Skip to main content
Erschienen in: Advances in Therapy 4/2017

Open Access 11.03.2017 | Review

A Review of the Ahmed Glaucoma Valve Implant and Comparison with Other Surgical Operations

verfasst von: Ivano Riva, Gloria Roberti, Andreas Katsanos, Francesco Oddone, Luciano Quaranta

Erschienen in: Advances in Therapy | Ausgabe 4/2017

Abstract

The Ahmed glaucoma valve (AGV) is a popular glaucoma drainage implant used for the control of intraocular pressure in patients with glaucoma. While in the past AGV implantation was reserved for glaucoma patients poorly controlled after one or more filtration procedures, mounting evidence has recently encouraged its use as a primary surgery in selected cases. AGV has been demonstrated to be safe and effective in reducing intraocular pressure in patients with primary or secondary refractory glaucoma. Compared to other glaucoma surgeries, AGV implantation has shown favorable efficacy and safety. The aim of this article is to review the results of studies directly comparing AGV with other surgical procedures in patients with glaucoma.
Hinweise
Enhanced content To view enhanced content for this article go to http://​www.​medengine.​com/​Redeem/​2297F0607309C4FD​.

Introduction

Glaucoma drainage implants (GDIs) have become a valuable tool for the surgical management of refractory glaucoma [1, 2]. GDIs have been demonstrated to be effective in reducing intraocular pressure (IOP) and are especially indicated in cases with high risk of failure after filtering procedures [3, 4]. Typically, GDIs are indicated for eyes that have already undergone one or more glaucoma surgeries, and for particular types of glaucoma in which filtering procedures are likely to fail (e.g., secondary glaucomas) [58]. The use of GDIs as a primary surgical procedure for glaucoma has also been investigated [9, 10]. Despite some concerns [11], early results are promising [12].
The Ahmed glaucoma valve (AGV), a GDI device, was approved by the FDA in 1993 for use in glaucoma patients with uncontrolled IOP [13]. Unlike the Molteno and Baerveldt implants, AGV provides a complex mechanism to control aqueous humor (AH) flow. Indeed, a valve mechanism has been introduced in an attempt to prevent complications sometimes encountered with non-valved GDIs, such as hypotony, shallow anterior chamber, choroidal effusion and choroidal detachment [14]. Several studies have shown that AGV is safe and effective in reducing IOP in patients with primary [1517], secondary [1824] and refractory glaucoma [2527]. However, the hypotensive efficacy of AGV in comparison with other anti-glaucoma interventions has not been completely elucidated. The aim of the present review is to describe the results of studies directly comparing AGV and other IOP-reducing procedures in patients with glaucoma. This article is based on previously conducted studies and does not involve any new studies of human or animal subjects performed by any of the authors.

Ahmed Glaucoma Valve: Device Description and Technical Data

AGV consists of three parts: (1) a plate which, depending on the model, is composed of medical grade silicone, polypropylene or polyethylene (Medpor; Porex, Atlanta, GA, USA; subsequently Stryker, Kalamazoo, MI, USA), (2) a drainage tube composed of medical grade silicone, and (3) a valve mechanism made of medical-grade silicone (Fig. 1a). Polypropylene is a rigid, inflexible plastic, highly resistant to torsional forces, while silicone is a flexible rubber. Medpor is a porous high-density polyethylene which allows for rapid tissue integration and vascular ingrowth [28].
The adult model (S2) of AGV provides 184 mm2 of total plate area, while the pediatric model (S3) provides a total area of 96 mm2. Obviously, a smaller plate facilitates positioning in infants and subjects with a small eye. A variant of the device with two plates (total filtration area: 364 mm2), and one with a clip for pars plana tube insertion have also been designed. The recently introduced M4 AGV model has a total plate area of 160 mm2 not including the surface area of pores (Table 1).
Table 1
Ahmed glaucoma implants available models
Type
Model
Surface area (mm2)
Material (plate)
Single plate
S2
184
Polypropylene
Single plate, pediatric size
S3
96
Polypropylene
Double plate
B1
364
Polypropylene
Single plate
FP7
184
Silicone
Single plate, pediatric size
FP8
96
Silicone
Double plate
FX1
364
Silicone
Single plate
M4
160
Polyethylene
Pars plana
PS2
184
Polypropylene
Pars plana, pediatric size
PS3
96
Polypropylene
Pars plana
PC7
184
Silicone
Pars plana, pediatric size
PC8
96
Silicone
The valve mechanism of AGV consists of thin silicone elastomer membranes measuring 8 mm in length and 7 mm in width which create a Venturi-type chamber. The membranes are pre-tensioned to open and close so that, at least in principle, flow in response to IOP variations is ensured in the range of 8–12 mmHg [29]. After implantation, AH flows slowly and continuously into the trapezoidal chamber of the valve (Fig. 1b). As the pressure reaches the pre-set threshold value, the valve opens, thus decreasing the IOP. Since the inlet cross-section of the chamber is wider than the outlet, a pressure differential is created across the chamber. This pressure differential enables the valve to remain open even with a small pressure differential between the AC and the subconjunctival spaces surrounding the device. In order for Bernoulli’s equation to be satisfied (fluid flowing into section A = fluid flowing out of section B; Fig. 1b), the velocity of the fluid has to increase as it leaves the chamber through the drainage tube. This increased velocity and the non-obtrusive flow account for better evacuation and smaller valve friction. The tension of the silicone membranes helps to reduce the risk of hypotony by closing the mechanism after the pressure has decreased to the desired level.

Ahmed Glaucoma Valve: Comparison with Other Glaucoma Surgeries

Ahmed Glaucoma Valve Versus Trabeculectomy

Trabeculectomy is recognized as the gold-standard surgery for glaucoma against which other procedures are evaluated [5, 6, 30]. A comparison between AGV and trabeculectomy may be biased by the fact that different perceived indications for each procedure may result in dissimilar groups being compared. For example, GDIs are frequently used as a second-line surgery after trabeculectomy has failed. Thus, when evaluating trials comparing trabeculectomy versus AGV implantation, special attention should be paid to sample characteristics.
Wilson et al. randomized 117 patients with primary open angle glaucoma (POAG), primary angle closure glaucoma (PACG) and secondary glaucoma to trabeculectomy (n = 62) or AGV implantation (n = 55) [31]. Totals of 44 of 62 eyes in the trabeculectomy group and 31 of 55 eyes in the AGV group had not undergone any prior surgical procedure for glaucoma. The mean follow-up time was 9.7 months, with a range of 6–13 months. After 1 year, the cumulative probability of success (IOP <21 mmHg and at least 15% IOP reduction from preoperative level) was 83.6% for the trabeculectomy group and 88.1% for the AGV group (p = 0.43). The AGV group required more frequently adjunctive medical therapy and the trabeculectomy group achieved significantly lower IOP from week 6 to 15 (12.6 vs. 16.4 mmHg) and from month 11 to 13 (11.4 vs. 17.2 mmHg).
Long-term results of AGV implantation versus trabeculectomy as initial surgical management in patients with POAG or PACG were later investigated by the same authors [15]. The study enrolled 123 patients who were followed-up for 31 months. Totals of 64 patients were randomized to trabeculectomy and 59 to AGV. In the first postoperative year, the trabeculectomy group generally obtained lower pressures, but differences became statistically significant in the period between months 11 and 13. Subsequently, the IOP in the two groups became similar through months 41–52, and differences were not significant except from months 34 to 40, when the AGV group achieved lower pressures. When the authors applied the same criteria of success as in their first study, the cumulative probability of success was 68.1% and 69.8% at months 41–52 for the AGV group and the trabeculectomy group, respectively. No difference in the rate of complications was found between the two groups.
Tran et al. compared the efficacy of trabeculectomy and AGV in a group of patients with POAG, exfoliative glaucoma and pigmentary glaucoma [17]. Totals of 61 of 78 patients in the AGV and 61 of 88 patients in the trabeculectomy group had already undergone at least one procedure for glaucoma. Applying the same success criteria as in the previous studies (i.e.: IOP ≤21 mmHg and at least 15% IOP reduction from preoperative level), at 5 years there was no difference in cumulative success between the groups (36% vs. 48% for the AGV and the trabeculectomy group, respectively; p = 0.094). However, when more stringent criteria were used (IOP ≤18, ≤15 and ≤12 mmHg, and a concomitant IOP reduction ≥20, ≥25 and ≥30% from baseline), trabeculectomy had significantly higher long-term probability of success than the AGV implant. The authors concluded that, when great IOP reduction and low IOP levels are needed, trabeculectomy achieves this goal more frequently. Doubts have been raised about the methodology of this retrospective study and in particular about the difference in visual acuity and IOP between the two groups at baseline (which may suggest that one group had less healthy eyes than the other) [32].
AGV and trabeculectomy were retrospectively compared by Shen et al. in a sample of 40 eyes with neovascular glaucoma (NVG) [33]. Surgical success was defined as an IOP ≤21 and ≥6 mmHg, with or without medications, and without the need of further glaucoma surgery. Only one eye (in the AGV group) had previously undergone a surgical procedure for glaucoma. Mean IOP values did not differ between the two groups at each visit follow-up, although the trabeculectomy group required a significantly greater number of medications at months 3 and 6. The respective cumulative success rates for the AGV and trabeculectomy groups were 90% and 85% at 6 months, 70% and 65% at 12 months, 60% and 60% at 18 months, and 60 and 55% at 24 months. No significant difference in the survival curves of the two groups was found (p = 0.815).
The advent of anti-vascular endothelial growth factor (anti-VEGF) drugs and their use as an adjuvant for glaucoma surgery in NVG has dramatically increased the success of trabeculectomy [3436]. In a prospective interventional study, Liu et al. randomized 37 eyes of 36 patients with NVG to AGV implantation (19 patients) or combined trabeculectomy with intravitreal ranibizumab injection (18 patients) [37]. The administration of ranibizumab (0.5 mg) was done 1 week before trabeculectomy. Complete success was defined as an IOP ≥6 and ≤21 mmHg without any antiglaucoma medication or further glaucoma surgery. Partial success was defined as an IOP <21 mmHg with topical antiglaucoma medications. In the combined trabeculectomy and ranibizumab group, 11 eyes (61.1%) achieved complete success, 6 eyes (33.3%) achieved partial success and 1 eye (5.6%) failed. In the AGV group, 11 eyes (57.9%) had complete success, 2 eyes (10.5%) had partial success, and 6 eyes (31.6%) failed. Complications were more common in the AGV group, especially in the early postoperative period (hypotony, shallow anterior chamber, hyphema). Although results from this study are exciting, additional studies with a larger sample size are needed to provide clear evidence about the optimal surgical treatment of NVG. Moreover, a clinical trial comparing the combination of AGV with ranibizumab versus the combination of trabeculectomy with ranibizumab would be useful.

