01.08.2014 | Review Article

An overview of home tonometry and telemetry for intraocular pressure monitoring in humans

verfasst von: Edward Yung, Valerie Trubnik, L. Jay Katz

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 8/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To review the existing technology for self-tonometry and evaluate methods for continuous monitoring of IOP currently undergoing development and clinical trials in humans.

Background

Glaucoma is one of the leading causes of blindness worldwide. Current glaucoma therapy is focused primarily on lowering intraocular pressures (IOP). Decisions to modify treatment regiments are primarily based on office IOP. Since IOP fluctuates throughout the day, values obtained in the office may be a poor representation of the patient’s disease. IOP measurements outside of the physician’s office environment would provide better knowledge of the disease state and allow for better-informed medical decision making.

Methods and results

We performed a literature search using Medline and IEEE database for studies investigating technologies that have been developed for continuous 24-hour IOP monitoring.

Conclusion

There is currently no technology that has been approved for use to allow for continuous monitoring of IOP fluctuations. New experimental technology being developed and currently undergoing clinical trials has demonstrated potential for changing the diagnosis and management of glaucoma.
Literatur
1.
Zurück zum Zitat Resnikoff S, Pascolini D, Etya’ale D et al. (2004) Global data on visual impairment in the year 2002. Bull World Health Organ 82(11):844–851PubMedCentralPubMed Resnikoff S, Pascolini D, Etya’ale D et al. (2004) Global data on visual impairment in the year 2002. Bull World Health Organ 82(11):844–851PubMedCentralPubMed
3.
4.
Zurück zum Zitat Wessels IF, Oh Y (1990) Tonometer utilization, accuracy, and calibration under field conditions. Arch Ophthalmol 108(12):1709–112PubMedCrossRef Wessels IF, Oh Y (1990) Tonometer utilization, accuracy, and calibration under field conditions. Arch Ophthalmol 108(12):1709–112PubMedCrossRef
5.
Zurück zum Zitat Tonnu PA, Ho T, Newson T et al. (2005) The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry. Br J Ophthalmol 89(7):851–854PubMedCentralPubMedCrossRef Tonnu PA, Ho T, Newson T et al. (2005) The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry. Br J Ophthalmol 89(7):851–854PubMedCentralPubMedCrossRef
6.
Zurück zum Zitat Collaer N, Zeyen T, Caprioli J (2005) Sequential office pressure measurements in the management of glaucoma. J Glaucoma 14(3):196–200PubMedCrossRef Collaer N, Zeyen T, Caprioli J (2005) Sequential office pressure measurements in the management of glaucoma. J Glaucoma 14(3):196–200PubMedCrossRef
7.
Zurück zum Zitat Asrani S, Zeimer R, Wilensky J et al. (2000) Large diurnal fluctuations in intraocular pressures are an independent risk factor in patients with glaucoma. J Glaucoma 9:134–142PubMedCrossRef Asrani S, Zeimer R, Wilensky J et al. (2000) Large diurnal fluctuations in intraocular pressures are an independent risk factor in patients with glaucoma. J Glaucoma 9:134–142PubMedCrossRef
8.
Zurück zum Zitat Caprioli J, Coleman AL (2008) Intraocular Pressure Fluctuation: A risk factor for visual field progression at low intraocular pressures in the Advanced Glaucoma Intervention Study. Ophthalmology 115(7):1123–1129PubMedCrossRef Caprioli J, Coleman AL (2008) Intraocular Pressure Fluctuation: A risk factor for visual field progression at low intraocular pressures in the Advanced Glaucoma Intervention Study. Ophthalmology 115(7):1123–1129PubMedCrossRef
9.
Zurück zum Zitat McLaren JW, Brubaker RF, FitzSimon JS (1996) Continuous measurement of intraocular pressure in rabbits by telemetry. Investig Ophthalmol Vis Sci 37(6):966–975 McLaren JW, Brubaker RF, FitzSimon JS (1996) Continuous measurement of intraocular pressure in rabbits by telemetry. Investig Ophthalmol Vis Sci 37(6):966–975
10.
