A new concept for verifying the isocentric alignment of the proton-rotational gantry for radiation control

verfasst von: Takahiro Kato, Yuhei Yamazaki, Tatsuhiko Sagara

Erschienen in: Radiological Physics and Technology | Ausgabe 1/2020

Einloggen, um Zugang zu erhalten

Abstract

The purpose of this study was to introduce the modified Winston–Lutz (mWL) test and to evaluate its feasibility. This is a new method to completely absorb the proton beam around the isocenter inside a phantom for radiation control. The mWL test was performed using a 14-cm-diameter acrylic Lucy 3D QA Phantom for a passive-scattering proton beam gantry. The energy of the unmodulated proton beam was adjusted such that the residual range was < 130 mm, and the energy of the proton beam was completely lost around the isocenter. The radiation field was formed with a multi-leaf collimator at 8.6 × 8.6 mm2 in the isocenter plane. The phantom was loaded with a 4-mm-diameter tungsten ball, and the EBT3 was set up at the isocenter. The proton beam was irradiated at gantry angles with 45° steps, and the isocenter deviation of the proton beam was measured and subsequently analyzed. Although the radiation field penumbra was blurred under the influence of scattered radiation due to placement in the phantom compared to the traditional WL test placed in the air, evaluation of the beam axis accuracy was possible. The results confirmed that the maximum total displacement was less than 0.9 mm, and the specifications of the device were met. The mWL test is feasible and effective to reduce the building activation in proton beam treatment facilities. Thus, it can be considered a useful method that sufficiently satisfies the shielding calculation conditions.
Literatur
1.
Zurück zum Zitat Klein EE, Hanley J, Bayouth J, Yin FF, Simon W, Dresser S, Serago C, Aguirre F, Ma L, Arjomandy B, Liu C, Sandin C, Holmes T; Task Group 142, American Association of Physicists in Medicine. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36(9):4197–212. Klein EE, Hanley J, Bayouth J, Yin FF, Simon W, Dresser S, Serago C, Aguirre F, Ma L, Arjomandy B, Liu C, Sandin C, Holmes T; Task Group 142, American Association of Physicists in Medicine. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36(9):4197–212.
2.
Zurück zum Zitat Moyers MF, Lesyna W. Isocenter characteristics of an external ring proton gantry. Int J Radiat Oncol Biol Phys. 2004;60(5):1622–30.CrossRef Moyers MF, Lesyna W. Isocenter characteristics of an external ring proton gantry. Int J Radiat Oncol Biol Phys. 2004;60(5):1622–30.CrossRef
3.
Zurück zum Zitat Barkhof J, Schut G, Flanz JB, Goitein M, Schippers JM. Verification of the alignment of a therapeutic radiation beam relative to its patient positioner. Med Phys. 1999;26(11):2429–37.CrossRef Barkhof J, Schut G, Flanz JB, Goitein M, Schippers JM. Verification of the alignment of a therapeutic radiation beam relative to its patient positioner. Med Phys. 1999;26(11):2429–37.CrossRef
4.
Zurück zum Zitat Ciangaru G, Yang JN, Oliver PJ, Bues M, Zhu M, Nakagawa F, Chiba H, Nakamura S, Yoshino H, Umezawa M, Smith AR. Verification procedure for isocentric alignment of proton beams. J Appl Clin Med Phys. 2007;8(4):2671.CrossRef Ciangaru G, Yang JN, Oliver PJ, Bues M, Zhu M, Nakagawa F, Chiba H, Nakamura S, Yoshino H, Umezawa M, Smith AR. Verification procedure for isocentric alignment of proton beams. J Appl Clin Med Phys. 2007;8(4):2671.CrossRef
5.
Zurück zum Zitat Fuse H, Sakae T, Terunuma T, Sato M, Aoki S. An isocenter detection and verification device for use in proton therapy. Rev. Sci. Instrum. 2012;83(12):125109.CrossRef Fuse H, Sakae T, Terunuma T, Sato M, Aoki S. An isocenter detection and verification device for use in proton therapy. Rev. Sci. Instrum. 2012;83(12):125109.CrossRef