Ahmed Glaucoma Valve Versus Ex-Press Shunts

The Ex-Press Mini Glaucoma Shunt (EXP) is a biocompatible, non-valved device developed by Optonol (Neve Ilan, Israel) to control IOP in glaucoma patients. Initially, EXP was implanted at the limbus so that the external portion of the device was only covered by the conjunctiva. This early technique was associated with a high rate of complications, including hypotony, erosion, extrusion and endophthalmitis [3842]. To avoid these complications, a modified guarded technique was adopted, with the creation of a partial-thickness scleral flap covering the device at the limbus [43]. EXP was developed in an effort to make filtration surgery more straightforward. Nonetheless, it seems that the efficacy and safety profile of the EXP are not superior to standard trabeculectomy [4447]. Only a few studies directly comparing AGV and EXP have been published.
Zhang et al. compared the efficacy and safety of EXP and AGV in a sample of 69 patients with refractory glaucoma (32 implanted with EXP and 37 with AGV) [48]. Qualified surgical success was defined as an IOP ≥5 mmHg and ≤21 mmHg, with or without topical glaucoma medications; complete surgical success was defined as an IOP ≥5 and ≤21 mmHg without any topical glaucoma medications. Baseline IOP was 36.6 ± 9.5 mmHg in the EXP group and 35.4 ± 9.1 mmHg in the AGV group. At the 9-month follow-up, patients implanted with EXP had qualified and complete success rates of 75% and 62.5%, respectively, and patients implanted with AGV had qualified and complete success rates of 62.1% and 51.3%, respectively (p > 0.05). No difference in postoperative complications was found between groups. Results from this study should be interpreted with caution, due to the short length of follow-up, the race of the patients (all patients were Asians) and the lack of information on baseline patient characteristics.
In a recent retrospective study, the efficacy of EXP and AGV was investigated in 64 eyes of 57 glaucoma patients (EXP implanted in 31 eyes and AGV in 33 eyes) [49]. At baseline, the AGV group had higher mean IOP (30.1 ± 10.4 vs. 23.5 ± 7.5 mmHg, p < 0.01), more pseudophakic eyes (72.7% vs. 41.9%, p = 0.02) and a higher number of previous surgeries, including trabeculectomy, vitrectomy and keratoplasty (p < 0.01). Failure was defined as an IOP >21 or ≤5 mmHg on two consecutive postoperative follow-up visits 3 months from surgery, reoperation for glaucoma or loss of light perception. Mean follow-up time was 2.6 ± 1.1 years in the EXP group and 3.3 ± 1.6 years in the AGV group. Failure rates were 16.1% in patients implanted with EXP and 24.2% in patients implanted with AGV (p = 0.696). A higher rate of complications was encountered in the AGV group (60.1%) than in the EXP group (32.3%). The results of this report should be interpreted with caution, due to methodological limitations and baseline differences between the two study groups.
Further trials directly comparing EXP and AGV are needed before more confident conclusions can be drawn. However, it is worth noting that the efficacy of EXP is typically affected by the same factors that limit the efficacy and prognosis of other filtration surgeries. Consequently, at least in theory, EXP may be a less-than-ideal option for secondary glaucomas at high risk of failure in which aggressive conjunctival healing and scarring is anticipated.

Ahmed Glaucoma Valve Versus Other Glaucoma Drainage Implants

Several glaucoma GDIs are currently available. Their characteristics differ in terms of size, shape, plate material and presence of a flow-limiting mechanism. Theoretically, a valve mechanism embedded in the implant may reduce postoperative hypotony-related complications. On the other hand, such a mechanism could limit AH drainage from the eye so that the efficacy of the implant would be compromised. Several comparison studies of AGV versus other implants have been published.