Zurück zum Zitat Carlson KH, McLaren JW, Topper JE et al. (1987) Effect of body position on intraocular pressure and aqueous flow. Invest Ophthalmol Vis Sci 28(8):1346–1352PubMed Carlson KH, McLaren JW, Topper JE et al. (1987) Effect of body position on intraocular pressure and aqueous flow. Invest Ophthalmol Vis Sci 28(8):1346–1352PubMed
11.
Zurück zum Zitat Yamabayashi S, Aguilar RN, Hosoda M et al. (1991) Postural change of intraocular and blood pressures in ocular hypertension and low tension glaucoma. Br J Ophthalmol 75(11):652–655PubMedCentralPubMedCrossRef Yamabayashi S, Aguilar RN, Hosoda M et al. (1991) Postural change of intraocular and blood pressures in ocular hypertension and low tension glaucoma. Br J Ophthalmol 75(11):652–655PubMedCentralPubMedCrossRef
12.
Zurück zum Zitat Kothe AC (1994) The effect of posture on intraocular pressure and pulsatile ocular blood flow in normal and glaucomatous eyes. Surv Ophthalmol 38:S191–S197PubMedCrossRef Kothe AC (1994) The effect of posture on intraocular pressure and pulsatile ocular blood flow in normal and glaucomatous eyes. Surv Ophthalmol 38:S191–S197PubMedCrossRef
13.
Zurück zum Zitat Smedt SD, Mermound A, Schnyder C (2012) 24-Hour intraocular pressure fluctuation monitoring using an ocular telemetry sensor: Tolerability and functionality in healthy subjects. J Glaucoma 21(8):539–544PubMedCrossRef Smedt SD, Mermound A, Schnyder C (2012) 24-Hour intraocular pressure fluctuation monitoring using an ocular telemetry sensor: Tolerability and functionality in healthy subjects. J Glaucoma 21(8):539–544PubMedCrossRef
14.
Zurück zum Zitat Hara T, Hara T, Tsuru T (2006) Increase of peak intraocular pressure during sleep in reproduced diurnal changes by posture. Arch Ophthalmol 124(2):165–168PubMedCrossRef Hara T, Hara T, Tsuru T (2006) Increase of peak intraocular pressure during sleep in reproduced diurnal changes by posture. Arch Ophthalmol 124(2):165–168PubMedCrossRef
15.
Zurück zum Zitat Liu JKH, Kripke DF, Hoffman RE et al. (1998) Nocturnal elevation of intraocular pressure in young adults. Invest Ophthalmol Vis Sci 39:2707–2712PubMed Liu JKH, Kripke DF, Hoffman RE et al. (1998) Nocturnal elevation of intraocular pressure in young adults. Invest Ophthalmol Vis Sci 39:2707–2712PubMed
16.
Zurück zum Zitat Liu JKH, Bouligny RP, Kripke DF et al. (2003) Nocturnal elevation of intraocular pressure is detectable in the sitting position. Invest Ophthalmol Vis Sci 44(10):4439–4442PubMedCrossRef Liu JKH, Bouligny RP, Kripke DF et al. (2003) Nocturnal elevation of intraocular pressure is detectable in the sitting position. Invest Ophthalmol Vis Sci 44(10):4439–4442PubMedCrossRef
17.
Zurück zum Zitat Brown B, Morris P, Muller C et al. (1988) Fluctuations in intra-ocular pressure with sleep: I. Time course of IOP increase after the onset of sleep. Ophthalmic Physiol Opt 8(3):246–248PubMed Brown B, Morris P, Muller C et al. (1988) Fluctuations in intra-ocular pressure with sleep: I. Time course of IOP increase after the onset of sleep. Ophthalmic Physiol Opt 8(3):246–248PubMed
18.
