Objective

We hypothesize that cut screws will deform in a manner that increases the core and outer diameters of the screw hole compared to uncut controls, and effects will be more pronounced in titanium screws.

Materials and methods

We used biomechanical polyurethane foam blocks to simulate cortical bone. We organized four groups of stainless steel and titanium cut and uncut screws. Blocks were fitted with a jig to ensure perpendicular screw insertion. We imaged the blocks using digital mammography and measured them using PACS software. Power analysis determined a power of 0.95 and an alpha error of 0.05.

Results

Highly statistically significant differences in core diameter were found after cutting stainless steel and titanium screws. Cutting stainless steel screws increased core diameter by 0.30 mm (95% CI, 0.16 to 0.45; p < .001). Titanium screws’ core diameter increased by 0.45 mm (95% CI, 0.30 to 0.61; p < .001). No significant differences were found in the outer diameters of stainless steel and titanium screws after cutting.

Conclusion

Titanium and stainless steel screw tracts demonstrated screw core diameter and screw thread pattern deformation after cutting. Titanium screws demonstrated more significant effects.
Literatur
5.
Zurück zum Zitat ASTM International. ASTM F1839-08: Standard specification for rigid polyurethane foam for use as a standard material for testing orthopaedic devices and instruments. American Society for Testing and Materials; 2021. ASTM International. ASTM F1839-08: Standard specification for rigid polyurethane foam for use as a standard material for testing orthopaedic devices and instruments. American Society for Testing and Materials; 2021.
6.
Zurück zum Zitat ASTM International. ASTM F543-17: Standard specification and test methods for metallic medical bone screws. American Society for Testing and Materials; 2017. ASTM International. ASTM F543-17: Standard specification and test methods for metallic medical bone screws. American Society for Testing and Materials; 2017.
13.
Zurück zum Zitat Selenia H. Hologic selenia dimensions mammography system data sheet. Hologic Inc.; 2016. Selenia H. Hologic selenia dimensions mammography system data sheet. Hologic Inc.; 2016.
14.
Zurück zum Zitat Fleiss JL. Statistical methods for rates and proportions. 2nd ed. New York: John Wiley; 1981. Fleiss JL. Statistical methods for rates and proportions. 2nd ed. New York: John Wiley; 1981.
17.
Metadaten
Titel
A radiologic determination of the different screw cutting patterns in cut and uncut orthopedic cortical screws using a novel imaging technique
verfasst von
Joshua W. Major
Andrew J. Ernst
Jonathan K. Kallevang
Alexander Kaeck
Stephen D. Grijalva
Thomas J. Douglas
Andrew W. Wilson
Conor B. Garry
Publikationsdatum
27.05.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
Skeletal Radiology / Ausgabe 12/2023
Print ISSN: 0364-2348
Elektronische ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-023-04368-7

Weitere Artikel der Ausgabe 12/2023

Skeletal Radiology 12/2023 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.