Ahmed Glaucoma Valve Versus Baerveldt Implant

Syed et al. retrospectively evaluated the efficacy of AGV and Baerveldt 350-mm2 implant in refractory glaucoma patients (64 eyes, 32 implanted with AGV and 32 with the Baerveldt device) [50]. Patients were matched for age, race, gender, glaucoma type, previous ocular history and preoperative IOP. A ligature with a 7.0 polyglactin suture was performed only in patients who underwent a Baerveldt implantation. The suture was tied around the tube 1–2 mm from the plate to limit filtration in the immediate postoperative period. During the first year from surgery, both the AGV and the Baerveldt group had a similar IOP profile. There was no significant difference in IOP between the two groups at any visit intervals. Success, defined as IOP decrease of at least 30% from baseline and IOP <22 mmHg with or without medications, was achieved in 25 of the 32 patients (75%) in the Baerveldt group and 24 of the 32 patients (75%) in the AGV group. The same rate of complications was recorded for both devices, with 3 patients (9.4%) in each group having complications that resulted in significant loss of vision or necessitated re-intervention.
Tsai retrospectively compared the efficacy and safety of AGV and Baerveldt implant in his personal single-surgeon series. Both medium-term (1 year) [51] and long-term (4 years) [52] results were reported. Surgical failure was defined as an IOP >21 or <6 mmHg at the last post-operative visit, reduction or loss of vision and additional surgical procedures. Totals of 70 patients underwent Baerveldt implantation (both 250- and 350-mm2 plate models) and 48 patients underwent AGV implantation. Patients implanted with the AGV had lower mean IOP 1 day and 1 week (p < 0.01) after surgery. Afterward, no differences were observed in IOP from 1 to 48 months. No significant difference was found between AGV and Baerveldt implant survival curves (p ≥ 0.05) up to 48 months of follow-up. The 1-, 2-, 3- and 4-year survival rates for AGV and Baerveldt groups were, respectively, 83 and 73% (p = 0.183), and 67% (p = 0.359), 71 and 64% (p = 0.458), and 62 and 64% (p = 0.843). Patients in the Baerveldt group were more likely to develop early postoperative hypotony-related complications and failure, whereas patients in the AGV group were more likely to be on additional glaucoma medications (starting at 18 months post-surgery) and to develop later failure. These results should be interpreted cautiously, considering the retrospective nature of the study and the absence of patient matching at baseline. Patients in the AGV group were older at baseline (69.2 vs. 62.3 years; p = 0.032), were more likely to have a diagnosis of glaucoma associated with inflammation (20.8% vs. 4.3%, p < 0.01) and had a higher pre-operative mean IOP (38.5 vs. 34.6 mmHg; p = 0.032).
The Ahmed versus Baerveldt study (AVB study) was a prospective, multicenter, randomized clinical trial comparing AGV (model FP7) and Baerveldt 350-mm2 implants in patients affected by refractory glaucoma [53]. Surgical procedures were standardized in all the centers involved in the study. No manipulation to limit filtration was performed in the AGV group, while a releasable intraluminal nylon cord or a polyglactin ligature suture was placed in the Baerveldt devices that were implanted. Patients were scheduled to be followed-up until 5 years and failure was prospectively defined as uncontrolled IOP (IOP >18 or <5 mmHg or IOP reduction from baseline <20% in two consecutive visits at or after 3 months), additional glaucoma surgery, or loss of light perception. Between October 2005 and March 2009, 238 patients were randomized to AGV (n = 124) or Baerveldt (n = 114) implantation. The mean number of previous surgical procedures was the same between the two groups. Overall, the mean number of previous interventions was 1.7 ± 1.2, the most common being cataract extraction (172 patients, 72%), trabeculectomy (89 patients, 37%), pars-plana vitrectomy (34 patients, 14%) and penetrating keratoplasty (19 patients, 8%). Only 27 patients had no previous surgery performed (11%) [53]. At 1-year follow-up, failure had occurred in 51 patients (43%) in the AGV group and 30 patients (28%) in the Baerveldt group (p = 0.02) [54]. The cumulative probability of failure was 43% in the AGV group and 28% in the Baerveldt group (p = 0.049). At 3-year follow-up, failure had occurred in 63 patients (51%) in the AGV group and 39 patients (34%) in the Baerveldt group (p = 0.013) [55]. The cumulative probability of failure at 3 years was 51% in the AGV group and 34% in the Baerveldt group (p = 0.03). Regarding the IOP, from month 1 onward, the Baerveldt group had lower IOP values than the AGV group, which reached statistical significance at the 12- and 18-month time-points (p < 0.01, Fig. 2a). Mean IOP at 3 years was 15.7 ± 4.8 mmHg in the AGV group and 14.4 ± 5.1 mmHg in the Baerveldt group. Although both groups had similar complication rates, the Baerveldt group experienced more frequently hypotony-related vision-threatening complications than the AGV group (6 vs. 0%, p < 0.01). The authors concluded that, although both devices are effective in reducing IOP in patients with refractory glaucoma, the Baerveldt implant achieves success in a higher percentage of cases, at the cost of a higher rate of hypotony-related complications.
The Ahmed Baerveldt Comparison study (ABC) was a randomized, multicenter, controlled clinical trial designed to prospectively compare the safety and efficacy of these two commonly implanted GDIs [56]. A total of 276 patients were enrolled by 16 centers worldwide between October 2006 and April 2008. Totals of 143 patients were randomized to AGV (model FP7) and 133 to Baerveldt implantation (350-mm2 plate). All operations were performed by experienced surgeons using the same standardized technique. The tube of the Baerveldt GDI was totally occluded in order to restrict AH flow until encapsulation of the plate was achieved. The primary outcome measure was failure, defined as follows: (1) IOP >21 mmHg or IOP reduction <20% from baseline in two consecutive visits (at least 3 months after surgery); (2) IOP <5 mmHg in two consecutive visits (at least 3 months after surgery); (3) additional glaucoma surgery including the removal of the implant; and (4) loss of light perception. Most common diagnoses at baseline were POAG (40%), NVG (29%), PACG (7%) and uveitic glaucoma (7%). Forty-two percent of patients had previously undergone trabeculectomy, while 20% of patients had no incisional surgery at enrollment. Results of the study were reported at 1, 3 and 5 years [5759]. In the AGV group, IOP was reduced from 31.2 ± 11.2 mmHg at baseline to 15.4 ± 5.5 and 14.3 ± 4.7 mmHg at 1-year and 3-year follow-up, respectively [57, 58]. In the Baerveldt group, IOP was reduced from 31.8 ± 12.5 mmHg at baseline to 13.2 ± 6.8 and 13.1 ± 4.5 mmHg at 1-year and 3-year follow-up, respectively [57, 58]. At 5 years, after censoring for patients who underwent additional surgery or loss of light perception, IOP was reduced from 29.6 ± 10.1 at baseline to 14.7 ± 4.4 mmHg in the AGV group and from 28.3 ± 9.3 at baseline to 12.7 ± 4.5 mmHg in the Baerveldt group [59]. The AGV group achieved lower IOP at 1-day and 1-week visits; however, mean IOP in the Baerveldt group was approximately 1–2 mmHg lower thereafter, except at the 2-year visit (Fig. 2b). The cumulative probability of failure was 44.7% in the AGV group and 39.4% in the Baerveldt group at 5 years (p = 0.65). While the total number of failures was similar, the reasons for failure were different between groups. Failure due to uncontrolled IOP or reoperation was recorded in 46 eyes of the AGV group and in 25 eyes of the Baerveldt group (p < 001). Only 11 eyes experienced persistent hypotony or loss of light perception in the AGV group, compared to 22 eyes in the Baerveldt group. No difference in the rate of complications was found at 5 years [60], with late complications developing in 56 patients in the AGV group and in 67 patients in the Baerveldt group (p = 0.082). However, tube occlusion (p = 0.037) and phthisis bulbi (p = 0.037) occurred more frequently in the Baerveldt group. Other complications resulting in reoperation or vision loss were also more common in the Baerveldt group (p = 0.034).
A recent meta-analysis by Wang et al. compared the efficacy and safety of AGV and Baerveldt GDIs in patients with glaucoma [61]. Ten controlled clinical trials (1048 eyes) were included in this meta-analysis, involving 2 randomized clinical trials and 8 retrospective comparative studies. Short-term and long-term results were analyzed separately. Definition of success was consistent with the original studies, and a criterion of IOP <21 mmHg was adopted. Over the short term (714 eyes), the success rate was 78.6% and 79.7% (OR 0.97, p = 0.90) for the AGV and the Baerveldt groups, respectively, while over the long term (835 eyes), the success rate was 59.2 and 68.4% (OR 0.73, p = 0.04) for the AGV and the Baerveldt groups, respectively. The AGV group had higher IOP values than the Baerveldt group both over the short term (6 studies, 685 eyes, weighted mean difference: 2.12 mmHg; 95% CI 0.72–3.52; p < 0.05) and the long term (7 studies, 659 eyes, weighted mean difference: 1.85 mmHg; 95% CI 0.43–3.28; p = 0.01). However, eyes in the AGV group experienced overall fewer complications (OR 0.67; 95% CI 0.50–0.90; p < 0.01), and especially fewer serious complications (OR 0.57; 95% CI 0.36–0.91, p = 0.02).