Zurück zum Zitat Brown B, Burton P, Mann S et al. (1988) Fluctuations in intra-ocular pressure with sleep: II. Time course of IOP decrease after waking from sleep. Ophthalmic Physiol Opt 8(3):249–252PubMed Brown B, Burton P, Mann S et al. (1988) Fluctuations in intra-ocular pressure with sleep: II. Time course of IOP decrease after waking from sleep. Ophthalmic Physiol Opt 8(3):249–252PubMed
19.
Zurück zum Zitat Orzalesi N, Rossetti L, Invernizzi T et al. (2000) Effect of timolol, latanoprost, and dorzolamide on circadian IOP in glaucoma or ocular hypertension. Invest Ophthal VisSci 41(9):2566–2573 Orzalesi N, Rossetti L, Invernizzi T et al. (2000) Effect of timolol, latanoprost, and dorzolamide on circadian IOP in glaucoma or ocular hypertension. Invest Ophthal VisSci 41(9):2566–2573
20.
Zurück zum Zitat Orzalesi N, Rossetti L, Bottoli A et al. (2003) The effect of latanoprost, brimonidine, and a fixed combination of timolol and dorzolamide on circadian intraocular pressure in patients with glaucoma or ocular hypertension. Arch Ophthalmol 121(4):453–457PubMedCrossRef Orzalesi N, Rossetti L, Bottoli A et al. (2003) The effect of latanoprost, brimonidine, and a fixed combination of timolol and dorzolamide on circadian intraocular pressure in patients with glaucoma or ocular hypertension. Arch Ophthalmol 121(4):453–457PubMedCrossRef
21.
Zurück zum Zitat De Alva M, Jervell J (1997) Self monitoring improves quality of life and prognosis of people with diabetes. Br Med J 315:184–185CrossRef De Alva M, Jervell J (1997) Self monitoring improves quality of life and prognosis of people with diabetes. Br Med J 315:184–185CrossRef
22.
Zurück zum Zitat Hughes E, Spry P, Diamond J (2003) 24-Hour monitoring of intraocular pressure in glaucoma management: a retrospective review. J Glaucoma 12:232–236PubMedCrossRef Hughes E, Spry P, Diamond J (2003) 24-Hour monitoring of intraocular pressure in glaucoma management: a retrospective review. J Glaucoma 12:232–236PubMedCrossRef
23.
Zurück zum Zitat Draeger J, Schwartz R, Groenhoff et al. (1993) Self-tonometry under microgravity conditions. Clin Investigator 71:700–703CrossRef Draeger J, Schwartz R, Groenhoff et al. (1993) Self-tonometry under microgravity conditions. Clin Investigator 71:700–703CrossRef
24.
Zurück zum Zitat Theofylaktopoulos I, Diestelhorst M, Krieglstein GK (1999) Self-tonometry with the Ocuton S versus Goldmann tonometry. Graefes Arch Clin Exp Ophthalmol 237:720–724PubMedCrossRef Theofylaktopoulos I, Diestelhorst M, Krieglstein GK (1999) Self-tonometry with the Ocuton S versus Goldmann tonometry. Graefes Arch Clin Exp Ophthalmol 237:720–724PubMedCrossRef
25.
Zurück zum Zitat Sacu S, Vass C, Schemper M et al. (2004) Self-tonometry with the Ocuton S: evaluation of accuracy in glaucoma patients. Acta Ophthalmol Scand 82:405–409PubMedCrossRef Sacu S, Vass C, Schemper M et al. (2004) Self-tonometry with the Ocuton S: evaluation of accuracy in glaucoma patients. Acta Ophthalmol Scand 82:405–409PubMedCrossRef
26.
Zurück zum Zitat Marchini G, Babighian S, Specchia L et al. (2002) Evaluation of the new Ocuton S tonometer. Acta Ophthalmol Scand 80:167–171PubMedCrossRef Marchini G, Babighian S, Specchia L et al. (2002) Evaluation of the new Ocuton S tonometer. Acta Ophthalmol Scand 80:167–171PubMedCrossRef
27.
