A Quantification Method for Breast Tissue Thickness and Iodine Concentration Using Photon-Counting Detector

verfasst von: Seokmin Han

Erschienen in: Journal of Imaging Informatics in Medicine | Ausgabe 5/2015

Einloggen, um Zugang zu erhalten

Abstract

The purpose of contrast-enhanced digital mammography (CEDM) is to facilitate detection and characterization of the lesions in the breast using intravenous injection of an iodinated contrast agent. CEDM produces iodine images with gray levels proportional to iodine concentration at each pixel, which can be considered as quantification of iodine. While dual-energy CEDM requires an accurate knowledge of the thickness of compressed breast for the quantification, it is known that the accuracy of the built-in thickness measurement is not satisfactory. Triple-energy CEDM, which can provide a third image, can alleviate the limitation of dual-energy CEDM. If triple exposure technique is applied, it can lead to increased risk of motion artifact. An energy-resolving photon-counting detector (PCD) that can acquire multispectral X-ray images can reduce the risk of motion artifact. In this research, an easily implementable method for iodine quantification in breast imaging was suggested, and it was applied to the images of breast phantom with various iodine concentrations. The iodine concentrations in breast phantom simulate lesions filled with different iodine concentrations in the breast. The result shows that the proposed method can quantify the iodine concentrations in breast phantom accurately.
Literatur
1.
Zurück zum Zitat Thomas R. Nelson, Laura I. Cerviño, John M. Boone, and Karen K. Lindfors: Classification of breast computed tomography data. Medical Physics, 2008 Thomas R. Nelson, Laura I. Cerviño, John M. Boone, and Karen K. Lindfors: Classification of breast computed tomography data. Medical Physics, 2008
2.
Zurück zum Zitat Clarisse Dromain, Corinne Balleyguier, Ghazal Adler, Jean Remi Garbay, Suzette Delaloge: Contrast-enhanced digital mammography. European Journal of Radiology, 2009 Clarisse Dromain, Corinne Balleyguier, Ghazal Adler, Jean Remi Garbay, Suzette Delaloge: Contrast-enhanced digital mammography. European Journal of Radiology, 2009
3.
Zurück zum Zitat John M. Lewin, Pamela K. Isaacs, Virginia Vance, Fred J. Larke: Dual-Energy Contrast enhanced Digital Subtraction Mammography: Feasibility. Radiology, October 2003 John M. Lewin, Pamela K. Isaacs, Virginia Vance, Fred J. Larke: Dual-Energy Contrast enhanced Digital Subtraction Mammography: Feasibility. Radiology, October 2003
4.
Zurück zum Zitat Clarisse Dromain, Fabienne Thibault, Felix Diekmann, Eva M. Fallenberg, Roberta A. Jong, Marcia Koomen, R. Edward Hendrick, Anne Tardivon and Alicia Toledano: Dual-energy contrast-enhanced digital mammography: initial clinical results of a multireader, multicase study. Breast Cancer Research, 2012 Clarisse Dromain, Fabienne Thibault, Felix Diekmann, Eva M. Fallenberg, Roberta A. Jong, Marcia Koomen, R. Edward Hendrick, Anne Tardivon and Alicia Toledano: Dual-energy contrast-enhanced digital mammography: initial clinical results of a multireader, multicase study. Breast Cancer Research, 2012
5.
Zurück zum Zitat Maxine Jochelson: Contrast-Enhanced Digital Mammography. Radiologic Clinics of North America, 2014 Maxine Jochelson: Contrast-Enhanced Digital Mammography. Radiologic Clinics of North America, 2014
6.
