09.07.2018 | Original Paper

Agreement and repeatability of central corneal thickness measurements by four different optical devices and an ultrasound pachymeter

verfasst von: Nesrin Buyuktortop Gokcinar, Erhan Yumusak, Nurgul Ornek, Serap Yorubulut, Zafer Onaran

Erschienen in: International Ophthalmology | Ausgabe 7/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To compare the repeatability and agreement of central corneal thickness (CCT) measurements by spectral-domain optical coherence tomography (OCT), corneal topography (CT) with a combined Scheimpflug–Placido system, optical biometry (OB), specular microscopy (SM), and ultrasound pachymetry (UP).

Methods

A single observer measured CCT twice in 150 eyes of 150 subjects with each of five devices: Nidek RS-3000 Advance OCT, CSO Sirius combined Scheimpflug–Placido disc system CT, Nidek AL-Scan partial coherence interferometry-based OB, Tomey EM-3000 SM, and Reichert iPac ultrasonic pachymeter. Pachymetry values corrected by the SM device software were also recorded. Levels of agreement between devices were evaluated by Bland–Altman plots with 95% limits of agreement, and repeatability for each device was analysed with intraclass correlation coefficients.

Results

The mean CCTs measured by OCT, CT, OB, SM, corrected SM, and UP were 544.60 ± 29.56, 536.19 ± 32.14, 528.29 ± 29.45, 524.88 ± 32.38, 537.88 ± 32.38, and 545.29 ± 30.75 μm, respectively. Mean CCT differed significantly between the devices (p < 0.05) apart from between OCT and UP, and between CT and corrected SM. Mean paired differences between devices ranged between 0.68 and 20.41 μm. Repeatability with all devices was excellent (> 0.99). The range of limits of agreement was the least between OCT and UP.