Zurück zum Zitat Kothy P, Vargha P, Hollo G (2001) Ocuton S self-tonometry vs Goldmann tonometry; a diurnal comparison study. Acta Ophthalmologica Scandanavica 79:294–297CrossRef Kothy P, Vargha P, Hollo G (2001) Ocuton S self-tonometry vs Goldmann tonometry; a diurnal comparison study. Acta Ophthalmologica Scandanavica 79:294–297CrossRef
28.
Zurück zum Zitat Lanfermann E, Jurgens C, Grossjohann R et al. (2009) Intraocular pressure measurements with the newly Ocuton S*TT-MV self-tonometer in comparison to Goldmann applanation tonometry in glaucoma patients. Med Sci Monit 15(11):CR556–CR562PubMed Lanfermann E, Jurgens C, Grossjohann R et al. (2009) Intraocular pressure measurements with the newly Ocuton S*TT-MV self-tonometer in comparison to Goldmann applanation tonometry in glaucoma patients. Med Sci Monit 15(11):CR556–CR562PubMed
29.
Zurück zum Zitat Rosenwasser GO, Holland S, Pflugfelder SC et al. (1990) Topical anesthetic abuse. Ophthalmology 97(8):967–972PubMedCrossRef Rosenwasser GO, Holland S, Pflugfelder SC et al. (1990) Topical anesthetic abuse. Ophthalmology 97(8):967–972PubMedCrossRef
30.
Zurück zum Zitat McGee HT, Fraunfelder FW (2007) Toxicities of topical ophthalmic anesthetics. Expert Opin Drug Saf 6(6):637–640PubMedCrossRef McGee HT, Fraunfelder FW (2007) Toxicities of topical ophthalmic anesthetics. Expert Opin Drug Saf 6(6):637–640PubMedCrossRef
31.
Zurück zum Zitat Yeniad B, Canturk S, Esin Ozdemir F et al. (2010) Toxic keratopathy due to abuse of topical anesthetic drugs. Cutan Ocul Toxicol 29(2):105–109PubMedCrossRef Yeniad B, Canturk S, Esin Ozdemir F et al. (2010) Toxic keratopathy due to abuse of topical anesthetic drugs. Cutan Ocul Toxicol 29(2):105–109PubMedCrossRef
32.
Zurück zum Zitat Lam DSC, Leung DYL, Chiu TYH et al. (2004) Pressure phosphene self-tonometry: A comparison with Goldmann Tonometry in Glaucoma Patients. Investig Ophthalmol Vis Sci 45(9):3131–3136CrossRef Lam DSC, Leung DYL, Chiu TYH et al. (2004) Pressure phosphene self-tonometry: A comparison with Goldmann Tonometry in Glaucoma Patients. Investig Ophthalmol Vis Sci 45(9):3131–3136CrossRef
33.
Zurück zum Zitat Tai MC, Chen PL, Wu JN et al. (2005) Clinical evaluation of the intraocular pressure in patients with glaucoma or ocular hypertension by a self-assessable tonometer. J Ocul Pharmacol Ther 21(1):55–61PubMedCrossRef Tai MC, Chen PL, Wu JN et al. (2005) Clinical evaluation of the intraocular pressure in patients with glaucoma or ocular hypertension by a self-assessable tonometer. J Ocul Pharmacol Ther 21(1):55–61PubMedCrossRef
34.
Zurück zum Zitat Danesh-Meyer HV, Niederer R, Gaskin BJ, Gamble (2004) Comparison of the Proview pressure phosphene tonometer performed by the patient and examiner with the Goldmann applanation tonometer. Clin Exp Ophthalmol 32(1):29–32CrossRef Danesh-Meyer HV, Niederer R, Gaskin BJ, Gamble (2004) Comparison of the Proview pressure phosphene tonometer performed by the patient and examiner with the Goldmann applanation tonometer. Clin Exp Ophthalmol 32(1):29–32CrossRef
35.