Zurück zum Zitat Yu Zou and Michael D. Silver: Analysis of Fast kV-switching in Dual Energy CT using a Prereconstruction Decomposition Technique. Proceeding of SPIE, March 2008 Yu Zou and Michael D. Silver: Analysis of Fast kV-switching in Dual Energy CT using a Prereconstruction Decomposition Technique. Proceeding of SPIE, March 2008
7.
Zurück zum Zitat Serghei Malkov, Jeff Wang, Karla Kerlikowske, Steven R. Cummings, and John A. Shepherd: Single x-ray absorptiometry method for the quantitative mammographic measure of fibroglandular tissue volume. Medical Physics, 2009 Serghei Malkov, Jeff Wang, Karla Kerlikowske, Steven R. Cummings, and John A. Shepherd: Single x-ray absorptiometry method for the quantitative mammographic measure of fibroglandular tissue volume. Medical Physics, 2009
8.
Zurück zum Zitat Albert H. Tyson, Gordon E. Mawdsley, and Martin J. Yaffe: Measurement of compressed breast thickness by optical stereoscopic photogrammetry. Medical Physics, 2009 Albert H. Tyson, Gordon E. Mawdsley, and Martin J. Yaffe: Measurement of compressed breast thickness by optical stereoscopic photogrammetry. Medical Physics, 2009
9.
Zurück zum Zitat A. Burch and J. Law: A method for estimating compressed breast thickness during mammography. British Journal of Radiology, 1995 A. Burch and J. Law: A method for estimating compressed breast thickness during mammography. British Journal of Radiology, 1995
10.
Zurück zum Zitat P.V. Granton, S.I. Pollmann, N.L. Ford, M. Drangova, and D.W. Holdsworth: Implementation of dual- and triple-energy cone-beam micro-CT for postreconstruction material decomposition. Medical Physics, 2008 P.V. Granton, S.I. Pollmann, N.L. Ford, M. Drangova, and D.W. Holdsworth: Implementation of dual- and triple-energy cone-beam micro-CT for postreconstruction material decomposition. Medical Physics, 2008
11.
Zurück zum Zitat J. P. Schlomka, E. Roessl, R. Dorscheid, S. Dill, G. Martens, T. Istel, C. Bäumer, C. Herrmann, R. Steadman, G. Zeitler, A. Livne and R. Proksa: Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography. Physics in Medicine and Biology, 2008 J. P. Schlomka, E. Roessl, R. Dorscheid, S. Dill, G. Martens, T. Istel, C. Bäumer, C. Herrmann, R. Steadman, G. Zeitler, A. Livne and R. Proksa: Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography. Physics in Medicine and Biology, 2008
12.
Zurück zum Zitat Markus Firsching, Patrick Takoukam Talla, Thilo Michel, Gisela Anton: Material resolving X-ray imaging using spectrum reconstruction with Medipix2. Nuclear Instruments and Methods in Physics Research Section A, 2008 Markus Firsching, Patrick Takoukam Talla, Thilo Michel, Gisela Anton: Material resolving X-ray imaging using spectrum reconstruction with Medipix2. Nuclear Instruments and Methods in Physics Research Section A, 2008
13.
Zurück zum Zitat P. S. Yeh, A. Macovski, and W. Brody: Noise analysis in isolation of iodine using three energies. Medical Physics, 1980 P. S. Yeh, A. Macovski, and W. Brody: Noise analysis in isolation of iodine using three energies. Medical Physics, 1980
14.
Zurück zum Zitat F. Kelcz. and C.A. Mistretta: Absorption-edge fluoroscopy using a three-spectrum technique. Medical Physics, 1976. F. Kelcz. and C.A. Mistretta: Absorption-edge fluoroscopy using a three-spectrum technique. Medical Physics, 1976.
15.
Zurück zum Zitat S. Puong, P. Milioni de Carvalho, and S. Muller: Triple-Energy Contrast Enhanced Digital Mammography. Proc. of SPIE, 2010 S. Puong, P. Milioni de Carvalho, and S. Muller: Triple-Energy Contrast Enhanced Digital Mammography. Proc. of SPIE, 2010
16.