Conclusions

Different CCT measurement techniques produce quite different results, so CCT evaluation and follow-up should be performed using the same device or devices with close compatibility.
Literatur
1.
Zurück zum Zitat Lackner B, Schmidinger G, Pieh S, Funovics MA, Skorpik C (2005) Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound. Optom Vis Sci 82(10):892–899CrossRefPubMed Lackner B, Schmidinger G, Pieh S, Funovics MA, Skorpik C (2005) Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound. Optom Vis Sci 82(10):892–899CrossRefPubMed
2.
Zurück zum Zitat Barkana Y, Gerber Y, Elbaz U et al (2005) Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg 31(9):1729–1735CrossRefPubMed Barkana Y, Gerber Y, Elbaz U et al (2005) Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg 31(9):1729–1735CrossRefPubMed
3.
Zurück zum Zitat Chakrabarti HS, Craig JP, Brahma A, Malik TY, McGhee CN (2001) Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes. J Cataract Refract Surg 27(11):1823–1828CrossRefPubMed Chakrabarti HS, Craig JP, Brahma A, Malik TY, McGhee CN (2001) Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes. J Cataract Refract Surg 27(11):1823–1828CrossRefPubMed
4.
Zurück zum Zitat Doughty MJ, Jonuscheit S (2010) The orbscan acoustic (correction) factor for central corneal thickness measures of normal human corneas. Eye Contact Lens 36(2):106–115CrossRefPubMed Doughty MJ, Jonuscheit S (2010) The orbscan acoustic (correction) factor for central corneal thickness measures of normal human corneas. Eye Contact Lens 36(2):106–115CrossRefPubMed
5.
Zurück zum Zitat Gonul S, Koktekir BE, Bakbak B, Gedik S (2014) Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera. Arq Bras Oftalmol 77(6):345–350PubMed Gonul S, Koktekir BE, Bakbak B, Gedik S (2014) Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera. Arq Bras Oftalmol 77(6):345–350PubMed
6.
Zurück zum Zitat Khaja WA, Grover S, Kelmenson AT, Ferguson LR, Sambhav K, Chalam KV (2015) Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and Orbscan. Clin Ophthalmol 9:1065–1070PubMedPubMedCentral Khaja WA, Grover S, Kelmenson AT, Ferguson LR, Sambhav K, Chalam KV (2015) Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and Orbscan. Clin Ophthalmol 9:1065–1070PubMedPubMedCentral
7.
Zurück zum Zitat Bechmann M, Thiel M, Roesen B et al (2000) Central corneal thickness determined with optical coherence tomography in various types of glaucoma. Br J Ophthalmol 84(11):1233–1237CrossRefPubMedPubMedCentral Bechmann M, Thiel M, Roesen B et al (2000) Central corneal thickness determined with optical coherence tomography in various types of glaucoma. Br J Ophthalmol 84(11):1233–1237CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Wirbelauer C, Scholz C, Hoerauf H et al (2002) Noncontact corneal pachymetry with slit lamp-adapted optical coherence tomography. Am J Ophthalmol 133(4):444–450CrossRefPubMed Wirbelauer C, Scholz C, Hoerauf H et al (2002) Noncontact corneal pachymetry with slit lamp-adapted optical coherence tomography. Am J Ophthalmol 133(4):444–450CrossRefPubMed
9.
Zurück zum Zitat Wong AC-M, Wong C-C, Yuen NS-Y et al (2002) Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetry. Eye 16(6):715–721CrossRefPubMed Wong AC-M, Wong C-C, Yuen NS-Y et al (2002) Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetry. Eye 16(6):715–721CrossRefPubMed
10.
Zurück zum Zitat Amano S, Honda N, Amano Y et al (2006) Comparison of central corneal thickness by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography. Ophthalmology 113(6):937–941CrossRefPubMed Amano S, Honda N, Amano Y et al (2006) Comparison of central corneal thickness by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography. Ophthalmology 113(6):937–941CrossRefPubMed
11.
Zurück zum Zitat Li H, Leung CK, Wong L et al (2008) Comparative study of central corneal thickness measurement with slit-lamp optical coherence tomography and visante optical coherence tomography. Ophthalmology 115(5):796–801CrossRefPubMed Li H, Leung CK, Wong L et al (2008) Comparative study of central corneal thickness measurement with slit-lamp optical coherence tomography and visante optical coherence tomography. Ophthalmology 115(5):796–801CrossRefPubMed
12.
Zurück zum Zitat Rao HL, Kumar AU, Kumar A et al (2011) Evaluation of central corneal thickness measurement with RTVue spectral domain optical coherence tomography in normal subjects. Cornea 30(2):121–126CrossRefPubMed Rao HL, Kumar AU, Kumar A et al (2011) Evaluation of central corneal thickness measurement with RTVue spectral domain optical coherence tomography in normal subjects. Cornea 30(2):121–126CrossRefPubMed
13.