Zurück zum Zitat Rai S, Moster MR, Kesen M et al. (2005) Level of disagreement between Proview Phosphene Tonometer and Goldmann Applanation Tonometer Intraocular Pressure Readings. J Glaucoma 14:120–123PubMedCrossRef Rai S, Moster MR, Kesen M et al. (2005) Level of disagreement between Proview Phosphene Tonometer and Goldmann Applanation Tonometer Intraocular Pressure Readings. J Glaucoma 14:120–123PubMedCrossRef
36.
Zurück zum Zitat Li J, Herndon LW, Asrani SG et al. (2004) Clinical comparison of the Proview Eye Pressure Monitor with Goldmann Applanation Tonometer and the TonoPen. Arch Ophthalmol 122:1117–1121PubMedCrossRef Li J, Herndon LW, Asrani SG et al. (2004) Clinical comparison of the Proview Eye Pressure Monitor with Goldmann Applanation Tonometer and the TonoPen. Arch Ophthalmol 122:1117–1121PubMedCrossRef
37.
Zurück zum Zitat Rosentreter A, Jablonski KS, Mellein AC et al. (2011) A new rebound tonometer for home monitoring of intraocular pressure. Graefes Arch Clin Exp Ophthalmol 249(11):1713–1719PubMedCrossRef Rosentreter A, Jablonski KS, Mellein AC et al. (2011) A new rebound tonometer for home monitoring of intraocular pressure. Graefes Arch Clin Exp Ophthalmol 249(11):1713–1719PubMedCrossRef
38.
Zurück zum Zitat Asrani S, Chatterjee A, Wallace DK et al. (2011) Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device. J Glaucoma 20:74–79PubMedCrossRef Asrani S, Chatterjee A, Wallace DK et al. (2011) Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device. J Glaucoma 20:74–79PubMedCrossRef
39.
Zurück zum Zitat Muttuvelu DV, Baggesen K, Ehlers N (2012) Precision and accuracy of the ICare tonometer – Peripheral and central IOP measurements by rebound tonometry. Acta Ophthalmol (Copenh) 90(4):322–326CrossRef Muttuvelu DV, Baggesen K, Ehlers N (2012) Precision and accuracy of the ICare tonometer – Peripheral and central IOP measurements by rebound tonometry. Acta Ophthalmol (Copenh) 90(4):322–326CrossRef
40.
Zurück zum Zitat Chiu W, Lam A, Chen D et al. (2008) The influence of corneal properties on rebound tonometry. Ophthalmology 115:80–84PubMedCrossRef Chiu W, Lam A, Chen D et al. (2008) The influence of corneal properties on rebound tonometry. Ophthalmology 115:80–84PubMedCrossRef
41.
Zurück zum Zitat Kupin TH, Shin DH, Juzych MS et al. (1993) Use of a Tono-Pen for long-term home tonometry. Am J Ophthalmol 116(5):643–644PubMed Kupin TH, Shin DH, Juzych MS et al. (1993) Use of a Tono-Pen for long-term home tonometry. Am J Ophthalmol 116(5):643–644PubMed
42.
Zurück zum Zitat Ruokonen PC, Schwenteck T, Draeger J (2007) Evaluation of the impedance tonometer TGDc-01 and iCare according to international ocular tonometer standards ISO 8612. Graefes Arch Clin Exp Ophthalmol 245:1259–1265PubMedCrossRef Ruokonen PC, Schwenteck T, Draeger J (2007) Evaluation of the impedance tonometer TGDc-01 and iCare according to international ocular tonometer standards ISO 8612. Graefes Arch Clin Exp Ophthalmol 245:1259–1265PubMedCrossRef
43.
Zurück zum Zitat Boles Carenini B, Brogliatti B, Tonetto C et al. (1992) The Pulsair-Keeler non-contact tonometer in self-tonometry: preliminary results. Int Ophthalmol 16:295–297PubMedCrossRef Boles Carenini B, Brogliatti B, Tonetto C et al. (1992) The Pulsair-Keeler non-contact tonometer in self-tonometry: preliminary results. Int Ophthalmol 16:295–297PubMedCrossRef
44.
