis of benefits and harms. Br J Sports Med 49:1229–1235CrossRef
25.
Zurück zum Zitat Cui Y, Dyvorne H, Besa C, Cooper N, Taouli B (2015) IVIM diffusion-weighted imaging of the liver at 3.0T: comparison with 1.5T. Eur J Radiol Open 2:123–128CrossRef Cui Y, Dyvorne H, Besa C, Cooper N, Taouli B (2015) IVIM diffusion-weighted imaging of the liver at 3.0T: comparison with 1.5T. Eur J Radiol Open 2:123–128CrossRef
26.
Zurück zum Zitat Chen CA, Lu W, John CT et al (2009) Multiecho IDEAL gradient-echo water-fat separation for rapid assessment of cartilage volume at 1.5 T: initial experience. Radiology 252:561–567CrossRef Chen CA, Lu W, John CT et al (2009) Multiecho IDEAL gradient-echo water-fat separation for rapid assessment of cartilage volume at 1.5 T: initial experience. Radiology 252:561–567CrossRef
27.
Zurück zum Zitat Jung JY, Jee WH, Park MY, Lee SY, Kim JM (2012) Meniscal tear configurations: categorization with 3D isotropic turbo spin-echo MRI compared with conventional MRI at 3 T. AJR Am J Roentgenol 198:W173–W180CrossRef Jung JY, Jee WH, Park MY, Lee SY, Kim JM (2012) Meniscal tear configurations: categorization with 3D isotropic turbo spin-echo MRI compared with conventional MRI at 3 T. AJR Am J Roentgenol 198:W173–W180CrossRef
Metadaten
Titel
3T MRI of the knee with optimised isotropic 3D sequences: Accurate delineation of intra-articular pathology without prolonged acquisition times
verfasst von
Osamah M. Abdulaal
Louise Rainford
Peter MacMahon
Eoin Kavanagh
Marie Galligan
James Cashman
Allison McGee
Publikationsdatum
21.04.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 11/2017
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-017-4816-x

Weitere Artikel der Ausgabe 11/2017

European Radiology 11/2017 Zur Ausgabe

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Wie toxische Männlichkeit der Gesundheit von Männern schadet

08.04.2024 Andrologie Nachrichten

Stark, erfolgreich, allzeit belastbar – das sind Erwartungen, die Jungen und Männer von der Gesellschaft spüren. Das kann sie „toxisch“ werden lassen – und letztlich sogar der Gesundheit schaden, mahnt Dr. Dirk Sander von der Deutschen Aidshilfe.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.