Zurück zum Zitat Beutelspacher SC, Serbecic N, Scheuerle AF (2011) Assessment of central corneal thickness using OCT, ultrasound, optical low coherence reflectometry and Scheimpflug pachymetry. Eur J Ophthalmol 21(2):132–137CrossRefPubMed Beutelspacher SC, Serbecic N, Scheuerle AF (2011) Assessment of central corneal thickness using OCT, ultrasound, optical low coherence reflectometry and Scheimpflug pachymetry. Eur J Ophthalmol 21(2):132–137CrossRefPubMed
14.
Zurück zum Zitat Kim HY, Budenz DL, Lee PS, Feuer WJ, Barton K (2008) Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry. Am J Ophthalmol 145(2):228–232CrossRefPubMed Kim HY, Budenz DL, Lee PS, Feuer WJ, Barton K (2008) Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry. Am J Ophthalmol 145(2):228–232CrossRefPubMed
15.
Zurück zum Zitat Bayhan HA, Aslan Bayhan S, Can I (2014) Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter. Int J Ophthalmol 7(2):302–308PubMedPubMedCentral Bayhan HA, Aslan Bayhan S, Can I (2014) Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter. Int J Ophthalmol 7(2):302–308PubMedPubMedCentral
16.
Zurück zum Zitat Randleman JB, Lynn MJ, Perez-Straziota CE, Weissman HM, Kim SW (2015) Comparison of central and peripheral corneal thickness measurements with scanning-slit, Scheimpflug and Fourier-domain ocular coherence tomography. Br J Ophthalmol 99(9):1176–1181CrossRefPubMed Randleman JB, Lynn MJ, Perez-Straziota CE, Weissman HM, Kim SW (2015) Comparison of central and peripheral corneal thickness measurements with scanning-slit, Scheimpflug and Fourier-domain ocular coherence tomography. Br J Ophthalmol 99(9):1176–1181CrossRefPubMed
17.
Zurück zum Zitat Sedaghat MR, Daneshvar R, Kargozar A, Derakhshan A, Daraei M (2010) Comparison of central corneal thickness measurement using ultrasonic pachymetry, rotating Scheimpflug camera, and scanning-slit topography. Am J Ophthalmol 150(6):780–789CrossRefPubMed Sedaghat MR, Daneshvar R, Kargozar A, Derakhshan A, Daraei M (2010) Comparison of central corneal thickness measurement using ultrasonic pachymetry, rotating Scheimpflug camera, and scanning-slit topography. Am J Ophthalmol 150(6):780–789CrossRefPubMed
18.
Zurück zum Zitat Módis L Jr, Szalai E, Németh G, Berta A (2011) Reliability of the corneal thickness measurements with the Pentacam HR imaging system and ultrasound pachymetry. Cornea 30(5):561–566CrossRefPubMed Módis L Jr, Szalai E, Németh G, Berta A (2011) Reliability of the corneal thickness measurements with the Pentacam HR imaging system and ultrasound pachymetry. Cornea 30(5):561–566CrossRefPubMed
19.
Zurück zum Zitat Bao F, Wang Q, Cheng S et al (2014) Comparison and evaluation of central corneal thickness using 2 new noncontact specular microscopes and conventional pachymetry devices. Cornea 33(6):576–581CrossRefPubMed Bao F, Wang Q, Cheng S et al (2014) Comparison and evaluation of central corneal thickness using 2 new noncontact specular microscopes and conventional pachymetry devices. Cornea 33(6):576–581CrossRefPubMed
20.
Zurück zum Zitat Scotto R, Bagnis A, Papadia M, Cutolo CA, Risso D, Traverso CE (2017) Comparison of central corneal thickness measurements using ultrasonic pachymetry, anterior segment OCT and noncontact specular microscopy. J Glaucoma 26(10):860–865CrossRefPubMed Scotto R, Bagnis A, Papadia M, Cutolo CA, Risso D, Traverso CE (2017) Comparison of central corneal thickness measurements using ultrasonic pachymetry, anterior segment OCT and noncontact specular microscopy. J Glaucoma 26(10):860–865CrossRefPubMed
21.
Zurück zum Zitat Pierro L, Iuliano L, Gagliardi M, Ambrosi A, Rama P, Bandello F (2016) Central corneal thickness reproducibility among ten different instruments. Optom Vis Sci 93(11):1371–1379CrossRefPubMed Pierro L, Iuliano L, Gagliardi M, Ambrosi A, Rama P, Bandello F (2016) Central corneal thickness reproducibility among ten different instruments. Optom Vis Sci 93(11):1371–1379CrossRefPubMed
22.
Zurück zum Zitat McAlinden C, Khadka J, Pesudovs K (2015) Precision (repeatability and reproducibility) studies and sample-size calculation. J Cataract Refract Surg 41(12):2598–2604CrossRefPubMed McAlinden C, Khadka J, Pesudovs K (2015) Precision (repeatability and reproducibility) studies and sample-size calculation. J Cataract Refract Surg 41(12):2598–2604CrossRefPubMed
Metadaten
Titel
Agreement and repeatability of central corneal thickness measurements by four different optical devices and an ultrasound pachymeter
verfasst von
Nesrin Buyuktortop Gokcinar
Erhan Yumusak
Nurgul Ornek
Serap Yorubulut
Zafer Onaran
Publikationsdatum
09.07.2018
Verlag
Springer Netherlands
Erschienen in
International Ophthalmology / Ausgabe 7/2019
Print ISSN: 0165-5701
Elektronische ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-018-0983-2

Weitere Artikel der Ausgabe 7/2019

International Ophthalmology 7/2019 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.