Zurück zum Zitat Hjortdal J, Jensen PK (1995) In vitro measurement of corneal strain, thickness, and curvature using digital image processing. Acta Ophthalmol Scand 73:5–11PubMedCrossRef Hjortdal J, Jensen PK (1995) In vitro measurement of corneal strain, thickness, and curvature using digital image processing. Acta Ophthalmol Scand 73:5–11PubMedCrossRef
45.
Zurück zum Zitat Van der Jagt LH, Jansonius NM (2005) Three portable tonometers, the TGDc-01, the ICARE, and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt 25:429–435PubMedCrossRef Van der Jagt LH, Jansonius NM (2005) Three portable tonometers, the TGDc-01, the ICARE, and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt 25:429–435PubMedCrossRef
46.
Zurück zum Zitat Resua CG, Fernandez MJG, Exposito AC et al. (2005) Clinical evaluation of the new TGDc-01 “PRA” palpebral tonometer: comparison with contact and non-contact tonometry. Optom Vis Sci 82(2):143–150CrossRef Resua CG, Fernandez MJG, Exposito AC et al. (2005) Clinical evaluation of the new TGDc-01 “PRA” palpebral tonometer: comparison with contact and non-contact tonometry. Optom Vis Sci 82(2):143–150CrossRef
47.
Zurück zum Zitat Cook JA, Botello AP, Elders A et al. (2012) Systematic review of the agreement of tonometers with Goldmann Applanation Tonometry. Ophthalmology 119:1552–1557PubMedCrossRef Cook JA, Botello AP, Elders A et al. (2012) Systematic review of the agreement of tonometers with Goldmann Applanation Tonometry. Ophthalmology 119:1552–1557PubMedCrossRef
48.
Zurück zum Zitat Lam AKC, Douthwaite WA (1997) The effect of an artificially elevated intraocular pressure on the central corneal curvature. Ophthalmic Physiol Opt 17:18–24PubMedCrossRef Lam AKC, Douthwaite WA (1997) The effect of an artificially elevated intraocular pressure on the central corneal curvature. Ophthalmic Physiol Opt 17:18–24PubMedCrossRef
49.
Zurück zum Zitat Leonardi M, Leuenberger P, Bertrand D et al. (2004) First steps towards noninvasive intraocular pressure monitoring with a sensing contact lens. Invest Ophthalmol Vis Sci 45(9):3113–3117PubMedCrossRef Leonardi M, Leuenberger P, Bertrand D et al. (2004) First steps towards noninvasive intraocular pressure monitoring with a sensing contact lens. Invest Ophthalmol Vis Sci 45(9):3113–3117PubMedCrossRef
50.
Zurück zum Zitat Leonardi M, Pitchon EM, Bertsch A et al. (2009) Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes. Acta Ophthalmol (Copenh) 87(4):433–437CrossRef Leonardi M, Pitchon EM, Bertsch A et al. (2009) Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes. Acta Ophthalmol (Copenh) 87(4):433–437CrossRef
51.
Zurück zum Zitat Faschinger C, Mossbock G (2010) Kontinuierliche 24-h-Aufzeichnung von Augendruckschwankungen mittels drahtlosem Kontaktlensensor Triggerfish. Ophthalmologe 107(10):918–922PubMedCrossRef Faschinger C, Mossbock G (2010) Kontinuierliche 24-h-Aufzeichnung von Augendruckschwankungen mittels drahtlosem Kontaktlensensor Triggerfish. Ophthalmologe 107(10):918–922PubMedCrossRef
52.
Zurück zum Zitat Mansouri K, Shaarawy T (2011) Continuous intraocular pressure monitoring with a wireless ocular telemetry sensor: initial clinical experience in patients with open angle glaucoma. Br J Ophthalmol 95:627–629PubMedCrossRef Mansouri K, Shaarawy T (2011) Continuous intraocular pressure monitoring with a wireless ocular telemetry sensor: initial clinical experience in patients with open angle glaucoma. Br J Ophthalmol 95:627–629PubMedCrossRef
53.