Zurück zum Zitat Seokmin Han, Dong-Goo Kang, Sunghoon Kang, Younghun Sung: Quantitative breast imaging using photon counting detector. Proc. SPIE. 8668, Medical Imaging, 2013 Seokmin Han, Dong-Goo Kang, Sunghoon Kang, Younghun Sung: Quantitative breast imaging using photon counting detector. Proc. SPIE. 8668, Medical Imaging, 2013
17.
Zurück zum Zitat Seokmin Han, Dong-Goo Kang: Tissue cancellation in dual energy mammography using a calibration phantom customized for direct mapping. IEEE Transactions on Medical Imaging, 2014 Seokmin Han, Dong-Goo Kang: Tissue cancellation in dual energy mammography using a calibration phantom customized for direct mapping. IEEE Transactions on Medical Imaging, 2014
18.
Zurück zum Zitat D.R. Dance, C.L. Skinner, K.C. Young, J.R. Beckett, C.J. Kotre: Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Physics in Medicine and Biology, 2000 D.R. Dance, C.L. Skinner, K.C. Young, J.R. Beckett, C.J. Kotre: Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Physics in Medicine and Biology, 2000
19.
Zurück zum Zitat D.R. Dance: Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose. Physics in Medicine and Biology, 1990 D.R. Dance: Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose. Physics in Medicine and Biology, 1990
20.
Zurück zum Zitat William P. Argo, Kathleen Hintenlang, and David E. Hintenlang: A tissue-equivalent phantom series for mammography dosimetry. Journal of Applied Clinical Medical Physics, Vol 5, No 4, 2004 William P. Argo, Kathleen Hintenlang, and David E. Hintenlang: A tissue-equivalent phantom series for mammography dosimetry. Journal of Applied Clinical Medical Physics, Vol 5, No 4, 2004
21.
Zurück zum Zitat Grant E. van der Ploeg, Robert T. Withers, and Joe Laforgia: Percent body fat via DEXA: comparison with a four-compartment model. Journal of Applied Physiology, February 2003 Grant E. van der Ploeg, Robert T. Withers, and Joe Laforgia: Percent body fat via DEXA: comparison with a four-compartment model. Journal of Applied Physiology, February 2003
22.
Zurück zum Zitat Angelo Pietrobelli, Zimian Wang, Carmelo Formica, and Steven B. Heymsfield: Dual-energy X-ray absorptiometry: fat estimation errors due to variation in soft tissue hydration. Americal Journal of Physiology Endocrinology and Metabolism, May 1998 Angelo Pietrobelli, Zimian Wang, Carmelo Formica, and Steven B. Heymsfield: Dual-energy X-ray absorptiometry: fat estimation errors due to variation in soft tissue hydration. Americal Journal of Physiology Endocrinology and Metabolism, May 1998
23.
Zurück zum Zitat J.T.M. Jansen,W.J.H. Veldkamp, M.A.O. Thijssen, S. van Woudenberg, and J. Zoetelief: Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography. Physics in Medicine and Biology, 2005 J.T.M. Jansen,W.J.H. Veldkamp, M.A.O. Thijssen, S. van Woudenberg, and J. Zoetelief: Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography. Physics in Medicine and Biology, 2005
24.
Zurück zum Zitat Christopher E Tromans, Mary R Cocker and Sir Michael Brady: Quantification and normalization of x-ray mammograms. Physics in Medicine and Biology, 2012 Christopher E Tromans, Mary R Cocker and Sir Michael Brady: Quantification and normalization of x-ray mammograms. Physics in Medicine and Biology, 2012
25.
Zurück zum Zitat P. Baldelli,A. Bravin, C. Di Maggio, G. Gennaro, A. Sarnelli, A. Taibi, M. Gambaccini: Evaluation of the minimum iodine concentration for contrast-enhanced subtraction. Physics in Medicine and Biology, 2006 P. Baldelli,A. Bravin, C. Di Maggio, G. Gennaro, A. Sarnelli, A. Taibi, M. Gambaccini: Evaluation of the minimum iodine concentration for contrast-enhanced subtraction. Physics in Medicine and Biology, 2006