Ahmed Glaucoma Valve Versus Molteno and Krupin Implant

In an earlier study, Ayyala et al. compared the efficacy of the double-plate Molteno GDI versus the efficacy of the AGV [62]. In their retrospective, case–control study, 60 advanced glaucoma patients matched for age, diagnosis, and previous surgeries underwent Molteno GDI (30 patients) or AGV implantation (30 patients). Minimum follow-up was 6 months and success was defined as follows: (1) IOP <22 and >4 mmHg; (2) a decrease of no more than 2 lines in visual acuity; and (3) no additional surgery to control IOP. At the 12- and 24-months follow-up, there was no difference in the rate of success between the two groups (73 and 56% for the Molteno group and 60 and 50% for the AGV group, at 12 and 24 months, respectively, p = 0.72). Nevertheless, IOP was significantly lower in patients implanted with the Molteno GDI at both time-points. At 12 and 24 months, mean IOP was 13.36 ± 5.2 and 13.3 ± 5.1 mmHg in the Molteno group, and 16.7 ± 5.6 and 19 ± 5.8 in the AGV group (p = 0.026 and p < 0.01 for the comparisons at 12 and 24 months, respectively).
The efficacy of the AGV and the single-plate Molteno GDI were compared in patients with NVG in a study by Yalvac et al. (65 patients, 38 with AGV and 27 with Molteno GDI) [20]. Patients were matched for age, pre-operative IOP, gender and follow-up. Surgical success was defined as an IOP <22 and >5 mmHg, without additional glaucoma surgery and no loss of light perception. Patients using topical medications for IOP control were not considered failures. The overall success rates of the AGV were 63.2% at 1 year, 56.1% at 2 years, 43.2% at 3 years, 37.8% at 4 years and 25.2% at 5 years. The success rates of the Molteno GDI were 37%, 29.6%, 29.6%, 29.6% and 29.6%, respectively at the 1-, 2-, 3-, 4- and 5-year follow-ups. Although success rates were always higher in the AGV group, the log-rank test indicated no significant difference between survival curves (p = 0.141). Patients implanted with the Molteno GDI had more complications, especially hypotony-related, in the early postoperative phase. Choroidal effusion was encountered in 5 eyes (18.5%) in the Molteno group and 2 eyes (5.3%) in the AGV group, shallow anterior chamber was seen in 4 eyes (14.8%) in the Molteno group and 2 eyes (5.3%) in the AGV group, and 1 case of suprachoroidal hemorrhage was observed in the Molteno group (3.7%).
In a prospective, randomized, multicenter study with a follow-up of 2 years, Nassiri et al. compared the efficacy of the AGV (FP7 model) and the Molteno GDI in a series of patients with refractory glaucoma [63]. The study included 46 eyes randomized to AGV and 46 eyes randomized to Molteno GDI implantation. Treatment success was defined as IOP between 6 and 21 mmHg, while failure was defined as IOP >21 mmHg on maximally tolerated medical therapy or IOP <6 mmHg on 2 consecutive visits. Phthisis bulbi, loss of light perception, devastating complications or reoperation were considered failures. Cumulative probabilities of success at 1 and 2 years were 88% and 82% in the AGV group and 93% and 84% in the Molteno group (log-rank test, p = 0.65). While IOP was significantly lower in the AGV group compared to the Molteno group at postoperative day 1 and week 1, the Molteno group had significantly lower pressures thereafter, up to 24 months. Complication rates were comparable in the two groups and no case of persistent hypotony or devastating complication was encountered.
The efficacy and safety of the double-plate Molteno, the Krupin implant and the AGV were compared in a retrospective study by Taglia et al. [64]. A total of 27 Molteno, 13 Krupin and 13 AGV implantations were included in this study. The patients were matched for age and diagnosis at baseline. However, patients in the AGV group had higher IOP and used more glaucoma medications than the other groups at baseline. Two criteria of success were adopted: (1) IOP >6 and <15 mmHg 3 months after surgery (medications permitted) without additional filtration surgery and without tube removal; and (2) IOP >6 and <21 mmHg 3 months after surgery (medications not permitted) without additional filtration surgery and without tube removal. The Molteno group was significantly more likely than the AGV group to satisfy the first criterion of success (p < 0.01, log-rank test), while a similar but statistically non-significant tendency was observed with the second criterion of success. After 1 year, success rates according to the first and the second criterion were, respectively, 80% and 57% for the Molteno group, 39% and 46% for the Krupin group and 35% and 25% for the AGV group. None of the patients in the AGV group had complications requiring reintervention. Thirteen of 27 patients (48%) in the Molteno group required additional surgery, in most cases during the first 3 months of follow-up (11 of 13 patients). Complications requiring re-intervention were generally related to hypotony despite the temporary ligation of the tube. Seven of 13 patients (54%) in the Krupin group had complications requiring re-intervention; 5 of these complications were hypotony-related. The results of this report should be interpreted with caution due to the small sample size and the retrospective nature of the study.

Ahmed Glaucoma Valve Versus Cyclodestructive Interventions

Cyclodestructive procedures have traditionally been reserved for eyes with refractory glaucoma and limited visual potential [6568]. Immediate and late complications of these procedures include pain, visual acuity reduction, corneal edema, persistent hypotony and phthisis bulbi [69, 70]. An endoscopic technique which allows direct visualization and laser treatment of the ciliary body has been described [71]. Such an endoscopic cyclophotocoagulation (ECP) device combines a laser source, an endoscope, and an illumination beam in the same probe [68, 70, 72]. The technique appears promising in terms of efficacy and safety [68, 70, 73].
In a study by Yildirim et al., 66 patients with NVG and visual acuity of hand motion or worse were randomized to AGV or transcleral diode cyclophotocoagulation [69]. A minimum 24-month follow-up was available for 58 of the 66 patients. Success was defined as an IOP <21 and >5 mmHg without additional glaucoma surgery and without loss of light perception. Post-operative medications were allowed and not considered a failure. Preoperative IOP was 43.4 ± 11.9 mmHg in the cyclophotocoagulation group and 43.3 ± 7.4 mmHg in the AGV group (p > 0.05). At 24 months, the IOP decreased to 18.72 ± 13.5 and 22.88 ± 7.3 mmHg in the transcleral cyclophotocoagulation and AGV groups, respectively (p > 0.05). Success rates at 24 months were similar in the two groups (61.18% and 59.26% for the cyclophotocoagulation and AGV groups, p > 0.05). The most frequent complications in the cyclodestructive group were anterior segment inflammation (5 eyes, 20%), neurotrophic keratitis (2 eyes, 8%) and hypotony (3 eyes, 12%). In the AGV group, hyphema was encountered in 5 eyes (15%) and tube occlusion in 3 eyes (9%). No devastating complications were noted in either group.
The safety and efficacy of ECP through the pars plana versus AGV were compared in a prospective study by Lima et al. [74]. Sixty-eight patients were randomized to ECP (n = 34) or AGV implantation (n = 34). To be eligible for inclusion patients had to be pseudophakic with IOP ≥35 mmHg on maximum tolerated medical therapy, have undergone at least 1 previous trabeculectomy, and have visual acuity better than light perception. Success was defined as IOP >6 and <21 mmHg at 24 months with or without medications. Preoperative IOP was 41.32 ± 2.03 mmHg in the AGV group and 41.61 ± 3.42 in the ECP group. At the 24-month follow-up, the IOP was reduced to 14.73 ± 6.44 and 14.07 ± 7.21 in the AGV and ECP groups, respectively (p = 0.07). Success at 24 months was similar in the two groups: 70.58% in the AGV group and 73.52% in the ECP group (p = 0.5). Interestingly, eyes in the AGV group had a greater incidence of visual acuity decline at 24 months. In the AGV group, a higher rate of complications was encountered, including choroidal detachment (17.64%) and shallow anterior chamber (17.64%). It should be noted that, in this study, laser treatment was performed along the full length of the ciliary processes and the anterior third of the pars plana over 210°. This rather extensive coagulation may have accounted for the favorable ECP results. Further studies with more patients and longer follow-up are needed to clarify the role of ECP in the treatment of patients with refractory glaucoma.

Conclusion

The role of GDIs has been traditionally limited to patients with refractory glaucoma and patients with high risk of failure after standard filtering surgeries. The existing evidence seems to suggest that AGV might have a more favorable safety profile when compared to non-valved GDIs [20, 51, 52, 54, 55, 60, 61, 64]. On the other hand, non-valved GDIs seem to achieve lower pressures than the AGV [54, 55, 5759, 6264]. AGV is a useful surgical choice and represents a reasonable compromise between efficacy and potential surgery-related complications. Due to its safety profile, AGV may be considered as the initial surgery in those glaucomas that are more prone to failure after standard filtering procedures. However, even in these cases, the risk–benefit ratio has to be carefully weighted and discussed with the patient. In conclusion, the existing literature suggests that the currently available AGV is probably the GDI that offers the most favorable risk-efficacy profile.

Acknowledgements

The contribution of the IRCCS Fondazione GB Bietti per l’Oftalmologia in this paper was supported by Ministry of Health and Fondazione Roma. All named authors meet the International Committee of Medical Journal Editors (ICMJE) criteria for authorship for this manuscript, take responsibility for the integrity of the work as a whole, and have given final approval to the version to be published.

Disclosures

Ivano Riva, Gloria Roberti, Andreas Katsanos, Francesco Oddone and Luciano Quaranta declare that they have nothing to disclose.

Compliance with Ethics Guidelines

This article is based on previously conducted studies, and does not involve any new studies of human or animal subjects performed by any of the authors.

Data Availability

Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study.

Open Access

This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://​creativecommons.​org/​licenses/​by-nc/​4.​0/​), which permits any noncommercial use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://​creativecommons.​org/​licenses/​by/​4.​0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