Zurück zum Zitat Mansouri K, Medeiros FA, Tafresh et al. (2012) Continuous 24-hour monitoring of intraocular pressure patterns with a contact lens sensor: Safety, tolerability, and reproducibility in patients with glaucoma. Arch Ophthalmol 130(12):1534–1539CrossRef Mansouri K, Medeiros FA, Tafresh et al. (2012) Continuous 24-hour monitoring of intraocular pressure patterns with a contact lens sensor: Safety, tolerability, and reproducibility in patients with glaucoma. Arch Ophthalmol 130(12):1534–1539CrossRef
54.
Zurück zum Zitat Lorenz K, Korb C, Herzog N et al. (2013) Tolerability of 24-hour intraocular pressure monitoring of a pressure-sensitive contact lens. J Glaucoma 22(4):311–316PubMedCrossRef Lorenz K, Korb C, Herzog N et al. (2013) Tolerability of 24-hour intraocular pressure monitoring of a pressure-sensitive contact lens. J Glaucoma 22(4):311–316PubMedCrossRef
55.
Zurück zum Zitat Matsumoto T, Nagata TR, Slashin M et al. (1978) Measurement by holographic interferometry of the deformation of the eye accompanying changes in intraocular pressure. Appl Opt 17(22):3538–3539PubMedCrossRef Matsumoto T, Nagata TR, Slashin M et al. (1978) Measurement by holographic interferometry of the deformation of the eye accompanying changes in intraocular pressure. Appl Opt 17(22):3538–3539PubMedCrossRef
56.
Zurück zum Zitat Sanchez I, Laukhin V, Moya A et al. (2011) Prototype of a nanostructured sensing contact lens for noninvasive intraocular pressure monitoring. Investig Ophthalmol Vis Sci 52(11):8310–8315CrossRef Sanchez I, Laukhin V, Moya A et al. (2011) Prototype of a nanostructured sensing contact lens for noninvasive intraocular pressure monitoring. Investig Ophthalmol Vis Sci 52(11):8310–8315CrossRef
57.
Zurück zum Zitat Collins CC (1967) Miniature passive pressure transensor for implanting in the eye. IEEE Trans Biomed Eng 14:74–83PubMedCrossRef Collins CC (1967) Miniature passive pressure transensor for implanting in the eye. IEEE Trans Biomed Eng 14:74–83PubMedCrossRef
58.
Zurück zum Zitat Collins CC (1967) Passive telemetry with glass transensors. Proc Nat Telemetring Conf 146–151 Collins CC (1967) Passive telemetry with glass transensors. Proc Nat Telemetring Conf 146–151
59.
Zurück zum Zitat Wolbarsht ML, Wortman J, Schwartz B et al. (1980) A sclera buckle pressure gauge for continuous monitoring of intraocular pressure. Int Ophthalmol 3:11–17PubMedCrossRef Wolbarsht ML, Wortman J, Schwartz B et al. (1980) A sclera buckle pressure gauge for continuous monitoring of intraocular pressure. Int Ophthalmol 3:11–17PubMedCrossRef
60.
Zurück zum Zitat Rizq RN, Choi WH, Eilers D et al. (2001) Intraocular pressure measurement at choroid surface: a feasibility study with implications for implantable microsystems. Br J Ophthalmol 85:868–871PubMedCentralPubMedCrossRef Rizq RN, Choi WH, Eilers D et al. (2001) Intraocular pressure measurement at choroid surface: a feasibility study with implications for implantable microsystems. Br J Ophthalmol 85:868–871PubMedCentralPubMedCrossRef
61.
Zurück zum Zitat Svedbergh B, Backlund Y, Hok B, Rosengren L (1992) The IOP-IOL. A prob into the eye. Acta Ophthalmol (Copenhagen) 70(2):266–268CrossRef Svedbergh B, Backlund Y, Hok B, Rosengren L (1992) The IOP-IOL. A prob into the eye. Acta Ophthalmol (Copenhagen) 70(2):266–268CrossRef
62.
