A method for monitoring and improving patient: ventilator interaction
verfasst von:
Magdy Younes, Laurent Brochard, Salvatore Grasso, John Kun, Jordi Mancebo, Marco Ranieri, Jean-Christophe Richard, Henry Younes
Erschienen in:
Intensive Care Medicine
|
Ausgabe 8/2007
Einloggen, um Zugang zu erhalten
Abstract
Objective
To evaluate a new approach for monitoring and improving patient-ventilator interaction that utilizes a signal generated by the equation of motion, using improvised values for resistance and elastance obtained noninvasively.
Design and setting
Observational study in intensive care units in five European centers.
Patients
We studied 21 stable patients instrumented with esophageal/gastric catheters for a previous study and ventilated alternately with pressure support (PSV) and proportional assist (PAV) ventilation with a Tyco 840 ventilator.
Measurements and results
Previously recorded digital files were analyzed in real-time by a prototype incorporating the new technology (PVI monitor, YRT, Winnipeg, Canada). Actual onsets (PDI-TONSET) and ends (PDI-TEND) of inspiratory efforts, ineffective efforts, and patient respiratory rate were identified visually from transdiaphragmatic or calculated respiratory muscle pressure. Monitor-identified TONSET occurred 0.107 ± 0.074 s after PDI-TONSET, substantially less than trigger delay observed with conventional triggering (0.326 ± 0.086 s). End of effort was identified 0.097 ± 0.096 s after PDI-TEND, significantly less than actual cycling-off delay during PSV (0.486 ± 0.307 s) or PAV (0.277 ± 0.084 s). The monitor detected 80% of ineffective efforts. There was excellent agreement between monitor-estimated respiratory rate and actual patient rate over a wide range (17–59/min) of patient rates (mean ( ± SD) of difference −0.2 ± 1.9/min for pressure support and 0.2 ± 0.9/min for proportional assist) even when large discrepancies existed (> 35/min) between patient and ventilator rates.
Conclusions
The proposed approach should make it possible to improve patient-ventilator interaction and to obtain accurate estimates of true patient respiratory rate when there is nonsynchrony.