e.Med Allgemeinmedizin

Kombi-Abonnement

Mit e.Med Allgemeinmedizin erhalten Sie Zugang zu allen CME-Fortbildungen und Premium-Inhalten der allgemeinmedizinischen Zeitschriften, inklusive einer gedruckten Allgemeinmedizin-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Minckler DS, Vedula SS, Li TJ et al. Aqueous shunts for glaucoma. Cochrane Database Syst Rev. 2006:CD004918. Minckler DS, Vedula SS, Li TJ et al. Aqueous shunts for glaucoma. Cochrane Database Syst Rev. 2006:CD004918.
2.
Zurück zum Zitat Assaad MH, Baerveldt G, Rockwood EJ. Glaucoma drainage devices: pros and cons. Curr Opin Ophthalmol. 1999;10:147–53.CrossRefPubMed Assaad MH, Baerveldt G, Rockwood EJ. Glaucoma drainage devices: pros and cons. Curr Opin Ophthalmol. 1999;10:147–53.CrossRefPubMed
3.
Zurück zum Zitat Gedde SJ, Schiffman JC, Feuer WJ, et al. Treatment outcomes in the Tube Versus Trabeculectomy (TVT) study after five years of follow-up. Am J Ophthalmol. 2012;153(789–803):e782. Gedde SJ, Schiffman JC, Feuer WJ, et al. Treatment outcomes in the Tube Versus Trabeculectomy (TVT) study after five years of follow-up. Am J Ophthalmol. 2012;153(789–803):e782.
4.
Zurück zum Zitat Huang MC, Netland PA, Coleman AL, et al. Intermediate-term clinical experience with the Ahmed glaucoma valve implant. Am J Ophthalmol. 1999;127:27–33.CrossRefPubMed Huang MC, Netland PA, Coleman AL, et al. Intermediate-term clinical experience with the Ahmed glaucoma valve implant. Am J Ophthalmol. 1999;127:27–33.CrossRefPubMed
5.
Zurück zum Zitat Prum BE Jr, Rosenberg LF, Gedde SJ, et al. Primary open-angle glaucoma preferred practice pattern((R)) guidelines. Ophthalmology. 2016;123:P41–111.CrossRefPubMed Prum BE Jr, Rosenberg LF, Gedde SJ, et al. Primary open-angle glaucoma preferred practice pattern((R)) guidelines. Ophthalmology. 2016;123:P41–111.CrossRefPubMed
6.
Zurück zum Zitat European Glaucoma Society. Terminology and guidelines for glaucoma. 4th ed. Savona: European Glaucoma Society; 2014. European Glaucoma Society. Terminology and guidelines for glaucoma. 4th ed. Savona: European Glaucoma Society; 2014.
7.
Zurück zum Zitat Chow K, Mora J. Practice preferences for glaucoma drainage device implantation and cyclodestruction in Australia and New Zealand. J Glaucoma. 2012;21:199–205.CrossRefPubMed Chow K, Mora J. Practice preferences for glaucoma drainage device implantation and cyclodestruction in Australia and New Zealand. J Glaucoma. 2012;21:199–205.CrossRefPubMed
8.
Zurück zum Zitat Joshi AB, Parrish RK 2nd, Feuer WF. 2002 survey of the American Glaucoma Society: practice preferences for glaucoma surgery and antifibrotic use. J Glaucoma. 2005;14:172–4.CrossRefPubMed Joshi AB, Parrish RK 2nd, Feuer WF. 2002 survey of the American Glaucoma Society: practice preferences for glaucoma surgery and antifibrotic use. J Glaucoma. 2005;14:172–4.CrossRefPubMed
9.
Zurück zum Zitat Molteno AC, Bevin TH, Herbison P, Husni MA. Long-term results of primary trabeculectomies and Molteno implants for primary open-angle glaucoma. Arch Ophthalmol. 2011;129:1444–50.CrossRefPubMed Molteno AC, Bevin TH, Herbison P, Husni MA. Long-term results of primary trabeculectomies and Molteno implants for primary open-angle glaucoma. Arch Ophthalmol. 2011;129:1444–50.CrossRefPubMed
10.
Zurück zum Zitat Gedde SJ, Singh K, Schiffman JC, Feuer WJ. Tube versus trabeculectomy study G. The tube versus trabeculectomy study: interpretation of results and application to clinical practice. Curr Opin Ophthalmol. 2012;23:118–26.CrossRefPubMed Gedde SJ, Singh K, Schiffman JC, Feuer WJ. Tube versus trabeculectomy study G. The tube versus trabeculectomy study: interpretation of results and application to clinical practice. Curr Opin Ophthalmol. 2012;23:118–26.CrossRefPubMed
11.
Zurück zum Zitat Caprioli J. The tube versus trabeculectomy study: why its findings may not change clinical practice? Am J Ophthalmol. 2011;151(742–744):e741. Caprioli J. The tube versus trabeculectomy study: why its findings may not change clinical practice? Am J Ophthalmol. 2011;151(742–744):e741.
12.
Zurück zum Zitat Nguyen QH. Primary surgical management refractory glaucoma: tubes as initial surgery. Curr Opin Ophthalmol. 2009;20:122–5.CrossRefPubMed Nguyen QH. Primary surgical management refractory glaucoma: tubes as initial surgery. Curr Opin Ophthalmol. 2009;20:122–5.CrossRefPubMed
13.
Zurück zum Zitat Coleman AL, Hill R, Wilson MR, et al. Initial clinical experience with the Ahmed glaucoma valve implant. Am J Ophthalmol. 1995;120:23–31.CrossRefPubMed Coleman AL, Hill R, Wilson MR, et al. Initial clinical experience with the Ahmed glaucoma valve implant. Am J Ophthalmol. 1995;120:23–31.CrossRefPubMed
14.
Zurück zum Zitat Patel S, Pasquale LR. Glaucoma drainage devices: a review of the past, present, and future. Semin Ophthalmol. 2010;25:265–70.CrossRefPubMed Patel S, Pasquale LR. Glaucoma drainage devices: a review of the past, present, and future. Semin Ophthalmol. 2010;25:265–70.CrossRefPubMed
15.
Zurück zum Zitat Wilson MR, Mendis U, Paliwal A, Haynatzka V. Long-term follow-up of primary glaucoma surgery with Ahmed glaucoma valve implant versus trabeculectomy. Am J Ophthalmol. 2003;136:464–70.CrossRefPubMed Wilson MR, Mendis U, Paliwal A, Haynatzka V. Long-term follow-up of primary glaucoma surgery with Ahmed glaucoma valve implant versus trabeculectomy. Am J Ophthalmol. 2003;136:464–70.CrossRefPubMed
16.
Zurück zum Zitat Quaranta L, Riva I, Floriani IC. Outcomes of using a sutureless bovine pericardial patch graft for Ahmed glaucoma valve implantation. Eur J Ophthalmol. 2013;23:738–42.CrossRefPubMed Quaranta L, Riva I, Floriani IC. Outcomes of using a sutureless bovine pericardial patch graft for Ahmed glaucoma valve implantation. Eur J Ophthalmol. 2013;23:738–42.CrossRefPubMed
17.
Zurück zum Zitat Tran DH, Souza C, Ang MJ, et al. Comparison of long-term surgical success of Ahmed valve implant versus trabeculectomy in open-angle glaucoma. Br J Ophthalmol. 2009;93:1504–9.CrossRefPubMed Tran DH, Souza C, Ang MJ, et al. Comparison of long-term surgical success of Ahmed valve implant versus trabeculectomy in open-angle glaucoma. Br J Ophthalmol. 2009;93:1504–9.CrossRefPubMed
18.
Zurück zum Zitat Da Mata A, Burk SE, Netland PA, et al. Management of uveitic glaucoma with Ahmed glaucoma valve implantation. Ophthalmology. 1999;106:2168–72.CrossRefPubMed Da Mata A, Burk SE, Netland PA, et al. Management of uveitic glaucoma with Ahmed glaucoma valve implantation. Ophthalmology. 1999;106:2168–72.CrossRefPubMed
19.
Zurück zum Zitat Papadaki TG, Zacharopoulos IP, Pasquale LR, et al. Long-term results of Ahmed glaucoma valve implantation for uveitic glaucoma. Am J Ophthalmol. 2007;144:62–9.CrossRefPubMed Papadaki TG, Zacharopoulos IP, Pasquale LR, et al. Long-term results of Ahmed glaucoma valve implantation for uveitic glaucoma. Am J Ophthalmol. 2007;144:62–9.CrossRefPubMed
20.
Zurück zum Zitat Yalvac IS, Eksioglu U, Satana B, Duman S. Long-term results of Ahmed glaucoma valve and Molteno implant in neovascular glaucoma. Eye (Lond). 2007;21:65–70.CrossRefPubMed Yalvac IS, Eksioglu U, Satana B, Duman S. Long-term results of Ahmed glaucoma valve and Molteno implant in neovascular glaucoma. Eye (Lond). 2007;21:65–70.CrossRefPubMed
21.
Zurück zum Zitat Netland PA, Ishida K, Boyle JW. The Ahmed glaucoma valve in patients with and without neovascular glaucoma. J Glaucoma. 2010;19:581–6.CrossRefPubMed Netland PA, Ishida K, Boyle JW. The Ahmed glaucoma valve in patients with and without neovascular glaucoma. J Glaucoma. 2010;19:581–6.CrossRefPubMed
22.
Zurück zum Zitat Ma KT, Yang JY, Kim JH, et al. Surgical results of Ahmed valve implantation with intraoperative bevacizumab injection in patients with neovascular glaucoma. J Glaucoma. 2012;21:331–6.CrossRefPubMed Ma KT, Yang JY, Kim JH, et al. Surgical results of Ahmed valve implantation with intraoperative bevacizumab injection in patients with neovascular glaucoma. J Glaucoma. 2012;21:331–6.CrossRefPubMed
23.
Zurück zum Zitat Ishida K, Ahmed II, Netland PA. Ahmed glaucoma valve surgical outcomes in eyes with and without silicone oil endotamponade. J Glaucoma. 2009;18:325–30.CrossRefPubMed Ishida K, Ahmed II, Netland PA. Ahmed glaucoma valve surgical outcomes in eyes with and without silicone oil endotamponade. J Glaucoma. 2009;18:325–30.CrossRefPubMed
24.
Zurück zum Zitat Gupta S, Chaurasia AK, Chawla R, et al. Long-term outcomes of glaucoma drainage devices for glaucoma post-vitreoretinal surgery with silicone oil insertion: a prospective evaluation. Graefes Arch Clin Exp Ophthalmol. 2016;254:2449–54.CrossRefPubMed Gupta S, Chaurasia AK, Chawla R, et al. Long-term outcomes of glaucoma drainage devices for glaucoma post-vitreoretinal surgery with silicone oil insertion: a prospective evaluation. Graefes Arch Clin Exp Ophthalmol. 2016;254:2449–54.CrossRefPubMed
25.
Zurück zum Zitat Das JC, Chaudhuri Z, Sharma P, Bhomaj S. The Ahmed glaucoma valve in refractory glaucoma: experiences in Indian eyes. Eye (Lond). 2005;19:183–90.CrossRef Das JC, Chaudhuri Z, Sharma P, Bhomaj S. The Ahmed glaucoma valve in refractory glaucoma: experiences in Indian eyes. Eye (Lond). 2005;19:183–90.CrossRef
26.