Zurück zum Zitat Walter P, Schnakenberg U, vom Bogel G et al. (2000) Development of a completely encapsulated intraocular pressure sensor. Ophthalmic Res 32:278–284PubMedCrossRef Walter P, Schnakenberg U, vom Bogel G et al. (2000) Development of a completely encapsulated intraocular pressure sensor. Ophthalmic Res 32:278–284PubMedCrossRef
63.
Zurück zum Zitat Hille K, Draeger J, Eggers T et al. (2001) Technischer aufbau, kalibrierung und ergebnisse mit einem neuen intraokularen drucksensor mit telemetrischer Ubertragung. Klin Monatsbl Augenheilkd 218(5):376–380PubMedCrossRef Hille K, Draeger J, Eggers T et al. (2001) Technischer aufbau, kalibrierung und ergebnisse mit einem neuen intraokularen drucksensor mit telemetrischer Ubertragung. Klin Monatsbl Augenheilkd 218(5):376–380PubMedCrossRef
64.
Zurück zum Zitat Chen PJ, Rodger DC, Humayun MS (2006) Unpowered spiral-tube parylene pressure sensor for intraocular pressure sensing. Sensors and Actuators A:Physical 127(2):276–282CrossRef Chen PJ, Rodger DC, Humayun MS (2006) Unpowered spiral-tube parylene pressure sensor for intraocular pressure sensing. Sensors and Actuators A:Physical 127(2):276–282CrossRef
65.
Zurück zum Zitat Chen PJ, Rodger DC, Meng E, et al. (2006) Implantable unpowered parylene MEMS intraocular pressure sensor. Technical Digest, IEEE Engineering in Medicine and Biology Society Special Topic Conference on Microtechnologies in Medicine and Biology. 256-259 Chen PJ, Rodger DC, Meng E, et al. (2006) Implantable unpowered parylene MEMS intraocular pressure sensor. Technical Digest, IEEE Engineering in Medicine and Biology Society Special Topic Conference on Microtechnologies in Medicine and Biology. 256-259
66.
Zurück zum Zitat Data provided courtesy of Implandata Ophthalmic Products GmbH Data provided courtesy of Implandata Ophthalmic Products GmbH
67.
Zurück zum Zitat Todani A, Behlau I, Fava MA et al. (2011) Intraocular pressure measurement by radio wave telemetry. Invest Ophthalmol Vis Sci 52(13):9573–9580PubMedCrossRef Todani A, Behlau I, Fava MA et al. (2011) Intraocular pressure measurement by radio wave telemetry. Invest Ophthalmol Vis Sci 52(13):9573–9580PubMedCrossRef
68.
Zurück zum Zitat Chen G, Ghead H, Haque R, et al. (2011) A cubic-millimeter energy-autonomous wireless intraocular pressure monitor. ISSCC Digest of Technical Papers, pp 310-312 Chen G, Ghead H, Haque R, et al. (2011) A cubic-millimeter energy-autonomous wireless intraocular pressure monitor. ISSCC Digest of Technical Papers, pp 310-312
69.
Zurück zum Zitat Lee Y, Kim G, Bang S, et al. (2012) A modular 1 mm3 die-stacked sensing platform with optical communication and multi-modal energy harvesting. ISSCC Digest of Technical Papers, pp 402-404 Lee Y, Kim G, Bang S, et al. (2012) A modular 1 mm3 die-stacked sensing platform with optical communication and multi-modal energy harvesting. ISSCC Digest of Technical Papers, pp 402-404
Metadaten
Titel
An overview of home tonometry and telemetry for intraocular pressure monitoring in humans
verfasst von
Edward Yung
Valerie Trubnik
L. Jay Katz
Publikationsdatum
01.08.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 8/2014
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-014-2668-5

Weitere Artikel der Ausgabe 8/2014

Graefe's Archive for Clinical and Experimental Ophthalmology 8/2014 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.