6.
Zurück zum Zitat Hansen P, Hu D. An isocenter estimation tool for proton gantry alignment. Rev. Sc.i Instrum. 2017;88(12):125102.CrossRef Hansen P, Hu D. An isocenter estimation tool for proton gantry alignment. Rev. Sc.i Instrum. 2017;88(12):125102.CrossRef
7.
Zurück zum Zitat Lutz W, Winston KR, Maleki N. A system for stereotactic radiosurgery with a linear accelerator. Int. J. Radiat. Oncol. Biol. Phys. 1988;14(2):373–81.CrossRef Lutz W, Winston KR, Maleki N. A system for stereotactic radiosurgery with a linear accelerator. Int. J. Radiat. Oncol. Biol. Phys. 1988;14(2):373–81.CrossRef
8.
Zurück zum Zitat Du W, Johnson JL, Jiang W, Kudchadker RJ. On the selection of gantry and collimator angles for isocenter localization using Winston–Lutz tests. J. Appl. Clin. Med. Phys. 2016;17(1):167–78.CrossRef Du W, Johnson JL, Jiang W, Kudchadker RJ. On the selection of gantry and collimator angles for isocenter localization using Winston–Lutz tests. J. Appl. Clin. Med. Phys. 2016;17(1):167–78.CrossRef
9.
Zurück zum Zitat Cadini F, Bolst D, Guatelli S, Beltran C, Jackson M, Rosenfeld AB. Neutron shielding for a new projected proton therapy facility: a Geant4 simulation study. Phys Med. 2016;32(12):1862–71.CrossRef Cadini F, Bolst D, Guatelli S, Beltran C, Jackson M, Rosenfeld AB. Neutron shielding for a new projected proton therapy facility: a Geant4 simulation study. Phys Med. 2016;32(12):1862–71.CrossRef
10.
Zurück zum Zitat Sakurai H, Ishikawa H, Okumura T. Proton beam therapy in Japan: current and future status. Jpn. J. Clin. Oncol. 2016;46(10):885–92.CrossRef Sakurai H, Ishikawa H, Okumura T. Proton beam therapy in Japan: current and future status. Jpn. J. Clin. Oncol. 2016;46(10):885–92.CrossRef
11.
Zurück zum Zitat Smith A, Gillin M, Bues M, Zhu XR, Suzuki K, Mohan R, Woo S, Lee A, Komaki R, Cox J, Hiramoto K, Akiyama H, Ishida T, Sasaki T, Matsuda K. The M. D. Anderson proton therapy system. Med Phys. 2009;36(9): 4068–83.CrossRef Smith A, Gillin M, Bues M, Zhu XR, Suzuki K, Mohan R, Woo S, Lee A, Komaki R, Cox J, Hiramoto K, Akiyama H, Ishida T, Sasaki T, Matsuda K. The M. D. Anderson proton therapy system. Med Phys. 2009;36(9): 4068–83.CrossRef
12.
Zurück zum Zitat Wang N, Ghebremedhin A, Patyal B. Commissioning of a proton gantry equipped with dual X-ray imagers and a robotic patient positioner, and evaluation of the accuracy of single-beam image registration for this system. Med. Phys. 2015;42(6):2979–91.CrossRef Wang N, Ghebremedhin A, Patyal B. Commissioning of a proton gantry equipped with dual X-ray imagers and a robotic patient positioner, and evaluation of the accuracy of single-beam image registration for this system. Med. Phys. 2015;42(6):2979–91.CrossRef
13.
Zurück zum Zitat Pidikiti R, Patel BC, Maynard MR, Dugas JP, Syh J, Sahoo N, Wu HT, Rosen LR. Commissioning of the world's first compact pencil-beam scanning proton therapy system. J. Appl. Clin. Med. Phys. 2018;19(1):94–105.CrossRef Pidikiti R, Patel BC, Maynard MR, Dugas JP, Syh J, Sahoo N, Wu HT, Rosen LR. Commissioning of the world's first compact pencil-beam scanning proton therapy system. J. Appl. Clin. Med. Phys. 2018;19(1):94–105.CrossRef
Metadaten
Titel
A new concept for verifying the isocentric alignment of the proton-rotational gantry for radiation control
verfasst von
Takahiro Kato
Yuhei Yamazaki
Tatsuhiko Sagara
Publikationsdatum
09.11.2019
Verlag
Springer Singapore
Erschienen in
Radiological Physics and Technology / Ausgabe 1/2020
Print ISSN: 1865-0333
Elektronische ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-019-00544-4

Weitere Artikel der Ausgabe 1/2020

Radiological Physics and Technology 1/2020 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.