Zurück zum Zitat Souza C, Tran DH, Loman J, et al. Long-term outcomes of Ahmed glaucoma valve implantation in refractory glaucomas. Am J Ophthalmol. 2007;144:893–900.CrossRefPubMed Souza C, Tran DH, Loman J, et al. Long-term outcomes of Ahmed glaucoma valve implantation in refractory glaucomas. Am J Ophthalmol. 2007;144:893–900.CrossRefPubMed
27.
Zurück zum Zitat Wishart PK, Choudhary A, Wong D. Ahmed glaucoma valves in refractory glaucoma: a 7-year audit. Br J Ophthalmol. 2010;94:1174–9.CrossRefPubMed Wishart PK, Choudhary A, Wong D. Ahmed glaucoma valves in refractory glaucoma: a 7-year audit. Br J Ophthalmol. 2010;94:1174–9.CrossRefPubMed
28.
Zurück zum Zitat Kim J, Allingham RR, Hall J, et al. Clinical experience with a novel glaucoma drainage implant. J Glaucoma. 2014;23:e91–7.CrossRefPubMed Kim J, Allingham RR, Hall J, et al. Clinical experience with a novel glaucoma drainage implant. J Glaucoma. 2014;23:e91–7.CrossRefPubMed
29.
Zurück zum Zitat Coleman AL, Smyth RJ, Wilson MR, Tam M. Initial clinical experience with the Ahmed glaucoma valve implant in pediatric patients. Arch Ophthalmol. 1997;115:186–91.CrossRefPubMed Coleman AL, Smyth RJ, Wilson MR, Tam M. Initial clinical experience with the Ahmed glaucoma valve implant in pediatric patients. Arch Ophthalmol. 1997;115:186–91.CrossRefPubMed
30.
Zurück zum Zitat Japanese Glaucoma Society. Guidelines for glaucoma. 2nd ed. Tokyo: Japan Glaucoma Society; 2006. Japanese Glaucoma Society. Guidelines for glaucoma. 2nd ed. Tokyo: Japan Glaucoma Society; 2006.
31.
Zurück zum Zitat Wilson MR, Mendis U, Smith SD, Paliwal A. Ahmed glaucoma valve implant vs trabeculectomy in the surgical treatment of glaucoma: a randomized clinical trial. Am J Ophthalmol. 2000;130:267–73.CrossRefPubMed Wilson MR, Mendis U, Smith SD, Paliwal A. Ahmed glaucoma valve implant vs trabeculectomy in the surgical treatment of glaucoma: a randomized clinical trial. Am J Ophthalmol. 2000;130:267–73.CrossRefPubMed
32.
Zurück zum Zitat Mahroo OA, Stanbury R, Lim KS. Were the groups in the trabeculectomy versus Ahmed valve study really comparable? Br J Ophthalmol. 2010;94:1551–2.CrossRefPubMed Mahroo OA, Stanbury R, Lim KS. Were the groups in the trabeculectomy versus Ahmed valve study really comparable? Br J Ophthalmol. 2010;94:1551–2.CrossRefPubMed
33.
Zurück zum Zitat Shen CC, Salim S, Du H, Netland PA. Trabeculectomy versus Ahmed glaucoma valve implantation in neovascular glaucoma. Clin Ophthalmol. 2011;5:281–6.PubMedPubMedCentral Shen CC, Salim S, Du H, Netland PA. Trabeculectomy versus Ahmed glaucoma valve implantation in neovascular glaucoma. Clin Ophthalmol. 2011;5:281–6.PubMedPubMedCentral
34.
Zurück zum Zitat Simha A, Braganza A, Abraham L, Samuel P, Lindsley K. Anti-vascular endothelial growth factor for neovascular glaucoma. Cochrane Database Syst Rev. 2013:CD007920. Simha A, Braganza A, Abraham L, Samuel P, Lindsley K. Anti-vascular endothelial growth factor for neovascular glaucoma. Cochrane Database Syst Rev. 2013:CD007920.
35.
Zurück zum Zitat Elmekawey H, Khafagy A. Intracameral ranibizumab and subsequent mitomycin C augmented trabeculectomy in neovascular glaucoma. J Glaucoma. 2014;23:437–40.CrossRefPubMed Elmekawey H, Khafagy A. Intracameral ranibizumab and subsequent mitomycin C augmented trabeculectomy in neovascular glaucoma. J Glaucoma. 2014;23:437–40.CrossRefPubMed
36.
Zurück zum Zitat Kitnarong N, Sriyakul C, Chinwattanakul S. A prospective study to evaluate intravitreous ranibizumab as adjunctive treatment for trabeculectomy in neovascular glaucoma. Ophthalmol Ther. 2015;4:33–41.CrossRefPubMedPubMedCentral Kitnarong N, Sriyakul C, Chinwattanakul S. A prospective study to evaluate intravitreous ranibizumab as adjunctive treatment for trabeculectomy in neovascular glaucoma. Ophthalmol Ther. 2015;4:33–41.CrossRefPubMedPubMedCentral
37.
Zurück zum Zitat Liu L, Xu Y, Huang Z, Wang X. Intravitreal ranibizumab injection combined trabeculectomy versus Ahmed valve surgery in the treatment of neovascular glaucoma: assessment of efficacy and complications. BMC Ophthalmol. 2016;16:65.CrossRefPubMedPubMedCentral Liu L, Xu Y, Huang Z, Wang X. Intravitreal ranibizumab injection combined trabeculectomy versus Ahmed valve surgery in the treatment of neovascular glaucoma: assessment of efficacy and complications. BMC Ophthalmol. 2016;16:65.CrossRefPubMedPubMedCentral
38.
Zurück zum Zitat Gandolfi S, Traverso CF, Bron A, et al. Short-term results of a miniature draining implant for glaucoma in combined surgery with phacoemulsification. Acta Ophthalmol Scand Suppl. 2002;236:66.CrossRefPubMed Gandolfi S, Traverso CF, Bron A, et al. Short-term results of a miniature draining implant for glaucoma in combined surgery with phacoemulsification. Acta Ophthalmol Scand Suppl. 2002;236:66.CrossRefPubMed
39.
Zurück zum Zitat Traverso CE, De Feo F, Messas-Kaplan A, et al. Long term effect on IOP of a stainless steel glaucoma drainage implant (Ex-PRESS) in combined surgery with phacoemulsification. Br J Ophthalmol. 2005;89:425–9.CrossRefPubMedPubMedCentral Traverso CE, De Feo F, Messas-Kaplan A, et al. Long term effect on IOP of a stainless steel glaucoma drainage implant (Ex-PRESS) in combined surgery with phacoemulsification. Br J Ophthalmol. 2005;89:425–9.CrossRefPubMedPubMedCentral
40.
Zurück zum Zitat Wamsley S, Moster MR, Rai S, Alvim HS, Fontanarosa J. Results of the use of the Ex-PRESS miniature glaucoma implant in technically challenging, advanced glaucoma cases: a clinical pilot study. Am J Ophthalmol. 2004;138:1049–51.CrossRefPubMed Wamsley S, Moster MR, Rai S, Alvim HS, Fontanarosa J. Results of the use of the Ex-PRESS miniature glaucoma implant in technically challenging, advanced glaucoma cases: a clinical pilot study. Am J Ophthalmol. 2004;138:1049–51.CrossRefPubMed
41.
Zurück zum Zitat Stewart RM, Diamond JG, Ashmore ED, Ayyala RS. Complications following ex-press glaucoma shunt implantation. Am J Ophthalmol. 2005;140:340–1.CrossRefPubMed Stewart RM, Diamond JG, Ashmore ED, Ayyala RS. Complications following ex-press glaucoma shunt implantation. Am J Ophthalmol. 2005;140:340–1.CrossRefPubMed
42.
Zurück zum Zitat Stein JD, Herndon LW, Brent Bond J, Challa P. Exposure of Ex-PRESS miniature glaucoma devices: case series and technique for tube shunt removal. J Glaucoma. 2007;16:704–6.CrossRefPubMed Stein JD, Herndon LW, Brent Bond J, Challa P. Exposure of Ex-PRESS miniature glaucoma devices: case series and technique for tube shunt removal. J Glaucoma. 2007;16:704–6.CrossRefPubMed
43.
Zurück zum Zitat Dahan E, Carmichael TR. Implantation of a miniature glaucoma device under a scleral flap. J Glaucoma. 2005;14:98–102.CrossRefPubMed Dahan E, Carmichael TR. Implantation of a miniature glaucoma device under a scleral flap. J Glaucoma. 2005;14:98–102.CrossRefPubMed
44.
Zurück zum Zitat Buys YM. Trabeculectomy with ExPRESS: weighing the benefits and cost. Curr Opin Ophthalmol. 2013;24:111–8.CrossRefPubMed Buys YM. Trabeculectomy with ExPRESS: weighing the benefits and cost. Curr Opin Ophthalmol. 2013;24:111–8.CrossRefPubMed
45.
Zurück zum Zitat Wagschal LD, Trope GE, Jinapriya D, Jin YP, Buys YM. Prospective randomized study comparing Ex-PRESS to trabeculectomy: 1-year results. J Glaucoma. 2015;24:624–9.CrossRefPubMed Wagschal LD, Trope GE, Jinapriya D, Jin YP, Buys YM. Prospective randomized study comparing Ex-PRESS to trabeculectomy: 1-year results. J Glaucoma. 2015;24:624–9.CrossRefPubMed
46.
Zurück zum Zitat Gonzalez-Rodriguez JM, Trope GE, Drori-Wagschal L, Jinapriya D, Buys YM. Comparison of trabeculectomy versus Ex-PRESS: 3-year follow-up. Br J Ophthalmol. 2016;100:1269–73.CrossRefPubMed Gonzalez-Rodriguez JM, Trope GE, Drori-Wagschal L, Jinapriya D, Buys YM. Comparison of trabeculectomy versus Ex-PRESS: 3-year follow-up. Br J Ophthalmol. 2016;100:1269–73.CrossRefPubMed
47.
Zurück zum Zitat Patel HY, Wagschal LD, Trope GE, Buys YM. Economic analysis of the Ex-PRESS miniature glaucoma device versus trabeculectomy. J Glaucoma. 2014;23:385–90.CrossRefPubMed Patel HY, Wagschal LD, Trope GE, Buys YM. Economic analysis of the Ex-PRESS miniature glaucoma device versus trabeculectomy. J Glaucoma. 2014;23:385–90.CrossRefPubMed
48.
Zurück zum Zitat Zhang M, Li B, Sun Y. EX-PRESS and ahmed glaucoma valve in treatment of refractory glaucoma. Acta Ophthalmol. 2016;94:e382–3.CrossRefPubMed Zhang M, Li B, Sun Y. EX-PRESS and ahmed glaucoma valve in treatment of refractory glaucoma. Acta Ophthalmol. 2016;94:e382–3.CrossRefPubMed
49.
Zurück zum Zitat Waisbourd M, Fischer N, Shalev H, et al. Trabeculectomy with Ex-PRESS implant versus Ahmed glaucoma valve implantation-a comparative study. Int J Ophthalmol. 2016;9:1415–20.PubMedPubMedCentral Waisbourd M, Fischer N, Shalev H, et al. Trabeculectomy with Ex-PRESS implant versus Ahmed glaucoma valve implantation-a comparative study. Int J Ophthalmol. 2016;9:1415–20.PubMedPubMedCentral
50.
Zurück zum Zitat Syed HM, Law SK, Nam SH, et al. Baerveldt-350 implant versus Ahmed valve for refractory glaucoma: a case-controlled comparison. J Glaucoma. 2004;13:38–45.CrossRefPubMed Syed HM, Law SK, Nam SH, et al. Baerveldt-350 implant versus Ahmed valve for refractory glaucoma: a case-controlled comparison. J Glaucoma. 2004;13:38–45.CrossRefPubMed
51.