26.
Zurück zum Zitat Robert A. Kruger: Estimation of the diameter of and iodine concentration within blood vessels using digital radiography devices. Med Phys, 1981 Robert A. Kruger: Estimation of the diameter of and iodine concentration within blood vessels using digital radiography devices. Med Phys, 1981
27.
Zurück zum Zitat T. P. Fuerst, M. S. Van Lysel: Measurement of absolute blood iodine concentration during digital subtraction ventriculography. Med Phys, 1993 T. P. Fuerst, M. S. Van Lysel: Measurement of absolute blood iodine concentration during digital subtraction ventriculography. Med Phys, 1993
28.
Zurück zum Zitat Felix Diekmann, Ulrich Bick: Tomosynthesis and contrast-enhanced digital mammography: recent advances in digital mammography. Eur Radiol, 2007 Felix Diekmann, Ulrich Bick: Tomosynthesis and contrast-enhanced digital mammography: recent advances in digital mammography. Eur Radiol, 2007
29.
Zurück zum Zitat Masatoshi Saito: Dual-energy approach to contrast-enhanced mammography using the balanced filter method: spectral optimization and preliminary phantom measurement. Med Phys, 2007 Masatoshi Saito: Dual-energy approach to contrast-enhanced mammography using the balanced filter method: spectral optimization and preliminary phantom measurement. Med Phys, 2007
30.
Zurück zum Zitat Robert Leithner, Thomas Knogler and Peter Homolka: Development and production of a prototype iodine contrast phantom for CEDEM. Phys Med Biol, 2013 Robert Leithner, Thomas Knogler and Peter Homolka: Development and production of a prototype iodine contrast phantom for CEDEM. Phys Med Biol, 2013
31.
Zurück zum Zitat S. Eva Singletary, Craig Allred, Pandora Ashley, Lawrence W. Bassett, Donald Berry, Kirby I. Bland, Patrick I. Borgen, Gary Clark, Stephen B. Edge, Daniel F. Hayes, Lorie L. Hughes, Robert V.P. Hutter, Monica Morrow, David L. Page, Abram Recht, Richard L. Theriault, Ann Thor, Donald L. Weaver, H. Samuel Wieand and Frederick L. Greene: Revision of the American Joint Committee on Cancer Staging System for Breast Cancer. Journal of Clinical Oncology, July 2002 S. Eva Singletary, Craig Allred, Pandora Ashley, Lawrence W. Bassett, Donald Berry, Kirby I. Bland, Patrick I. Borgen, Gary Clark, Stephen B. Edge, Daniel F. Hayes, Lorie L. Hughes, Robert V.P. Hutter, Monica Morrow, David L. Page, Abram Recht, Richard L. Theriault, Ann Thor, Donald L. Weaver, H. Samuel Wieand and Frederick L. Greene: Revision of the American Joint Committee on Cancer Staging System for Breast Cancer. Journal of Clinical Oncology, July 2002
32.
Zurück zum Zitat E. Fredenberg, M. Hemmendorff, B. Cederström, M. Åslund, and M. Danielsson: Contrast-enhanced spectral mammography with a photon-counting detector. Medical Physics, 2010 E. Fredenberg, M. Hemmendorff, B. Cederström, M. Åslund, and M. Danielsson: Contrast-enhanced spectral mammography with a photon-counting detector. Medical Physics, 2010
Metadaten
Titel
A Quantification Method for Breast Tissue Thickness and Iodine Concentration Using Photon-Counting Detector
verfasst von
Seokmin Han
Publikationsdatum
01.10.2015
Verlag
Springer US
Erschienen in
Journal of Imaging Informatics in Medicine / Ausgabe 5/2015
Print ISSN: 2948-2925
Elektronische ISSN: 2948-2933
DOI
https://doi.org/10.1007/s10278-015-9784-6

Weitere Artikel der Ausgabe 5/2015

Journal of Digital Imaging 5/2015 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.