Zurück zum Zitat Tsai JC, Johnson CC, Dietrich MS. The Ahmed shunt versus the Baerveldt shunt for refractory glaucoma: a single-surgeon comparison of outcome. Ophthalmology. 2003;110:1814–21.CrossRefPubMed Tsai JC, Johnson CC, Dietrich MS. The Ahmed shunt versus the Baerveldt shunt for refractory glaucoma: a single-surgeon comparison of outcome. Ophthalmology. 2003;110:1814–21.CrossRefPubMed
52.
Zurück zum Zitat Tsai JC, Johnson CC, Kammer JA, Dietrich MS. The Ahmed shunt versus the Baerveldt shunt for refractory glaucoma II: longer-term outcomes from a single surgeon. Ophthalmology. 2006;113:913–7.CrossRefPubMed Tsai JC, Johnson CC, Kammer JA, Dietrich MS. The Ahmed shunt versus the Baerveldt shunt for refractory glaucoma II: longer-term outcomes from a single surgeon. Ophthalmology. 2006;113:913–7.CrossRefPubMed
53.
Zurück zum Zitat Christakis PG, Tsai JC, Zurakowski D, et al. The Ahmed versus Baerveldt study: design, baseline patient characteristics, and intraoperative complications. Ophthalmology. 2011;118:2172–9.CrossRefPubMed Christakis PG, Tsai JC, Zurakowski D, et al. The Ahmed versus Baerveldt study: design, baseline patient characteristics, and intraoperative complications. Ophthalmology. 2011;118:2172–9.CrossRefPubMed
54.
Zurück zum Zitat Christakis PG, Kalenak JW, Zurakowski D, et al. The Ahmed versus Baerveldt study: one-year treatment outcomes. Ophthalmology. 2011;118:2180–9.CrossRefPubMed Christakis PG, Kalenak JW, Zurakowski D, et al. The Ahmed versus Baerveldt study: one-year treatment outcomes. Ophthalmology. 2011;118:2180–9.CrossRefPubMed
55.
Zurück zum Zitat Christakis PG, Tsai JC, Kalenak JW, et al. The Ahmed versus Baerveldt study: three-year treatment outcomes. Ophthalmology. 2013;120:2232–40.CrossRefPubMed Christakis PG, Tsai JC, Kalenak JW, et al. The Ahmed versus Baerveldt study: three-year treatment outcomes. Ophthalmology. 2013;120:2232–40.CrossRefPubMed
56.
Zurück zum Zitat Barton K, Gedde SJ, Budenz DL, et al. The Ahmed Baerveldt comparison study methodology, baseline patient characteristics, and intraoperative complications. Ophthalmology. 2011;118:435–42.CrossRefPubMed Barton K, Gedde SJ, Budenz DL, et al. The Ahmed Baerveldt comparison study methodology, baseline patient characteristics, and intraoperative complications. Ophthalmology. 2011;118:435–42.CrossRefPubMed
57.
Zurück zum Zitat Budenz DL, Barton K, Feuer WJ, et al. Treatment outcomes in the Ahmed Baerveldt Comparison Study after 1 year of follow-up. Ophthalmology. 2011;118:443–52.CrossRefPubMed Budenz DL, Barton K, Feuer WJ, et al. Treatment outcomes in the Ahmed Baerveldt Comparison Study after 1 year of follow-up. Ophthalmology. 2011;118:443–52.CrossRefPubMed
58.
Zurück zum Zitat Barton K, Feuer WJ, Budenz DL, et al. Three-year treatment outcomes in the Ahmed Baerveldt comparison study. Ophthalmology. 2014;121(1547–1557):e1541. Barton K, Feuer WJ, Budenz DL, et al. Three-year treatment outcomes in the Ahmed Baerveldt comparison study. Ophthalmology. 2014;121(1547–1557):e1541.
59.
Zurück zum Zitat Budenz DL, Barton K, Gedde SJ, et al. Five-year treatment outcomes in the Ahmed Baerveldt comparison study. Ophthalmology. 2015;122:308–16.CrossRefPubMed Budenz DL, Barton K, Gedde SJ, et al. Five-year treatment outcomes in the Ahmed Baerveldt comparison study. Ophthalmology. 2015;122:308–16.CrossRefPubMed
60.
Zurück zum Zitat Budenz DL, Feuer WJ, Barton K, et al. Postoperative complications in the Ahmed Baerveldt comparison study during five years of follow-up. Am J Ophthalmol. 2016;163(75–82):e73. Budenz DL, Feuer WJ, Barton K, et al. Postoperative complications in the Ahmed Baerveldt comparison study during five years of follow-up. Am J Ophthalmol. 2016;163(75–82):e73.
61.
Zurück zum Zitat Wang YW, Wang PB, Zeng C, Xia XB. Comparison of the Ahmed glaucoma valve with the Baerveldt glaucoma implant: a meta-analysis. BMC Ophthalmol. 2015;15:132.CrossRefPubMedPubMedCentral Wang YW, Wang PB, Zeng C, Xia XB. Comparison of the Ahmed glaucoma valve with the Baerveldt glaucoma implant: a meta-analysis. BMC Ophthalmol. 2015;15:132.CrossRefPubMedPubMedCentral
62.
Zurück zum Zitat Ayyala RS, Zurakowski D, Monshizadeh R, et al. Comparison of double-plate Molteno and Ahmed glaucoma valve in patients with advanced uncontrolled glaucoma. Ophthalmic Surg Lasers. 2002;33:94–101.PubMed Ayyala RS, Zurakowski D, Monshizadeh R, et al. Comparison of double-plate Molteno and Ahmed glaucoma valve in patients with advanced uncontrolled glaucoma. Ophthalmic Surg Lasers. 2002;33:94–101.PubMed
63.
Zurück zum Zitat Nassiri N, Kamali G, Rahnavardi M, et al. Ahmed glaucoma valve and single-plate Molteno implants in treatment of refractory glaucoma: a comparative study. Am J Ophthalmol. 2010;149:893–902.CrossRefPubMed Nassiri N, Kamali G, Rahnavardi M, et al. Ahmed glaucoma valve and single-plate Molteno implants in treatment of refractory glaucoma: a comparative study. Am J Ophthalmol. 2010;149:893–902.CrossRefPubMed
64.
Zurück zum Zitat Taglia DP, Perkins TW, Gangnon R, Heatley GA, Kaufman PL. Comparison of the Ahmed Glaucoma valve, the Krupin eye valve with disk, and the double-plate Molteno implant. J Glaucoma. 2002;11:347–53.CrossRefPubMed Taglia DP, Perkins TW, Gangnon R, Heatley GA, Kaufman PL. Comparison of the Ahmed Glaucoma valve, the Krupin eye valve with disk, and the double-plate Molteno implant. J Glaucoma. 2002;11:347–53.CrossRefPubMed
65.
Zurück zum Zitat Bloom PA, Tsai JC, Sharma K, et al. “Cyclodiode”. Trans-scleral diode laser cyclophotocoagulation in the treatment of advanced refractory glaucoma. Ophthalmology. 1997;104:1508–19 (discussion 1519–1520).CrossRefPubMed Bloom PA, Tsai JC, Sharma K, et al. “Cyclodiode”. Trans-scleral diode laser cyclophotocoagulation in the treatment of advanced refractory glaucoma. Ophthalmology. 1997;104:1508–19 (discussion 1519–1520).CrossRefPubMed
66.
Zurück zum Zitat Schlote T, Derse M, Rassmann K, et al. Efficacy and safety of contact transscleral diode laser cyclophotocoagulation for advanced glaucoma. J Glaucoma. 2001;10:294–301.CrossRefPubMed Schlote T, Derse M, Rassmann K, et al. Efficacy and safety of contact transscleral diode laser cyclophotocoagulation for advanced glaucoma. J Glaucoma. 2001;10:294–301.CrossRefPubMed
67.
Zurück zum Zitat Feldman RM, el-Harazi SM, LoRusso FJ, et al. Histopathologic findings following contact transscleral semiconductor diode laser cyclophotocoagulation in a human eye. J Glaucoma. 1997;6:139–40.PubMed Feldman RM, el-Harazi SM, LoRusso FJ, et al. Histopathologic findings following contact transscleral semiconductor diode laser cyclophotocoagulation in a human eye. J Glaucoma. 1997;6:139–40.PubMed
69.
Zurück zum Zitat Yildirim N, Yalvac IS, Sahin A, Ozer A, Bozca T. A comparative study between diode laser cyclophotocoagulation and the Ahmed glaucoma valve implant in neovascular glaucoma: a long-term follow-up. J Glaucoma. 2009;18:192–6.CrossRefPubMed Yildirim N, Yalvac IS, Sahin A, Ozer A, Bozca T. A comparative study between diode laser cyclophotocoagulation and the Ahmed glaucoma valve implant in neovascular glaucoma: a long-term follow-up. J Glaucoma. 2009;18:192–6.CrossRefPubMed
70.
Zurück zum Zitat Lin SC. Endoscopic and transscleral cyclophotocoagulation for the treatment of refractory glaucoma. J Glaucoma. 2008;17:238–47.CrossRefPubMed Lin SC. Endoscopic and transscleral cyclophotocoagulation for the treatment of refractory glaucoma. J Glaucoma. 2008;17:238–47.CrossRefPubMed
71.
Zurück zum Zitat Patel A, Thompson JT, Michels RG, Quigley HA. Endolaser treatment of the ciliary body for uncontrolled glaucoma. Ophthalmology. 1986;93:825–30.CrossRefPubMed Patel A, Thompson JT, Michels RG, Quigley HA. Endolaser treatment of the ciliary body for uncontrolled glaucoma. Ophthalmology. 1986;93:825–30.CrossRefPubMed
72.
Zurück zum Zitat Uram M. Ophthalmic laser microendoscope endophotocoagulation. Ophthalmology. 1992;99:1829–32.CrossRefPubMed Uram M. Ophthalmic laser microendoscope endophotocoagulation. Ophthalmology. 1992;99:1829–32.CrossRefPubMed
73.
Zurück zum Zitat Chen J, Cohn RA, Lin SC, Cortes AE, Alvarado JA. Endoscopic photocoagulation of the ciliary body for treatment of refractory glaucomas. Am J Ophthalmol. 1997;124:787–96.CrossRefPubMed Chen J, Cohn RA, Lin SC, Cortes AE, Alvarado JA. Endoscopic photocoagulation of the ciliary body for treatment of refractory glaucomas. Am J Ophthalmol. 1997;124:787–96.CrossRefPubMed
74.
Zurück zum Zitat Lima FE, Magacho L, Carvalho DM, Susanna R Jr, Avila MP. A prospective, comparative study between endoscopic cyclophotocoagulation and the Ahmed drainage implant in refractory glaucoma. J Glaucoma. 2004;13:233–7.CrossRefPubMed Lima FE, Magacho L, Carvalho DM, Susanna R Jr, Avila MP. A prospective, comparative study between endoscopic cyclophotocoagulation and the Ahmed drainage implant in refractory glaucoma. J Glaucoma. 2004;13:233–7.CrossRefPubMed
Metadaten
Titel
A Review of the Ahmed Glaucoma Valve Implant and Comparison with Other Surgical Operations
verfasst von
Ivano Riva
Gloria Roberti
Andreas Katsanos
Francesco Oddone
Luciano Quaranta
Publikationsdatum
11.03.2017
Verlag
Springer Healthcare
Erschienen in
Advances in Therapy / Ausgabe 4/2017
Print ISSN: 0741-238X
Elektronische ISSN: 1865-8652
DOI
https://doi.org/10.1007/s12325-017-0503-1

Weitere Artikel der Ausgabe 4/2017

Advances in Therapy 4/2017 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

„Jeder Fall von plötzlichem Tod muss obduziert werden!“

17.05.2024 Plötzlicher Herztod Nachrichten

Ein signifikanter Anteil der Fälle von plötzlichem Herztod ist genetisch bedingt. Um ihre Verwandten vor diesem Schicksal zu bewahren, sollten jüngere Personen, die plötzlich unerwartet versterben, ausnahmslos einer Autopsie unterzogen werden.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Schlechtere Vorhofflimmern-Prognose bei kleinem linken Ventrikel

17.05.2024 Vorhofflimmern Nachrichten

Nicht nur ein vergrößerter, sondern auch ein kleiner linker Ventrikel ist bei Vorhofflimmern mit einer erhöhten Komplikationsrate assoziiert. Der Zusammenhang besteht nach Daten aus China unabhängig von anderen Risikofaktoren.

Semaglutid bei Herzinsuffizienz: Wie erklärt sich die Wirksamkeit?

17.05.2024 Herzinsuffizienz Nachrichten

Bei adipösen Patienten mit Herzinsuffizienz des HFpEF-Phänotyps ist Semaglutid von symptomatischem Nutzen. Resultiert dieser Benefit allein aus der Gewichtsreduktion oder auch aus spezifischen Effekten auf die Herzinsuffizienz-Pathogenese? Eine neue Analyse gibt Aufschluss.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.